Volumn 433, Issue , 2014, Pages 28-33

Evaluation of accuracy of FAD-GDH- and mutant Q-GDH-based blood glucose monitors in multi-patient populations

Author keywords

Arterial blood; Blood glucose monitoring system; Glucose dehydrogenase; Hemodialysis; ISO 15197:2013; Neonate

Indexed keywords

FLAVIN ADENINE DINUCLEOTIDE GLUCOSE DEHYDROGENASE; GLUCOSE; GLUCOSE DEHYDROGENASE; QUINOPROTEIN GLUCOSE DEHYDROGENASE MUTANT VARIANT; UNCLASSIFIED DRUG; FLAVINE ADENINE NUCLEOTIDE; GLUCOSE BLOOD LEVEL; GLUCOSE DEHYDROGENASE (PYRROLOQUINOLINE-QUINONE);

EID: 84896124490     PISSN: 00098981     EISSN: 18733492     Source Type: Journal    
DOI: 10.1016/j.cca.2014.02.023     Document Type: Article
Times cited : (12)

References (37)
  • 1
    • 10244223555 scopus 로고
    • Electrode systems for continuous monitoring in cardiovascular surgery
    • Clark L.C., Lyons C. Electrode systems for continuous monitoring in cardiovascular surgery. Ann N Y Acad Sci 1962, 102:29-45.
    • (1962) Ann N Y Acad Sci , vol.102 , pp. 29-45
    • Clark, L.C.1    Lyons, C.2
  • 2
    • 84857405775 scopus 로고    scopus 로고
    • Review of glucose oxidases and glucose dehydrogenases: a bird's eye view of glucose sensing enzymes
    • Ferri S., Kojima K., Sode K. Review of glucose oxidases and glucose dehydrogenases: a bird's eye view of glucose sensing enzymes. J Diabetes Sci Technol 2011, 5:1068-1076.
    • (2011) J Diabetes Sci Technol , vol.5 , pp. 1068-1076
    • Ferri, S.1    Kojima, K.2    Sode, K.3
  • 4
    • 0034844040 scopus 로고    scopus 로고
    • Glucose biosensors: 40years of advances and challenges
    • Wang J. Glucose biosensors: 40years of advances and challenges. Electroanalysis 2001, 13:983-988.
    • (2001) Electroanalysis , vol.13 , pp. 983-988
    • Wang, J.1
  • 5
    • 0035023293 scopus 로고    scopus 로고
    • Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing
    • Tang Z., Louie R.F., Lee J.H., Lee D.M., Miller E.E., Kost G.J. Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med 2001, 29:1062-1070.
    • (2001) Crit Care Med , vol.29 , pp. 1062-1070
    • Tang, Z.1    Louie, R.F.2    Lee, J.H.3    Lee, D.M.4    Miller, E.E.5    Kost, G.J.6
  • 6
    • 43549095197 scopus 로고    scopus 로고
    • Glucose oxidase: natural occurrence, function, properties and industrial applications
    • Wong C.M., Wong K.H., Chen X.D. Glucose oxidase: natural occurrence, function, properties and industrial applications. Appl Microbiol Biotechnol 2008, 78:927-938.
    • (2008) Appl Microbiol Biotechnol , vol.78 , pp. 927-938
    • Wong, C.M.1    Wong, K.H.2    Chen, X.D.3
  • 7
    • 84858706476 scopus 로고    scopus 로고
    • Characterization of different FAD-dependent glucose dehydrogenases for possible use in glucose-based biosensors and biofuel cells
    • Zafar M.N., Beden N., Leech D., Sygmund C., Ludwig R., Gorton L. Characterization of different FAD-dependent glucose dehydrogenases for possible use in glucose-based biosensors and biofuel cells. Anal Bioanal Chem 2012, 402:2069-2077.
    • (2012) Anal Bioanal Chem , vol.402 , pp. 2069-2077
    • Zafar, M.N.1    Beden, N.2    Leech, D.3    Sygmund, C.4    Ludwig, R.5    Gorton, L.6
  • 9
    • 52449111730 scopus 로고    scopus 로고
    • A disposable electrochemical glucose sensor using catalytic subunit of novel thermostable glucose dehydrogenase
    • Yamaoka H., Sode K. A disposable electrochemical glucose sensor using catalytic subunit of novel thermostable glucose dehydrogenase. Open Biotechnol J 2007, 1:26-30.
    • (2007) Open Biotechnol J , vol.1 , pp. 26-30
    • Yamaoka, H.1    Sode, K.2
  • 10
    • 70349495829 scopus 로고    scopus 로고
    • Analytical interferences in point-of-care testing glucometers by icodextrin and its metabolites: an overview. Peritoneal dialysis international
    • Flore K.M., Delanghe J.R. Analytical interferences in point-of-care testing glucometers by icodextrin and its metabolites: an overview. Peritoneal dialysis international. J Int Soc Perit Dial 2009, 29:377-383.
    • (2009) J Int Soc Perit Dial , vol.29 , pp. 377-383
    • Flore, K.M.1    Delanghe, J.R.2
  • 11
    • 77953072186 scopus 로고    scopus 로고
    • Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars
    • Frias J.P., Lim C.G., Ellison J.M., Montandon C.M. Review of adverse events associated with false glucose readings measured by GDH-PQQ-based glucose test strips in the presence of interfering sugars. Diabetes Care 2010, 33:728-729.
    • (2010) Diabetes Care , vol.33 , pp. 728-729
    • Frias, J.P.1    Lim, C.G.2    Ellison, J.M.3    Montandon, C.M.4
  • 12
    • 84896128586 scopus 로고    scopus 로고
    • FDA Public Health Notification: potentially fatal errors with GDH-PQQ glucose monitoring technology
    • FDA
    • FDA FDA Public Health Notification: potentially fatal errors with GDH-PQQ glucose monitoring technology. US Food Drug Adm 2009, http://www.fda.gov/medicaldevices/safety/alertsandnotices/publichealthnotifications/ucm176992.htm.
    • (2009) US Food Drug Adm
  • 13
    • 2042548037 scopus 로고    scopus 로고
    • Engineering PQQ glucose dehydrogenase with improved substrate specificity. Site-directed mutagenesis studies on the active center of PQQ glucose dehydrogenase
    • Igarashi S., Hirokawa T., Sode K. Engineering PQQ glucose dehydrogenase with improved substrate specificity. Site-directed mutagenesis studies on the active center of PQQ glucose dehydrogenase. Biomol Eng 2004, 21:81-89.
    • (2004) Biomol Eng , vol.21 , pp. 81-89
    • Igarashi, S.1    Hirokawa, T.2    Sode, K.3
  • 14
    • 0033517714 scopus 로고    scopus 로고
    • Construction and characterization of mutant water-soluble PQQ glucose dehydrogenases with altered K(m) values-site-directed mutagenesis studies on the putative active site
    • Igarashi S., Ohtera T., Yoshida H., Witarto A.B., Sode K. Construction and characterization of mutant water-soluble PQQ glucose dehydrogenases with altered K(m) values-site-directed mutagenesis studies on the putative active site. Biochem Biophys Res Commun 1999, 264:820-824.
    • (1999) Biochem Biophys Res Commun , vol.264 , pp. 820-824
    • Igarashi, S.1    Ohtera, T.2    Yoshida, H.3    Witarto, A.B.4    Sode, K.5
  • 15
    • 33645227125 scopus 로고    scopus 로고
    • Novel FAD-dependent glucose dehydrogenase for a dioxygen-insensitive glucose biosensor
    • Tsujimura S., Kojima S., Kano K., et al. Novel FAD-dependent glucose dehydrogenase for a dioxygen-insensitive glucose biosensor. Biosci Biotechnol Biochem 2006, 70:654-659.
    • (2006) Biosci Biotechnol Biochem , vol.70 , pp. 654-659
    • Tsujimura, S.1    Kojima, S.2    Kano, K.3
  • 16
    • 77249122160 scopus 로고    scopus 로고
    • System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197
    • Freckmann G., Baumstark A., Jendrike N., et al. System accuracy evaluation of 27 blood glucose monitoring systems according to DIN EN ISO 15197. Diabetes Technol Ther 2010, 12:221-231.
    • (2010) Diabetes Technol Ther , vol.12 , pp. 221-231
    • Freckmann, G.1    Baumstark, A.2    Jendrike, N.3
  • 17
    • 84875805117 scopus 로고    scopus 로고
    • System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197
    • Freckmann G., Schmid C., Baumstark A., Pleus S., Link M., Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol 2012, 6:1060-1075.
    • (2012) J Diabetes Sci Technol , vol.6 , pp. 1060-1075
    • Freckmann, G.1    Schmid, C.2    Baumstark, A.3    Pleus, S.4    Link, M.5    Haug, C.6
  • 18
    • 84875803673 scopus 로고    scopus 로고
    • Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197
    • Baumstark A., Pleus S., Schmid C., Link M., Haug C., Freckmann G. Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J Diabetes Sci Technol 2012, 6:1076-1086.
    • (2012) J Diabetes Sci Technol , vol.6 , pp. 1076-1086
    • Baumstark, A.1    Pleus, S.2    Schmid, C.3    Link, M.4    Haug, C.5    Freckmann, G.6
  • 19
    • 60649086908 scopus 로고    scopus 로고
    • Assessing the quality of Bionime self-monitoring blood glucose system Rightest GM110: a critical evaluation of interference and ambient circumstances
    • Hsu C.T., Hsiao H.C., Lee M.S., et al. Assessing the quality of Bionime self-monitoring blood glucose system Rightest GM110: a critical evaluation of interference and ambient circumstances. Clin Chim Acta Int J Clin Chem 2009, 402:119-123.
    • (2009) Clin Chim Acta Int J Clin Chem , vol.402 , pp. 119-123
    • Hsu, C.T.1    Hsiao, H.C.2    Lee, M.S.3
  • 20
    • 79955557642 scopus 로고    scopus 로고
    • Accuracy and precision evaluation of seven self-monitoring blood glucose systems
    • Kuo C.Y., Hsu C.T., Ho C.S., Su T.E., Wu M.H., Wang C.J. Accuracy and precision evaluation of seven self-monitoring blood glucose systems. Diabetes Technol Ther 2011, 13:596-600.
    • (2011) Diabetes Technol Ther , vol.13 , pp. 596-600
    • Kuo, C.Y.1    Hsu, C.T.2    Ho, C.S.3    Su, T.E.4    Wu, M.H.5    Wang, C.J.6
  • 22
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland J.M., Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986, 1:307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 23
    • 84896132269 scopus 로고    scopus 로고
    • Points to consider for portable blood glucose monitoring devices intended for bedside use in the neonate nursery
    • FDA
    • FDA Points to consider for portable blood glucose monitoring devices intended for bedside use in the neonate nursery. US Food Drug Adm 1996, http://www.fda.gov/downloads/MedicalDevices/DeviceRegulationandGuidance/GuidanceDocuments/UCM094125.pdf.
    • (1996) US Food Drug Adm
  • 24
    • 34447137321 scopus 로고    scopus 로고
    • Glucose measurement: confounding issues in setting targets for inpatient management: response to Dungan et al
    • Mahoney J., Ellison J. Glucose measurement: confounding issues in setting targets for inpatient management: response to Dungan et al. Diabetes Care 2007, 30:e71.
    • (2007) Diabetes Care , vol.30
    • Mahoney, J.1    Ellison, J.2
  • 25
    • 38949097440 scopus 로고    scopus 로고
    • Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters
    • Karon B.S., Griesmann L., Scott R., et al. Evaluation of the impact of hematocrit and other interference on the accuracy of hospital-based glucose meters. Diabetes Technol Ther 2008, 10:111-120.
    • (2008) Diabetes Technol Ther , vol.10 , pp. 111-120
    • Karon, B.S.1    Griesmann, L.2    Scott, R.3
  • 26
    • 0033849382 scopus 로고    scopus 로고
    • Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing
    • Tang Z., Lee J.H., Louie R.F., Kost G.J. Effects of different hematocrit levels on glucose measurements with handheld meters for point-of-care testing. Arch Pathol Lab Med 2000, 124:1135-1140.
    • (2000) Arch Pathol Lab Med , vol.124 , pp. 1135-1140
    • Tang, Z.1    Lee, J.H.2    Louie, R.F.3    Kost, G.J.4
  • 27
    • 0031761237 scopus 로고    scopus 로고
    • Positive interference of icodextrin metabolites in some enzymatic glucose methods
    • Janssen W., Harff G., Caers M., Schellekens A. Positive interference of icodextrin metabolites in some enzymatic glucose methods. Clin Chem 1998, 44:2379-2380.
    • (1998) Clin Chem , vol.44 , pp. 2379-2380
    • Janssen, W.1    Harff, G.2    Caers, M.3    Schellekens, A.4
  • 28
    • 19944365525 scopus 로고    scopus 로고
    • Screening guidelines for newborns at risk for low blood glucose
    • Aziz K., Dancey P. Screening guidelines for newborns at risk for low blood glucose. J Paediatr Child Health 2004, 9:723-729.
    • (2004) J Paediatr Child Health , vol.9 , pp. 723-729
    • Aziz, K.1    Dancey, P.2
  • 29
    • 79952203402 scopus 로고    scopus 로고
    • Postnatal glucose homeostasis in late-preterm and term infants
    • Adamkin D.H. Postnatal glucose homeostasis in late-preterm and term infants. Pediatrics 2011, 127:575-579.
    • (2011) Pediatrics , vol.127 , pp. 575-579
    • Adamkin, D.H.1
  • 30
    • 21444450978 scopus 로고    scopus 로고
    • The investigation and management of neonatal hypoglycaemia
    • Deshpande S., Ward Platt M. The investigation and management of neonatal hypoglycaemia. Semin Fetal Neonatal Med 2005, 10:351-361.
    • (2005) Semin Fetal Neonatal Med , vol.10 , pp. 351-361
    • Deshpande, S.1    Ward, P.M.2
  • 31
    • 0030662966 scopus 로고    scopus 로고
    • Neonatal hypoglycemia, Part I: background and definition
    • Halamek L.P., Benaron D.A., Stevenson D.K. Neonatal hypoglycemia, Part I: background and definition. Clin Pediatr 1997, 36:675-680.
    • (1997) Clin Pediatr , vol.36 , pp. 675-680
    • Halamek, L.P.1    Benaron, D.A.2    Stevenson, D.K.3
  • 32
    • 34249902075 scopus 로고    scopus 로고
    • Assessing the quality of glucose monitor studies: a critical evaluation of published reports
    • Mahoney J., Ellison J. Assessing the quality of glucose monitor studies: a critical evaluation of published reports. Clin Chem 2007, 53:1122-1128.
    • (2007) Clin Chem , vol.53 , pp. 1122-1128
    • Mahoney, J.1    Ellison, J.2
  • 33
    • 60249099082 scopus 로고    scopus 로고
    • Consensus report of the coalition for clinical research-self-monitoring of blood glucose
    • Klonoff D.C., Bergenstal R., Blonde L., et al. Consensus report of the coalition for clinical research-self-monitoring of blood glucose. J Diabetes Sci Technol 2008, 2:1030-1053.
    • (2008) J Diabetes Sci Technol , vol.2 , pp. 1030-1053
    • Klonoff, D.C.1    Bergenstal, R.2    Blonde, L.3
  • 34
    • 84875203570 scopus 로고    scopus 로고
    • Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems
    • Thorpe G.H. Assessing the quality of publications evaluating the accuracy of blood glucose monitoring systems. Diabetes Technol Ther 2013, 15:253-259.
    • (2013) Diabetes Technol Ther , vol.15 , pp. 253-259
    • Thorpe, G.H.1
  • 35
    • 20544458073 scopus 로고    scopus 로고
    • What is accuracy and how close must the agreement be?
    • Nichols J.H. What is accuracy and how close must the agreement be?. Diabetes Technol Ther 2005, 7:558-562.
    • (2005) Diabetes Technol Ther , vol.7 , pp. 558-562
    • Nichols, J.H.1
  • 36
    • 84865146825 scopus 로고    scopus 로고
    • Accuracy in blood glucose measurement: what will a tightening of requirements yield?
    • Heinemann L., Lodwig V., Freckmann G. Accuracy in blood glucose measurement: what will a tightening of requirements yield?. J Diabetes Sci Technol 2012, 6:435-443.
    • (2012) J Diabetes Sci Technol , vol.6 , pp. 435-443
    • Heinemann, L.1    Lodwig, V.2    Freckmann, G.3
  • 37
    • 36348960172 scopus 로고    scopus 로고
    • Assessing glucose monitor performance-a standardized approach
    • Mahoney J.J., Ellison J.M. Assessing glucose monitor performance-a standardized approach. Diabetes Technol Ther 2007, 9:545-552.
    • (2007) Diabetes Technol Ther , vol.9 , pp. 545-552
    • Mahoney, J.J.1    Ellison, J.M.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.