System accuracy evaluation of 43 blood glucose monitoring systems for self-monitoring of blood glucose according to DIN en ISO 15197

Author keywords

Blood glucose monitoring systems; Conformit europ enne mark; DIN EN ISO 15197:2003; Self monitoring of blood glucose; System accuracy

Indexed keywords

ARTICLE; BLOOD GLUCOSE METER; BLOOD GLUCOSE MONITORING; CLINICAL ASSESSMENT; COMFORT; DIAGNOSTIC ACCURACY; DIAGNOSTIC TEST ACCURACY STUDY; EVALUATION STUDY; GLUCOSE BLOOD LEVEL; HUMAN; PACKAGING; QUALITY CONTROL; SELF MONITORING; STANDARD; TEST STRIP; ADOLESCENT; ADULT; AGED; BLOOD; COMPARATIVE STUDY; DEVICE APPROVAL; DEVICES; DIABETES MELLITUS; EUROPEAN UNION; EVALUATION; FEMALE; INSTRUMENTATION; INSULIN INFUSION; LEGAL ASPECT; MALE; MIDDLE AGED; MULTICENTER STUDY; REPRODUCIBILITY;

EID: 84875805117     PISSN: None     EISSN: 19322968     Source Type: Journal    
DOI: 10.1177/193229681200600510     Document Type: Article
Times cited : (212)

References (39)
  • 1
    • 0027370108 scopus 로고
    • The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
    • The Diabetes Control and Complications Trial Research Group
    • The Diabetes Control and Complications Trial Research Group. The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med. 1993;329(14):977-86.
    • (1993) N Engl J Med , vol.329 , Issue.14 , pp. 977-986
  • 2
    • 26844470094 scopus 로고    scopus 로고
    • Current evidence regarding the value of self-monitored blood glucose testing
    • DOI 10.1016/j.amjmed.2005.07.053, PII S0002934305006650
    • Blonde L, Karter AJ. Current evidence regarding the value of self-monitored blood glucose testing. Am J Med. 2005;118(Suppl 9A):20S-6S. (Pubitemid 41446939)
    • (2005) American Journal of Medicine , vol.118 , Issue.9 SUPPL.
    • Blonde, L.1    Karter, A.J.2
  • 3
    • 33744828023 scopus 로고    scopus 로고
    • Global Guideline for Type 2 Diabetes: Recommendations for standard, comprehensive, and minimal care
    • IDF Clinical Guidelines Task Force
    • IDF Clinical Guidelines Task Force. Global Guideline for Type 2 Diabetes: recommendations for standard, comprehensive, and minimal care. Diabet Med. 2006;23(6):579-93.
    • (2006) Diabet Med , vol.23 , Issue.6 , pp. 579-593
  • 5
    • 33745249928 scopus 로고    scopus 로고
    • Continuous subcutaneous glucose monitoring in children with type 1 diabetes mellitus: A single-blind, randomized, controlled trial
    • Lagarde WH, Barrows FP, Davenport ML, Kang M, Guess HA, Calikoglu AS. Continuous subcutaneous glucose monitoring in children with type 1 diabetes mellitus: a single-blind, randomized, controlled trial. Pediatr Diabetes. 2006;7(3):159-64.
    • (2006) Pediatr Diabetes , vol.7 , Issue.3 , pp. 159-164
    • Lagarde, W.H.1    Barrows, F.P.2    Davenport, M.L.3    Kang, M.4    Guess, H.A.5    Calikoglu, A.S.6
  • 6
    • 26844457935 scopus 로고    scopus 로고
    • The role of self-monitoring of blood glucose in the care of people with diabetes: Report of a global consensus conference
    • DOI 10.1016/j.amjmed.2005.07.055, PII S0002934305006625
    • Bergenstal RM, Gavin JR 3rd; Global Consensus Conference on Glucose Monitoring Panel. The role of self-monitoring of blood glucose in the care of people with diabetes: report of a global consensus conference. Am J Med. 2005;118(Suppl 9A):1S-6S. (Pubitemid 41446936)
    • (2005) American Journal of Medicine , vol.118 , Issue.9 SUPPL.
    • Bergenstal, R.M.1    Gavin III, J.R.2
  • 7
    • 31844437604 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose in type 2 diabetes and long-term outcome: An epidemiological cohort study
    • DOI 10.1007/s00125-005-0083-5
    • Martin S, Schneider B, Heinemann L, Lodwig V, Kurth HJ, Kolb H, Scherbaum WA. Self-monitoring of blood glucose in type 2 diabetes and long-term outcome: an epidemiological cohort study. Diabetologia. 2006;49(2):271-8. (Pubitemid 43185395)
    • (2006) Diabetologia , vol.49 , Issue.2 , pp. 271-278
    • Martin, S.1    Schneider, B.2    Heinemann, L.3    Lodwig, V.4    Kurth, H.-J.5    Kolb, H.6    Scherbaum, W.A.7
  • 8
    • 79951717523 scopus 로고    scopus 로고
    • Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: Results from the Structured Testing Program study
    • Polonsky WH, Fisher L, Schikman CH, Hinnen DA, Parkin CG, Jelsovsky Z, Petersen B, Schweitzer M, Wagner RS. Structured self-monitoring of blood glucose significantly reduces A1C levels in poorly controlled, noninsulin-treated type 2 diabetes: results from the Structured Testing Program study. Diabetes Care. 2011;34(2):262-7.
    • (2011) Diabetes Care , vol.34 , Issue.2 , pp. 262-267
    • Polonsky, W.H.1    Fisher, L.2    Schikman, C.H.3    Hinnen, D.A.4    Parkin, C.G.5    Jelsovsky, Z.6    Petersen, B.7    Schweitzer, M.8    Wagner, R.S.9
  • 9
    • 78649336430 scopus 로고    scopus 로고
    • Maltose interference-free test strips for blood glucose testing at point-of-care: A laboratory performance evaluation
    • Ng WY, Tiong CC, Jacob E. Maltose interference-free test strips for blood glucose testing at point-of-care: a laboratory performance evaluation. Diabetes Technol Ther. 2010;12(11):889-93.
    • (2010) Diabetes Technol Ther , vol.12 , Issue.11 , pp. 889-893
    • Ng, W.Y.1    Tiong, C.C.2    Jacob, E.3
  • 10
    • 77954670191 scopus 로고    scopus 로고
    • Individuals achieve more accurate results with meters that are codeless and employ dynamic electrochemistry
    • Rao A, Wiley M, Iyengar S, Nadeau D, Carnevale J. Individuals achieve more accurate results with meters that are codeless and employ dynamic electrochemistry. J Diabetes Sci Technol. 2010;4(1):145-50.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.1 , pp. 145-150
    • Rao, A.1    Wiley, M.2    Iyengar, S.3    Nadeau, D.4    Carnevale, J.5
  • 11
    • 84857437572 scopus 로고    scopus 로고
    • Dynamic electrochemistry: A step in the right direction
    • Rice MJ. Dynamic electrochemistry: a step in the right direction. J Diabetes Sci Technol. 2011;5(5):1176-8.
    • (2011) J Diabetes Sci Technol , vol.5 , Issue.5 , pp. 1176-1178
    • Rice, M.J.1
  • 12
    • 84857431968 scopus 로고    scopus 로고
    • Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices
    • Musholt PB, Schipper C, Thomé N, Ramljak S, Schmidt M, Forst T, Pfützner A. Dynamic electrochemistry corrects for hematocrit interference on blood glucose determinations with patient self-measurement devices. J Diabetes Sci Technol. 2011;5(5):1167-75.
    • (2011) J Diabetes Sci Technol , vol.5 , Issue.5 , pp. 1167-1175
    • Musholt, P.B.1    Schipper, C.2    Thomé, N.3    Ramljak, S.4    Schmidt, M.5    Forst, T.6    Pfützner, A.7
  • 13
    • 84859636768 scopus 로고    scopus 로고
    • Dynamic electrochemistry: An innovative method for high-quality blood glucose monitoring
    • Kempf K, Heinemann L. Dynamic electrochemistry: an innovative method for high-quality blood glucose monitoring. Diabetologie und Stoffwechsel. 2012;7(2):121-6.
    • (2012) Diabetologie und Stoffwechsel , vol.7 , Issue.2 , pp. 121-126
    • Kempf, K.1    Heinemann, L.2
  • 17
    • 58549091951 scopus 로고    scopus 로고
    • Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen
    • Bundesärztekammer. Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Dtsch Ärztebl. 2008;105(7):341-55.
    • (2008) Dtsch Ärztebl , vol.105 , Issue.7 , pp. 341-355
  • 18
    • 0030053820 scopus 로고    scopus 로고
    • Evaluation of the YSI 2300 glucose analyzer: Algorithm-corrected results are accurate and specific
    • DOI 10.1016/0009-9120(95)02010-1
    • Astles JR, Sedor FA, Toffaletti JG. Evaluation of the YSI 2300 glucose analyzer: algorithm-corrected results are accurate and specific. Clin Biochem. 1996;29(1):27-31. (Pubitemid 26073095)
    • (1996) Clinical Biochemistry , vol.29 , Issue.1 , pp. 27-31
    • Astles, J.R.1    Sedor, F.A.2    Toffaletti, J.G.3
  • 19
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • PII S0140673686908378
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307-10. (Pubitemid 16134762)
    • (1986) Lancet , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 20
    • 77951733801 scopus 로고    scopus 로고
    • A review of standards and statistics used to describe blood glucose monitor performance
    • Krouwer JS, Cembrowski GS. A review of standards and statistics used to describe blood glucose monitor performance. J Diabetes Sci Technol. 2010;4(1):75-83.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.1 , pp. 75-83
    • Krouwer, J.S.1    Cembrowski, G.S.2
  • 21
    • 84865146825 scopus 로고    scopus 로고
    • Accuracy in blood glucose measurement: What will a tightening of requirements yield?
    • Heinemann L, Lodwig V, Freckmann G. Accuracy in blood glucose measurement: what will a tightening of requirements yield? J Diabetes Sci Technol. 2012;6(2):435-43.
    • (2012) J Diabetes Sci Technol , vol.6 , Issue.2 , pp. 435-443
    • Heinemann, L.1    Lodwig, V.2    Freckmann, G.3
  • 23
    • 80051699538 scopus 로고    scopus 로고
    • Blutzuckerselbstmessung heute: Sind alle Geräte gleich?
    • Freckmann G, Haug C, Heinemann L. Blutzuckerselbstmessung heute: Sind alle Geräte gleich? Diabetes Stoffwechsel Herz. 2011;20(4):235-41.
    • (2011) Diabetes Stoffwechsel Herz , vol.20 , Issue.4 , pp. 235-241
    • Freckmann, G.1    Haug, C.2    Heinemann, L.3
  • 24
    • 1442275314 scopus 로고    scopus 로고
    • The Need for Separate Performance Goals for Glucose Sensors in the Hypoglycemic, Normoglycemic, and Hyperglycemic Ranges
    • DOI 10.2337/diacare.27.3.834
    • Klonoff DC. The need for separate performance goals for glucose sensors in the hypoglycemic, normoglycemic, and hyperglycemic ranges. Diabetes Care. 2004;27(3):834-6. (Pubitemid 38280561)
    • (2004) Diabetes Care , vol.27 , Issue.3 , pp. 834-836
    • Klonoff, D.C.1
  • 25
    • 79952199087 scopus 로고    scopus 로고
    • We need tighter regulatory standards for blood glucose monitoring, but they should be for accuracy disclosure
    • Ginsberg BH. We need tighter regulatory standards for blood glucose monitoring, but they should be for accuracy disclosure. J Diabetes Sci Technol. 2010;4(5):1265-8.
    • (2010) J Diabetes Sci Technol , vol.4 , Issue.5 , pp. 1265-1268
    • Ginsberg, B.H.1
  • 26
    • 84865134562 scopus 로고    scopus 로고
    • New Criteria for Assessing the Accuracy of Blood Glucose Monitors Meeting, October 28, 2011
    • Walsh J, Roberts R, Vigersky RA, Schwartz F. New Criteria for Assessing the Accuracy of Blood Glucose Monitors Meeting, October 28, 2011. J Diabetes Sci Technol. 2012;6(2):466-74.
    • (2012) J Diabetes Sci Technol , vol.6 , Issue.2 , pp. 466-474
    • Walsh, J.1    Roberts, R.2    Vigersky, R.A.3    Schwartz, F.4
  • 27
    • 3142583462 scopus 로고    scopus 로고
    • Plasma glucose measurement with the Yellow Spring Glucose 2300 STAT and the Olympus AU640
    • DOI 10.1136/jcp.2003.013417
    • Twomey PJ. Plasma glucose measurement with the Yellow Springs Glucose 2300 STAT and the Olympus AU640. J Clin Pathol. 2004;57(7):752-4. (Pubitemid 38901492)
    • (2004) Journal of Clinical Pathology , vol.57 , Issue.7 , pp. 752-754
    • Twomey, P.J.1
  • 29
    • 70350209515 scopus 로고    scopus 로고
    • Factors affecting blood glucose monitoring: Sources of errors in measurement
    • Ginsberg BH. Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol. 2009;3(4):903-13.
    • (2009) J Diabetes Sci Technol , vol.3 , Issue.4 , pp. 903-913
    • Ginsberg, B.H.1
  • 30
    • 44249100012 scopus 로고    scopus 로고
    • The technology behind glucose meters: Test strips
    • DOI 10.1089/dia.2008.0005
    • Hönes J, Müller P, Surridge N. The technology behind glucose meters: test strips. Diabetes Technol Ther. 2008;10(S1):S10-26. (Pubitemid 351725046)
    • (2008) Diabetes Technology and Therapeutics , vol.10 , Issue.SUPPL. 1
    • Hones, J.1    Muller, P.2    Surridge, N.3
  • 31
    • 0035023293 scopus 로고    scopus 로고
    • Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing
    • Tang Z, Louie RF, Lee JH, Lee DM, Miller EE, Kost GJ. Oxygen effects on glucose meter measurements with glucose dehydrogenase- and oxidase-based test strips for point-of-care testing. Crit Care Med. 2001;29(5):1062-70. (Pubitemid 32424364)
    • (2001) Critical Care Medicine , vol.29 , Issue.5 , pp. 1062-1070
    • Tang, Z.1    Louie, R.F.2    Lee, J.H.3    Lee, D.M.4    Miller, E.E.5    Kost, G.J.6
  • 32
    • 0033764601 scopus 로고    scopus 로고
    • Oxygen effects on glucose measurements with a reference analyzer and three handheld meters
    • Tang Z, Louie RF, Payes M, Chang KC, Kost GJ. Oxygen effects on glucose measurements with a reference analyzer and three handheld meters. Diabetes Technol Ther. 2000;2(3):349-62.
    • (2000) Diabetes Technol Ther , vol.2 , Issue.3 , pp. 349-362
    • Tang, Z.1    Louie, R.F.2    Payes, M.3    Chang, K.C.4    Kost, G.J.5
  • 33
    • 0031028799 scopus 로고    scopus 로고
    • Influence of blood sample oxygen tension on blood glucose concentration measured using an enzyme-electrode method
    • DOI 10.1097/00003246-199702000-00006
    • Kurahashi K, Maruta H, Usuda Y, Ohtsuka M. Influence of blood sample oxygen tension on blood glucose concentration measured using an enzyme-electrode method. Crit Care Med. 1997;25(2):231-5. (Pubitemid 27086055)
    • (1997) Critical Care Medicine , vol.25 , Issue.2 , pp. 231-235
    • Kurahashi, K.1    Maruta, H.2    Usuda, Y.3    Ohtsuka, M.4
  • 34
    • 57349106743 scopus 로고    scopus 로고
    • Standardized evaluation of nine instruments for self-monitoring of blood glucose
    • Kristensen GB, Monsen G, Skeie S, Sandberg S. Standardized evaluation of nine instruments for self-monitoring of blood glucose. Diabetes Technol Ther. 2008;10(6):467-77.
    • (2008) Diabetes Technol Ther , vol.10 , Issue.6 , pp. 467-477
    • Kristensen, G.B.1    Monsen, G.2    Skeie, S.3    Sandberg, S.4
  • 35
    • 79955557642 scopus 로고    scopus 로고
    • Accuracy and precision evaluation of seven self-monitoring blood glucose systems
    • Kuo CY, Hsu CT, Ho CS, Su TE, Wu MH, Wang CJ. Accuracy and precision evaluation of seven self-monitoring blood glucose systems. Diabetes Technol Ther. 2011;13(5):596-600.
    • (2011) Diabetes Technol Ther , vol.13 , Issue.5 , pp. 596-600
    • Kuo, C.Y.1    Hsu, C.T.2    Ho, C.S.3    Su, T.E.4    Wu, M.H.5    Wang, C.J.6
  • 36
    • 84859313151 scopus 로고    scopus 로고
    • Accuracy evaluation of five blood glucose monitoring systems obtained from the pharmacy: A european multicenter study with 453 subjects
    • Tack C, Pohlmeier H, Behnke T, Schmid V, Grenningloh M, Forst T, Pfützner A. Accuracy evaluation of five blood glucose monitoring systems obtained from the pharmacy: a european multicenter study with 453 subjects. Diabetes Technol Ther. 2012;14(4):330-7.
    • (2012) Diabetes Technol Ther , vol.14 , Issue.4 , pp. 330-337
    • Tack, C.1    Pohlmeier, H.2    Behnke, T.3    Schmid, V.4    Grenningloh, M.5    Forst, T.6    Pfützner, A.7
  • 38


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.