Task-based assessment of breast tomosynthesis: Effect of acquisition parameters and quantum noise

Author keywords

Objective assessment; Prewhitening observer model; Structured background; Tomosynthesis

Indexed keywords

BIOMEDICAL SIGNAL PROCESSING; COMPUTERIZED TOMOGRAPHY; ESTIMATION; ITERATIVE METHODS; MAXIMUM LIKELIHOOD; MAXIMUM PRINCIPLE; MEDICAL IMAGING; QUANTUM NOISE; SIGNAL SAMPLING;

EID: 77950553921     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3357288     Document Type: Article
Times cited : (176)

References (48)
  • 2
    • 0035049369 scopus 로고    scopus 로고
    • Human observer detection experiments with mammograms and power-law noise
    • DOI 10.1118/1.1355308
    • A. E. Burgess, F. L. Jacobson, and P. F. Judy, " Human observer detection experiments with mammograms and power-law noise.," Med. Phys. 0094-2405 28, 419-437 (2001). 10.1118/1.1355308 (Pubitemid 32323293)
    • (2001) Medical Physics , vol.28 , Issue.4 , pp. 419-437
    • Burgess, A.E.1    Jacobson, F.L.2    Judy, P.F.3
  • 3
    • 0141928682 scopus 로고    scopus 로고
    • Digital x-ray tomosynthesis: Current state of the art and clinical potential
    • DOI 10.1088/0031-9155/48/19/R01, PII S0031915503325618
    • J. T. Dobbins III and D. J. Godfrey, " Digital x-ray tomosynthesis: Current state of the art and clinical potential., Phys. Med. Biol. 0031-9155 48, R65-R106 (2003). 10.1088/0031-9155/48/19/R01 (Pubitemid 37258356)
    • (2003) Physics in Medicine and Biology , vol.48 , Issue.19
    • Dobbins III, J.T.1    Godfrey, D.J.2
  • 5
    • 0030711575 scopus 로고    scopus 로고
    • Digital tomosynthesis in breast imaging
    • 0033-8419.
    • L. T. Niklason, " Digital tomosynthesis in breast imaging.," Radiology 0033-8419 205, 399-406 (1997).
    • (1997) Radiology , vol.205 , pp. 399-406
    • Niklason, L.T.1
  • 8
    • 0042376270 scopus 로고    scopus 로고
    • Practical strategies for the clinical implementation of matrix inversion tomosynthesis (MITS)
    • 0277-786X,. 10.1117/12.480352
    • D. J. Godfrey, A. Radar, and J. T. Dobbins III, " Practical strategies for the clinical implementation of matrix inversion tomosynthesis (MITS).," Proc. SPIE 0277-786X 5030, 379-391 (2003). 10.1117/12.480352
    • (2003) Proc. SPIE , vol.5030 , pp. 379-391
    • Godfrey, D.J.1    Radar, A.2    Dobbins III, J.T.3
  • 9
    • 65449183123 scopus 로고    scopus 로고
    • Stochastic noise characteristics in matrix inversion tomosynthesis (MITS)
    • 0094-2405,. 10.1118/1.3103399
    • D. J. Godfrey, H. P. McAdams, and J. T. Dobbins III, " Stochastic noise characteristics in matrix inversion tomosynthesis (MITS).," Med. Phys. 0094-2405 36, 1521-1532 (2009). 10.1118/1.3103399
    • (2009) Med. Phys. , vol.36 , pp. 1521-1532
    • Godfrey, D.J.1    McAdams, H.P.2    Dobbins III, J.T.3
  • 10
    • 23844452192 scopus 로고    scopus 로고
    • Design and performance of the prototype full field breast tomosynthesis system with selenium based flat panel detector
    • DOI 10.1117/12.595833, 61, Medical Imaging 2005 - Physics of Medical Imaging
    • B. Ren, C. Ruth, J. Stein, A. Smith, I. Shaw, and Z. Jing, " Design and performance of the prototype full field breast tomosynthesis system with selenium-based flat panel detector.," Proc. SPIE 0277-786X 5745, 550-561 (2005). 10.1117/12.595833 (Pubitemid 41146402)
    • (2005) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.5745 , Issue.I , pp. 550-561
    • Ren, B.1    Ruth, C.2    Stein, J.3    Smith, A.4    Shaw, I.5    Jing, Z.6
  • 11
    • 33745351039 scopus 로고    scopus 로고
    • Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device
    • DOI 10.1117/12.651380, Medical Imaging 2006: Physics of Medical Imaging
    • T. Mertelmeier, J. Orman, W. Haerer, and M. K. Dudam, " Optimizing filtered backprojection reconstruction for a breast tomosynthesis prototype device.," Proc. SPIE 0277-786X 6142, 61420F (2006). 10.1117/12.651380 (Pubitemid 43944866)
    • (2006) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.I6142
    • Mertelmeier, T.1    Orman, J.2    Haerer, W.3    Dudam, M.K.4
  • 12
    • 33749404804 scopus 로고    scopus 로고
    • A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis
    • DOI 10.1118/1.2237543
    • Y. H. Zhang, H. P. Chan, B. Sahiner, J. Wei, M. M. Goodsitt, L. M. Hadjiiski, J. Ge, and C. A. Zhou, " A comparative study of limited-angle cone-beam reconstruction methods for breast tomosynthesis.," Med. Phys. 0094-2405 33, 3781-3795 (2006). 10.1118/1.2237543 (Pubitemid 44505103)
    • (2006) Medical Physics , vol.33 , Issue.10 , pp. 3781-3795
    • Zhang, Y.1    Chan, H.-P.2    Sahiner, B.3    Wei, J.4    Goodsitt, M.M.5    Hadjiiski, L.M.6    Ge, J.7    Zhou, C.8
  • 13
    • 70350718948 scopus 로고    scopus 로고
    • Enhanced imaging of micro-calcifications in digital breast tomosynthesis through improved image reconstruction algorithms
    • 0094-2405,. 10.1118/1.3232211
    • E. Y. Sidky, X. C. Pan, I. Reiser, R. M. Nishikawa, R. H. Moore, and D. B. Kopans, " Enhanced imaging of micro-calcifications in digital breast tomosynthesis through improved image reconstruction algorithms.," Med. Phys. 0094-2405 36, 4920-4932 (2009). 10.1118/1.3232211
    • (2009) Med. Phys. , vol.36 , pp. 4920-4932
    • Sidky, E.Y.1    Pan, X.C.2    Reiser, I.3    Nishikawa, R.M.4    Moore, R.H.5    Kopans, D.B.6
  • 14
    • 35148840705 scopus 로고    scopus 로고
    • Image reconstruction in digital breast tomosynthesis by total variation minimization
    • DOI 10.1117/12.713663, Medical Imaging 2007: Physics of Medical Imaging
    • I. Reiser, J. Bian, R. M. Nishikawa, E. Y. Sidky, and X. Pan, " Comparison of reconstruction algorithms for digital breast tomosynthesis.," in Proceedings of the Ninth International Meeting on Fully Three-Dimensional Image Reconstruction in Radiology and Nuclear Medicine, 2007, pp. 155-158 (available at http://arXiv.org/abs/0908.2610v1). (Pubitemid 47550183)
    • (2007) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.6510 , Issue.PART 2 , pp. 651027
    • Sidky, E.Y.1    Reiser, I.S.2    Nishikawa, R.3    Pan, X.4
  • 16
    • 63849136026 scopus 로고    scopus 로고
    • Optimization of the acquisition geometry in digital tomosynthesis of the breast
    • 0094-2405,. 10.1118/1.3090889
    • I. Sechopoulos and C. Ghetti, " Optimization of the acquisition geometry in digital tomosynthesis of the breast.," Med. Phys. 0094-2405 36, 1199-1207 (2009). 10.1118/1.3090889
    • (2009) Med. Phys. , vol.36 , pp. 1199-1207
    • Sechopoulos, I.1    Ghetti, C.2
  • 17
    • 56749165474 scopus 로고    scopus 로고
    • Three-dimensional linear systems analysis for breast tomosynthesis
    • 0094-2405,. 10.1118/1.2996014
    • B. Zhao and W. Zhao, " Three-dimensional linear systems analysis for breast tomosynthesis.," Med. Phys. 0094-2405 35, 5219-5232 (2008). 10.1118/1.2996014
    • (2008) Med. Phys. , vol.35 , pp. 5219-5232
    • Zhao, B.1    Zhao, W.2
  • 18
    • 58149136863 scopus 로고    scopus 로고
    • Experimental validation of a three-dimensional linear system model for breast tomosynthesis
    • 0094-2405,. 10.1118/1.3040178
    • B. Zhao, J. Zhou, Y. H. Hu, T. Mertelmeier, J. Ludwig, and W. Zhao, " Experimental validation of a three-dimensional linear system model for breast tomosynthesis.," Med. Phys. 0094-2405 36, 240-251 (2009). 10.1118/1.3040178
    • (2009) Med. Phys. , vol.36 , pp. 240-251
    • Zhao, B.1    Zhou, J.2    Hu, Y.H.3    Mertelmeier, T.4    Ludwig, J.5    Zhao, W.6
  • 19
    • 33748303079 scopus 로고    scopus 로고
    • Resolution at oblique incidence angles of a flat panel imager for breast tomosynthesis
    • DOI 10.1118/1.2241994
    • J. G. Mainprize, A. K. Bloomquist, M. R. Kempston, and M. J. Yaffe, " Resolution at oblique incidence angles of a flat panel imager for breast tomosynthesis.," Med. Phys. 0094-2405 33, 3159-3164 (2006). 10.1118/1.2241994 (Pubitemid 44327416)
    • (2006) Medical Physics , vol.33 , Issue.9 , pp. 3159-3164
    • Mainprize, J.G.1    Bloomquist, A.K.2    Kempston, M.P.3    Yaffe, M.J.4
  • 20
    • 70350745246 scopus 로고    scopus 로고
    • Optimized image acquisition for breast tomosynthesis in projection and reconstruction space
    • 0094-2405,. 10.1118/1.3231814
    • A. S. Chawla, J. Y. Lo, J. A. Baker, and E. Samei, " Optimized image acquisition for breast tomosynthesis in projection and reconstruction space.," Med. Phys. 0094-2405 36, 4859-4869 (2009). 10.1118/1.3231814
    • (2009) Med. Phys. , vol.36 , pp. 4859-4869
    • Chawla, A.S.1    Lo, J.Y.2    Baker, J.A.3    Samei, E.4
  • 23
    • 0031012181 scopus 로고    scopus 로고
    • Simulation of subtle lung nodules in projection chest radiography
    • E. Samei, M. J. Flynn, and W. R. Eyler, " Simulation of subtle lung nodules in projection chest radiography.," Radiology 0033-8419 202, 117-124 (1997). (Pubitemid 27024116)
    • (1997) Radiology , vol.202 , Issue.1 , pp. 117-124
    • Samei, E.1    Flynn, M.J.2    Eyler, W.R.3
  • 25
    • 4644334876 scopus 로고    scopus 로고
    • A comparison of reconstruction algorithms for breast tomosynthesis
    • DOI 10.1118/1.1786692
    • T. Wu, R. H. Moore, E. A. Rafferty, and D. B. Kopans, " A comparison of reconstruction algorithms for breast tomosynthesis.," Med. Phys. 0094-2405 31, 2636-2647 (2004). 10.1118/1.1786692 (Pubitemid 39287881)
    • (2004) Medical Physics , vol.31 , Issue.9 , pp. 2636-2647
    • Wu, T.1    Moore, R.H.2    Rafferty, E.A.3    Kopans, D.B.4
  • 28
    • 65449139477 scopus 로고    scopus 로고
    • Comparison of power spectra for tomosynthesis projections and reconstructed images
    • 0094-2405,. 10.1118/1.3116774
    • E. Engstrom, I. Reiser, and R. M. Nishikawa, " Comparison of power spectra for tomosynthesis projections and reconstructed images.," Med. Phys. 0094-2405 36, 1753-1758 (2009). 10.1118/1.3116774
    • (2009) Med. Phys. , vol.36 , pp. 1753-1758
    • Engstrom, E.1    Reiser, I.2    Nishikawa, R.M.3
  • 31
    • 54949086053 scopus 로고    scopus 로고
    • Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images
    • 0094-2405,. 10.1118/1.2988161
    • S. Richard and J. H. Siewerdsen, " Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images.," Med. Phys. 0094-2405 35, 5043-5053 (2008). 10.1118/1.2988161
    • (2008) Med. Phys. , vol.35 , pp. 5043-5053
    • Richard, S.1    Siewerdsen, J.H.2
  • 32
    • 0034132171 scopus 로고    scopus 로고
    • Visual signal detection in structured backgrounds. III. Calculation of figures of merit for model observers in statistically nonstationary backgrounds
    • 10.1364/JOSAA.17.000193 1084-7529
    • F. O. Bochud, C. K. Abbey, and M. P. Eckstein, " Visual signal detection in structured backgrounds. III. Calculation of figures of merit for model observers in statistically nonstationary backgrounds.," J. Opt. Soc. Am. A Opt. Image Sci. Vis. 17, 193-205 (2000). 10.1364/JOSAA.17.000193 1084-7529
    • (2000) J. Opt. Soc. Am. A Opt. Image Sci. Vis. , vol.17 , pp. 193-205
    • Bochud, F.O.1    Abbey, C.K.2    Eckstein, M.P.3
  • 33
    • 0032783631 scopus 로고    scopus 로고
    • Estimation of the noisy component of anatomical backgrounds
    • DOI 10.1118/1.598632
    • F. O. Bochud, J. F. Valley, F. R. Verdun, C. Hessler, and P. Schnyder, " Estimation of the noisy component of anatomical backgrounds.," Med. Phys. 0094-2405 26, 1365-1370 (1999). 10.1118/1.598632 (Pubitemid 29334805)
    • (1999) Medical Physics , vol.26 , Issue.7 , pp. 1365-1370
    • Bochud, F.O.1    Valley, J.-F.2    Verdun, F.R.3    Hessler, C.4    Schnyder, P.5
  • 35
    • 0034046904 scopus 로고    scopus 로고
    • Monte Carlo validation in diagnostic radiological imaging
    • DOI 10.1118/1.599007
    • J. M. Boone, M. H. Buonocore, and V. N. Cooper, " Monte Carlo validation in diagnostic radiological imaging.," Med. Phys. 0094-2405 27, 1294-1304 (2000). 10.1118/1.599007 (Pubitemid 30416098)
    • (2000) Medical Physics , vol.27 , Issue.6 , pp. 1294-1304
    • Boone, J.M.1    Buonocore, M.H.2    Cooper III, V.N.3
  • 36
    • 33845927292 scopus 로고    scopus 로고
    • Computation of the glandular radiation dose in digital tomosynthesis of the breast
    • DOI 10.1118/1.2400836
    • I. Sechopoulos, S. Suryanarayanan, S. Vedantham, C. J. D'Orsi, and A. Karellas, " Computation of the glandular radiation dose in digital tomosynthesis of the breast.," Med. Phys. 0094-2405 34, 221-232 (2007). 10.1118/1.2400836 (Pubitemid 46035245)
    • (2007) Medical Physics , vol.34 , Issue.1 , pp. 221-232
    • Sechopoulos, I.1    Suryanarayanan, S.2    Vedantham, S.3    D'Orsi, C.4    Karellas, A.5
  • 37
    • 33645524546 scopus 로고    scopus 로고
    • A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging
    • 0094-2405,. 10.1118/1.2174127
    • X. Gong, S. J. Glick, B. Liu, A. A. Vdula, and S. Thacker, " A computer simulation study comparing lesion detection accuracy with digital mammography, breast tomosynthesis, and cone-beam CT breast imaging.," Med. Phys. 0094-2405 33, 1041-1052 (2006). 10.1118/1.2174127
    • (2006) Med. Phys. , vol.33 , pp. 1041-1052
    • Gong, X.1    Glick, S.J.2    Liu, B.3    Vdula, A.A.4    Thacker, S.5
  • 39
    • 34047275404 scopus 로고    scopus 로고
    • Fractal analysis of mammographic parenchymal patterns in breast cancer risk assessment
    • DOI 10.1016/j.acra.2007.02.003, PII S107663320700089X
    • H. Li, M. L. Giger, O. I. Olopade, and M. R. Chinander, " Power spectral analysis of mammographic parenchymal patterns for breast cancer risk assessment.," J. Digit Imaging 0897-1889 21, 145-152 (2008). 10.1007/s10278-007-9093-9 (Pubitemid 46551761)
    • (2007) Academic Radiology , vol.14 , Issue.5 , pp. 513-521
    • Li, H.1    Giger, M.L.2    Olopade, O.I.3    Lan, L.4
  • 40
    • 0023196389 scopus 로고
    • X-ray characterisation of normal and neoplastic breast tissues
    • 0031-9155,. 10.1088/0031-9155/32/6/002
    • P. C. Johns and M. J. Yaffe, " X-ray characterisation of normal and neoplastic breast tissues.," Phys. Med. Biol. 0031-9155 32, 675-695 (1987). 10.1088/0031-9155/32/6/002
    • (1987) Phys. Med. Biol. , vol.32 , pp. 675-695
    • Johns, P.C.1    Yaffe, M.J.2
  • 44
    • 0032594435 scopus 로고    scopus 로고
    • Advances in high-speed mammographic image quality
    • 0277-786X,. 10.1117/12.349485
    • P. C. Bunch, " Advances in high-speed mammographic image quality.," Proc. SPIE 0277-786X 3659, 120-130 (1999). 10.1117/12.349485
    • (1999) Proc. SPIE , vol.3659 , pp. 120-130
    • Bunch, P.C.1
  • 45
    • 0031536511 scopus 로고    scopus 로고
    • Improvements on Cross-Validation: The .632+ Bootstrap Method
    • B. Efron and R. Tibshirani, " Improvements on cross-validation: The.632+ bootstrap method.," J. Am. Stat. Assoc. 0162-1459 92, 548-56 (1997). 10.2307/2965703 (Pubitemid 127094695)
    • (1997) Journal of the American Statistical Association , vol.92 , Issue.438 , pp. 548-560
    • Efron, B.1    Tibshirani, R.2
  • 47
    • 33745430674 scopus 로고    scopus 로고
    • Low-contrast lesion detection in tomosynthetic breast imaging using a realistic breast phantom
    • DOI 10.1117/12.652343, Medical Imaging 2006: Physics of Medical Imaging
    • L. Zhou, J. Oldan, P. Fisher, and G. Gindi, " Low-contrast lesion detection in tomosynthetic breast imaging using a realistic breast phantom.," Proc. SPIE 0277-786X 6142, 61425A (2006). 10.1117/12.652343 (Pubitemid 43945032)
    • (2006) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.III6142
    • Zhou, L.1    Oldan, J.2    Fisher, P.3    Gindi, G.4
  • 48
    • 66749084526 scopus 로고    scopus 로고
    • Effect of non-isotropic detector blur on microcalcification detectability in tomosynthesis
    • 0277-786X,. 10.1117/12.813808
    • I. Reiser, R. M. Nishikawa, and B. A. Lau, " Effect of non-isotropic detector blur on microcalcification detectability in tomosynthesis.," Proc. SPIE 0277-786X 7258, 72585Z (2009). 10.1117/12.813808
    • (2009) Proc. SPIE , vol.7258
    • Reiser, I.1    Nishikawa, R.M.2    Lau, B.A.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.