Volumn 19, Issue 2, 2008, Pages 122-126

New ways to measure intraocular pressure

Author keywords

Corneal thickness; Glaucoma; Intraocular pressure; Tonometry

Indexed keywords

ANXIETY; BIOMECHANICS; CORNEA; CORNEA THICKNESS; GLAUCOMA; HUMAN; HYSTERESIS; INTERMETHOD COMPARISON; INTRAOCULAR PRESSURE; OCULOGRAPHY; OCULOPLETHYSMOGRAPHY; PRESSURE MEASUREMENT; PRIORITY JOURNAL; REVIEW; VISION;

EID: 40049104749     PISSN: 10408738     EISSN: None     Source Type: Journal    
DOI: 10.1097/ICU.0b013e3282f391ae     Document Type: Review
Times cited : (49)

References (37)
  • 2
    • 0027269723 scopus 로고
    • The effect of corneal thickness on applanation tonometry
    • Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 1993; 115:592-596.
    • (1993) Am J Ophthalmol , vol.115 , pp. 592-596
    • Whitacre, M.M.1    Stein, R.A.2    Hassanein, K.3
  • 3
    • 0036272299 scopus 로고    scopus 로고
    • The Ocular Hypertension Treatment Study: Baseline factors that predict the onset of primary open-angle glaucoma
    • discussion 829-830
    • Gordon MO, Beiser JA, Brandt JD, et al. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol 2002; 120:714-720; discussion 829-830.
    • (2002) Arch Ophthalmol , vol.120 , pp. 714-720
    • Gordon, M.O.1    Beiser, J.A.2    Brandt, J.D.3
  • 4
    • 0034448216 scopus 로고    scopus 로고
    • Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma
    • Asrani S, Zeimer R, Wilensky J, et al. Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma. J Glaucoma 2000; 9:134-142.
    • (2000) J Glaucoma , vol.9 , pp. 134-142
    • Asrani, S.1    Zeimer, R.2    Wilensky, J.3
  • 5
    • 13844317411 scopus 로고    scopus 로고
    • Determining in vivo biomechanical properties of the cornea with an ocular response analyzer
    • Luce DA. Determining in vivo biomechanical properties of the cornea with an ocular response analyzer. J Cataract Refract Surg 2005; 31:156-162.
    • (2005) J Cataract Refract Surg , vol.31 , pp. 156-162
    • Luce, D.A.1
  • 6
    • 33748924156 scopus 로고    scopus 로고
    • Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer
    • ORA corneal-compensated intraocular pressure measurements are less influenced by corneal thickness than Goldmann pressure measurements. This study was conducted in nonglaucoma patients
    • Medeiros FA, Weinreb RN. Evaluation of the influence of corneal biomechanical properties on intraocular pressure measurements using the ocular response analyzer. J Glaucoma 2006; 15:364-370. ORA corneal-compensated intraocular pressure measurements are less influenced by corneal thickness than Goldmann pressure measurements. This study was conducted in nonglaucoma patients.
    • (2006) J Glaucoma , vol.15 , pp. 364-370
    • Medeiros, F.A.1    Weinreb, R.N.2
  • 7
    • 33750591541 scopus 로고    scopus 로고
    • Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements
    • ORA overestimates intraocular pressure compared with Goldmann tonometry. ORA readings seem to be affected by corneal thickness. This study was performed in glaucoma patients
    • Martinez-de-la-Casa JM, Garcia-Feijoo J, Fernandez-Vidal A, et al. Ocular response analyzer versus Goldmann applanation tonometry for intraocular pressure measurements. Invest Ophthalmol Vis Sci 2006; 47:4410-4414. ORA overestimates intraocular pressure compared with Goldmann tonometry. ORA readings seem to be affected by corneal thickness. This study was performed in glaucoma patients.
    • (2006) Invest Ophthalmol Vis Sci , vol.47 , pp. 4410-4414
    • Martinez-de-la-Casa, J.M.1    Garcia-Feijoo, J.2    Fernandez-Vidal, A.3
  • 8
    • 34248347895 scopus 로고    scopus 로고
    • Kotecha A, Elsheikh A, Roberts CR, et al. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006; 47:5337-5347. This study defines a new corneal measurement from the ORA called the corneal constant factor. This factor explains more of the interindividual variation in Goldmann intraocular pressure than does corneal thickness alone. ORA pressure measurements normalized for this new factor are not associated with corneal thickness.
    • Kotecha A, Elsheikh A, Roberts CR, et al. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006; 47:5337-5347. This study defines a new corneal measurement from the ORA called the corneal constant factor. This factor explains more of the interindividual variation in Goldmann intraocular pressure than does corneal thickness alone. ORA pressure measurements normalized for this new factor are not associated with corneal thickness.
  • 9
    • 25644460363 scopus 로고    scopus 로고
    • Dynamic contour tonometry: Presentation of a new tonometer
    • Kanngiesser HE, Kniestedt C, Robert YC. Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 2005; 14:344-350.
    • (2005) J Glaucoma , vol.14 , pp. 344-350
    • Kanngiesser, H.E.1    Kniestedt, C.2    Robert, Y.C.3
  • 10
    • 4444321895 scopus 로고    scopus 로고
    • Dynamic contour tonometry: A comparative study on human cadaver eyes
    • Kniestedt C, Nee M, Stamper RL. Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol 2004; 122:1287-1293.
    • (2004) Arch Ophthalmol , vol.122 , pp. 1287-1293
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 11
    • 20444489648 scopus 로고    scopus 로고
    • Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states
    • Kniestedt C, Nee M, Stamper RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 2005; 243:359-366.
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 359-366
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 12
    • 33751505840 scopus 로고    scopus 로고
    • Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry
    • Both rebound tonometry and dynamic contour tonometry pressure readings show excellent correlation with Goldmann tonometry. Both tonometers overestimate intraocular pressure compared with Goldmann tonometry
    • Martinez-de-la-Casa JM, Garcia-Feijoo J, Vico E, et al. Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry. Ophthalmology 2006; 113:2156-2162. Both rebound tonometry and dynamic contour tonometry pressure readings show excellent correlation with Goldmann tonometry. Both tonometers overestimate intraocular pressure compared with Goldmann tonometry.
    • (2006) Ophthalmology , vol.113 , pp. 2156-2162
    • Martinez-de-la-Casa, J.M.1    Garcia-Feijoo, J.2    Vico, E.3
  • 13
    • 33845499055 scopus 로고    scopus 로고
    • Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry
    • Dynamic contour tonometry pressure readings are relatively unchanged after LASIK, in comparison with Goldmann tonometry and measures obtained with ORA. There was no statistical change in the ocular pulse amplitude after LASIK
    • Pepose JS, Feigenbaum SK, Qazi MA, et al. Changes in corneal biomechanics and intraocular pressure following LASIK using static, dynamic, and noncontact tonometry. Am J Ophthalmol 2007; 143:39-47. Dynamic contour tonometry pressure readings are relatively unchanged after LASIK, in comparison with Goldmann tonometry and measures obtained with ORA. There was no statistical change in the ocular pulse amplitude after LASIK.
    • (2007) Am J Ophthalmol , vol.143 , pp. 39-47
    • Pepose, J.S.1    Feigenbaum, S.K.2    Qazi, M.A.3
  • 14
    • 33845807890 scopus 로고    scopus 로고
    • Francis BA, Hsieh A, Lai MY, et al. Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry. Ophthalmology 2007; 114:20-26. With 2157 participants, this is the largest study to look at dynamic contour tonometry in comparison with Goldmann tonometry. Dynamic contour tonometry was less affected by corneal thickness but was affected by corneal curvature. This effect was lower than the effect of corneal thickness on Goldmann tonometry, however.
    • Francis BA, Hsieh A, Lai MY, et al. Effects of corneal thickness, corneal curvature, and intraocular pressure level on Goldmann applanation tonometry and dynamic contour tonometry. Ophthalmology 2007; 114:20-26. With 2157 participants, this is the largest study to look at dynamic contour tonometry in comparison with Goldmann tonometry. Dynamic contour tonometry was less affected by corneal thickness but was affected by corneal curvature. This effect was lower than the effect of corneal thickness on Goldmann tonometry, however.
  • 15
    • 33750088433 scopus 로고    scopus 로고
    • Intraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometry
    • Dynamic contour tonometry may measure an intraocular pressure that is independent of corneal thickness
    • Punjabi OS, Ho HK, Kniestedt C, et al. Intraocular pressure and ocular pulse amplitude comparisons in different types of glaucoma using dynamic contour tonometry. Curr Eye Res 2006; 31:851-862. Dynamic contour tonometry may measure an intraocular pressure that is independent of corneal thickness.
    • (2006) Curr Eye Res , vol.31 , pp. 851-862
    • Punjabi, O.S.1    Ho, H.K.2    Kniestedt, C.3
  • 16
    • 33947600358 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry and Goldmann applanation tonometry in African American subjects
    • Dynamic contour tonometry measurements in African-Americans seem to be less dependent on corneal properties than Goldmann tonometry
    • Medeiros FA, Sample PA, Weinreb RN. Comparison of dynamic contour tonometry and Goldmann applanation tonometry in African American subjects. Ophthalmology 2007; 114:658-665. Dynamic contour tonometry measurements in African-Americans seem to be less dependent on corneal properties than Goldmann tonometry.
    • (2007) Ophthalmology , vol.114 , pp. 658-665
    • Medeiros, F.A.1    Sample, P.A.2    Weinreb, R.N.3
  • 17
    • 33748983782 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects
    • Dynamic contour tonometry measurements correlated strongly with Goldmann tonometry in both glaucomatous and healthy eyes
    • Barleon L, Hoffmann EM, Berres M, et al. Comparison of dynamic contour tonometry and Goldmann applanation tonometry in glaucoma patients and healthy subjects. Am J Ophthalmol 2006; 142:583-590. Dynamic contour tonometry measurements correlated strongly with Goldmann tonometry in both glaucomatous and healthy eyes.
    • (2006) Am J Ophthalmol , vol.142 , pp. 583-590
    • Barleon, L.1    Hoffmann, E.M.2    Berres, M.3
  • 19
    • 37349016169 scopus 로고    scopus 로고
    • The clinical utility of dynamic contour tonometry and ocular pulse amplitude
    • in press
    • Weizer JS, Asrani S, Stinnett SS, Herndon LW. The clinical utility of dynamic contour tonometry and ocular pulse amplitude. J Glaucoma (in press).
    • J Glaucoma
    • Weizer, J.S.1    Asrani, S.2    Stinnett, S.S.3    Herndon, L.W.4
  • 20
    • 33845245066 scopus 로고    scopus 로고
    • Rebound tonometry: New opportunities and limitations of noninvasive determination of intraocular pressure
    • Cervino A. Rebound tonometry: new opportunities and limitations of noninvasive determination of intraocular pressure. Br J Ophthalmol 2006; 90:1444-1446.
    • (2006) Br J Ophthalmol , vol.90 , pp. 1444-1446
    • Cervino, A.1
  • 21
    • 0030318702 scopus 로고    scopus 로고
    • A new electromechanical method for measuring intraocular pressure
    • Kontiola A. A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol 1996; 93:265-276.
    • (1996) Doc Ophthalmol , vol.93 , pp. 265-276
    • Kontiola, A.1
  • 22
    • 33646901276 scopus 로고    scopus 로고
    • Measuring intraocular pressure-adjustments for corneal thickness and new technologies
    • Herndon LW. Measuring intraocular pressure-adjustments for corneal thickness and new technologies. Curr Opin Ophthalmol 2006; 17:115-119.
    • (2006) Curr Opin Ophthalmol , vol.17 , pp. 115-119
    • Herndon, L.W.1
  • 23
    • 40049093896 scopus 로고    scopus 로고
    • Abraham LM, Epasinghe NC, Selva D, Casson R. Comparison of the ICare rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists. Eye 2007; 8 December. ICare rebound tonometry compares reasonably to Goldmann tonometry in both experienced and inexperienced hands.
    • Abraham LM, Epasinghe NC, Selva D, Casson R. Comparison of the ICare rebound tonometer with the Goldmann applanation tonometer by experienced and inexperienced tonometrists. Eye 2007; 8 December. ICare rebound tonometry compares reasonably to Goldmann tonometry in both experienced and inexperienced hands.
  • 24
    • 34247633160 scopus 로고    scopus 로고
    • Reproducibility and tolerability of the ICare rebound tonometer in school children
    • Schoolchildren are able to tolerate well intraocular pressure measurements with the iCare tonometer. Pressure measurements in this population were also highly reproducible
    • Sahin A, Basmak H, Niyaz L, Yildirim N. Reproducibility and tolerability of the ICare rebound tonometer in school children. JGlaucoma 2007; 16:185-188. Schoolchildren are able to tolerate well intraocular pressure measurements with the iCare tonometer. Pressure measurements in this population were also highly reproducible.
    • (2007) JGlaucoma , vol.16 , pp. 185-188
    • Sahin, A.1    Basmak, H.2    Niyaz, L.3    Yildirim, N.4
  • 25
    • 38849093096 scopus 로고    scopus 로고
    • The influence of corneal properties on rebound tonometry
    • Epub ahead of print
    • Chui WS, Lam A, Chen D, Chiu R. The influence of corneal properties on rebound tonometry. Ophthalmology 2007; [Epub ahead of print].
    • (2007) Ophthalmology
    • Chui, W.S.1    Lam, A.2    Chen, D.3    Chiu, R.4
  • 26
    • 34247567827 scopus 로고    scopus 로고
    • Moreno-Montanes J, Garcia N, Fernandez-Hortelano A, Garcia-Layana A. Rebound tonometer compared with Goldmann tonometer in normal and pathologic corneas. Cornea 2007; 26:427-430. Rebound tonometry measurements were similar to Goldmann measures in pathologic corneas. In some corneas, rebound tonometry was able to obtain a pressure measurement whereas Goldmann tonometry was not.
    • Moreno-Montanes J, Garcia N, Fernandez-Hortelano A, Garcia-Layana A. Rebound tonometer compared with Goldmann tonometer in normal and pathologic corneas. Cornea 2007; 26:427-430. Rebound tonometry measurements were similar to Goldmann measures in pathologic corneas. In some corneas, rebound tonometry was able to obtain a pressure measurement whereas Goldmann tonometry was not.
  • 27
    • 34447631350 scopus 로고    scopus 로고
    • Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients
    • In glaucoma patients, rebound tonometry pressure measurements seem to be more affected by corneal thickness than Goldmann measurements
    • Sahin A, Niyaz L, Yildirim N. Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients. Clin Expeiment Ophthalmol 2007; 35:335-339. In glaucoma patients, rebound tonometry pressure measurements seem to be more affected by corneal thickness than Goldmann measurements.
    • (2007) Clin Expeiment Ophthalmol , vol.35 , pp. 335-339
    • Sahin, A.1    Niyaz, L.2    Yildirim, N.3
  • 28
    • 33746277068 scopus 로고    scopus 로고
    • Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers
    • The ICare tonometer is influenced by corneal thickness and agrees well with applanation tonometers
    • Nakamura M, Darhad U, Tatsumi Y, et al. Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. Am J Ophthalmol 2006; 142:332-334. The ICare tonometer is influenced by corneal thickness and agrees well with applanation tonometers.
    • (2006) Am J Ophthalmol , vol.142 , pp. 332-334
    • Nakamura, M.1    Darhad, U.2    Tatsumi, Y.3
  • 29
    • 33745667304 scopus 로고    scopus 로고
    • Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness
    • Iliev ME, Goldblum D, Katsoulis K, et al. Comparison of rebound tonometry with Goldmann applanation tonometry and correlation with central corneal thickness. Br J Ophthalmol 2006; 90:833-835.
    • (2006) Br J Ophthalmol , vol.90 , pp. 833-835
    • Iliev, M.E.1    Goldblum, D.2    Katsoulis, K.3
  • 31
    • 0032471383 scopus 로고    scopus 로고
    • A new tonometer: The pressure phosphene tonometer: clinical comparison with Goldman tonometry
    • Fresco BB. A new tonometer: the pressure phosphene tonometer: clinical comparison with Goldman tonometry. Ophthalmology 1998; 105:2123-2126.
    • (1998) Ophthalmology , vol.105 , pp. 2123-2126
    • Fresco, B.B.1
  • 32
    • 15244345047 scopus 로고    scopus 로고
    • Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intraocular pressure readings
    • Rai S, Moster MR, Kesen M, et al. Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intraocular pressure readings. J Glaucoma 2005; 14:120-123.
    • (2005) J Glaucoma , vol.14 , pp. 120-123
    • Rai, S.1    Moster, M.R.2    Kesen, M.3
  • 33
    • 2942601243 scopus 로고    scopus 로고
    • The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice
    • Alvarez TL, Gollance SA, Thomas GA, et al. The Proview phosphene tonometer fails to measure ocular pressure accurately in clinical practice. Ophthalmology 2004; 111:1077-1085.
    • (2004) Ophthalmology , vol.111 , pp. 1077-1085
    • Alvarez, T.L.1    Gollance, S.A.2    Thomas, G.A.3
  • 34
    • 3543128942 scopus 로고    scopus 로고
    • Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen
    • Li J, Herndon LW, Asrani SG, et al. Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen. Arch Ophthalmol 2004; 122:1117-1121.
    • (2004) Arch Ophthalmol , vol.122 , pp. 1117-1121
    • Li, J.1    Herndon, L.W.2    Asrani, S.G.3
  • 35
    • 34547292477 scopus 로고    scopus 로고
    • Evaluation of some factors affecting the agreement between the Proview Eye Pressure Monitor and the Goldmann applanation tonometer measurements
    • Overall agreement between Goldmann tonometry and the Proview phosphene tonometer was poor. Additional training did not influence Proview tonometer measurements
    • Gunvant P, Lievens CW, Newman JM 3rd, et al. Evaluation of some factors affecting the agreement between the Proview Eye Pressure Monitor and the Goldmann applanation tonometer measurements. Clin Exp Optom 2007; 90:290-295. Overall agreement between Goldmann tonometry and the Proview phosphene tonometer was poor. Additional training did not influence Proview tonometer measurements.
    • (2007) Clin Exp Optom , vol.90 , pp. 290-295
    • Gunvant, P.1    Lievens, C.W.2    Newman 3rd, J.M.3
  • 36
    • 33751052080 scopus 로고    scopus 로고
    • A clinical evaluation of Proview pressure phosphene tonometry in children
    • Chung I, Bartolone A, Swanson WH, Thau AP. A clinical evaluation of Proview pressure phosphene tonometry in children. Optom Vis Sci 2006; 83:817-822.
    • (2006) Optom Vis Sci , vol.83 , pp. 817-822
    • Chung, I.1    Bartolone, A.2    Swanson, W.H.3    Thau, A.P.4
  • 37
    • 28244464854 scopus 로고    scopus 로고
    • The Proview Eye Pressure Monitor: Influence of clinical factors on accuracy and agreement with the Goldmann tonometer
    • Herse P, Hans A, Hall J, et al. The Proview Eye Pressure Monitor: influence of clinical factors on accuracy and agreement with the Goldmann tonometer. Ophthalmic Physiol Opt 2005; 25:416-420.
    • (2005) Ophthalmic Physiol Opt , vol.25 , pp. 416-420
    • Herse, P.1    Hans, A.2    Hall, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.