Volumn 273, Issue 3, 2014, Pages 657-672

How I do it: Managing radiation dose in CT

Author keywords

[No Author keywords available]

Indexed keywords

ALGORITHM; CLINICAL PROTOCOL; COMPUTER ASSISTED TOMOGRAPHY; DOSE LENGTH PRODUCT; EFFECTIVE TUBE CURRENT TIME PRODUCT; FILTERED BACK PROJECTION; HUMAN; IMAGE ANALYSIS; IMAGE QUALITY; IMAGE RECONSTRUCTION; ITERATIVE RECONSTRUCTION PROJECTION; MEDICAL EDUCATION; MEDICAL TERMINOLOGY; NON INVASIVE PROCEDURE; PATIENT EDUCATION; RADIATION ABSORPTION; RADIATION DOSE; RADIATION DOSE REDUCTION; RADIATION EXPOSURE; RADIATION HAZARD; RADIATION MONITORING; RADIATION RESPONSE; RADIATION SAFETY; RADIOLOGICAL PARAMETERS; REVIEW; RISK ASSESSMENT; SIZE SPECIFIC DOSE ESTIMATE; TUBE CURRENT MODULATION; TUBE CURRENT TIME PRODUCT; TUBE VOLTAGE; TUBE VOLTAGE MODIFICATION; VOLUME COMPUTED TOMOGRAPHY DOSE INDEX; COMPUTER ASSISTED DIAGNOSIS; EDUCATION; NOMENCLATURE; ORGANIZATION AND MANAGEMENT; PATIENT CARE; PATIENT SAFETY; PROCEDURES; RADIATION PROTECTION; RADIOLOGY; RADIOMETRY;

EID: 84916624353     PISSN: 00338419     EISSN: 15271315     Source Type: Journal    
DOI: 10.1148/radiol.14132328     Document Type: Review
Times cited : (154)

References (84)
  • 1
    • 0035142347 scopus 로고    scopus 로고
    • Taking care of children: Check out the parameters used for helical CT
    • Rogers LF. Taking care of children: Check out the parameters used for helical CT. AJR Am J Roentgenol 2001;176(2):287
    • (2001) AJR Am J Roentgenol , vol.176 , Issue.2 , pp. 287
    • Rogers, L.F.1
  • 2
    • 0012444335 scopus 로고    scopus 로고
    • Radiation exposure in CT why so high?
    • Rogers LF. Radiation exposure in CT: why so high? AJR Am J Roentgenol 2001;177(2):277
    • (2001) AJR Am J Roentgenol , vol.177 , Issue.2 , pp. 277
    • Rogers, L.F.1
  • 3
    • 36549007486 scopus 로고    scopus 로고
    • Computed tomography: An increasing source of radiation exposure
    • Brenner DJ, Hall EJ. Computed tomography: An increasing source of radiation exposure. N Engl J Med 2007;357(22):2277-2284
    • (2007) N Engl J Med , vol.357 , Issue.22 , pp. 2277-2284
    • Brenner, D.J.1    Hall, E.J.2
  • 4
    • 67650536250 scopus 로고    scopus 로고
    • National Council on Radiation Protection and Measurements, NCRP report. Bethesda Md: National Council on Radiation Protection and Measurements
    • National Council on Radiation Protection and Measurements. Ionizing radiation exposure of the population of the United States. NCRP report no. 160. Bethesda, Md: National Council on Radiation Protection and Measurements, 2009
    • (2009) Ionizing radiation exposure of the population of the United States
  • 5
    • 84928100651 scopus 로고    scopus 로고
    • Medical radiation soars, with risks often overlooked
    • Brody J. Medical radiation soars, with risks often overlooked. New York Times. August 20, 2012:D7.http://well.blogs.nytimes. com/2012/08/20/medical-radiation-soarswith-risks-often-overlooked/
    • (2012) New York Times. August , vol.20 , pp. D7
    • Brody, J.1
  • 6
    • 84899680666 scopus 로고    scopus 로고
    • We are giving ourselves cancer
    • Redberg R, Smith-Bindman R. We are giving ourselves cancer. New York Times. January 31, 2014:A27.http://www.nytimes. com/2014/01/31/opinion/we-are-givingourselves-cancer.html?-r=0
    • (2014) New York Times. January , vol.31 , pp. A27
    • Redberg, R.1    Smith-Bindman, R.2
  • 7
    • 78149490137 scopus 로고    scopus 로고
    • U.S. Food and Drug Administration Web site.Published February 16, 2010. Updated December 14
    • White paper: Initiative to reduce unnecessary radiation exposure from medical imaging. U.S. Food and Drug Administration Web site.http://www.fda.gov/Radiation-EmittingProducts/RadiationSafety/RadiationDoseReduction/ucm199994.htm. Published February 16, 2010. Updated December 14, 2010
    • (2010) White paper: Initiative to reduce unnecessary radiation exposure from medical imaging
  • 9
    • 84916599953 scopus 로고    scopus 로고
    • Joint Commission Web site, Published December 20
    • Joint Commission announces new and revised diagnostic imaging standards. Joint Commission Web site. http://www.jointcommission.org/joint-commission-announces-new-and-revised-diagnostic-imaging-standards/. Published December 20 2013
    • (2013) Joint Commission announces new and revised diagnostic imaging standards
  • 10
    • 79952299747 scopus 로고    scopus 로고
    • Managing radiation use in medical imaging: A multifaceted challenge
    • Hricak H, Brenner DJ, Adelstein SJ, et al. Managing radiation use in medical imaging: A multifaceted challenge. Radiology 2011;258(3):889-905
    • (2011) Radiology. , vol.258 , Issue.3 , pp. 889-905
    • Hricak, H.1    Brenner, D.J.2    Adelstein, S.J.3
  • 11
    • 81455139639 scopus 로고    scopus 로고
    • Strategies for reducing radiation exposure in multi-detector row CT
    • Sodickson A. Strategies for reducing radiation exposure in multi-detector row CT. Radiol Clin North Am 2012;50(1):1-14
    • (2012) Radiol Clin North Am , vol.50 , Issue.1 , pp. 1-14
    • Sodickson, A.1
  • 12
    • 80053082299 scopus 로고    scopus 로고
    • CT radiation dose: Trending in the right direction
    • Amis ES Jr. CT radiation dose: Trending in the right direction. Radiology 2011;261(1):5-8
    • (2011) Radiology. , vol.261 , Issue.1 , pp. 5-8
    • Amis, E.S.1
  • 13
    • 84928096790 scopus 로고    scopus 로고
    • American College of Radiology. ACR white paper on radiation dose in medicine: Three years later
    • Amis ES Jr, Butler PF; American College of Radiology. ACR white paper on radiation dose in medicine: Three years later. J Am Coll Radiol 2010;7(11):865-870
    • (2010) J Am Coll Radiol , vol.7 , Issue.11 , pp. 865-870
    • Amis, E.S.1    Butler, P.F.2
  • 14
    • 84864373822 scopus 로고    scopus 로고
    • Achieving routine submillisievert CT scanning: Report from the summit on management of radiation dose in CT
    • McCollough CH, Chen GH, Kalender W, et al. Achieving routine submillisievert CT scanning: Report from the summit on management of radiation dose in CT. Radiology 2012;264(2):567-580
    • (2012) Radiology. , vol.264 , Issue.2 , pp. 567-580
    • McCollough, C.H.1    Chen, G.H.2    Kalender, W.3
  • 15
    • 33645562681 scopus 로고    scopus 로고
    • CT dose reduction and dose management tools: Overview of available options
    • McCollough CH, Bruesewitz MR, Kofler JM Jr. CT dose reduction and dose management tools: Overview of available options. RadioGraphics 2006;26(2):503-512
    • (2006) RadioGraphics , vol.26 , Issue.2 , pp. 503-512
    • McCollough, C.H.1    Bruesewitz, M.R.2    Kofler, J.M.3
  • 16
    • 79952272541 scopus 로고    scopus 로고
    • CT radiation dose what can you do right now in your practice?
    • Coakley FV, Gould R, Yeh BM, Arenson RL. CT radiation dose: what can you do right now in your practice? AJR Am J Roentgenol 2011;196(3):619-625
    • (2011) AJR Am J Roentgenol , vol.196 , Issue.3 , pp. 619-625
    • Coakley, F.V.1    Gould, R.2    Yeh, B.M.3    Arenson, R.L.4
  • 17
    • 64949094007 scopus 로고    scopus 로고
    • Risks associated with low doses and low dose rates of ionizing radiation: Why linearity may be (almost) the best we can do
    • Little MP, Wakeford R, Tawn EJ, Bouffler SD, Berrington de Gonzalez A. Risks associated with low doses and low dose rates of ionizing radiation: why linearity may be (almost) the best we can do. Radiology 2009;251(1):6-12
    • (2009) Radiology , vol.251 , Issue.1 , pp. 6-12
    • Little, M.P.1    Wakeford, R.2    Tawn, E.J.3    Bouffler, S.D.4    Berrington De Gonzalez, A.5
  • 18
    • 64949191402 scopus 로고    scopus 로고
    • The linear no-threshold re-lationship is inconsistent with radiation biologic and experimental data
    • Tubiana M, Feinendegen LE, Yang C, Kaminski JM. The linear no-threshold re-lationship is inconsistent with radiation biologic and experimental data. Radiology 2009;251(1):13-22
    • (2009) Radiology , vol.251 , Issue.1 , pp. 13-22
    • Tubiana, M.1    Feinendegen, L.E.2    Yang, C.3    Kaminski, J.M.4
  • 19
    • 34548581922 scopus 로고    scopus 로고
    • Solid cancer incidence in atomic bomb survivors 1958-1998
    • Preston DL, Ron E, Tokuoka S, et al. Solid cancer incidence in atomic bomb survivors: 1958-1998. Radiation Research 2007;168(1): 1-64
    • (2007) Radiation Research , vol.168 , Issue.1 , pp. 1-64
    • Preston, D.L.1    Ron, E.2    Tokuoka, S.3
  • 21
    • 58149240065 scopus 로고    scopus 로고
    • Mortality and cancer incidence following occupational radiation exposure: Third analysis of the National Registry for Radiation Workers
    • Muirhead CR, O'Hagan JA, Haylock RG, et al. Mortality and cancer incidence following occupational radiation exposure: Third analysis of the National Registry for Radiation Workers. British J Cancer 2009;100(1):206-12
    • (2009) British J Cancer , vol.100 , Issue.1 , pp. 206-12
    • Muirhead, C.R.1    O'Hagan, J.A.2    Haylock, R.G.3
  • 22
    • 0036838837 scopus 로고    scopus 로고
    • Cancer risk from low-level radiation
    • Cohen BL. Cancer risk from low-level radiation. AJR Am J Roentgeno 2002;179(5): 1137-1143
    • (2002) AJR Am J Roentgeno , vol.179 , Issue.5 , pp. 1137-1143
    • Cohen, B.L.1
  • 23
    • 84867912726 scopus 로고    scopus 로고
    • Radiation exposure from CT scans: How to close our knowledge gaps, monitor and safeguard exposure-proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24-25, 2011
    • Boone JM, Hendee WR, McNitt-Gray MF, Seltzer SE. Radiation exposure from CT scans: How to close our knowledge gaps, monitor and safeguard exposure-proceedings and recommendations of the Radiation Dose Summit, sponsored by NIBIB, February 24-25, 2011. Radiology 2012;265(2):544-554
    • (2012) Radiology , vol.265 , Issue.2 , pp. 544-554
    • Boone, J.M.1    Hendee, W.R.2    McNitt-Gray, M.F.3    Seltzer, S.E.4
  • 25
    • 84555178181 scopus 로고    scopus 로고
    • Systematic review and meta-analysis of immediate total-body computed tomography compared with selective radiological imaging of injured patients
    • Sierink JC, Saltzherr TP, Reitsma JB, Van Delden OM, Luitse JS, Goslings JC. Systematic review and meta-analysis of immediate total-body computed tomography compared with selective radiological imaging of injured patients. Br J Surg 2012;99(Suppl 1): 52-58
    • (2012) Br J Surg , vol.99 , pp. 52-58
    • Sierink, J.C.1    Saltzherr, T.P.2    Reitsma, J.B.3    Van Delden, O.M.4    Luitse, J.S.5    Goslings, J.C.6
  • 26
    • 0032518878 scopus 로고    scopus 로고
    • Effect of computed tomography of the appendix on treatment of patients and use of hospital resources
    • Rao PM, Rhea JT, Novelline RA, Mostafavi AA, McCabe CJ. Effect of computed tomography of the appendix on treatment of patients and use of hospital resources. N Engl J Med 1998;338(3):141-146
    • (1998) N Engl J Med , vol.338 , Issue.3 , pp. 141-146
    • Rao, P.M.1    Rhea, J.T.2    Novelline, R.A.3    Mostafavi, A.A.4    McCabe, C.J.5
  • 27
    • 77958598531 scopus 로고    scopus 로고
    • Cost-effectiveness of CT perfusion for selecting patients for intravenous thrombolysis: A US hospital perspective
    • Jackson D, Earnshaw SR, Farkouh R, Schwamm L. Cost-effectiveness of CT perfusion for selecting patients for intravenous thrombolysis: A US hospital perspective. AJNR Am J Neuroradiol 2010;31(9):1669-1674
    • (2010) AJNR Am J Neuroradiol , vol.31 , Issue.9 , pp. 1669-1674
    • Jackson, D.1    Earnshaw, S.R.2    Farkouh, R.3    Schwamm, L.4
  • 28
    • 84876928134 scopus 로고    scopus 로고
    • Body CT scanning in young adults: Examination indications, patient outcomes, and risk of radiation-induced cancer
    • Zondervan RL, Hahn PF, Sadow CA, Liu B, Lee SI. Body CT scanning in young adults: Examination indications, patient outcomes, and risk of radiation-induced cancer. Radiology 2013;267(2):460-469
    • (2013) Radiology. , vol.267 , Issue.2 , pp. 460-469
    • Zondervan, R.L.1    Hahn, P.F.2    Sadow, C.A.3    Liu, B.4    Lee, S.I.5
  • 29
    • 73149110471 scopus 로고    scopus 로고
    • Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer
    • Smith-Bindman R, Lipson J, Marcus R, et al. Radiation dose associated with common computed tomography examinations and the associated lifetime attributable risk of cancer. Arch Intern Med 2009;169(22): 2078-2086
    • (2009) Arch Intern Med , vol.169 , Issue.22 , pp. 2078-2086
    • Smith-Bindman, R.1    Lipson, J.2    Marcus, R.3
  • 30
    • 77954230504 scopus 로고    scopus 로고
    • Is computed tomography safe?
    • Smith-Bindman R. Is computed tomography safe? N Engl J Med 2010;363(1):1-4
    • (2010) N Engl J Med , vol.363 , Issue.1 , pp. 1-4
    • Smith-Bindman, R.1
  • 31
    • 73149113560 scopus 로고    scopus 로고
    • Projected cancer risks from computed tomographic scans performed in the United States in 2007
    • Berrington de González A, Mahesh M, Kim KP, et al. Projected cancer risks from computed tomographic scans performed in the United States in 2007. Arch Intern Med 2009;169(22):2071-2077
    • (2009) Arch Intern Med , vol.169 , Issue.22 , pp. 2071-2077
    • Berrington De González, A.1    Mahesh, M.2    Kim, K.P.3
  • 32
    • 84926231441 scopus 로고    scopus 로고
    • Utilization strategies for cumulative dose estimates: A review and rational assessment
    • Durand DJ, Dixon RL, Morin RL. Utilization strategies for cumulative dose estimates: A review and rational assessment. J Am Coll Radiol 2012;9(7):480-485
    • (2012) J Am Coll Radiol , vol.9 , Issue.7 , pp. 480-485
    • Durand, D.J.1    Dixon, R.L.2    Morin, R.L.3
  • 33
    • 84928096409 scopus 로고    scopus 로고
    • The choosing wisely campaign and its potential impact on diagnostic radiation burden
    • Mahesh M, Durand DJ. The choosing wisely campaign and its potential impact on diagnostic radiation burden. J Am Coll Radiol 2013;10(1):65-66
    • (2013) J Am Coll Radiol , vol.10 , Issue.1 , pp. 65-66
    • Mahesh, M.1    Durand, D.J.2
  • 34
    • 44349176789 scopus 로고    scopus 로고
    • Radiation safety in the era of helical CT: A patient-based protection program currently in place in two community hospitals in New Hampshire
    • Birnbaum S. Radiation safety in the era of helical CT: A patient-based protection program currently in place in two community hospitals in New Hampshire. J Am Coll Radiol 2008;5(6):714-718, e5
    • (2008) J Am Coll Radiol , vol.5 , Issue.6 , pp. 714-718+e5
    • Birnbaum, S.1
  • 35
    • 84916633245 scopus 로고    scopus 로고
    • ACR appropriateness criteria. American College of Radiology Web site.Updated May 2014
    • ACR appropriateness criteria. American College of Radiology Web site.www.acr. org/Quality-Safety/Appropriateness-Criteria. Updated May 2014
  • 36
    • 84863267489 scopus 로고    scopus 로고
    • The Royal College of Radiologists, Royal College of Radiologists Web site. Accessed January 13
    • The Royal College of Radiologists. RCR referral guidelines. Royal College of Radiologists Web site.http://www.rcr.ac.uk/content.aspx?PageID=995. Accessed January 13, 2014
    • (2014) RCR referral guidelines
  • 39
    • 84916633244 scopus 로고    scopus 로고
    • ABIM Foundation Choosing Wisely Web site.
    • ABIM Foundation. Choosing Wisely Web site.http://www.choosingwisely.org
  • 40
    • 84928096084 scopus 로고    scopus 로고
    • Radiology order entry with decision support: Initial clinical experience
    • Rosenthal DI, Weilburg JB, Schultz T, et al. Radiology order entry with decision support: Initial clinical experience. J Am Coll Radiol 2006;3(10):799-806
    • (2006) J Am Coll Radiol , vol.3 , Issue.10 , pp. 799-806
    • Rosenthal, D.I.1    Weilburg, J.B.2    Schultz, T.3
  • 41
    • 64949091695 scopus 로고    scopus 로고
    • Effect of computerized order entry with integrated decision support on the growth of outpatient procedure volumes: Seven-year time series analysis
    • Sistrom CL, Dang PA, Weilburg JB, Dreyer KJ, Rosenthal DI, Thrall JH. Effect of computerized order entry with integrated decision support on the growth of outpatient procedure volumes: Seven-year time series analysis. Radiology 2009;251(1):147-155
    • (2009) Radiology , vol.251 , Issue.1 , pp. 147-155
    • Sistrom, C.L.1    Dang, P.A.2    Weilburg, J.B.3    Dreyer, K.J.4    Rosenthal, D.I.5    Thrall, J.H.6
  • 42
  • 43
    • 79960689293 scopus 로고    scopus 로고
    • Outside imaging in emergency department transfer patients: CD import reduces rates of subsequent imaging utilization
    • Sodickson A, Opraseuth J, Ledbetter S. Outside imaging in emergency department transfer patients: CD import reduces rates of subsequent imaging utilization. Radiology 2011; 260(2):408-413
    • (2011) Radiology. , vol.260 , Issue.2 , pp. 408-413
    • Sodickson, A.1    Opraseuth, J.2    Ledbetter, S.3
  • 44
    • 84928096604 scopus 로고    scopus 로고
    • CT radiation dose reduction by modifying primary factors
    • Singh S, Kalra MK, Thrall JH, Mahesh M. CT radiation dose reduction by modifying primary factors. J Am Coll Radiol 2011;8(5): 369-372
    • (2011) J Am Coll Radiol , vol.8 , Issue.5 , pp. 369-372
    • Singh, S.1    Kalra, M.K.2    Thrall, J.H.3    Mahesh, M.4
  • 45
    • 78650629641 scopus 로고    scopus 로고
    • Volume CT dose index and dose-length product displayed during CT: What good are they?
    • Huda W, Mettler FA. Volume CT dose index and dose-length product displayed during CT: what good are they? Radiology 2011; 258(1):236-242
    • (2011) Radiology. , vol.258 , Issue.1 , pp. 236-242
    • Huda, W.1    Mettler, F.A.2
  • 46
    • 73649148717 scopus 로고    scopus 로고
    • Automatic selection of tube potential for radiation dose reduction in CT: A general strategy
    • Yu L, Li H, Fletcher JG, McCollough CH. Automatic selection of tube potential for radiation dose reduction in CT: A general strategy. Med Phys 2010;37(1):234-243
    • (2010) Med Phys , vol.37 , Issue.1 , pp. 234-243
    • Yu, L.1    Li, H.2    Fletcher, J.G.3    McCollough, C.H.4
  • 47
    • 81855170190 scopus 로고    scopus 로고
    • Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: Improved dose effectiveness
    • Winklehner A, Goetti R, Baumueller S, et al. Automated attenuation-based tube potential selection for thoracoabdominal computed tomography angiography: Improved dose effectiveness. Invest Radiol 2011;46(12): 767-773
    • (2011) Invest Radiol , vol.46 , Issue.12 , pp. 767-773
    • Winklehner, A.1    Goetti, R.2    Baumueller, S.3
  • 48
    • 84868035910 scopus 로고    scopus 로고
    • Lowering kilovoltage to reduce radiation dose in contrast-enhanced abdominal CT: Initial assessment of a prototype automated kilovoltage selection tool
    • Hough DM, Fletcher JG, Grant KL, et al. Lowering kilovoltage to reduce radiation dose in contrast-enhanced abdominal CT: Initial assessment of a prototype automated kilovoltage selection tool. AJR Am J Roentgenol 2012;199(5):1070-1077
    • (2012) AJR Am J Roentgenol , vol.199 , Issue.5 , pp. 1070-1077
    • Hough, D.M.1    Fletcher, J.G.2    Grant, K.L.3
  • 49
    • 33845741615 scopus 로고    scopus 로고
    • 16-MDCT angiography in living kidney donors at various tube potentials: Impact on image quality and radiation dose
    • Sahani DV, Kalva SP, Hahn PF, Saini S. 16-MDCT angiography in living kidney donors at various tube potentials: Impact on image quality and radiation dose. AJR Am J Roentgenol 2007;188(1):115-120
    • (2007) AJR Am J Roentgenol , vol.188 , Issue.1 , pp. 115-120
    • Sahani, D.V.1    Kalva, S.P.2    Hahn, P.F.3    Saini, S.4
  • 50
    • 84865588644 scopus 로고    scopus 로고
    • Automated attenuation-based kilovoltage selection: Preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta
    • Goetti R, Winklehner A, Gordic S, et al. Automated attenuation-based kilovoltage selection: Preliminary observations in patients after endovascular aneurysm repair of the abdominal aorta. AJR Am J Roentgenol 2012;199(3):W380-W385
    • (2012) AJR Am J Roentgenol , vol.199 , Issue.3 , pp. W380-W385
    • Goetti, R.1    Winklehner, A.2    Gordic, S.3
  • 51
    • 84867914269 scopus 로고    scopus 로고
    • Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT
    • Lee KH, Lee JM, Moon SK, et al. Attenuation-based automatic tube voltage selection and tube current modulation for dose reduction at contrast-enhanced liver CT. Radiology 2012;265(2):437-447
    • (2012) Radiology. , vol.265 , Issue.2 , pp. 437-447
    • Lee, K.H.1    Lee, J.M.2    Moon, S.K.3
  • 52
    • 0035187964 scopus 로고    scopus 로고
    • Dose and pitch relationship for a particular multislice CT scanner
    • Mahesh M, Scatarige JC, Cooper J, Fishman EK. Dose and pitch relationship for a particular multislice CT scanner. AJR Am J Roentgenol 2001;177(6):1273-1275
    • (2001) AJR Am J Roentgenol , vol.177 , Issue.6 , pp. 1273-1275
    • Mahesh, M.1    Scatarige, J.C.2    Cooper, J.3    Fishman, E.K.4
  • 53
    • 84858293727 scopus 로고    scopus 로고
    • Iterative reconstruction methods in X-ray CT
    • Beister M, Kolditz D, Kalender WA. Iterative reconstruction methods in X-ray CT. Phys Med 2012;28(2):94-108
    • (2012) Phys Med , vol.28 , Issue.2 , pp. 94-108
    • Beister, M.1    Kolditz, D.2    Kalender, W.A.3
  • 54
    • 84886661454 scopus 로고    scopus 로고
    • Model-based iterative reconstruction and adaptive statistical iterative reconstruction techniques in abdominal CT: Comparison of image quality in the detection of colorectal liver metastases
    • Volders D, Bols A, Haspeslagh M, Coenegrachts K. Model-based iterative reconstruction and adaptive statistical iterative reconstruction techniques in abdominal CT: Comparison of image quality in the detection of colorectal liver metastases. Radiology 2013;269(2):469-474
    • (2013) Radiology. , vol.269 , Issue.2 , pp. 469-474
    • Volders, D.1    Bols, A.2    Haspeslagh, M.3    Coenegrachts, K.4
  • 55
    • 69949096664 scopus 로고    scopus 로고
    • Iterative reconstruction technique for reducing body radia tion dose at CT: Feasibility study
    • Hara AK, Paden RG, Silva AC, Kujak JL, Lawder HJ, Pavlicek W. Iterative reconstruction technique for reducing body radia tion dose at CT: feasibility study. AJR Am J Roentgenol 2009;193(3):764-771
    • (2009) AJR Am J Roentgenol , vol.193 , Issue.3 , pp. 764-771
    • Hara, A.K.1    Paden, R.G.2    Silva, A.C.3    Kujak, J.L.4    Lawder, H.J.5    Pavlicek, W.6
  • 56
    • 35649025312 scopus 로고    scopus 로고
    • A three-dimensional statistical approach to improved image quality for multislice helical CT
    • Thibault JB, Sauer KD, Bouman CA, Hsieh J. A three-dimensional statistical approach to improved image quality for multislice helical CT. Med Phys 2007;34(11): 4526-4544
    • (2007) Med Phys , vol.34 , Issue.11 , pp. 4526-4544
    • Thibault, J.B.1    Sauer, K.D.2    Bouman, C.A.3    Hsieh, J.4
  • 57
    • 84866635102 scopus 로고    scopus 로고
    • Radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: How low can we go?
    • Kulkarni NM, Uppot RN, Eisner BH, Sahani DV. Radiation dose reduction at multidetector CT with adaptive statistical iterative reconstruction for evaluation of urolithiasis: How low can we go? Radiology 2012;265(1): 158-166
    • (2012) Radiology. , vol.265 , Issue.1 , pp. 158-166
    • Kulkarni, N.M.1    Uppot, R.N.2    Eisner, B.H.3    Sahani, D.V.4
  • 58
    • 84863666413 scopus 로고    scopus 로고
    • Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: A pilot study
    • Yamada Y, Jinzaki M, Tanami Y, et al. Model-based iterative reconstruction technique for ultralow-dose computed tomography of the lung: A pilot study. Invest Radiol 2012;47(8):482-489
    • (2012) Invest Radiol , vol.47 , Issue.8 , pp. 482-489
    • Yamada, Y.1    Jinzaki, M.2    Tanami, Y.3
  • 59
    • 84863721418 scopus 로고    scopus 로고
    • Radiation dose reduction in abdominal computed tomography during the late hepatic arterial phase using a model-based iterative reconstruction algorithm: How low can we go?
    • Husarik DB, Marin D, Samei E, et al. Radiation dose reduction in abdominal computed tomography during the late hepatic arterial phase using a model-based iterative reconstruction algorithm: How low can we go? Invest Radiol 2012;47(8):468-474
    • (2012) Invest Radiol , vol.47 , Issue.8 , pp. 468-474
    • Husarik, D.B.1    Marin, D.2    Samei, E.3
  • 60
    • 84870903463 scopus 로고    scopus 로고
    • Abdominal CT with model-based iterative reconstruction (MBIR): Initial results of a prospective trial comparing ultralow-dose with standard-dose imaging
    • Pickhardt PJ, Lubner MG, Kim DH, et al. Abdominal CT with model-based iterative reconstruction (MBIR): Initial results of a prospective trial comparing ultralow-dose with standard-dose imaging. AJR Am J Roentgenol 2012;199(6):1266-1274
    • (2012) AJR Am J Roentgenol , vol.199 , Issue.6 , pp. 1266-1274
    • Pickhardt, P.J.1    Lubner, M.G.2    Kim, D.H.3
  • 62
    • 80052292210 scopus 로고    scopus 로고
    • National Electrical Manufacturers Association Web site.Published October 1, Accessed December 23 2013
    • Computed tomography dose check: NEMA XR 25-2010. National Electrical Manufacturers Association Web site.http://www.nema.org/standards/Pages/Computed-Tomography-Dose-Check.aspx. Published October 1, 2010. Accessed December 23, 2013
    • (2010) Computed tomography dose check: NEMA XR 25-2010
  • 63
    • 84916637187 scopus 로고    scopus 로고
    • American College of Radiology Web site.Updated November 18
    • Computed tomography accreditation. American College of Radiology Web site.http://http://www.acr.org/Quality-Safety/Accreditation/CT. Updated November 18, 2013
    • (2013) Computed tomography accreditation
  • 64
    • 47249166093 scopus 로고    scopus 로고
    • Effective doses in radiology and diagnostic nuclear medicine: A catalog
    • Mettler FA Jr, Huda W, Yoshizumi TT, Mahesh M. Effective doses in radiology and diagnostic nuclear medicine: A catalog. Radiology 2008;248(1):254-263
    • (2008) Radiology , vol.248 , Issue.1 , pp. 254-263
    • Mettler, F.A.1    Huda, W.2    Yoshizumi, T.T.3    Mahesh, M.4
  • 65
    • 77950600954 scopus 로고    scopus 로고
    • How effective is effective dose as a predictor of radiation risk?
    • McCollough CH, Christner JA, Kofler JM. How effective is effective dose as a predictor of radiation risk? AJR Am J Roentgenol 2010;194(4):890-896
    • (2010) AJR Am J Roentgenol , vol.194 , Issue.4 , pp. 890-896
    • McCollough, C.H.1    Christner, J.A.2    Kofler, J.M.3
  • 66
    • 51549112607 scopus 로고    scopus 로고
    • American Association of Physicists in Medicine, Task group 96. College Park, Md: American Association of Physicists in Medicine
    • American Association of Physicists in Medicine. The measurement, reporting and management of radiation dose in CT. Task group 96. College Park, Md: American Association of Physicists in Medicine, 2008
    • (2008) The measurement, reporting and management of radiation dose in CT
  • 67
    • 51549088875 scopus 로고    scopus 로고
    • Converting dose-length product to effective dose at CT
    • Huda W, Ogden KM, Khorasani MR. Converting dose-length product to effective dose at CT. Radiology 2008;248(3):995-1003
    • (2008) Radiology , vol.248 , Issue.3 , pp. 995-1003
    • Huda, W.1    Ogden, K.M.2    Khorasani, M.R.3
  • 68
    • 84870020657 scopus 로고    scopus 로고
    • Size-specific dose estimation for CT: How should it be used and what does it mean?
    • Brink JA, Morin RL. Size-specific dose estimation for CT: How should it be used and what does it mean? Radiology 2012;265(3): 666-668
    • (2012) Radiology. , vol.265 , Issue.3 , pp. 666-668
    • Brink, J.A.1    Morin, R.L.2
  • 69
    • 84862505274 scopus 로고    scopus 로고
    • American Association of Physicists in Medicine, Task Group 204. College Park Md: American Association of Physicists in Medicine
    • American Association of Physicists in Medicine. Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations. Task Group 204. College Park, Md: American Association of Physicists in Medicine, 2011
    • (2011) Size-specific dose estimates (SSDE) in pediatric and adult body CT examinations
  • 71
    • 81255184726 scopus 로고    scopus 로고
    • Quality initiatives: CT radiation dose reduction: How to implement change without sacrificing diagnostic quality
    • Tamm EP, Rong XJ, Cody DD, Ernst RD, Fitzgerald NE, Kundra V. Quality initiatives: CT radiation dose reduction: How to implement change without sacrificing diagnostic quality. RadioGraphics 2011;31(7):1823-1832
    • (2011) RadioGraphics , vol.31 , Issue.7 , pp. 1823-1832
    • Tamm, E.P.1    Rong, X.J.2    Cody, D.D.3    Ernst, R.D.4    Fitzgerald, N.E.5    Kundra, V.6
  • 72
    • 84884690196 scopus 로고    scopus 로고
    • American College of Radiology Web site.
    • Dose index registry. American College of Radiology Web site.http://www.acr.org/Quality-Safety/National-Radiology-Data-Registry/Dose-Index-Registry
    • Dose index registry
  • 73
    • 84890438755 scopus 로고    scopus 로고
    • Multidetector CT radiation dose optimisation in adults: Short-and long-term effects of a clinical audit
    • Tack D, Jahnen A, Kohler S, et al. Multidetector CT radiation dose optimisation in adults: Short-and long-term effects of a clinical audit. Eur Radiol 2014;24(1):169-175
    • (2014) Eur Radiol , vol.24 , Issue.1 , pp. 169-175
    • Tack, D.1    Jahnen, A.2    Kohler, S.3
  • 74
    • 84916633243 scopus 로고    scopus 로고
    • SB No. 1237
    • California 2013 Radiation Protection Act, SB No. 1237.http://www.leginfo.ca.gov/pub/09-10/bill/sen/sb-1201-1250/sb-1237-bill-20100929-chaptered.html
    • California 2013 Radiation Protection Act
  • 75
    • 84916636448 scopus 로고    scopus 로고
    • American Association of Physicists in Medicine Web site
    • AAPM CT scan protocols. American Association of Physicists in Medicine Web site.http://www.aapm.org/pubs/CTProtocols
    • AAPM CT scan protocols
  • 76
    • 84916633242 scopus 로고    scopus 로고
    • Rhode Island Hospital Web site.
    • Brown Medical School CT protocols. Rhode Island Hospital Web site.http://www.rhodeislandhospital.org/services/diagnosticimaging/computed-tomography/protocols/ct-protocols/
    • Brown Medical School CT protocols
  • 77
    • 84916633241 scopus 로고    scopus 로고
    • CTisus Web site.
    • CTisus CT protocols. CTisus Web site.http: //www.ctisus.org/ctprotocols/protocolsin dex.html
    • CTisus CT protocols
  • 78
    • 84874301989 scopus 로고    scopus 로고
    • Reducing radiation dose at CT colonography: Decreasing tube voltage to 100 kVp
    • Chang KJ, Caovan DB, Grand DJ, Huda W, Mayo-Smith WW. Reducing radiation dose at CT colonography: Decreasing tube voltage to 100 kVp. Radiology 2013;266(3):791-800
    • (2013) Radiology. , vol.266 , Issue.3 , pp. 791-800
    • Chang, K.J.1    Caovan, D.B.2    Grand, D.J.3    Huda, W.4    Mayo-Smith, W.W.5
  • 79
    • 70449135326 scopus 로고    scopus 로고
    • Efficient use of automatic exposure control systems in computed tomography requires correct patient positioning
    • Gudjonsdottir J, Svensson JR, Campling S, Brennan PC, Jonsdottir B. Efficient use of automatic exposure control systems in computed tomography requires correct patient positioning. Acta Radiol 2009;50(9): 1035-1041
    • (2009) Acta Radiol , vol.50 , Issue.9 , pp. 1035-1041
    • Gudjonsdottir, J.1    Svensson, J.R.2    Campling, S.3    Brennan, P.C.4    Jonsdottir, B.5
  • 81
    • 84862170887 scopus 로고    scopus 로고
    • Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies
    • Habibzadeh MA, Ay MR, Asl AR, Ghadiri H, Zaidi H. Impact of miscentering on patient dose and image noise in x-ray CT imaging: Phantom and clinical studies. Phys Med 2012;28(3):191-199
    • (2012) Phys Med , vol.28 , Issue.3 , pp. 191-199
    • Habibzadeh, M.A.1    Ay, M.R.2    Asl, A.R.3    Ghadiri, H.4    Zaidi, H.5
  • 82
    • 34547583507 scopus 로고    scopus 로고
    • Split-bolus MDCT urography with synchronous nephrographic and excretory phase enhancement
    • Chow LC, Kwan SW, Olcott EW, Sommer G. Split-bolus MDCT urography with synchronous nephrographic and excretory phase enhancement. AJR Am J Roentgenol 2007;189(2):314-322
    • (2007) AJR Am J Roentgenol , vol.189 , Issue.2 , pp. 314-322
    • Chow, L.C.1    Kwan, S.W.2    Olcott, E.W.3    Sommer, G.4
  • 83
    • 84862639230 scopus 로고    scopus 로고
    • Virtual nonenhanced dual-energy CT urography with tin-filter technology: Determinants of detection of urinary calculi in the renal collecting system
    • Mangold S, Thomas C, Fenchel M, et al. Virtual nonenhanced dual-energy CT urography with tin-filter technology: Determinants of detection of urinary calculi in the renal collecting system. Radiology 2012;264(1): 119-125
    • (2012) Radiology. , vol.264 , Issue.1 , pp. 119-125
    • Mangold, S.1    Thomas, C.2    Fenchel, M.3
  • 84
    • 56149093386 scopus 로고    scopus 로고
    • CT dose: How to measure, how to reduce
    • McCollough C.H.
    • McCollough CH. CT dose: How to measure, how to reduce. Health Phys 2008;95(5): 508-517.
    • (2008) Health Phys , vol.95 , Issue.5 , pp. 508-517


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.