Automated non-invasive blood pressure devices: Are they suitable for use?

Author keywords

Algorithms; Blood pressure; Blood pressure determination; Blood pressure monitors; Healthcare market; Sphygmomanometers

Indexed keywords

AMBULATORY MONITORING; ARM; ARTICLE; AUTOMATION; BLOOD PRESSURE; BLOOD PRESSURE MEASUREMENT; DEFIBRILLATOR; DEVICE; MONITOR; PRIORITY JOURNAL; QUESTIONNAIRE; SPHYGMOMANOMETER; VALIDATION PROCESS; WRIST;

EID: 27144452976     PISSN: 13595237     EISSN: None     Source Type: Journal    
DOI: 10.1097/01.mbp.0000173487.82766.be     Document Type: Article
Times cited : (18)

References (37)
  • 1
    • 0035857994 scopus 로고    scopus 로고
    • Measurement of blood pressure: An evidence based review
    • McAlister FA, Straus SE. Measurement of blood pressure: an evidence based review. BMJ 2001; 322:908-911.
    • (2001) BMJ , vol.322 , pp. 908-911
    • McAlister, F.A.1    Straus, S.E.2
  • 2
    • 0035962451 scopus 로고    scopus 로고
    • Blood pressure measurement. Part II - Conventional sphygmomanometry: Technique of auscultatory pressure measurement
    • Beevers G, Lip GYH, O'Brien E. Blood pressure measurement. Part II - conventional sphygmomanometry: technique of auscultatory pressure measurement. BMJ 2001; 322:1043-1047.
    • (2001) BMJ , vol.322 , pp. 1043-1047
    • Beevers, G.1    Lip, G.Y.H.2    O'Brien, E.3
  • 5
    • 27144498194 scopus 로고    scopus 로고
    • Blood pressure measurement
    • Izzo JL Jr, Black H (eds). Philadelphia, USA: Lippincott Williams & Wilkins
    • Grim CM, Grim CE. Blood pressure measurement. In: Izzo JL Jr, Black H (eds). Hypertension primer. 3rd edn. Philadelphia, USA: Lippincott Williams & Wilkins; 2003. p. 321.
    • (2003) Hypertension Primer. 3rd Edn. , pp. 321
    • Grim, C.M.1    Grim, C.E.2
  • 6
    • 0036854179 scopus 로고    scopus 로고
    • Use of automated home blood pressure monitoring in pregnancy: Is it safe?
    • Lo C, Taylor RS, Gamble G, McCowan L, North RA. Use of automated home blood pressure monitoring in pregnancy: is it safe? Am J Obstet Gynecol 2002; 187:1321-1328.
    • (2002) Am J Obstet Gynecol , vol.187 , pp. 1321-1328
    • Lo, C.1    Taylor, R.S.2    Gamble, G.3    McCowan, L.4    North, R.A.5
  • 7
    • 0033770477 scopus 로고    scopus 로고
    • Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer
    • van Popele NM, Bos WJ, de Beer NA, van Der Kuip DA, Hofman A, Grobbee DE, et al. Arterial stiffness as underlying mechanism of disagreement between an oscillometric blood pressure monitor and a sphygmomanometer. Hypertension 2000; 36:484-488.
    • (2000) Hypertension , vol.36 , pp. 484-488
    • Van Popele, N.M.1    Bos, W.J.2    De Beer, N.A.3    Van Der Kuip, D.A.4    Hofman, A.5    Grobbee, D.E.6
  • 8
    • 0022839965 scopus 로고
    • Reliability of automatic oscillometric monitoring of arterial pressure. Effect of hypotension and arrhythmias
    • Caramella JP, Desmonts JM. Reliability of automatic oscillometric monitoring of arterial pressure. Effect of hypotension and arrhythmias. Arch Mal Coeur Vaiss 1986; 79:1794-1799.
    • (1986) Arch Mal Coeur Vaiss , vol.79 , pp. 1794-1799
    • Caramella, J.P.1    Desmonts, J.M.2
  • 9
    • 12444277817 scopus 로고    scopus 로고
    • European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement
    • O'Brien E, Asmar R, Beilin L, Imai Y, Maillon JM, Mancia G, et al. European Society of Hypertension recommendations for conventional, ambulatory and home blood pressure measurement. J Hypertens 2003; 21:821-848.
    • (2003) J Hypertens , vol.21 , pp. 821-848
    • O'Brien, E.1    Asmar, R.2    Beilin, L.3    Imai, Y.4    Maillon, J.M.5    Mancia, G.6
  • 10
    • 0030659535 scopus 로고    scopus 로고
    • Validation of the Omron R3 blood pressure self measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation
    • Eckert S, Gleichmann U, Zagorski O, Klapp A. Validation of the Omron R3 blood pressure self measuring device through simultaneous comparative invasive measurements according to protocol 58130 of the German Institute for Validation. Blood Press Monit 1997; 2:189-192.
    • (1997) Blood Press Monit , vol.2 , pp. 189-192
    • Eckert, S.1    Gleichmann, U.2    Zagorski, O.3    Klapp, A.4
  • 11
    • 0027218269 scopus 로고
    • The British Hypertension Society protocol for the evaluation of blood pressure measuring devices
    • O'Brien E, Petrie J, Littler WA, de Swiet M, Padfield PL, Altman DG, et al. The British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens 1993; 11(Suppl. 2):S43-S63.
    • (1993) J Hypertens , vol.11 , Issue.SUPPL. 2
    • O'Brien, E.1    Petrie, J.2    Littler, W.A.3    De Swiet, M.4    Padfield, P.L.5    Altman, D.G.6
  • 12
    • 0035992025 scopus 로고    scopus 로고
    • Working group on blood pressure monitoring of the European Society of Hypertension international protocol for validation of blood pressure measuring devices in adults
    • O'Brien E, Pickering T, Asmar R, Myers M, Parati G, Staessen J, et al. Working group on blood pressure monitoring of the European Society of Hypertension international protocol for validation of blood pressure measuring devices in adults. Blood Press Monit 2002; 7:3-17.
    • (2002) Blood Press Monit , vol.7 , pp. 3-17
    • O'Brien, E.1    Pickering, T.2    Asmar, R.3    Myers, M.4    Parati, G.5    Staessen, J.6
  • 13
    • 0003556412 scopus 로고
    • Arlington, Virginia, USA: Association for the Advancement of Medical Instrumentation. American National Standard ANSI/AAMI SP10-1992
    • Association for the Advancement of Medical Instrumentation. Electronic or automated sphygmomanometers. Arlington, Virginia, USA: Association for the Advancement of Medical Instrumentation; 1993. American National Standard ANSI/AAMI SP10-1992.
    • (1993) Electronic or Automated Sphygmomanometers
  • 15
    • 16044368606 scopus 로고    scopus 로고
    • Berlin, Germany: Beuth Verlag; DIN Standard 58130
    • Deutsches Institut für Normierung eV. Nichtinvasive Blutdruckmeßgeräte - Klinische Prüfung. Berlin, Germany: Beuth Verlag; 1996. DIN Standard 58130.
    • (1996) Nichtinvasive Blutdruckmeßgeräte - Klinische Prüfung
  • 17
    • 0035799041 scopus 로고    scopus 로고
    • Blood pressure measuring devices: Recommendations of the European Society of Hypertension
    • O'Brien E, Waeber B, Parati G, Staessan J, Myers MG. Blood pressure measuring devices: recommendations of the European Society of Hypertension. Br Med J 2001; 322:531-536.
    • (2001) Br Med J , vol.322 , pp. 531-536
    • O'Brien, E.1    Waeber, B.2    Parati, G.3    Staessan, J.4    Myers, M.G.5
  • 18
    • 0142061633 scopus 로고    scopus 로고
    • A website for blood pressure monitoring devices: Dableducational.com
    • O'Brien E. A website for blood pressure monitoring devices: dableducational.com. Blood Press Monit 2003; 8:177.
    • (2003) Blood Press Monit , vol.8 , pp. 177
    • O'Brien, E.1
  • 19
    • 0034521052 scopus 로고    scopus 로고
    • A simulation-based evaluation of nine oscillometric blood pressure monitors for self-measurement
    • Ng KG. A simulation-based evaluation of nine oscillometric blood pressure monitors for self-measurement. Blood Press Monit 2000; 5:5-6.
    • (2000) Blood Press Monit , vol.5 , pp. 5-6
    • Ng, K.G.1
  • 20
    • 0031434311 scopus 로고    scopus 로고
    • Substitution of simulators for human subjects
    • Mieke S. Substitution of simulators for human subjects. Blood Press Monit 1997; 2:251-256.
    • (1997) Blood Press Monit , vol.2 , pp. 251-256
    • Mieke, S.1
  • 21
    • 0027298076 scopus 로고
    • Assessment of oscillometric non-invasive blood pressure monitors using the Dynatech Nevada Cuff Link analyzer
    • Amoore JN. Assessment of oscillometric non-invasive blood pressure monitors using the Dynatech Nevada Cuff Link analyzer. J Med Eng Technol 1993; 17:25-31.
    • (1993) J Med Eng Technol , vol.17 , pp. 25-31
    • Amoore, J.N.1
  • 22
    • 18844415051 scopus 로고    scopus 로고
    • Oscillometric blood pressure devices and simulators: Measurements of repeatability and differences between models
    • Sims AJ, Reay CA, Bousfield DR, Menes JA, Murray A. Oscillometric blood pressure devices and simulators: measurements of repeatability and differences between models. J Med Eng Technol 2005.
    • (2005) J Med Eng Technol
    • Sims, A.J.1    Reay, C.A.2    Bousfield, D.R.3    Menes, J.A.4    Murray, A.5
  • 23
    • 20044372008 scopus 로고    scopus 로고
    • Low-cost oscillometric blood pressure monitors: Device repeatability and device differences
    • Sims AJ, Reay CA, Bousfield DR, Menes JA, Murray A. Low-cost oscillometric blood pressure monitors: device repeatability and device differences. Physiol Meas 2005; 26:441-445.
    • (2005) Physiol Meas , vol.26 , pp. 441-445
    • Sims, A.J.1    Reay, C.A.2    Bousfield, D.R.3    Menes, J.A.4    Murray, A.5
  • 24
    • 0033742990 scopus 로고    scopus 로고
    • Validation of the AD UA-767 device for the self-measurement of blood pressure
    • Rogoza AN, Pavlova TS, Sergeeva MV. Validation of the AD UA-767 device for the self-measurement of blood pressure. Blood Press Monit 2000; 5:227-231.
    • (2000) Blood Press Monit , vol.5 , pp. 227-231
    • Rogoza, A.N.1    Pavlova, T.S.2    Sergeeva, M.V.3
  • 25
    • 0036694398 scopus 로고    scopus 로고
    • Validation of the AD UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance
    • Longo D, Bertolo O, Toffanin G, Frezza P, Palatini P. Validation of the AD UA-631 (UA-779 Life Source) device for self-measurement of blood pressure and relationship between its performance and large artery compliance. Blood Press Monit 2002; 7:243-248.
    • (2002) Blood Press Monit , vol.7 , pp. 243-248
    • Longo, D.1    Bertolo, O.2    Toffanin, G.3    Frezza, P.4    Palatini, P.5
  • 26
    • 0011706255 scopus 로고    scopus 로고
    • Evaluation and validation of Omron HEM 705 CP and HEM 706/711 monitors for self-measurement of blood pressure
    • Artigao LM, Llavador JJ, Puras A, López-Abril J, Rubio MM, Torres C, et al. Evaluation and validation of Omron HEM 705 CP and HEM 706/711 monitors for self-measurement of blood pressure. Aten Primaria 2000; 25:96-102.
    • (2000) Aten Primaria , vol.25 , pp. 96-102
    • Artigao, L.M.1    Llavador, J.J.2    Puras, A.3    López-Abril, J.4    Rubio, M.M.5    Torres, C.6
  • 27
    • 0031775127 scopus 로고    scopus 로고
    • Evaluation of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation
    • Anwar YA, Giacco S, McCabe EJ, Tendler BE, White WB. Evaluation of the Omron HEM-737 Intellisense device for use on adults according to the recommendations of the Association for the Advancement of Medical Instrumentation. Blood Press Monit 1998; 3:261-266.
    • (1998) Blood Press Monit , vol.3 , pp. 261-266
    • Anwar, Y.A.1    Giacco, S.2    McCabe, E.J.3    Tendler, B.E.4    White, W.B.5
  • 30
    • 0042383300 scopus 로고    scopus 로고
    • Evaluation of two devices for self-measurement of blood pressure according to the International Protocol: The Omron M5-I and the Omron 705IT
    • El Assaad MA, Topouchian JA, Asmar RA. Evaluation of two devices for self-measurement of blood pressure according to the International Protocol: the Omron M5-I and the Omron 705IT. Blood Press Monit 2003; 8:127-133.
    • (2003) Blood Press Monit , vol.8 , pp. 127-133
    • El Assaad, M.A.1    Topouchian, J.A.2    Asmar, R.A.3
  • 32
    • 0033741933 scopus 로고    scopus 로고
    • Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist blood pressure monitor for use on adults versus auscultation as defined by ANSI/AAMI SP10-1992
    • Wessig K, Hollinger S, Schmalzhaf I, Lenz T. Clinical evaluation of the efficacy of the Braun PrecisionSensor oscillometric wrist blood pressure monitor for use on adults versus auscultation as defined by ANSI/AAMI SP10-1992. Blood Press Monit 2000; 5:239-245.
    • (2000) Blood Press Monit , vol.5 , pp. 239-245
    • Wessig, K.1    Hollinger, S.2    Schmalzhaf, I.3    Lenz, T.4
  • 35
    • 0346689984 scopus 로고    scopus 로고
    • Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled
    • Yarrows SA. Comparison of the Omron HEM-637 wrist monitor to the auscultation method with the wrist position sensor on or disabled. Am J Hypertens 2004; 17:54-58.
    • (2004) Am J Hypertens , vol.17 , pp. 54-58
    • Yarrows, S.A.1
  • 36
    • 0036802104 scopus 로고    scopus 로고
    • Wrist blood pressure-measuring devices: A comparative study of accuracy with a standard auscultatory method using a mercury manometer
    • Altunkan S, Yildiz S, Azer S. Wrist blood pressure-measuring devices: a comparative study of accuracy with a standard auscultatory method using a mercury manometer. Blood Press Monit 2002; 7:281-284.
    • (2002) Blood Press Monit , vol.7 , pp. 281-284
    • Altunkan, S.1    Yildiz, S.2    Azer, S.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.