Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197

Author keywords

Blood glucose monitoring systems; Conformit europ enne mark; DIN EN ISO 15197:2003; Lot to lot variability; Self monitoring of blood glucose; System accuracy

Indexed keywords

GLUCOSE; GLUCOSE OXIDASE; HEXOKINASE;

EID: 84875803673     PISSN: None     EISSN: 19322968     Source Type: Journal    
DOI: 10.1177/193229681200600511     Document Type: Article
Times cited : (85)

References (26)
  • 1
    • 0027370108 scopus 로고
    • The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
    • The Diabetes Control and Complications Trial Research Group
    • The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977-86.
    • (1993) N Engl J Med , vol.329 , Issue.14 , pp. 977-986
  • 2
    • 26844470094 scopus 로고    scopus 로고
    • Current evidence regarding the value of self-monitored blood glucose testing
    • DOI 10.1016/j.amjmed.2005.07.053, PII S0002934305006650
    • Blonde L, Karter AJ. Current evidence regarding the value of self-monitored blood glucose testing. Am J Med. 2005;118(Suppl 9A):20S-6S. (Pubitemid 41446939)
    • (2005) American Journal of Medicine , vol.118 , Issue.9 SUPPL.
    • Blonde, L.1    Karter, A.J.2
  • 3
    • 33744828023 scopus 로고    scopus 로고
    • Global Guideline for Type 2 Diabetes: Recommendations for standard, comprehensive, and minimal care
    • IDF Clinical Guidelines Task Force
    • IDF Clinical Guidelines Task Force. Global Guideline for Type 2 Diabetes: recommendations for standard, comprehensive, and minimal care. Diabet Med. 2006;23(6):579-93.
    • (2006) Diabet Med , vol.23 , Issue.6 , pp. 579-593
  • 5
    • 26844457935 scopus 로고    scopus 로고
    • The role of self-monitoring of blood glucose in the care of people with diabetes: Report of a global consensus conference
    • DOI 10.1016/j.amjmed.2005.07.055, PII S0002934305006625
    • Bergenstal RM, Gavin JR 3rd; Global Consensus Conference on Glucose Monitoring Panel. The role of self-monitoring of blood glucose in the care of people with diabetes: report of a global consensus conference. Am J Med. 2005;118(Suppl 9A):1S-6S. (Pubitemid 41446936)
    • (2005) American Journal of Medicine , vol.118 , Issue.9 SUPPL.
    • Bergenstal, R.M.1    Gavin III, J.R.2
  • 6
    • 31844437604 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose in type 2 diabetes and long-term outcome: An epidemiological cohort study
    • DOI 10.1007/s00125-005-0083-5
    • Martin S, Schneider B, Heinemann L, Lodwig V, Kurth HJ, Kolb H, Scherbaum WA. Self-monitoring of blood glucose in type 2 diabetes and long-term outcome: an epidemiological cohort study. Diabetologia. 2006;49(2):271-8. (Pubitemid 43185395)
    • (2006) Diabetologia , vol.49 , Issue.2 , pp. 271-278
    • Martin, S.1    Schneider, B.2    Heinemann, L.3    Lodwig, V.4    Kurth, H.-J.5    Kolb, H.6    Scherbaum, W.A.7
  • 7
    • 79951717523 scopus 로고    scopus 로고
    • Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: Results from the Structured Testing Program study
    • Polonsky WH, Fisher L, Schikman CH, Hinnen DA, Parkin CG, Jelsovsky Z, Petersen B, Schweitzer M, Wagner RS. Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: results from the Structured Testing Program study. Diabetes Care. 2011;34(2):262-7.
    • (2011) Diabetes Care , vol.34 , Issue.2 , pp. 262-267
    • Polonsky, W.H.1    Fisher, L.2    Schikman, C.H.3    Hinnen, D.A.4    Parkin, C.G.5    Jelsovsky, Z.6    Petersen, B.7    Schweitzer, M.8    Wagner, R.S.9
  • 10
    • 0031753863 scopus 로고    scopus 로고
    • Clinical evaluation of a new, miniaturized biosensor for self- monitoring of blood glucose
    • Albertson C, Davis C, Ellison J, Chu C. Clinical evaluation of a new, miniaturized biosensor for self-monitoring of blood glucose. Clin Chem. 1998;44(9):2056-7. (Pubitemid 28405790)
    • (1998) Clinical Chemistry , vol.44 , Issue.9 , pp. 2056-2057
    • Albertson, C.1    Davis, C.2    Ellison, J.3    Chu, C.4
  • 11
    • 0036092613 scopus 로고    scopus 로고
    • Instruments for self-monitoring of blood glucose: Comparisons of testing quality achieved by patients and a technician
    • Skeie S, Thue G, Nerhus K, Sandberg S. Instruments for self-monitoring of blood glucose: comparisons of testing quality achieved by patients and a technician. Clin Chem. 2002;48(7):994-1003. (Pubitemid 34680893)
    • (2002) Clinical Chemistry , vol.48 , Issue.7 , pp. 994-1003
    • Skeie, S.1    Thue, G.2    Nerhus, K.3    Sandberg, S.4
  • 13
    • 57349106743 scopus 로고    scopus 로고
    • Standardized evaluation of nine instruments for self-monitoring of blood glucose
    • Kristensen GB, Monsen G, Skeie S, Sandberg S. Standardized evaluation of nine instruments for self-monitoring of blood glucose. Diabetes Technol Ther. 2008;10(6):467-77.
    • (2008) Diabetes Technol Ther , vol.10 , Issue.6 , pp. 467-477
    • Kristensen, G.B.1    Monsen, G.2    Skeie, S.3    Sandberg, S.4
  • 14
    • 24144435337 scopus 로고    scopus 로고
    • Between-lot variation in external quality assessment of glucose: Clinical importance and effect on participant performance evaluation
    • DOI 10.1373/clinchem.2005.049080
    • Kristensen GB, Christensen NG, Thue G, Sandberg S. Betweenlot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem. 2005;51(9):1632-6. (Pubitemid 41233098)
    • (2005) Clinical Chemistry , vol.51 , Issue.9 , pp. 1632-1636
    • Kristensen, G.B.B.1    Christensen, N.G.2    Thue, G.3    Sandberg, S.4
  • 15
    • 0030561530 scopus 로고    scopus 로고
    • A comparison of statistical techniques to evaluate the performance of the glucometer Elite blood glucose meter
    • DOI 10.1016/S0009-9120(96)00092-6
    • Harrison B, Markes R, Bradley P, Ismail IA. A comparison of statistical techniques to evaluate the performance of the Glucometer Elite blood glucose meter. Clin Biochem. 1996;29(6):521-7. (Pubitemid 26398904)
    • (1996) Clinical Biochemistry , vol.29 , Issue.6 , pp. 521-527
    • Harrison, B.1    Markes, R.2    Bradley, P.3    Ismail, I.A.4
  • 16
    • 84875805117 scopus 로고    scopus 로고
    • System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197
    • Freckmann G, Schmid C, Baumstark A, Pleus S, Link M, Haug C. System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN EN ISO 15197. J Diabetes Sci Technol. 2012;6(5):1060-75.
    • (2012) J Diabetes Sci Technol , vol.6 , Issue.5 , pp. 1060-1075
    • Freckmann, G.1    Schmid, C.2    Baumstark, A.3    Pleus, S.4    Link, M.5    Haug, C.6
  • 17
    • 0030053820 scopus 로고    scopus 로고
    • Evaluation of the YSI 2300 glucose analyzer: Algorithm-corrected results are accurate and specific
    • DOI 10.1016/0009-9120(95)02010-1
    • Astles JR, Sedor FA, Toffaletti JG. Evaluation of the YSI 2300 glucose analyzer: algorithm-corrected results are accurate and specific. Clin Biochem. 1996;29(1):27-31. (Pubitemid 26073095)
    • (1996) Clinical Biochemistry , vol.29 , Issue.1 , pp. 27-31
    • Astles, J.R.1    Sedor, F.A.2    Toffaletti, J.G.3
  • 18
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • PII S0140673686908378
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. (Pubitemid 16134762)
    • (1986) Lancet , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 19
    • 3142583462 scopus 로고    scopus 로고
    • Plasma glucose measurement with the Yellow Spring Glucose 2300 STAT and the Olympus AU640
    • DOI 10.1136/jcp.2003.013417
    • Twomey PJ. Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640. J Clin Pathol. 2004;57(7):752-4. (Pubitemid 38901492)
    • (2004) Journal of Clinical Pathology , vol.57 , Issue.7 , pp. 752-754
    • Twomey, P.J.1
  • 20
    • 70350209515 scopus 로고    scopus 로고
    • Factors affecting blood glucose monitoring: Sources of errors in measurement
    • Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol. 2009;3(4):903-13.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.4 , pp. 903-913
    • Ginsberg, B.H.1
  • 21
    • 80052051806 scopus 로고    scopus 로고
    • Effect of ambient temperature on analytical performance of self-monitoring blood glucose systems
    • Nerhus K, Rustad P, Sandberg S. Effect of ambient temperature on analytical performance of self-monitoring blood glucose systems. Diabetes Technol Ther. 2011;13(9):883-92.
    • (2011) Diabetes Technol Ther , vol.13 , Issue.9 , pp. 883-892
    • Nerhus, K.1    Rustad, P.2    Sandberg, S.3
  • 23
    • 79956160504 scopus 로고    scopus 로고
    • Glucose monitoring after fruit peeling: Pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: Wash your hands with tap water before you check blood glucose level
    • Hirose T, Mita T, Fujitani Y, Kawamori R, Watada H. Glucose monitoring after fruit peeling: pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: wash your hands with tap water before you check blood glucose level. Diabetes Care. 2011;34(3):596-7.
    • (2011) Diabetes Care , vol.34 , Issue.3 , pp. 596-597
    • Hirose, T.1    Mita, T.2    Fujitani, Y.3    Kawamori, R.4    Watada, H.5
  • 25
    • 79955557642 scopus 로고    scopus 로고
    • Accuracy and precision evaluation of seven self-monitoring blood glucose systems
    • Kuo CY, Hsu CT, Ho CS, Su TE, Wu MH, Wang CJ. Accuracy and precision evaluation of seven self-monitoring blood glucose systems. Diabetes Technol Ther. 2011;13(5):596-600.
    • (2011) Diabetes Technol Ther , vol.13 , Issue.5 , pp. 596-600
    • Kuo, C.Y.1    Hsu, C.T.2    Ho, C.S.3    Su, T.E.4    Wu, M.H.5    Wang, C.J.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.