Volumn 17, Issue 2, 2012, Pages 80-88

Oscillometric sphygmomanometers: A critical appraisal of current technology

Author keywords

Accuracy; Blood pressure measurement; Oscillometric; Sphygmomanometer; Validation

Indexed keywords

ACCURACY; ALGORITHM; AMPLIFIER; ARTERIAL STIFFNESS; AUSCULTATION; BLOOD PRESSURE CUFF; BLOOD PRESSURE MEASUREMENT; HUMAN; OSCILLOMETRIC SPHYGMOMANOMETER; OSCILLOMETRY; PRIORITY JOURNAL; REVIEW; SIGNAL PROCESSING; SIMULATOR; SPHYGMOMANOMETER; THEORETICAL MODEL; TRANSDUCER; VALIDATION PROCESS;

EID: 84860837397     PISSN: 13595237     EISSN: 14735725     Source Type: Journal    
DOI: 10.1097/MBP.0b013e32835026b0     Document Type: Review
Times cited : (39)

References (66)
  • 1
    • 0025041825 scopus 로고
    • Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure
    • DOI 10.1097/00004872-199007000-00005
    • O'Brien E, Mee F, Atkins N, O'Malley K. Inaccuracy of seven popular sphygmomanometers for home measurement of blood pressure. J Hypertens 1990; 8:621-634. (Pubitemid 20248493)
    • (1990) Journal of Hypertension , vol.8 , Issue.7 , pp. 621-634
    • O'Brien, E.1    Mee, F.2    Atkins, N.3    O'Malley, K.4
  • 2
    • 0027479012 scopus 로고
    • National standard for measurement of resting and ambulatory blood pressures with automated sphygmomanometers
    • White WB, Berson AS, Robbins C, Jamieson MJ, Prisant LM, Roccella E, et al. National standard for measurement of resting and ambulatory blood pressure with automated sphygmomanometers. Hypertension 1993; 21:504-509. (Pubitemid 23095165)
    • (1993) Hypertension , vol.21 , Issue.4 , pp. 504-509
    • White, W.B.1    Berson, A.S.2    Robbins, C.3    Jamieson, M.J.4    Prisant, L.M.5    Roccella, E.6    Sheps, S.G.7
  • 3
    • 0001849046 scopus 로고
    • The British Hypertension Society protocol for the evaluation of blood pressure measuring devices
    • O'Brien E, Petrie J, Littler R, De Sweit M, Padfield PL, Altman D, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11 (Suppl 2):S43-S63.
    • (1993) J Hypertens , vol.11 , Issue.SUPPL. 2
    • O'Brien, E.1    Petrie, J.2    Littler, R.3    De Sweit, M.4    Padfield, P.L.5    Altman, D.6
  • 4
    • 0035527552 scopus 로고    scopus 로고
    • Oscillometric blood pressure measurement: Progress and problems
    • DOI 10.1097/00126097-200112000-00004
    • van Montfrans GA. Oscillometric blood pressure measurement: progress and problems. Blood Press Monit 2001; 6:287-290. (Pubitemid 34194881)
    • (2001) Blood Pressure Monitoring , vol.6 , Issue.6 , pp. 287-290
    • Van Montfrans, G.A.1
  • 5
    • 13444255966 scopus 로고    scopus 로고
    • Recommendations for blood pressure measurement in humans and experimental animals: Part 1: Blood pressure measurement in humans - A statement for professionals from the Subcommittee of Professional and Public Education of the American Heart Association Council on high blood pressure research
    • DOI 10.1161/01.CIR.0000154900.76284.F6
    • Pickering TG, Hall JE, Appel LJ, Falkner BE, Graves J, Hill MN, et al. Recommendations for blood pressure measurement in humans and experimental animals part 1: blood pressure measurement in humans. A statement for professionals from the subcommittee of professional and public education of the American Heart Association Council on high blood pressure research. Circulation 2005; 111:697-716. (Pubitemid 40216499)
    • (2005) Circulation , vol.111 , Issue.5 , pp. 697-716
    • Pickering, T.G.1    Hall, J.E.2    Appel, L.J.3    Falkner, B.E.4    Graves, J.5    Hill, M.N.6    Jones, D.W.7    Kurtz, T.8    Sheps, S.G.9    Roccella, E.J.10
  • 6
    • 70749154760 scopus 로고    scopus 로고
    • Unreliable oscillometric blood pressure measurement: Prevalence, repeatability and characteristics of the phenomenon
    • Stergiou GS, Lourida P, Tzamouranis D, Baibas NM. Unreliable oscillometric blood pressure measurement: prevalence, repeatability and characteristics of the phenomenon. J Human Hypertens 2009; 23:794-800.
    • (2009) J Human Hypertens , vol.23 , pp. 794-800
    • Stergiou, G.S.1    Lourida, P.2    Tzamouranis, D.3    Baibas, N.M.4
  • 7
    • 76449109838 scopus 로고    scopus 로고
    • European Society of Hypertension International Protocol for the validation of blood pressure monitors: A critical review of its application and rationale for revision
    • Stergiou GS, Karpettas N, Atkins N, O'Brien E. European Society of Hypertension International Protocol for the validation of blood pressure monitors: A critical review of its application and rationale for revision. Blood Press Monit 2010; 15:39-48.
    • (2010) Blood Press Monit , vol.15 , pp. 39-48
    • Stergiou, G.S.1    Karpettas, N.2    Atkins, N.3    O'Brien, E.4
  • 8
    • 80053017818 scopus 로고    scopus 로고
    • Errors of oscillometric blood pressure measurement as predicted by simulation
    • Raamat R, Talts J, Jagomägi K, Kivastik J. Errors of oscillometric blood pressure measurement as predicted by simulation. Blood Press Monit 2011; 16:238-245.
    • (2011) Blood Press Monit , vol.16 , pp. 238-245
    • Raamat, R.1    Talts, J.2    Jagomägi, K.3    Kivastik, J.4
  • 9
    • 0031874834 scopus 로고    scopus 로고
    • Determinants of the limits of agreement between the sphygmomanometer and the SpaceLabs 90207 device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patients
    • DOI 10.1097/00004872-199816080-00007
    • van Ittersum FJ, Wijering RM, Lambert J, Donker AJ, Stehouwer CD. Determinants of the limits of agreement between the sphygmomanometer and the Spacelabs 90207 device for blood pressure measurement in healthy volunteers and insulin-dependent diabetic patients. J Hypertens 1998; 16:1125-1130. (Pubitemid 28334013)
    • (1998) Journal of Hypertension , vol.16 , Issue.8 , pp. 1125-1130
    • Van Ittersum, F.J.1    Wijering, R.M.J.2    Lambert, J.3    Donker, A.J.M.4    Stehouwer, C.D.A.5
  • 11
    • 0029956336 scopus 로고    scopus 로고
    • A comparison of non-invasive methods of blood pressure measurement in normotensive and hypertensive pregnant women
    • DOI 10.1016/S0959-289X(96)80026-4
    • Arfeen ZU, Maran NJ, Simon EJ, McClure JH. A comparison of non-invasive methods of blood pressure measurement in normotensive and hypertensive pregnant women. Int J Obstet Anesth 1996; 5:168-171. (Pubitemid 26229618)
    • (1996) International Journal of Obstetric Anesthesia , vol.5 , Issue.3 , pp. 168-171
    • Arfeen, Z.U.1    Maran, N.J.2    Simon, E.J.3    McClure, J.H.4
  • 13
    • 0035061763 scopus 로고    scopus 로고
    • Blood pressure monitoring during pregnancy
    • Feldman DB. Blood pressure monitoring during pregnancy. Blood Press Monit 2001; 6:1-7.
    • (2001) Blood Press Monit , vol.6 , pp. 1-7
    • Feldman, D.B.1
  • 14
    • 0030020672 scopus 로고    scopus 로고
    • Oscillometric blood pressure measurements by different devices are not interchangeable
    • DOI 10.1097/00000539-199602000-00029
    • Kaufmann MA, Pargger H, Drop LJ. Oscillometric blood pressure measurements by different devices are not interchangeable. Anesth Anal 1996; 82:377-381. (Pubitemid 26055031)
    • (1996) Anesthesia and Analgesia , vol.82 , Issue.2 , pp. 377-381
    • Kaufmann, M.A.1    Pargger, H.2    Drop, L.J.3
  • 15
    • 0033936833 scopus 로고    scopus 로고
    • Can simulators evaluate systematic differences between oscillometric non-invasive blood-pressure monitors?
    • Amoore JN, Scott DHT. Can simulators evaluate systematic differences between oscillometric NIBP monitors? Blood Press Monit 2000; 5:81-89. (Pubitemid 30426705)
    • (2000) Blood Pressure Monitoring , vol.5 , Issue.2 , pp. 81-89
    • Amoore, J.N.1    Scott, D.H.T.2
  • 18
    • 0020307159 scopus 로고
    • Characterization of the oscillometric method for measuring indirect blood pressure
    • Geddes LA, Voelz MH, Combs C, Reiner D, Babbs CF. Characterisation of the oscillometric method for measuring indirect blood pressure. Ann Biomed Eng 1982; 10:271-280. (Pubitemid 13088565)
    • (1982) Annals of Biomedical Engineering , vol.10 , Issue.6 , pp. 271-280
    • Geddes, L.A.1    Voelz, M.2    Combs, C.3
  • 19
    • 0028243132 scopus 로고
    • Theory of the oscillometric maximum and the systolic and diastolic detection ratios
    • DOI 10.1007/BF02368225
    • Drzewiecki GM, Hood R, Apple H. Theory of the oscillometric maximum and the systolic and diastolic detection ratios. Ann Biomed Eng 1994; 22:88-96. (Pubitemid 24189259)
    • (1994) Annals of Biomedical Engineering , vol.22 , Issue.1 , pp. 88-96
    • Drzewiecki, G.1    Hood, R.2    Apple, H.3
  • 20
    • 0029366530 scopus 로고
    • Automated noninvasive blood pressure measurement
    • Ng K-G. Automated noninvasive blood pressure measurement. Intensive Care World 1995; 12:89-90, 92-102.
    • (1995) Intensive Care World , vol.12 , Issue.89-90 , pp. 92-102
    • Ng, K.-G.1
  • 21
    • 33745922093 scopus 로고    scopus 로고
    • Oscillometric noninvasive blood pressure measurement devices and simulators
    • PII 0000466920060400000026
    • Amoore JN, Murray IC. Oscillometric non-invasive blood pressure measurement devices and simulators. J Clin Eng 2006; 31:85-95. (Pubitemid 44046441)
    • (2006) Journal of Clinical Engineering , vol.31 , Issue.2 , pp. 85-95
    • Amoore, J.N.1    Murray, I.C.2
  • 22
    • 0042080747 scopus 로고
    • A simple and accurate form of sphygmomanometer or arterial pressure gauge contrived for clinical use
    • Hill L, Barnard H. A simple and accurate form of sphygmomanometer or arterial pressure gauge contrived for clinical use. BMJ 1897; 2:904.
    • (1897) BMJ , vol.2 , pp. 904
    • Hill, L.1    Barnard, H.2
  • 23
    • 77955713153 scopus 로고    scopus 로고
    • Instruments and methods for calibration of oscillometric blood pressure measurement devices
    • Balestrieri E, Rapuano S. Instruments and methods for calibration of oscillometric blood pressure measurement devices. IEEE Trans Instrum Meas 2010; 59:2391-2404.
    • (2010) IEEE Trans Instrum Meas , vol.59 , pp. 2391-2404
    • Balestrieri, E.1    Rapuano, S.2
  • 24
    • 0029831681 scopus 로고    scopus 로고
    • Review: A century of confusion; which bladder for accurate blood pressure measurement?
    • O'Brien E. A century of confusion: which bladder for accurate blood pressure measurement? J Human Hypertens 1996; 10:565-572. (Pubitemid 26402311)
    • (1996) Journal of Human Hypertension , vol.10 , Issue.9 , pp. 565-572
    • O'Brien, E.1
  • 25
    • 0030221353 scopus 로고    scopus 로고
    • A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement
    • DOI 10.1109/10.508540
    • Ursino M, Cristalli C. A mathematical study of some biomechanical factors affecting the oscillometric blood pressure measurement. IEEE Trans Biomed Eng 1996; 43:761-778. (Pubitemid 26284797)
    • (1996) IEEE Transactions on Biomedical Engineering , vol.43 , Issue.8 , pp. 761-778
    • Ursino, M.1    Cristalli, C.2
  • 26
    • 0035059859 scopus 로고    scopus 로고
    • Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives
    • DOI 10.1097/00126097-200102000-00003
    • Graves JW. Prevalence of blood pressure cuff sizes in a referral practice of 430 consecutive adult hypertensives. Blood Press Monit 2001; 6:17-20. (Pubitemid 32269926)
    • (2001) Blood Pressure Monitoring , vol.6 , Issue.1 , pp. 17-20
    • Graves, J.W.1
  • 27
    • 0033995223 scopus 로고    scopus 로고
    • The mercury sphygmomanometer should be abandoned before it is proscribed
    • DOI 10.1038/sj.jhh.1000932
    • Markandu ND, Whitcher F, Arnold A, Carney C. The mercury sphygmomanometer should be abandoned before it is proscribed. J Human Hypertension 2000; 14:31-36. (Pubitemid 30113628)
    • (2000) Journal of Human Hypertension , vol.14 , Issue.1 , pp. 31-36
    • Markandu, N.D.1    Whitcher, F.2    Arnold, A.3    Carney, C.4
  • 28
    • 23944491162 scopus 로고    scopus 로고
    • Design flaw in Walgreen's blood pressure cuff for home measurement
    • Reynolds RD. Design flaw in Walgreen's blood pressure cuff for home measurement. Blood Press Monit 2005; 10:227-228.
    • (2005) Blood Press Monit , vol.10 , pp. 227-228
    • Reynolds, R.D.1
  • 29
    • 51249083245 scopus 로고    scopus 로고
    • The trouble with blood pressure cuffs
    • Bellamy JE, Pugh H, Sanders DJ. The trouble with blood pressure cuffs. BMJ 2008; 337:a431.
    • (2008) BMJ , vol.337
    • Bellamy, J.E.1    Pugh, H.2    Sanders, D.J.3
  • 30
    • 1542648293 scopus 로고    scopus 로고
    • Accuracy and applicability of the Terumo ES-H55 double-cuff sphygmomanometer for hospital use
    • DOI 10.1097/01.mbp.0000097242.01232.8c
    • Tochikubo O, Nishijima K, Ohshige K, Kimura K. Accuracy and applicability of the Terumo ES-H55 double-cuff sphygmomanometer for hospital use. Blood Press Monit 2003; 8:203-209. (Pubitemid 41623179)
    • (2003) Blood Pressure Monitoring , vol.8 , Issue.5 , pp. 203-209
    • Tochikubo, O.1    Nishijima, K.2    Ohshige, K.3    Kimura, K.4
  • 31
    • 70350393596 scopus 로고    scopus 로고
    • Blood pressure measurement: Arm circumference and cuff size availability
    • Veiga EV, Arcuri EA, Cloutier L, Santos JL. Blood pressure measurement: Arm circumference and cuff size availability. Rev Lat Am Enfermagem 2009; 17:455-461.
    • (2009) Rev Lat Am Enfermagem , vol.17 , pp. 455-461
    • Veiga, E.V.1    Arcuri, E.A.2    Cloutier, L.3    Santos, J.L.4
  • 32
    • 77955311383 scopus 로고    scopus 로고
    • Obesity, hypertension and home sphygmomanometer cuffs
    • Akpolat T. Obesity, hypertension and home sphygmomanometer cuffs. Eur J Intern Med 2010; 21:338-340.
    • (2010) Eur J Intern Med , vol.21 , pp. 338-340
    • Akpolat, T.1
  • 35
    • 0346207561 scopus 로고    scopus 로고
    • American National Standard ANSI/AAMI, SP10. American national standard. ANSI/AAMI SP 10-2002, Arlington, VA
    • American National Standard ANSI/AAMI, SP10. 2002 American national standard. Manual, electronic or automated sphygmomanometers. ANSI/AAMI SP 10-2002, Arlington, VA.
    • (2002) Manual, Electronic or Automated Sphygmomanometers
  • 37
    • 40949161444 scopus 로고    scopus 로고
    • Lax sphygmomanometer standard causes overdetection and underdetection of hypertension: A computer simulation study
    • DOI 10.1097/MBP.0b013e3282f7691c, PII 0012609720080400000005
    • Turner MJ, van Schalkwyk JM, Irwig L. Lax sphygmomanometer standard causes overdetection and underdetection of hypertension: A computer simulation study. Blood Press Monit 2008; 13:91-99. (Pubitemid 351416909)
    • (2008) Blood Pressure Monitoring , vol.13 , Issue.2 , pp. 91-99
    • Turner, M.J.1    Van Schalkwyk, J.M.2    Irwig, L.3
  • 38
    • 14644439181 scopus 로고    scopus 로고
    • Do SpaceLabs ambulatory non-invasive blood pressure recorders measure blood pressure consistently over several years use?
    • DOI 10.1097/00126097-200502000-00009
    • Amoore JN, Dewar D, Gough K, Padfield PL. Do Spacelabs ambulatory NIBP recorders measure blood pressure consistently over several years use? Blood Press Monit 2005; 10:51-56. (Pubitemid 40316152)
    • (2005) Blood Pressure Monitoring , vol.10 , Issue.1 , pp. 51-56
    • Amoore, J.N.1    Dewar, D.2    Gough, K.3    Padfield, P.L.4
  • 39
    • 33748282746 scopus 로고    scopus 로고
    • Extracting oscillometric pulses from the cuff pressure: Does it affect the pressures determined by oscillometric blood pressure monitors?
    • DOI 10.1097/01.mbp.0000217999.04592.7c, PII 0012609720061000000005
    • Amoore JN. Extracting oscillometric pulses from the cuff pressure: does it affect the pressures determined by oscillometric blood pressure monitors? Blood Press Monit 2006; 11:269-279. (Pubitemid 44319046)
    • (2006) Blood Pressure Monitoring , vol.11 , Issue.5 , pp. 269-279
    • Amoore, J.N.1
  • 41
    • 0026886469 scopus 로고
    • Cardiac cycle phase uncertainty: Another source of error in indirect blood pressure measurement
    • Hatsel CP. Cardiac cycle phase uncertainty: Another source of error in indirect blood pressure measurement. J Med Eng Technol 1992; 16: 157-158.
    • (1992) J Med Eng Technol , vol.16 , pp. 157-158
    • Hatsel, C.P.1
  • 42
    • 37349087844 scopus 로고    scopus 로고
    • Automatic blood pressure measurement: The oscillometric waveform shape is a potential contributor to differences between oscillometric and auscultatory pressure measurements
    • DOI 10.1097/HJH.0b013e3282f06ec7, PII 0000487220080100000009
    • Amoore JN, Lemesre Y, Murray IC, Mieke S, King ST, Smith FE, Murray A. Automatic blood pressure measurement: The oscillometric waveform shape is a potential contributor to the differences between oscillometric and auscultatory pressure measurements. J Hypertens 2008; 26:35-43. (Pubitemid 350294847)
    • (2008) Journal of Hypertension , vol.26 , Issue.1 , pp. 35-43
    • Amoore, J.N.1    Lemesre, Y.2    Murray, I.C.3    Mieke, S.4    King, S.T.5    Smith, F.E.6    Murray, A.7
  • 43
    • 0036528828 scopus 로고    scopus 로고
    • Determination of accuracy in neonates for non-invasive blood pressure device using an improved algorithm
    • DOI 10.1097/00126097-200204000-00006
    • Nelson RM, Stebor AD, Groh CM, Timoney PM, Theobold KS, Friedman BA. Determination of accuracy in neonates for non-invasive blood pressure using an improved algorithm. Blood Press Monit 2002; 7:123-129. (Pubitemid 34913421)
    • (2002) Blood Pressure Monitoring , vol.7 , Issue.2 , pp. 123-129
    • Nelson, R.M.1    Stebor, A.D.2    Groh, C.M.3    Timoney, P.M.4    Theobald, K.S.5    Friedman, B.A.6
  • 44
    • 3042791251 scopus 로고    scopus 로고
    • Faster oscillometric manometry does not sacrifice the accuracy of blood pressure determination
    • DOI 10.1097/01.mbp.0000130430.78823.42
    • Sugimachi M, Okamoto H, Hoka S, Sunagawa K. Faster oscillometric manometry does not sacrifice the accuracy of blood pressure determination. Blood Press Monit 2004; 9:135-141. (Pubitemid 38888054)
    • (2004) Blood Pressure Monitoring , vol.9 , Issue.3 , pp. 135-141
    • Sugimachi, M.1    Okamoto, H.2    Hoka, S.3    Sunagawa, K.4
  • 45
    • 20044380065 scopus 로고    scopus 로고
    • Differences in oscillometric non-invasive blood pressure measurements recorded by different revisions of the Philips Component Monitoring System
    • DOI 10.1097/01.mbp.0000168399.64296.0f
    • Murray I, Amoore JN, Scott DHT. Differences in oscillometric non-invasive blood pressure measurements recorded by different revisions of the Philips Component Monitoring System. Blood Press Monit 2005; 10:215-222. (Pubitemid 41192343)
    • (2005) Blood Pressure Monitoring , vol.10 , Issue.4 , pp. 215-222
    • Murray, I.C.1    Amoore, J.N.2    Scott, D.H.T.3
  • 46
    • 1542344282 scopus 로고
    • Standard algorithm of blood-pressure measurement by the oscillometric method. (Letter
    • Sapinski A. Standard algorithm of blood-pressure measurement by the oscillometric method. (Letter). Med Biol Eng Comput 1992; 30:671.
    • (1992) Med Biol Eng Comput , vol.30 , pp. 671
    • Sapinski, A.1
  • 47
    • 0034889766 scopus 로고    scopus 로고
    • Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor
    • DOI 10.1097/00126097-200106000-00008
    • Wright JM, Mattu GS, Perry TL Jr, Gelferc ME, Strange KD, Zorn A, Chen Y. Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitor. Blood Press Monit 2001; 6:161-165. (Pubitemid 32744590)
    • (2001) Blood Pressure Monitoring , vol.6 , Issue.3 , pp. 161-165
    • Wright, J.M.1    Mattu, G.S.2    Perry Jr., T.L.3    Gelferc, M.E.4    Strange, K.D.5    Zorn, A.6    Chen, Y.7
  • 48
    • 84860841378 scopus 로고    scopus 로고
    • Anon. SunTech Medical (507 Airport Blvd., Suite 117, Morrisville, NC 27560, USA). Reference document 82-0048-00 Rev A. Available at: [Accessed 13 December]
    • Anon. The Rationale for Clinical Monitors and Validation of the Oscillometric Method. SunTech Medical (507 Airport Blvd., Suite 117, Morrisville, NC 27560, USA). Reference document 82-0048-00 Rev A. Available at: http://www.suntechmed.com. [Accessed 13 December 2011].
    • (2011) The Rationale for Clinical Monitors and Validation of the Oscillometric Method
  • 49
    • 0036327796 scopus 로고    scopus 로고
    • New algorithm for oscillometric noninvasive automatic arterial pressure measurement in patients with atrial fibrillation
    • Sugimachi M, Sunagawa K, Okamoto H, Hoka S. New algorithm for oscillometric noninvasive automatic arterial pressure measurement in patients with atrial fibrillation. Masui 2002; 51:784-790, Japanese. (Pubitemid 34832330)
    • (2002) Japanese Journal of Anesthesiology , vol.51 , Issue.7 , pp. 784-790
    • Sugimachi, M.1    Sunagawa, K.2    Okamoto, H.3    Hoka, S.4
  • 50
    • 27144452976 scopus 로고    scopus 로고
    • Automated non-invasive blood pressure devices: Are they suitable for use?
    • DOI 10.1097/01.mbp.0000173487.82766.be
    • Sims AJ, Menes JA, Bousfield DR, Reay CA, Murray A. Automated noninvasive blood pressure devices: Are they suitable for use? Blood Press Monit 2005; 10:275-281. (Pubitemid 41502886)
    • (2005) Blood Pressure Monitoring , vol.10 , Issue.5 , pp. 275-281
    • Sims, A.J.1    Menes, J.A.2    Bousfield, D.R.3    Reay, C.A.4    Murray, A.5
  • 51
    • 0035799041 scopus 로고    scopus 로고
    • Blood pressure measuring devices: Recommendations of the European Society of Hypertension
    • O'Brien E, Waeber B, Parati G, Staessen J, Myers MG. Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ 2001; 322:531-536. (Pubitemid 32230920)
    • (2001) British Medical Journal , vol.322 , Issue.7285 , pp. 531-536
    • O'Brien, E.1    Waeber, B.2    Parati, G.3    Staessen, J.4    Myers, M.G.5
  • 52
    • 84860839103 scopus 로고    scopus 로고
    • dablsEducational Trust. Available at. [Accessed 19 May]
    • dablsEducational Trust. Available at http://www.dableducational.org. [Accessed 19 May 2011].
    • (2011)
  • 53
    • 84860841377 scopus 로고    scopus 로고
    • British Hypertension Society. Available at. [Accessed 19 May]
    • British Hypertension Society. Available at http://www.bhsoc.org. [Accessed 19 May 2011].
    • (2011)
  • 54
    • 76449087188 scopus 로고    scopus 로고
    • European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults
    • O'Brien E, Atkins N, Stergiou G, Karpettas N, Parati G, Asmar R, et al. European Society of Hypertension International Protocol revision 2010 for the validation of blood pressure measuring devices in adults. Blood Press Monit 2010; 15:23-38.
    • (2010) Blood Press Monit , vol.15 , pp. 23-38
    • O'Brien, E.1    Atkins, N.2    Stergiou, G.3    Karpettas, N.4    Parati, G.5    Asmar, R.6
  • 55
    • 0035123774 scopus 로고    scopus 로고
    • Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor
    • Jones CR, Taylor K, Poston L, Shennan AH. Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor. J Hum Hypertens 2001; 15:191-195. (Pubitemid 32181485)
    • (2001) Journal of Human Hypertension , vol.15 , Issue.3 , pp. 191-195
    • Jones, C.R.1    Taylor, K.2    Poston, L.3    Shennan, A.H.4
  • 56
    • 0036592609 scopus 로고    scopus 로고
    • Accuracy of the Welch Allyn Vital Signs Monitor 52000 automatic blood pressure measuring device according to a modified British Hypertension Society Protocol
    • DOI 10.1097/00126097-200206000-00007
    • Braam RL, de Maat C, Thien TH. Accuracy of theWelch Allyn Vital Signs Monitor 52000 automatic blood pressure measuring device according to a modified British Hypertension Society protocol. Blood Press Monit 2002; 7:185-189. (Pubitemid 35001783)
    • (2002) Blood Pressure Monitoring , vol.7 , Issue.3 , pp. 185-189
    • Braam, R.L.1    De Maat, C.2    Thien, T.3
  • 59
    • 0031434311 scopus 로고    scopus 로고
    • Substitution of simulators for human subjects
    • Mieke S. Substitution of simulators for human subjects. Blood Press Monit 1997; 2:251-256. (Pubitemid 28000079)
    • (1997) Blood Pressure Monitoring , vol.2 , Issue.5 , pp. 251-256
    • Mieke, S.1
  • 60
    • 0031464019 scopus 로고    scopus 로고
    • Considerations for simulation-based evaluation of noninvasive blood pressure monitors
    • Ng K-G. Considerations for simulation-based evaluation of noninvasive blood pressure monitors. Blood Press Monit 1997; 2:256-268.
    • (1997) Blood Press Monit , vol.2 , pp. 256-268
    • Ng, K.-G.1
  • 61
    • 33744508715 scopus 로고    scopus 로고
    • Can a simulator that regenerates physiological waveforms evaluate oscillometric non-invasive blood pressure devices?
    • DOI 10.1097/01.mbp.0000200482.72410.e2, PII 0012609720060400000004
    • Amoore JN, Vacher E, Murray IC, Mieke S, King ST, Smith FE, Murray A. Can a simulator that regenerates physiological waveforms evaluate oscillometric non-invasive blood pressure devices? Blood Press Monit 2006; 11:63-67. (Pubitemid 44090018)
    • (2006) Blood Pressure Monitoring , vol.11 , Issue.2 , pp. 63-67
    • Amoore, J.N.1    Vacher, E.2    Murray, I.C.3    Mieke, S.4    King, S.T.5    Smith, F.E.6    Murray, A.7
  • 62
    • 0033918279 scopus 로고    scopus 로고
    • A simulation study of the consistency of oscillometric blood pressure measurements with and without artefacts
    • Amoore JN. A simulation study of the consistency of oscillometric blood pressure measurements with and without artefacts. Blood Press Monit 2000; 5:69-79. (Pubitemid 30426704)
    • (2000) Blood Pressure Monitoring , vol.5 , Issue.2 , pp. 69-79
    • Amoore, J.N.1
  • 63
    • 0027397159 scopus 로고
    • ZUR MESSSICHERHEIT NICHTINVASIVER OSZILLOMETRISCHER BLUTDRUCKMESSGERATE
    • Mieke S, Grob H, Ulbrich M, Papadopoulos G, Frucht U. On the performance of non-invasive oscillometric sphygmomanometers. Anaesthesist 1993; 42:38-43, [in German]. (Pubitemid 23037518)
    • (1993) Anaesthesist , vol.42 , Issue.1 , pp. 38-43
    • Mieke, S.1    Gross, H.2    Ulbrich, M.3    Papadopoulos, G.4    Frucht, U.5
  • 64
    • 0027298076 scopus 로고
    • Assessment of oscillometric non-invasive blood pressure monitors using the Dynatech Nevada CuffLink analyser
    • Amoore JN. Assessment of oscillometric non-invasive blood pressure monitors using the Dynatech Nevada CuffLink analyser. J Med Eng Technol 1993; 17:25-31. (Pubitemid 23165077)
    • (1993) Journal of Medical Engineering and Technology , vol.17 , Issue.1 , pp. 25-31
    • Amoore, J.N.1
  • 65
    • 18844415051 scopus 로고    scopus 로고
    • Oscillometric blood pressure devices and simulators: Measurements of repeatability and differences between models
    • DOI 10.1080/03091900500067983
    • Sims AJ, Reay CA, Bousfield DR, Menes JA, Murray A. Oscillometric blood pressure devices and simulators: measurements of repeatability and differences between models. J Med Eng Technol 2005; 29:112-118. (Pubitemid 40695371)
    • (2005) Journal of Medical Engineering and Technology , vol.29 , Issue.3 , pp. 112-118
    • Sims, A.J.1    Reay, C.A.2    Bousfield, D.R.3    Menes, J.A.4    Murray, A.5
  • 66
    • 0030946816 scopus 로고    scopus 로고
    • Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator
    • Amoore JN, Geake WB. Evaluation of the Critikon 8100 and Spacelabs 90207 non-invasive blood pressure monitors using a test simulator. J Human Hypertens 1997; 11:163-169. (Pubitemid 27234676)
    • (1997) Journal of Human Hypertension , vol.11 , Issue.3 , pp. 163-169
    • Amoore, J.N.1    Geake, W.B.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.