메뉴 건너뛰기




Volumn 28, Issue 1, 2012, Pages 195-212

Micron: An actively stabilized handheld tool for microsurgery

Author keywords

Compensation; medical robotics; optical tracking; piezoelectric devices

Indexed keywords

ERROR METRICES; EYE-HAND COORDINATION; FORCE CAPABILITIES; HAND MOTION; HANDHELDS; HUMAN VISUAL; HUMAN/MACHINE SYSTEMS; KEY COMPONENT; MANIPULATION TASK; MEDICAL ROBOTICS; OPTICAL TRACKING; POSITION ERRORS; RANGE OF MOTIONS;

EID: 84857031089     PISSN: 15523098     EISSN: None     Source Type: Journal    
DOI: 10.1109/TRO.2011.2169634     Document Type: Review
Times cited : (223)

References (58)
  • 2
    • 63949087940 scopus 로고    scopus 로고
    • The role of posture, magnification, and grip force on microscopic accuracy
    • B. Safwat et al., "The role of posture, magnification, and grip force on microscopic accuracy," Ann. Biomed. Eng. , vol. 37, pp. 997-1006, 2009.
    • (2009) Ann. Biomed. Eng. , vol.37 , pp. 997-1006
    • Safwat, B.1
  • 3
    • 0033774033 scopus 로고    scopus 로고
    • Macular hole surgery with and without internal limiting membrane peeling
    • H. L. Brooks, Jr., "Macular hole surgery with and without internal limiting membrane peeling," Ophthalmol. , vol. 107, pp. 1939-1948, 2000.
    • (2000) Ophthalmol. , vol.107 , pp. 1939-1948
    • Brooks Jr., H.L.1
  • 4
    • 0033506013 scopus 로고    scopus 로고
    • Surgical decompression of branch retinal vein occlusion via arteriovenous crossing sheathotomy: A prospective review of 15 cases
    • E. M. Opremcak and R. A. Bruce, "Surgical decompression of branch retinal vein occlusion via arteriovenous crossing sheathotomy," Retina , vol. 19, pp. 1-5, 1999. (Pubitemid 30387533)
    • (1999) Retina , vol.19 , Issue.1 , pp. 1-5
    • Opremcak, E.M.1    Bruce, R.A.2
  • 5
    • 0033989115 scopus 로고    scopus 로고
    • A study of surgical approaches to retinal vascular occlusions
    • W. M. Tang and D. P. Han, "A study of surgical approaches to retinal vascular occlusions," Arch. Ophthalmol. , vol. 118, pp. 138-143, 2000. (Pubitemid 30036459)
    • (2000) Archives of Ophthalmology , vol.118 , Issue.1 , pp. 138-143
    • Tang, W.M.1    Han, D.P.2
  • 7
    • 0027408246 scopus 로고
    • A complete second gut induced by transplanted micromeres in the sea urchin embryo
    • A. Ransick and E. H. Davidson, "A complete second gut induced by transplanted micromeres in the sea urchin embryo," Science , vol. 259, pp. 1134-1138, 1993. (Pubitemid 23085351)
    • (1993) Science , vol.259 , Issue.5098 , pp. 1134-1138
    • Ransick, A.1    Davidson, E.H.2
  • 8
    • 20744460765 scopus 로고    scopus 로고
    • Micro/macro force-servoed gripper for precision photonics assembly and analysis
    • DOI 10.1017/S0263574704000839
    • R. M. Voyles and S. Hulst, "Micro/macro force-servoed gripper for precision photonics assembly and analysis," Robotica , vol. 23, pp. 401-408, 2005. (Pubitemid 40850412)
    • (2005) Robotica , vol.23 , Issue.4 , pp. 401-408
    • Voyles, R.M.1    Hulst, S.2
  • 9
    • 0004210199 scopus 로고
    • Baltimore, MD: The Johns Hopkins Univ. Press
    • R. J. Elble and W. C. Koller, Tremor. Baltimore, MD: The Johns Hopkins Univ. Press, 1990.
    • (1990) Tremor
    • Elble, R.J.1    Koller, W.C.2
  • 10
    • 0031960224 scopus 로고    scopus 로고
    • Interaction of tremor and magnification in a motor performance task with visual feedback
    • K. Vasilakos et al., "Interaction of tremor and magnification in a motor performance taskwith visual feedback," J. Motor Behav. , vol. 30, pp. 158-168, 1998. (Pubitemid 28228325)
    • (1998) Journal of Motor Behavior , vol.30 , Issue.2 , pp. 158-168
    • Vasilakos, K.1    Glass, L.2    Beuter, A.3
  • 11
    • 0030273144 scopus 로고    scopus 로고
    • Quantitative analysis of tremor time series
    • J. Timmer et al., "Quantitative analysis of tremor time series, " Electroencephalogr. Clin. Neurophysiol. , vol. 101, pp. 461-468, 1996.
    • (1996) Electroencephalogr. Clin. Neurophysiol. , vol.101 , pp. 461-468
    • Timmer, J.1
  • 12
    • 0018910137 scopus 로고
    • Mechanical and neurons feedback factors in postural hand tremor of normal subjects
    • R. N. Stiles, "Mechanical and neural feedback factors in postural hand tremor of normal subjects, " J. Neurophysiol., vo l. 44, pp. 40-59, 1980. (Pubitemid 10069519)
    • (1980) Journal of Neurophysiology , vol.44 , Issue.1 , pp. 40-59
    • Stiles, R.N.1
  • 13
    • 0034307947 scopus 로고    scopus 로고
    • Determinants of physiologic tremor in a large normal population
    • J. Raethjen et al., "Determinants of physiologic tremor in a large normal population," Clin. Neurophysiol. , vol. 111, pp. 1825-1837, 2000.
    • (2000) Clin. Neurophysiol. , vol.111 , pp. 1825-1837
    • Raethjen, J.1
  • 14
    • 77950888562 scopus 로고    scopus 로고
    • Relationship between hand stability and the 10-Hz physiological tremor during various manual tasks
    • H. Endo and K. Kawahara, "Relationship between hand stability and the 10-Hz physiological tremor during various manual tasks," Ergonom. , vol. 53, pp. 491-501, 2010.
    • (2010) Ergonom. , vol.53 , pp. 491-501
    • Endo, H.1    Kawahara, K.2
  • 17
    • 33750743698 scopus 로고    scopus 로고
    • The frequency of human, manual adjustments in balancing an inverted pendulum is constrained by intrinsic physiological factors
    • DOI 10.1113/jphysiol.2006.118786
    • I. D. Loram et al., "The frequency of human, manual adjustments in balancing an inverted pendulum is constrained by intrinsic physiological factors, " J. Physiol., Lond., vol. 577, pp. 417-432, 2006. (Pubitemid 44698265)
    • (2006) Journal of Physiology , vol.577 , Issue.1 , pp. 417-432
    • Loram, I.D.1    Gawthrop, P.J.2    Lakie, M.3
  • 18
    • 0017388349 scopus 로고
    • Frequency response of the human controller for various processes and stochastic testsignals
    • B. Kuchen et al., "Frequency response of the human controller for various processes and stochastic testsignals," Biol. Cybern. , vol. 27, pp. 33-39, 1977. (Pubitemid 8139452)
    • (1977) Biological Cybernetics , vol.27 , Issue.1 , pp. 33-39
    • Kuchen, B.1    Henning, K.2    Rake, H.3    Schaefer, O.4
  • 19
    • 84857030804 scopus 로고    scopus 로고
    • A review of medical robotics for minimally invasive soft tissue surgery
    • G. Dogangil et al., "A review of medical robotics for minimally invasive soft tissue surgery," in Proc. Inst. Mech. Eng., H , 2009, vol. 223, pp. 1-27.
    • (2009) Proc. Inst. Mech. Eng., H , vol.223 , pp. 1-27
    • Dogangil, G.1
  • 20
    • 0348110262 scopus 로고    scopus 로고
    • The da Vinci telerobotic surgical system: The virtual operative field and telepresence surgery
    • DOI 10.1016/S0039-6109(03)00164-6
    • G. Ballantyne and F. Moll, "The da Vinci telerobotic surgical system: the virtual operative field and telepresence surgery," Surg. Clin. North Amer., vol. 83, pp. 1293-1304, 2003. (Pubitemid 37543098)
    • (2003) Surgical Clinics of North America , vol.83 , Issue.6 , pp. 1293-1304
    • Ballantyne, G.H.1    Moll, F.2
  • 21
    • 38049164250 scopus 로고    scopus 로고
    • Feasibility study of intraocular robotic surgery with the da Vinci surgical system
    • D. H. Bourla et al., "Feasibility study of intraocular robotic surgery with the da Vinci surgical system," Retina , vol. 28, pp. 154-158, 2008.
    • (2008) Retina , vol.28 , pp. 154-158
    • Bourla, D.H.1
  • 22
    • 67549144555 scopus 로고    scopus 로고
    • A prototype surgical manipulator for robotic intraocular micro surgery
    • A. P. Mulgaonkar et al., "A prototype surgical manipulator for robotic intraocular micro surgery," Stud. Health Technol. Inform. , vol. 142, pp. 215-217, 2009.
    • (2009) Stud. Health Technol. Inform. , vol.142 , pp. 215-217
    • Mulgaonkar, A.P.1
  • 23
    • 0030700673 scopus 로고    scopus 로고
    • Toward robot-assisted vascular microsurgery in the retina
    • DOI 10.1007/BF01880668
    • P. S. Jensen et al., "Toward robot-assisted vascular microsurgery in the retina," Graefes Arch. Clin. Exp. Ophthalmol. , vol. 235, pp. 696-701, 1997. (Pubitemid 27497048)
    • (1997) Graefe's Archive for Clinical and Experimental Ophthalmology , vol.235 , Issue.11 , pp. 696-701
    • Jensen, P.S.1
  • 24
    • 0001254603 scopus 로고
    • A teleoperated microsurgical robot and associated virtual environment for eye surgery
    • I. W. Hunter et al., "A teleoperated microsurgical robot and associated virtual environment for eye surgery," Presence , vol. 2, pp. 265-280, 1993.
    • (1993) Presence , vol.2 , pp. 265-280
    • Hunter, I.W.1
  • 25
    • 67949117111 scopus 로고    scopus 로고
    • Robot-assisted vitreoretinal surgery: Development of a prototype and feasibility studies in an animal model
    • T. Ueta et al., "Robot-assisted vitreoretinal surgery: Development of a prototype and feasibility studies in an animal model," Ophthalmol. , vol. 116, pp. 1538-1543, 2009.
    • (2009) Ophthalmol. , vol.116 , pp. 1538-1543
    • Ueta, T.1
  • 26
    • 0032801309 scopus 로고    scopus 로고
    • Evaluation of a telerobotic system to assist surgeons in microsurgery
    • DOI 10.1002/(SICI)1097-0150(1999)4:1<15::AID-IGS2>3.0.CO;2-0
    • H. Das et al., "Evaluation of a telerobotic system to assist surgeons in microsurgery," Comput. Aided Surg. , vol. 4, pp. 15-25, 1999. (Pubitemid 29370258)
    • (1999) Computer Aided Surgery , vol.4 , Issue.1 , pp. 15-25
    • Das, H.1    Zak, H.2    Johnson, J.3    Crouch, J.4    Frambach, D.5
  • 29
    • 84957065500 scopus 로고    scopus 로고
    • Surgical forces and tactile perception during retinal microsurgery
    • P. Gupta et al., "Surgical forces and tactile perception during retinal microsurgery," Lect. Notes Comput. Sci. , vol. 1679, pp. 1218-1225, 1999.
    • (1999) Lect. Notes Comput. Sci. , vol.1679 , pp. 1218-1225
    • Gupta, P.1
  • 30
    • 84956685953 scopus 로고    scopus 로고
    • Performance measurement in scaled teleoperation for microsurgery
    • CVRMed-MRCAS'97: First Joint Conference Computer Vision, Virtual Reality and Robotics in Medicine and Medical Robotics and Computer-Assisted Surgery
    • S. E. Salcudean et al., "Performance measurement in scaled teleoperation for microsurgery," Lect. Notes Comput. Sci. , vol. 1205, pp. 789-798, 1997. (Pubitemid 127045618)
    • (1997) Lecture Notes in Computer Science , Issue.1205 , pp. 789-798
    • Salcudean, S.E.1    Ku, S.2    Bell, G.3
  • 31
    • 58149215973 scopus 로고    scopus 로고
    • Haptic feedback in robot-assisted minimally invasive surgery
    • A. M. Okamura, "Haptic feedback in robot-assisted minimally invasive surgery," Curr. Opin. Urol. , vol. 19, pp. 102-107, 2009.
    • (2009) Curr. Opin. Urol. , vol.19 , pp. 102-107
    • Okamura, A.M.1
  • 32
    • 32044466197 scopus 로고    scopus 로고
    • Haptic interaction in robot-assisted endoscopic surgery: A sensorized end-effector
    • M. Tavakoli et al., "Haptic interaction in robot-assisted endoscopic surgery: A sensorized end-effector," Int. J. Med. Robot. Comput. Assist. Surg. , vol. 1, p. 53, 2005.
    • (2005) Int. J. Med. Robot. Comput. Assist. Surg. , vol.1 , pp. 53
    • Tavakoli, M.1
  • 33
    • 67749130638 scopus 로고    scopus 로고
    • A sub-millimetric, 0.25mNr esolution fully integrated fiber-optic force-sensing tool for retinal microsurgery
    • I. Iordachita et al., "A sub-millimetric, 0.25mNr esolution fully integrated fiber-optic force-sensing tool for retinal microsurgery," Int. J. Comput. Assist. Radiol. Surg., vol. 4, pp. 383-390, 2009.
    • (2009) Int. J. Comput. Assist. Radiol. Surg. , vol.4 , pp. 383-390
    • Iordachita, I.1
  • 35
    • 0142196055 scopus 로고    scopus 로고
    • A miniature microsurgical instrument tip force sensor for enhanced force feedback during robot-assisted manipulation
    • Oct.
    • P. J. Berkelman et al., "A miniature microsurgical instrument tip force sensor for enhanced force feedback during robot-assisted manipulation, " IEEE Trans. Robot. Autom. , vol. 19, no. 5, pp. 917-921, Oct. 2003.
    • (2003) IEEE Trans. Robot. Autom. , vol.19 , Issue.5 , pp. 917-921
    • Berkelman, P.J.1
  • 36
    • 33746869732 scopus 로고    scopus 로고
    • Active tremor compensation in handheld instrument for microsurgery
    • Ph.D. dissertation Carnegie Mellon Univ., Pittsburgh, PA
    • W. T. Ang, "Active tremor compensation in handheld instrument for microsurgery, " Ph.D. dissertation, Robot. Inst., Carnegie Mellon Univ., Pittsburgh, PA, 2004.
    • (2004) Robot. Inst.
    • Ang, W.T.1
  • 38
    • 84903807072 scopus 로고    scopus 로고
    • Test of tracing performance with an active handheld micromanipulator
    • Aug.
    • D. Y. Choi et al., "Test of tracing performance with an active handheld micromanipulator," in Proc. 29th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., Aug. 2007, p. 3638-3641.
    • (2007) Proc. 29th Annu. Int. Conf. IEEE Eng. Med. Biol. Soc , pp. 3638-3641
    • Choi, D.Y.1
  • 39
    • 79951521613 scopus 로고    scopus 로고
    • Compact sensing design of a hand-held active tremor compensation instrument
    • Dec.
    • W. T. Latt et al., "Compact sensing design of a hand-held active tremor compensation instrument," IEEE Sens. J. , vol. 9, no. 12, pp. 1864-1871, Dec. 2009.
    • (2009) IEEE Sens. J. , vol.9 , Issue.12 , pp. 1864-1871
    • Latt, W.T.1
  • 40
    • 0030152291 scopus 로고    scopus 로고
    • Modeling and canceling tremor in human-machine interfaces
    • DOI 10.1109/51.499755
    • C. N. Riviere and N.V. Thakor, "Modeling and canceling tremor in humanmachine interfaces," IEEE Eng. Med. Biol.Mag., vol. 15, no. 3, pp. 29-36, May/Jun. 1996. (Pubitemid 26157212)
    • (1996) IEEE Engineering in Medicine and Biology Magazine , vol.15 , Issue.3 , pp. 29-36
    • Riviere, C.N.1    Thakor, N.V.2
  • 41
    • 0142258223 scopus 로고    scopus 로고
    • Toward active tremor canceling in handheld microsurgical instruments
    • Oct.
    • C. N. Riviere et al., "Toward active tremor canceling in handheld microsurgical instruments," IEEE Trans. Robot. Autom. , vol. 19, no. 5, pp. 793-800, Oct. 2003.
    • (2003) IEEE Trans. Robot. Autom. , vol.19 , Issue.5 , pp. 793-800
    • Riviere, C.N.1
  • 42
    • 0024029224 scopus 로고
    • Kinematic analysis of a three-degrees-offreedom in-parallel actuated manipulator
    • Jun.
    • K. M. Lee and D. K. Shah, "Kinematic analysis of a three-degrees-offreedom in-parallel actuated manipulator," IEEE J. Robot. Autom. , vol. 4, no. 3, pp. 354-360, Jun. 1988.
    • (1988) IEEE J. Robot. Autom. , vol.4 , Issue.3 , pp. 354-360
    • Lee, K.M.1    Shah, D.K.2
  • 43
    • 67349220000 scopus 로고    scopus 로고
    • High-speed microscale optical tracking using digital frequency-domain multiplexing
    • Jun.
    • R. A. MacLachlan and C. N. Riviere, "High-speed microscale optical tracking using digital frequency-domain multiplexing," IEEE Trans. Instrum. Meas. , vol. 58, no. 6, pp. 1991-2001, Jun. 2009.
    • (2009) IEEE Trans. Instrum. Meas. , vol.58 , Issue.6 , pp. 1991-2001
    • MacLachlan, R.A.1    Riviere, C.N.2
  • 44
    • 84975553202 scopus 로고
    • Closed-form solution of absolute orientation using unit quaternions
    • B. K. P. Horn, "Closed-form solution of absolute orientation using unit quaternions," J. Opt. Soc. Am. A. , vol. 4, pp. 629-642, 1987.
    • (1987) J. Opt. Soc. Am. A. , vol.4 , pp. 629-642
    • Horn, B.K.P.1
  • 45
    • 46349094959 scopus 로고    scopus 로고
    • Human-machine interface for robotic surgery and stereotaxy
    • Jun.
    • A. D. Greer et al., "Human-machine interface for robotic surgery and stereotaxy," IEEE/ASME Trans. Mechatron. , vol. 13, no. 3, pp. 355-361, Jun. 2008.
    • (2008) IEEE/ASME Trans. Mechatron. , vol.13 , Issue.3 , pp. 355-361
    • Greer, A.D.1
  • 46
    • 0030148431 scopus 로고    scopus 로고
    • Performance of feedforward and feedback systems for active control
    • PII S1063667696040667
    • S. J. Elliott and T. J. Sutton, "Performance of feedforward and feedback systems for active control," IEEE Trans. Speech Audio Process. , vol. 4, no. 3, pp. 214-223, May 1996. (Pubitemid 126753007)
    • (1996) IEEE Transactions on Speech and Audio Processing , vol.4 , Issue.3 , pp. 214-223
    • Elliott, S.J.1    Sutton, T.J.2
  • 49
    • 0028416822 scopus 로고
    • State-space interpretation of model predictive control
    • DOI 10.1016/0005-1098(94)90159-7
    • J. H. Lee et al., "State-space interpretation of model predictive control," Automatica , vol. 30, pp. 707-717, 1994. (Pubitemid 124008344)
    • (1994) Automatica , vol.30 , Issue.4 , pp. 707-717
    • Lee, J.H.1    Morari, M.2    Garcia, C.E.3
  • 50
    • 84929912614 scopus 로고
    • Anti-windup design for internal model control
    • A. Zheng et al., "Anti-windup design for internal model control," Int. J. Control , vol. 60, pp. 1015-1024, 1994.
    • (1994) Int. J. Control , vol.60 , pp. 1015-1024
    • Zheng, A.1
  • 51
    • 44349129968 scopus 로고    scopus 로고
    • Smith predictor control and internal model control: A tutorial
    • N. Abe and K. Yamanaka, "Smith predictor control and internal model control: A tutorial," in Proc. Annu. Conf. Soc. Instrum. Control Eng., 2003, vol. 2, pp. 1383-1387.
    • (2003) Proc. Annu. Conf. Soc. Instrum. Control Eng. , vol.2 , pp. 1383-1387
    • Abe, N.1    Yamanaka, K.2
  • 52
    • 34247335777 scopus 로고    scopus 로고
    • Feedforward controller with inverse rate-dependent model for piezoelectric actuators in trajectory-tracking applications
    • DOI 10.1109/TMECH.2007.892824
    • W. T. Ang et al., "Feedforward controller with inverse rate-dependent model for piezoelectric actuators in trajectory-tracking applications, " IEEE/ASME Trans. Mechatron. , vol. 12, no. 2, pp. 134-142, Apr. 2007. (Pubitemid 46629536)
    • (2007) IEEE/ASME Transactions on Mechatronics , vol.12 , Issue.2 , pp. 134-142
    • Ang, W.T.1    Khosla, P.K.2    Riviere, C.N.3
  • 53
    • 0020128096 scopus 로고
    • Improving the linearity of piezoelectric ceramic actuators
    • C. V. Newcomb and I. Flinn, "Improving the linearity of piezoelectric ceramic actuators," Electron. Lett. , vol. 18, pp. 442-444, 1982. (Pubitemid 12524802)
    • (1982) Electronics Letters , vol.18 , Issue.11 , pp. 442-444
    • Newcomb, C.V.1    Flinn, I.2
  • 54
    • 33750510737 scopus 로고    scopus 로고
    • Wideband and hysteresis-free regulation of piezoelectric actuator based on induced current for high-speed scanning probe microscopy
    • M. Kageshima et al., "Wideband and hysteresis-free regulation of piezoelectric actuator based on induced current for high-speed scanning probe microscopy, " Rev. Sci. Instrum. , vol. 77, art no. 103701, 6 pp., 2006.
    • (2006) Rev. Sci. Instrum. , vol.77 , pp. 1037016
    • Kageshima, M.1
  • 56
    • 34547796984 scopus 로고    scopus 로고
    • Challenges in image-guided therapy system design
    • S. DiMaio et al., "Challenges in image-guided therapy system design," Neuroimage , vol. 37, pp. S144-S151, 2007.
    • (2007) Neuroimage , vol.37
    • Dimaio, S.1
  • 57
    • 78650806539 scopus 로고    scopus 로고
    • Cell micromanipulation with an active handheld micromanipulator
    • IEEE Eng. Med. Biol. Soc., Aug./Sep.
    • J. C. Tabaŕes et al., "Cell micromanipulation with an active handheld micromanipulator," in Proc. 32nd Annu. Int. Conf. IEEE Eng. Med. Biol. Soc., Aug./Sep. 2010, pp. 4363-4366.
    • (2010) Proc. 32nd Annu. Int. Conf , pp. 4363-4366
    • Tabaŕes, J.C.1
  • 58
    • 77950535810 scopus 로고    scopus 로고
    • Semiautomated intraocular laser surgery using handheld instruments
    • B. C. Becker et al. Semiautomated intraocular laser surgery using handheld instruments Lasers Surg. Med. 42 264-273 2010
    • (2010) Lasers Surg. Med. , vol.42 , pp. 264-273
    • Becker, B.C.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.