Volumn 35, Issue 4, 2007, Pages 335-339

Comparison of the rebound tonometer with the Goldmann applanation tonometer in glaucoma patients

Author keywords

Glaucoma; Goldmann applanation tonometer; Intraocular pressure; Rebound tonometer

Indexed keywords

ADULT; AGED; ARTICLE; CONTROLLED STUDY; CORNEA THICKNESS; CORRELATION ANALYSIS; DIAGNOSTIC ACCURACY; DIAGNOSTIC VALUE; FEMALE; GLAUCOMA; HUMAN; INTERMETHOD COMPARISON; INTRAOCULAR PRESSURE; LINEAR REGRESSION ANALYSIS; MAJOR CLINICAL STUDY; MALE; PRESSURE MEASUREMENT; STUDENT T TEST; TONOMETER; TONOMETRY; COMPARATIVE STUDY; CORNEA; ECHOGRAPHY; HISTOLOGY; METHODOLOGY; MIDDLE AGED; OCULOPLETHYSMOGRAPHY; REPRODUCIBILITY;

EID: 34447631350     PISSN: 14426404     EISSN: 14429071     Source Type: Journal    
DOI: 10.1111/j.1442-9071.2007.01451.x     Document Type: Article
Times cited : (41)

References (15)
  • 1
    • 0020955893 scopus 로고
    • The epidemiology of open-angle glaucoma: A review
    • Leske MC. The epidemiology of open-angle glaucoma: A review. Am J Epidemiol 1983; 118: 166-91.
    • (1983) Am J Epidemiol , vol.118 , pp. 166-191
    • Leske, M.C.1
  • 2
    • 33646240338 scopus 로고    scopus 로고
    • Contamination of disposable tonometer prisms during tonometry
    • Rajak SN, Paul J, Sharma V, Vickers S. Contamination of disposable tonometer prisms during tonometry. Eye 2006; 20: 358-61.
    • (2006) Eye , vol.20 , pp. 358-361
    • Rajak, S.N.1    Paul, J.2    Sharma, V.3    Vickers, S.4
  • 3
    • 0032083291 scopus 로고    scopus 로고
    • Sterilisation of tonometers and gonioscopes
    • Sood D, Honavar SG. Sterilisation of tonometers and gonioscopes. Indian J Ophthalmol 1998; 46: 113-16.
    • (1998) Indian J Ophthalmol , vol.46 , pp. 113-116
    • Sood, D.1    Honavar, S.G.2
  • 5
    • 33745282996 scopus 로고    scopus 로고
    • Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients
    • Brusini P, Salvetat ML, Zeppieri M, Tosoni C, Parisi L. Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma 2006; 15: 213-17.
    • (2006) J Glaucoma , vol.15 , pp. 213-217
    • Brusini, P.1    Salvetat, M.L.2    Zeppieri, M.3    Tosoni, C.4    Parisi, L.5
  • 7
    • 33746277068 scopus 로고    scopus 로고
    • Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers
    • Nakamura M, Darhad U, Tatsumi Y et al. Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. Am J Ophthalmol 2006; 142: 332-4.
    • (2006) Am J Ophthalmol , vol.142 , pp. 332-334
    • Nakamura, M.1    Darhad, U.2    Tatsumi, Y.3
  • 8
    • 33751505840 scopus 로고    scopus 로고
    • Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry
    • Martinez-de-la-Casa JM, Garcia-Feijoo J, Vico E et al. Effect of corneal thickness on dynamic contour, rebound, and Goldmann tonometry. Ophthalmology 2006; 113: 2156-62.
    • (2006) Ophthalmology , vol.113 , pp. 2156-2162
    • Martinez-de-la-Casa, J.M.1    Garcia-Feijoo, J.2    Vico, E.3
  • 9
    • 0021823424 scopus 로고
    • Statistical analysis of multi-eye data in ophthalmic research
    • Ray WA, O'Day DM. Statistical analysis of multi-eye data in ophthalmic research. Invest Ophthalmol Vis Sci 1985; 26: 1186-8.
    • (1985) Invest Ophthalmol Vis Sci , vol.26 , pp. 1186-1188
    • Ray, W.A.1    O'Day, D.M.2
  • 10
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-10.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 11
    • 0030318702 scopus 로고    scopus 로고
    • A new electromechanical method for measuring intraocular pressure
    • Kontiola A. A new electromechanical method for measuring intraocular pressure. Doc Ophthalmol 1996; 93: 265-76.
    • (1996) Doc Ophthalmol , vol.93 , pp. 265-276
    • Kontiola, A.1
  • 12
    • 0034115306 scopus 로고    scopus 로고
    • A new induction-based impact method for measuring intraocular pressure
    • Kontiola AI. A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol Scand 2000; 78: 142-5.
    • (2000) Acta Ophthalmol Scand , vol.78 , pp. 142-145
    • Kontiola, A.I.1
  • 14
    • 28244471011 scopus 로고    scopus 로고
    • Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry
    • van der Jagt LH, Jansonius NM. Three portable tonometers, the TGDc-01, the ICARE and the Tonopen XL, compared with each other and with Goldmann applanation tonometry. Ophthalmic Physiol Opt 2005; 25: 429-35.
    • (2005) Ophthalmic Physiol Opt , vol.25 , pp. 429-435
    • van der Jagt, L.H.1    Jansonius, N.M.2
  • 15
    • 33644629030 scopus 로고    scopus 로고
    • Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes
    • Garcia-Resua C, Gonzalez-Meijome JM, Gilino J, Yebra-Pimentel E. Accuracy of the new ICare rebound tonometer vs. other portable tonometers in healthy eyes. Optom Vis Sci 2006; 83: 102-7.
    • (2006) Optom Vis Sci , vol.83 , pp. 102-107
    • Garcia-Resua, C.1    Gonzalez-Meijome, J.M.2    Gilino, J.3    Yebra-Pimentel, E.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.