메뉴 건너뛰기




Volumn 28, Issue 3, 2013, Pages 469-473

Unsatisfactory Accuracy as Determined by Computer Navigation of VISIONAIRE Patient-Specific Instrumentation for Total Knee Arthroplasty

Author keywords

Accuracy; Navigation; Patient specific instrumentation; Total knee arthroplasty

Indexed keywords

ACCURACY; ADULT; AGED; ARTHROTOMY; ARTICLE; COMPUTER AIDED DESIGN; COMPUTER ASSISTED SURGERY; COMPUTER SYSTEM; FEMALE; FEMUR; HUMAN; KNEE OSTEOARTHRITIS; MAJOR CLINICAL STUDY; MALE; PREOPERATIVE PERIOD; RANGE OF MOTION; SURGICAL EQUIPMENT; SURGICAL TECHNIQUE; TIBIA; TOTAL KNEE REPLACEMENT;

EID: 84874562813     PISSN: 08835403     EISSN: 15328406     Source Type: Journal    
DOI: 10.1016/j.arth.2012.07.012     Document Type: Article
Times cited : (92)

References (29)
  • 1
    • 0036903403 scopus 로고    scopus 로고
    • The effects of varus tibial alignment on proximal tibial surface strain in total knee arthroplasty: the posteromedial hot spot
    • Green G.V., Berend K.R., Berend M.E., et al. The effects of varus tibial alignment on proximal tibial surface strain in total knee arthroplasty: the posteromedial hot spot. J Arthroplasty 2002, 17:1033.
    • (2002) J Arthroplasty , vol.17 , pp. 1033
    • Green, G.V.1    Berend, K.R.2    Berend, M.E.3
  • 3
    • 0028044658 scopus 로고
    • Postoperative alignment of total knee replacement. Its effect on survival
    • Ritter M.A., Faris P.M., Keating E.M., et al. Postoperative alignment of total knee replacement. Its effect on survival. Clin Orthop Relat Res 1994, 299:153.
    • (1994) Clin Orthop Relat Res , vol.299 , pp. 153
    • Ritter, M.A.1    Faris, P.M.2    Keating, E.M.3
  • 4
    • 68749101913 scopus 로고    scopus 로고
    • Coronal alignment in total knee arthroplasty: just how important is it?
    • Fang D.M., Ritter M.A., Davis K.E. Coronal alignment in total knee arthroplasty: just how important is it?. J Arthroplasty 2009, 24:39.
    • (2009) J Arthroplasty , vol.24 , pp. 39
    • Fang, D.M.1    Ritter, M.A.2    Davis, K.E.3
  • 5
    • 58149343701 scopus 로고    scopus 로고
    • Patient-specific approach in total knee arthroplasty
    • Lombardi A.V., Berend K.R., Adams J.B. Patient-specific approach in total knee arthroplasty. Orthopedics 2008, 31:927.
    • (2008) Orthopedics , vol.31 , pp. 927
    • Lombardi, A.V.1    Berend, K.R.2    Adams, J.B.3
  • 6
    • 84865102501 scopus 로고    scopus 로고
    • Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA
    • Ng V.Y., DeClaire J.H., Berend K.R., et al. Improved accuracy of alignment with patient-specific positioning guides compared with manual instrumentation in TKA. Clin Orthop Relat Res 2012, 470:99.
    • (2012) Clin Orthop Relat Res , vol.470 , pp. 99
    • Ng, V.Y.1    DeClaire, J.H.2    Berend, K.R.3
  • 7
    • 85056066533 scopus 로고    scopus 로고
    • The reproducibility of a kinematically-derived axis of the knee versus digitized anatomical landmarks using a knee navigation system
    • Doro L.C., Hughes R.E., Miller J.D., et al. The reproducibility of a kinematically-derived axis of the knee versus digitized anatomical landmarks using a knee navigation system. Open Biomed Eng J 2008, 2:52.
    • (2008) Open Biomed Eng J , vol.2 , pp. 52
    • Doro, L.C.1    Hughes, R.E.2    Miller, J.D.3
  • 8
    • 0027231237 scopus 로고
    • Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty
    • Dennis D.A., Channer M., Susman M.H., et al. Intramedullary versus extramedullary tibial alignment systems in total knee arthroplasty. J Arthroplasty 1993, 8:43.
    • (1993) J Arthroplasty , vol.8 , pp. 43
    • Dennis, D.A.1    Channer, M.2    Susman, M.H.3
  • 9
    • 0025350822 scopus 로고
    • Comparative experience with intramedullary and extramedullary alignment in total knee arthroplasty
    • Engh G.A., Petersen T.L. Comparative experience with intramedullary and extramedullary alignment in total knee arthroplasty. J Arthroplasty 1990, 5:1.
    • (1990) J Arthroplasty , vol.5 , pp. 1
    • Engh, G.A.1    Petersen, T.L.2
  • 10
    • 36749020620 scopus 로고    scopus 로고
    • Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery
    • Mason J.B., Fehring T.K., Estok R., et al. Meta-analysis of alignment outcomes in computer-assisted total knee arthroplasty surgery. J Arthroplasty 2007, 22:1097.
    • (2007) J Arthroplasty , vol.22 , pp. 1097
    • Mason, J.B.1    Fehring, T.K.2    Estok, R.3
  • 11
    • 84863409483 scopus 로고    scopus 로고
    • Meta-analysis of navigation vs conventional total knee arthroplasty
    • Hetaimish B.M., Khan M.M., Simunovic N., et al. Meta-analysis of navigation vs conventional total knee arthroplasty. J Arthroplasty 2012, 27(6):1177.
    • (2012) J Arthroplasty , vol.27 , Issue.6 , pp. 1177
    • Hetaimish, B.M.1    Khan, M.M.2    Simunovic, N.3
  • 12
    • 20944451265 scopus 로고    scopus 로고
    • Sagittal laxity in vivo after total knee arthroplasty
    • Ishii Y., Matsuda Y., Ishii R., et al. Sagittal laxity in vivo after total knee arthroplasty. Arch Orthop Trauma Surg 2005, 125:249.
    • (2005) Arch Orthop Trauma Surg , vol.125 , pp. 249
    • Ishii, Y.1    Matsuda, Y.2    Ishii, R.3
  • 13
    • 80052808313 scopus 로고    scopus 로고
    • Instability in primary total knee arthroplasty
    • Del Gaizo D.J., Della Valle C.J. Instability in primary total knee arthroplasty. Orthopedics 2011, 34:519.
    • (2011) Orthopedics , vol.34 , pp. 519
    • Del Gaizo, D.J.1    Della Valle, C.J.2
  • 14
    • 84861342996 scopus 로고    scopus 로고
    • Early revision for isolated internal malrotation of the femoral component in total knee arthroplasty
    • Pietsch M., Hofmann S. Early revision for isolated internal malrotation of the femoral component in total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 2012, 20(6):1057.
    • (2012) Knee Surg Sports Traumatol Arthrosc , vol.20 , Issue.6 , pp. 1057
    • Pietsch, M.1    Hofmann, S.2
  • 15
    • 56749133134 scopus 로고    scopus 로고
    • The effect of femoral component malrotation on patellar biomechanics
    • Kessler O., Patil S., Colwell C.W., et al. The effect of femoral component malrotation on patellar biomechanics. J Biomech 2008, 41:3332.
    • (2008) J Biomech , vol.41 , pp. 3332
    • Kessler, O.1    Patil, S.2    Colwell, C.W.3
  • 16
    • 79955895576 scopus 로고    scopus 로고
    • Biomechanical effects of total knee arthroplasty component malrotation: a computational simulation
    • Thompson J.A., Hast M.W., Granger J.F., et al. Biomechanical effects of total knee arthroplasty component malrotation: a computational simulation. J Orthop Res 2011, 29:969.
    • (2011) J Orthop Res , vol.29 , pp. 969
    • Thompson, J.A.1    Hast, M.W.2    Granger, J.F.3
  • 17
    • 37349078592 scopus 로고    scopus 로고
    • Custom-fit total knee arthroplasty (OtisKnee) results in malalignment
    • Klatt B.A., Goyal N., Austin M.S., et al. Custom-fit total knee arthroplasty (OtisKnee) results in malalignment. J Arthroplasty 2008, 23:26.
    • (2008) J Arthroplasty , vol.23 , pp. 26
    • Klatt, B.A.1    Goyal, N.2    Austin, M.S.3
  • 18
    • 68849090271 scopus 로고    scopus 로고
    • In vivo adduction and reverse axial rotation (external) of the tibial component can be minimized
    • Howell S.M., Hodapp E.E., Kuznik K., et al. In vivo adduction and reverse axial rotation (external) of the tibial component can be minimized. Orthopedics 2009, 32:319.
    • (2009) Orthopedics , vol.32 , pp. 319
    • Howell, S.M.1    Hodapp, E.E.2    Kuznik, K.3
  • 19
    • 84861188144 scopus 로고    scopus 로고
    • Custom-fit total knee arthroplasty: our initial experience in 32 knees
    • Bali K., Walker P., Bruce W. Custom-fit total knee arthroplasty: our initial experience in 32 knees. J Arthroplasty 2012, 27(6):1149.
    • (2012) J Arthroplasty , vol.27 , Issue.6 , pp. 1149
    • Bali, K.1    Walker, P.2    Bruce, W.3
  • 20
    • 83555176169 scopus 로고    scopus 로고
    • The value of patient-matched instrumentation in total knee arthroplasty
    • Noble J.W., Moore C.A., Liu N. The value of patient-matched instrumentation in total knee arthroplasty. J Arthroplasty 2012, 27:153.
    • (2012) J Arthroplasty , vol.27 , pp. 153
    • Noble, J.W.1    Moore, C.A.2    Liu, N.3
  • 21
    • 84856954337 scopus 로고    scopus 로고
    • Are patient-specific cutting blocks cost-effective for total knee arthroplasty?
    • Nunley R.M., Ellison B.S., Ruh E.L., et al. Are patient-specific cutting blocks cost-effective for total knee arthroplasty?. Clin Orthop Relat Res 2012, 470:889.
    • (2012) Clin Orthop Relat Res , vol.470 , pp. 889
    • Nunley, R.M.1    Ellison, B.S.2    Ruh, E.L.3
  • 22
    • 84856958576 scopus 로고    scopus 로고
    • Do patient-specific guides improve coronal alignment in total knee arthroplasty?
    • Nunley R.M., Ellison B.S., Zhu J., et al. Do patient-specific guides improve coronal alignment in total knee arthroplasty?. Clin Orthop Relat Res 2012, 470:895.
    • (2012) Clin Orthop Relat Res , vol.470 , pp. 895
    • Nunley, R.M.1    Ellison, B.S.2    Zhu, J.3
  • 23
    • 57349117781 scopus 로고    scopus 로고
    • The geometry of the tibial plateau and its influence on the biomechanics of the tibiofemoral joint
    • Hashemi J., Chandrashekar N., Gill B., et al. The geometry of the tibial plateau and its influence on the biomechanics of the tibiofemoral joint. J Bone Joint Surg Am 2008, 90:2724.
    • (2008) J Bone Joint Surg Am , vol.90 , pp. 2724
    • Hashemi, J.1    Chandrashekar, N.2    Gill, B.3
  • 24
    • 23044447057 scopus 로고    scopus 로고
    • Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system
    • Jenny J.Y., Boeri C., Picard F., et al. Reproducibility of intra-operative measurement of the mechanical axes of the lower limb during total knee replacement with a non-image-based navigation system. Comput Aided Surg 2004, 9:161.
    • (2004) Comput Aided Surg , vol.9 , pp. 161
    • Jenny, J.Y.1    Boeri, C.2    Picard, F.3
  • 25
    • 78649913833 scopus 로고    scopus 로고
    • The accuracy of acquisition of an imageless computer-assisted system and its implication for knee arthroplasty
    • Lustig S., Fleury C., Goy D., et al. The accuracy of acquisition of an imageless computer-assisted system and its implication for knee arthroplasty. Knee 2012, 18:15.
    • (2012) Knee , vol.18 , pp. 15
    • Lustig, S.1    Fleury, C.2    Goy, D.3
  • 26
    • 33646839119 scopus 로고    scopus 로고
    • Accuracy of a computer-assisted navigation system for total knee replacement
    • Pitto R.P., Graydon A.J., Bradley L., et al. Accuracy of a computer-assisted navigation system for total knee replacement. J Bone Joint Surg Br 2006, 88:601.
    • (2006) J Bone Joint Surg Br , vol.88 , pp. 601
    • Pitto, R.P.1    Graydon, A.J.2    Bradley, L.3
  • 27
    • 79955093511 scopus 로고    scopus 로고
    • The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction
    • Hirschmann M.T., Konala P., Amsler F., et al. The position and orientation of total knee replacement components: a comparison of conventional radiographs, transverse 2D-CT slices and 3D-CT reconstruction. J Bone Joint Surg Br 2011, 93:629.
    • (2011) J Bone Joint Surg Br , vol.93 , pp. 629
    • Hirschmann, M.T.1    Konala, P.2    Amsler, F.3
  • 28
    • 68549141006 scopus 로고    scopus 로고
    • How precise can bony landmarks be determined on a CT scan of the knee?
    • Victor J., Van Doninck D., Labey L., et al. How precise can bony landmarks be determined on a CT scan of the knee?. Knee 2009, 16:358.
    • (2009) Knee , vol.16 , pp. 358
    • Victor, J.1    Van Doninck, D.2    Labey, L.3
  • 29
    • 84866712897 scopus 로고    scopus 로고
    • Are MRI-based, patient matched cutting jigs as accurate as the tibial guides?
    • Conteduca F., Iorio R., Mazza D., et al. Are MRI-based, patient matched cutting jigs as accurate as the tibial guides?. Int Orthop 2012, 36(8):1589.
    • (2012) Int Orthop , vol.36 , Issue.8 , pp. 1589
    • Conteduca, F.1    Iorio, R.2    Mazza, D.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.