Volumn 39, Issue 6, 2012, Pages 3051-3059

On using 3D γ-analysis for IMRT and VMAT pretreatment plan QA

Author keywords

3D Gamma; EPID; IMRT; QA; VMAT

Indexed keywords

3D-GAMMA; BACK PROJECTION; EPID; GAMMA ANALYSIS; IMRT; PRE-TREATMENT; STATISTICAL CORRELATION; VMAT;

EID: 84863527560     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.4711755     Document Type: Article
Times cited : (32)

References (37)
  • 3
    • 0031861573 scopus 로고    scopus 로고
    • A technique for the quantitative evaluation of dose distributions
    • 10.1118/1.598248
    • D. A. Low, W. B. Hams, S. Mutic, and J. A. Purdy, A technique for the quantitative evaluation of dose distributions., Med. Phys. 25, 656-661 (1998). 10.1118/1.598248
    • (1998) Med. Phys. , vol.25 , pp. 656-661
    • Low, D.A.1    Hams, W.B.2    Mutic, S.3    Purdy, J.A.4
  • 4
    • 0141520680 scopus 로고    scopus 로고
    • Evaluation of the Gamma dose distribution comparison method
    • 10.1118/1.1598711
    • D. A. Low and J. F. Dempsey, Evaluation of the Gamma dose distribution comparison method., Med. Phys. 30, 2455-2464 (2003). 10.1118/1.1598711
    • (2003) Med. Phys. , vol.30 , pp. 2455-2464
    • Low, D.A.1    Dempsey, J.F.2
  • 5
    • 14544300121 scopus 로고    scopus 로고
    • Interpretation and evaluation of the Gamma index and the Gamma index angle for the verification of IMRT hybrid plans
    • 10.1088/0031-9155/50/3/001
    • M. Stock, B. Kroupa, and D. Georg, Interpretation and evaluation of the Gamma index and the Gamma index angle for the verification of IMRT hybrid plans., Phys. Med. Biol. 50 (3), 399-411 (2005). 10.1088/0031-9155/50/3/001
    • (2005) Phys. Med. Biol. , vol.50 , Issue.3 , pp. 399-411
    • Stock, M.1    Kroupa, B.2    Georg, D.3
  • 6
    • 13244254064 scopus 로고    scopus 로고
    • Quantitative analysis of patient-specific dosimetric IMRT verification
    • 10.1088/0031-9155/50/1/009
    • G. J. Budgell, B. A. Perrin, J. H. L. Mott, J. Fairfoul, and R. I. Mackay, Quantitative analysis of patient-specific dosimetric IMRT verification., Phys. Med. Biol. 50, 103-119 (2005). 10.1088/0031-9155/50/1/009
    • (2005) Phys. Med. Biol. , vol.50 , pp. 103-119
    • Budgell, G.J.1    Perrin, B.A.2    Mott, J.H.L.3    Fairfoul, J.4    MacKay, R.I.5
  • 7
    • 33750729320 scopus 로고    scopus 로고
    • Dosimetric pre-treatment verification of IMRT using an EPID; Clinical experience
    • 10.1016/j.radonc.2006.09.008
    • M. Van Zijtveld, M. L. P. Dirkx, H. C. de Boer, and B. J. M. Heijmen, Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience., Radiother. Oncol. 81, 168-175 (2006). 10.1016/j.radonc.2006.09.008
    • (2006) Radiother. Oncol. , vol.81 , pp. 168-175
    • Van Zijtveld, M.1    Dirkx, M.L.P.2    De Boer, H.C.3    Heijmen, B.J.M.4
  • 8
    • 34249947757 scopus 로고    scopus 로고
    • A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy
    • S. Both, L. M. Alecu, A. R. Stan, M. Alecu, A. Ciura, J. M. Hansen, and R. Alecu, A study to establish reasonable action limits for patient-specific quality assurance in intensity-modulated radiation therapy., J. Appl. Clin. Med. Phys. 8 (2), 1-8 (2007).
    • (2007) J. Appl. Clin. Med. Phys. , vol.8 , Issue.2 , pp. 1-8
    • Both, S.1    Alecu, L.M.2    Stan, A.R.3    Alecu, M.4    Ciura, A.5    Hansen, J.M.6    Alecu, R.7
  • 9
    • 65449118359 scopus 로고    scopus 로고
    • Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    • 10.1118/1.3117563
    • D. Letourneau, J. Publicover, J. Kozelka, D. J. Mosley, and D. A. Jaffray, Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy., Med. Phys. 36 (5), 1813-1821 (2009). 10.1118/1.3117563
    • (2009) Med. Phys. , vol.36 , Issue.5 , pp. 1813-1821
    • Letourneau, D.1    Publicover, J.2    Kozelka, J.3    Mosley, D.J.4    Jaffray, D.A.5
  • 10
    • 79951785189 scopus 로고    scopus 로고
    • Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array
    • 10.1186/1748-717X-6-21
    • D. Wagner and H. Vorwerk, Two years experience with quality assurance protocol for patient related Rapid Arc treatment plan verification using a two dimensional ionization chamber array., Radiat. Oncol. 6, 1-8 (2011). 10.1186/1748-717X-6-21
    • (2011) Radiat. Oncol. , vol.6 , pp. 1-8
    • Wagner, D.1    Vorwerk, H.2
  • 13
    • 0035044866 scopus 로고    scopus 로고
    • A feasible method for clinical delivery of verification and dose reconstruction in Tomotherapy
    • 10.1118/1.1352579
    • J. M. Kapatoes, G. H. Olivear, K. J. Ruchala, J. B. Smilowitz, P. J. Reckwerdt, and T. R. Mackie, A feasible method for clinical delivery of verification and dose reconstruction in Tomotherapy., Med. Phys. 28, 528-542 (2001). 10.1118/1.1352579
    • (2001) Med. Phys. , vol.28 , pp. 528-542
    • Kapatoes, J.M.1    Olivear, G.H.2    Ruchala, K.J.3    Smilowitz, J.B.4    Reckwerdt, P.J.5    MacKie, T.R.6
  • 14
    • 84892779521 scopus 로고    scopus 로고
    • White Paper published by Sun Nuclear Cor, Melbourne, FL, USA, downloadable from www.sunnuclear.com
    • On the accuracy of the planned dose perturbation algorithm., White Paper published by Sun Nuclear Corp., Melbourne, FL, USA, downloadable from www.sunnuclear.com (2010).
    • (2010) On the Accuracy of the Planned Dose Perturbation Algorithm
  • 15
    • 0029790894 scopus 로고    scopus 로고
    • Modeling dose distributions from portal dose images using the convolutionsuperposition method
    • 10.1118/1.597872
    • T. R. McNutt, T. R. Mackie, P. J. Reckwerdt, and B. R. Paliwal, Modeling dose distributions from portal dose images using the convolutionsuperposition method., Med. Phys. 23, 1381-1392 (1996). 10.1118/1.597872
    • (1996) Med. Phys. , vol.23 , pp. 1381-1392
    • McNutt, T.R.1    MacKie, T.R.2    Reckwerdt, P.J.3    Paliwal, B.R.4
  • 16
    • 85024993983 scopus 로고    scopus 로고
    • A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction
    • 10.1120/jacm2026.25359
    • W. D. Renner, K. Norton, and T. Holmes, A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction., J. Appl. Clin. Med. Phys. 6, 22-39 (2005). 10.1120/jacmp.2026. 25359
    • (2005) J. Appl. Clin. Med. Phys. , vol.6 , pp. 22-39
    • Renner, W.D.1    Norton, K.2    Holmes, T.3
  • 17
    • 14644390214 scopus 로고    scopus 로고
    • Three-dimensional IMRT verification with a flat-panel EPID
    • 10.1118/1.1843471
    • S. Steciw, B. Warkentin, S. Rathee, and B. G. Fallone, Three-dimensional IMRT verification with a flat-panel EPID., Med. Phys. 32 (2), 600-612 (2005). 10.1118/1.1843471
    • (2005) Med. Phys. , vol.32 , Issue.2 , pp. 600-612
    • Steciw, S.1    Warkentin, B.2    Rathee, S.3    Fallone, B.G.4
  • 18
    • 31844436637 scopus 로고    scopus 로고
    • Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
    • 10.1118/1.2147744
    • M. Wendling, R. J. W. Louwe, L. N. McDermott, J. Sonke, M. Van Herk, and B. J. Mijnheer, Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method., Med. Phys. 33 (2), 259-273 (2006). 10.1118/1.2147744
    • (2006) Med. Phys. , vol.33 , Issue.2 , pp. 259-273
    • Wendling, M.1    Louwe, R.J.W.2    McDermott, L.N.3    Sonke, J.4    Van Herk, M.5    Mijnheer, B.J.6
  • 19
    • 34547781371 scopus 로고    scopus 로고
    • 3D dose reconstruction to ensure correct external beam treatment of patients
    • 10.1016/j.meddos.2007.02.005
    • W. D. Renner, 3D dose reconstruction to ensure correct external beam treatment of patients., Med. Dosim. 32, 157-165 (2007). 10.1016/j.meddos.2007. 02.005
    • (2007) Med. Dosim. , vol.32 , pp. 157-165
    • Renner, W.D.1
  • 20
    • 67649583518 scopus 로고    scopus 로고
    • A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments
    • 10.1118/1.3148482
    • M. Wendling, L. N. McDermott, A. Mans, J. Sonke, M. Van Herk, and B. J. Mijnheer, A simple backprojection algorithm for 3D in vivo EPID dosimetry of IMRT treatments., Med. Phys. 36 (7), 3310-3321 (2009). 10.1118/1.3148482
    • (2009) Med. Phys. , vol.36 , Issue.7 , pp. 3310-3321
    • Wendling, M.1    McDermott, L.N.2    Mans, A.3    Sonke, J.4    Van Herk, M.5    Mijnheer, B.J.6
  • 21
    • 77953815615 scopus 로고    scopus 로고
    • Dose reconstruction for volumetric modulated arc therapy (VMAT) using con-beam CT and dynamic log files
    • 10.1088/0031-9155/55/13/002
    • J. Qian, L. Lee, W. Liu, K. Chu, E. Mok, G. Luxton, Q. Le, and L. Xing, Dose reconstruction for volumetric modulated arc therapy (VMAT) using con-beam CT and dynamic log files., Phys. Med. Biol. 55, 3597-3610 (2010). 10.1088/0031-9155/55/13/002
    • (2010) Phys. Med. Biol. , vol.55 , pp. 3597-3610
    • Qian, J.1    Lee, L.2    Liu, W.3    Chu, K.4    Mok, E.5    Luxton, G.6    Le, Q.7    Xing, L.8
  • 23
    • 84863515711 scopus 로고    scopus 로고
    • Accurate modeling of Calypso kVue couch for PINNACLE SmartArc based ELEKTA VMAT prostate treatments
    • C. Wu, C. Tunnicliff, and D. R. Asche, Accurate modeling of Calypso kVue couch for PINNACLE SmartArc based ELEKTA VMAT prostate treatments., Med. Phys. 37 (6), 3221 (2010); AAPM Annual Meeting, Philadelphia, PA, 072010. 10.1118/1.3468547
    • (2010) Med. Phys. , vol.37 , Issue.6 , pp. 3221
    • Wu, C.1    Tunnicliff, C.2    Asche, D.R.3
  • 24
    • 0029064833 scopus 로고
    • Limitations of a pencil beam approach to photon dose calculations in lung tissue
    • 10.1088/0031-9155/40/9/002
    • T. Knoos, A. Ahnesjo, P. Nilsson, and L. Weber, Limitations of a pencil beam approach to photon dose calculations in lung tissue., Phys. Med. Biol. 40, 1411-1420 (1995). 10.1088/0031-9155/40/9/002
    • (1995) Phys. Med. Biol. , vol.40 , pp. 1411-1420
    • Knoos, T.1    Ahnesjo, A.2    Nilsson, P.3    Weber, L.4
  • 26
    • 0033970872 scopus 로고    scopus 로고
    • Iterative approaches to dose optimization in tomotherapy
    • 10.1088/0031-9155/45/1/306
    • D. M. Shepard, G. H. Olivera, P. J. Recjwerdt, and T. R. Mackie, Iterative approaches to dose optimization in tomotherapy., Phys. Med. Biol. 45, 69-90 (2000). 10.1088/0031-9155/45/1/306
    • (2000) Phys. Med. Biol. , vol.45 , pp. 69-90
    • Shepard, D.M.1    Olivera, G.H.2    Recjwerdt, P.J.3    MacKie, T.R.4
  • 27
    • 34548231706 scopus 로고    scopus 로고
    • A survey on planar IMRT QA analysis
    • 10.1120/jacmv8i3.2448
    • B. E. Nelms and J. A. Simon, A survey on planar IMRT QA analysis., J. Appl. Clin. Med. Phys. 8 (3), 76-90 (2007). 10.1120/jacmp.v8i3.2448
    • (2007) J. Appl. Clin. Med. Phys. , vol.8 , Issue.3 , pp. 76-90
    • Nelms, B.E.1    Simon, J.A.2
  • 28
    • 82755190904 scopus 로고    scopus 로고
    • Towards a better understanding of the γ index: Investigation of parameters with a surface-based distance method
    • 10.1118/1.3659707
    • H. Li, L. Dong, L. Zhang, J. Yang, M. T. Gillin, and X. R. Zhu, Towards a better understanding of the γ index: Investigation of parameters with a surface-based distance method., Med. Phys. 38 (12), 6730-6741 (2011). 10.1118/1.3659707
    • (2011) Med. Phys. , vol.38 , Issue.12 , pp. 6730-6741
    • Li, H.1    Dong, L.2    Zhang, L.3    Yang, J.4    Gillin, M.T.5    Zhu, X.R.6
  • 29
    • 77953518833 scopus 로고    scopus 로고
    • MapCheck used for rotational IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc
    • 10.1118/1.3431994
    • P. A. Jursinic, R. Sharma, and J. Reuter, MapCheck used for rotational IMRT measurements: Step-and-shoot, Tomotherapy, RapidArc., Med. Phys. 37 (6), 2837-2846 (2010). 10.1118/1.3431994
    • (2010) Med. Phys. , vol.37 , Issue.6 , pp. 2837-2846
    • Jursinic, P.A.1    Sharma, R.2    Reuter, J.3
  • 30
    • 85168628815 scopus 로고    scopus 로고
    • Quantitative evaluation and correction of the gantry angular dependence of MapCheck2 for patient QA measurement
    • rd Annual Meeting, Vancouver, BC, Canada, 072011. 10.1118/1.3612121
    • (2011) Med. Phys. , vol.38 , Issue.6 , pp. 3525
    • Zhou, H.1    Wu, C.2
  • 31
    • 23944453854 scopus 로고    scopus 로고
    • Consequences of the spectral response of an a-Si EPID and implications for dosimetric calibration
    • 10.1118/1.1984335
    • C. Kirkby and R. Sloboda, Consequences of the spectral response of an a-Si EPID and implications for dosimetric calibration., Med. Phys. 32 (8), 2649-2658 (2005). 10.1118/1.1984335
    • (2005) Med. Phys. , vol.32 , Issue.8 , pp. 2649-2658
    • Kirkby, C.1    Sloboda, R.2
  • 32
    • 79551672470 scopus 로고    scopus 로고
    • Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors
    • 10.1118/1.3544657
    • B. E. Nelms, H. Zhen, and W. Tome, Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors., Med. Phys. 38 (2), 1037-1044 (2011). 10.1118/1.3544657
    • (2011) Med. Phys. , vol.38 , Issue.2 , pp. 1037-1044
    • Nelms, B.E.1    Zhen, H.2    Tome, W.3
  • 33
    • 34748821714 scopus 로고    scopus 로고
    • Off-axis dose response characteristics of an amorphous silicon electric portal imaging device
    • 10.1118/1.2779944
    • P. B. Greer, Off-axis dose response characteristics of an amorphous silicon electric portal imaging device., Med. Phys. 34 (10), 3815-3824 (2007). 10.1118/1.2779944
    • (2007) Med. Phys. , vol.34 , Issue.10 , pp. 3815-3824
    • Greer, P.B.1
  • 34
    • 69549135345 scopus 로고    scopus 로고
    • An effective correction algorithm for off-axis portal dosimetry errors
    • 10.1118/1.3187785
    • D. W. Bailey, L. Kumaraswamy, and M. B. Podgorsak, An effective correction algorithm for off-axis portal dosimetry errors., Med. Phys. 36 (9), 4089-4094 (2009). 10.1118/1.3187785
    • (2009) Med. Phys. , vol.36 , Issue.9 , pp. 4089-4094
    • Bailey, D.W.1    Kumaraswamy, L.2    Podgorsak, M.B.3
  • 35
    • 2542441748 scopus 로고    scopus 로고
    • Investigation of the optimal backscatter for an aSi electronic portal imaging device
    • 10.1088/0031-9155/49/9/010
    • K. Lung, J. O. Kim, and J. V. Siebers, Investigation of the optimal backscatter for an aSi electronic portal imaging device., Phys. Med. Biol. 49, 1723-1738 (2004). 10.1088/0031-9155/49/9/010
    • (2004) Phys. Med. Biol. , vol.49 , pp. 1723-1738
    • Lung, K.1    Kim, J.O.2    Siebers, J.V.3
  • 36
    • 20044368303 scopus 로고    scopus 로고
    • Verification of the optimal backscatter for an aSi electronic portal imaging device
    • 10.1088/0031-9155/50/10/011
    • J. A. Moore and J. V. Siebers Verification of the optimal backscatter for an aSi electronic portal imaging device., Phys. Med. Biol. 50, 2341-2350 (2005). 10.1088/0031-9155/50/10/011
    • (2005) Phys. Med. Biol. , vol.50 , pp. 2341-2350
    • Moore, J.A.1    Siebers, J.V.2
  • 37
    • 80053595703 scopus 로고    scopus 로고
    • Moving from γ passing rates to patients DVH-based QA metrics in pretreatment dose QA
    • 10.1118/1.3633904
    • H. Zhen, B. E. Nelms, and W. A. Tome, Moving from γ passing rates to patients DVH-based QA metrics in pretreatment dose QA., Med. Phys. 38 (10), 5477-5489 (2011). 10.1118/1.3633904
    • (2011) Med. Phys. , vol.38 , Issue.10 , pp. 5477-5489
    • Zhen, H.1    Nelms, B.E.2    Tome, W.A.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.