Volumn 100, Issue 3, 2011, Pages 446-452

Clinical introduction of a linac head-mounted 2D detector array based quality assurance system in head and neck IMRT

Author keywords

2D ionization chamber array; Dose reconstruction; Film dosimetry; Gamma index analysis; IMRT QA

Indexed keywords

ARTICLE; CANCER RADIOTHERAPY; CLINICAL ARTICLE; CONTROLLED STUDY; DATA BASE; DOSIMETRY; GAMMA RADIATION; HEAD AND NECK CANCER; HUMAN; IMAGING SYSTEM; INTENSITY MODULATED RADIATION THERAPY; LINEAR ACCELERATOR; PHANTOM; PRIORITY JOURNAL; QUALITY CONTROL; RADIATION DETECTOR; VALIDATION PROCESS;

EID: 80054086840     PISSN: 01678140     EISSN: 18790887     Source Type: Journal    
DOI: 10.1016/j.radonc.2011.09.007     Document Type: Article
Times cited : (34)

References (38)
  • 1
    • 0035889335 scopus 로고    scopus 로고
    • Intensity-modulated radiotherapy: Current status and issues of interest
    • A.L. Boyer, E.B. Butler, and T.A. DiPetrillo Intensity-modulated radiotherapy: current status and issues of interest Int J Radiat Oncol Biol Phys 51 2001 880 914
    • (2001) Int J Radiat Oncol Biol Phys , vol.51 , pp. 880-914
    • Boyer, A.L.1    Butler, E.B.2    Dipetrillo, T.A.3
  • 2
    • 10744231470 scopus 로고    scopus 로고
    • Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee
    • DOI 10.1118/1.1591194
    • G.A. Ezzell, J.M. Galvin, and D. Low Guidance document on delivery, treatment planning, and clinical implementation of IMRT: report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee Med Phys 30 2003 2089 2115 (Pubitemid 37022015)
    • (2003) Medical Physics , vol.30 , Issue.8 , pp. 2089-2115
    • Ezzell, G.A.1    Galvin, J.M.2    Low, D.3    Palta, J.R.4    Rosen, I.5    Sharpe, M.B.6    Xia, P.7    Xiao, Y.8    Xing, L.9    Yu, C.X.10
  • 4
    • 0031762329 scopus 로고    scopus 로고
    • Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy
    • DOI 10.1118/1.598381
    • T. Losasso, C.S. Chui, and C.C. Ling Physical and dosimetric aspects of a multileaf collimation system used in the dynamic mode for implementing intensity modulated radiotherapy Med Phys 25 1998 1919 1927 (Pubitemid 28485214)
    • (1998) Medical Physics , vol.25 , Issue.10 , pp. 1919-1927
    • LoSasso, T.1    Chui, C.-S.2    Ling, C.C.3
  • 5
    • 33744899892 scopus 로고    scopus 로고
    • Improved peripheral dose calculation accuracy for a small MLC field brought by the latest commercial treatment planning system
    • DOI 10.1017/S1460396906000161, PII S1460396906000161
    • J.C.L. Chow, G.N. Grigorov, and R. Jiang Improved peripheral dose calculation accuracy for a small MLC field brought by the latest commercial treatment planning system J Radiother Pract 5 2006 121 128 (Pubitemid 43842441)
    • (2006) Journal of Radiotherapy in Practice , vol.5 , Issue.2 , pp. 121-128
    • Chow, J.C.L.1    Grigorov, G.N.2    Jiang, R.3
  • 7
    • 38049001015 scopus 로고    scopus 로고
    • Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer
    • G. Mu, E. Ludlum, and P. Xia Impact of MLC leaf position errors on simple and complex IMRT plans for head and neck cancer Phys Med Biol 53 2008 77 88
    • (2008) Phys Med Biol , vol.53 , pp. 77-88
    • Mu, G.1    Ludlum, E.2    Xia, P.3
  • 8
    • 70049110836 scopus 로고    scopus 로고
    • On the sensitivity of patient-specific IMRT QA to MLC positioning errors
    • G. Yan, C. Liu, T.A. Simon, L.C. Peng, C. Fox, and J.G. Li On the sensitivity of patient-specific IMRT QA to MLC positioning errors J Appl Clin Med Phys 10 2009 2915
    • (2009) J Appl Clin Med Phys , vol.10 , pp. 2915
    • Yan, G.1    Liu, C.2    Simon, T.A.3    Peng, L.C.4    Fox, C.5    Li, J.G.6
  • 10
    • 0034499349 scopus 로고    scopus 로고
    • Monitor unit settings for intensity modulated beams delivered using a step-and-shoot approach
    • DOI 10.1118/1.1328383
    • M.B. Sharpe, B.M. Miller, D. Yan, and J.W. Wong Monitor unit settings for intensity modulated beams delivered using a step-and-shoot approach Med Phys 27 2000 2719 2725 (Pubitemid 32063451)
    • (2000) Medical Physics , vol.27 , Issue.12 , pp. 2719-2725
    • Sharpe, M.B.1    Miller, B.M.2    Yan, D.3    Wong, J.W.4
  • 11
    • 54949136185 scopus 로고    scopus 로고
    • Dosimetric characteristics of linear accelerator photon beams with small monitor unit settings
    • S.K. Kang, K.H. Cheong, and T. Hwang Dosimetric characteristics of linear accelerator photon beams with small monitor unit settings Med Phys 35 2008 5172 5178
    • (2008) Med Phys , vol.35 , pp. 5172-5178
    • Kang, S.K.1    Cheong, K.H.2    Hwang, T.3
  • 14
    • 46149099592 scopus 로고    scopus 로고
    • Radiochromic film dosimetry with flatbed scanners: A fast and accurate method for dose calibration and uniformity correction with single film exposure
    • DOI 10.1118/1.2936334
    • L. Menegotti, A. Delana, and A. Martignano Radiochromic film dosimetry with flatbed scanners: a fast and accurate method for dose calibration and uniformity correction with single film exposure Med Phys 35 2008 3078 3085 (Pubitemid 351904511)
    • (2008) Medical Physics , vol.35 , Issue.7 , pp. 3078-3085
    • Menegotti, L.1    Delana, A.2    Martignano, A.3
  • 15
    • 38849173037 scopus 로고    scopus 로고
    • Accurate dosimetry with GafChromic™ EBT film of a 6 MV photon beam in water: What level is achievable?
    • DOI 10.1118/1.2828196
    • L.J. van Battum, D. Hoffmans, H. Piersma, and S. Heukelom Accurate dosimetry with GafChromic EBT film of a 6 MV photon beam in water: what level is achievable? Med Phys 35 2008 704 716 (Pubitemid 351190890)
    • (2008) Medical Physics , vol.35 , Issue.2 , pp. 704-716
    • Van Battum, L.J.1    Hoffmans, D.2    Piersma, H.3    Heukelom, S.4
  • 16
    • 1542603236 scopus 로고    scopus 로고
    • Evaluation of a 2D diode array for IMRT quality assurance
    • DOI 10.1016/j.radonc.2003.10.014, PII S0167814003004171
    • D. Letourneau, M. Gulam, D. Yan, M. Oldham, and J.W. Wong Evaluation of a 2D diode array for IMRT quality assurance Radiother Oncol 70 2004 199 206 (Pubitemid 38353020)
    • (2004) Radiotherapy and Oncology , vol.70 , Issue.2 , pp. 199-206
    • Letourneau, D.1    Gulam, M.2    Yan, D.3    Oldham, M.4    Wong, J.W.5
  • 17
    • 25444490701 scopus 로고    scopus 로고
    • D-IMRT verification with a 2D pixel ionization chamber: Dosimetric and clinical results in head and neck cancer
    • M. Stasi, S. Giordanengo, and R. Cirio D-IMRT verification with a 2D pixel ionization chamber: dosimetric and clinical results in head and neck cancer Phys Med Biol 50 2005 4681 4694
    • (2005) Phys Med Biol , vol.50 , pp. 4681-4694
    • Stasi, M.1    Giordanengo, S.2    Cirio, R.3
  • 19
    • 59249091676 scopus 로고    scopus 로고
    • The DOSIMAP, a high spatial resolution tissue equivalent 2D dosimeter for LINAC QA and IMRT verification
    • V. Collomb-Patton, P. Boher, T. Leroux, J.M. Fontbonne, A. Vela, and A. Batalla The DOSIMAP, a high spatial resolution tissue equivalent 2D dosimeter for LINAC QA and IMRT verification Med Phys 36 2009 317 328
    • (2009) Med Phys , vol.36 , pp. 317-328
    • Collomb-Patton, V.1    Boher, P.2    Leroux, T.3    Fontbonne, J.M.4    Vela, A.5    Batalla, A.6
  • 21
  • 22
    • 77954277352 scopus 로고    scopus 로고
    • Clinical performance of a transmission detector array for the permanent supervision of IMRT deliveries
    • B. Poppe, H.K. Looe, N. Chofor, A. Ruhmann, D. Harder, and K.C. Willborn Clinical performance of a transmission detector array for the permanent supervision of IMRT deliveries Radiother Oncol 95 2010 158 165
    • (2010) Radiother Oncol , vol.95 , pp. 158-165
    • Poppe, B.1    Looe, H.K.2    Chofor, N.3    Ruhmann, A.4    Harder, D.5    Willborn, K.C.6
  • 23
    • 38149131763 scopus 로고    scopus 로고
    • The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations
    • W. van Elmpt, S. Nijsten, B. Mijnheer, A. Dekker, and P. Lambin The next step in patient-specific QA: 3D dose verification of conformal and intensity-modulated RT based on EPID dosimetry and Monte Carlo dose calculations Radiother Oncol 86 2008 86 92
    • (2008) Radiother Oncol , vol.86 , pp. 86-92
    • Van Elmpt, W.1    Nijsten, S.2    Mijnheer, B.3    Dekker, A.4    Lambin, P.5
  • 24
    • 0031854569 scopus 로고    scopus 로고
    • Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation
    • DOI 10.1088/0031-9155/43/8/004, PII S003191559889386X
    • K.L. Pasma, M. Kroonwijk, J.C. de Boer, A.G. Visser, and B.J. Heijmen Accurate portal dose measurement with a fluoroscopic electronic portal imaging device (EPID) for open and wedged beams and dynamic multileaf collimation Phys Med Biol 43 1998 2047 2060 (Pubitemid 28372607)
    • (1998) Physics in Medicine and Biology , vol.43 , Issue.8 , pp. 2047-2060
    • Pasma, K.L.1    Kroonwijk, M.2    De Boer, J.C.J.3    Visser, A.G.4    Heijmen, B.J.M.5
  • 26
    • 33847018108 scopus 로고    scopus 로고
    • 3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID
    • DOI 10.1016/j.radonc.2006.12.010, PII S0167814007000126
    • M. van Zijtveld, M.L. Dirkx, H.C. de Boer, and B.J. Heijmen 3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID Radiother Oncol 82 2007 201 207 (Pubitemid 46275137)
    • (2007) Radiotherapy and Oncology , vol.82 , Issue.2 , pp. 201-207
    • Van Zijtveld, M.1    Dirkx, M.L.P.2    De Boer, H.C.J.3    Heijmen, B.J.M.4
  • 27
    • 77955715285 scopus 로고    scopus 로고
    • Evaluation of the 'dose of the day' for IMRT prostate cancer patients derived from portal dose measurements and cone-beam CT
    • M. van Zijtveld, M. Dirkx, M. Breuers, R. Kuipers, and B. Heijmen Evaluation of the 'dose of the day' for IMRT prostate cancer patients derived from portal dose measurements and cone-beam CT Radiother Oncol 96 2010 172 177
    • (2010) Radiother Oncol , vol.96 , pp. 172-177
    • Van Zijtveld, M.1    Dirkx, M.2    Breuers, M.3    Kuipers, R.4    Heijmen, B.5
  • 29
    • 79551672470 scopus 로고    scopus 로고
    • Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors
    • B.E. Nelms, H. Zhen, and W.A. Tome Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors Med Phys 38 2011 1037 1044
    • (2011) Med Phys , vol.38 , pp. 1037-1044
    • Nelms, B.E.1    Zhen, H.2    Tome, W.A.3
  • 30
    • 0345256650 scopus 로고    scopus 로고
    • A dose delivery verification method for conventional and intensity-modulated radiation therapy using measured field fluence distributions
    • DOI 10.1118/1.1610771
    • W.D. Renner, M. Sarfaraz, M.A. Earl, and C.X. Yu A dose delivery verification method for conventional and intensity-modulated radiation therapy using measured field fluence distributions Med Phys 30 2003 2996 3005 (Pubitemid 37444310)
    • (2003) Medical Physics , vol.30 , Issue.11 , pp. 2996-3005
    • Renner, W.D.1    Sarfaraz, M.2    Earl, M.A.3    Yu, C.X.4
  • 31
    • 0021907462 scopus 로고
    • A convolution method of calculating dose for 15-MV x rays
    • DOI 10.1118/1.595774
    • T.R. Mackie, J.W. Scrimger, and J.J. Battista A convolution method of calculating dose for 15-MV X-rays Med Phys 12 1985 188 196 (Pubitemid 15089002)
    • (1985) Medical Physics , vol.12 , Issue.2 , pp. 188-196
    • Mackie, T.R.1    Scrimger, J.W.2    Battista, J.J.3
  • 32
    • 0024424112 scopus 로고
    • Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media
    • DOI 10.1118/1.596360
    • A. Ahnesjo Collapsed cone convolution of radiant energy for photon dose calculation in heterogeneous media Med Phys 16 1989 577 592 (Pubitemid 19224308)
    • (1989) Medical Physics , vol.16 , Issue.4 , pp. 577-592
    • Ahnesjo, A.1
  • 33
    • 0037340926 scopus 로고    scopus 로고
    • A virtual photon energy fluence model for Monte Carlo dose calculation
    • DOI 10.1118/1.1543152
    • M. Fippel, F. Haryanto, O. Dohm, F. Nusslin, and S. Kriesen A virtual photon energy fluence model for Monte Carlo dose calculation Med Phys 30 2003 301 311 (Pubitemid 36337095)
    • (2003) Medical Physics , vol.30 , Issue.3 , pp. 301-311
    • Fippel, M.1    Haryanto, F.2    Dohm, O.3    Nusslin, F.4    Kriesen, S.5
  • 35
    • 14544300121 scopus 로고    scopus 로고
    • Interpretation and evaluation of the γ index and the γ index angle for the verification of IMRT hybrid plans
    • DOI 10.1088/0031-9155/50/3/001
    • M. Stock, B. Kroupa, and D. Georg Interpretation and evaluation of the gamma index and the gamma index angle for the verification of IMRT hybrid plans Phys Med Biol 50 2005 399 411 (Pubitemid 40297915)
    • (2005) Physics in Medicine and Biology , vol.50 , Issue.3 , pp. 399-411
    • Stock, M.1    Kroupa, B.2    Georg, D.3
  • 36
    • 0035192124 scopus 로고    scopus 로고
    • Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode
    • DOI 10.1118/1.1410123
    • T. Losasso, C.S. Chui, and C.C. Ling Comprehensive quality assurance for the delivery of intensity modulated radiotherapy with a multileaf collimator used in the dynamic mode Med Phys 28 2001 2209 2219 (Pubitemid 33095551)
    • (2001) Medical Physics , vol.28 , Issue.11 , pp. 2209-2219
    • LoSasso, T.1    Chui, C.-S.2    Ling, C.C.3
  • 37
    • 34248226518 scopus 로고    scopus 로고
    • Spatial resolution of 2D ionization chamber arrays for IMRT dose verification: Single-detector size and sampling step width
    • B. Poppe, A. Djouguela, A. Blechschmidt, K. Willborn, A. Ruhmann, and D. Harder Spatial resolution of 2D ionization chamber arrays for IMRT dose verification: single-detector size and sampling step width Phys Med Biol 52 2007 2921 2935
    • (2007) Phys Med Biol , vol.52 , pp. 2921-2935
    • Poppe, B.1    Djouguela, A.2    Blechschmidt, A.3    Willborn, K.4    Ruhmann, A.5    Harder, D.6
  • 38
    • 79961087515 scopus 로고    scopus 로고
    • Reconstruction of high-resolution 3D dose from matrix measurements: Error detection capability of the COMPASS correction kernel method
    • J. Godart, E.W. Korevaar, R. Visser, D.J. Wauben, and A.A. van't Veld Reconstruction of high-resolution 3D dose from matrix measurements: error detection capability of the COMPASS correction kernel method Phys Med Biol 56 2011 5029 5043
    • (2011) Phys Med Biol , vol.56 , pp. 5029-5043
    • Godart, J.1    Korevaar, E.W.2    Visser, R.3    Wauben, D.J.4    Van'T Veld, A.A.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.