Aerosol therapy in patients receiving noninvasive positive pressure ventilation

Author keywords

acute exacerbation of COPD; bronchodilator therapy; continuous positive airway pressure; inhalation therapy; mechanical ventilation; metered dose inhalers; nebulizers; noninvasive ventilation; pressure support ventilation

Indexed keywords

AEROSOL; ARTIFICIAL VENTILATION; ASTHMA; BREATHING; CHRONIC OBSTRUCTIVE LUNG DISEASE; DRUG DELIVERY SYSTEM; FACE MASK; FORCED EXPIRATORY VOLUME; FORCED VITAL CAPACITY; GAS FLOW; HUMAN; HUMIDITY; NASOGASTRIC TUBE; NEBULIZER; NONINVASIVE POSITIVE PRESSURE VENTILATION; PATIENT SELECTION; PEAK INSPIRATORY FLOW; POSITIVE END EXPIRATORY PRESSURE; RESPIRATORY AIRFLOW; RESPIRATORY DISTRESS; REVIEW;

EID: 84859386983     PISSN: 19412711     EISSN: 19412703     Source Type: Journal    
DOI: 10.1089/jamp.2011.0929     Document Type: Review
Times cited : (73)

References (132)
  • 1
    • 67650367543 scopus 로고    scopus 로고
    • Non-invasive ventilation in acute respiratory failure
    • Nava S, and Hill NS: Non-invasive ventilation in acute respiratory failure. Lancet. 2009;374:250-259.
    • (2009) Lancet , vol.374 , pp. 250-259
    • Nava, S.1    Hill, N.S.2
  • 2
    • 0035881158 scopus 로고    scopus 로고
    • Physiologic response to pressure support ventilation delivered before and after extubation in patients not capable of totally spontaneous breathing
    • Vitacca M, Ambrosino N, Clini E, Porta R, Rampulla C, Lanini B, and Nava S: Physiologic response to pressure support ventilation delivered before and after extubation in patients not capable of totally spontaneous breathing. Am J Respir Crit Care Med. 2001;164:638-641.
    • (2001) Am. J. Respir. Crit. Care Med. , vol.164 , pp. 638-641
    • Vitacca, M.1    Ambrosino, N.2    Clini, E.3    Porta, R.4    Rampulla, C.5    Lanini, B.6    Nava, S.7
  • 3
    • 42949119771 scopus 로고    scopus 로고
    • Noninvasive positive pressure ventilation in the acute care setting: Where are we
    • Ambrosino N, and Vagheggini G: Noninvasive positive pressure ventilation in the acute care setting: where are we? Eur Respir J. 2008;31:874-886.
    • (2008) Eur. Respir. J. , vol.31 , pp. 874-886
    • Ambrosino, N.1    Vagheggini, G.2
  • 7
    • 0029018269 scopus 로고
    • Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failure
    • Kramer N, Meyer TJ, Meharg J, Cece RD, and Hill NS: Randomized, prospective trial of noninvasive positive pressure ventilation in acute respiratory failure. Am J Respir Crit Care Med. 1995;151:1799-1806.
    • (1995) Am. J. Respir. Crit. Care Med. , vol.151 , pp. 1799-1806
    • Kramer, N.1    Meyer, T.J.2    Meharg, J.3    Cece, R.D.4    Hill, N.S.5
  • 8
    • 84923222206 scopus 로고    scopus 로고
    • Noninvasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary disease
    • CD004104
    • Ram FSF, Picot J, Lightowler J, and Wedzicha JA: Noninvasive positive pressure ventilation for treatment of respiratory failure due to exacerbations of chronic obstructive pulmonary disease. Cochrane Database Syst Rev. 2004;3: CD004104.
    • (2004) Cochrane Database Syst. Rev. , vol.3
    • Ram, F.S.F.1    Picot, J.2    Lightowler, J.3    Wedzicha, J.A.4
  • 9
    • 0035064877 scopus 로고    scopus 로고
    • And the srlf Collaborative group on Mechanical Ventilation: Noninvasive versus conventional mechanical ventilation: An epidemiologic survey
    • Carlucci A, Richard JC, Wysocki M, Lepage E, Brochard L, and the SRLF Collaborative group on Mechanical Ventilation: Noninvasive versus conventional mechanical ventilation: an epidemiologic survey. Am J Respir Crit Care Med. 2001;163:874-880.
    • (2001) Am. J. Respir. Crit. Care Med. , vol.163 , pp. 874-880
    • Carlucci, A.1    Richard, J.C.2    Wysocki, M.3    Lepage, E.4    Brochard, L.5
  • 10
    • 68049086912 scopus 로고    scopus 로고
    • Effects of inhalational bronchodilator treatment during noninvasive ventilation in severe chronic obstructive pulmonary disease exacerbations
    • Mukhopadhyay A, Dela Pena E, Wadden B, Procyshyn M, and Lim TK: Effects of inhalational bronchodilator treatment during noninvasive ventilation in severe chronic obstructive pulmonary disease exacerbations. J Crit Care. 2009;24:474.e1-474.e5.
    • (2009) J. Crit. Care. , vol.24
    • Mukhopadhyay, A.1    Dela Pena, E.2    Wadden, B.3    Procyshyn, M.4    Lim, T.K.5
  • 11
    • 33846038094 scopus 로고    scopus 로고
    • Inhalation therapy in invasive and noninvasive mechanical ventilation
    • Dhand R: Inhalation therapy in invasive and noninvasive mechanical ventilation. Curr Opin Crit Care. 2007;13:27-38.
    • (2007) Curr. Opin. Crit. Care. , vol.13 , pp. 27-38
    • Dhand, R.1
  • 13
    • 34247611809 scopus 로고    scopus 로고
    • The mask for noninvasive ventilation: Principles of design and effects on aerosol delivery
    • Hess DR: The mask for noninvasive ventilation: principles of design and effects on aerosol delivery. J Aerosol Med. 2007;20(Suppl 1):S85-S99.
    • (2007) J. Aerosol. Med. , vol.20 SUPPL1
    • Hess, D.R.1
  • 14
    • 4644303572 scopus 로고    scopus 로고
    • Aerosol delivery to ventilated infant and pediatric patients
    • Fink JB: Aerosol delivery to ventilated infant and pediatric patients. Respir Care. 2004;49:653-665.
    • (2004) Respir. Care , vol.49 , pp. 653-665
    • Fink, J.B.1
  • 15
    • 0031020173 scopus 로고    scopus 로고
    • Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask
    • Parkes SN, Bersten AD: Aerosol kinetics and bronchodilator efficacy during continuous positive airway pressure delivered by face mask. Thorax. 1997;52:171-175
    • (1997) Thorax , vol.52 , pp. 171-175
    • Parkes, S.N.1    Bersten, A.D.2
  • 16
    • 2642518054 scopus 로고    scopus 로고
    • Factors affecting oxygen delivery with bi-level positive airway pressure
    • Schwartz AR, Kacmarek RM, and Hess DR: Factors affecting oxygen delivery with bi-level positive airway pressure. Respir Care. 2004;49:270-275.
    • (2004) Respir. Care , vol.49 , pp. 270-275
    • Schwartz, A.R.1    Kacmarek, R.M.2    Hess, D.R.3
  • 17
    • 34748905281 scopus 로고    scopus 로고
    • Noninvasive ventilation in neuromuscular disease: Equipment and application
    • 896- 911; discussion
    • Hess DR: Noninvasive ventilation in neuromuscular disease: equipment and application. Respir Care. 2006;51:896- 911; discussion 911-912.
    • (2006) Respir. Care , vol.51 , pp. 911-912
    • Hess, D.R.1
  • 18
    • 0028785413 scopus 로고
    • Comparative effects of pressure support ventilation and intermittent positive pressure breathing (IPPB) in non-intubated healthy subjects
    • Mancebo J, Isabey D, Lorino H, Lofaso F, Lemaire F, and Brochard L: Comparative effects of pressure support ventilation and intermittent positive pressure breathing (IPPB) in non-intubated healthy subjects. Eur Respir J. 1995;8: 1901-1909.
    • (1995) Eur. Respir J. , vol.8 , pp. 1901-1909
    • Mancebo, J.1    Isabey, D.2    Lorino, H.3    Lofaso, F.4    Lemaire, F.5    Brochard, L.6
  • 19
    • 29544442582 scopus 로고    scopus 로고
    • And Expó sito JA: Noninvasive ventilation in acute cardiogenic pulmonary edema: Systematic review and metaanalysis
    • Masip J, Roque M, Sánchez B, Fernández R, Subirana M, and Expó sito JA: Noninvasive ventilation in acute cardiogenic pulmonary edema: systematic review and metaanalysis. JAMA. 2005;294:3124-3130.
    • (2005) JAMA , vol.294 , pp. 3124-3130
    • Masip, J.1    Roque, M.2    Sánchez, B.3    Fernández, R.4    Subirana, M.5
  • 20
    • 33645979354 scopus 로고    scopus 로고
    • Effect of non-invasive positive pressure ventilation (NIPPV) on mortality in patients with acute cardiogenic pulmonary oedema: A meta-analysis
    • Peter JV, Moran JL, Phillips-Hughes J, Graham P, and Bersten AD: Effect of non-invasive positive pressure ventilation (NIPPV) on mortality in patients with acute cardiogenic pulmonary oedema: a meta-analysis. Lancet. 2006;367:1155-1163.
    • (2006) Lancet , vol.367 , pp. 1155-1163
    • Peter, J.V.1    Moran, J.L.2    Phillips-Hughes, J.3    Graham, P.4    Bersten, A.D.5
  • 21
    • 0030002808 scopus 로고    scopus 로고
    • Noninvasive mechanical ventilation in patients with acute respiratory failure
    • Abou-Shala N, and Meduri U: Noninvasive mechanical ventilation in patients with acute respiratory failure. Crit Care Med. 1996;24:705-715.
    • (1996) Crit. Care Med. , vol.24 , pp. 705-715
    • Abou-Shala, N.1    Meduri, U.2
  • 22
    • 0031464945 scopus 로고    scopus 로고
    • Effects of non-invasive ventilation on pulmonary gas exchange and hemodynamics during acute hypercapnic exacerbations of chronic obstructive airways disease
    • Diaz O, Iglesia R, Ferrer M, Zavala E, Santos C, Wagner PD, Roca J, and Rodriguez-Roisin R: Effects of non-invasive ventilation on pulmonary gas exchange and hemodynamics during acute hypercapnic exacerbations of chronic obstructive airways disease. Am J Respir Crit Care Med. 1997;156:1840-1845.
    • (1997) Am. J. Respir. Crit. Care Med. , vol.156 , pp. 1840-1845
    • Diaz, O.1    Iglesia, R.2    Ferrer, M.3    Zavala, E.4    Santos, C.5    Wagner, P.D.6    Roca, J.7    Rodriguez-Roisin, R.8
  • 23
    • 0031010891 scopus 로고    scopus 로고
    • Comparative physiologic effects of non-invasive assist control and pressure support ventilation in acute hypercapnic respiratory failure
    • Girault C, Richard JC, Chevron V, Tamion F, Pasquis P, Leroy J, and Bonmarchand G: Comparative physiologic effects of non-invasive assist control and pressure support ventilation in acute hypercapnic respiratory failure. Chest. 1997;111:1639-1648.
    • (1997) Chest , vol.111 , pp. 1639-1648
    • Girault, C.1    Richard, J.C.2    Chevron, V.3    Tamion, F.4    Pasquis, P.5    Leroy, J.6    Bonmarchand, G.7
  • 24
    • 0024339856 scopus 로고
    • The effect of intermittent positive-pressure hyperinflation in restrictive chest wall disease
    • Simonds AK, Parker RA, and Branthwaite MA: The effect of intermittent positive-pressure hyperinflation in restrictive chest wall disease. Respiration. 1989;55:136-143.
    • (1989) Respiration , vol.55 , pp. 136-143
    • Simonds, A.K.1    Parker, R.A.2    Branthwaite, M.A.3
  • 25
    • 0028296316 scopus 로고
    • Physiologic effects of positive end-expiratory pressure and mask pressure support during exacerbations of chronic obstructive pulmonary disease
    • Appendini L, Patessio A, Zanaboni S, Carone M, Gukov B, Donner CF, and Rossi A: Physiologic effects of positive end-expiratory pressure and mask pressure support during exacerbations of chronic obstructive pulmonary disease. Am J Respir Crit Care Med. 1994;149:1069-1076.
    • (1994) Am. J. Respir. Crit. Care Med. , vol.149 , pp. 1069-1076
    • Appendini, L.1    Patessio, A.2    Zanaboni, S.3    Carone, M.4    Gukov, B.5    Donner, C.F.6    Rossi, A.7
  • 26
    • 0035508509 scopus 로고    scopus 로고
    • Noninvasive proportional assist ventilation for acute respiratory insufficiency comparison with pressure support ventilation
    • Gay PC, Hess D, and Hill N: Noninvasive proportional assist ventilation for acute respiratory insufficiency comparison with pressure support ventilation. Am J Respir Crit Care Med. 2001;164:1606-1611.
    • (2001) Am. J. Respir. Crit. Care Med. , vol.164 , pp. 1606-1611
    • Gay, P.C.1    Hess, D.2    Hill, N.3
  • 27
    • 1242351697 scopus 로고    scopus 로고
    • Comparison of lung deposition in two types of nebulization. Intrapulmonary percussive ventilation vs. jet nebulization
    • Reychler G, Keyeux A, Cremers C, Veriter C, Rodenstein DO, and Liistro G: Comparison of lung deposition in two types of nebulization. Intrapulmonary percussive ventilation vs. jet nebulization. Chest. 2004;125:502-508.
    • (2004) Chest , vol.125 , pp. 502-508
    • Reychler, G.1    Keyeux, A.2    Cremers, C.3    Veriter, C.4    Rodenstein, D.O.5    Liistro, G.6
  • 28
    • 0023811790 scopus 로고
    • Aerosol deposition in the human lung: Effect of highfrequency scillation on the deposition characteristics of an inhaled nebulized aerosol
    • Thomas SH, Langford JA, George RJ, and Geddes DM: Aerosol deposition in the human lung: effect of highfrequency scillation on the deposition characteristics of an inhaled nebulized aerosol. Clin Sci (Lond). 1988;75:535-542.
    • (1988) Clin. Sci. (Lond) , vol.75 , pp. 535-542
    • Thomas, S.H.1    Langford, J.A.2    George, R.J.3    Geddes, D.M.4
  • 29
    • 0036305760 scopus 로고    scopus 로고
    • Effect of a nonrebreathing exhalation valve on long-term ventilation using a bilevel device
    • Hill NN, Carlisle C, and Kramer NR: Effect of a nonrebreathing exhalation valve on long-term ventilation using a bilevel device. Chest. 2002;122:84-91.
    • (2002) Chest , vol.122 , pp. 84-91
    • Hill, N.N.1    Carlisle, C.2    Kramer, N.R.3
  • 30
    • 0346607207 scopus 로고    scopus 로고
    • Dynamic dead space in face masks used with noninvasive ventilators: A lung model study
    • Saatci E, Miller DM, Stell IM, Lee KC, and Moxham J: Dynamic dead space in face masks used with noninvasive ventilators: a lung model study. Eur Respir J. 2004;23:129-135.
    • (2004) Eur. Respir J. , vol.23 , pp. 129-135
    • Saatci, E.1    Miller, D.M.2    Stell, I.M.3    Lee, K.C.4    Moxham, J.5
  • 31
    • 0030845450 scopus 로고    scopus 로고
    • Inhaled drug therapy in mechanically- ventilated patients
    • Dhand R and Tobin MJ: Inhaled drug therapy in mechanically- ventilated patients. Am J Respir Crit Care Med. 1997;156:3-10.
    • (1997) Am. J. Respir. Crit. Care Med. , vol.156 , pp. 3-10
    • Dhand, R.1    Tobin, M.J.2
  • 32
    • 0032900977 scopus 로고    scopus 로고
    • Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors
    • Fink JB, Dhand R, Grychowski J, Fahey PJ, and Tobin MJ: Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors. Am J Respir Crit Care Med. 1999;159:63-68.
    • (1999) Am. J. Respir. Crit. Care Med. , vol.159 , pp. 63-68
    • Fink, J.B.1    Dhand, R.2    Grychowski, J.3    Fahey, P.J.4    Tobin, M.J.5
  • 34
    • 0032086343 scopus 로고    scopus 로고
    • Heat and moisture exchangers: Structure and function
    • Wilkes AR: Heat and moisture exchangers: structure and function. Respir Care Clin N Am. 1998;4:261-279.
    • (1998) Respir. Care Clin. N. Am. , vol.4 , pp. 261-279
    • Wilkes, A.R.1
  • 35
    • 0034113778 scopus 로고    scopus 로고
    • Overcoming the adverse effect of humidity in aerosol delivery via pressurized metereddose inhalers during mechanical ventilation
    • Lange CF, and Finlay WH: Overcoming the adverse effect of humidity in aerosol delivery via pressurized metereddose inhalers during mechanical ventilation. Am J Respir Crit Care Med. 2000;161:1614-1618.
    • (2000) Am. J. Respir. Crit. Care Med. , vol.161 , pp. 1614-1618
    • Lange, C.F.1    Finlay, W.H.2
  • 40
    • 0029845395 scopus 로고    scopus 로고
    • Changes in airway resistance induced by nasal inhalation of cold dry, dry, or moist air in normal individuals
    • Fontanari P, Burnet H, Zattara-Hartmann MC, and Jammes Y: Changes in airway resistance induced by nasal inhalation of cold dry, dry, or moist air in normal individuals. J Appl Physiol. 1996;81:1739-1743.
    • (1996) J. Appl. Physiol. , vol.81 , pp. 1739-1743
    • Fontanari, P.1    Burnet, H.2    Zattara-Hartmann, M.C.3    Jammes, Y.4
  • 41
    • 0032755850 scopus 로고    scopus 로고
    • Beneficial effects of helium:oxygen versus air:oxygen noninvasive pressure support in patients with decompensated chronic obstructive pulmonary disease
    • Jolliet P, Tassaux D, Thouret JM, and Chevrolet JC: Beneficial effects of helium:oxygen versus air:oxygen noninvasive pressure support in patients with decompensated chronic obstructive pulmonary disease. Crit Care Med. 1999;27:2422-2429.
    • (1999) Crit. Care Med. , vol.27 , pp. 2422-2429
    • Jolliet, P.1    Tassaux, D.2    Thouret, J.M.3    Chevrolet, J.C.4
  • 45
    • 48249123463 scopus 로고    scopus 로고
    • Aerosol deposition in the human lung periphery is increased by reduceddensity gas breathing
    • Peterson JB, Prisk GK, and Darquenne C: Aerosol deposition in the human lung periphery is increased by reduceddensity gas breathing. J Aerosol Med Pulm Drug Deliv. 2008;21:159-168.
    • (2008) J. Aerosol. Med. Pulm. Drug. Deliv. , vol.21 , pp. 159-168
    • Peterson, J.B.1    Prisk, G.K.2    Darquenne, C.3
  • 46
    • 0035138503 scopus 로고    scopus 로고
    • Improvement in aerosol delivery with helium-oxygen mixtures during mechanical ventilation
    • Goode ML, Fink JB, Dhand R, and Tobin MJ: Improvement in aerosol delivery with helium-oxygen mixtures during mechanical ventilation. Am J Respir Crit Care Med. 2001; 163:109-114.
    • (2001) Am. J. Respir. Crit. Care Med. , vol.163 , pp. 109-114
    • Goode, M.L.1    Fink, J.B.2    Dhand, R.3    Tobin, M.J.4
  • 47
    • 0032962998 scopus 로고    scopus 로고
    • The effect of heliox on nebulizer function using a beta-agonist bronchodilator
    • Hess DR, Acosta FL, Ritz RH, Kacmarek RM, and Camargo CA Jr: The effect of heliox on nebulizer function using a beta-agonist bronchodilator. Chest. 1999;115:184-189.
    • (1999) Chest , vol.115 , pp. 184-189
    • Hess, D.R.1    Acosta, F.L.2    Ritz, R.H.3    Kacmarek, R.M.4    Camargo, Jr.C.A.5
  • 48
    • 61549135790 scopus 로고    scopus 로고
    • Interfaces and humidification for non-invasive mechanical ventilation
    • Nava S, Navalesi P, and Gregoretti C: Interfaces and humidification for non-invasive mechanical ventilation. Respir Care. 2009;54:71-82.
    • (2009) Respir. Care , vol.54 , pp. 71-82
    • Nava, S.1    Navalesi, P.2    Gregoretti, C.3
  • 49
    • 0006413989 scopus 로고    scopus 로고
    • Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of mask in patients with chronic respiratory failure
    • Navalesi P, Fanfulla F, Frigerio P, Gregoretti C, and Nava S: Physiologic evaluation of noninvasive mechanical ventilation delivered with three types of mask in patients with chronic respiratory failure. Crit Care Med. 2000;28:1785-1790.
    • (2000) Crit. Care Med. , vol.28 , pp. 1785-1790
    • Navalesi, P.1    Fanfulla, F.2    Frigerio, P.3    Gregoretti, C.4    Nava, S.5
  • 50
    • 0037329116 scopus 로고    scopus 로고
    • Controlled trial of oronasal mask ventilation in the treatment of acute respiratory failure
    • Kwok H, McCormack J, Cece R, Houtchens J, and Hill NS: Controlled trial of oronasal mask ventilation in the treatment of acute respiratory failure. Crit Care Med. 2003; 31:468-473.
    • (2003) Crit. Care Med. , vol.31 , pp. 468-473
    • Kwok, H.1    McCormack, J.2    Cece, R.3    Houtchens, J.4    Hill, N.S.5
  • 51
    • 59649095107 scopus 로고    scopus 로고
    • Interface strategy during non-invasive positive pressure ventilation for hypercapnic acute respiratory failure
    • Girault C, Briel A, Benichou J, Hellot MF, Dachraoui F, Tamion F, and Bonmarchand G: Interface strategy during non-invasive positive pressure ventilation for hypercapnic acute respiratory failure. Crit Care Med. 2009;37:124-131.
    • (2009) Crit. Care Med. , vol.37 , pp. 124-131
    • Girault, C.1    Briel, A.2    Benichou, J.3    Hellot, M.F.4    Dachraoui, F.5    Tamion, F.6    Bonmarchand, G.7
  • 52
    • 62849092828 scopus 로고    scopus 로고
    • Physiologic effects of different interfaces during noninvasive ventilation for acute respiratory failure
    • Fraticelli AT, Lellouche F, LHer E, Taille S, Mancebo J, and Brochard L: Physiologic effects of different interfaces during noninvasive ventilation for acute respiratory failure. Crit Care Med. 2009;37:939-945.
    • (2009) Crit. Care Med. , vol.37 , pp. 939-945
    • Fraticelli, A.T.1    Lellouche, F.2    Lher, E.3    Taille, S.4    Mancebo, J.5    Brochard, L.6
  • 54
    • 0027425866 scopus 로고
    • Comparison of nebulised aerosol deposition in the lungs of healthy adults following oral and nasal inhalation
    • Everard ML, Hardy JG, and Milner AD: Comparison of nebulised aerosol deposition in the lungs of healthy adults following oral and nasal inhalation. Thorax. 1993;48:1045-1046.
    • (1993) Thorax , vol.48 , pp. 1045-1046
    • Everard, M.L.1    Hardy, J.G.2    Milner, A.D.3
  • 55
    • 0033803102 scopus 로고    scopus 로고
    • Breath-synchronized nebulization diminishes the impact of patient-device interfaces (face mask or mouthpiece) on the inhaled mass of nebulized budesonide
    • Nikander K, Agertoft L, and Pedersen S: Breath-synchronized nebulization diminishes the impact of patient-device interfaces (face mask or mouthpiece) on the inhaled mass of nebulized budesonide. J Asthma. 2000;37:451-459.
    • (2000) J. Asthma. , vol.37 , pp. 451-459
    • Nikander, K.1    Agertoft, L.2    Pedersen, S.3
  • 57
    • 4644324201 scopus 로고    scopus 로고
    • Aerosol deposition in the extrathoracic region
    • Chen YS: Aerosol deposition in the extrathoracic region. Aerosol Sci Technol. 2003;37:659-671.
    • (2003) Aerosol. Sci. Technol. , vol.37 , pp. 659-671
    • Chen, Y.S.1
  • 58
    • 11344277827 scopus 로고    scopus 로고
    • Particle deposition in human nasal airway replicas manufactured by different methods. Part I: Inertial regime particles
    • Kelly JT, Asgharian B, Kimbell JS, and Wong B: Particle deposition in human nasal airway replicas manufactured by different methods. Part I: inertial regime particles. Aerosol Sci Technol. 2004;38:1063-1071.
    • (2004) Aerosol. Sci. Technol. , vol.38 , pp. 1063-1071
    • Kelly, J.T.1    Asgharian, B.2    Kimbell, J.S.3    Wong, B.4
  • 62
    • 33644878657 scopus 로고    scopus 로고
    • Ipratropiumbromide- induced acute anisocoria in the intensive care setting due to ill-fitting face masks
    • Bisquerra RA, Botz GH, and Nates JL: Ipratropiumbromide- induced acute anisocoria in the intensive care setting due to ill-fitting face masks. Respir Care. 2005;50: 1662-1664.
    • (2005) Respir. Care , vol.50 , pp. 1662-1664
    • Bisquerra, R.A.1    Botz, G.H.2    Nates, J.L.3
  • 63
    • 79953148714 scopus 로고    scopus 로고
    • A preliminary randomized controlled trial to assess effectiveness of nasal high-flow oxygen in intensive care patients
    • Parke RL, McGuinness SP, and Eccleston ML: A preliminary randomized controlled trial to assess effectiveness of nasal high-flow oxygen in intensive care patients. Respir Care. 2011;56:265-270.
    • (2011) Respir. Care , vol.56 , pp. 265-270
    • Parke, R.L.1    McGuinness, S.P.2    Eccleston, M.L.3
  • 66
    • 77955469508 scopus 로고    scopus 로고
    • Evaluation of enhanced condensational growth (ECG) for controlled respiratory drug delivery in a mouth-throat and upper tracheobronchial model
    • Hindle M, and Longest PW: Evaluation of enhanced condensational growth (ECG) for controlled respiratory drug delivery in a mouth-throat and upper tracheobronchial model. Pharmaceut Res. 2010;27:1800-1811.
    • (2010) Pharmaceut Res. , vol.27 , pp. 1800-1811
    • Hindle, M.1    Longest, P.W.2
  • 67
    • 79953890457 scopus 로고    scopus 로고
    • Improving the lung delivery of nasally administered aerosols during noninvasive ventilation-an application of enhanced condensational growth (ECG)
    • Longest PW, Tian G, and Hindle M: Improving the lung delivery of nasally administered aerosols during noninvasive ventilation-an application of enhanced condensational growth (ECG). J Aerosol Med Pulm Drug Deliv. 2011;24:103-118.
    • (2011) J. Aerosol. Med. Pulm. Drug. Deliv. , vol.24 , pp. 103-118
    • Longest, P.W.1    Tian, G.2    Hindle, M.3
  • 68
    • 0029081287 scopus 로고
    • Albuterol delivery in a model of mechanical ventilation. Comparison of metereddose inhaler and nebulizer efficiency
    • Diot P, Morra L, and Smaldone GC: Albuterol delivery in a model of mechanical ventilation. Comparison of metereddose inhaler and nebulizer efficiency. Am J Respir Crit Care Med. 1995;152:1391-1394.
    • (1995) Am. J. Respir. Crit. Care Med. , vol.152 , pp. 1391-1394
    • Diot, P.1    Morra, L.2    Smaldone, G.C.3
  • 69
    • 0025647073 scopus 로고
    • Metered dose inhaler aerosol characteristics are affected by the endotracheal tube actuator/adapter used
    • Bishop MJ, Larson RP, and Buschman DL: Metered dose inhaler aerosol characteristics are affected by the endotracheal tube actuator/adapter used. Anesthesiology. 1990;73: 1263-1265.
    • (1990) Anesthesiology , vol.73 , pp. 1263-1265
    • Bishop, M.J.1    Larson, R.P.2    Buschman, D.L.3
  • 70
    • 0026782880 scopus 로고
    • Evaluation of a reservoir device for metered-dose bronchodilator delivery to intubated adults
    • Rau JL, Harwood RJ, and Groff JL: Evaluation of a reservoir device for metered-dose bronchodilator delivery to intubated adults. An in vitro study. Chest. 1992; 102:924-930.
    • (1992) An. In. Vitro. Study. Chest. , vol.102 , pp. 924-930
    • Rau, J.L.1    Harwood, R.J.2    Groff, J.L.3
  • 71
    • 0028147668 scopus 로고
    • Efficiency of bronchodilator aerosol delivery to the lungs om the metered dose inhaler in mechanically ventilated patients. A study comparing four different actuator devices
    • Fuller HD, Dolovich MB, Turpie FH, and Newhouse MT: Efficiency of bronchodilator aerosol delivery to the lungs om the metered dose inhaler in mechanically ventilated patients. A study comparing four different actuator devices. Chest. 1994;105:214-218.
    • (1994) Chest , vol.105 , pp. 214-218
    • Fuller, H.D.1    Dolovich, M.B.2    Turpie, F.H.3    Newhouse, M.T.4
  • 72
    • 0029048880 scopus 로고
    • Bronchodilator delivery by metered-dose inhaler in ventilator-supported patients
    • Dhand R, Jubran A, and Tobin MJ: Bronchodilator delivery by metered-dose inhaler in ventilator-supported patients. Am J Respir Crit Care Med. 1995;151:1827-1833.
    • (1995) Am. J. Respir. Crit. Care Med. , vol.151 , pp. 1827-1833
    • Dhand, R.1    Jubran, A.2    Tobin, M.J.3
  • 74
    • 0032438394 scopus 로고    scopus 로고
    • Improved airway targeting with the CFC-free HFA-beclomethasone metered-dose inhaler compared with the CFC-beclomethasone
    • Leach CL, Davidson PJ, and Boudreau RJ: Improved airway targeting with the CFC-free HFA-beclomethasone metered-dose inhaler compared with the CFC-beclomethasone. Eur Respir J. 1998;12:1346-1353.
    • (1998) Eur. Respir. J. , vol.12 , pp. 1346-1353
    • Leach, C.L.1    Davidson, P.J.2    Boudreau, R.J.3
  • 75
    • 79952763932 scopus 로고    scopus 로고
    • Aerosol drug delivery: Developments in device design and clinical use
    • Dolovich MB, and Dhand R: Aerosol drug delivery: developments in device design and clinical use. Lancet. 2011;377:1032-1045.
    • (2011) Lancet , vol.377 , pp. 1032-1045
    • Dolovich, M.B.1    Dhand, R.2
  • 77
    • 0031983971 scopus 로고    scopus 로고
    • Salbutamol delivery froma hydrofluoralkane pressurized metered-dose inhaler in pediatric ventilator circuits
    • Wildhaber JH, Hayden MJ, Dore ND, Devadason SG, and LeSou ëf PN: Salbutamol delivery froma hydrofluoralkane pressurized metered-dose inhaler in pediatric ventilator circuits. Chest. 1998;113:186-191.
    • (1998) Chest , vol.113 , pp. 186-191
    • Wildhaber, J.H.1    Hayden, M.J.2    Dore, N.D.3    Devadason, S.G.4    Lesouëf, P.N.5
  • 78
    • 0035076815 scopus 로고    scopus 로고
    • Albuterol delivery in a neonatal ventilated lung model: Nebulization versus chlorofluorocarbon- and hydrofluoroalkane- pressurized metered dose inhalers
    • Lugo RA, Kenney JK, Keenan J, Salyer JW, Ballard J, and Ward RM: Albuterol delivery in a neonatal ventilated lung model: nebulization versus chlorofluorocarbon- and hydrofluoroalkane- pressurized metered dose inhalers. Pediatr Pulmonol 2001;31:247-254.
    • (2001) Pediatr. Pulmonol , vol.31 , pp. 247-254
    • Lugo, R.A.1    Kenney, J.K.2    Keenan, J.3    Salyer, J.W.4    Ballard, J.5    Ward, R.M.6
  • 79
    • 1542543550 scopus 로고    scopus 로고
    • The delivery of chlorofluorocarbon-propelled versus hydrofluoroalkane- propelled beclomethasone dipropionate aerosol to the mechanically ventilated patient: A laboratory study
    • Mitchell JP, Nagel MW, Wiersema KJ, Doyle CC, and Migounov VA: The delivery of chlorofluorocarbon-propelled versus hydrofluoroalkane-propelled beclomethasone dipropionate aerosol to the mechanically ventilated patient: a laboratory study. Respir Care. 2003;48:1025-1032.
    • (2003) Respir. Care , vol.48 , pp. 1025-1032
    • Mitchell, J.P.1    Nagel, M.W.2    Wiersema, K.J.3    Doyle, C.C.4    Migounov, V.A.5
  • 80
    • 33644873498 scopus 로고    scopus 로고
    • Albuterol delivery during noninvasive ventilation
    • Branconnier MP, and Hess DR: Albuterol delivery during noninvasive ventilation. Respir Care. 2005; 50:1649-1653.
    • (2005) Respir. Care , vol.50 , pp. 1649-1653
    • Branconnier, M.P.1    Hess, D.R.2
  • 82
    • 0026728835 scopus 로고
    • Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Effect of ventilator settings and nebulizer type, position, and volume of fill
    • ODoherty MJ, Thomas SH, Page CJ, Treacher DF, and Nunan TO: Delivery of a nebulized aerosol to a lung model during mechanical ventilation. Effect of ventilator settings and nebulizer type, position, and volume of fill. Am Rev Respir Dis. 1992;146:383-388.
    • (1992) Am. Rev. Respir. Dis. , vol.146 , pp. 383-388
    • Odoherty, M.J.1    Thomas, S.H.2    Page, C.J.3    Treacher, D.F.4    Nunan, T.O.5
  • 83
    • 0023473104 scopus 로고
    • Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency
    • Hughes JM, Saez J: Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency. Respir Care. 1987;32:1131-1135.
    • (1987) Respir. Care , vol.32 , pp. 1131-1135
    • Hughes, J.M.1    Saez, J.2
  • 84
    • 77955480004 scopus 로고    scopus 로고
    • Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified. circuits during adult mechanical ventilation
    • Ari A, Areabi H, and Fink JB: Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified. circuits during adult mechanical ventilation. Respir Care. 2010;55:837-844.
    • (2010) Respir. Care , vol.55 , pp. 837-844
    • Ari, A.1    Areabi, H.2    Fink, J.B.3
  • 85
    • 0029763316 scopus 로고    scopus 로고
    • Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand
    • Hess D, Fisher D, Williams P, Pooler S, and Kacmarek RM: Medication nebulizer performance. Effects of diluent volume, nebulizer flow, and nebulizer brand. Chest. 1996;110: 498-505.
    • (1996) Chest , vol.110 , pp. 498-505
    • Hess, D.1    Fisher, D.2    Williams, P.3    Pooler, S.4    Kacmarek, R.M.5
  • 86
    • 0036835672 scopus 로고    scopus 로고
    • Design principles of liquid nebulization devices currently in use
    • 1257-1275; discussion
    • Rau JL: Design principles of liquid nebulization devices currently in use. Respir Care. 2002;47:1257-1275; discussion 1275-1278.
    • (2002) Respir. Care , vol.47 , pp. 1275-1278
    • Rau, J.L.1
  • 88
    • 77954909965 scopus 로고    scopus 로고
    • In-vitro characterization of the nebulized dose during non-invasive ventilation
    • Abdelrahim ME, Plant P, and Chrystyn H: In-vitro characterization of the nebulized dose during non-invasive ventilation. J Pharm Pharmacol. 2010;62:966-972.
    • (2010) J. Pharm. Pharmacol. , vol.62 , pp. 966-972
    • Abdelrahim, M.E.1    Plant, P.2    Chrystyn, H.3
  • 89
    • 0028911383 scopus 로고
    • Testing of dry powder aerosol formulations in different environmental conditions
    • Jashnani RN, Byron PR, and Dalby RN: Testing of dry powder aerosol formulations in different environmental conditions. Int J Pharm. 1995;113:123-130.
    • (1995) Int. J. Pharm. , vol.113 , pp. 123-130
    • Jashnani, R.N.1    Byron, P.R.2    Dalby, R.N.3
  • 90
    • 0028040181 scopus 로고
    • Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery
    • ORiordan TG, Palmer LB, and Smaldone GC: Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery. Am J Respir Crit Care Med. 1994;149:214-219.
    • (1994) Am. J. Respir. Crit. Care Med. , vol.149 , pp. 214-219
    • Oriordan, T.G.1    Palmer, L.B.2    Smaldone, G.C.3
  • 91
    • 34548224200 scopus 로고    scopus 로고
    • Effect of continuous positive airway pressure combined to nebulization on lung deposition measured by urinary excretion of amikacin
    • Reychler G, Leal T, Roeseler J, Thys F, Delvau N, and Liistro G: Effect of continuous positive airway pressure combined to nebulization on lung deposition measured by urinary excretion of amikacin. Respir Med. 2007;101:2051-2055.
    • (2007) Respir. Med. , vol.101 , pp. 2051-2055
    • Reychler, G.1    Leal, T.2    Roeseler, J.3    Thys, F.4    Delvau, N.5    Liistro, G.6
  • 92
    • 0025290635 scopus 로고
    • Droplets produced by medical nebulizers. Some factors affecting their size and solute concentration
    • Phipps PR, and Gonda I: Droplets produced by medical nebulizers. Some factors affecting their size and solute concentration. Chest. 1990;97:1327-1332.
    • (1990) Chest , vol.97 , pp. 1327-1332
    • Phipps, P.R.1    Gonda, I.2
  • 93
    • 33644841077 scopus 로고    scopus 로고
    • Positive expiratory pressure changes aerosol distribution in patients with cystic fibrosis
    • Laube BL, Geller DE, Lin TC, Dalby RN, West-Diener M, and Zeitlin PL: Positive expiratory pressure changes aerosol distribution in patients with cystic fibrosis. Respir Care. 2005;50:1438-1444.
    • (2005) Respir. Care , vol.50 , pp. 1438-1444
    • Laube, B.L.1    Geller, D.E.2    Lin, T.C.3    Dalby, R.N.4    West-Diener, M.5    Zeitlin, P.L.6
  • 94
    • 0036852058 scopus 로고    scopus 로고
    • In vitro evaluation of aerosol bronchodilator delivery during noninvasive positive pressure ventilation: Effect of ventilator settings and nebulizer position
    • Chatmongkolchart S, Schettino GP, Dillman C, Kacmarek RM, and Hess DR: In vitro evaluation of aerosol bronchodilator delivery during noninvasive positive pressure ventilation: effect of ventilator settings and nebulizer position. Crit Care Med. 2002;30:2515-2519.
    • (2002) Crit. Care Med. , vol.30 , pp. 2515-2519
    • Chatmongkolchart, S.1    Schettino, G.P.2    Dillman, C.3    Kacmarek, R.M.4    Hess, D.R.5
  • 95
    • 0041666298 scopus 로고    scopus 로고
    • Maximizing aerosol delivery during mechanical ventilation: Go with the flow and go slow
    • Dhand R: Maximizing aerosol delivery during mechanical ventilation: go with the flow and go slow. Intensive Care Med. 2003;29:1041-1042.
    • (2003) Intensive. Care Med. , vol.29 , pp. 1041-1042
    • Dhand, R.1
  • 96
    • 0242552206 scopus 로고    scopus 로고
    • Aerosol therapy during mechanical ventilation: Getting ready for prime time
    • Dhand R: Aerosol therapy during mechanical ventilation: getting ready for prime time. Am J Respir Crit Care Med. 2003;168:1148-1149.
    • (2003) Am. J. Respir. Crit. Care Med. , vol.168 , pp. 1148-1149
    • Dhand, R.1
  • 97
    • 0033930984 scopus 로고    scopus 로고
    • Influence of inspiratory flow rate, particle size, and airway caliber on aerosolized drug delivery to the lung
    • Dolovich MA: Influence of inspiratory flow rate, particle size, and airway caliber on aerosolized drug delivery to the lung. Respir Care. 2000;45:597-608.
    • (2000) Respir. Care , vol.45 , pp. 597-608
    • Dolovich, M.A.1
  • 98
    • 0024567059 scopus 로고
    • Homogeneity of bronchopulmonary distribution of 99mTc aerosol in normal subjects and in cystic fibrosis patients
    • Laube BL, Links JM, LaFrance ND, Wagner HN, and Rosenstein BJ: Homogeneity of bronchopulmonary distribution of 99mTc aerosol in normal subjects and in cystic fibrosis patients. Chest. 1989;95:822-830.
    • (1989) Chest , vol.95 , pp. 822-830
    • Laube, B.L.1    Links, J.M.2    Lafrance, N.D.3    Wagner, H.N.4    Rosenstein, B.J.5
  • 99
    • 33846604218 scopus 로고    scopus 로고
    • Assessing new technologies: Patient-device interactions and deposition
    • Smaldone GC: Assessing new technologies: patient-device interactions and deposition. Respir Care. 2005;50:1151-1160.
    • (2005) Respir. Care , vol.50 , pp. 1151-1160
    • Smaldone, G.C.1
  • 100
    • 0025190161 scopus 로고
    • Continuous positive airway pressure reduces work of breathing and dyspnea during weaning from mechanical ventilation in severe chronic obstructive pulmonary disease
    • Petrof BJ, Legare M, Goldberg P, Milic-Emili J, and Gottfried SB: Continuous positive airway pressure reduces work of breathing and dyspnea during weaning from mechanical ventilation in severe chronic obstructive pulmonary disease. Am Rev Respir Dis. 1990;141:281-289.
    • (1990) Am. Rev. Respir. Dis. , vol.141 , pp. 281-289
    • Petrof, B.J.1    Legare, M.2    Goldberg, P.3    Milic-Emili, J.4    Gottfried, S.B.5
  • 102
    • 0015899897 scopus 로고
    • The fate of nebulized salbutamol (albuterol) administered by intermittent positive pressure respiration to asthmatic patients
    • Shenfield GM, Evans ME, Walker SR, and Paterson JW: The fate of nebulized salbutamol (albuterol) administered by intermittent positive pressure respiration to asthmatic patients. Am Rev Respir Dis. 1973;108:501-505.
    • (1973) Am. Rev. Respir. Dis. , vol.108 , pp. 501-505
    • Shenfield, G.M.1    Evans, M.E.2    Walker, S.R.3    Paterson, J.W.4
  • 103
    • 0017362241 scopus 로고
    • Pulmonary aerosol deposition in chronic bronchitis: Intermittent positive pressure breathing versus quiet breathing
    • Dolovich MB, Killian D, Wolff RK, Obminski G, and Newhouse MT: Pulmonary aerosol deposition in chronic bronchitis: intermittent positive pressure breathing versus quiet breathing. Am Rev Respir Dis. 1977;115:397-402.
    • (1977) Am. Rev. Respir. Dis. , vol.115 , pp. 397-402
    • Dolovich, M.B.1    Killian, D.2    Wolff, R.K.3    Obminski, G.4    Newhouse, M.T.5
  • 106
    • 0028040181 scopus 로고
    • Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery
    • ORiordan TG, Palmer LB, and Smaldone GC: Aerosol deposition in mechanically ventilated patients. Optimizing nebulizer delivery. Am J Respir Crit Care Med. 1994;149: 214-219.
    • (1994) Am. J. Respir. Crit. Care Med. , vol.149 , pp. 214-219
    • Oriordan, T.G.1    Palmer, L.B.2    Smaldone, G.C.3
  • 107
    • 33745629370 scopus 로고    scopus 로고
    • Comparison of lung deposition of amikacin by intrapulmonary percussive ventilation and jet nebulization by urinary monitoring
    • Reychler G, Wallemacq P, Rodenstein DO, Cumps J, Leal T, and Liistro G: Comparison of lung deposition of amikacin by intrapulmonary percussive ventilation and jet nebulization by urinary monitoring. J Aerosol Med. 2006;19:199-207.
    • (2006) J. Aerosol. Med. , vol.19 , pp. 199-207
    • Reychler, G.1    Wallemacq, P.2    Rodenstein, D.O.3    Cumps, J.4    Leal, T.5    Liistro, G.6
  • 109
    • 0026331545 scopus 로고
    • Treatment of bronchial asthma with terbutaline inhaled by conespacer combined with positive expiratory pressure mask
    • Frischknecht-Christensen E, Norregard O, and Dahl R: Treatment of bronchial asthma with terbutaline inhaled by conespacer combined with positive expiratory pressure mask. Chest. 1991;100:317-321.
    • (1991) Chest , vol.100 , pp. 317-321
    • Frischknecht-Christensen, E.1    Norregard, O.2    Dahl, R.3
  • 110
    • 0028842697 scopus 로고
    • Treatment of acute bronchospasm with beta-adrenergic agonist aerosols delivered by a nasal bilevel positive airway pressure circuit
    • Pollack CV Jr, Fleisch KB, and Dowsey K: Treatment of acute bronchospasm with beta-adrenergic agonist aerosols delivered by a nasal bilevel positive airway pressure circuit. Ann Emerg Med. 1995;26:552-557
    • (1995) Ann. Emerg. Med. , vol.26 , pp. 552-557
    • Pollack, Jr.C.V.1    Fleisch, K.B.2    Dowsey, K.3
  • 111
    • 0037390456 scopus 로고    scopus 로고
    • A pilot prospective, randomized, placebo-controlled trial of bilevel positive airway pressure in acute asthmatic attack
    • Soroksky A, Stav D, Shpirer I. A pilot prospective, randomized, placebo-controlled trial of bilevel positive airway pressure in acute asthmatic attack. Chest. 2003;123:1018-1025.
    • (2003) Chest , vol.123 , pp. 1018-1025
    • Soroksky, A.1    Stav, D.2    Shpirer, I.3
  • 112
    • 67649826855 scopus 로고    scopus 로고
    • Reversal of bronchial obstruction with bi-level positive airway pressure and nebulization in patients with acute asthma
    • Brandao DC, Lima VM, Filho VG, Silva TS, Campos TF, Dean E, and De Andrade AD: Reversal of bronchial obstruction with bi-level positive airway pressure and nebulization in patients with acute asthma. J Asthma. 2009;46: 356-361.
    • (2009) J. Asthma. , vol.46 , pp. 356-361
    • Brandao, D.C.1    Lima, V.M.2    Filho, V.G.3    Silva, T.S.4    Campos, T.F.5    Dean, E.6    De Andrade, A.D.7
  • 113
    • 77954589738 scopus 로고    scopus 로고
    • A prospective randomized controlled trial on the efficacy of noninvasive ventilation in severe acute asthma
    • Gupta D, Nath A, Agarwal R, and Behera D: A prospective randomized controlled trial on the efficacy of noninvasive ventilation in severe acute asthma. Respir Care 2010;55:536-543.
    • (2010) Respir. Care , vol.55 , pp. 536-543
    • Gupta, D.1    Nath, A.2    Agarwal, R.3    Behera, D.4
  • 114
    • 0021037448 scopus 로고
    • Intermittent Positive Pressure Breathing Trial Group: Intermittent positive pressure breathing therapy of chronic obstructive pulmonary disease: A clinical trial
    • Intermittent Positive Pressure Breathing Trial Group: Intermittent positive pressure breathing therapy of chronic obstructive pulmonary disease: a clinical trial. Ann Intern Med. 1983;99:612-620.
    • (1983) Ann. Intern. Med. , vol.99 , pp. 612-620
  • 115
    • 0034764689 scopus 로고    scopus 로고
    • Salbutamol delivery during non-invasive mechanical ventilation in patients with chronic obstructive pulmonary disease: A randomized, controlled study
    • Nava S, Karakurt S, Rampulla C, Braschi A, and Fanfulla F: Salbutamol delivery during non-invasive mechanical ventilation in patients with chronic obstructive pulmonary disease: a randomized, controlled study. Intensive Care Med 2001;27:1627-1635.
    • (2001) Intensive. Care Med , vol.27 , pp. 1627-1635
    • Nava, S.1    Karakurt, S.2    Rampulla, C.3    Braschi, A.4    Fanfulla, F.5
  • 116
    • 0029775579 scopus 로고    scopus 로고
    • Aerosol delivery from a metered-dose inhaler during. mechanical ventilation. An in vitro model
    • Fink JB, Dhand R, Duarte AG, Jenne JW, and Tobin MJ: Aerosol delivery from a metered-dose inhaler during. mechanical ventilation. An in vitro model. Am J Respir Crit Care Med. 1996;154:382-387.
    • (1996) Am. J. Respir. Crit. Care Med. , vol.154 , pp. 382-387
    • Fink, J.B.1    Dhand, R.2    Duarte, A.G.3    Jenne, J.W.4    Tobin, M.J.5
  • 117
    • 33846172670 scopus 로고    scopus 로고
    • Influence of respiratory efforts on b2- agonist induced bronchodilation in mechanically ventilated COPD patients: A prospective clinical study
    • Malliotakis P, Mouloudi E, Prinianakis G, Kondili E, and Georgopoulos D: Influence of respiratory efforts on b2- agonist induced bronchodilation in mechanically ventilated COPD patients: a prospective clinical study. Respir Med. 2007;101:300-307.
    • (2007) Respir. Med. , vol.101 , pp. 300-307
    • Malliotakis, P.1    Mouloudi, E.2    Prinianakis, G.3    Kondili, E.4    Georgopoulos, D.5
  • 118
    • 23944439929 scopus 로고    scopus 로고
    • The effects of nebulized salbutamol, external positive end-expiratory pressure, and their combination on respiratory mechanics, hemodynamics, and gas exchange in mechanically ventilated chronic obstructive pulmonary disease patients
    • Tzoufi M, Mentzelopoulos SD, Roussos C, and Armaganidis A: The effects of nebulized salbutamol, external positive end-expiratory pressure, and their combination on respiratory mechanics, hemodynamics, and gas exchange in mechanically ventilated chronic obstructive pulmonary disease patients. Anesth Analg. 2005;101:843-850.
    • (2005) Anesth. Analg. , vol.101 , pp. 843-850
    • Tzoufi, M.1    Mentzelopoulos, S.D.2    Roussos, C.3    Armaganidis, A.4
  • 123
    • 0023137604 scopus 로고
    • Systemic cardiovascular and metabolic effects associated with the inhalation of an increased dose of albuterol. Influence of mouth rinsing and gargling
    • Kung M, Croley SW, and Phillips BA: Systemic cardiovascular and metabolic effects associated with the inhalation of an increased dose of albuterol. Influence of mouth rinsing and gargling. Chest. 1987;91:382-387.
    • (1987) Chest , vol.91 , pp. 382-387
    • Kung, M.1    Croley, S.W.2    Phillips, B.A.3
  • 125
    • 0025054476 scopus 로고
    • Albuterol and spacerinduced atrial fibrillation
    • Breeden CC, and Safirstein BH: Albuterol and spacerinduced atrial fibrillation. Chest. 1990;98:762-763.
    • (1990) Chest , vol.98 , pp. 762-763
    • Breeden, C.C.1    Safirstein, B.H.2
  • 126
    • 0027445729 scopus 로고
    • A prospective, randomized study of continuous versus intermittent nebulized albuterol for severe status asthmaticus in children
    • Papo MC, Frank J, and Thompson AE: A prospective, randomized study of continuous versus intermittent nebulized albuterol for severe status asthmaticus in children. Crit Care Med. 1993;21:1479-1486.
    • (1993) Crit. Care Med. , vol.21 , pp. 1479-1486
    • Papo, M.C.1    Frank, J.2    Thompson, A.E.3
  • 127
    • 0029888165 scopus 로고    scopus 로고
    • Continuous vs. intermittent albuterol, at high and low doses, in the treatment of severe acute asthma in adults
    • Shrestha M, Bidadi K, Gourlay S, and Hayes J: Continuous vs. intermittent albuterol, at high and low doses, in the treatment of severe acute asthma in adults. Chest. 1996;110: 42-47.
    • (1996) Chest , vol.110 , pp. 42-47
    • Shrestha, M.1    Bidadi, K.2    Gourlay, S.3    Hayes, J.4
  • 128
    • 0027770791 scopus 로고
    • Metered- dose inhaler versus nebulized albuterol in mechanically ventilated patients
    • Manthous CA, Hall JB, Schmidt GA, and Wood LD: Metered- dose inhaler versus nebulized albuterol in mechanically ventilated patients. Am Rev Respir Dis. 1993;148: 1567-1570.
    • (1993) Am. Rev. Respir. Dis. , vol.148 , pp. 1567-1570
    • Manthous, C.A.1    Hall, J.B.2    Schmidt, G.A.3    Wood, L.D.4
  • 130
    • 0023229571 scopus 로고
    • Effects of an anticholinergic bronchodilator on arterial blood gases of hypoxemic patients with chronic obstructive pulmonary disease. Comparison with a beta-adrenergic agent
    • Gross NJ, and Bankwala Z. Effects of an anticholinergic bronchodilator on arterial blood gases of hypoxemic patients with chronic obstructive pulmonary disease. Comparison with a beta-adrenergic agent. Am Rev Respir Dis. 1987;136:1091-1094.
    • (1987) Am. Rev. Respir. Dis. , vol.136 , pp. 1091-1094
    • Gross, N.J.1    Bankwala, Z.2
  • 131
    • 0032863007 scopus 로고    scopus 로고
    • Effects of salmeterol on arterial blood gases in patients with stable chronic obstructive pulmonary disease. Comparison with albuterol and ipratropium
    • Khoukaz G, and Gross NJ: Effects of salmeterol on arterial blood gases in patients with stable chronic obstructive pulmonary disease. Comparison with albuterol and ipratropium. Am J Respir Crit Care Med. 1999;160:1028-1030.
    • (1999) Am. J. Respir. Crit. Care Med. , vol.160 , pp. 1028-1030
    • Khoukaz, G.1    Gross, N.J.2
  • 132
    • 0024551382 scopus 로고
    • Nebulized albuterol for acute hyperkalemia in patients on hemodialysis
    • Allon M, Dunlay R, and Copkney C. Nebulized albuterol for acute hyperkalemia in patients on hemodialysis. Ann Intern Med. 1989;110:426-429.
    • (1989) Ann. Intern. Med. , vol.110 , pp. 426-429
    • Allon, M.1    Dunlay, R.2    Copkney, C.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.