Evaluation of aerosol generator devices at 3 locations in humidified and non-humidified circuits during adult mechanical ventilation

Author keywords

Aerosol drug deposition; Aerosols; Drug administration; Heat; Humidity; Jet nebulizer; Mechanical ventilation; Metered dose inhaler; Nebulizers; pMDI; Ultrasonic nebulizer; Ventilator circuit; Vibrating mesh nebulizer

Indexed keywords

SALBUTAMOL SULFATE; BRONCHODILATING AGENT; SALBUTAMOL;

EID: 77955480004     PISSN: 00201324     EISSN: None     Source Type: Journal    
DOI: None     Document Type: Article
Times cited : (137)

References (26)
  • 1
    • 0033840918 scopus 로고    scopus 로고
    • Aerosol therapy in mechanically ventilated patients: Recent advances and new techniques
    • Fink J, Dhand R. Aerosol therapy in mechanically ventilated patients: recent advances and new techniques. Semin Respir Crit Care Med 2000;21(3):183-201.
    • (2000) Semin Respir Crit Care Med , vol.21 , Issue.3 , pp. 183-201
    • Fink, J.1    Dhand, R.2
  • 2
    • 0033027165 scopus 로고    scopus 로고
    • Bronchodilator therapy in mechanically ventilated patients
    • Fink J, Tobin M, Dhand R. Bronchodilator therapy in mechanically ventilated patients. Respir Care 1999;44(1):53-69.
    • (1999) Respir Care , vol.44 , Issue.1 , pp. 53-69
    • Fink, J.1    Tobin, M.2    Dhand, R.3
  • 3
    • 0035164687 scopus 로고    scopus 로고
    • Inhalation therapy during mechanical ventilation
    • Duarte AG, Fink JB, Dhand R. Inhalation therapy during mechanical ventilation. Respir Care Clin N Am 2001;7(2):233-260.
    • (2001) Respir Care Clin N Am , vol.7 , Issue.2 , pp. 233-260
    • Duarte, A.G.1    Fink, J.B.2    Dhand, R.3
  • 4
    • 0242552206 scopus 로고    scopus 로고
    • Aerosol therapy during mechanical ventilation: Getting ready for prime time
    • Dhand R. Aerosol therapy during mechanical ventilation: getting ready for prime time. Am J Respir Crit Care Med 2003;168(10):1148-1149.
    • (2003) Am J Respir Crit Care Med , vol.168 , Issue.10 , pp. 1148-1149
    • Dhand, R.1
  • 5
    • 0035165202 scopus 로고    scopus 로고
    • Laboratory evaluation of metered-dose inhalers with models that simulate interaction with the patient
    • Fink J, Dhand R. Laboratory evaluation of metered-dose inhalers with models that simulate interaction with the patient. Respir Care Clin N Am 2001;7(2):303-317.
    • (2001) Respir Care Clin N Am , vol.7 , Issue.2 , pp. 303-317
    • Fink, J.1    Dhand, R.2
  • 6
    • 85012519236 scopus 로고
    • Aerosol delivery during mechanical ventilation: A predictive in-vitro lung model
    • Fuller H, Dolovich M, Chambers C, Newhouse M. Aerosol delivery during mechanical ventilation: a predictive in-vitro lung model. J Aerosol Med 1992;5(4):251-259.
    • (1992) J Aerosol Med , vol.5 , Issue.4 , pp. 251-259
    • Fuller, H.1    Dolovich, M.2    Chambers, C.3    Newhouse, M.4
  • 7
    • 0041666298 scopus 로고    scopus 로고
    • Maximizing aerosol delivery during mechanical ventilation: Go with the flow and go slow
    • Dhand R. Maximizing aerosol delivery during mechanical ventilation: go with the flow and go slow. Intensive Care Med 2003;29(7):1041-1042.
    • (2003) Intensive Care Med , vol.29 , Issue.7 , pp. 1041-1042
    • Dhand, R.1
  • 8
    • 0027454256 scopus 로고
    • Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation
    • Thomas SH, O'Doherty MJ, Page CJ, Treacher DF, Nunan TO. Delivery of ultrasonic nebulized aerosols to a lung model during mechanical ventilation. Am Rev Respir Dis 1993;148(4):872-877.
    • (1993) Am Rev Respir Dis , vol.148 , Issue.4 , pp. 872-877
    • Thomas, S.H.1    O'Doherty, M.J.2    Page, C.J.3    Treacher, D.F.4    Nunan, T.O.5
  • 9
    • 0029081287 scopus 로고
    • Salbutamol delivery in a model of mechanical ventilation: Comparison of metered-dose inhaler and nebulizer efficiency
    • Diot P, Morra L, Smaldone G. Salbutamol delivery in a model of mechanical ventilation: comparison of metered-dose inhaler and nebulizer efficiency. Am J Respir Crit Care Med 1995;152(4):1391-1394.
    • (1995) Am J Respir Crit Care Med , vol.152 , Issue.4 , pp. 1391-1394
    • Diot, P.1    Morra, L.2    Smaldone, G.3
  • 10
    • 0029775579 scopus 로고    scopus 로고
    • Aerosol delivery from a metered-dose inhaler during mechanical ventilation. An in-vitro model
    • Fink J, Dhand R, Duarte A, Jenne J, Tobin M. Aerosol delivery from a metered-dose inhaler during mechanical ventilation. An in-vitro model. Am J Respir Crit Care Med 1996;154(2):382-387.
    • (1996) Am J Respir Crit Care Med , vol.154 , Issue.2 , pp. 382-387
    • Fink, J.1    Dhand, R.2    Duarte, A.3    Jenne, J.4    Tobin, M.5
  • 11
    • 0034113778 scopus 로고    scopus 로고
    • Overcoming the adverse effect of humidity in aerosol delivery via pressured metered-dose inhaler during mechanical ventilation
    • Lange C, Finlay W. Overcoming the adverse effect of humidity in aerosol delivery via pressured metered-dose inhaler during mechanical ventilation. Am J Respir Crit Care Med 2000;161(5):1614-1618.
    • (2000) Am J Respir Crit Care Med , vol.161 , Issue.5 , pp. 1614-1618
    • Lange, C.1    Finlay, W.2
  • 12
    • 0023473104 scopus 로고
    • Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency
    • Hughes J, Saez J. Effects of nebulizer mode and position in a mechanical ventilator circuit on dose efficiency. Respir Care 1987;32(12):1131-1135.
    • (1987) Respir Care , vol.32 , Issue.12 , pp. 1131-1135
    • Hughes, J.1    Saez, J.2
  • 13
    • 0026683191 scopus 로고
    • Effect of a new nebulizer position on aerosol delivery during mechanical ventilation
    • Quinn W. Effect of a new nebulizer position on aerosol delivery during mechanical ventilation. Respir Care 1992;37(5):423-431.
    • (1992) Respir Care , vol.37 , Issue.5 , pp. 423-431
    • Quinn, W.1
  • 14
    • 0026728835 scopus 로고
    • Delivery of a nebulized aerosol to a lung model during mechanical ventilation: Effect of ventilator settings and nebulizer type, position, and volume of fill
    • O'Doherty MJ, Thomas SH, Page CJ, Treacher DF, Nunan TO. Delivery of a nebulized aerosol to a lung model during mechanical ventilation: effect of ventilator settings and nebulizer type, position, and volume of fill. Am Rev Respir Dis 1992;146(2):383-388.
    • (1992) Am Rev Respir Dis , vol.146 , Issue.2 , pp. 383-388
    • O'Doherty, M.J.1    Thomas, S.H.2    Page, C.J.3    Treacher, D.F.4    Nunan, T.O.5
  • 15
    • 0028873969 scopus 로고
    • Effect of a spacer on pulmonary aerosol deposition from a jet nebulizer during mechanical ventilation
    • Harvey C, O'Doherty M, Page C, Thomas S, Nunan T, Treacher D. Effect of a spacer on pulmonary aerosol deposition from a jet nebulizer during mechanical ventilation. Thorax 1995;50(1):50-53.
    • (1995) Thorax , vol.50 , Issue.1 , pp. 50-53
    • Harvey, C.1    O'Doherty, M.2    Page, C.3    Thomas, S.4    Nunan, T.5    Treacher, D.6
  • 16
    • 0026782880 scopus 로고
    • Evaluation of a reservoir device for metered-dose bronchodilator delivery to intubated adults: An in-vitro study
    • Rau J, Harwood R, Groff J. Evaluation of a reservoir device for metered-dose bronchodilator delivery to intubated adults: an in-vitro study. Chest 1992;102(3):924-930.
    • (1992) Chest , vol.102 , Issue.3 , pp. 924-930
    • Rau, J.1    Harwood, R.2    Groff, J.3
  • 17
    • 0032900977 scopus 로고    scopus 로고
    • Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors
    • Fink JB, Dhand R, Grychowski J, Fahey PJ, Tobin MJ. Reconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors. Am J Respir Crit Care Med 1999;159(1):63-68.
    • (1999) Am J Respir Crit Care Med , vol.159 , Issue.1 , pp. 63-68
    • Fink, J.B.1    Dhand, R.2    Grychowski, J.3    Fahey, P.J.4    Tobin, M.J.5
  • 18
    • 0030887597 scopus 로고    scopus 로고
    • Comparison of jet and ultrasonic nebulizer pulmonary aerosol deposition during mechanical ventilation
    • Harvey C, O'Doherty M, Page C, Thomas S, Nunan T, Treacher D. Comparison of jet and ultrasonic nebulizer pulmonary aerosol deposition during mechanical ventilation. Eur Respir J 1997;10(4):905-909.
    • (1997) Eur Respir J , vol.10 , Issue.4 , pp. 905-909
    • Harvey, C.1    O'Doherty, M.2    Page, C.3    Thomas, S.4    Nunan, T.5    Treacher, D.6
  • 19
    • 0029781461 scopus 로고    scopus 로고
    • Dose response to bronchodilator delivered by metered-dose inhaler in ventilator supported patients
    • Dhand R, Duarte A, Jubran A, Jenne J, Fink J, Fahey R, Tobin M. Dose response to bronchodilator delivered by metered-dose inhaler in ventilator supported patients. Am J Respir Crit Care Med 1996;154(2):388-393.
    • (1996) Am J Respir Crit Care Med , vol.154 , Issue.2 , pp. 388-393
    • Dhand, R.1    Duarte, A.2    Jubran, A.3    Jenne, J.4    Fink, J.5    Fahey, R.6    Tobin, M.7
  • 20
    • 0028040181 scopus 로고
    • Aerosol deposition in mechanically ventilated patients: Optimizing nebulizer delivery
    • O'Riordan T, Palmer L, Smaldone G. Aerosol deposition in mechanically ventilated patients: optimizing nebulizer delivery. Am J Respir Crit Care Med 1994;149(1):214-219.
    • (1994) Am J Respir Crit Care Med , vol.149 , Issue.1 , pp. 214-219
    • O'Riordan, T.1    Palmer, L.2    Smaldone, G.3
  • 23
    • 0028147668 scopus 로고
    • Efficiency of bronchodilator aerosol delivery to the lungs from metered-dose inhaler in mechanically ventilated patients: A study comparing four different actuator devices
    • Fuller H, Dolovich M, Turpie F, Newhouse M. Efficiency of bronchodilator aerosol delivery to the lungs from metered-dose inhaler in mechanically ventilated patients: a study comparing four different actuator devices. Chest 1994;105(1):214-218.
    • (1994) Chest , vol.105 , Issue.1 , pp. 214-218
    • Fuller, H.1    Dolovich, M.2    Turpie, F.3    Newhouse, M.4
  • 24
    • 0027770791 scopus 로고
    • Metered-dose inhaler versus nebulized salbutamol in mechanically ventilated patients
    • Manthous C, Hall C, Schmidt G, Wood L. Metered-dose inhaler versus nebulized salbutamol in mechanically ventilated patients. Am Rev Respir Dis 1993;148(6):1567-1570.
    • (1993) Am Rev Respir Dis , vol.148 , Issue.6 , pp. 1567-1570
    • Manthous, C.1    Hall, C.2    Schmidt, G.3    Wood, L.4
  • 25
    • 0034223152 scopus 로고    scopus 로고
    • Bronchodilator therapy with metered dose inhaler and spacer versus nebulizer in mechanically ventilated patients: Comparison of magnitude and duration of response
    • Duarte A, Momi K, Bidani A. Bronchodilator therapy with metered dose inhaler and spacer versus nebulizer in mechanically ventilated patients: comparison of magnitude and duration of response. Respir Care 2000;45(7):817-823.
    • (2000) Respir Care , vol.45 , Issue.7 , pp. 817-823
    • Duarte, A.1    Momi, K.2    Bidani, A.3
  • 26
    • 0030466295 scopus 로고    scopus 로고
    • Serum salbutamol levels after metered-dose inhaler administration to ventilated patients and healthy controls
    • Duarte A, Dhand R, Reid D, Tobin M, Jenne J. Serum salbutamol levels after metered-dose inhaler administration to ventilated patients and healthy controls. Am J Respir Crit Care Med 1996;154(6):1658-1663.
    • (1996) Am J Respir Crit Care Med , vol.154 , Issue.6 , pp. 1658-1663
    • Duarte, A.1    Dhand, R.2    Reid, D.3    Tobin, M.4    Jenne, J.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.