Volumn 57, Issue 22, 2012, Pages 7629-7650

Measurements of system sharpness for two digital breast tomosynthesis systems

Author keywords

[No Author keywords available]

Indexed keywords

CALCULATED VALUES; DETECTOR RESOLUTION; DIGITAL BREAST TOMOSYNTHESIS; FULL WIDTH HALF MAXIMUM; IMAGING PERFORMANCE; MULTIPLE PINHOLES; POINT-SPREAD FUNCTIONS; PROJECTION DATA; PROJECTION IMAGE; QUALITY ASSURANCE PROTOCOLS; SIEMENS; TEST OBJECT; TUNGSTEN WIRES;

EID: 84869068599     PISSN: 00319155     EISSN: 13616560     Source Type: Journal    
DOI: 10.1088/0031-9155/57/22/7629     Document Type: Article
Times cited : (58)

References (47)
  • 1
    • 79955759140 scopus 로고    scopus 로고
    • Investigating the potential for super-resolution in digital breast tomosynthesis
    • Acciavatti R J and Maidment A D A 2011 Investigating the potential for super-resolution in digital breast tomosynthesis Proc SPIE 7691 79615K-1-12
    • (2011) Proc SPIE , vol.7691
    • Acciavatti, R.J.1    Maidment, A.D.A.2
  • 3
    • 56349132626 scopus 로고    scopus 로고
    • Breast tomosynthesis and digital mammography: A comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings
    • 10.1007/s00330-008-1076-9 0938-7994
    • Andersson I, Ikeda D M, Zackrisson S, Ruschin M, Svahn T, Timberg P and Tinberg A 2008 Breast tomosynthesis and digital mammography: a comparison of breast cancer visibility and BIRADS classification in a population of cancers with subtle mammographic findings Eur. Radiol. 18 2817-25
    • (2008) Eur. Radiol. , vol.18 , Issue.12 , pp. 2817-2825
    • Andersson, I.1    Ikeda, D.M.2    Zackrisson, S.3    Ruschin, M.4    Svahn, T.5    Timberg, P.6    Tinberg, A.7
  • 4
    • 0002050483 scopus 로고    scopus 로고
    • Applied linear-systems theory
    • Cunningham I A 2000 Applied linear-systems theory Physics and Psychophysics vol 1 ed J Beutel, H L Kundel and R L Van Metter (Bellingham, WA: SPIE) pp 79-159
    • (2000) Physics and Psychophysics , vol.1 , pp. 79-159
    • Cunningham, I.A.1
  • 5
    • 35148817140 scopus 로고    scopus 로고
    • Spatial resolution of x-ray tomosynthesis in relation to computed tomography for coronal/sagittal images of the knee
    • Flynn M J, McGee R and Blechinger J 2007 Spatial resolution of x-ray tomosynthesis in relation to computed tomography for coronal/sagittal images of the knee Proc. SPIE 65100D-1-9
    • (2007) Proc. SPIE
    • Flynn, M.J.1    McGee Blechinger R, J.2
  • 6
    • 77952363141 scopus 로고    scopus 로고
    • Anatomical background and generalized detectability in tomosynthesis and cone-beam CT
    • 10.1118/1.3352586 0094-2405
    • Gang G J, Tward D J, Lee J and Siewerdsen J H 2010 Anatomical background and generalized detectability in tomosynthesis and cone-beam CT Med. Phys. 37 1948-65
    • (2010) Med. Phys. , vol.37 , Issue.5 , pp. 1948-1965
    • Gang, G.J.1    Tward, D.J.2    Lee, J.3    Siewerdsen, J.H.4
  • 7
    • 77954818863 scopus 로고    scopus 로고
    • Digital breast tomosynthesis versus digital mammography: A clinical performance study
    • 10.1007/s00330-009-1699-5 0938-7994
    • Gennaro G et al 2010 Digital breast tomosynthesis versus digital mammography: a clinical performance study Eur. Radiol. 20 1545-53
    • (2010) Eur. Radiol. , vol.20 , Issue.7 , pp. 1545-1553
    • Gennaro, G.1
  • 9
    • 56749186140 scopus 로고    scopus 로고
    • Image artifacts in digital breast tomosynthesis: Investigation of the effects of system geometry and reconstruction parameters using a linear system approach
    • 10.1118/1.2996110 0094-2405
    • Hu Y-H, Zhao B and Zhao W 2008 Image artifacts in digital breast tomosynthesis: investigation of the effects of system geometry and reconstruction parameters using a linear system approach Med. Phys. 35 5242-52
    • (2008) Med. Phys. , vol.35 , Issue.12 , pp. 5242-5252
    • Hu, Y.-H.1    Zhao, B.2    Zhao, W.3
  • 10
    • 79955767400 scopus 로고    scopus 로고
    • The effect of angular dose distribution on the detection of microcalcifications in digital breast tomosynthesis
    • 10.1118/1.3570580 0094-2405
    • Hu Y-H and Zhao W 2011 The effect of angular dose distribution on the detection of microcalcifications in digital breast tomosynthesis Med. Phys. 38 2455-66
    • (2011) Med. Phys. , vol.38 , Issue.5 , pp. 2455-2466
    • Hu, Y.-H.1    Zhao, W.2
  • 12
    • 84869048318 scopus 로고    scopus 로고
    • Medical electricalequipment - Characteristics of digital x-ray image devices: Part 1. Determination of the detective quantum efficiency
    • IEC 2005 Medical electricalequipment - characteristics of digital x-ray image devices: Part 1. Determination of the detective quantum efficiency International Standard IEC 62220-1-2 (Geneva, Switzerland: International Electrotechnical Commission)
    • (2005) International Standard IEC 62220-1-2
  • 16
    • 84869035567 scopus 로고    scopus 로고
    • Towards an international consensus strategy for periodic quality control of digital breast tomosynthesis systems
    • Jacobs J, Marshall N W, Cockmartin L, Zanca F, van Engen R, Young K, Bosmans H and Samei E 2010 Towards an international consensus strategy for periodic quality control of digital breast tomosynthesis systems Proc. SPIE 7622 7622-15
    • (2010) Proc. SPIE , vol.7622 , pp. 7622-7615
    • Jacobs, J.1    Marshall, N.W.2    Cockmartin, L.3    Zanca, F.4    Van Engen, R.5    Young, K.6    Bosmans, H.7    Samei, E.8
  • 18
    • 80855139823 scopus 로고    scopus 로고
    • Further development of a resolution modification routine for the simulation of the modulation transfer function of digital x-ray detectors
    • 10.1118/1.3644845 0094-2405
    • Konstantinidis A C, Olivo A and Speller R D 2011 Further development of a resolution modification routine for the simulation of the modulation transfer function of digital x-ray detectors Med. Phys. 38 5916-20
    • (2011) Med. Phys. , vol.38 , Issue.11 , pp. 5916-5920
    • Konstantinidis, A.C.1    Olivo, A.2    Speller, R.D.3
  • 19
    • 66849126255 scopus 로고    scopus 로고
    • The effect of lag on image quality for a digital breast tomosynthesis system
    • Mainprize J G, Wang X and Jaffe M J 2009 The effect of lag on image quality for a digital breast tomosynthesis system Proc. SPIE 7258 72580R1-12
    • (2009) Proc. SPIE , vol.7258
    • Mainprize, J.G.1    Wang, X.2    Jaffe, M.J.3
  • 20
    • 33646485572 scopus 로고    scopus 로고
    • A comparison between objective and subjective image quality measurements for a full field digital mammography system
    • 10.1088/0031-9155/51/10/006 0031-9155
    • Marshall N W 2006a A comparison between objective and subjective image quality measurements for a full field digital mammography system Phys. Med. Biol. 51 2441-63
    • (2006) Phys. Med. Biol. , vol.51 , Issue.10 , pp. 2441-2463
    • Marshall, N.W.1
  • 21
    • 33751172327 scopus 로고    scopus 로고
    • Retrospective analysis of a detector fault for a full field digital mammography system
    • DOI 10.1088/0031-9155/51/21/018, PII S0031915506259219, 018
    • Marshall N W 2006b Retrospective analysis of a detector fault for a full field digital mammography system Phys. Med. Biol. 51 5655-73 (Pubitemid 44773338)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.21 , pp. 5655-5673
    • Marshall, N.W.1
  • 22
    • 35848968890 scopus 로고    scopus 로고
    • Early experience in the use of quantitative image quality measurements for the quality assurance of full field digital mammography x-ray systems
    • DOI 10.1088/0031-9155/52/18/006, PII S003191550749251X
    • Marshall N W 2007 Early experience in the use of quantitative image quality measurements for the quality assurance of full field digital mammography x-ray systems Phys. Med. Biol. 52 5545-68 (Pubitemid 350056683)
    • (2007) Physics in Medicine and Biology , vol.52 , Issue.18 , pp. 5545-5568
    • Marshall, N.W.1
  • 23
    • 77954654610 scopus 로고    scopus 로고
    • Technical evaluation of a digital breast tomosynthesis system
    • 10.1007/978-3-642-13666-5-47 0302-9743
    • Marshall N W, Jacobs J, Cockmartin L and Bosmans H 2010 Technical evaluation of a digital breast tomosynthesis system Digital Mammography (IWDM 2010) (Lecture Notes in Computer Science vol 6136) ed J Martí (Berlin: Springer) pp 350-6
    • (2010) Digital Mammography (IWDM 2010) , vol.6136 , pp. 350-356
    • Marshall, N.W.1    Jacobs, J.2    Cockmartin, L.3    Bosmans, H.4
  • 24
    • 33745351039 scopus 로고    scopus 로고
    • Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device
    • Mertelmeier T, Orman J, Haerer W and Dudam M 2006 Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device Proc SPIE 6142 61420F-1-12
    • (2006) Proc SPIE , vol.6142
    • Mertelmeier, T.1    Orman, J.2    Haerer, W.3    Dudam, M.4
  • 25
    • 79960374186 scopus 로고    scopus 로고
    • Image quality assessment in digital mammography: Part II. NPWE as a validated alternative for contrast detail analysis
    • 10.1088/0031-9155/56/14/003 0031-9155
    • Monnin P, Marshall N W, Bosmans H, Bochud F O and Verdun F R 2011 Image quality assessment in digital mammography: part II. NPWE as a validated alternative for contrast detail analysis Phys. Med. Biol. 56 4221-38
    • (2011) Phys. Med. Biol. , vol.56 , Issue.14 , pp. 4221-4238
    • Monnin, P.1    Marshall, N.W.2    Bosmans, H.3    Bochud, F.O.4    Verdun, F.R.5
  • 26
  • 28
    • 0025324064 scopus 로고
    • Mammographic resolution: Influence of focal spot intensity distribution and geometry
    • DOI 10.1118/1.596482
    • Nickoloff E L, Donnelly E, Eve L, Atherton J V and Asch T 1990 Mammographic resolution: influence of focal spot intensity distribution and geometry Med. Phys. 17 436-47 (Pubitemid 20226006)
    • (1990) Medical Physics , vol.17 , Issue.3 , pp. 436-447
    • Nickoloff, E.L.1    Donnelly, E.2    Eve, L.3    Atherton, J.V.4    Asch, T.5
  • 30
    • 78650159058 scopus 로고    scopus 로고
    • A statistical task-based evaluation method for three-dimensional x-ray breast imaging using variable background phantoms
    • 10.1118/1.3488910 0094-2405
    • Park S, Jennings R, Liu H, Badano A and Myers K 2010 A statistical task-based evaluation method for three-dimensional x-ray breast imaging using variable background phantoms Med. Phys. 37 6253-70
    • (2010) Med. Phys. , vol.37 , Issue.12 , pp. 6253-6270
    • Park, S.1    Jennings, R.2    Liu, H.3    Badano, A.4    Myers, K.5
  • 32
    • 77950553921 scopus 로고    scopus 로고
    • Task-based assessment of breast tomosynthesis: Effect of acquisition parameters and quantum noise
    • 10.1118/1.3357288 0094-2405
    • Reiser I and Nishikawa R M 2010 Task-based assessment of breast tomosynthesis: effect of acquisition parameters and quantum noise Med. Phys. 37 1591-600
    • (2010) Med. Phys. , vol.37 , Issue.4 , pp. 1591-1600
    • Reiser, I.1    Nishikawa, R.M.2
  • 33
    • 79955766837 scopus 로고    scopus 로고
    • 3D task-based performance assessment metrics for optimization of performance and dose in breast tomosynthesis of performance and dose in breast tomosynthesis
    • Richard S and Samei E 2011 3D task-based performance assessment metrics for optimization of performance and dose in breast tomosynthesis of performance and dose in breast tomosynthesis Proc. SPIE 7961 79611A-1-6
    • (2011) Proc. SPIE , vol.7961
    • Richard, S.1    Samei, E.2
  • 34
    • 0031950889 scopus 로고    scopus 로고
    • A method for measuring the presampled MTF of digital radiographic systems using an edge test device
    • DOI 10.1118/1.598165
    • Samei E, Flynn M J and Reimann D A 1998 A method for measuring the presampled MTF of digital radiographic systems using an edge test device Med. Phys. 25 102-13 (Pubitemid 28063694)
    • (1998) Medical Physics , vol.25 , Issue.1 , pp. 102-113
    • Samei, E.1    Flynn, M.J.2    Reimann, D.A.3
  • 35
    • 33646438839 scopus 로고    scopus 로고
    • Intercomparison of methods for image quality characterization: I. Modulation transfer function
    • 10.1118/1.2188816 0094-2405
    • Samei E, Ranger N T, Dobbins J T III and Chen Y 2006 Intercomparison of methods for image quality characterization: I. Modulation transfer function Med. Phys. 33 1454-65
    • (2006) Med. Phys. , vol.33 , Issue.5 , pp. 1454-1465
    • Samei, E.1    Ranger, N.T.2    Dobbins, J.T.3    Chen, Y.4
  • 36
    • 35148860311 scopus 로고    scopus 로고
    • Method for the determination of the modulation transfer function (MTF) in 3D x-ray imaging systems with focus on correction for finite extent of test objects
    • Schäfer D, Wiegert J and Bertram M 2007 Method for the determination of the modulation transfer function (MTF) in 3D x-ray imaging systems with focus on correction for finite extent of test objects Proc. SPIE 6510 65104Q-1-10
    • (2007) Proc. SPIE , vol.6510
    • Schäfer, D.1    Wiegert, J.2    Bertram, M.3
  • 37
    • 63849136026 scopus 로고    scopus 로고
    • Optimization of the acquisition geometry in digital breast tomosynthesis
    • 10.1118/1.3090889 0094-2405
    • Sechopoulos I and Ghetti C 2009 Optimization of the acquisition geometry in digital breast tomosynthesis Med. Phys. 36 1199-207
    • (2009) Med. Phys. , vol.36 , Issue.4 , pp. 1199-1207
    • Sechopoulos, I.1    Ghetti, C.2
  • 38
    • 79955782661 scopus 로고    scopus 로고
    • Investigation of the effect of tube motion in breast tomosynthesis: Continuous or step and shoot?
    • Shaheen E, Marshall N and Bosmans H 2011a Investigation of the effect of tube motion in breast tomosynthesis: continuous or step and shoot? Proc. SPIE 7961 79611E-1-9
    • (2011) Proc. SPIE , vol.7961
    • Shaheen, E.1    Marshall, N.2    Bosmans, H.3
  • 40
    • 1842865422 scopus 로고    scopus 로고
    • A framework for noise-power spectrum analysis of multidimensional images
    • DOI 10.1118/1.1513158
    • Siewerdsen J H, Cunningham I A and Jaffray D A 2002 A framework for noise-power spectrum analysis of multidimensional images Med. Phys. 29 2655-71 (Pubitemid 35375465)
    • (2002) Medical Physics , vol.29 , Issue.11 , pp. 2655-2671
    • Siewerdsen, J.H.1    Cunningham, I.A.2    Jaffray, D.A.3
  • 43
    • 33745368904 scopus 로고    scopus 로고
    • Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom
    • Thornton M M and Flynn M J 2006 Measurement of the spatial resolution of a clinical volumetric computed tomography scanner using a sphere phantom Proc SPIE 6142 61421Z-1-10
    • (2006) Proc SPIE , vol.6142
    • Thornton, M.M.1    Flynn, M.J.2
  • 45
    • 58149136863 scopus 로고    scopus 로고
    • Experimental validation of a three-dimensional linear system model for breast tomosynthesis
    • 10.1118/1.3040178 0094-2405
    • Zhao B, Zhou J, Hu Y-H, Mertelmeier T, Ludwig J and Zhao W 2009 Experimental validation of a three-dimensional linear system model for breast tomosynthesis Med. Phys. 36 240-51
    • (2009) Med. Phys. , vol.36 , Issue.1 , pp. 240-251
    • Zhao, B.1    Zhou, J.2    Hu, Y.-H.3    Mertelmeier, T.4    Ludwig, J.5    Zhao, W.6
  • 46
    • 56749165474 scopus 로고    scopus 로고
    • Three-dimensional linear system analysis for breast tomosynthesis
    • 10.1118/1.2996014 0094-2405
    • Zhao B and Zhao W 2008 Three-dimensional linear system analysis for breast tomosynthesis Med. Phys. 35 5129-232
    • (2008) Med. Phys. , vol.35 , Issue.12 , pp. 5219-5232
    • Zhao, B.1    Zhao, W.2
  • 47
    • 33847676189 scopus 로고    scopus 로고
    • A computer simulation platform for the optimization of a breast tomosynthesis system
    • DOI 10.1118/1.2558160
    • Zhou J, Zhao B and Zhao W 2007 A computer simulation platform for the optimization of a breast tomosynthesis system Med. Phys. 34 1098-109 (Pubitemid 46355644)
    • (2007) Medical Physics , vol.34 , Issue.3 , pp. 1098-1109
    • Zhou, J.1    Zhao, B.2    Zhao, W.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.