Comparison of Dynamic Contour Tonometry and Goldmann Applanation Tonometry in Glaucoma Patients and Healthy Subjects

Author keywords

[No Author keywords available]

Indexed keywords

ADULT; AGED; ARTICLE; CONTROLLED STUDY; CORRELATION ANALYSIS; EVALUATION; FEMALE; GLAUCOMA; HUMAN; MAJOR CLINICAL STUDY; MALE; PRIORITY JOURNAL; PROSPECTIVE STUDY; REGRESSION ANALYSIS; TONOMETRY;

EID: 33748983782     PISSN: 00029394     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.ajo.2006.05.030     Document Type: Article
Times cited : (66)

References (35)
  • 1
    • 1842504236 scopus 로고    scopus 로고
    • Corneal thickness in glaucoma screening, diagnosis, and management
    • Brandt J.D. Corneal thickness in glaucoma screening, diagnosis, and management. Curr Opin Ophthalmol 15 (2004) 85-89
    • (2004) Curr Opin Ophthalmol , vol.15 , pp. 85-89
    • Brandt, J.D.1
  • 2
    • 0028273231 scopus 로고
    • Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies
    • Dielemans I., Vingerling J.R., Hofman A., et al. Reliability of intraocular pressure measurement with the Goldmann applanation tonometer in epidemiological studies. Graefes Arch Clin Exp Ophthalmol 232 (1994) 141-144
    • (1994) Graefes Arch Clin Exp Ophthalmol , vol.232 , pp. 141-144
    • Dielemans, I.1    Vingerling, J.R.2    Hofman, A.3
  • 3
    • 70449195078 scopus 로고
    • Über Applanationsttonometrie
    • Goldmann H., and Schmidt T. Über Applanationsttonometrie. Ophthalmologica 134 (1957) 221-242
    • (1957) Ophthalmologica , vol.134 , pp. 221-242
    • Goldmann, H.1    Schmidt, T.2
  • 4
    • 0025614886 scopus 로고
    • Tonometer utilization, accuracy, and calibration under field conditions
    • Wessels I.F., and Oh Y. Tonometer utilization, accuracy, and calibration under field conditions. Arch Ophthalmol 108 (1990) 1709-1712
    • (1990) Arch Ophthalmol , vol.108 , pp. 1709-1712
    • Wessels, I.F.1    Oh, Y.2
  • 5
    • 0027291601 scopus 로고
    • Sources of error with use of Goldmann-type tonometers
    • Whitacre M.M., and Stein R. Sources of error with use of Goldmann-type tonometers. Surv Ophthalmol 38 (1993) 1-30
    • (1993) Surv Ophthalmol , vol.38 , pp. 1-30
    • Whitacre, M.M.1    Stein, R.2
  • 6
    • 0030963090 scopus 로고    scopus 로고
    • Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study
    • Wolfs R.C., Klaver C.C., Vingerling J.R., et al. Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study. Am J Ophthalmol 123 (1997) 767-772
    • (1997) Am J Ophthalmol , vol.123 , pp. 767-772
    • Wolfs, R.C.1    Klaver, C.C.2    Vingerling, J.R.3
  • 8
    • 0027269723 scopus 로고
    • The effect of corneal thickness on applanation tonometry
    • Whitacre M.M., Stein R.A., and Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol 115 (1993) 592-596
    • (1993) Am J Ophthalmol , vol.115 , pp. 592-596
    • Whitacre, M.M.1    Stein, R.A.2    Hassanein, K.3
  • 9
    • 2942596237 scopus 로고    scopus 로고
    • Intraocular pressure and ocular pulse amplitude using dynamic contour tonometry and contact lens tonometry
    • Hoffmann E.M., Grus F.H., and Pfeiffer N. Intraocular pressure and ocular pulse amplitude using dynamic contour tonometry and contact lens tonometry. BMC Ophthalmol 4 (2004) 4
    • (2004) BMC Ophthalmol , vol.4 , pp. 4
    • Hoffmann, E.M.1    Grus, F.H.2    Pfeiffer, N.3
  • 10
    • 20444489648 scopus 로고    scopus 로고
    • Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states
    • Kniestedt C., Nee M., and Stamper R.L. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol 243 (2005) 359-366
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 359-366
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 11
    • 0022780097 scopus 로고
    • Comparison of methods of measuring blood pressure
    • Altman D.G., and Bland J.M. Comparison of methods of measuring blood pressure. J Epidemiol Commun Health 40 (1986) 274-277
    • (1986) J Epidemiol Commun Health , vol.40 , pp. 274-277
    • Altman, D.G.1    Bland, J.M.2
  • 13
    • 0033498048 scopus 로고    scopus 로고
    • Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic
    • Shah S., Chatterjee A., Mathai M., et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology 106 (1999) 2154-2160
    • (1999) Ophthalmology , vol.106 , pp. 2154-2160
    • Shah, S.1    Chatterjee, A.2    Mathai, M.3
  • 14
    • 0031848594 scopus 로고    scopus 로고
    • Applanation tonometry and correction according to corneal thickness
    • Stodtmeister R. Applanation tonometry and correction according to corneal thickness. Acta Ophthalmol Scand 76 (1998) 319-324
    • (1998) Acta Ophthalmol Scand , vol.76 , pp. 319-324
    • Stodtmeister, R.1
  • 15
    • 3042667876 scopus 로고    scopus 로고
    • Dynamic contour tonometry for post-LASIK intraocular pressure measurements
    • Duba I., and Wirthlin A.C. Dynamic contour tonometry for post-LASIK intraocular pressure measurements. Klin Monatsbl Augenheilkd 221 (2004) 347-350
    • (2004) Klin Monatsbl Augenheilkd , vol.221 , pp. 347-350
    • Duba, I.1    Wirthlin, A.C.2
  • 16
    • 0041360373 scopus 로고    scopus 로고
    • Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis
    • Kaufmann C., Bachmann L.M., and Thiel M.A. Intraocular pressure measurements using dynamic contour tonometry after laser in situ keratomileusis. Invest Ophthalmol Vis Sci 44 (2003) 3790-3794
    • (2003) Invest Ophthalmol Vis Sci , vol.44 , pp. 3790-3794
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 17
    • 13844318981 scopus 로고    scopus 로고
    • Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK
    • Siganos D.S., Papastergiou G.I., and Moedas C. Assessment of the Pascal dynamic contour tonometer in monitoring intraocular pressure in unoperated eyes and eyes after LASIK. J Cataract Refract Surg 30 (2004) 746-751
    • (2004) J Cataract Refract Surg , vol.30 , pp. 746-751
    • Siganos, D.S.1    Papastergiou, G.I.2    Moedas, C.3
  • 18
    • 0029609715 scopus 로고
    • Ocular hypertension and central corneal thickness
    • Argus W.A. Ocular hypertension and central corneal thickness. Ophthalmology 102 (1995) 1810-1812
    • (1995) Ophthalmology , vol.102 , pp. 1810-1812
    • Argus, W.A.1
  • 19
    • 0034803965 scopus 로고    scopus 로고
    • Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS)
    • Brandt J.D., Beiser J.A., Kass M.A., et al. Central corneal thickness in the Ocular Hypertension Treatment Study (OHTS). Ophthalmology 108 (2001) 1779-1788
    • (2001) Ophthalmology , vol.108 , pp. 1779-1788
    • Brandt, J.D.1    Beiser, J.A.2    Kass, M.A.3
  • 20
    • 0141682012 scopus 로고    scopus 로고
    • Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African-Americans
    • Shimmyo M., Ross A.J., Moy A., et al. Intraocular pressure, Goldmann applanation tension, corneal thickness, and corneal curvature in Caucasians, Asians, Hispanics, and African-Americans. Am J Ophthalmol 136 (2003) 603-613
    • (2003) Am J Ophthalmol , vol.136 , pp. 603-613
    • Shimmyo, M.1    Ross, A.J.2    Moy, A.3
  • 21
    • 25644460363 scopus 로고    scopus 로고
    • Dynamic contour tonometry: presentation of a new tonometer
    • Kanngiesser H.E., Kniestedt C., and Robert Y.C. Dynamic contour tonometry: presentation of a new tonometer. J Glaucoma 14 (2005) 344-350
    • (2005) J Glaucoma , vol.14 , pp. 344-350
    • Kanngiesser, H.E.1    Kniestedt, C.2    Robert, Y.C.3
  • 22
    • 4344659799 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with Goldmann applanation tonometry
    • Kaufmann C., Bachmann L.M., and Thiel M.A. Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci 45 (2004) 3118-3121
    • (2004) Invest Ophthalmol Vis Sci , vol.45 , pp. 3118-3121
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 23
    • 4444321895 scopus 로고    scopus 로고
    • Dynamic contour tonometry: a comparative study on human cadaver eyes
    • Kniestedt C., Nee M., and Stamper R.L. Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol 122 (2004) 1287-1293
    • (2004) Arch Ophthalmol , vol.122 , pp. 1287-1293
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 24
    • 33748986481 scopus 로고    scopus 로고
    • Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry
    • Forthcoming
    • Ku J.Y., Danesh-Meyer H.V., Craig J.P., et al. Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye (2005) Forthcoming
    • (2005) Eye
    • Ku, J.Y.1    Danesh-Meyer, H.V.2    Craig, J.P.3
  • 25
    • 22544486553 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness
    • Doyle A., and Lachkar Y. Comparison of dynamic contour tonometry with Goldman applanation tonometry over a wide range of central corneal thickness. J Glaucoma 14 (2005) 288-292
    • (2005) J Glaucoma , vol.14 , pp. 288-292
    • Doyle, A.1    Lachkar, Y.2
  • 26
    • 23744463890 scopus 로고    scopus 로고
    • Dynamic contour tonometry for intraocular pressure measurement
    • Kamppeter B.A., and Jonas J.B. Dynamic contour tonometry for intraocular pressure measurement. Am J Ophthalmol 140 (2005) 318-320
    • (2005) Am J Ophthalmol , vol.140 , pp. 318-320
    • Kamppeter, B.A.1    Jonas, J.B.2
  • 27
    • 27744519845 scopus 로고    scopus 로고
    • Clinical comparison of contour and applanation tonometry and their relationship to pachymetry
    • Kniestedt C., Lin S., Choe J., et al. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol 123 (2005) 1532-1557
    • (2005) Arch Ophthalmol , vol.123 , pp. 1532-1557
    • Kniestedt, C.1    Lin, S.2    Choe, J.3
  • 28
    • 28444475214 scopus 로고    scopus 로고
    • The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    • Kotecha A., White E.T., Shewry J.M., et al. The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry. Br J Ophthalmol 89 (2005) 1572-1575
    • (2005) Br J Ophthalmol , vol.89 , pp. 1572-1575
    • Kotecha, A.1    White, E.T.2    Shewry, J.M.3
  • 29
    • 33748973706 scopus 로고    scopus 로고
    • Swiss-Microtechnology-AG. PASCAL. Dynamic contour tonometer. User manual, Version 2.0, 2004:1-61.
  • 31
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland J.M., and Altman D.G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1 (1986) 307-310
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 32
    • 0036016618 scopus 로고    scopus 로고
    • Statistical techniques for comparing measurers and methods of measurement: a critical review
    • Ludbrook J. Statistical techniques for comparing measurers and methods of measurement: a critical review. Clin Exp Pharmacol Physiol 29 (2002) 527-536
    • (2002) Clin Exp Pharmacol Physiol , vol.29 , pp. 527-536
    • Ludbrook, J.1
  • 33
    • 24944580479 scopus 로고    scopus 로고
    • Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study
    • Pache M., Wilmsmeyer S., Lautebach S., et al. Dynamic contour tonometry versus Goldmann applanation tonometry: a comparative study. Graefes Arch Clin Exp Ophthalmol 243 (2005) 763-767
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 763-767
    • Pache, M.1    Wilmsmeyer, S.2    Lautebach, S.3
  • 34
    • 0029092032 scopus 로고
    • A comparison of methods for correlated ordinal measures with ophthalmic applications
    • Gange S.J., Linton K.L., Scott A.J., et al. A comparison of methods for correlated ordinal measures with ophthalmic applications. Stat Med 14 (1995) 1961-1974
    • (1995) Stat Med , vol.14 , pp. 1961-1974
    • Gange, S.J.1    Linton, K.L.2    Scott, A.J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.