Volumn 43, Issue 4, 2016, Pages 1982-1994

IMRT QA: Selecting gamma criteria based on error detection sensitivity

Author keywords

error curve method; gamma criteria sensitivity; gamma passing rates; IMRT QA; IMRT QA tolerances

Indexed keywords

PIXELS; QUALITY CONTROL; RADIOTHERAPY; RANDOM ERRORS; SYSTEMATIC ERRORS;

EID: 84963770158     PISSN: 00942405     EISSN: 24734209     Source Type: Journal    
DOI: 10.1118/1.4943953     Document Type: Article
Times cited : (51)

References (49)
  • 1
    • 10744231470 scopus 로고    scopus 로고
    • Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee
    • G. A. Ezzell, J. M. Galvin, D. Low, J. R. Palta, I. Rosen, M. B. Sharpe, P. Xia, Y. Xiao, L. Xing, and X. Y. Cedric, "Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT Subcommittee of the AAPM Radiation Therapy Committee," Med. Phys. 30, 2089-2115 (2003). 10.1118/1.1591194
    • (2003) Med. Phys. , vol.30 , pp. 2089-2115
    • Ezzell, G.A.1    Galvin, J.M.2    Low, D.3    Palta, J.R.4    Rosen, I.5    Sharpe, M.B.6    Xia, P.7    Xiao, Y.8    Xing, L.9    Cedric, X.Y.10
  • 2
    • 79952172951 scopus 로고    scopus 로고
    • Dosimetry tools and techniques for IMRT
    • D. A. Low, J. M. Moran, J. F. Dempsey, L. Dong, and M. Oldham, "Dosimetry tools and techniques for IMRT," Med. Phys. 38, 1313-1338 (2011). 10.1118/1.3514120
    • (2011) Med. Phys. , vol.38 , pp. 1313-1338
    • Low, D.A.1    Moran, J.M.2    Dempsey, J.F.3    Dong, L.4    Oldham, M.5
  • 4
    • 84870187312 scopus 로고    scopus 로고
    • American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) practice guideline for intensity-modulated radiation therapy (IMRT)
    • A. C. Hartford, J. M. Galvin, D. C. Beyer, T. J. Eichler, G. S. Ibbott, B. Kavanagh, C. J. Schultz, and S. A. Rosenthal, "American College of Radiology (ACR) and American Society for Radiation Oncology (ASTRO) practice guideline for intensity-modulated radiation therapy (IMRT)," Am. J. Clin. Oncol. 35, 612-617 (2012). 10.1097/COC.0b013e31826e0515
    • (2012) Am. J. Clin. Oncol. , vol.35 , pp. 612-617
    • Hartford, A.C.1    Galvin, J.M.2    Beyer, D.C.3    Eichler, T.J.4    Ibbott, G.S.5    Kavanagh, B.6    Schultz, C.J.7    Rosenthal, S.A.8
  • 5
    • 0031861573 scopus 로고    scopus 로고
    • A technique for the quantitative evaluation of dose distributions
    • D. A. Low, W. B. Harms, S. Mutic, and J. A. Purdy, "A technique for the quantitative evaluation of dose distributions," Med. Phys. 25, 656-661 (1998). 10.1118/1.598248
    • (1998) Med. Phys. , vol.25 , pp. 656-661
    • Low, D.A.1    Harms, W.B.2    Mutic, S.3    Purdy, J.A.4
  • 6
    • 0141520680 scopus 로고    scopus 로고
    • Evaluation of the gamma dose distribution comparison method
    • D. A. Low and J. F. Dempsey, "Evaluation of the gamma dose distribution comparison method," Med. Phys. 30, 2455-2464 (2003). 10.1118/1.1598711
    • (2003) Med. Phys. , vol.30 , pp. 2455-2464
    • Low, D.A.1    Dempsey, J.F.2
  • 8
    • 34249947757 scopus 로고    scopus 로고
    • A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy
    • S. Both, I. M. Alecu, A. R. Stan, M. Alecu, A. Ciura, J. M. Hansen, and R. Alecu, "A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy," J. Appl. Clin. Med. Phys. 8, 1-8 (2007). 10.1120/jacmp.v8i2.2374
    • (2007) J. Appl. Clin. Med. Phys. , vol.8 , pp. 1-8
    • Both, S.1    Alecu, I.M.2    Stan, A.R.3    Alecu, M.4    Ciura, A.5    Hansen, J.M.6    Alecu, R.7
  • 9
    • 57049164216 scopus 로고    scopus 로고
    • Establishing action levels for EPID-based QA for IMRT
    • R. M. Howell, I. P. Smith, and C. S. Jarrio, "Establishing action levels for EPID-based QA for IMRT," J. Appl. Clin. Med. Phys. 9, 16-25 (2008). 10.1120/jacmp.v9i3.2721
    • (2008) J. Appl. Clin. Med. Phys. , vol.9 , pp. 16-25
    • Howell, R.M.1    Smith, I.P.2    Jarrio, C.S.3
  • 10
    • 44349152982 scopus 로고    scopus 로고
    • An analysis of tolerance levels in IMRT quality assurance procedures
    • P. S. Basran and M. K. Woo, "An analysis of tolerance levels in IMRT quality assurance procedures," Med. Phys. 35, 2300-2307 (2008). 10.1118/1.2919075
    • (2008) Med. Phys. , vol.35 , pp. 2300-2307
    • Basran, P.S.1    Woo, M.K.2
  • 11
    • 78049325395 scopus 로고    scopus 로고
    • Practical guidelines for routine intensity-modulated radiotherapy verification: Pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry
    • A. J. Vinall, A. J. Williams, V. E. Currie, A. Van Esch, and D. Huyskens, "Practical guidelines for routine intensity-modulated radiotherapy verification: Pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry," Br. J. Radiol. 83, 949-957 (2010). 10.1259/bjr/31573847
    • (2010) Br. J. Radiol. , vol.83 , pp. 949-957
    • Vinall, A.J.1    Williams, A.J.2    Currie, V.E.3    Van Esch, A.4    Huyskens, D.5
  • 12
    • 84864701518 scopus 로고    scopus 로고
    • 3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification
    • P. Carrasco, N. Jornet, A. Latorre, T. Eudaldo, A. Ruiz, and M. Ribas, "3D DVH-based metric analysis versus per-beam planar analysis in IMRT pretreatment verification," Med. Phys. 39, 5040-5049 (2012). 10.1118/1.4736949
    • (2012) Med. Phys. , vol.39 , pp. 5040-5049
    • Carrasco, P.1    Jornet, N.2    Latorre, A.3    Eudaldo, T.4    Ruiz, A.5    Ribas, M.6
  • 13
    • 84860705093 scopus 로고    scopus 로고
    • Sensitivity of a helical diode array device to delivery errors in IMRT treatment and establishment of tolerance level for pretreatment QA
    • F. Garcia-Vicente, V. Fernandez, R. Bermudez, A. Gomez, L. Perez, A. Zapatero, and J. J. Torres, "Sensitivity of a helical diode array device to delivery errors in IMRT treatment and establishment of tolerance level for pretreatment QA," J. Appl. Clin. Med. Phys. 13, 111-123 (2012). 10.1120/jacmp.v13i1.3660
    • (2012) J. Appl. Clin. Med. Phys. , vol.13 , pp. 111-123
    • Garcia-Vicente, F.1    Fernandez, V.2    Bermudez, R.3    Gomez, A.4    Perez, L.5    Zapatero, A.6    Torres, J.J.7
  • 15
    • 84923463366 scopus 로고    scopus 로고
    • Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification
    • X. Jin, H. Yan, C. Han, Y. Zhou, J. Yi, and C. Xie, "Correlation between gamma index passing rate and clinical dosimetric difference for pre-treatment 2D and 3D volumetric modulated arc therapy dosimetric verification," Br. J. Radiol. 88, 20140577 (2015). 10.1259/bjr.20140577
    • (2015) Br. J. Radiol. , vol.88 , pp. 20140577
    • Jin, X.1    Yan, H.2    Han, C.3    Zhou, Y.4    Yi, J.5    Xie, C.6
  • 16
    • 84922842255 scopus 로고    scopus 로고
    • The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery
    • A. K. Templeton, J. C. Chu, and J. V. Turian, "The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery," J. Appl. Clin. Med. Phys. 16, 32-39 (2015). 10.1120/jacmp.v16i1.4814
    • (2015) J. Appl. Clin. Med. Phys. , vol.16 , pp. 32-39
    • Templeton, A.K.1    Chu, J.C.2    Turian, J.V.3
  • 17
    • 84948718537 scopus 로고    scopus 로고
    • Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors
    • L. Vieillevigne, J. Molinier, T. Brun, and R. Ferrand, "Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors," Phys. Med. 31, 720-725 (2015). 10.1016/j.ejmp.2015.05.016
    • (2015) Phys. Med. , vol.31 , pp. 720-725
    • Vieillevigne, L.1    Molinier, J.2    Brun, T.3    Ferrand, R.4
  • 18
    • 70049110836 scopus 로고    scopus 로고
    • On the sensitivity of patient-specific IMRT QA to MLC positioning errors
    • G. Yan, C. Liu, T. Simon, L.-C. Peng, C. Fox, and J. Li, "On the sensitivity of patient-specific IMRT QA to MLC positioning errors," J. Appl. Clin. Med. Phys. 10, 120-128 (2009). 10.1120/jacmp.v10i1.2915
    • (2009) J. Appl. Clin. Med. Phys. , vol.10 , pp. 120-128
    • Yan, G.1    Liu, C.2    Simon, T.3    Peng, L.-C.4    Fox, C.5    Li, J.6
  • 19
    • 79551672470 scopus 로고    scopus 로고
    • Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors
    • B. E. Nelms, H. Zhen, and W. A. Tomé, "Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors," Med. Phys. 38, 1037-1044 (2011). 10.1118/1.3544657
    • (2011) Med. Phys. , vol.38 , pp. 1037-1044
    • Nelms, B.E.1    Zhen, H.2    Tomé, W.A.3
  • 20
    • 80053595703 scopus 로고    scopus 로고
    • Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA
    • H. Zhen, B. E. Nelms, and W. A. Tomé, "Moving from gamma passing rates to patient DVH-based QA metrics in pretreatment dose QA," Med. Phys. 38, 5477-5489 (2011). 10.1118/1.3633904
    • (2011) Med. Phys. , vol.38 , pp. 5477-5489
    • Zhen, H.1    Nelms, B.E.2    Tomé, W.A.3
  • 21
    • 84870943985 scopus 로고    scopus 로고
    • Pretreatment patient-specific IMRT quality assurance: A correlation study between gamma index and patient clinical dose volume histogram
    • M. Stasi, S. Bresciani, A. Miranti, A. Maggio, V. Sapino, and P. Gabriele, "Pretreatment patient-specific IMRT quality assurance: A correlation study between gamma index and patient clinical dose volume histogram," Med. Phys. 39, 7626-7634 (2012). 10.1118/1.4767763
    • (2012) Med. Phys. , vol.39 , pp. 7626-7634
    • Stasi, M.1    Bresciani, S.2    Miranti, A.3    Maggio, A.4    Sapino, V.5    Gabriele, P.6
  • 22
    • 84889647084 scopus 로고    scopus 로고
    • Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levels
    • 15pp.
    • B. E. Nelms, M. F. Chan, G. Jarry, M. Lemire, J. Lowden, C. Hampton, and V. Feygelman, "Evaluating IMRT and VMAT dose accuracy: Practical examples of failure to detect systematic errors when applying a commonly used metric and action levels," Med. Phys. 40, 111722 (15pp.) (2013). 10.1118/1.4826166
    • (2013) Med. Phys. , vol.40 , pp. 111722
    • Nelms, B.E.1    Chan, M.F.2    Jarry, G.3    Lemire, M.4    Lowden, J.5    Hampton, C.6    Feygelman, V.7
  • 23
    • 84883444329 scopus 로고    scopus 로고
    • Real-world examples of sensitivity failures of the 3%/3 mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning
    • B. Nelms, G. Jarry, M. Chan, C. Hampton, Y. Watanabe, and V. Feygelman, "Real-world examples of sensitivity failures of the 3%/3 mm pass rate metric and published action levels when used in IMRT/VMAT system commissioning," J. Phys.: Conf. Ser. 444, 012086 (2013). 10.1088/1742-6596/444/1/012086
    • (2013) J. Phys.: Conf. Ser. , vol.444 , pp. 012086
    • Nelms, B.1    Jarry, G.2    Chan, M.3    Hampton, C.4    Watanabe, Y.5    Feygelman, V.6
  • 24
    • 84892773055 scopus 로고    scopus 로고
    • Using a novel dose QA tool to quantify the impact of systematic errors otherwise undetected by conventional QA methods: Clinical head and neck case studies
    • M. F. Chan, J. Li, K. Schupak, and C. Burman, "Using a novel dose QA tool to quantify the impact of systematic errors otherwise undetected by conventional QA methods: Clinical head and neck case studies," Technol. Cancer Res. Treat. 13, 57-67 (2014). 10.7785/tcrt.2012.500353
    • (2014) Technol. Cancer Res. Treat. , vol.13 , pp. 57-67
    • Chan, M.F.1    Li, J.2    Schupak, K.3    Burman, C.4
  • 25
    • 84947474571 scopus 로고    scopus 로고
    • Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations
    • C. Bojechko and E. C. Ford, "Quantifying the performance of in vivo portal dosimetry in detecting four types of treatment parameter variations," Med. Phys. 42, 6912-6918 (2015). 10.1118/1.4935093
    • (2015) Med. Phys. , vol.42 , pp. 6912-6918
    • Bojechko, C.1    Ford, E.C.2
  • 27
    • 77953487894 scopus 로고    scopus 로고
    • On the insensitivity of single field planar dosimetry to IMRT inaccuracies
    • J. J. Kruse, "On the insensitivity of single field planar dosimetry to IMRT inaccuracies," Med. Phys. 37, 2516-2524 (2010). 10.1118/1.3425781
    • (2010) Med. Phys. , vol.37 , pp. 2516-2524
    • Kruse, J.J.1
  • 28
    • 84922463784 scopus 로고    scopus 로고
    • Toward optimizing patient-specific IMRT QA techniques in the accurate detection of dosimetrically acceptable and unacceptable patient plans
    • (14 p).
    • E. M. McKenzie, P. A. Balter, F. C. Stingo, J. Jones, D. S. Followill, and S. F. Kry, "Toward optimizing patient-specific IMRT QA techniques in the accurate detection of dosimetrically acceptable and unacceptable patient plans," Med. Phys. 41, 121702 (14 pp.) (2014). 10.1118/1.4899177
    • (2014) Med. Phys. , vol.41 , pp. 121702
    • McKenzie, E.M.1    Balter, P.A.2    Stingo, F.C.3    Jones, J.4    Followill, D.S.5    Kry, S.F.6
  • 29
    • 34548231706 scopus 로고    scopus 로고
    • A survey on IMRT QA analysis
    • B. E. Nelms and J. A. Simon, "A survey on IMRT QA analysis," J. Appl. Clin. Med. Phys. 8, 76-90 (2007). 10.1120/jacmp.v8i3.2448
    • (2007) J. Appl. Clin. Med. Phys. , vol.8 , pp. 76-90
    • Nelms, B.E.1    Simon, J.A.2
  • 30
    • 84890121955 scopus 로고    scopus 로고
    • A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems
    • M. Hussein, P. Rowshanfarzad, M. A. Ebert, A. Nisbet, and C. H. Clark, "A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems," Radiother. Oncol. 109, 370-376 (2013). 10.1016/j.radonc.2013.08.048
    • (2013) Radiother. Oncol. , vol.109 , pp. 370-376
    • Hussein, M.1    Rowshanfarzad, P.2    Ebert, M.A.3    Nisbet, A.4    Clark, C.H.5
  • 31
    • 84877760238 scopus 로고    scopus 로고
    • Statistical process control analysis for patient-specific IMRT and VMAT QA
    • T. Sanghangthum, S. Suriyapee, S. Srisatit, and T. Pawlicki, "Statistical process control analysis for patient-specific IMRT and VMAT QA," J. Radiat. Res. 54, 546-552 (2013). 10.1093/jrr/rrs112
    • (2013) J. Radiat. Res. , vol.54 , pp. 546-552
    • Sanghangthum, T.1    Suriyapee, S.2    Srisatit, S.3    Pawlicki, T.4
  • 33
    • 84864240475 scopus 로고    scopus 로고
    • Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments
    • G. M. Mancuso, J. D. Fontenot, J. P. Gibbons, and B. C. Parker, "Comparison of action levels for patient-specific quality assurance of intensity modulated radiation therapy and volumetric modulated arc therapy treatments," Med. Phys. 39, 4378-4385 (2012). 10.1118/1.4729738
    • (2012) Med. Phys. , vol.39 , pp. 4378-4385
    • Mancuso, G.M.1    Fontenot, J.D.2    Gibbons, J.P.3    Parker, B.C.4
  • 35
    • 84873148965 scopus 로고    scopus 로고
    • Delivery quality assurance with ArcCHECK
    • C. Neilson, M. Klein, R. Barnett, and S. Yartsev, "Delivery quality assurance with ArcCHECK," Med. Dosim. 38, 77-80 (2013). 10.1016/j.meddos.2012.07.004
    • (2013) Med. Dosim. , vol.38 , pp. 77-80
    • Neilson, C.1    Klein, M.2    Barnett, R.3    Yartsev, S.4
  • 36
    • 84907022995 scopus 로고    scopus 로고
    • A six-year review of more than 13,000 patient-specific IMRT QA results from 13 different treatment sites
    • K. B. Pulliam, D. Followill, L. Dong, M. Gillin, K. Prado, and S. F. Kry, "A six-year review of more than 13,000 patient-specific IMRT QA results from 13 different treatment sites," J. Appl. Clin. Med. Phys. 15, 196-206 (2014). 10.1120/jacmp.v15i5.4935
    • (2014) J. Appl. Clin. Med. Phys. , vol.15 , pp. 196-206
    • Pulliam, K.B.1    Followill, D.2    Dong, L.3    Gillin, M.4    Prado, K.5    Kry, S.F.6
  • 37
    • 84890057550 scopus 로고    scopus 로고
    • Tomotherapy treatment plan quality assurance: The impact of applied criteria on passing rate in gamma index method
    • 6 pp.
    • S. Bresciani, A. Di Dia, A. Maggio, C. Cutaia, A. Miranti, E. Infusino, and M. Stasi, "Tomotherapy treatment plan quality assurance: The impact of applied criteria on passing rate in gamma index method," Med. Phys. 40, 121711 (6 pp.) (2013). 10.1118/1.4829515
    • (2013) Med. Phys. , vol.40 , pp. 121711
    • Bresciani, S.1    Di Dia, A.2    Maggio, A.3    Cutaia, C.4    Miranti, A.5    Infusino, E.6    Stasi, M.7
  • 38
    • 84930936465 scopus 로고    scopus 로고
    • Patient specific IMRT quality assurance with film, ionization chamber and detector arrays: Our institutional experience
    • W. Cruz, G. Narayanasamy, M. Regan, P. Mavroidis, N. Papanikolaou, C. S. Ha, and S. Stathakis, "Patient specific IMRT quality assurance with film, ionization chamber and detector arrays: Our institutional experience," Radiat. Phys. Chem. 115, 12-16 (2015). 10.1016/j.radphyschem.2015.06.002
    • (2015) Radiat. Phys. Chem. , vol.115 , pp. 12-16
    • Cruz, W.1    Narayanasamy, G.2    Regan, M.3    Mavroidis, P.4    Papanikolaou, N.5    Ha, C.S.6    Stathakis, S.7
  • 39
    • 79960552723 scopus 로고    scopus 로고
    • Evaluation of a new VMAT QA device, or the 'X' and 'O' array geometries
    • V. Feygelman, G. Zhang, C. Stevens, and B. E. Nelms, "Evaluation of a new VMAT QA device, or the 'X' and 'O' array geometries," J. Appl. Clin. Med. Phys. 12, 146-168 (2011). 10.1120/jacmp.v12i2.3346
    • (2011) J. Appl. Clin. Med. Phys. , vol.12 , pp. 146-168
    • Feygelman, V.1    Zhang, G.2    Stevens, C.3    Nelms, B.E.4
  • 40
    • 80855128877 scopus 로고    scopus 로고
    • Statistical variability and confidence intervals for planar dose QA pass rates
    • D. W. Bailey, B. E. Nelms, K. Attwood, L. Kumaraswamy, and M. B. Podgorsak, "Statistical variability and confidence intervals for planar dose QA pass rates," Med. Phys. 38, 6053-6064 (2011). 10.1118/1.3651695
    • (2011) Med. Phys. , vol.38 , pp. 6053-6064
    • Bailey, D.W.1    Nelms, B.E.2    Attwood, K.3    Kumaraswamy, L.4    Podgorsak, M.B.5
  • 41
    • 65449118359 scopus 로고    scopus 로고
    • Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    • D. Létourneau, J. Publicover, J. Kozelka, D. J. Moseley, and D. A. Jaffray, "Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy," Med. Phys. 36, 1813-1821 (2009). 10.1118/1.3117563
    • (2009) Med. Phys. , vol.36 , pp. 1813-1821
    • Létourneau, D.1    Publicover, J.2    Kozelka, J.3    Moseley, D.J.4    Jaffray, D.A.5
  • 42
    • 73649086212 scopus 로고    scopus 로고
    • Calibration of a novel four-dimensional diode array
    • G. Yan, B. Lu, J. Kozelka, C. Liu, and J. G. Li, "Calibration of a novel four-dimensional diode array," Med. Phys. 37, 108-115 (2010). 10.1118/1.3266769
    • (2010) Med. Phys. , vol.37 , pp. 108-115
    • Yan, G.1    Lu, B.2    Kozelka, J.3    Liu, C.4    Li, J.G.5
  • 43
    • 80052401496 scopus 로고    scopus 로고
    • Optimizing the accuracy of a helical diode array dosimeter: A comprehensive calibration methodology coupled with a novel virtual inclinometer
    • J. Kozelka, J. Robinson, B. Nelms, G. Zhang, D. Savitskij, and V. Feygelman, "Optimizing the accuracy of a helical diode array dosimeter: A comprehensive calibration methodology coupled with a novel virtual inclinometer," Med. Phys. 38, 5021-5032 (2011). 10.1118/1.3622823
    • (2011) Med. Phys. , vol.38 , pp. 5021-5032
    • Kozelka, J.1    Robinson, J.2    Nelms, B.3    Zhang, G.4    Savitskij, D.5    Feygelman, V.6
  • 44
    • 78650062818 scopus 로고    scopus 로고
    • Accuracy of out-of-field dose calculations by a commercial treatment planning system
    • R. M. Howell, S. B. Scarboro, S. Kry, and D. Z. Yaldo, "Accuracy of out-of-field dose calculations by a commercial treatment planning system," Phys. Med. Biol. 55, 6999-7008 (2010). 10.1088/0031-9155/55/23/S03
    • (2010) Phys. Med. Biol. , vol.55 , pp. 6999-7008
    • Howell, R.M.1    Scarboro, S.B.2    Kry, S.3    Yaldo, D.Z.4
  • 45
    • 84963746890 scopus 로고    scopus 로고
    • E-MPHYA6-43-028604 for additional error curve data
    • See supplementary material at http://dx.doi.org/10.1118/1.4943953 E-MPHYA6-43-028604 for additional error curve data.
  • 48
    • 80054081009 scopus 로고    scopus 로고
    • An analytical approach to acceptance criteria for quality assurance of intensity modulated radiotherapy
    • A. Carver, M. Gilmore, S. Riley, J. Uzan, and P. Mayles, "An analytical approach to acceptance criteria for quality assurance of intensity modulated radiotherapy," Radiother. Oncol. 100, 453-455 (2011). 10.1016/j.radonc.2011.08.045
    • (2011) Radiother. Oncol. , vol.100 , pp. 453-455
    • Carver, A.1    Gilmore, M.2    Riley, S.3    Uzan, J.4    Mayles, P.5
  • 49
    • 84904785092 scopus 로고    scopus 로고
    • The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT
    • J. I. Kim, S. Y. Park, H. J. Kim, J. H. Kim, S. J. Ye, and J. M. Park, "The sensitivity of gamma-index method to the positioning errors of high-definition MLC in patient-specific VMAT QA for SBRT," Radiat. Oncol. 9, 167-178 (2014). 10.1186/1748-717X-9-167
    • (2014) Radiat. Oncol. , vol.9 , pp. 167-178
    • Kim, J.I.1    Park, S.Y.2    Kim, H.J.3    Kim, J.H.4    Ye, S.J.5    Park, J.M.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.