Gamma index comparison of three VMAT QA systems and evaluation of their sensitivity to delivery errors

Author keywords

Delivery error; Gamma index; Quality assurance; RapidArc

Indexed keywords

ACCURACY; ARTICLE; COLLIMATOR; CONTROLLED STUDY; DOSIMETRY; GAMMA RADIATION; HUMAN; INTERMETHOD COMPARISON; MAJOR CLINICAL STUDY; RADIATION DETECTION; RADIATION DETECTOR; RADIATION DOSE DISTRIBUTION; REPRODUCIBILITY; SENSITIVITY ANALYSIS; TREATMENT PLANNING; VOLUMETRIC MODULATED ARC THERAPY; COMPARATIVE STUDY; DEVICES; EVALUATION STUDY; INTENSITY MODULATED RADIATION THERAPY; MEDICAL ERROR; NEOPLASMS; PREVENTION AND CONTROL; QUALITY CONTROL; RADIOMETRY; ROTATION; STANDARDS;

EID: 84948718537     PISSN: 11201797     EISSN: 1724191X     Source Type: Journal    
DOI: 10.1016/j.ejmp.2015.05.016     Document Type: Article
Times cited : (67)

References (25)
  • 1
    • 0029113269 scopus 로고
    • Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy
    • Yu Cedric X. Intensity-modulated arc therapy with dynamic multileaf collimation: an alternative to tomotherapy. Phys Med Biol 1995, 1435-1449.
    • (1995) Phys Med Biol , pp. 1435-1449
    • Yu Cedric, X.1
  • 3
    • 37549049393 scopus 로고    scopus 로고
    • Volumetric modulated arc therapy: IMRT in a single gantry arc
    • Otto K. Volumetric modulated arc therapy: IMRT in a single gantry arc. Med Phys 2008, 35(1):310-317.
    • (2008) Med Phys , vol.35 , Issue.1 , pp. 310-317
    • Otto, K.1
  • 4
    • 70350726816 scopus 로고    scopus 로고
    • IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119
    • Ezzell A., Burmeister J.W., Dogan N., LoSasso T.J., Mechalakos J.G., Mihailidis D., et al. IMRT commissioning: multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119. Med Phys 2009, 36(11):5359-5373.
    • (2009) Med Phys , vol.36 , Issue.11 , pp. 5359-5373
    • Ezzell, A.1    Burmeister, J.W.2    Dogan, N.3    LoSasso, T.J.4    Mechalakos, J.G.5    Mihailidis, D.6
  • 5
    • 65449118359 scopus 로고    scopus 로고
    • Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    • Létourneau D., Publicover J., Kozelka J., Moseley D.J., Jaffray D.A. Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy. Med Phys 2009, 36(5):1813-1821.
    • (2009) Med Phys , vol.36 , Issue.5 , pp. 1813-1821
    • Létourneau, D.1    Publicover, J.2    Kozelka, J.3    Moseley, D.J.4    Jaffray, D.A.5
  • 6
    • 84965093881 scopus 로고    scopus 로고
    • Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA
    • Chaswal V., Weldon M., Gupta N., Chakravarti A., Rong Y. Commissioning and comprehensive evaluation of the ArcCHECK cylindrical diode array for VMAT pretreatment delivery QA. J Appl Clin Med Phys 2015, 15(4):212-225.
    • (2015) J Appl Clin Med Phys , vol.15 , Issue.4 , pp. 212-225
    • Chaswal, V.1    Weldon, M.2    Gupta, N.3    Chakravarti, A.4    Rong, Y.5
  • 8
    • 84890121955 scopus 로고    scopus 로고
    • A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems
    • Hussein M., Rowshanfarzad P., Ebert M.A., Nisbet A., Clark C.H. A comparison of the gamma index analysis in various commercial IMRT/VMAT QA systems. Radiother Oncol 2013, 109(3):370-376.
    • (2013) Radiother Oncol , vol.109 , Issue.3 , pp. 370-376
    • Hussein, M.1    Rowshanfarzad, P.2    Ebert, M.A.3    Nisbet, A.4    Clark, C.H.5
  • 9
    • 79551661964 scopus 로고    scopus 로고
    • Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems
    • 612
    • Masi L., Casamassima F., Doro R., Francescon P. Quality assurance of volumetric modulated arc therapy: evaluation and comparison of different dosimetric systems. Med Phys 2011, 38(2). 612.
    • (2011) Med Phys , vol.38 , Issue.2
    • Masi, L.1    Casamassima, F.2    Doro, R.3    Francescon, P.4
  • 10
    • 84888413988 scopus 로고    scopus 로고
    • A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification
    • Hussein M., Adams E.J., Jordan T.J., Clark C.H., Nisbet A. A critical evaluation of the PTW 2D-ARRAY seven29 and OCTAVIUS II phantom for IMRT and VMAT verification. J Appl Clin Med Phys 2013, 14(6):274-292.
    • (2013) J Appl Clin Med Phys , vol.14 , Issue.6 , pp. 274-292
    • Hussein, M.1    Adams, E.J.2    Jordan, T.J.3    Clark, C.H.4    Nisbet, A.5
  • 11
    • 84874780814 scopus 로고    scopus 로고
    • Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors
    • 031716
    • Fredh A., Scherman J.B., Fog L.S., Munck af Rosenschöld P. Patient QA systems for rotational radiation therapy: a comparative experimental study with intentional errors. Med Phys 2013, 40(3). 031716.
    • (2013) Med Phys , vol.40 , Issue.3
    • Fredh, A.1    Scherman, J.B.2    Fog, L.S.3    Munck af Rosenschöld, P.4
  • 12
    • 84889680232 scopus 로고    scopus 로고
    • Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics
    • Coleman L., Skourou C. Sensitivity of volumetric modulated arc therapy patient specific QA results to multileaf collimator errors and correlation to dose volume histogram based metrics. Med Phys 2013, 40(11):111715.
    • (2013) Med Phys , vol.40 , Issue.11 , pp. 111715
    • Coleman, L.1    Skourou, C.2
  • 13
    • 84874766976 scopus 로고    scopus 로고
    • On the sensitivity of common gamma-index evaluation methods to MLC misalignments in rapidarc quality assurance
    • 031702
    • Heilemann G., Poppe B., Laub W. On the sensitivity of common gamma-index evaluation methods to MLC misalignments in rapidarc quality assurance. Med Phys 2013, 40(3). 031702.
    • (2013) Med Phys , vol.40 , Issue.3
    • Heilemann, G.1    Poppe, B.2    Laub, W.3
  • 14
    • 0031861573 scopus 로고    scopus 로고
    • A technique for the quantitative evaluation of dose distributions
    • Low D.A., Harms W.B., Mutic S., Purdy J.A. A technique for the quantitative evaluation of dose distributions. Med Phys 1998, 25(5):656-661.
    • (1998) Med Phys , vol.25 , Issue.5 , pp. 656-661
    • Low, D.A.1    Harms, W.B.2    Mutic, S.3    Purdy, J.A.4
  • 16
    • 33746606839 scopus 로고    scopus 로고
    • GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications
    • Nicolini G., Fogliata A., Vanetti E., Clivio A., Cozzi L. GLAaS: an absolute dose calibration algorithm for an amorphous silicon portal imager. Applications to IMRT verifications. Med Phys 2006, 33(8):2839-2851.
    • (2006) Med Phys , vol.33 , Issue.8 , pp. 2839-2851
    • Nicolini, G.1    Fogliata, A.2    Vanetti, E.3    Clivio, A.4    Cozzi, L.5
  • 18
    • 53149135069 scopus 로고    scopus 로고
    • The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy
    • Nicolini G., Vanetti E., Clivio A., Fogliata A., Korreman S., Bocanek J., et al. The GLAaS algorithm for portal dosimetry and quality assurance of RapidArc, an intensity modulated rotational therapy. Radiat Oncol 2008, 3:24.
    • (2008) Radiat Oncol , vol.3 , pp. 24
    • Nicolini, G.1    Vanetti, E.2    Clivio, A.3    Fogliata, A.4    Korreman, S.5    Bocanek, J.6
  • 19
    • 84922842255 scopus 로고    scopus 로고
    • The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery II
    • Templeton A.K., Chu J.C.H., Turian J.V. The sensitivity of ArcCHECK-based gamma analysis to manufactured errors in helical tomotherapy radiation delivery II. J Appl Clin Med Phys 2015, 16(1):2-7.
    • (2015) J Appl Clin Med Phys , vol.16 , Issue.1 , pp. 2-7
    • Templeton, A.K.1    Chu, J.C.H.2    Turian, J.V.3
  • 20
    • 84884153727 scopus 로고    scopus 로고
    • Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT)
    • Li G., Bai S., Chen N., Henderson L., Wu K., Xiao J., et al. Evaluation of the sensitivity of two 3D diode array dosimetry systems to setup error for quality assurance (QA) of volumetric-modulated arc therapy (VMAT). J Appl Clin Med Phys 2013, 14(5):13-24.
    • (2013) J Appl Clin Med Phys , vol.14 , Issue.5 , pp. 13-24
    • Li, G.1    Bai, S.2    Chen, N.3    Henderson, L.4    Wu, K.5    Xiao, J.6
  • 21
    • 70049110836 scopus 로고    scopus 로고
    • On the sensitivity of patient-specific IMRT QA to MLC positioning errors
    • Yan G., Liu C., Simon T.A., Peng L.C., Fox C., Li J.G. On the sensitivity of patient-specific IMRT QA to MLC positioning errors. J Appl Clin Med Phys 2009, 10(1):120-128.
    • (2009) J Appl Clin Med Phys , vol.10 , Issue.1 , pp. 120-128
    • Yan, G.1    Liu, C.2    Simon, T.A.3    Peng, L.C.4    Fox, C.5    Li, J.G.6
  • 22
    • 79951785189 scopus 로고    scopus 로고
    • Two years experience with quality assurance protocol for patient related rapid arc treatment plan verification using a two dimensional ionization chamber array
    • Wagner D., Vorwerk H. Two years experience with quality assurance protocol for patient related rapid arc treatment plan verification using a two dimensional ionization chamber array. Radiat Oncol 2011, 6:1-8.
    • (2011) Radiat Oncol , vol.6 , pp. 1-8
    • Wagner, D.1    Vorwerk, H.2
  • 23
    • 79551672470 scopus 로고    scopus 로고
    • Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors
    • Nelms B.E., Zhen H., Tomé W.A. Per-beam, planar IMRT QA passing rates do not predict clinically relevant patient dose errors. Med Phys 2011, 38(2):1037-1044.
    • (2011) Med Phys , vol.38 , Issue.2 , pp. 1037-1044
    • Nelms, B.E.1    Zhen, H.2    Tomé, W.A.3
  • 24
    • 84899490264 scopus 로고    scopus 로고
    • Clinically relevant quality assurance (QA) for prostate RapidArc plans: gamma maps and DVH-based evaluation
    • Cozzolino M., Oliviero C., Califano G., Clemente S., Pedicini P., Caivano R., et al. Clinically relevant quality assurance (QA) for prostate RapidArc plans: gamma maps and DVH-based evaluation. Phys Medica 2014, 30(4):462-472.
    • (2014) Phys Medica , vol.30 , Issue.4 , pp. 462-472
    • Cozzolino, M.1    Oliviero, C.2    Califano, G.3    Clemente, S.4    Pedicini, P.5    Caivano, R.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.