Volumn 2, Issue 2, 2007, Pages 109-123

Current Trends in PET and Combined (PET/CT and PET/MR) Systems Design

Author keywords

[No Author keywords available]

Indexed keywords

CONTRAST MEDIUM; FLUORODEOXYGLUCOSE F 18;

EID: 46849087258     PISSN: 15568598     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.cpet.2007.10.004     Document Type: Review
Times cited : (31)

References (150)
  • 1
    • 0036448742 scopus 로고    scopus 로고
    • Molecular imaging with PET-the future challenges
    • Jones T. Molecular imaging with PET-the future challenges. Br J Radiol 75 (2002) 6S-15S
    • (2002) Br J Radiol , vol.75
    • Jones, T.1
  • 2
    • 0034126959 scopus 로고    scopus 로고
    • PET: the merging of biology and imaging into molecular imaging
    • Phelps M.E. PET: the merging of biology and imaging into molecular imaging. J Nucl Med 41 (2000) 661-681
    • (2000) J Nucl Med , vol.41 , pp. 661-681
    • Phelps, M.E.1
  • 3
    • 0036203522 scopus 로고    scopus 로고
    • The history of positron emission tomography
    • Nutt R. The history of positron emission tomography. Mol Imaging Biol 4 (2002) 11-26
    • (2002) Mol Imaging Biol , vol.4 , pp. 11-26
    • Nutt, R.1
  • 4
    • 0036156690 scopus 로고    scopus 로고
    • Guest editorial: the conception of FDG-PET imaging
    • Alavi A., and Reivich M. Guest editorial: the conception of FDG-PET imaging. Semin Nucl Med 32 (2002) 2-5
    • (2002) Semin Nucl Med , vol.32 , pp. 2-5
    • Alavi, A.1    Reivich, M.2
  • 5
    • 33947586234 scopus 로고    scopus 로고
    • Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006
    • Czernin J., Allen-Auerbach M., and Schelbert H.R. Improvements in cancer staging with PET/CT: literature-based evidence as of September 2006. J Nucl Med 48 (2007) 78S-88S
    • (2007) J Nucl Med , vol.48
    • Czernin, J.1    Allen-Auerbach, M.2    Schelbert, H.R.3
  • 6
    • 0033870513 scopus 로고    scopus 로고
    • A combined PET/CT scanner for clinical oncology
    • Beyer T., Townsend D.W., Brun T., et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 41 (2000) 1369-1379
    • (2000) J Nucl Med , vol.41 , pp. 1369-1379
    • Beyer, T.1    Townsend, D.W.2    Brun, T.3
  • 8
    • 33748192085 scopus 로고    scopus 로고
    • Multimodality in vivo imaging systems: twice the power or double the trouble?
    • Cherry S.R. Multimodality in vivo imaging systems: twice the power or double the trouble?. Annu Rev Biomed Eng 8 (2006) 35-62
    • (2006) Annu Rev Biomed Eng , vol.8 , pp. 35-62
    • Cherry, S.R.1
  • 9
    • 33845477269 scopus 로고
    • The use of positron-emitting radioisotopes for the localization of brain tumors
    • Wrenn Jr. F.R., Good M.L., and Handler P. The use of positron-emitting radioisotopes for the localization of brain tumors. Science 113 (1951) 525-527
    • (1951) Science , vol.113 , pp. 525-527
    • Wrenn Jr., F.R.1    Good, M.L.2    Handler, P.3
  • 10
    • 0000605030 scopus 로고
    • The uses of nuclear disintegration in the diagnosis and treatment of brain tumor
    • Sweet W.H. The uses of nuclear disintegration in the diagnosis and treatment of brain tumor. N Engl J Med 245 (1951) 875-878
    • (1951) N Engl J Med , vol.245 , pp. 875-878
    • Sweet, W.H.1
  • 11
    • 0001835009 scopus 로고
    • Localization of brain tumors with positron emitters
    • Brownell G.L., and Sweet W.H. Localization of brain tumors with positron emitters. Nucleonics 11 (1953) 40-45
    • (1953) Nucleonics , vol.11 , pp. 40-45
    • Brownell, G.L.1    Sweet, W.H.2
  • 12
    • 84941475787 scopus 로고
    • Design considerations for a positron emission transaxial tomograph (PETT III)
    • Phelps M.E., Hoffman E.J., Mullani N.A., et al. Design considerations for a positron emission transaxial tomograph (PETT III). IEEE Tran Nucl Sci 23 (1976) 516-522
    • (1976) IEEE Tran Nucl Sci , vol.23 , pp. 516-522
    • Phelps, M.E.1    Hoffman, E.J.2    Mullani, N.A.3
  • 13
    • 0017134455 scopus 로고
    • Design and performance characteristics of a whole-body transaxial tomograph
    • Hoffman E.J., Phelps M.E., Mullani N.A., et al. Design and performance characteristics of a whole-body transaxial tomograph. J Nucl Med 17 (1976) 493-502
    • (1976) J Nucl Med , vol.17 , pp. 493-502
    • Hoffman, E.J.1    Phelps, M.E.2    Mullani, N.A.3
  • 14
    • 0017227407 scopus 로고
    • The continuous inhalation of oxygen-15 for assessing regional oxygen extraction in the brain of man
    • Jones T., Chesler D.A., and Ter-Pogossian M.M. The continuous inhalation of oxygen-15 for assessing regional oxygen extraction in the brain of man. Br J Radiol 49 (1976) 339-343
    • (1976) Br J Radiol , vol.49 , pp. 339-343
    • Jones, T.1    Chesler, D.A.2    Ter-Pogossian, M.M.3
  • 15
    • 0017928649 scopus 로고
    • Positome II: a high efficiency positron imaging device for dynamic brain studies
    • Thompson C.J., Yamamoto Y.L., and Meyer E. Positome II: a high efficiency positron imaging device for dynamic brain studies. IEEE Trans Nucl Sci 26 (1979) 583-589
    • (1979) IEEE Trans Nucl Sci , vol.26 , pp. 583-589
    • Thompson, C.J.1    Yamamoto, Y.L.2    Meyer, E.3
  • 16
    • 0018076873 scopus 로고
    • Design considerations for a positron emission transverse tomograph (PETT V) for imaging of the brain
    • Ter-Pogossian M.M., Mullani N.A., Hood J.T., et al. Design considerations for a positron emission transverse tomograph (PETT V) for imaging of the brain. J Comput Assist Tomogr 2 (1978) 539-544
    • (1978) J Comput Assist Tomogr , vol.2 , pp. 539-544
    • Ter-Pogossian, M.M.1    Mullani, N.A.2    Hood, J.T.3
  • 17
    • 0020072892 scopus 로고
    • PETT VI: a positron emission tomograph utilizing cesium fluoride scintillation detectors
    • Ter-Pogossian M.M., Ficke D.C., Hood Sr. J.T., et al. PETT VI: a positron emission tomograph utilizing cesium fluoride scintillation detectors. J Comput Assist Tomogr 6 (1982) 125-133
    • (1982) J Comput Assist Tomogr , vol.6 , pp. 125-133
    • Ter-Pogossian, M.M.1    Ficke, D.C.2    Hood Sr., J.T.3
  • 18
    • 0036464422 scopus 로고    scopus 로고
    • The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph
    • Wienhard K., Schmand M., Casey M.E., et al. The ECAT HRRT: performance and first clinical application of the new high resolution research tomograph. IEEE Trans Nucl Sci 49 (2002) 104-110
    • (2002) IEEE Trans Nucl Sci , vol.49 , pp. 104-110
    • Wienhard, K.1    Schmand, M.2    Casey, M.E.3
  • 19
    • 64749091451 scopus 로고    scopus 로고
    • Instrumentation for positron emission mammography
    • Thompson C.J. Instrumentation for positron emission mammography. PET Clinics 1 (2006) 33-38
    • (2006) PET Clinics , vol.1 , pp. 33-38
    • Thompson, C.J.1
  • 21
    • 0038043944 scopus 로고    scopus 로고
    • Investigation of Lanthanum scintillators for 3-D PET
    • Surti S., Karp J., Muehllehner G., et al. Investigation of Lanthanum scintillators for 3-D PET. IEEE Trans Nucl Sci 50 (2003) 348-354
    • (2003) IEEE Trans Nucl Sci , vol.50 , pp. 348-354
    • Surti, S.1    Karp, J.2    Muehllehner, G.3
  • 22
    • 11344253041 scopus 로고    scopus 로고
    • Perspectives on the future development of new scintillators
    • Melcher C.L. Perspectives on the future development of new scintillators. Nucl Instrum Methods Phys Res A 537 (2005) 6-14
    • (2005) Nucl Instrum Methods Phys Res A , vol.537 , pp. 6-14
    • Melcher, C.L.1
  • 23
    • 34247230296 scopus 로고    scopus 로고
    • Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities
    • Surti S., Kuhn A., Werner M.E., et al. Performance of Philips Gemini TF PET/CT scanner with special consideration for its time-of-flight imaging capabilities. J Nucl Med 48 (2007) 471-480
    • (2007) J Nucl Med , vol.48 , pp. 471-480
    • Surti, S.1    Kuhn, A.2    Werner, M.E.3
  • 24
    • 0142179913 scopus 로고    scopus 로고
    • Detector technology challenges for nuclear medicine and PET
    • Marsden P.K. Detector technology challenges for nuclear medicine and PET. Nucl Instr Methods A 513 (2003) 1-7
    • (2003) Nucl Instr Methods A , vol.513 , pp. 1-7
    • Marsden, P.K.1
  • 26
    • 0037149996 scopus 로고    scopus 로고
    • Inorganic scintillators in medical imaging
    • van Eijk C.W.E. Inorganic scintillators in medical imaging. Phys Med Biol 47 (2002) R85-R106
    • (2002) Phys Med Biol , vol.47
    • van Eijk, C.W.E.1
  • 27
    • 0032099862 scopus 로고    scopus 로고
    • A YSO/LSO phoswich array detector for single and coincidence photon imaging
    • Dahlbom M., MacDonald L.R., Schmand M., et al. A YSO/LSO phoswich array detector for single and coincidence photon imaging. IEEE Trans Nucl Sci 45 (1998) 1128-1132
    • (1998) IEEE Trans Nucl Sci , vol.45 , pp. 1128-1132
    • Dahlbom, M.1    MacDonald, L.R.2    Schmand, M.3
  • 28
    • 3242807530 scopus 로고    scopus 로고
    • Feasibility of a novel design of high-resolution parallax-free Compton enhanced PET scanner dedicated to brain research
    • Braem A., Chamizo Llatas M., Chesi E., et al. Feasibility of a novel design of high-resolution parallax-free Compton enhanced PET scanner dedicated to brain research. Phys Med Biol 49 (2004) 2547-2562
    • (2004) Phys Med Biol , vol.49 , pp. 2547-2562
    • Braem, A.1    Chamizo Llatas, M.2    Chesi, E.3
  • 29
    • 33748775475 scopus 로고    scopus 로고
    • Novel geometrical concept of high performance brain PET scanner: principle, design and performance estimates
    • Seguinot J., Braem A., Chesi E., et al. Novel geometrical concept of high performance brain PET scanner: principle, design and performance estimates. Nuovo Cimento C 29 (2006) 429-463
    • (2006) Nuovo Cimento C , vol.29 , pp. 429-463
    • Seguinot, J.1    Braem, A.2    Chesi, E.3
  • 30
    • 0031854801 scopus 로고    scopus 로고
    • Time-of-flight PET
    • Lewellen T.K. Time-of-flight PET. Semin Nucl Med 28 (1998) 268-275
    • (1998) Semin Nucl Med , vol.28 , pp. 268-275
    • Lewellen, T.K.1
  • 31
    • 0142126690 scopus 로고    scopus 로고
    • Time of flight in PET revisited
    • Moses W.W. Time of flight in PET revisited. IEEE Trans Nucl Sci 50 (2003) 1325-1330
    • (2003) IEEE Trans Nucl Sci , vol.50 , pp. 1325-1330
    • Moses, W.W.1
  • 33
    • 25444442095 scopus 로고    scopus 로고
    • First experimental results of time-of-flight reconstruction on an LSO PET scanner
    • Conti M., Bendriem B., Casey M., et al. First experimental results of time-of-flight reconstruction on an LSO PET scanner. Phys Med Biol 50 (2005) 4507-4526
    • (2005) Phys Med Biol , vol.50 , pp. 4507-4526
    • Conti, M.1    Bendriem, B.2    Casey, M.3
  • 34
    • 33846562178 scopus 로고    scopus 로고
    • Harrison RL, Gillispie SB, Alessio AM, et al. The effects of object size, attenuation, scatter, and random coincidences on signal to noise ratio in simulations of time-of-flight positron emission tomography. IEEE Nuclear Science Symposium Conference Record. October 23-29, 2005, Puerto Rico.
    • Harrison RL, Gillispie SB, Alessio AM, et al. The effects of object size, attenuation, scatter, and random coincidences on signal to noise ratio in simulations of time-of-flight positron emission tomography. IEEE Nuclear Science Symposium Conference Record. October 23-29, 2005, Puerto Rico.
  • 35
    • 33646388372 scopus 로고    scopus 로고
    • Investigation of time-of-flight benefit for fully 3-D PET
    • Surti S., Karp J.S., Popescu L.M., et al. Investigation of time-of-flight benefit for fully 3-D PET. IEEE Trans Med Imaging 25 (2006) 529-538
    • (2006) IEEE Trans Med Imaging , vol.25 , pp. 529-538
    • Surti, S.1    Karp, J.S.2    Popescu, L.M.3
  • 36
    • 2642514922 scopus 로고    scopus 로고
    • NEMA NU 2 performance tests for scanners with intrinsic radioactivity
    • Watson C.C., Casey M.E., Erikksson L., et al. NEMA NU 2 performance tests for scanners with intrinsic radioactivity. J Nucl Med 45 (2004) 822-826
    • (2004) J Nucl Med , vol.45 , pp. 822-826
    • Watson, C.C.1    Casey, M.E.2    Erikksson, L.3
  • 37
    • 2942584985 scopus 로고    scopus 로고
    • Performance evaluation of the new whole-body PET/CT scanner: discovery ST
    • Bettinardi V., Danna M., Savi A., et al. Performance evaluation of the new whole-body PET/CT scanner: discovery ST. Eur J Nucl Med Mol Imaging 31 (2004) 867-881
    • (2004) Eur J Nucl Med Mol Imaging , vol.31 , pp. 867-881
    • Bettinardi, V.1    Danna, M.2    Savi, A.3
  • 38
    • 16544364015 scopus 로고    scopus 로고
    • Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes
    • Mawlawi O., Podoloff D.A., Kohlmyer S., et al. Performance characteristics of a newly developed PET/CT scanner using NEMA standards in 2D and 3D modes. J Nucl Med 45 (2004) 1734-1742
    • (2004) J Nucl Med , vol.45 , pp. 1734-1742
    • Mawlawi, O.1    Podoloff, D.A.2    Kohlmyer, S.3
  • 39
    • 2642558094 scopus 로고    scopus 로고
    • PET performance measurements for an LSO-based combined PET/CT scanner using the National Electrical Manufacturers Association NU 2-2001 standard
    • Erdi Y.E., Nehmeh S.A., Mulnix T., et al. PET performance measurements for an LSO-based combined PET/CT scanner using the National Electrical Manufacturers Association NU 2-2001 standard. J Nucl Med 45 (2004) 813-821
    • (2004) J Nucl Med , vol.45 , pp. 813-821
    • Erdi, Y.E.1    Nehmeh, S.A.2    Mulnix, T.3
  • 40
    • 3242717954 scopus 로고    scopus 로고
    • Imaging characteristics of a 3-dimensional GSO whole-body PET camera
    • Surti S., and Karp J.S. Imaging characteristics of a 3-dimensional GSO whole-body PET camera. J Nucl Med 45 (2004) 1040-1049
    • (2004) J Nucl Med , vol.45 , pp. 1040-1049
    • Surti, S.1    Karp, J.S.2
  • 41
    • 33644874714 scopus 로고    scopus 로고
    • Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard
    • Brambilla M., Secco C., Dominietto M., et al. Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard. J Nucl Med 46 (2005) 2083-2091
    • (2005) J Nucl Med , vol.46 , pp. 2083-2091
    • Brambilla, M.1    Secco, C.2    Dominietto, M.3
  • 42
    • 34247204476 scopus 로고    scopus 로고
    • NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes
    • Kemp B.J., Kim C., Williams J.J., et al. NEMA NU 2-2001 performance measurements of an LYSO-based PET/CT system in 2D and 3D acquisition modes. J Nucl Med 47 (2006) 1960-1967
    • (2006) J Nucl Med , vol.47 , pp. 1960-1967
    • Kemp, B.J.1    Kim, C.2    Williams, J.J.3
  • 43
    • 34247242955 scopus 로고    scopus 로고
    • Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard
    • Matsumoto K., Kitamura K., Mizuta T., et al. Performance characteristics of a new 3-dimensional continuous-emission and spiral-transmission high-sensitivity and high-resolution PET camera evaluated with the NEMA NU 2-2001 standard. J Nucl Med 47 (2006) 83-90
    • (2006) J Nucl Med , vol.47 , pp. 83-90
    • Matsumoto, K.1    Kitamura, K.2    Mizuta, T.3
  • 45
    • 0036624342 scopus 로고    scopus 로고
    • A new high-resolution PET scanner dedicated to brain research
    • Watanabe M., Shimizu K., Omura T., et al. A new high-resolution PET scanner dedicated to brain research. IEEE Trans Nucl Sci 49 (2002) 634-639
    • (2002) IEEE Trans Nucl Sci , vol.49 , pp. 634-639
    • Watanabe, M.1    Shimizu, K.2    Omura, T.3
  • 46
    • 0141557961 scopus 로고    scopus 로고
    • Performance of a brain PET camera based on anger-logic gadolinium oxyorthosilicate detectors
    • Karp J.S., Surti S., Daube-Witherspoon M.E., et al. Performance of a brain PET camera based on anger-logic gadolinium oxyorthosilicate detectors. J Nucl Med 44 (2003) 1340-1349
    • (2003) J Nucl Med , vol.44 , pp. 1340-1349
    • Karp, J.S.1    Surti, S.2    Daube-Witherspoon, M.E.3
  • 47
    • 3342966076 scopus 로고    scopus 로고
    • Novel design of a parallax free Compton enhanced PET scanner
    • Braem A., Chamizo Llatas M., Chesi E., et al. Novel design of a parallax free Compton enhanced PET scanner. Nucl Instr Methods A 525 (2004) 268-274
    • (2004) Nucl Instr Methods A , vol.525 , pp. 268-274
    • Braem, A.1    Chamizo Llatas, M.2    Chesi, E.3
  • 48
    • 33646446682 scopus 로고    scopus 로고
    • The new challenges of brain PET imaging technology
    • Zaidi H., and Montandon M.-L. The new challenges of brain PET imaging technology. Curr Med Imag Rev 2 (2006) 3-13
    • (2006) Curr Med Imag Rev , vol.2 , pp. 3-13
    • Zaidi, H.1    Montandon, M.-L.2
  • 49
    • 0035355148 scopus 로고    scopus 로고
    • maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer
    • Doshi N.K., Silverman R.W., Shao Y., et al. maxPET, a dedicated mammary and axillary region PET imaging system for breast cancer. IEEE Trans Nucl Sci 48 (2001) 811-815
    • (2001) IEEE Trans Nucl Sci , vol.48 , pp. 811-815
    • Doshi, N.K.1    Silverman, R.W.2    Shao, Y.3
  • 50
    • 0031471309 scopus 로고    scopus 로고
    • Dedicated PET scanners for breast imaging
    • Freifelder R., and Karp J.S. Dedicated PET scanners for breast imaging. Phys Med Biol 42 (1997) 2463-2480
    • (1997) Phys Med Biol , vol.42 , pp. 2463-2480
    • Freifelder, R.1    Karp, J.S.2
  • 51
    • 0037263417 scopus 로고    scopus 로고
    • Positron emission mammography: initial clinical results
    • Levine E.A., Freimanis R.I., Perrier N.D., et al. Positron emission mammography: initial clinical results. Ann Surg Oncol 10 (2003) 86-91
    • (2003) Ann Surg Oncol , vol.10 , pp. 86-91
    • Levine, E.A.1    Freimanis, R.I.2    Perrier, N.D.3
  • 52
    • 2942711894 scopus 로고    scopus 로고
    • Instrumentation optimization for positron emission mammography
    • Moses W.W., and Qi J. Instrumentation optimization for positron emission mammography. Nucl Instr Meth A 527 (2004) 76-82
    • (2004) Nucl Instr Meth A , vol.527 , pp. 76-82
    • Moses, W.W.1    Qi, J.2
  • 53
    • 0034026990 scopus 로고    scopus 로고
    • Positron emission mammographic instrument: initial results
    • Murthy K., Aznar M., Bergman A.M., et al. Positron emission mammographic instrument: initial results. Radiology 215 (2000) 280-285
    • (2000) Radiology , vol.215 , pp. 280-285
    • Murthy, K.1    Aznar, M.2    Bergman, A.M.3
  • 54
    • 0033864256 scopus 로고    scopus 로고
    • The potential role of positron emission mammography for detection of breast cancer: a phantom study
    • Raylman R.R., Majewski S., Wojcik R., et al. The potential role of positron emission mammography for detection of breast cancer: a phantom study. Med Phys 27 (2000) 1943-1954
    • (2000) Med Phys , vol.27 , pp. 1943-1954
    • Raylman, R.R.1    Majewski, S.2    Wojcik, R.3
  • 55
    • 3042819646 scopus 로고    scopus 로고
    • Positron emission mammography with tomographic acquisition using dual planar detectors: initial evaluations
    • Smith M.F., Raylman R.R., Majewski S., et al. Positron emission mammography with tomographic acquisition using dual planar detectors: initial evaluations. Phys Med Biol 49 (2004) 2437-2452
    • (2004) Phys Med Biol , vol.49 , pp. 2437-2452
    • Smith, M.F.1    Raylman, R.R.2    Majewski, S.3
  • 57
    • 19944430403 scopus 로고    scopus 로고
    • Positron emission mammography: high-resolution biochemical breast imaging
    • Weinberg I.N., Beylin D., Zavarzin V., et al. Positron emission mammography: high-resolution biochemical breast imaging. Technol Cancer Res Treat 4 (2005) 55-60
    • (2005) Technol Cancer Res Treat , vol.4 , pp. 55-60
    • Weinberg, I.N.1    Beylin, D.2    Zavarzin, V.3
  • 58
    • 33846643984 scopus 로고    scopus 로고
    • Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation
    • Zhang J., Olcott P.D., Chinn G., et al. Study of the performance of a novel 1 mm resolution dual-panel PET camera design dedicated to breast cancer imaging using Monte Carlo simulation. Med Phys 34 (2007) 689-702
    • (2007) Med Phys , vol.34 , pp. 689-702
    • Zhang, J.1    Olcott, P.D.2    Chinn, G.3
  • 59
    • 3342901946 scopus 로고    scopus 로고
    • Detector development for a novel positron emission mammography scanner based on YAP:Ce crystals
    • Belcari N., Camarda M., Del Guerra A., et al. Detector development for a novel positron emission mammography scanner based on YAP:Ce crystals. Nucl Instr Meth A 525 (2004) 258-262
    • (2004) Nucl Instr Meth A , vol.525 , pp. 258-262
    • Belcari, N.1    Camarda, M.2    Del Guerra, A.3
  • 61
    • 0030980637 scopus 로고    scopus 로고
    • Three-dimensional imaging characteristics of the HEAD PENN-PET scanner
    • Karp J.S., Freifelder R., Geagan M.J., et al. Three-dimensional imaging characteristics of the HEAD PENN-PET scanner. J Nucl Med 38 (1997) 636-643
    • (1997) J Nucl Med , vol.38 , pp. 636-643
    • Karp, J.S.1    Freifelder, R.2    Geagan, M.J.3
  • 62
    • 0035659148 scopus 로고    scopus 로고
    • Performance of a whole-body PET scanner using curve-plate NaI(Tl) detectors
    • Adam L.E., Karp J.S., Daube-Witherspoon M.E., et al. Performance of a whole-body PET scanner using curve-plate NaI(Tl) detectors. J Nucl Med 42 (2001) 1821-1830
    • (2001) J Nucl Med , vol.42 , pp. 1821-1830
    • Adam, L.E.1    Karp, J.S.2    Daube-Witherspoon, M.E.3
  • 63
    • 0002416579 scopus 로고    scopus 로고
    • The changing design of positron imaging systems
    • Phelps M.E., and Cherry S.R. The changing design of positron imaging systems. Clin Positron Imaging 1 (1998) 31-45
    • (1998) Clin Positron Imaging , vol.1 , pp. 31-45
    • Phelps, M.E.1    Cherry, S.R.2
  • 64
    • 26244468593 scopus 로고    scopus 로고
    • Dual-modality imaging: more than the sum of its components
    • Zaidi H. (Ed), Springer, New York
    • Hasegawa B., and Zaidi H. Dual-modality imaging: more than the sum of its components. In: Zaidi H. (Ed). Quantitative analysis in nuclear medicine imaging (2006), Springer, New York 35-81
    • (2006) Quantitative analysis in nuclear medicine imaging , pp. 35-81
    • Hasegawa, B.1    Zaidi, H.2
  • 65
    • 0025626968 scopus 로고
    • Description of a simultaneous emission-transmission CT system
    • Hasegawa B.H., Gingold E.L., Reilly S.M., et al. Description of a simultaneous emission-transmission CT system. Proc SPIE 1231 (1990) 50-60
    • (1990) Proc SPIE , vol.1231 , pp. 50-60
    • Hasegawa, B.H.1    Gingold, E.L.2    Reilly, S.M.3
  • 68
    • 2442477114 scopus 로고    scopus 로고
    • Medical image registration using mutual information
    • Maes F., Vandermeulen D., and Suetens P. Medical image registration using mutual information. Proc IEEE 91 (2003) 1699-1722
    • (2003) Proc IEEE , vol.91 , pp. 1699-1722
    • Maes, F.1    Vandermeulen, D.2    Suetens, P.3
  • 69
    • 33846637951 scopus 로고    scopus 로고
    • Extension of single scatter simulation to scatter correction of time-of-flight PET
    • Watson C.C. Extension of single scatter simulation to scatter correction of time-of-flight PET. Conference Record of the IEEE Nuclear Science Symposium 5 (2005) 2492-2496
    • (2005) Conference Record of the IEEE Nuclear Science Symposium , vol.5 , pp. 2492-2496
    • Watson, C.C.1
  • 71
    • 0028004662 scopus 로고
    • An interactive technique for three-dimensional image registration: validation for PET, SPECT, MRI and CT brain studies
    • Pietrzyk U., Herholz K., Fink G., et al. An interactive technique for three-dimensional image registration: validation for PET, SPECT, MRI and CT brain studies. J Nucl Med 35 (1994) 2011-2018
    • (1994) J Nucl Med , vol.35 , pp. 2011-2018
    • Pietrzyk, U.1    Herholz, K.2    Fink, G.3
  • 72
    • 0031910991 scopus 로고    scopus 로고
    • Automated image registration. I. General methods and intrasubject, intramodality validation
    • Woods R.P., Grafton S.T., Holmes C.J., et al. Automated image registration. I. General methods and intrasubject, intramodality validation. J Comput Assist Tomogr 22 (1998) 139-152
    • (1998) J Comput Assist Tomogr , vol.22 , pp. 139-152
    • Woods, R.P.1    Grafton, S.T.2    Holmes, C.J.3
  • 73
    • 27144473297 scopus 로고    scopus 로고
    • Comparison of alignment of computer-registered data sets: combined PET/CT versus independent PET and CT of the thorax
    • Krishnasetty V., Fischman A.J., Halpern E.L., et al. Comparison of alignment of computer-registered data sets: combined PET/CT versus independent PET and CT of the thorax. Radiology 237 (2005) 635-639
    • (2005) Radiology , vol.237 , pp. 635-639
    • Krishnasetty, V.1    Fischman, A.J.2    Halpern, E.L.3
  • 74
    • 1642586981 scopus 로고    scopus 로고
    • Software approach to merging molecular with anatomic information
    • Slomka P.J. Software approach to merging molecular with anatomic information. J Nucl Med 45 Suppl 1 (2004) 36S-45S
    • (2004) J Nucl Med , vol.45 , Issue.SUPPL. 1
    • Slomka, P.J.1
  • 76
    • 0037792992 scopus 로고    scopus 로고
    • Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography
    • Lardinois D., Weder W., Hany T.F., et al. Staging of non-small-cell lung cancer with integrated positron-emission tomography and computed tomography. N Engl J Med 348 (2003) 2500-2507
    • (2003) N Engl J Med , vol.348 , pp. 2500-2507
    • Lardinois, D.1    Weder, W.2    Hany, T.F.3
  • 77
    • 0037811001 scopus 로고    scopus 로고
    • X-ray-based attenuation correction for positron emission tomography/computed tomography scanners
    • Kinahan P.E., Hasegawa B.H., and Beyer T. X-ray-based attenuation correction for positron emission tomography/computed tomography scanners. Semin Nucl Med 33 (2003) 166-179
    • (2003) Semin Nucl Med , vol.33 , pp. 166-179
    • Kinahan, P.E.1    Hasegawa, B.H.2    Beyer, T.3
  • 78
    • 2442555874 scopus 로고    scopus 로고
    • Scatter modelling and compensation in emission tomography
    • Zaidi H., and Koral K.F. Scatter modelling and compensation in emission tomography. Eur J Nucl Med Mol Imaging 31 (2004) 761-782
    • (2004) Eur J Nucl Med Mol Imaging , vol.31 , pp. 761-782
    • Zaidi, H.1    Koral, K.F.2
  • 79
    • 0036144958 scopus 로고    scopus 로고
    • Clinically feasible reconstruction of 3D whole-body PET/CT data using blurred anatomical labels
    • Comtat C., Kinahan P.E., Fessler J.A., et al. Clinically feasible reconstruction of 3D whole-body PET/CT data using blurred anatomical labels. Phys Med Biol 47 (2002) 1-20
    • (2002) Phys Med Biol , vol.47 , pp. 1-20
    • Comtat, C.1    Kinahan, P.E.2    Fessler, J.A.3
  • 80
    • 0037369944 scopus 로고    scopus 로고
    • Respiration-induced attenuation artifact at PET/CT: technical considerations
    • Goerres G.W., Burger C., Kamel E., et al. Respiration-induced attenuation artifact at PET/CT: technical considerations. Radiology 226 (2003) 906-910
    • (2003) Radiology , vol.226 , pp. 906-910
    • Goerres, G.W.1    Burger, C.2    Kamel, E.3
  • 81
    • 0037884881 scopus 로고    scopus 로고
    • Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT
    • Osman M.M., Cohade C., Nakamoto Y., et al. Respiratory motion artifacts on PET emission images obtained using CT attenuation correction on PET-CT. Eur J Nucl Med Mol Imaging 30 (2003) 603-606
    • (2003) Eur J Nucl Med Mol Imaging , vol.30 , pp. 603-606
    • Osman, M.M.1    Cohade, C.2    Nakamoto, Y.3
  • 82
    • 28544440972 scopus 로고    scopus 로고
    • Respiratory motion artifact in the liver dome on FDG PET/CT: comparison of attenuation correction with CT and a caesium external source
    • Papathanassiou D., Becker S., Amir R., et al. Respiratory motion artifact in the liver dome on FDG PET/CT: comparison of attenuation correction with CT and a caesium external source. Eur J Nucl Med Mol Imaging 32 (2005) 1422-1428
    • (2005) Eur J Nucl Med Mol Imaging , vol.32 , pp. 1422-1428
    • Papathanassiou, D.1    Becker, S.2    Amir, R.3
  • 83
    • 34447306296 scopus 로고    scopus 로고
    • Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections
    • Gould K.L., Pan T., Loghin C., et al. Frequent diagnostic errors in cardiac PET/CT due to misregistration of CT attenuation and emission PET images: a definitive analysis of causes, consequences, and corrections. J Nucl Med 48 (2007) 1112-1121
    • (2007) J Nucl Med , vol.48 , pp. 1112-1121
    • Gould, K.L.1    Pan, T.2    Loghin, C.3
  • 84
    • 34447313914 scopus 로고    scopus 로고
    • Attenuation correction in myocardial perfusion SPECT/CT: effects of misregistration and value of reregistration
    • Goetze S., Brown T.L., Lavely W.C., et al. Attenuation correction in myocardial perfusion SPECT/CT: effects of misregistration and value of reregistration. J Nucl Med 48 (2007) 1090-1095
    • (2007) J Nucl Med , vol.48 , pp. 1090-1095
    • Goetze, S.1    Brown, T.L.2    Lavely, W.C.3
  • 85
    • 27744443466 scopus 로고    scopus 로고
    • Attenuation correction of PET images with respiration-averaged CT images in PET/CT
    • Pan T., Mawlawi O., Nehmeh S.A., et al. Attenuation correction of PET images with respiration-averaged CT images in PET/CT. J Nucl Med 46 (2005) 1481-1487
    • (2005) J Nucl Med , vol.46 , pp. 1481-1487
    • Pan, T.1    Mawlawi, O.2    Nehmeh, S.A.3
  • 86
    • 22844449456 scopus 로고    scopus 로고
    • Reconstruction from truncated projections in CT using adaptive detruncation
    • Sourbelle K., Kachelriess M., and Kalender W.A. Reconstruction from truncated projections in CT using adaptive detruncation. Eur Radiol 15 (2005) 1008-1014
    • (2005) Eur Radiol , vol.15 , pp. 1008-1014
    • Sourbelle, K.1    Kachelriess, M.2    Kalender, W.A.3
  • 88
    • 33745515808 scopus 로고    scopus 로고
    • Truncation artifact on PET/CT: impact on measurements of activity concentration and assessment of a correction algorithm
    • Mawlawi O., Erasmus J.J., Pan T., et al. Truncation artifact on PET/CT: impact on measurements of activity concentration and assessment of a correction algorithm. AJR Am J Roentgenol 186 (2006) 1458-1467
    • (2006) AJR Am J Roentgenol , vol.186 , pp. 1458-1467
    • Mawlawi, O.1    Erasmus, J.J.2    Pan, T.3
  • 89
    • 33745851216 scopus 로고    scopus 로고
    • Whole-body 18F-FDG PET/CT in the presence of truncation artifacts
    • Beyer T., Bockisch A., Kuhl H., et al. Whole-body 18F-FDG PET/CT in the presence of truncation artifacts. J Nucl Med 47 (2006) 91-99
    • (2006) J Nucl Med , vol.47 , pp. 91-99
    • Beyer, T.1    Bockisch, A.2    Kuhl, H.3
  • 90
    • 0037945457 scopus 로고    scopus 로고
    • Cause and magnitude of the error induced by oral CT contrast agent in CT-based attenuation correction of PET emission studies
    • Dizendorf E., Hany T.F., Buck A., et al. Cause and magnitude of the error induced by oral CT contrast agent in CT-based attenuation correction of PET emission studies. J Nucl Med 44 (2003) 732-738
    • (2003) J Nucl Med , vol.44 , pp. 732-738
    • Dizendorf, E.1    Hany, T.F.2    Buck, A.3
  • 91
    • 0036788467 scopus 로고    scopus 로고
    • Focal tracer uptake: a potential artifact in contrast-enhanced dual-modality PET/CT scans
    • Antoch G., Freudenberg L.S., Egelhof T., et al. Focal tracer uptake: a potential artifact in contrast-enhanced dual-modality PET/CT scans. J Nucl Med 43 (2002) 1339-1342
    • (2002) J Nucl Med , vol.43 , pp. 1339-1342
    • Antoch, G.1    Freudenberg, L.S.2    Egelhof, T.3
  • 92
    • 1642586985 scopus 로고    scopus 로고
    • To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT
    • Antoch G., Freudenberg L.S., Beyer T., et al. To enhance or not to enhance? 18F-FDG and CT contrast agents in dual-modality 18F-FDG PET/CT. J Nucl Med 45 Suppl 1 (2004) 56S-65S
    • (2004) J Nucl Med , vol.45 , Issue.SUPPL. 1
    • Antoch, G.1    Freudenberg, L.S.2    Beyer, T.3
  • 93
    • 10744231206 scopus 로고    scopus 로고
    • Dual-modality PET/CT scanning with negative oral contrast agent to avoid artifacts: introduction and evaluation
    • Antoch G., Kuehl H., Kanja J., et al. Dual-modality PET/CT scanning with negative oral contrast agent to avoid artifacts: introduction and evaluation. Radiology 230 (2004) 879-885
    • (2004) Radiology , vol.230 , pp. 879-885
    • Antoch, G.1    Kuehl, H.2    Kanja, J.3
  • 94
    • 1542681656 scopus 로고    scopus 로고
    • Correction for oral contrast artifacts in CT attenuation-corrected PET images obtained by combined PET/CT
    • Nehmeh S.A., Erdi Y.E., Kalaigian H., et al. Correction for oral contrast artifacts in CT attenuation-corrected PET images obtained by combined PET/CT. J Nucl Med 44 (2003) 1940-1944
    • (2003) J Nucl Med , vol.44 , pp. 1940-1944
    • Nehmeh, S.A.1    Erdi, Y.E.2    Kalaigian, H.3
  • 95
    • 0037950269 scopus 로고    scopus 로고
    • Effects of nonionic intravenous contrast agents in PET/CT imaging: phantom and canine studies
    • Nakamoto Y., Chin B.B., Kraitchman D.L., et al. Effects of nonionic intravenous contrast agents in PET/CT imaging: phantom and canine studies. Radiology 227 (2003) 817-824
    • (2003) Radiology , vol.227 , pp. 817-824
    • Nakamoto, Y.1    Chin, B.B.2    Kraitchman, D.L.3
  • 96
    • 15844389922 scopus 로고    scopus 로고
    • Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error?
    • Yau Y.-Y., Chan W.-S., Tam Y.-M., et al. Application of intravenous contrast in PET/CT: does it really introduce significant attenuation correction error?. J Nucl Med 46 (2005) 283-291
    • (2005) J Nucl Med , vol.46 , pp. 283-291
    • Yau, Y.-Y.1    Chan, W.-S.2    Tam, Y.-M.3
  • 97
    • 25144468360 scopus 로고    scopus 로고
    • PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients
    • Berthelsen A.K., Holm S., Loft A., et al. PET/CT with intravenous contrast can be used for PET attenuation correction in cancer patients. Eur J Nucl Med Mol Imaging 32 (2005) 1167-1175
    • (2005) Eur J Nucl Med Mol Imaging , vol.32 , pp. 1167-1175
    • Berthelsen, A.K.1    Holm, S.2    Loft, A.3
  • 98
    • 25144473983 scopus 로고    scopus 로고
    • Oral contrast medium in PET/CT: should you or shouldn't you?
    • Groves A.M., Kayani I., Dickson J.C., et al. Oral contrast medium in PET/CT: should you or shouldn't you?. Eur J Nucl Med Mol Imaging 32 (2005) 1160-1166
    • (2005) Eur J Nucl Med Mol Imaging , vol.32 , pp. 1160-1166
    • Groves, A.M.1    Kayani, I.2    Dickson, J.C.3
  • 99
    • 17844386601 scopus 로고    scopus 로고
    • Optimized intravenous contrast administration for diagnostic whole-body 18F-FDG PET/CT
    • Beyer T., Antoch G., Bockisch A., et al. Optimized intravenous contrast administration for diagnostic whole-body 18F-FDG PET/CT. J Nucl Med 46 (2005) 429-435
    • (2005) J Nucl Med , vol.46 , pp. 429-435
    • Beyer, T.1    Antoch, G.2    Bockisch, A.3
  • 100
    • 33747599964 scopus 로고    scopus 로고
    • Attenuation-correction induced artifact in F-18 FDG PET imaging following total knee replacement
    • Heiba S.I., Luo J., Sadek S., et al. Attenuation-correction induced artifact in F-18 FDG PET imaging following total knee replacement. Clin Positron Imaging 3 (2000) 237-239
    • (2000) Clin Positron Imaging , vol.3 , pp. 237-239
    • Heiba, S.I.1    Luo, J.2    Sadek, S.3
  • 101
    • 1542399588 scopus 로고    scopus 로고
    • Cardiac pacemakers and central venous lines can induce focal artifacts on CT-corrected PET images
    • Halpern B.S., Dahlbom M., Waldherr C., et al. Cardiac pacemakers and central venous lines can induce focal artifacts on CT-corrected PET images. J Nucl Med 45 (2004) 290-293
    • (2004) J Nucl Med , vol.45 , pp. 290-293
    • Halpern, B.S.1    Dahlbom, M.2    Waldherr, C.3
  • 102
    • 17844369745 scopus 로고    scopus 로고
    • Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?
    • DiFilippo F.P., and Brunken R.C. Do implanted pacemaker leads and ICD leads cause metal-related artifact in cardiac PET/CT?. J Nucl Med 46 (2005) 436-443
    • (2005) J Nucl Med , vol.46 , pp. 436-443
    • DiFilippo, F.P.1    Brunken, R.C.2
  • 103
    • 0036201958 scopus 로고    scopus 로고
    • Head and neck imaging with PET and PET/CT: artifacts from dental metallic implants
    • Goerres G.W., Hany T.F., Kamel E., et al. Head and neck imaging with PET and PET/CT: artifacts from dental metallic implants. Eur J Nucl Med Mol Imaging 29 (2002) 367-370
    • (2002) Eur J Nucl Med Mol Imaging , vol.29 , pp. 367-370
    • Goerres, G.W.1    Hany, T.F.2    Kamel, E.3
  • 104
    • 0037294323 scopus 로고    scopus 로고
    • Artifacts at PET and PET/CT caused by metallic hip prosthetic material
    • Goerres G.W., Ziegler S.I., Burger C., et al. Artifacts at PET and PET/CT caused by metallic hip prosthetic material. Radiology 226 (2003) 577-584
    • (2003) Radiology , vol.226 , pp. 577-584
    • Goerres, G.W.1    Ziegler, S.I.2    Burger, C.3
  • 105
    • 0037936886 scopus 로고    scopus 로고
    • Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner
    • Kamel E.M., Burger C., Buck A., et al. Impact of metallic dental implants on CT-based attenuation correction in a combined PET/CT scanner. Eur Radiol 13 (2003) 724-728
    • (2003) Eur Radiol , vol.13 , pp. 724-728
    • Kamel, E.M.1    Burger, C.2    Buck, A.3
  • 106
    • 0037908633 scopus 로고    scopus 로고
    • Artifactual 2-deoxy-2-[(18)F]fluoro-D-glucose localization surrounding metallic objects in a PET/CT scanner using CT-based attenuation correction
    • Bujenovic S., Mannting F., Chakrabarti R., et al. Artifactual 2-deoxy-2-[(18)F]fluoro-D-glucose localization surrounding metallic objects in a PET/CT scanner using CT-based attenuation correction. Mol Imaging Biol 5 (2003) 20-22
    • (2003) Mol Imaging Biol , vol.5 , pp. 20-22
    • Bujenovic, S.1    Mannting, F.2    Chakrabarti, R.3
  • 108
    • 0033983855 scopus 로고    scopus 로고
    • An iterative approach to the beam hardening correction in cone beam CT
    • Hsieh J., Molthen R.C., Dawson C.A., et al. An iterative approach to the beam hardening correction in cone beam CT. Med Phys 27 (2000) 23-29
    • (2000) Med Phys , vol.27 , pp. 23-29
    • Hsieh, J.1    Molthen, R.C.2    Dawson, C.A.3
  • 109
    • 33846624378 scopus 로고    scopus 로고
    • Kachelrieb M, Kalender WA. Improving PET/CT attenuation correction with iterative CT beam hardening correction. IEEE Nuclear Science Symposium Conference Record. October 23-29, 2005, Puerto Rico.
    • Kachelrieb M, Kalender WA. Improving PET/CT attenuation correction with iterative CT beam hardening correction. IEEE Nuclear Science Symposium Conference Record. October 23-29, 2005, Puerto Rico.
  • 110
    • 33748163247 scopus 로고    scopus 로고
    • Assessment of errors caused by x-ray scatter and use of contrast medium when using CT-based attenuation correction in PET
    • Ay M., and Zaidi H. Assessment of errors caused by x-ray scatter and use of contrast medium when using CT-based attenuation correction in PET. Eur J Nucl Med Mol Imaging 33 (2006) 1301-1313
    • (2006) Eur J Nucl Med Mol Imaging , vol.33 , pp. 1301-1313
    • Ay, M.1    Zaidi, H.2
  • 111
    • 31044451240 scopus 로고    scopus 로고
    • A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT
    • Siewerdsen J.H., Daly M.J., Bakhtiar B., et al. A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT. Med Phys 33 (2006) 187-197
    • (2006) Med Phys , vol.33 , pp. 187-197
    • Siewerdsen, J.H.1    Daly, M.J.2    Bakhtiar, B.3
  • 112
    • 21244491581 scopus 로고    scopus 로고
    • Effect of scatter on reconstructed image quality in cone beam computed tomography: evaluation of a scatter-reduction optimisation function
    • Malusek A., Seger M.M., Sandborg M., et al. Effect of scatter on reconstructed image quality in cone beam computed tomography: evaluation of a scatter-reduction optimisation function. Radiat Prot Dosimetry 114 (2005) 337-340
    • (2005) Radiat Prot Dosimetry , vol.114 , pp. 337-340
    • Malusek, A.1    Seger, M.M.2    Sandborg, M.3
  • 113
    • 33746402970 scopus 로고    scopus 로고
    • Disruptive technology: the conflict over PET/CT
    • Wiley G. Disruptive technology: the conflict over PET/CT. Imaging Economics 18 (2005) 17-20
    • (2005) Imaging Economics , vol.18 , pp. 17-20
    • Wiley, G.1
  • 114
    • 39049158018 scopus 로고    scopus 로고
    • The quest for the ideal anato-molecular imaging fusion tool
    • Zaidi H. The quest for the ideal anato-molecular imaging fusion tool. Biomed Imaging Interv J 2 (2006) e47
    • (2006) Biomed Imaging Interv J , vol.2
    • Zaidi, H.1
  • 115
  • 116
    • 33746445585 scopus 로고    scopus 로고
    • Putting 'clear' into nuclear medicine: a decade of PET/CT development
    • Beyer T., and Townsend D. Putting 'clear' into nuclear medicine: a decade of PET/CT development. Eur J Nucl Med Mol Imaging 33 (2006) 857-861
    • (2006) Eur J Nucl Med Mol Imaging , vol.33 , pp. 857-861
    • Beyer, T.1    Townsend, D.2
  • 117
    • 1642618344 scopus 로고    scopus 로고
    • PET/CT: panacea, redundancy, or something in between?
    • Vogel W.V., Oyen W.J., Barentsz J.O., et al. PET/CT: panacea, redundancy, or something in between?. J Nucl Med 45 Suppl 1 (2004) 15S-24S
    • (2004) J Nucl Med , vol.45 , Issue.SUPPL. 1
    • Vogel, W.V.1    Oyen, W.J.2    Barentsz, J.O.3
  • 118
    • 19844377102 scopus 로고    scopus 로고
    • Issues, controversies, and clinical utility of combined PET/CT imaging: what is the interpreting physician facing?
    • Blodgett T.M., Casagranda B., Townsend D.W., et al. Issues, controversies, and clinical utility of combined PET/CT imaging: what is the interpreting physician facing?. AJR Am J Roentgenol 184 (2005) S138-S145
    • (2005) AJR Am J Roentgenol , vol.184
    • Blodgett, T.M.1    Casagranda, B.2    Townsend, D.W.3
  • 119
    • 21144434055 scopus 로고    scopus 로고
    • The value of PET/CT is being over-sold as a clinical tool in radiation oncology
    • Xing L., and Wessels B. The value of PET/CT is being over-sold as a clinical tool in radiation oncology. Med Phys 32 (2005) 1457-1459
    • (2005) Med Phys , vol.32 , pp. 1457-1459
    • Xing, L.1    Wessels, B.2
  • 121
    • 4544239680 scopus 로고    scopus 로고
    • Retrospective digital image fusion of multidetector CT and 18F-FDG PET: clinical value in pancreatic lesions-a prospective study with 104 patients
    • Lemke A.J., Niehues S.M., Hosten N., et al. Retrospective digital image fusion of multidetector CT and 18F-FDG PET: clinical value in pancreatic lesions-a prospective study with 104 patients. J Nucl Med 45 (2004) 1279-1286
    • (2004) J Nucl Med , vol.45 , pp. 1279-1286
    • Lemke, A.J.1    Niehues, S.M.2    Hosten, N.3
  • 122
    • 34247593919 scopus 로고    scopus 로고
    • Point/counterpoint. Simultaneous PET/MR will replace PET/CT as the molecular multimodality imaging platform of choice
    • Zaidi H., Mawlawi O., and Orton C.G. Point/counterpoint. Simultaneous PET/MR will replace PET/CT as the molecular multimodality imaging platform of choice. Med Phys 34 (2007) 1525-1528
    • (2007) Med Phys , vol.34 , pp. 1525-1528
    • Zaidi, H.1    Mawlawi, O.2    Orton, C.G.3
  • 123
    • 0028606312 scopus 로고
    • Use of a magnetic field to increase the spatial resolution of positron emission tomography
    • Hammer B.E., Christensen N.L., and Heil B.G. Use of a magnetic field to increase the spatial resolution of positron emission tomography. Med Phys 21 (1994) 1917-1920
    • (1994) Med Phys , vol.21 , pp. 1917-1920
    • Hammer, B.E.1    Christensen, N.L.2    Heil, B.G.3
  • 124
    • 0028923795 scopus 로고
    • Positron emission tomography within a magnetic field using photomultiplier tubes and light-guides
    • Christensen N.L., Hammer B.E., Heil B.G., et al. Positron emission tomography within a magnetic field using photomultiplier tubes and light-guides. Phys Med Biol 40 (1995) 691-697
    • (1995) Phys Med Biol , vol.40 , pp. 691-697
    • Christensen, N.L.1    Hammer, B.E.2    Heil, B.G.3
  • 125
    • 0030215777 scopus 로고    scopus 로고
    • Combined MRI-PET scanner: a Monte-Carlo evaluation of the improvements in PET resolution due to the effects of a static homogeneous magnetic field
    • Raylman R.R., Hammer B.E., and Christensen N.L. Combined MRI-PET scanner: a Monte-Carlo evaluation of the improvements in PET resolution due to the effects of a static homogeneous magnetic field. IEEE Trans Nucl Sci 43 (1996) 2406-2412
    • (1996) IEEE Trans Nucl Sci , vol.43 , pp. 2406-2412
    • Raylman, R.R.1    Hammer, B.E.2    Christensen, N.L.3
  • 126
    • 0030801355 scopus 로고    scopus 로고
    • Simultaneous PET and MR imaging
    • Shao Y., Cherry S.R., Farahani K., et al. Simultaneous PET and MR imaging. Phys Med Biol 42 (1997) 1965-1970
    • (1997) Phys Med Biol , vol.42 , pp. 1965-1970
    • Shao, Y.1    Cherry, S.R.2    Farahani, K.3
  • 127
    • 0032765446 scopus 로고    scopus 로고
    • A study of artifacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner
    • Slates R., Farahani K., Shao Y., et al. A study of artifacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. Phys Med Biol 44 (1999) 2015-2027
    • (1999) Phys Med Biol , vol.44 , pp. 2015-2027
    • Slates, R.1    Farahani, K.2    Shao, Y.3
  • 128
    • 29144437800 scopus 로고    scopus 로고
    • Design and development of an MR-compatible PET scanner for imaging small animals
    • Mackewn J.E., Strul D., Hallett W.A., et al. Design and development of an MR-compatible PET scanner for imaging small animals. IEEE Trans Nucl Sci 52 (2005) 1376-1380
    • (2005) IEEE Trans Nucl Sci , vol.52 , pp. 1376-1380
    • Mackewn, J.E.1    Strul, D.2    Hallett, W.A.3
  • 129
    • 17644384354 scopus 로고    scopus 로고
    • A block detector for a multislice, depth-of-interaction MR-compatible PET
    • Yamamoto S., Takamatsu S., Murayama H., et al. A block detector for a multislice, depth-of-interaction MR-compatible PET. EEE Trans Nucl Sci 52 (2005) 33-37
    • (2005) EEE Trans Nucl Sci , vol.52 , pp. 33-37
    • Yamamoto, S.1    Takamatsu, S.2    Murayama, H.3
  • 130
    • 33846873900 scopus 로고    scopus 로고
    • Simultaneous MRI and PET imaging of a rat brain
    • Raylman R.R., Majewski S., Lemieux S.K., et al. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol 51 (2006) 6371-6379
    • (2006) Phys Med Biol , vol.51 , pp. 6371-6379
    • Raylman, R.R.1    Majewski, S.2    Lemieux, S.K.3
  • 131
    • 34250002531 scopus 로고    scopus 로고
    • Simultaneous acquisition of magnetic resonance spectroscopy (MRS) data and positron emission tomography (PET) images with a prototype MR-compatible, small animal PET imager
    • Raylman R.R., Majewski S., Velan S.S., et al. Simultaneous acquisition of magnetic resonance spectroscopy (MRS) data and positron emission tomography (PET) images with a prototype MR-compatible, small animal PET imager. J Magn Reson 186 (2007) 305-310
    • (2007) J Magn Reson , vol.186 , pp. 305-310
    • Raylman, R.R.1    Majewski, S.2    Velan, S.S.3
  • 133
    • 33646485976 scopus 로고    scopus 로고
    • Simulation of scattering and attenuation of 511 keV photons in a combined PET/field-cycled MRI system
    • Handler W.B., Gilbert K.M., Peng H., et al. Simulation of scattering and attenuation of 511 keV photons in a combined PET/field-cycled MRI system. Phys Med Biol 51 (2006) 2479-2491
    • (2006) Phys Med Biol , vol.51 , pp. 2479-2491
    • Handler, W.B.1    Gilbert, K.M.2    Peng, H.3
  • 134
    • 64749091450 scopus 로고    scopus 로고
    • Initial performance assessment of the LabPET APD-based digital PET scanner
    • [abstract]
    • Dumouchel T., Bergeron M., Cadorette J., et al. Initial performance assessment of the LabPET APD-based digital PET scanner. [abstract]. J Nucl Med 48 (2007) 39P
    • (2007) J Nucl Med , vol.48
    • Dumouchel, T.1    Bergeron, M.2    Cadorette, J.3
  • 135
    • 34247199659 scopus 로고    scopus 로고
    • Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner
    • Catana C., Wu Y., Judenhofer M.S., et al. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 47 (2006) 1968-1976
    • (2006) J Nucl Med , vol.47 , pp. 1968-1976
    • Catana, C.1    Wu, Y.2    Judenhofer, M.S.3
  • 136
    • 33745111311 scopus 로고    scopus 로고
    • Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI
    • Pichler B.J., Judenhofer M.S., Catana C., et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med 47 (2006) 639-647
    • (2006) J Nucl Med , vol.47 , pp. 639-647
    • Pichler, B.J.1    Judenhofer, M.S.2    Catana, C.3
  • 137
    • 34548087667 scopus 로고    scopus 로고
    • Simultaneous PET/MR images, acquired with a compact MRI compatible PET detector in a 7 Tesla magnet
    • Judenhofer M.S., Catana C., Swann B.K., et al. Simultaneous PET/MR images, acquired with a compact MRI compatible PET detector in a 7 Tesla magnet. Radiology 244 (2007) 807-814
    • (2007) Radiology , vol.244 , pp. 807-814
    • Judenhofer, M.S.1    Catana, C.2    Swann, B.K.3
  • 138
    • 33745403606 scopus 로고    scopus 로고
    • Status report on Silicon photomultiplier development and its applications
    • Dolgoshein B., et al. Status report on Silicon photomultiplier development and its applications. Nuclear Instr Meth A 563 (2006) 368-376
    • (2006) Nuclear Instr Meth A , vol.563 , pp. 368-376
    • Dolgoshein, B.1
  • 139
    • 33144468591 scopus 로고    scopus 로고
    • A detector head design for small-animal PET with silicon photomultipliers (SiPM)
    • Moehrs S., Del Guerra A., Herbert D.J., et al. A detector head design for small-animal PET with silicon photomultipliers (SiPM). Phys Med Biol 51 (2006) 1113-1127
    • (2006) Phys Med Biol , vol.51 , pp. 1113-1127
    • Moehrs, S.1    Del Guerra, A.2    Herbert, D.J.3
  • 142
    • 39049143111 scopus 로고    scopus 로고
    • Simultaneous MR/PET for brain imaging: first patient scans
    • [abstract]
    • Schlemmer H., Pichler B.J., Wienhard K., et al. Simultaneous MR/PET for brain imaging: first patient scans. [abstract]. J Nucl Med 48 (2007) 45P
    • (2007) J Nucl Med , vol.48
    • Schlemmer, H.1    Pichler, B.J.2    Wienhard, K.3
  • 143
    • 34548165833 scopus 로고    scopus 로고
    • Functional imaging of cancer with emphasis on molecular techniques
    • Torigian D.A., Huang S.S., Houseni M., et al. Functional imaging of cancer with emphasis on molecular techniques. CA Cancer J Clin 57 (2007) 206-224
    • (2007) CA Cancer J Clin , vol.57 , pp. 206-224
    • Torigian, D.A.1    Huang, S.S.2    Houseni, M.3
  • 144
  • 145
    • 0141642224 scopus 로고    scopus 로고
    • Medicine. The NIH Roadmap
    • Zerhouni E. Medicine. The NIH Roadmap. Science 302 (2003) 63-72
    • (2003) Science , vol.302 , pp. 63-72
    • Zerhouni, E.1
  • 146
    • 33747005619 scopus 로고    scopus 로고
    • Diffusion changes precede size reduction in neoadjuvant treatment of breast cancer
    • Pickles M.D., Gibbs P., Lowry M., et al. Diffusion changes precede size reduction in neoadjuvant treatment of breast cancer. Magn Reson Imaging 24 (2006) 843-847
    • (2006) Magn Reson Imaging , vol.24 , pp. 843-847
    • Pickles, M.D.1    Gibbs, P.2    Lowry, M.3
  • 147
    • 13944275976 scopus 로고    scopus 로고
    • Non-PET functional imaging techniques: optical
    • Intes X., and Chance B. Non-PET functional imaging techniques: optical. Radiol Clin North Am 43 (2005) 221-234
    • (2005) Radiol Clin North Am , vol.43 , pp. 221-234
    • Intes, X.1    Chance, B.2
  • 148
    • 13944274934 scopus 로고    scopus 로고
    • MR techniques for in vivo molecular and cellular imaging
    • Delikatny E.J., and Poptani H. MR techniques for in vivo molecular and cellular imaging. Radiol Clin North Am 43 (2005) 205-220
    • (2005) Radiol Clin North Am , vol.43 , pp. 205-220
    • Delikatny, E.J.1    Poptani, H.2
  • 149
    • 34548145140 scopus 로고    scopus 로고
    • Development of a cost-effective modular SPECT/CT scanner
    • Bailey D., Roach P., Bailey E., et al. Development of a cost-effective modular SPECT/CT scanner. Eur J Nucl Med Mol Imaging 34 (2007) 1415-1426
    • (2007) Eur J Nucl Med Mol Imaging , vol.34 , pp. 1415-1426
    • Bailey, D.1    Roach, P.2    Bailey, E.3