Dosimetric properties of an amorphous-silicon EPID used in continuous acquisition mode for application to dynamic and arc IMRT

Author keywords

Continuous acquisition mode; EPID dosimetry; Exit dosimetry; Portal image dosimetry; Real time dose verification

Indexed keywords

AMORPHOUS SILICON; DOSIMETRY; IONIZATION CHAMBERS; PIXELS; RADIOTHERAPY;

EID: 67649622802     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3148822     Document Type: Article
Times cited : (75)

References (31)
  • 1
    • 34247095688 scopus 로고    scopus 로고
    • Investigation of an amorphous silicon EPID for measurement and quality assurance of enhanced dynamic wedge
    • 0031-9155,. 10.1088/0031-9155/52/4/014
    • P. B. Greer and M. P. Barnes, " Investigation of an amorphous silicon EPID for measurement and quality assurance of enhanced dynamic wedge.," Phys. Med. Biol. 0031-9155 52, 1075-1087 (2007). 10.1088/0031-9155/52/4/014
    • (2007) Phys. Med. Biol. , vol.52 , pp. 1075-1087
    • Greer, P.B.1    Barnes, M.P.2
  • 2
    • 0038033842 scopus 로고    scopus 로고
    • Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy
    • 0094-2405,. 10.1118/1.1582469
    • P. B. Greer and C. C. Popescu, " Dosimetric properties of an amorphous silicon electronic portal imaging device for verification of dynamic intensity modulated radiation therapy.," Med. Phys. 0094-2405 30, 1618-1627 (2003). 10.1118/1.1582469
    • (2003) Med. Phys. , vol.30 , pp. 1618-1627
    • Greer, P.B.1    Popescu, C.C.2
  • 3
    • 1242271376 scopus 로고    scopus 로고
    • Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device
    • 0094-2405,. 10.1118/1.1637969
    • L. N. McDermott, R. J. W. Louwe, J. J. Sonke, M. B. Van Herk, and B. J. Mijnheer, " Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device.," Med. Phys. 0094-2405 31, 285-295 (2004). 10.1118/1.1637969
    • (2004) Med. Phys. , vol.31 , pp. 285-295
    • McDermott, L.N.1    Louwe, R.J.W.2    Sonke, J.J.3    Van Herk, M.B.4    Mijnheer, B.J.5
  • 4
    • 0031860585 scopus 로고    scopus 로고
    • X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays
    • 0094-2405,. 10.1118/1.598252
    • P. Munro and D. C. Bouius, " X-ray quantum limited portal imaging using amorphous silicon flat-panel arrays.," Med. Phys. 0094-2405 25, 689-702 (1998). 10.1118/1.598252
    • (1998) Med. Phys. , vol.25 , pp. 689-702
    • Munro, P.1    Bouius, D.C.2
  • 5
    • 27244439568 scopus 로고    scopus 로고
    • Dose-response characteristics of an amorphous silicon EPID
    • 0094-2405,. 10.1118/1.2040711
    • P. Winkler, A. Hefner, and D. Georg, " Dose-response characteristics of an amorphous silicon EPID.," Med. Phys. 0094-2405 32, 3095-3105 (2005). 10.1118/1.2040711
    • (2005) Med. Phys. , vol.32 , pp. 3095-3105
    • Winkler, P.1    Hefner, A.2    Georg, D.3
  • 6
    • 33748303072 scopus 로고    scopus 로고
    • Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans
    • 0094-2405,. 10.1118/1.2241997
    • W. Ansbacher, " Three-dimensional portal image-based dose reconstruction in a virtual phantom for rapid evaluation of IMRT plans.," Med. Phys. 0094-2405 33, 3369-3382 (2006). 10.1118/1.2241997
    • (2006) Med. Phys. , vol.33 , pp. 3369-3382
    • Ansbacher, W.1
  • 7
    • 17144426504 scopus 로고    scopus 로고
    • Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration
    • 0031-9155,. 10.1088/0031-9155/50/7/003
    • S. J. K. Baker, G. J. Budgell, and R. I. MacKay, " Use of an amorphous silicon electronic portal imaging device for multileaf collimator quality control and calibration.," Phys. Med. Biol. 0031-9155 50, 1377-1392 (2005). 10.1088/0031-9155/50/7/003
    • (2005) Phys. Med. Biol. , vol.50 , pp. 1377-1392
    • Baker, S.J.K.1    Budgell, G.J.2    MacKay, R.I.3
  • 8
    • 28044439624 scopus 로고    scopus 로고
    • Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager
    • 0094-2405,. 10.1118/1.2074227
    • G. J. Budgell, Q. Zhang, R. J. Trouncer, and R. I. Mackay, " Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager.," Med. Phys. 0094-2405 32, 3267-3278 (2005). 10.1118/1.2074227
    • (2005) Med. Phys. , vol.32 , pp. 3267-3278
    • Budgell, G.J.1    Zhang, Q.2    Trouncer, R.J.3    MacKay, R.I.4
  • 9
    • 38049169060 scopus 로고    scopus 로고
    • Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle
    • 0031-9155,. 10.1088/0031-9155/53/2/013
    • M. F. Clarke and G. J. Budgell, " Use of an amorphous silicon EPID for measuring MLC calibration at varying gantry angle.," Phys. Med. Biol. 0031-9155 53, 473-485 (2008). 10.1088/0031-9155/53/2/013
    • (2008) Phys. Med. Biol. , vol.53 , pp. 473-485
    • Clarke, M.F.1    Budgell, G.J.2
  • 10
    • 34147152661 scopus 로고    scopus 로고
    • Patient-specific dosimetry of conventional and intensity modulated radiation therapy using a novel full Monte Carlo phase space reconstruction method from electronic portal images
    • 0031-9155,. 10.1088/0031-9155/52/8/016
    • G. Jarry and F. Verhaegen, " Patient-specific dosimetry of conventional and intensity modulated radiation therapy using a novel full Monte Carlo phase space reconstruction method from electronic portal images.," Phys. Med. Biol. 0031-9155 52, 2277-2299 (2007). 10.1088/0031-9155/52/8/016
    • (2007) Phys. Med. Biol. , vol.52 , pp. 2277-2299
    • Jarry, G.1    Verhaegen, F.2
  • 11
    • 0034977377 scopus 로고    scopus 로고
    • Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device
    • 0094-2405,. 10.1118/1.1374244
    • B. M. C. McCurdy, K. Luchka, and S. Pistorius, " Dosimetric investigation and portal dose image prediction using an amorphous silicon electronic portal imaging device.," Med. Phys. 0094-2405 28, 911-924 (2001). 10.1118/1.1374244
    • (2001) Med. Phys. , vol.28 , pp. 911-924
    • McCurdy, B.M.C.1    Luchka, K.2    Pistorius, S.3
  • 12
    • 33749393203 scopus 로고    scopus 로고
    • Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification
    • 0094-2405,. 10.1118/1.2230810
    • L. N. McDermott, M. Wendling, B. Van Asselen, J. Stroom, J. J. Sonke, M. Van Herk, and B. J. Mijnheer, " Clinical experience with EPID dosimetry for prostate IMRT pre-treatment dose verification.," Med. Phys. 0094-2405 33, 3921-3930 (2006). 10.1118/1.2230810
    • (2006) Med. Phys. , vol.33 , pp. 3921-3930
    • McDermott, L.N.1    Wendling, M.2    Van Asselen, B.3    Stroom, J.4    Sonke, J.J.5    Van Herk, M.6    Mijnheer, B.J.7
  • 13
    • 33746606839 scopus 로고    scopus 로고
    • GLAaS: An absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications
    • 0094-2405,. 10.1118/1.2218314
    • G. Nicolini, A. Fogliata, E. Vanetti, A. Clivio, and L. Cozzi, " GLAaS: An absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications.," Med. Phys. 0094-2405 33, 2839-2851 (2006). 10.1118/1.2218314
    • (2006) Med. Phys. , vol.33 , pp. 2839-2851
    • Nicolini, G.1    Fogliata, A.2    Vanetti, E.3    Clivio, A.4    Cozzi, L.5
  • 14
    • 34249737441 scopus 로고    scopus 로고
    • Amorphous silicon EPID calibration for dosimetric applications: Comparison of a method based on Monte Carlo prediction of response with existing techniques
    • 0031-9155,. 10.1088/0031-9155/52/12/003
    • L. Parent, A. L. Fielding, D. R. Dance, J. Seco, and P. M. Evans, " Amorphous silicon EPID calibration for dosimetric applications: Comparison of a method based on Monte Carlo prediction of response with existing techniques.," Phys. Med. Biol. 0031-9155 52, 3351-3368 (2007). 10.1088/0031-9155/52/12/003
    • (2007) Phys. Med. Biol. , vol.52 , pp. 3351-3368
    • Parent, L.1    Fielding, A.L.2    Dance, D.R.3    Seco, J.4    Evans, P.M.5
  • 15
    • 0036703540 scopus 로고    scopus 로고
    • IMRT verification by three-dimensional dose reconstruction from portal beam measurements
    • 0094-2405,. 10.1118/1.1494988
    • M. Partridge, M. Ebert, and B. M. Hesse, " IMRT verification by three-dimensional dose reconstruction from portal beam measurements.," Med. Phys. 0094-2405 29, 1847-1858 (2002). 10.1118/1.1494988
    • (2002) Med. Phys. , vol.29 , pp. 1847-1858
    • Partridge, M.1    Ebert, M.2    Hesse, B.M.3
  • 17
    • 14644390214 scopus 로고    scopus 로고
    • Three-dimensional IMRT verification with a flat-panel EPID
    • 0094-2405,. 10.1118/1.1843471
    • S. Steciw, B. Warkentin, S. Rathee, and B. G. Fallone, " Three-dimensional IMRT verification with a flat-panel EPID.," Med. Phys. 0094-2405 32, 600-612 (2005). 10.1118/1.1843471
    • (2005) Med. Phys. , vol.32 , pp. 600-612
    • Steciw, S.1    Warkentin, B.2    Rathee, S.3    Fallone, B.G.4
  • 18
    • 1942478371 scopus 로고    scopus 로고
    • The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields
    • 0167-8140,. 10.1016/j.radonc.2004.02.018
    • A. Van Esch, T. Depuydt, and D. P. Huyskens, " The use of an aSi-based EPID for routine absolute dosimetric pre-treatment verification of dynamic IMRT fields.," Radiother. Oncol. 0167-8140 71, 223-234 (2004). 10.1016/j.radonc.2004.02.018
    • (2004) Radiother. Oncol. , vol.71 , pp. 223-234
    • Van Esch, A.1    Depuydt, T.2    Huyskens, D.P.3
  • 19
    • 31844436637 scopus 로고    scopus 로고
    • Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
    • 0094-2405,. 10.1118/1.2147744
    • M. Wendling, R. J. W. Louwe, L. N. McDermott, J. J. Sonke, M. Van Herk, and B. J. Mijnheer, " Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method.," Med. Phys. 0094-2405 33, 259-273 (2006). 10.1118/1.2147744
    • (2006) Med. Phys. , vol.33 , pp. 259-273
    • Wendling, M.1    Louwe, R.J.W.2    McDermott, L.N.3    Sonke, J.J.4    Van Herk, M.5    Mijnheer, B.J.6
  • 20
    • 34547115940 scopus 로고    scopus 로고
    • Implementation and validation of portal dosimetry with an amorphous silicon EPID in the energy range from 6 to 25 MV
    • 0031-9155,. 10.1088/0031-9155/52/15/N05
    • P. Winkler, A. Hefner, and D. Georg, " Implementation and validation of portal dosimetry with an amorphous silicon EPID in the energy range from 6 to 25 MV.," Phys. Med. Biol. 0031-9155 52, N355-N365 (2007). 10.1088/0031-9155/52/15/N05
    • (2007) Phys. Med. Biol. , vol.52
    • Winkler, P.1    Hefner, A.2    Georg, D.3
  • 21
    • 51449112251 scopus 로고    scopus 로고
    • Commissioning and quality assurance of RapidArc radiotherapy delivery system
    • 0360-3016,. 10.1016/j.ijrob2008.05.060
    • C. C. Ling, P. Zhang, Y. Archambault, J. Bocanek, G. Tang, and T. Losasso, " Commissioning and quality assurance of RapidArc radiotherapy delivery system.," Int. J. Radiat. Oncol., Biol., Phys. 0360-3016 72, 575-581 (2008). 10.1016/j.ijrobp.2008.05.060
    • (2008) Int. J. Radiat. Oncol., Biol., Phys. , vol.72 , pp. 575-581
    • Ling, C.C.1    Zhang, P.2    Archambault, Y.3    Bocanek, J.4    Tang, G.5    Losasso, T.6
  • 22
    • 53149135069 scopus 로고    scopus 로고
    • The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy
    • 10.1186/1748-717X-3-24 1748-717X
    • G. Nicolini, E. Vanetti, A. Clivio, A. Fogliata, S. Korreman, J. Bocanek, and L. Cozzi, " The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy.," Radiat. Oncol. 3, 24 (2008). 10.1186/1748-717X-3-24 1748-717X
    • (2008) Radiat. Oncol. , vol.3 , pp. 24
    • Nicolini, G.1    Vanetti, E.2    Clivio, A.3    Fogliata, A.4    Korreman, S.5    Bocanek, J.6    Cozzi, L.7
  • 23
    • 0033965317 scopus 로고    scopus 로고
    • Verification of dynamic multileaf collimation using an electronic portal imaging device
    • 0031-9155,. 10.1088/0031-9155/45/2/316
    • H. V. James, S. Atherton, G. J. Budgell, M. C. Kirby, and P. C. Williams, " Verification of dynamic multileaf collimation using an electronic portal imaging device.," Phys. Med. Biol. 0031-9155 45, 495-509 (2000). 10.1088/0031-9155/45/2/316
    • (2000) Phys. Med. Biol. , vol.45 , pp. 495-509
    • James, H.V.1    Atherton, S.2    Budgell, G.J.3    Kirby, M.C.4    Williams, P.C.5
  • 24
    • 1242293815 scopus 로고    scopus 로고
    • Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager
    • 0094-2405,. 10.1118/1.1639125
    • J. J. Sonke, L. S. Ploeger, B. Brand, M. H. Smitsmans, and M. van Herk, " Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager.," Med. Phys. 0094-2405 31, 389-395 (2004). 10.1118/1.1639125
    • (2004) Med. Phys. , vol.31 , pp. 389-395
    • Sonke, J.J.1    Ploeger, L.S.2    Brand, B.3    Smitsmans, M.H.4    Van Herk, M.5
  • 25
    • 0033796474 scopus 로고    scopus 로고
    • Computer verification of fluence map for intensity modulated radiation therapy
    • 0094-2405,. 10.1118/1.1289374
    • L. Xing and J. G. Li, " Computer verification of fluence map for intensity modulated radiation therapy.," Med. Phys. 0094-2405 27, 2084-2092 (2000). 10.1118/1.1289374
    • (2000) Med. Phys. , vol.27 , pp. 2084-2092
    • Xing, L.1    Li, J.G.2
  • 26
    • 11244294832 scopus 로고    scopus 로고
    • On the use of EPID-based implanted marker tracking for 4D radiotherapy
    • 0094-2405,. 10.1118/1.1812608
    • P. J. Keall, A. D. Todor, S. S. Vedam, C. L. Bartee, J. V. Siebers, V. R. Kini, and R. Mohan, " On the use of EPID-based implanted marker tracking for 4D radiotherapy.," Med. Phys. 0094-2405 31, 3492-3499 (2004). 10.1118/1.1812608
    • (2004) Med. Phys. , vol.31 , pp. 3492-3499
    • Keall, P.J.1    Todor, A.D.2    Vedam, S.S.3    Bartee, C.L.4    Siebers, J.V.5    Kini, V.R.6    Mohan, R.7
  • 27
    • 41849134623 scopus 로고    scopus 로고
    • Analysis of reproducibility of respiration-triggered gated radiotherapy for lung tumors
    • 0167-8140,. 10.1016/j.radonc.2008.02.008
    • F. O. Spoelstra, J. R. van Sornsen de Koste, J. P. Cuijpers, F. J. Lagerwaard, B. J. Slotman, and S. Senan, " Analysis of reproducibility of respiration-triggered gated radiotherapy for lung tumors.," Radiother. Oncol. 0167-8140 87, 59-64 (2008). 10.1016/j.radonc.2008.02.008
    • (2008) Radiother. Oncol. , vol.87 , pp. 59-64
    • Spoelstra, F.O.1    Van Sornsen De Koste, J.R.2    Cuijpers, J.P.3    Lagerwaard, F.J.4    Slotman, B.J.5    Senan, S.6
  • 29
    • 33644999533 scopus 로고    scopus 로고
    • Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry
    • 0094-2405,. 10.1118/1.2128498
    • P. B. Greer, " Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry.," Med. Phys. 0094-2405 32, 3558-3568 (2005). 10.1118/1.2128498
    • (2005) Med. Phys. , vol.32 , pp. 3558-3568
    • Greer, P.B.1
  • 30
    • 34748821714 scopus 로고    scopus 로고
    • Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device
    • 0094-2405,. 10.1118/1.2779944
    • P. B. Greer, " Off-axis dose response characteristics of an amorphous silicon electronic portal imaging device.," Med. Phys. 0094-2405 34, 3815-3824 (2007). 10.1118/1.2779944
    • (2007) Med. Phys. , vol.34 , pp. 3815-3824
    • Greer, P.B.1
  • 31
    • 85027227047 scopus 로고    scopus 로고
    • (personal communication).
    • H. Riems (personal communication).
    • Riems, H.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.