Volumn 17, Issue 9, 2015, Pages 625-634

Accuracy Evaluation of Four Blood Glucose Monitoring Systems in Unaltered Blood Samples in the Low Glycemic Range and Blood Samples in the Concentration Range Defined by ISO 15197

Author keywords

[No Author keywords available]

Indexed keywords

GLUCOSE; GLUCOSE OXIDASE; GLUCOSE BLOOD LEVEL;

EID: 84939426959     PISSN: 15209156     EISSN: 15578593     Source Type: Journal    
DOI: 10.1089/dia.2015.0043     Document Type: Article
Times cited : (34)

References (37)
  • 1
    • 84892649479 scopus 로고    scopus 로고
    • Standards of medical care in diabetes-2014
    • American Diabetes Association: Standards of medical care in diabetes-2014. Diabetes Care 2014;37(Suppl 1):S14-S80.
    • (2014) Diabetes Care , vol.37 , pp. S14-S80
    • American Diabetes Association1
  • 2
    • 57349104419 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose (SMBG) in insulin- and non-insulin-using adults with diabetes: Consensus recommendations for improving SMBG accuracy, utilization, and research
    • Hirsch IB, Bode BW, Childs BP, Close KL, Fisher WA, Gavin JR, Ginsberg BH, Raine CH, Verderese CA: Self-monitoring of blood glucose (SMBG) in insulin- and non-insulin-using adults with diabetes: consensus recommendations for improving SMBG accuracy, utilization, and research. Diabetes Technol Ther 2008;10:419-439.
    • (2008) Diabetes Technol Ther , vol.10 , pp. 419-439
    • Hirsch, I.B.1    Bode, B.W.2    Childs, B.P.3    Close, K.L.4    Fisher, W.A.5    Gavin, J.R.6    Ginsberg, B.H.7    Raine, C.H.8    Verderese, C.A.9
  • 3
    • 0034939686 scopus 로고    scopus 로고
    • Self-monitoring of blood glucose levels and glycemic control: The Northern California Kaiser Permanente Diabetes Registry
    • Karter AJ, Ackerson LM, Darbinian JA, D'Agostino RB Jr, Ferrara A, Liu J, Selby JV: Self-monitoring of blood glucose levels and glycemic control: the Northern California Kaiser Permanente Diabetes Registry. Am J Med 2001;111: 1-9.
    • (2001) Am J Med , vol.111 , pp. 1-9
    • Karter, A.J.1    Ackerson, L.M.2    Darbinian, J.A.3    D'Agostino, R.B.4    Ferrara, A.5    Liu, J.6    Selby, J.V.7
  • 4
    • 0027370108 scopus 로고
    • The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus
    • The Diabetes Control and Complications Trial Research Group: The effect of intensive treatment of diabetes on the development and progression of long-term complications in insulin-dependent diabetes mellitus. N Engl J Med 1993; 329:977-986.
    • (1993) N Engl J Med , vol.329 , pp. 977-986
  • 5
    • 84891446217 scopus 로고    scopus 로고
    • Clinical implications and economic impact of accuracy differences among commercially available blood glucose monitoring systems
    • Budiman ES, Samant N, Resch A: Clinical implications and economic impact of accuracy differences among commercially available blood glucose monitoring systems. J Diabetes Sci Technol 2013;7:365-380.
    • (2013) J Diabetes Sci Technol , vol.7 , pp. 365-380
    • Budiman, E.S.1    Samant, N.2    Resch, A.3
  • 6
    • 84858133080 scopus 로고    scopus 로고
    • Glucose meter inaccuracy and the impact on the care of patients
    • Hellman R: Glucose meter inaccuracy and the impact on the care of patients. Diabetes Metab Res Rev 2012;28:207-209.
    • (2012) Diabetes Metab Res Rev , vol.28 , pp. 207-209
    • Hellman, R.1
  • 9
    • 84930749375 scopus 로고    scopus 로고
    • Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control
    • DeSalvo DJ, Shanmugham S, Ly TT, Wilson DM, Buckingham BA: Accuracy evaluation of blood glucose monitoring systems in children on overnight closed-loop control. J Diabetes Sci Technol 2014;8:969-973.
    • (2014) J Diabetes Sci Technol , vol.8 , pp. 969-973
    • DeSalvo, D.J.1    Shanmugham, S.2    Ly, T.T.3    Wilson, D.M.4    Buckingham, B.A.5
  • 10
    • 84920860821 scopus 로고    scopus 로고
    • Accuracy evaluation of Contour Next compared with five blood glucose monitoring systems across a wide range of blood glucose concentrations occurring in a clinical research setting
    • Klaff LJ, Brazg R, Hughes K, Tideman AM, Schachner HC, Stenger P, Pardo S, Dunne N, Parkes JL: Accuracy evaluation of Contour Next compared with five blood glucose monitoring systems across a wide range of blood glucose concentrations occurring in a clinical research setting. Diabetes Technol Ther 2015;17:8-15.
    • (2015) Diabetes Technol Ther , vol.17 , pp. 8-15
    • Klaff, L.J.1    Brazg, R.2    Hughes, K.3    Tideman, A.M.4    Schachner, H.C.5    Stenger, P.6    Pardo, S.7    Dunne, N.8    Parkes, J.L.9
  • 13
    • 84859313151 scopus 로고    scopus 로고
    • Accuracy evaluation of five blood glucose monitoring systems obtained from the pharmacy: A European multicenter study with 453 subjects
    • Tack C, Pohlmeier H, Behnke T, Schmid V, Grenningloh M, Forst T, Pfützner A: Accuracy evaluation of five blood glucose monitoring systems obtained from the pharmacy: a European multicenter study with 453 subjects. Diabetes Technol Ther 2012;14:1-8.
    • (2012) Diabetes Technol Ther , vol.14 , pp. 1-8
    • Tack, C.1    Pohlmeier, H.2    Behnke, T.3    Schmid, V.4    Grenningloh, M.5    Forst, T.6    Pfützner, A.7
  • 14
    • 84864535171 scopus 로고    scopus 로고
    • Assessment of three frequently used blood glucose monitoring devices in clinical routine
    • Zueger T, Schuler V, Stettler C, Diem P, Christ E: Assessment of three frequently used blood glucose monitoring devices in clinical routine. Swiss Med Wkly 2012;142: w13631.
    • (2012) Swiss Med Wkly , vol.142
    • Zueger, T.1    Schuler, V.2    Stettler, C.3    Diem, P.4    Christ, E.5
  • 16
    • 84905901716 scopus 로고    scopus 로고
    • A comparative effectiveness analysis of three continuous glucose monitors: The Navigator, G4 Platinum, and Enlite
    • Damiano ER, McKeon K, El-Khatib FH, Zheng H, Nathan DM, Russell SJ: A comparative effectiveness analysis of three continuous glucose monitors: the Navigator, G4 Platinum, and Enlite. J Diabetes Sci Technol 2014;8:699-708.
    • (2014) J Diabetes Sci Technol , vol.8 , pp. 699-708
    • Damiano, E.R.1    McKeon, K.2    El-Khatib, F.H.3    Zheng, H.4    Nathan, D.M.5    Russell, S.J.6
  • 17
    • 84889635916 scopus 로고    scopus 로고
    • Performance evaluation of three continuous glucose monitoring systems: Comparison of six sensors per subject in parallel
    • Freckmann G, Pleus S, Link M, Zschornack E, Klotzer HM, Haug C: Performance evaluation of three continuous glucose monitoring systems: comparison of six sensors per subject in parallel. J Diabetes Sci Technol 2013;7:842-853.
    • (2013) J Diabetes Sci Technol , vol.7 , pp. 842-853
    • Freckmann, G.1    Pleus, S.2    Link, M.3    Zschornack, E.4    Klotzer, H.M.5    Haug, C.6
  • 18
    • 84898461270 scopus 로고    scopus 로고
    • Continuous glucose monitoring in the subcutaneous tissue over a 14- day sensor wear period
    • Hoss U, Budiman ES, Liu H, Christiansen MP: Continuous glucose monitoring in the subcutaneous tissue over a 14- day sensor wear period. J Diabetes Sci Technol 2013;7: 1210-1219.
    • (2013) J Diabetes Sci Technol , vol.7 , pp. 1210-1219
    • Hoss, U.1    Budiman, E.S.2    Liu, H.3    Christiansen, M.P.4
  • 19
    • 84861879435 scopus 로고    scopus 로고
    • Hypoglycemia, diabetes, and cardiovascular disease
    • Snell-Bergeon JK, Wadwa RP: Hypoglycemia, diabetes, and cardiovascular disease. Diabetes Technol Ther 2012; 14(Suppl 1):S-51-S-58.
    • (2012) Diabetes Technol Ther , vol.14 , pp. S51-S58
    • Snell-Bergeon, J.K.1    Wadwa, R.P.2
  • 21
    • 80053569972 scopus 로고    scopus 로고
    • Acute hypoglycemia decreases myocardial blood flow reserve in patients with type 1 diabetes mellitus and in healthy humans
    • Rana O, Byrne CD, Kerr D, Coppini DV, Zouwail S, Senior R, Begley J, Walker JJ, Greaves K: Acute hypoglycemia decreases myocardial blood flow reserve in patients with type 1 diabetes mellitus and in healthy humans. Circulation 2011;124:1548-1556.
    • (2011) Circulation , vol.124 , pp. 1548-1556
    • Rana, O.1    Byrne, C.D.2    Kerr, D.3    Coppini, D.V.4    Zouwail, S.5    Senior, R.6    Begley, J.7    Walker, J.J.8    Greaves, K.9
  • 22
    • 77957933451 scopus 로고    scopus 로고
    • Effects of acute insulin-induced hypoglycemia on indices of inflammation: Putative mechanism for aggravating vascular disease in diabetes
    • Wright RJ, Newby DE, Stirling D, Ludlam CA, Macdonald IA, Frier BM: Effects of acute insulin-induced hypoglycemia on indices of inflammation: putative mechanism for aggravating vascular disease in diabetes. Diabetes Care 2010;33:1591-1597.
    • (2010) Diabetes Care , vol.33 , pp. 1591-1597
    • Wright, R.J.1    Newby, D.E.2    Stirling, D.3    Ludlam, C.A.4    Macdonald, I.A.5    Frier, B.M.6
  • 26
    • 58549091951 scopus 로고    scopus 로고
    • Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen
    • Bundesärztekammer: Richtlinie der Bundesärztekammer zur Qualitätssicherung laboratoriumsmedizinischer Untersuchungen. Deutsch Ärzteblatt 2008;105:341-355.
    • (2008) Deutsch Ärzteblatt , vol.105 , pp. 341-355
    • Bundesärztekammer1
  • 30
    • 84884557685 scopus 로고    scopus 로고
    • System accuracy of blood glucose monitoring systems: Impact of use by patients and ambient conditions
    • Schmid C, Haug C, Heinemann L, Freckmann G: System accuracy of blood glucose monitoring systems: impact of use by patients and ambient conditions. Diabetes Technol Ther 2013;15:889-896.
    • (2013) Diabetes Technol Ther , vol.15 , pp. 889-896
    • Schmid, C.1    Haug, C.2    Heinemann, L.3    Freckmann, G.4
  • 32
    • 79956160504 scopus 로고    scopus 로고
    • Glucose monitoring after fruit peeling: Pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: Wash your hands with tap water before you check blood glucose level
    • Hirose T, Mita T, Fujitani Y, Kawamori R, Watada H: Glucose monitoring after fruit peeling: pseudohyperglycemia when neglecting hand washing before fingertip blood sampling: wash your hands with tap water before you check blood glucose level. Diabetes Care 2011; 34:596-597.
    • (2011) Diabetes Care , vol.34 , pp. 596-597
    • Hirose, T.1    Mita, T.2    Fujitani, Y.3    Kawamori, R.4    Watada, H.5
  • 33
    • 70350209515 scopus 로고    scopus 로고
    • Factors affecting blood glucose monitoring: Sources of errors in measurement
    • Ginsberg BH: Factors affecting blood glucose monitoring: sources of errors in measurement. J Diabetes Sci Technol 2009;3:903-913.
    • (2009) J Diabetes Sci Technol , vol.3 , pp. 903-913
    • Ginsberg, B.H.1
  • 34
    • 57349106743 scopus 로고    scopus 로고
    • Standardized evaluation of nine instruments for self-monitoring of blood glucose
    • Kristensen GB, Monsen G, Skeie S, Sandberg S: Standardized evaluation of nine instruments for self-monitoring of blood glucose. Diabetes Technol Ther 2008;10:467-477.
    • (2008) Diabetes Technol Ther , vol.10 , pp. 467-477
    • Kristensen, G.B.1    Monsen, G.2    Skeie, S.3    Sandberg, S.4
  • 35
    • 85016650386 scopus 로고    scopus 로고
    • System accuracy evaluation of four systems for self-monitoring of blood glucose following ISO 15197 using a glucose oxidase and a hexokinase-based comparison method
    • April 14 [Epub ahead of print]. pii: 1932296815580161
    • Link M, Schmid C, Pleus S, Baumstark A, Rittmeyer D, Haug C, Freckmann G: System accuracy evaluation of four systems for self-monitoring of blood glucose following ISO 15197 using a glucose oxidase and a hexokinase-based comparison method. J Diabetes Sci Technol 2015 April 14 [Epub ahead of print]. pii: 1932296815580161.
    • (2015) J Diabetes Sci Technol
    • Link, M.1    Schmid, C.2    Pleus, S.3    Baumstark, A.4    Rittmeyer, D.5    Haug, C.6    Freckmann, G.7
  • 36
    • 84875803673 scopus 로고    scopus 로고
    • Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197
    • Baumstark A, Pleus S, Schmid C, Link M, Haug C, Freckmann G: Lot-to-lot variability of test strips and accuracy assessment of systems for self-monitoring of blood glucose according to ISO 15197. J Diabetes Sci Technol 2012;6:1076-1086.
    • (2012) J Diabetes Sci Technol , vol.6 , pp. 1076-1086
    • Baumstark, A.1    Pleus, S.2    Schmid, C.3    Link, M.4    Haug, C.5    Freckmann, G.6
  • 37
    • 24144435337 scopus 로고    scopus 로고
    • Between-lot variation in external quality assessment of glucose: Clinical importance and effect on participant performance evaluation
    • Kristensen GB, Christensen NG, Thue G, Sandberg S: Between-lot variation in external quality assessment of glucose: clinical importance and effect on participant performance evaluation. Clin Chem 2005;51:1632-1636.
    • (2005) Clin Chem , vol.51 , pp. 1632-1636
    • Kristensen, G.B.1    Christensen, N.G.2    Thue, G.3    Sandberg, S.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.