Volumn 4, Issue 3, 2014, Pages 242-250

Continuous Cardiac Output Monitoring: What Do Validation Studies Tell Us?

Author keywords

Bland Altman; Cardiac output; Monitoring; Pulse contour; Thermodilution; Trending

Indexed keywords


EID: 84930009469     PISSN: 21676275     EISSN: 15233855     Source Type: Journal    
DOI: 10.1007/s40140-014-0062-9     Document Type: Article
Times cited : (9)

References (61)
  • 1
    • 0022979027 scopus 로고
    • Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance
    • COI: 1:STN:280:DyaL2s%2FgtlGrsw%3D%3D, PID: 3757531
    • Bernstein DP. Continuous noninvasive real-time monitoring of stroke volume and cardiac output by thoracic electrical bioimpedance. Crit Care Med. 1986;14(10):898–901. DOI: 10.1097/00003246-198610000-00015
    • (1986) Crit Care Med. , vol.14 , Issue.10 , pp. 898-901
    • Bernstein, D.P.1
  • 2
    • 84953385787 scopus 로고    scopus 로고
    • Evaluation of a cardiac output monitor (chapter 14)
    • Cannesson M, Pearse R, (eds), Cambridge University Press, Cambridge
    • Critchley LA. Evaluation of a cardiac output monitor (chapter 14). In: Cannesson M, Pearse R, editors. Perioperative hemodynamic monitoring and goal directed therapy. Cambridge: Cambridge University Press; 2014. p. 120–131
    • (2014) Perioperative hemodynamic monitoring and goal directed therapy , pp. 120-131
    • Critchley, L.A.1
  • 3
    • 0024430214 scopus 로고
    • On the source of peak first time derivative (dZ/dt) during impedance cardiography
    • COI: 1:STN:280:DyaK3c7itVajtw%3D%3D, PID: 2610417
    • Kubicek WG. On the source of peak first time derivative (dZ/dt) during impedance cardiography. Ann Biomed Eng. 1989;17(5):459–62. DOI: 10.1007/BF02368065
    • (1989) Ann Biomed Eng. , vol.17 , Issue.5 , pp. 459-462
    • Kubicek, W.G.1
  • 4
    • 0031830393 scopus 로고    scopus 로고
    • Impedance cardiography. The impact of new technology
    • COI: 1:STN:280:DyaK1cvksVWitw%3D%3D, PID: 9771176
    • Critchley LA. Impedance cardiography. The impact of new technology. Anaesthesia. 1998;53(7):677–84. DOI: 10.1046/j.1365-2044.1998.437-az0550.x
    • (1998) Anaesthesia. , vol.53 , Issue.7 , pp. 677-684
    • Critchley, L.A.1
  • 5
    • 26444571427 scopus 로고    scopus 로고
    • An investigation to show the effect of lung fluid on impedance cardiac output in the anaesthetized dog
    • PID: 16051651
    • Peng ZY, Critchley LA, Fok BS. An investigation to show the effect of lung fluid on impedance cardiac output in the anaesthetized dog. Br J Anaesth. 2005;95(4):458–64. DOI: 10.1093/bja/aei206
    • (2005) Br J Anaesth. , vol.95 , Issue.4 , pp. 458-464
    • Peng, Z.Y.1    Critchley, L.A.2    Fok, B.S.3
  • 6
    • 19044397672 scopus 로고    scopus 로고
    • The effect of peripheral resistance on impedance cardiography measurements in the anesthetized dog
    • PID: 15920200
    • Critchley LA, et al. The effect of peripheral resistance on impedance cardiography measurements in the anesthetized dog. Anesth Analg. 2005;100(6):1708–12. DOI: 10.1213/01.ANE.0000150602.40554.EB
    • (2005) Anesth Analg. , vol.100 , Issue.6 , pp. 1708-1712
    • Critchley, L.A.1
  • 7
    • 0029932049 scopus 로고    scopus 로고
    • Abdominal surgery alters the calibration of bioimpedance cardiac output measurement
    • COI: 1:STN:280:DyaK28zjvVGqsA%3D%3D, PID: 8738594
    • Critchley LA, Leung DH, Short TG. Abdominal surgery alters the calibration of bioimpedance cardiac output measurement. Int J Clin Monit Comput. 1996;13(1):1–8. DOI: 10.1007/BF02918206
    • (1996) Int J Clin Monit Comput. , vol.13 , Issue.1 , pp. 1-8
    • Critchley, L.A.1    Leung, D.H.2    Short, T.G.3
  • 8
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • COI: 1:STN:280:DyaL287hslWqsQ%3D%3D, PID: 2868172
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1(8476):307–10. DOI: 10.1016/S0140-6736(86)90837-8
    • (1986) Lancet. , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 9
    • 84872137892 scopus 로고    scopus 로고
    • Noninvasive cardiac output monitors: A state-of the-art review
    • This article provides an excellent review of the CO technology
    • ∙∙ Marik, P.E., Noninvasive cardiac output monitors: a state-of the-art review. J Cardiothorac Vasc Anesth. 2013;27(1):121–34. This article provides an excellent review of the CO technology.
    • (2013) J Cardiothorac Vasc Anesth , vol.27 , Issue.1 , pp. 121-134
    • Marik, P.E.1
  • 10
    • 33846035330 scopus 로고    scopus 로고
    • Comparison of blunt versus surgical dissection for aortic flow probe placement in the dog
    • Peng Z, Critchley L, James A. Comparison of blunt versus surgical dissection for aortic flow probe placement in the dog. Scand J Lab Anim Sci. 2006;33(4):227–35.
    • (2006) Scand J Lab Anim Sci. , vol.33 , Issue.4 , pp. 227-235
    • Peng, Z.1    Critchley, L.2    James, A.3
  • 11
    • 0033003166 scopus 로고    scopus 로고
    • A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    • COI: 1:STN:280:DC%2BD3s%2FmslCruw%3D%3D, PID: 12578081
    • Critchley LA, Critchley JA. A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques. J Clin Monit Comput. 1999;15(2):85–91. DOI: 10.1023/A:1009982611386
    • (1999) J Clin Monit Comput. , vol.15 , Issue.2 , pp. 85-91
    • Critchley, L.A.1    Critchley, J.A.2
  • 12
    • 78049343101 scopus 로고    scopus 로고
    • Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision
    • PID: 20881596
    • Peyton PJ, Chong SW. Minimally invasive measurement of cardiac output during surgery and critical care: a meta-analysis of accuracy and precision. Anesthesiology. 2010;113(5):1220–35. DOI: 10.1097/ALN.0b013e3181ee3130
    • (2010) Anesthesiology. , vol.113 , Issue.5 , pp. 1220-1235
    • Peyton, P.J.1    Chong, S.W.2
  • 13
    • 77949305811 scopus 로고    scopus 로고
    • Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives
    • PID: 20185659
    • Reuter DA, et al. Cardiac output monitoring using indicator-dilution techniques: basics, limits, and perspectives. Anesth Analg. 2010;110(3):799–811. DOI: 10.1213/ANE.0b013e3181cc885a
    • (2010) Anesth Analg. , vol.110 , Issue.3 , pp. 799-811
    • Reuter, D.A.1
  • 14
    • 78650944099 scopus 로고    scopus 로고
    • Determination of the precision error of the pulmonary artery thermodilution catheter using an in vitro continuous flow test rig
    • PID: 21127281
    • Yang XX, Critchley LA, Joynt GM. Determination of the precision error of the pulmonary artery thermodilution catheter using an in vitro continuous flow test rig. Anesth Analg. 2011;112(1):70–7. DOI: 10.1213/ANE.0b013e3181ff475e
    • (2011) Anesth Analg. , vol.112 , Issue.1 , pp. 70-77
    • Yang, X.X.1    Critchley, L.A.2    Joynt, G.M.3
  • 15
    • 57349199888 scopus 로고    scopus 로고
    • Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery
    • COI: 1:STN:280:DC%2BD1cjptVyisg%3D%3D, PID: 19059920
    • Biancofiore G, et al. Evaluation of an uncalibrated arterial pulse contour cardiac output monitoring system in cirrhotic patients undergoing liver surgery. Br J Anaesth. 2009;102(1):47–54. DOI: 10.1093/bja/aen343
    • (2009) Br J Anaesth. , vol.102 , Issue.1 , pp. 47-54
    • Biancofiore, G.1
  • 16
    • 78149469431 scopus 로고    scopus 로고
    • A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output
    • This article outlines the study design and statistics needed to show trending ability of a CO monitor
    • ∙∙ Critchley, L.A., A. Lee, and A.M. Ho, A critical review of the ability of continuous cardiac output monitors to measure trends in cardiac output. Anesth Analg. 2010;111(5):1180–92. This article outlines the study design and statistics needed to show trending ability of a CO monitor.
    • (2010) Anesth Analg , vol.111 , Issue.5 , pp. 1180-1192
    • Critchley, L.A.A.1    Leeho, A.M.2
  • 17
    • 84858775816 scopus 로고    scopus 로고
    • Thinking outside the (cardiac output) box
    • PID: 22425843
    • Michard F. Thinking outside the (cardiac output) box. Crit Care Med. 2012;40(4):1361–2. DOI: 10.1097/CCM.0b013e31824112fa
    • (2012) Crit Care Med. , vol.40 , Issue.4 , pp. 1361-1362
    • Michard, F.1
  • 18
    • 79957736830 scopus 로고    scopus 로고
    • Assessment of trending ability of cardiac output monitors by polar plot methodology
    • PID: 21419654
    • Critchley LA, Yang XX, Lee A. Assessment of trending ability of cardiac output monitors by polar plot methodology. J Cardiothorac Vasc Anesth. 2011;25(3):536–46. DOI: 10.1053/j.jvca.2011.01.003
    • (2011) J Cardiothorac Vasc Anesth. , vol.25 , Issue.3 , pp. 536-546
    • Critchley, L.A.1    Yang, X.X.2    Lee, A.3
  • 19
    • 85041088427 scopus 로고    scopus 로고
    • Minimally invasive cardiac output monitoring in the Year 2012 (chapter 4)
    • Aronow WS, (ed), InTech, Rijeka, Croatia
    • Critchley LA. Minimally invasive cardiac output monitoring in the Year 2012 (chapter 4). In: Aronow WS, editor. Arterial bypass. Rijeka, Croatia: InTech; 2013. p. 45–80.
    • (2013) Arterial bypass , pp. 45-80
    • Critchley, L.A.1
  • 20
    • 0025130516 scopus 로고
    • Continuous measurement of cardiac output with the use of stochastic system identification techniques
    • COI: 1:STN:280:DyaK3M%2FjvVWmug%3D%3D, PID: 2230861
    • Yelderman M. Continuous measurement of cardiac output with the use of stochastic system identification techniques. J Clin Monit. 1990;6(4):322–32. DOI: 10.1007/BF02842492
    • (1990) J Clin Monit. , vol.6 , Issue.4 , pp. 322-332
    • Yelderman, M.1
  • 21
    • 0031928196 scopus 로고    scopus 로고
    • A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system
    • COI: 1:STN:280:DyaK1czntlyltA%3D%3D, PID: 9710092
    • Mihm FG, et al. A multicenter evaluation of a new continuous cardiac output pulmonary artery catheter system. Crit Care Med. 1998;26(8):1346–50. DOI: 10.1097/00003246-199808000-00018
    • (1998) Crit Care Med. , vol.26 , Issue.8 , pp. 1346-1350
    • Mihm, F.G.1
  • 22
    • 59649101944 scopus 로고    scopus 로고
    • The cardiac output from blood pressure algorithms trial
    • PID: 19112280
    • Sun JX, et al. The cardiac output from blood pressure algorithms trial. Crit Care Med. 2009;37(1):72–80. DOI: 10.1097/CCM.0b013e3181930174
    • (2009) Crit Care Med. , vol.37 , Issue.1 , pp. 72-80
    • Sun, J.X.1
  • 23
    • 0027195184 scopus 로고
    • Computation of aortic flow from pressure in humans using a nonlinear, three-element model
    • COI: 1:STN:280:DyaK3szivVKhtw%3D%3D
    • Wesseling KH, et al. Computation of aortic flow from pressure in humans using a nonlinear, three-element model. J Appl Physiol (1985). 1993;74(5):2566–73.
    • (1993) J Appl Physiol (1985). , vol.74 , Issue.5 , pp. 2566-2573
    • Wesseling, K.H.1
  • 24
    • 42449103530 scopus 로고    scopus 로고
    • Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis
    • PID: 18420863, table of contents
    • Biais M, et al. Cardiac output measurement in patients undergoing liver transplantation: pulmonary artery catheter versus uncalibrated arterial pressure waveform analysis. Anesth Analg. 2008;106(5):1480–6 table of contents. DOI: 10.1213/ane.0b013e318168b309
    • (2008) Anesth Analg. , vol.106 , Issue.5 , pp. 1480-1486
    • Biais, M.1
  • 25
    • 85141002616 scopus 로고    scopus 로고
    • ∙ Meng L, et al. The impact of phenylephrine, ephedrine, and increased preload on third-generation Vigileo-FloTrac and esophageal doppler cardiac output measurements. Anesth Analg. 2011;113(4):751–7. This article very nicely showed how vasopressors adversely affected FloTrac readings.
    • (2011)
  • 26
    • 84896985335 scopus 로고    scopus 로고
    • Systematic review of uncalibrated arterial pressure waveform analysis to determine cardiac output and stroke volume variation
    • COI: 1:STN:280:DC%2BC2czlvVSksQ%3D%3D, PID: 24431387
    • Slagt C, Malagon I, Groeneveld AB. Systematic review of uncalibrated arterial pressure waveform analysis to determine cardiac output and stroke volume variation. Br J Anaesth. 2014;112(4):626–37. DOI: 10.1093/bja/aet429
    • (2014) Br J Anaesth. , vol.112 , Issue.4 , pp. 626-637
    • Slagt, C.1    Malagon, I.2    Groeneveld, A.B.3
  • 27
    • 80052230646 scopus 로고    scopus 로고
    • Evaluation of a new software version of the FloTrac/Vigileo (version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery
    • PID: 21680855
    • Biancofiore G, et al. Evaluation of a new software version of the FloTrac/Vigileo (version 3.02) and a comparison with previous data in cirrhotic patients undergoing liver transplant surgery. Anesth Analg. 2011;113(3):515–22.
    • (2011) Anesth Analg. , vol.113 , Issue.3 , pp. 515-522
    • Biancofiore, G.1
  • 28
    • 84878282847 scopus 로고    scopus 로고
    • Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre study
    • PID: 23558909
    • Scheeren TW, et al. Goal-directed intraoperative fluid therapy guided by stroke volume and its variation in high-risk surgical patients: a prospective randomized multicentre study. J Clin Monit Comput. 2013;27(3):225–33. DOI: 10.1007/s10877-013-9461-6
    • (2013) J Clin Monit Comput. , vol.27 , Issue.3 , pp. 225-233
    • Scheeren, T.W.1
  • 29
    • 0030832772 scopus 로고    scopus 로고
    • Lithium dilution cardiac output measurement: a comparison with thermodilution
    • COI: 1:STN:280:DyaK1c%2FjtlGmtw%3D%3D, PID: 9366760
    • Linton R, et al. Lithium dilution cardiac output measurement: a comparison with thermodilution. Crit Care Med. 1997;25(11):1796–800. DOI: 10.1097/00003246-199711000-00015
    • (1997) Crit Care Med. , vol.25 , Issue.11 , pp. 1796-1800
    • Linton, R.1
  • 30
    • 84885941630 scopus 로고    scopus 로고
    • Reliability of lithium dilution cardiac output in anaesthetized sheep
    • COI: 1:CAS:528:DC%2BC3sXhs1WnsLzL, PID: 23838805
    • Axiak Flammer SM, et al. Reliability of lithium dilution cardiac output in anaesthetized sheep. Br J Anaesth. 2013;111(5):833–9. DOI: 10.1093/bja/aet220
    • (2013) Br J Anaesth. , vol.111 , Issue.5 , pp. 833-839
    • Axiak Flammer, S.M.1
  • 31
    • 78549248199 scopus 로고    scopus 로고
    • Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters
    • PID: 21092290
    • Hadian M, et al. Cross-comparison of cardiac output trending accuracy of LiDCO, PiCCO, FloTrac and pulmonary artery catheters. Crit Care. 2010;14(6):R212. DOI: 10.1186/cc9335
    • (2010) Crit Care. , vol.14 , Issue.6 , pp. R212
    • Hadian, M.1
  • 32
    • 84887019652 scopus 로고    scopus 로고
    • Comparison of stroke volume and fluid responsiveness measurements in commonly used technologies for goal-directed therapy
    • PID: 23965199
    • Davies SJ, et al. Comparison of stroke volume and fluid responsiveness measurements in commonly used technologies for goal-directed therapy. J Clin Anesth. 2013;25(6):466–74. DOI: 10.1016/j.jclinane.2013.04.010
    • (2013) J Clin Anesth. , vol.25 , Issue.6 , pp. 466-474
    • Davies, S.J.1
  • 34
    • 23844520689 scopus 로고    scopus 로고
    • Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states
    • COI: 1:STN:280:DC%2BD2Mzlt1emtw%3D%3D, PID: 15894561
    • Scolletta S, et al. Pressure recording analytical method (PRAM) for measurement of cardiac output during various haemodynamic states. Br J Anaesth. 2005;95(2):159–65. DOI: 10.1093/bja/aei154
    • (2005) Br J Anaesth. , vol.95 , Issue.2 , pp. 159-165
    • Scolletta, S.1
  • 35
    • 79960547791 scopus 로고    scopus 로고
    • Comparison between an uncalibrated pulse contour method and thermodilution technique for cardiac output estimation in septic patients
    • COI: 1:STN:280:DC%2BC3MVagtA%3D%3D, PID: 21665901
    • Franchi F, et al. Comparison between an uncalibrated pulse contour method and thermodilution technique for cardiac output estimation in septic patients. Br J Anaesth. 2011;107(2):202–8. DOI: 10.1093/bja/aer123
    • (2011) Br J Anaesth. , vol.107 , Issue.2 , pp. 202-208
    • Franchi, F.1
  • 36
    • 84901503420 scopus 로고    scopus 로고
    • Validation of cardiac output studies from the Mostcare compared to a pulmonary artery catheter in septic patients
    • COI: 1:STN:280:DC%2BC2czivF2nsQ%3D%3D, PID: 24398442
    • Gopal S, et al. Validation of cardiac output studies from the Mostcare compared to a pulmonary artery catheter in septic patients. Minerva Anestesiol. 2014;80(3):314–23.
    • (2014) Minerva Anestesiol. , vol.80 , Issue.3 , pp. 314-323
    • Gopal, S.1
  • 37
    • 52149102727 scopus 로고    scopus 로고
    • The effects of vasodilation on cardiac output measured by PiCCO
    • PID: 18922424
    • Yamashita K, et al. The effects of vasodilation on cardiac output measured by PiCCO. J Cardiothorac Vasc Anesth. 2008;22(5):688–92. DOI: 10.1053/j.jvca.2008.04.007
    • (2008) J Cardiothorac Vasc Anesth. , vol.22 , Issue.5 , pp. 688-692
    • Yamashita, K.1
  • 38
    • 84887731414 scopus 로고    scopus 로고
    • Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients
    • COI: 1:STN:280:DC%2BC3srovFentw%3D%3D, PID: 23625132
    • Bendjelid K, et al. Performance of a new pulse contour method for continuous cardiac output monitoring: validation in critically ill patients. Br J Anaesth. 2013;111(4):573–9. DOI: 10.1093/bja/aet116
    • (2013) Br J Anaesth. , vol.111 , Issue.4 , pp. 573-579
    • Bendjelid, K.1
  • 40
    • 85047680055 scopus 로고
    • Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre
    • COI: 1:STN:280:DyaL1M7jsF2ntQ%3D%3D, PID: 3224351
    • Imholz BP, et al. Continuous non-invasive blood pressure monitoring: reliability of Finapres device during the Valsalva manoeuvre. Cardiovasc Res. 1988;22(6):390–7. DOI: 10.1093/cvr/22.6.390
    • (1988) Cardiovasc Res. , vol.22 , Issue.6 , pp. 390-397
    • Imholz, B.P.1
  • 41
    • 84861144517 scopus 로고    scopus 로고
    • Comparison of noninvasive cardiac output measurements using the Nexfin monitoring device and the esophageal Doppler
    • PID: 22608581
    • Chen G, et al. Comparison of noninvasive cardiac output measurements using the Nexfin monitoring device and the esophageal Doppler. J Clin Anesth. 2012;24(4):275–83. DOI: 10.1016/j.jclinane.2011.08.014
    • (2012) J Clin Anesth. , vol.24 , Issue.4 , pp. 275-283
    • Chen, G.1
  • 42
    • 84890179458 scopus 로고    scopus 로고
    • Validation of noninvasive pulse contour cardiac output using finger arterial pressure in cardiac surgery patients requiring fluid therapy
    • PID: 24475498
    • Hofhuizen C, et al. Validation of noninvasive pulse contour cardiac output using finger arterial pressure in cardiac surgery patients requiring fluid therapy. J Crit Care. 2014;29(1):161–5. DOI: 10.1016/j.jcrc.2013.09.005
    • (2014) J Crit Care. , vol.29 , Issue.1 , pp. 161-165
    • Hofhuizen, C.1
  • 43
    • 84881370424 scopus 로고    scopus 로고
    • Noninvasive continuous cardiac output by the Nexfin before and after preload-modifying maneuvers: a comparison with intermittent thermodilution cardiac output
    • PID: 23757471
    • Bubenek-Turconi SI, et al. Noninvasive continuous cardiac output by the Nexfin before and after preload-modifying maneuvers: a comparison with intermittent thermodilution cardiac output. Anesth Analg. 2013;117(2):366–72. DOI: 10.1213/ANE.0b013e31829562c3
    • (2013) Anesth Analg. , vol.117 , Issue.2 , pp. 366-372
    • Bubenek-Turconi, S.I.1
  • 44
    • 84867900309 scopus 로고    scopus 로고
    • The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge
    • PID: 23107227
    • Monnet X, et al. The estimation of cardiac output by the Nexfin device is of poor reliability for tracking the effects of a fluid challenge. Crit Care. 2012;16(5):R212. DOI: 10.1186/cc11846
    • (2012) Crit Care. , vol.16 , Issue.5 , pp. R212
    • Monnet, X.1
  • 45
    • 84882236591 scopus 로고    scopus 로고
    • Evaluation of cardiac output in intensive care using a non-invasive arterial pulse contour technique (Nexfin((R))) compared with echocardiography
    • COI: 1:STN:280:DC%2BC3sjpt1CrsA%3D%3D, PID: 23837860
    • Taton O, et al. Evaluation of cardiac output in intensive care using a non-invasive arterial pulse contour technique (Nexfin((R))) compared with echocardiography. Anaesthesia. 2013;68(9):917–23. DOI: 10.1111/anae.12341
    • (2013) Anaesthesia. , vol.68 , Issue.9 , pp. 917-923
    • Taton, O.1
  • 46
    • 34547132055 scopus 로고    scopus 로고
    • Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance
    • COI: 1:CAS:528:DC%2BD2sXovVyhsLg%3D, PID: 17384132
    • Keren H, Burkhoff D, Squara P. Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol. 2007;293(1):H583–9. DOI: 10.1152/ajpheart.00195.2007
    • (2007) Am J Physiol Heart Circ Physiol. , vol.293 , Issue.1 , pp. H583-H589
    • Keren, H.1    Burkhoff, D.2    Squara, P.3
  • 47
    • 41949125848 scopus 로고    scopus 로고
    • Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique
    • PID: 18340540
    • Raval NY, et al. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22(2):113–9. DOI: 10.1007/s10877-008-9112-5
    • (2008) J Clin Monit Comput. , vol.22 , Issue.2 , pp. 113-119
    • Raval, N.Y.1
  • 48
    • 34447131664 scopus 로고    scopus 로고
    • Noninvasive cardiac output monitoring (NICOM): a clinical validation
    • PID: 17458538
    • Squara P, et al. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007;33(7):1191–4. DOI: 10.1007/s00134-007-0640-0
    • (2007) Intensive Care Med. , vol.33 , Issue.7 , pp. 1191-1194
    • Squara, P.1
  • 49
    • 80054847610 scopus 로고    scopus 로고
    • Noninvasive cardiac output monitoring with bioreactance as an alternative to invasive instrumentation for preclinical drug evaluation in beagles
    • COI: 1:CAS:528:DC%2BC3MXhtlKjtb7M, PID: 21440649
    • Heerdt PM, et al. Noninvasive cardiac output monitoring with bioreactance as an alternative to invasive instrumentation for preclinical drug evaluation in beagles. J Pharmacol Toxicol Methods. 2011;64(2):111–8. DOI: 10.1016/j.vascn.2011.03.006
    • (2011) J Pharmacol Toxicol Methods. , vol.64 , Issue.2 , pp. 111-118
    • Heerdt, P.M.1
  • 50
    • 84884473073 scopus 로고    scopus 로고
    • A comparison of noninvasive bioreactance with oesophageal Doppler estimation of stroke volume during open abdominal surgery: an observational study
    • COI: 1:CAS:528:DC%2BC3sXhtVyksLjO, PID: 23549128
    • Conway DH, Hussain OA, Gall I. A comparison of noninvasive bioreactance with oesophageal Doppler estimation of stroke volume during open abdominal surgery: an observational study. Eur J Anaesthesiol. 2013;30(8):501–8. DOI: 10.1097/EJA.0b013e3283603250
    • (2013) Eur J Anaesthesiol. , vol.30 , Issue.8 , pp. 501-508
    • Conway, D.H.1    Hussain, O.A.2    Gall, I.3
  • 51
    • 34347243198 scopus 로고    scopus 로고
    • Comparison of electrical velocimetry and transpulmonary thermodilution for measuring cardiac output in piglets
    • PID: 17596220
    • Osthaus WA, et al. Comparison of electrical velocimetry and transpulmonary thermodilution for measuring cardiac output in piglets. Paediatr Anaesth. 2007;17(8):749–55. DOI: 10.1111/j.1460-9592.2007.02210.x
    • (2007) Paediatr Anaesth. , vol.17 , Issue.8 , pp. 749-755
    • Osthaus, W.A.1
  • 52
    • 20944444823 scopus 로고    scopus 로고
    • A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors
    • PID: 15957621
    • Ishihara H, et al. A new non-invasive continuous cardiac output trend solely utilizing routine cardiovascular monitors. J Clin Monit Comput. 2004;18(5–6):313–20. DOI: 10.1007/s10877-005-2452-5
    • (2004) J Clin Monit Comput. , vol.18 , Issue.5-6 , pp. 313-320
    • Ishihara, H.1
  • 53
    • 84862986509 scopus 로고    scopus 로고
    • Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output
    • PID: 22467885
    • Yamada T, et al. Multicenter study verifying a method of noninvasive continuous cardiac output measurement using pulse wave transit time: a comparison with intermittent bolus thermodilution cardiac output. Anesth Analg. 2012;115(1):82–7. DOI: 10.1213/ANE.0b013e31824e2b6c
    • (2012) Anesth Analg. , vol.115 , Issue.1 , pp. 82-87
    • Yamada, T.1
  • 54
    • 84877106251 scopus 로고    scopus 로고
    • The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm
    • PID: 22854918
    • Ishihara H, et al. The ability of a new continuous cardiac output monitor to measure trends in cardiac output following implementation of a patient information calibration and an automated exclusion algorithm. J Clin Monit Comput. 2012;26(6):465–71. DOI: 10.1007/s10877-012-9384-7
    • (2012) J Clin Monit Comput. , vol.26 , Issue.6 , pp. 465-471
    • Ishihara, H.1
  • 55
    • 84889027773 scopus 로고    scopus 로고
    • Pulse wave transit time measurements of cardiac output in patients undergoing partial hepatectomy: a comparison of the esCCO system with thermodilution
    • PID: 24257379
    • Tsutsui M, et al. Pulse wave transit time measurements of cardiac output in patients undergoing partial hepatectomy: a comparison of the esCCO system with thermodilution. Anesth Analg. 2013;117(6):1307–12. DOI: 10.1213/ANE.0b013e3182a44c87
    • (2013) Anesth Analg. , vol.117 , Issue.6 , pp. 1307-1312
    • Tsutsui, M.1
  • 56
    • 0024394464 scopus 로고
    • Continuous hemodynamic monitoring by esophageal Doppler
    • COI: 1:STN:280:DyaL1M3gvF2hsQ%3D%3D, PID: 2651004
    • Singer M, Clarke J, Bennett ED. Continuous hemodynamic monitoring by esophageal Doppler. Crit Care Med. 1989;17(5):447–52. DOI: 10.1097/00003246-198905000-00014
    • (1989) Crit Care Med. , vol.17 , Issue.5 , pp. 447-452
    • Singer, M.1    Clarke, J.2    Bennett, E.D.3
  • 57
    • 84895074457 scopus 로고    scopus 로고
    • USCOM-window to the circulation: utility of supra-sternal Doppler in an elderly anaesthetized patient for a robotic cystectomy
    • PID: 23907210
    • Critchley LA, Huang L. USCOM-window to the circulation: utility of supra-sternal Doppler in an elderly anaesthetized patient for a robotic cystectomy. J Clin Monit Comput. 2014;28(1):83–93. DOI: 10.1007/s10877-013-9499-5
    • (2014) J Clin Monit Comput. , vol.28 , Issue.1 , pp. 83-93
    • Critchley, L.A.1    Huang, L.2
  • 58
    • 11144335979 scopus 로고    scopus 로고
    • The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults
    • PID: 15368037
    • Dark PM, Singer M. The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults. Intensive Care Med. 2004;30(11):2060–6. DOI: 10.1007/s00134-004-2430-2
    • (2004) Intensive Care Med. , vol.30 , Issue.11 , pp. 2060-2066
    • Dark, P.M.1    Singer, M.2
  • 59
    • 0031871297 scopus 로고    scopus 로고
    • Noninvasive monitoring of cardiac output in critically ill patients using transesophageal Doppler
    • COI: 1:STN:280:DyaK1czisFartw%3D%3D, PID: 9655710
    • Valtier B, et al. Noninvasive monitoring of cardiac output in critically ill patients using transesophageal Doppler. Am J Respir Crit Care Med. 1998;158(1):77–83. DOI: 10.1164/ajrccm.158.1.9707031
    • (1998) Am J Respir Crit Care Med. , vol.158 , Issue.1 , pp. 77-83
    • Valtier, B.1
  • 60
    • 84867205980 scopus 로고    scopus 로고
    • A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM)
    • COI: 1:STN:280:DC%2BC38bhvFerug%3D%3D, PID: 22928650
    • Chong SW, Peyton PJ. A meta-analysis of the accuracy and precision of the ultrasonic cardiac output monitor (USCOM). Anaesthesia. 2012;67(11):1266–71. DOI: 10.1111/j.1365-2044.2012.07311.x
    • (2012) Anaesthesia. , vol.67 , Issue.11 , pp. 1266-1271
    • Chong, S.W.1    Peyton, P.J.2
  • 61
    • 14544292461 scopus 로고    scopus 로고
    • Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs
    • PID: 15728064
    • Critchley LA, et al. Testing the reliability of a new ultrasonic cardiac output monitor, the USCOM, by using aortic flowprobes in anesthetized dogs. Anesth Analg. 2005;100(3):748–53. DOI: 10.1213/01.ANE.0000144774.42408.05
    • (2005) Anesth Analg. , vol.100 , Issue.3 , pp. 748-753
    • Critchley, L.A.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.