Challenges associated with using bovine serum in wear testing orthopaedic biopolymers

Author keywords

arthroplasty; biodegradation; in vitro test; joint replacement; protein; wear mechanism

Indexed keywords

ARTIFICIAL JOINT; BOVINE SERUM; DIRECT IMPACT; IN-VITRO; IN-VITRO TESTS; INTERNATIONAL STANDARDS; JOINT REPLACEMENT; LITERATURE SURVEY; PRECIPITATION FORMATION; PROTEIN DEGRADATION; SIMULATOR TESTING; SYNOVIAL FLUID; WEAR MECHANISMS; WEAR PROCESS; WEAR RATES; WEAR-TESTING;

EID: 84856605902     PISSN: 09544119     EISSN: 20413033     Source Type: Journal    
DOI: 10.1177/0954411911416047     Document Type: Conference Paper
Times cited : (45)

References (104)
  • 2
    • 0028325196 scopus 로고
    • Osteolysis and particle disease in hip replacement: A review
    • Harris W. H. Osteolysis and particle disease in hip replacement: a review. Acta Orthop. 1994 ; 65 (1). 113-123
    • (1994) Acta Orthop. , vol.65 , Issue.1 , pp. 113-123
    • Harris, W.H.1
  • 3
    • 0033029624 scopus 로고    scopus 로고
    • Relationship of acetabular wear to osteolysis and loosening in to total hip arthroplasty
    • Sochart D. H. Relationship of acetabular wear to osteolysis and loosening in to total hip arthroplasty. Clin. Orthop. 1999 ; 363: 135-150
    • (1999) Clin. Orthop. , vol.363 , pp. 135-150
    • Sochart, D.H.1
  • 4
    • 2442419655 scopus 로고    scopus 로고
    • Early failure due to osteolysis in highly cross-linked ultra high molecular weight polyethylene: A case report
    • Bradford L., Kurland M., Sankaran H., Kim H., Pruitt L., Ries M. Early failure due to osteolysis in highly cross-linked ultra high molecular weight polyethylene: a case report. J. Bone Joint Surg. Am. 2004 ; 86: 1051-1056
    • (2004) J. Bone Joint Surg. Am. , vol.86 , pp. 1051-1056
    • Bradford, L.1    Kurland, M.2    Sankaran, H.3    Kim, H.4    Pruitt, L.5    Ries, M.6
  • 5
    • 1842473877 scopus 로고    scopus 로고
    • Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: A case report
    • Halley D., Glassman A., Crowninshield R. D. Recurrent dislocation after revision total hip replacement with a large prosthetic femoral head: a case report. J. Bone Joint Surg. Am. 2004 ; 86: 827-830
    • (2004) J. Bone Joint Surg. Am. , vol.86 , pp. 827-830
    • Halley, D.1    Glassman, A.2    Crowninshield, R.D.3
  • 6
    • 0022504260 scopus 로고
    • The problem of surface damage in polyethylene total knee components
    • Wright T. M., Bartel D. L. The problem of surface damage in polyethylene total knee components. Clin. Orthop. 1986 ; 205: 67-74
    • (1986) Clin. Orthop. , vol.205 , pp. 67-74
    • Wright, T.M.1    Bartel, D.L.2
  • 7
    • 0026346472 scopus 로고
    • Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components
    • Collier J. P., Mayor M. B., McNamara J. L., Surprenant V. A., Jensen R. E. Analysis of the failure of 122 polyethylene inserts from uncemented tibial knee components. Clin. Orthop. 1991 ; 273: 232-242
    • (1991) Clin. Orthop. , vol.273 , pp. 232-242
    • Collier, J.P.1    Mayor, M.B.2    McNamara, J.L.3    Surprenant, V.A.4    Jensen, R.E.5
  • 9
    • 0027972449 scopus 로고
    • Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty
    • Wasielewski R. C., Galante J. O., Leighty R. M., Natarajan R. N., Rosenberg A. G. Wear patterns on retrieved polyethylene tibial inserts and their relationship to technical considerations during total knee arthroplasty. Clin. Orthop. 1994 ; 299: 31-43
    • (1994) Clin. Orthop. , vol.299 , pp. 31-43
    • Wasielewski, R.C.1    Galante, J.O.2    Leighty, R.M.3    Natarajan, R.N.4    Rosenberg, A.G.5
  • 10
    • 0028008566 scopus 로고
    • Posteromedial tibial polyethylene failure in total knee replacements
    • Lewis P., Rorabeck C. H., Bourne R. B., Devane P. Posteromedial tibial polyethylene failure in total knee replacements. Clin. Orthop. 1994 ; 299: 11-17
    • (1994) Clin. Orthop. , vol.299 , pp. 11-17
    • Lewis, P.1    Rorabeck, C.H.2    Bourne, R.B.3    Devane, P.4
  • 11
    • 4444233039 scopus 로고    scopus 로고
    • Deformation, yielding, fracture and fatigue behaviour of conventional and highly cross-linked ultra high molecular weight polyethylene
    • Pruitt L. A. Deformation, yielding, fracture and fatigue behaviour of conventional and highly cross-linked ultra high molecular weight polyethylene. Biomaterials. 2005 ; 26: 905-915
    • (2005) Biomaterials , vol.26 , pp. 905-915
    • Pruitt, L.A.1
  • 13
    • 0141993638 scopus 로고    scopus 로고
    • Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: A randomized prospective trial
    • 55-59
    • Martell J. M., Verner J. J., Incavo S. J. Clinical performance of a highly cross-linked polyethylene at two years in total hip arthroplasty: a randomized prospective trial. J. Arthropl. 2003 ; 18 (7). S1 55-59
    • (2003) J. Arthropl. , vol.18 , Issue.7 , pp. 1
    • Martell, J.M.1    Verner, J.J.2    Incavo, S.J.3
  • 14
    • 58249094140 scopus 로고    scopus 로고
    • Five year comparative study of highly cross-linked (cross fire) and traditional polyethylene
    • Rajadhyaksha A. D., Brotea C., Cheung Y., Kuhn C., Ramakrishna R., Zelicof S. B. Five year comparative study of highly cross-linked (cross fire) and traditional polyethylene. J. Arthropl. 2009 ; 24 (2). 161-167
    • (2009) J. Arthropl. , vol.24 , Issue.2 , pp. 161-167
    • Rajadhyaksha, A.D.1    Brotea, C.2    Cheung, Y.3    Kuhn, C.4    Ramakrishna, R.5    Zelicof, S.B.6
  • 15
    • 78951469231 scopus 로고    scopus 로고
    • The incidence of acetabular osteolysis in young patients with conventional versus highly cross linked polyethylene
    • Mall N. A., Nunley R. M., Zhu J. J., Maloney W. J., Barrack R. L., Clohisy J. C. The incidence of acetabular osteolysis in young patients with conventional versus highly cross linked polyethylene. Clin. Orthop. Relat. Res. 2011 ; 469: 372-381
    • (2011) Clin. Orthop. Relat. Res. , vol.469 , pp. 372-381
    • Mall, N.A.1    Nunley, R.M.2    Zhu, J.J.3    Maloney, W.J.4    Barrack, R.L.5    Clohisy, J.C.6
  • 16
    • 0018034893 scopus 로고
    • Wear characteristics of UHMWPE: A method for accurately measuring extremely low wear rates
    • McKellop H., Clarke I. C., Markolf K. L., Amstutz H. C. Wear characteristics of UHMWPE: a method for accurately measuring extremely low wear rates. J. Biomed. Mater. Res. 1978 ; 12: 895-927
    • (1978) J. Biomed. Mater. Res. , vol.12 , pp. 895-927
    • McKellop, H.1    Clarke, I.C.2    Markolf, K.L.3    Amstutz, H.C.4
  • 17
    • 0027724402 scopus 로고
    • The effect of transfer film and surface roughness on the wear of lubricated UHMWPE
    • Cooper J. R., Dowson D., Fisher J. The effect of transfer film and surface roughness on the wear of lubricated UHMWPE. Clin. Mater. 1993 ; 14: 295-302
    • (1993) Clin. Mater. , vol.14 , pp. 295-302
    • Cooper, J.R.1    Dowson, D.2    Fisher, J.3
  • 19
    • 70349663249 scopus 로고    scopus 로고
    • Biopolymer wear screening rig validated to ASTM F732-00 and against clinical data
    • Joyce T. J. Biopolymer wear screening rig validated to ASTM F732-00 and against clinical data. Tribol. Mat., Surf. Inter. 2007 ; 1 (2). 63-67
    • (2007) Tribol. Mat., Surf. Inter. , vol.1 , Issue.2 , pp. 63-67
    • Joyce, T.J.1
  • 20
  • 21
    • 0042957553 scopus 로고    scopus 로고
    • Lubrication and wear of the Charnley, Charnley-Muller, and McKee-Farrar prostheses with special regard to their clinical behavior
    • Simon S. R., Radin E. L. Lubrication and wear of the Charnley, Charnley-Muller, and McKee-Farrar prostheses with special regard to their clinical behavior. Proceedings of First Open Scientific Meeting of the Hip Society ; Mosby, St Louis ;. 33.
    • Proceedings of First Open Scientific Meeting of the Hip Society
    • Simon, S.R.1    Radin, E.L.2
  • 22
    • 0019897146 scopus 로고
    • The wear characteristics of ultrahigh molecular weight polyethylene against a high density alumina ceramic under wet (distilled water) and dry conditions
    • Dowson D., Harding R. T. The wear characteristics of ultrahigh molecular weight polyethylene against a high density alumina ceramic under wet (distilled water) and dry conditions. Wear. 1982 ; 75: 313-331
    • (1982) Wear , vol.75 , pp. 313-331
    • Dowson, D.1    Harding, R.T.2
  • 23
    • 0025834740 scopus 로고
    • Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy
    • Kumar P., Oka M., Ikeuchi K., Shimizu K., Yamamuro T., Okumura H., Kotoura Y. Low wear rate of UHMWPE against zirconia ceramic (Y-PSZ) in comparison to alumina ceramic and SUS 316L alloy. J. Biomed. Mater. Res. 1991 ; 25: 813-828
    • (1991) J. Biomed. Mater. Res. , vol.25 , pp. 813-828
    • Kumar, P.1    Oka, M.2    Ikeuchi, K.3    Shimizu, K.4    Yamamuro, T.5    Okumura, H.6    Kotoura, Y.7
  • 24
    • 0030248625 scopus 로고    scopus 로고
    • Wear testing of materials and surfaces for total knee replacement
    • Walker P. S., Blunn G. W., Lilley P. A. Wear testing of materials and surfaces for total knee replacement. J. Biomed. Mater. Res. 1996 ; 33: 159-175
    • (1996) J. Biomed. Mater. Res. , vol.33 , pp. 159-175
    • Walker, P.S.1    Blunn, G.W.2    Lilley, P.A.3
  • 26
    • 0032126428 scopus 로고    scopus 로고
    • A multidirectional motion pin-on-disk wear test method for prosthetic joint materials
    • Saikko V. A multidirectional motion pin-on-disk wear test method for prosthetic joint materials. J. Biomed. Mater. Res. 1998 ; 41: 58-64
    • (1998) J. Biomed. Mater. Res. , vol.41 , pp. 58-64
    • Saikko, V.1
  • 27
    • 0033635152 scopus 로고    scopus 로고
    • A multi-directional wear screening device and preliminary results of UHMWPE articulating against stainless steel
    • Joyce T. J., Monk D., Scholes S. C., Unsworth A. A multi-directional wear screening device and preliminary results of UHMWPE articulating against stainless steel. Biomed. Mater. Eng. 2000 ; 10: 241-249
    • (2000) Biomed. Mater. Eng. , vol.10 , pp. 241-249
    • Joyce, T.J.1    Monk, D.2    Scholes, S.C.3    Unsworth, A.4
  • 32
    • 0042069555 scopus 로고    scopus 로고
    • The influences of lubricant and material on polymer/CoCr sliding friction
    • Yao Q., Laurent M. P., Johnson T. S., Blanchard C. R., Crowninshield R. D. The influences of lubricant and material on polymer/CoCr sliding friction. Wear. 2003 ; 255: 780-784
    • (2003) Wear , vol.255 , pp. 780-784
    • Yao, Q.1    Laurent, M.P.2    Johnson, T.S.3    Blanchard, C.R.4    Crowninshield, R.D.5
  • 33
    • 0031253559 scopus 로고    scopus 로고
    • Wear of prosthetic joint materials in various lubricants
    • Ahlroos T., Saikko V. Wear of prosthetic joint materials in various lubricants. Wear. 1997 ; 211: 113-119
    • (1997) Wear , vol.211 , pp. 113-119
    • Ahlroos, T.1    Saikko, V.2
  • 34
    • 85006577810 scopus 로고    scopus 로고
    • A review of lubrication conditions for wear simulation in artificial hip replacements
    • Brown S. S., Clarke I. C. A review of lubrication conditions for wear simulation in artificial hip replacements. Tribol. Trans. 2006 ; 49: 72-78
    • (2006) Tribol. Trans. , vol.49 , pp. 72-78
    • Brown, S.S.1    Clarke, I.C.2
  • 35
    • 0346209911 scopus 로고
    • The proteins of synovial fluid: A study of the α1/α2 globulin ratio
    • Binette J. P., Schmid K. The proteins of synovial fluid: a study of the α1/α2 globulin ratio. Arthritis Rheum. 1965 ; 8 (1). 14-28
    • (1965) Arthritis Rheum. , vol.8 , Issue.1 , pp. 14-28
    • Binette, J.P.1    Schmid, K.2
  • 38
    • 0030627862 scopus 로고    scopus 로고
    • Frictional heating of bearing materials tested in hip joint wear simulator
    • Lu Z., McKellop H. A. Frictional heating of bearing materials tested in hip joint wear simulator. Proc. IMechE, Part H: J. Engineering in Medicine. 1997 ; 211: 101-108
    • (1997) Proc. IMechE, Part H: J. Engineering in Medicine , vol.211 , pp. 101-108
    • Lu, Z.1    McKellop, H.A.2
  • 41
    • 84862884881 scopus 로고    scopus 로고
    • Weariest: Influence of monoclinic content at the surface of zirconia heads
    • Cales B. Weariest: influence of monoclinic content at the surface of zirconia heads. Norton Desmarquest Fine Ceramics ; Evreux Cedex, France ;.
    • Norton Desmarquest Fine Ceramics
    • Cales, B.1
  • 42
    • 0016942059 scopus 로고
    • The wear of plastic materials in the hip-joint
    • Charnley J. The wear of plastic materials in the hip-joint. Plast. Rubber. 1976 ; 1: 59-63
    • (1976) Plast. Rubber , vol.1 , pp. 59-63
    • Charnley, J.1
  • 44
    • 0028439173 scopus 로고
    • Comparative study of the wear of UHMWPE with zirconia ceramic and stainless steel femoral heads in artificial hip joints
    • Derbyshire B., Fisher J., Downson D., Hardaker C., Brummitt K. Comparative study of the wear of UHMWPE with zirconia ceramic and stainless steel femoral heads in artificial hip joints. Med. Eng. Phys. 1994 ; 16 (3). 229-236
    • (1994) Med. Eng. Phys. , vol.16 , Issue.3 , pp. 229-236
    • Derbyshire, B.1    Fisher, J.2    Downson, D.3    Hardaker, C.4    Brummitt, K.5
  • 45
    • 0030152397 scopus 로고    scopus 로고
    • Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditions
    • Wang A., Essner A., Stark C., Dumbleton J. H. Comparison of the size and morphology of UHMWPE wear debris produced by a hip joint simulator under serum and water lubricated conditions. Biomaterials. 1996 ; 17: 865-871
    • (1996) Biomaterials , vol.17 , pp. 865-871
    • Wang, A.1    Essner, A.2    Stark, C.3    Dumbleton, J.H.4
  • 46
    • 0027437359 scopus 로고
    • Wear of polyethylene acetabular cup. Metallic and ceramic heads compared in a hip simulator
    • Saikko V., Paavolainen P. O., Slatis P. Wear of polyethylene acetabular cup. Metallic and ceramic heads compared in a hip simulator. Acta Orthop. Scand. 1993 ; 64: 391-403
    • (1993) Acta Orthop. Scand. , vol.64 , pp. 391-403
    • Saikko, V.1    Paavolainen, P.O.2    Slatis, P.3
  • 48
    • 0029255360 scopus 로고
    • Wear mechanisms of UHMWPE in total joint replacements
    • Wang A., Sun D. C., Stark C., Dumbleton J. H. Wear mechanisms of UHMWPE in total joint replacements. Wear. 1995 ; 181 183: 241-249
    • (1995) Wear , vol.181-183 , pp. 241-249
    • Wang, A.1    Sun, D.C.2    Stark, C.3    Dumbleton, J.H.4
  • 49
    • 0019620861 scopus 로고
    • Friction and wear properties of polymer, metal and ceramic prosthetic joint materials evaluated on a multi channel screening device
    • McKellop H., Clarke I. C., Markolf K., Amstutz H. Friction and wear properties of polymer, metal and ceramic prosthetic joint materials evaluated on a multi channel screening device. J. Biomed. Mater. Res. 1981 ; 15: 619-653
    • (1981) J. Biomed. Mater. Res. , vol.15 , pp. 619-653
    • McKellop, H.1    Clarke, I.C.2    Markolf, K.3    Amstutz, H.4
  • 50
    • 25144435848 scopus 로고    scopus 로고
    • Friction and wear mechanisms in hip prosthesis: Comparison of joint materials behaviour in several lubricants
    • Gispert M. P., Serro A. P., Colco R., Saramago B. Friction and wear mechanisms in hip prosthesis: comparison of joint materials behaviour in several lubricants. Wear. 2006 ; 260: 149-158
    • (2006) Wear , vol.260 , pp. 149-158
    • Gispert, M.P.1    Serro, A.P.2    Colco, R.3    Saramago, B.4
  • 51
    • 0030560772 scopus 로고    scopus 로고
    • Water and bovine serum lubrication compared in a simulator PTFE/CoCr wear model
    • Clark I. C. Water and bovine serum lubrication compared in a simulator PTFE/CoCr wear model. J. Appl. Biomater. 1996 ; 33: 275-283
    • (1996) J. Appl. Biomater. , vol.33 , pp. 275-283
    • Clark, I.C.1
  • 52
    • 0030560772 scopus 로고    scopus 로고
    • Water and bovine serum lubrication compared in simulator PTFE/CoCr wear model
    • Good V. D., Clarke I. C., Anissian L. Water and bovine serum lubrication compared in simulator PTFE/CoCr wear model. J. Biomed. Mater. Res. 1996 ; 33 (4). 275-283
    • (1996) J. Biomed. Mater. Res. , vol.33 , Issue.4 , pp. 275-283
    • Good, V.D.1    Clarke, I.C.2    Anissian, L.3
  • 55
    • 77954791566 scopus 로고    scopus 로고
    • Biochemical comparisons of osteoarthritic human synovial fluid with calf sera used in knee simulator wear testing
    • Brandt J. M., Briere L. K., Marr J., MacDonald S. J., Bourne R. B., Medley J. B. Biochemical comparisons of osteoarthritic human synovial fluid with calf sera used in knee simulator wear testing. J. Biomed. Mater. Res. A. 2010 ; 94A (3). 961-971
    • (2010) J. Biomed. Mater. Res. A , vol.94 , Issue.3 , pp. 961-971
    • Brandt, J.M.1    Briere, L.K.2    Marr, J.3    MacDonald, S.J.4    Bourne, R.B.5    Medley, J.B.6
  • 57
    • 84862884874 scopus 로고    scopus 로고
    • The impact of lubricant protein concentration on the outcome of hip joint simulator testing
    • Wang A., Polineni V. K., Essner A., Stark C., Dumbleton J. H. The impact of lubricant protein concentration on the outcome of hip joint simulator testing. 45th ORS ; Anaheim, CA ;.
    • 45th ORS
    • Wang, A.1    Polineni, V.K.2    Essner, A.3    Stark, C.4    Dumbleton, J.H.5
  • 59
    • 84862893645 scopus 로고    scopus 로고
    • Sigma-Aldrich. Bovine Calf Serum Data Sheet Lot 117K8404", Sigmaaldrich
    • Sigma-Aldrich. Bovine Calf Serum Data Sheet Lot 117K8404", Sigmaaldrich.
  • 60
    • 0346725975 scopus 로고    scopus 로고
    • The effect of lubricant composition on in vitro wear testing of polymeric acetabular components
    • Wang A., Essner A., Schmidig G. The effect of lubricant composition on in vitro wear testing of polymeric acetabular components. J. Biomed. Mater. Res., Part B: Appl. Biomater. 2004 ; 68B: 45-52
    • (2004) J. Biomed. Mater. Res., Part B: Appl. Biomater. , vol.68 , pp. 45-52
    • Wang, A.1    Essner, A.2    Schmidig, G.3
  • 61
    • 0042230975 scopus 로고    scopus 로고
    • Effect of lubricant protein concentration on the wear of ultrahigh molecular weight polyethylene sliding against a CoCr counterface
    • Saikko V. Effect of lubricant protein concentration on the wear of ultrahigh molecular weight polyethylene sliding against a CoCr counterface. Trans ASME. 2003 ; 125: 638-642
    • (2003) Trans ASME , vol.125 , pp. 638-642
    • Saikko, V.1
  • 62
    • 77955536914 scopus 로고    scopus 로고
    • The effect of serum protein concentration on wear rates in a hip simulator
    • John K. S. The effect of serum protein concentration on wear rates in a hip simulator. J. Biomat. Appl. 2010 ; 25: 145-159
    • (2010) J. Biomat. Appl. , vol.25 , pp. 145-159
    • John, K.S.1
  • 64
    • 0008293323 scopus 로고
    • Comparative distribution of proteins and glycoproteins of serum and synovial fluid
    • Decker B., McKenzie B. F., McGuckin W. F., Slocumb C. H. Comparative distribution of proteins and glycoproteins of serum and synovial fluid. Arthritis Rheum. 1959 ; 2 (2). 162-177
    • (1959) Arthritis Rheum. , vol.2 , Issue.2 , pp. 162-177
    • Decker, B.1    McKenzie, B.F.2    McGuckin, W.F.3    Slocumb, C.H.4
  • 66
    • 0035876256 scopus 로고    scopus 로고
    • Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon
    • Saikko V., Ahlroos T., Calonius O., Keränen J. Wear simulation of total hip prostheses with polyethylene against CoCr, alumina and diamond-like carbon. Biomaterials. 2001 ; 22: 1507-514
    • (2001) Biomaterials , vol.22 , pp. 1507-1514
    • Saikko, V.1    Ahlroos, T.2    Calonius, O.3    Keränen, J.4
  • 67
    • 0031686498 scopus 로고    scopus 로고
    • Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups
    • Wang A., Polineni V. K., Stark C., Dumbleton J. H. Effect of femoral head surface roughness on the wear of ultrahigh molecular weight polyethylene acetabular cups. J. Arthropl. 1998 ; 13: 615-620
    • (1998) J. Arthropl. , vol.13 , pp. 615-620
    • Wang, A.1    Polineni, V.K.2    Stark, C.3    Dumbleton, J.H.4
  • 68
    • 0032616832 scopus 로고    scopus 로고
    • Type of motion and lubricant in wear simulation of polyethylene acetabular cup
    • Saikko V., Ahlroos T. Type of motion and lubricant in wear simulation of polyethylene acetabular cup. Proc. IMechE, Part H: J. Engineering in Medicine. 1999 ; 213: 301-310
    • (1999) Proc. IMechE, Part H: J. Engineering in Medicine , vol.213 , pp. 301-310
    • Saikko, V.1    Ahlroos, T.2
  • 71
    • 77955384560 scopus 로고    scopus 로고
    • Characterization of protein degradation in serum based lubricants during simulation wear testing of metal-on-metal hip prostheses
    • Maskiewicz V. K., Williams P. A., Prates S. J., Bowsher J. G., Clarke I. C. Characterization of protein degradation in serum based lubricants during simulation wear testing of metal-on-metal hip prostheses. J. Biomed. Mat. Res., Part B: Appl. Biomat. 2010 ; 94B (2). 429-440
    • (2010) J. Biomed. Mat. Res., Part B: Appl. Biomat. , vol.94 , Issue.2 , pp. 429-440
    • Maskiewicz, V.K.1    Williams, P.A.2    Prates, S.J.3    Bowsher, J.G.4    Clarke, I.C.5
  • 73
    • 84862884877 scopus 로고    scopus 로고
    • Available from: http://www.bbraun.com.au/index. cfm66AB39392A5AE626642337DB5AC48B8F.
  • 74
    • 84862888291 scopus 로고    scopus 로고
    • Available from: http://www.doctissimo.fr/medicament-PLASMION.htm
  • 75
    • 34548414633 scopus 로고    scopus 로고
    • Rheological properties of synovial fluid
    • Fam H., Bryant J. T., Kontopoulou M. Rheological properties of synovial fluid. Biorheology. 2007 ; 44: 59-74
    • (2007) Biorheology , vol.44 , pp. 59-74
    • Fam, H.1    Bryant, J.T.2    Kontopoulou, M.3
  • 77
    • 0141600293 scopus 로고    scopus 로고
    • 13, vol. 1, 1997, pp. 81-102 (Williams and Williams, Baltimore). Koopman W. J., ed. Baltimore: Williams and Williams
    • McCarty D. J. Arthritis allied conditions: a text book of rheumatology. 13, vol. 1, 1997, pp. 81-102 (Williams and Williams, Baltimore). Koopman W. J., ed. Baltimore: Williams and Williams ; 1997: 81-102.
    • (1997) Arthritis Allied Conditions: A Text Book of Rheumatology , pp. 81-102
    • McCarty, D.J.1
  • 79
    • 0031171676 scopus 로고    scopus 로고
    • Phospholipids as boundary lubricants in wear tests of prosthetic joint materials
    • Saikko V., Ahlroos T. Phospholipids as boundary lubricants in wear tests of prosthetic joint materials. Wear. 1997 ; 207: 86-91
    • (1997) Wear , vol.207 , pp. 86-91
    • Saikko, V.1    Ahlroos, T.2
  • 82
    • 0036937210 scopus 로고    scopus 로고
    • Highly viscous sodium hyaluronate and joint lubrication
    • Mori S., Naito M., Moriyama S. Highly viscous sodium hyaluronate and joint lubrication. Int. Orthop. 2002 ; 26: 116-121
    • (2002) Int. Orthop. , vol.26 , pp. 116-121
    • Mori, S.1    Naito, M.2    Moriyama, S.3
  • 83
    • 33748475338 scopus 로고    scopus 로고
    • Increased total knee arthroplasty ultra-high molecular weight polyethylene wear using a clinically relevant hyaluronic acid simulator lubricant
    • DesJardins J., Aurora A., Tanner S. L., Pace T. B., Acampora K. B., LaBerge M. Increased total knee arthroplasty ultra-high molecular weight polyethylene wear using a clinically relevant hyaluronic acid simulator lubricant. Proc. IMechE, Part H: J. Engineering in Medicine. 2006 ; 220 (H1). 609-623
    • (2006) Proc. IMechE, Part H: J. Engineering in Medicine , vol.220 , Issue.H1 , pp. 609-623
    • Desjardins, J.1    Aurora, A.2    Tanner, S.L.3    Pace, T.B.4    Acampora, K.B.5    Laberge, M.6
  • 85
    • 33749878090 scopus 로고    scopus 로고
    • Effect of synovia constituents on friction and wear of ultra high molecular weight polyethylene sliding against prosthetic joint materials
    • Sawae Y., Murakami T., Chen J. Effect of synovia constituents on friction and wear of ultra high molecular weight polyethylene sliding against prosthetic joint materials. Wear. 1998 ; 77: 1-7
    • (1998) Wear , vol.77 , pp. 1-7
    • Sawae, Y.1    Murakami, T.2    Chen, J.3
  • 86
    • 77049140451 scopus 로고
    • The physiological function of hyaluronic acid in synovial fluid; Viscous, elastic and lubricant properties
    • Ogston A. G., Stainer J. E. The physiological function of hyaluronic acid in synovial fluid; viscous, elastic and lubricant properties. J. Physiol. 1953 ; 119: 244-252
    • (1953) J. Physiol. , vol.119 , pp. 244-252
    • Ogston, A.G.1    Stainer, J.E.2
  • 89
    • 16644389660 scopus 로고    scopus 로고
    • The role of joint fluid in the tribology of total joint arthoplasty
    • Mazzucco D., Spector M. The role of joint fluid in the tribology of total joint arthoplasty. Clin. Ortho. 2004 ; 429: 17-32
    • (2004) Clin. Ortho. , vol.429 , pp. 17-32
    • Mazzucco, D.1    Spector, M.2
  • 90
    • 1642351728 scopus 로고    scopus 로고
    • Composition of joint fluid in patients undergoing total knee replacement and revision arthroplasty: Correlation with flow properties
    • Mazzucco D., Scott R., Spector M. Composition of joint fluid in patients undergoing total knee replacement and revision arthroplasty: correlation with flow properties. Biomaterials. 2004 ; 25: 4433-4445
    • (2004) Biomaterials , vol.25 , pp. 4433-4445
    • Mazzucco, D.1    Scott, R.2    Spector, M.3
  • 91
    • 0022343146 scopus 로고
    • Concentration and molecular weight of sodium hyaluronate in synovial fluid from patients with rheumatoid arthritis and other arthropathies
    • Dahl L. B., Dahl I. M., Engstrom-Laurent A., Granath K. Concentration and molecular weight of sodium hyaluronate in synovial fluid from patients with rheumatoid arthritis and other arthropathies. Ann. Rheum. Dis. 1985 ; 44: 817-822
    • (1985) Ann. Rheum. Dis. , vol.44 , pp. 817-822
    • Dahl, L.B.1    Dahl, I.M.2    Engstrom-Laurent, A.3    Granath, K.4
  • 92
    • 0031466954 scopus 로고    scopus 로고
    • Assay of synovial fluid parameters: Hyaluronan concentration as a potential marker for joint diseases
    • Praest B. M, Greiling H., Kock R. Assay of synovial fluid parameters: hyaluronan concentration as a potential marker for joint diseases. Clin. Chim. Act. 1997 ; 266: 117-128
    • (1997) Clin. Chim. Act. , vol.266 , pp. 117-128
    • Praest, B.M.1    Greiling, H.2    Kock, R.3
  • 94
    • 0014118432 scopus 로고
    • Hyaluronic acid in synovial fluid. I. Molecular parameters of hyaluronic acid in normal and arthritis human fluids
    • Balazs E. A., Watson D., Duff I. F., Roseman S. Hyaluronic acid in synovial fluid. I. Molecular parameters of hyaluronic acid in normal and arthritis human fluids. Arthritis Rheum. 1967 ; 10: 357-376
    • (1967) Arthritis Rheum. , vol.10 , pp. 357-376
    • Balazs, E.A.1    Watson, D.2    Duff, I.F.3    Roseman, S.4
  • 95
    • 0027817288 scopus 로고
    • Comparisons of the oscillatory shear viscoelasticity and composition of pathological fluids
    • Gomez J. E., Thurston G. B. Comparisons of the oscillatory shear viscoelasticity and composition of pathological fluids. Biorheology. 1993 ; 30: 409-427
    • (1993) Biorheology , vol.30 , pp. 409-427
    • Gomez, J.E.1    Thurston, G.B.2
  • 96
    • 11544311243 scopus 로고
    • The intrinsic viscosity of synovial fluid hyaluronic acid
    • Bollet A. J. The intrinsic viscosity of synovial fluid hyaluronic acid. J. Lab. Clin. Med. 1956 ; 48: 712-728
    • (1956) J. Lab. Clin. Med. , vol.48 , pp. 712-728
    • Bollet, A.J.1
  • 97
    • 0027213271 scopus 로고
    • Viscoelastic and transient network properties of HA as function of the concentration
    • De Smedt S., Deketser P., Ribitsch V., Lauwers A., Demeester J. Viscoelastic and transient network properties of HA as function of the concentration. Biorheology. 1993 ; 30: 31-41
    • (1993) Biorheology , vol.30 , pp. 31-41
    • De Smedt, S.1    Deketser, P.2    Ribitsch, V.3    Lauwers, A.4    Demeester, J.5
  • 98
    • 0000327139 scopus 로고    scopus 로고
    • Rheology of synovial fluids and substitute polymer
    • Schurz J. Rheology of synovial fluids and substitute polymer. J. Macromol. Sci.: Pure. 1996 ; A33: 1249-1262
    • (1996) J. Macromol. Sci.: Pure. , vol.33 , pp. 1249-1262
    • Schurz, J.1
  • 99
    • 0003929035 scopus 로고
    • Studies on the nature and significance of macromolecular complexes in the rheology of synovial fluid from normal and diseased human joints
    • Nuki G., Ferguson J. Studies on the nature and significance of macromolecular complexes in the rheology of synovial fluid from normal and diseased human joints. Rheol. Acta. 1971 ; 10: 8-14
    • (1971) Rheol. Acta. , vol.10 , pp. 8-14
    • Nuki, G.1    Ferguson, J.2
  • 100
    • 34548392714 scopus 로고
    • New methods in synovial fluid rheology and macromolecular network structure
    • Zeidler H., Altmann S. New methods in synovial fluid rheology and macromolecular network structure. Biorheology. 1981 ; 18: 122-127
    • (1981) Biorheology , vol.18 , pp. 122-127
    • Zeidler, H.1    Altmann, S.2
  • 101
    • 70349659732 scopus 로고    scopus 로고
    • Wear tests of orthopaedic biopolymers with the biolubricant augmented by a visco-supplement
    • Joyce T. J. Wear tests of orthopaedic biopolymers with the biolubricant augmented by a visco-supplement. Proc. IMechE, Part J: J. Engineering Tribology. 2009 ; 233: 297-302
    • (2009) Proc. IMechE, Part J: J. Engineering Tribology , vol.233 , pp. 297-302
    • Joyce, T.J.1
  • 102
    • 0021212341 scopus 로고
    • Surfactants identified in synovial fluid and their ability to act as boundary lubricants
    • Hills B. A., Butler B. D. Surfactants identified in synovial fluid and their ability to act as boundary lubricants. Ann. Rheum. Dis. 1984 ; 43 (4). 641-648
    • (1984) Ann. Rheum. Dis. , vol.43 , Issue.4 , pp. 641-648
    • Hills, B.A.1    Butler, B.D.2
  • 103
    • 0035222432 scopus 로고    scopus 로고
    • The influence of phospholipids concentration in protein containing lubricants on the wear of ultra-high molecular weight polyethylene in artificial hip joints
    • Bell J., Tipper J. L., Ingham E., Stone M. H., Fisher J. The influence of phospholipids concentration in protein containing lubricants on the wear of ultra-high molecular weight polyethylene in artificial hip joints. Proc. IMechE, Part H: J. Engineering in Medicine. 2000 ; 215: 259-263
    • (2000) Proc. IMechE, Part H: J. Engineering in Medicine , vol.215 , pp. 259-263
    • Bell, J.1    Tipper, J.L.2    Ingham, E.3    Stone, M.H.4    Fisher, J.5
  • 104
    • 0018569849 scopus 로고
    • Lipid composition of the tissues of human knee joints. I. Observations in normal joints (articular cartilage, meniscus, ligaments, synovial fluid, synovium, intra-articular fat pad and bone marrow)
    • Rabinowitz J. L., Gregg J. R., Nixon J. E., Schumacher H. R. Lipid composition of the tissues of human knee joints. I. Observations in normal joints (articular cartilage, meniscus, ligaments, synovial fluid, synovium, intra-articular fat pad and bone marrow). Clin. Orthop. Relat. Res. 1979 ; 143: 260-265
    • (1979) Clin. Orthop. Relat. Res. , vol.143 , pp. 260-265
    • Rabinowitz, J.L.1    Gregg, J.R.2    Nixon, J.E.3    Schumacher, H.R.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.