Volumn 22, Issue 2, 2009, Pages 193-200

Prosthetic screw detorque values in implants retained as cast bar superstructures or bars modified by the cresco ti precision technique-a comparative in vivo study

Author keywords

[No Author keywords available]

Indexed keywords

ARTICLE; CLINICAL TRIAL; CONTROLLED CLINICAL TRIAL; CONTROLLED STUDY; DENTAL CARE; DENTURE; EDENTULOUSNESS; EQUIPMENT DESIGN; HUMAN; INSTRUMENTATION; MANDIBLE; PROSPECTIVE STUDY; RANDOMIZED CONTROLLED TRIAL; TOOTH IMPLANTATION; TOOTH PROSTHESIS; TORQUE;

EID: 67650359881     PISSN: 08932174     EISSN: None     Source Type: Journal    
DOI: None     Document Type: Article
Times cited : (9)

References (62)
  • 1
    • 0037928812 scopus 로고    scopus 로고
    • A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years
    • Berglundh T, Persson L, Klinge B. A systematic review of the incidence of biological and technical complications in implant dentistry reported in prospective longitudinal studies of at least 5 years. J Clin Periodontol 2002;29 (suppl 3):197-212.
    • (2002) J Clin Periodontol , vol.29 , Issue.SUPPL 3 , pp. 197-212
    • Berglundh, T.1    Persson, L.2    Klinge, B.3
  • 2
    • 0035296434 scopus 로고    scopus 로고
    • Effectiveness of maxillary overdentures supported by implants: Maintenance and prosthetic complications
    • Kiener P, Oetterli M, Mericske E, Mericske-Stern R. Effectiveness of maxillary overdentures supported by implants: Maintenance and prosthetic complications. Int J Prosthodont 2001;14:133-140.
    • (2001) Int J Prosthodont , vol.14 , pp. 133-140
    • Kiener, P.1    Oetterli, M.2    Mericske, E.3    Mericske-Stern, R.4
  • 3
    • 27644447305 scopus 로고    scopus 로고
    • Biologic response of immediately versus delayed loaded implants supporting ill-fitting prostheses: An animal study
    • Duyck J, Vrielinck L, Lambrichts I, et al. Biologic response of immediately versus delayed loaded implants supporting ill-fitting prostheses: An animal study. Clin Implant Dent Relat Res 2005;7:150-158.
    • (2005) Clin Implant Dent Relat Res , vol.7 , pp. 150-158
    • Duyck, J.1    Vrielinck, L.2    Lambrichts, I.3
  • 4
    • 0030237874 scopus 로고    scopus 로고
    • Prosthesis misfit and marginal bone loss in edentulous implant patients
    • Jemt T, Book K. Prosthesis misfit and marginal bone loss in edentulous implant patients. Int J Oral Maxillofac Implants 1996;11:620-625.
    • (1996) Int J Oral Maxillofac Implants , vol.11 , pp. 620-625
    • Jemt, T.1    Book, K.2
  • 5
    • 0030277405 scopus 로고    scopus 로고
    • The response of bone in primates around unloaded dental implants supporting prostheses with different levels of fit
    • Carr AB, Gerard DA, Larsen PE. The response of bone in primates around unloaded dental implants supporting prostheses with different levels of fit J Prosthet Dent 1996;76:500-509.
    • (1996) J Prosthet Dent , vol.76 , pp. 500-509
    • Carr, A.B.1    Gerard, D.A.2    Larsen, P.E.3
  • 6
    • 0030097958 scopus 로고    scopus 로고
    • In vivo measurements of precision of fit involving implant- supported prostheses in the edentulous jaw
    • Jemt T. In vivo measurements of precision of fit involving implant- supported prostheses in the edentulous jaw. Int J Oral Maxillofac Implants 1996;11:151-158.
    • (1996) Int J Oral Maxillofac Implants , vol.11 , pp. 151-158
    • Jemt, T.1
  • 7
    • 0031995934 scopus 로고    scopus 로고
    • Complication rate with prosthodontic reconstructions on ITI and IMZ dental implants Internationales Team für Implantologie
    • Behr M, Lang R, Leibrock A, Rosentritt M, Handel G. Complication rate with prosthodontic reconstructions on ITI and IMZ dental implants. Internationales Team für Implantologie. Clin Oral Implants Res 1998;9:51-58.
    • (1998) Clin Oral Implants Res , vol.9 , pp. 51-58
    • Behr, M.1    Lang, R.2    Leibrock, A.3    Rosentritt, M.4    Handel, G.5
  • 8
    • 0026218667 scopus 로고
    • Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: A study of treatment from the time of prosthesis placement to the first annual checkup
    • Jemt T. Failures and complications in 391 consecutively inserted fixed prostheses supported by Brånemark implants in edentulous jaws: A study of treatment from the time of prosthesis placement to the first annual checkup. Int J Oral Maxillofac Implants 1991;6:270-276.
    • (1991) Int J Oral Maxillofac Implants , vol.6 , pp. 270-276
    • Jemt, T.1
  • 9
    • 0026832538 scopus 로고
    • Failures and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: From prosthetic treatment to first annual checkup
    • Jemt T, Lindén B, Lekholm U. Failures and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants: From prosthetic treatment to first annual checkup. Int J Oral Maxillofac Implants 1992;7:40-44.
    • (1992) Int J Oral Maxillofac Implants , vol.7 , pp. 40-44
    • Jemt, T.1    Lindén, B.2    Lekholm, U.3
  • 10
    • 0028390316 scopus 로고
    • Loose gold screws frequently occur in fullarch fixed prostheses supported by osseointegrated implants after5years
    • Kallus T, Bessing C. Loose gold screws frequently occur in fullarch fixed prostheses supported by osseointegrated implants after5years. Int J Oral Maxillofac Implants 1994;9:169-178.
    • (1994) Int J Oral Maxillofac Implants , vol.9 , pp. 169-178
    • Kallus, T.1    Bessing, C.2
  • 11
    • 0036366379 scopus 로고    scopus 로고
    • Changes in prosthetic screw stability because of misfit of implant-supported prostheses
    • al-Turki LE, Chai J, Lautenschlager EP, Hutten MC. Changes in prosthetic screw stability because of misfit of implant-supported prostheses. Int J Prosthodont 2002;15:38-42.
    • (2002) Int J Prosthodont , vol.15 , pp. 38-42
    • Al-Turki, L.E.1    Chai, J.2    Lautenschlager, E.P.3    Hutten, M.C.4
  • 12
    • 0033124057 scopus 로고    scopus 로고
    • Manual closing torque in five implant abutment systems: An in vitro comparative study
    • Gross M, Kozak D, Laufer BZ, Weiss El. Manual closing torque in five implant abutment systems: An in vitro comparative study. J Prosthet Dent 1999;81:574-578.
    • (1999) J Prosthet Dent , vol.81 , pp. 574-578
    • Gross, M.1    Kozak, D.2    Laufer, B.Z.3    El, W.4
  • 14
    • 0028185281 scopus 로고
    • Osseointegrated implants for single-tooth replacement: Progress report from a multicenter prospective study after 3 years
    • Laney WR, Jemt T, Harris D, et al. Osseointegrated implants for single-tooth replacement: Progress report from a multicenter prospective study after 3 years. Int J Oral Maxillofac Implants 1994;9:49-54.
    • (1994) Int J Oral Maxillofac Implants , vol.9 , pp. 49-54
    • Laney, W.R.1    Jemt, T.2    Harris, D.3
  • 15
    • 0038505307 scopus 로고    scopus 로고
    • Cyclic loading of implant-supported prostheses: Changes in component fit over time
    • Hecker DM, Eckert SE. Cyclic loading of implant-supported prostheses: Changes in component fit over time. J Prosthet Dent 2003;89:346-351.
    • (2003) J Prosthet Dent , vol.89 , pp. 346-351
    • Hecker, D.M.1    Eckert, S.E.2
  • 16
    • 33845618691 scopus 로고    scopus 로고
    • Influence of impression material and time on the 3-dimensional accuracy of implant impressions
    • Hoist S, Blatz MB, Bergler M, Goellner M, Wichmann M. Influence of impression material and time on the 3-dimensional accuracy of implant impressions. Quintessence Int 2007;38:67-73.
    • (2007) Quintessence Int , vol.38 , pp. 67-73
    • Hoist, S.1    Blatz, M.B.2    Bergler, M.3    Goellner, M.4    Wichmann, M.5
  • 17
    • 0033154855 scopus 로고    scopus 로고
    • Photogrammetry-An alternative to conventional impressions in implant dentistry? A clinical pilot study
    • Jemt T, Bäck T, Petersson A. Photogrammetry-An alternative to conventional impressions in implant dentistry? A clinical pilot study. Int J Prosthodont 1999;12:363-368.
    • (1999) Int J Prosthodont , vol.12 , pp. 363-368
    • Jemt, T.1    Bäck, T.2    Petersson, A.3
  • 18
    • 0033755243 scopus 로고    scopus 로고
    • Clinical long-term results and 10-year Kaplan- Meier analysis of Cerec restorations
    • Reiss B, Walther W. Clinical long-term results and 10-year Kaplan- Meier analysis of Cerec restorations. Int J Comput Dent 2000;3:9-23.
    • (2000) Int J Comput Dent , vol.3 , pp. 9-23
    • Reiss, B.1    Walther, W.2
  • 19
    • 0028186392 scopus 로고
    • The influence of spruing technique on the development of tension in a cast partial denture framework
    • Augthun M, Zyfuss M, Spiekermann H. The influence of spruing technique on the development of tension in a cast partial denture framework. Int J Prosthodont 1994;7:72-76.
    • (1994) Int J Prosthodont , vol.7 , pp. 72-76
    • Augthun, M.1    Zyfuss, M.2    Spiekermann, H.3
  • 20
    • 34648846773 scopus 로고    scopus 로고
    • Characteristics of commercial quick-heating phosphate-bonded investments for the accelerated casting technique
    • Yang CC, Yang HH, Ding SJ, Huang TH, Kao CT, Yan M. Characteristics of commercial quick-heating phosphate-bonded investments for the accelerated casting technique. Quintessence Int 2007;38:e271-e278.
    • (2007) Quintessence Int , vol.38
    • Yang, C.C.1    Yang, H.H.2    Ding, S.J.3    Huang, T.H.4    Kao, C.T.5    Yan, M.6
  • 22
    • 0032018774 scopus 로고    scopus 로고
    • Description and evaluation of a simplified method to achieve passive fit between cast titanium frameworks and implants
    • Helldén LB, Démand T. Description and evaluation of a simplified method to achieve passive fit between cast titanium frameworks and implants. Int J Oral Maxillofac Implants 1998;13:190-196.
    • (1998) Int J Oral Maxillofac Implants , vol.13 , pp. 190-196
    • Helldén, L.B.1    Démand, T.2
  • 23
    • 0033122309 scopus 로고    scopus 로고
    • Precision of CNC-milled titanium frameworks for implant treatment in the edentulous jaw
    • Jemt T, Bäck T, Petersson A. Precision of CNC-milled titanium frameworks for implant treatment in the edentulous jaw. Int J Prosthodont 1999;12:209-215.
    • (1999) Int J Prosthodont , vol.12 , pp. 209-215
    • Jemt, T.1    Bäck, T.2    Petersson, A.3
  • 24
    • 0029377214 scopus 로고
    • Accuracy of implant-supported prostheses in the edentulous jaw: Analysis of precision of fit between cast gold-alloy frameworks and master casts by means of a three-dimensional photogrammetric technique
    • Jemt T, Lie A. Accuracy of implant-supported prostheses in the edentulous jaw: Analysis of precision of fit between cast gold-alloy frameworks and master casts by means of a three-dimensional photogrammetric technique. Clin Oral Implants Res 1995;6: 172-180.
    • (1995) Clin Oral Implants Res , vol.6 , pp. 172-180
    • Jemt, T.1    Lie, A.2
  • 25
    • 0033182713 scopus 로고    scopus 로고
    • Distortion of three- unit implant frameworks during casting, soldering, and simulated porcelain firings
    • Zervas PJ, Papazoglou E, Beck FM, Carr AB. Distortion of three- unit implant frameworks during casting, soldering, and simulated porcelain firings. J Prosthodont 1999;8:171-179.
    • (1999) J Prosthodont , vol.8 , pp. 171-179
    • Zervas, P.J.1    Papazoglou, E.2    Beck, F.M.3    Carr, A.B.4
  • 26
    • 4143149751 scopus 로고    scopus 로고
    • In vitro comparative analysis of the fit of gold alloy or commercially pure titanium implant-supported prostheses before and after elec- troerosion
    • Sartori IA, Ribeiro RF, Francischone CE, de Mattos Mda G. In vitro comparative analysis of the fit of gold alloy or commercially pure titanium implant-supported prostheses before and after elec- troerosion. J Prosthet Dent 2004;92:132-138.
    • (2004) J Prosthet Dent , vol.92 , pp. 132-138
    • Sartori, I.A.1    Ribeiro, R.F.2    Francischone, C.E.3    De Mattos Mda, G.4
  • 27
    • 33749241323 scopus 로고    scopus 로고
    • The evolution of the CEREC system
    • Mörmann WH. The evolution of the CEREC system. J Am Dent Assoc 2006;137 suppl:7S-13S.
    • (2006) J Am Dent Assoc , vol.137 , Issue.SUPPL
    • Mörmann, W.H.1
  • 28
    • 0038488528 scopus 로고    scopus 로고
    • Comparisons of precision of fit between cast and CNC-milled titanium implant frameworks for the edentulous mandible
    • Ortorp A, Jemt T, Bäck T, Jälevik T. Comparisons of precision of fit between cast and CNC-milled titanium implant frameworks for the edentulous mandible. Int J Prosthodont 2003;16:194-200.
    • (2003) Int J Prosthodont , vol.16 , pp. 194-200
    • Ortorp, A.1    Jemt, T.2    Bäck, T.3    Jälevik, T.4
  • 29
    • 33746383047 scopus 로고    scopus 로고
    • Marginal accuracy of titanium copings fabricated by casting and CAD/CAM techniques
    • Witkowski S, Komine F, Gerds T. Marginal accuracy of titanium copings fabricated by casting and CAD/CAM techniques. J Prosthet Dent 2006;96:47-52.
    • (2006) J Prosthet Dent , vol.96 , pp. 47-52
    • Witkowski, S.1    Komine, F.2    Gerds, T.3
  • 30
    • 0031301642 scopus 로고    scopus 로고
    • Fit of implant frameworks fabricated by different techniques
    • Riedy SJ, Lang BR, Lang BE. Fit of implant frameworks fabricated by different techniques. J Prosthet Dent 1997;78:596-604.
    • (1997) J Prosthet Dent , vol.78 , pp. 596-604
    • Riedy, S.J.1    Lang, B.R.2    Lang, B.E.3
  • 32
    • 9644275499 scopus 로고    scopus 로고
    • Clinical performance of a method for the fabrication of implant-supported precisely fitting titanium frameworks: A retrospective 5- to 8-year clinical follow-up study
    • Hedkvist L, Mattsson T, Hellden LB. Clinical performance of a method for the fabrication of implant-supported precisely fitting titanium frameworks: A retrospective 5- to 8-year clinical follow- up study. Clin Implant Dent Relat Res 2004;6:174-180.
    • (2004) Clin Implant Dent Relat Res , vol.6 , pp. 174-180
    • Hedkvist, L.1    Mattsson, T.2    Hellden, L.B.3
  • 33
    • 13444301257 scopus 로고    scopus 로고
    • The Cresco Bridge and implant concept: Presentation of a technology for fabrication of abutment-free, passively fitting superstructures
    • Helldén LB, Ericson G, Olsson CO. The Cresco Bridge and implant concept: Presentation of a technology for fabrication of abutment-free, passively fitting superstructures. Int J Periodontics Restorative Dent 2005;25:89-94.
    • (2005) Int J Periodontics Restorative Dent , vol.25 , pp. 89-94
    • Helldén, L.B.1    Ericson, G.2    Olsson, C.O.3
  • 34
    • 35648989009 scopus 로고    scopus 로고
    • A simplified method to assess precision of fit between framework and supporting implants: A preliminary study
    • Calderini A, Maiorana C, Garlini G, Abbondanza T. A simplified method to assess precision of fit between framework and supporting implants: A preliminary study. Int J Oral Maxillofac Implants 2007;22:831-838.
    • (2007) Int J Oral Maxillofac Implants , vol.22 , pp. 831-838
    • Calderini, A.1    Maiorana, C.2    Garlini, G.3    Abbondanza, T.4
  • 35
    • 31344468073 scopus 로고    scopus 로고
    • A 3-year retrospective study of Cresco frameworks: Preload and complications
    • Hjalmarsson L, Smedberg Jl. A 3-year retrospective study of Cresco frameworks: Preload and complications. Clin Implant Dent Relat Res 2005;7:189-199.
    • (2005) Clin Implant Dent Relat Res , vol.7 , pp. 189-199
    • Hjalmarsson, L.1    Jl, S.2
  • 36
    • 0028451094 scopus 로고
    • Patterson EA Tightening characteristics for screwed joints in osseointegrated dental implants
    • Burguete RL, Johns RB, King T, Patterson EA Tightening characteristics for screwed joints in osseointegrated dental implants. J Prosthet Dent 1994:71:592-599.
    • (1994) J Prosthet Dent , vol.71 , pp. 592-599
    • Burguete, R.L.1    Johns, R.B.2    King, T.3
  • 37
    • 33645959184 scopus 로고    scopus 로고
    • Effect of casting procedures on screw loosening in UCLA-type abutments
    • Kano SC, Binon P, Bonfante G, Curtis DA. Effect of casting procedures on screw loosening in UCLA-type abutments. J Prosthodont 2006;15:77-81.
    • (2006) J Prosthodont , vol.15 , pp. 77-81
    • Kano, S.C.1    Binon, P.2    Bonfante, G.3    Curtis, D.A.4
  • 38
    • 0033074393 scopus 로고    scopus 로고
    • Assessment of cold welding properties of the internal conical interface of two commercially available implant systems
    • Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-166.
    • (1999) J Prosthet Dent , vol.81 , pp. 159-166
    • Norton, M.R.1
  • 39
    • 0036673514 scopus 로고    scopus 로고
    • The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws
    • Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant prosthetic screws. J Prosthet Dent 2002;88:183-191.
    • (2002) J Prosthet Dent , vol.88 , pp. 183-191
    • Tzenakis, G.K.1    Nagy, W.W.2    Fournelle, R.A.3    Dhuru, V.B.4
  • 40
    • 0038350018 scopus 로고    scopus 로고
    • The influence of functional forces on the biomechanics of implant-supported prostheses-A review
    • Sahin S, Cehreli MC, Yalçin E. The influence of functional forces on the biomechanics of implant-supported prostheses-A review. J Dent 2002;30:271-282.
    • (2002) J Dent , vol.30 , pp. 271-282
    • Sahin, S.1    Cehreli, M.C.2    Yalçin, E.3
  • 41
    • 0020776467 scopus 로고
    • Biomechanical considerations in osseointegrated prostheses
    • Skalak R. Biomechanical considerations in osseointegrated prostheses. J Prosthet Dent 1983;49:843-848.
    • (1983) J Prosthet Dent , vol.49 , pp. 843-848
    • Skalak, R.1
  • 42
    • 0027015903 scopus 로고
    • A study of 589 consecutive implants supporting complete fixed prostheses. Part II: Prosthetic aspects
    • Naert I, Quirynen M, van Steenberghe D, Darius P. A study of 589 consecutive implants supporting complete fixed prostheses. Part II: Prosthetic aspects. J Prosthet Dent 1992;68:949-956.
    • (1992) J Prosthet Dent , vol.68 , pp. 949-956
    • Naert, I.1    Quirynen, M.2    Van Steenberghe, D.3    Darius, P.4
  • 43
    • 0025471133 scopus 로고
    • The longitudinal clinical effectiveness of osseointegrated dental implants: The Toronto study. Part III: Problems and complications encountered
    • Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: The Toronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-194.
    • (1990) J Prosthet Dent , vol.64 , pp. 185-194
    • Zarb, G.A.1    Schmitt, A.2
  • 44
    • 0033210999 scopus 로고    scopus 로고
    • The CrescoTi Precision method: Description of a simplified method to fabricate titanium superstructures with passive fit to osseointegrated implants
    • Helldén LB, Dérand T, Johansson S, Lindberg A. The CrescoTi Precision method: Description of a simplified method to fabricate titanium superstructures with passive fit to osseointegrated implants. J Prosthet Dent 1999;82:487-491.
    • (1999) J Prosthet Dent , vol.82 , pp. 487-491
    • Helldén, L.B.1    Dérand, T.2    Johansson, S.3    Lindberg, A.4
  • 45
    • 14844303580 scopus 로고    scopus 로고
    • Structure and mechanical properties of Cresco-Ti laser-welded joints and stress analyses using finite element models of fixed distal extension and fixed partial prosthetic designs
    • Uysal H, Kurtoglu C, Gurbuz R, Tutuncu N. Structure and mechanical properties of Cresco-Ti laser-welded joints and stress analyses using finite element models of fixed distal extension and fixed partial prosthetic designs. J Prosthet Dent 2005;93:235-244.
    • (2005) J Prosthet Dent , vol.93 , pp. 235-244
    • Uysal, H.1    Kurtoglu, C.2    Gurbuz, R.3    Tutuncu, N.4
  • 46
    • 0029396277 scopus 로고
    • Joining titanium materials with tungsten inert gas welding, laser welding, and infrared brazing
    • Wang RR, Welsch GE. Joining titanium materials with tungsten inert gas welding, laser welding, and infrared brazing. J Prosthet Dent 1995;74:521-530.
    • (1995) J Prosthet Dent , vol.74 , pp. 521-530
    • Wang, R.R.1    Welsch, G.E.2
  • 47
    • 33644899699 scopus 로고    scopus 로고
    • Over-refractory casting technique as an alternative to one-piece multi-unit fixed partial denture frameworks
    • de Oliveira Correa G, Henriques GE, Mesquita MF, Sobrinho LC. Over-refractory casting technique as an alternative to one-piece multi-unit fixed partial denture frameworks. J Prosthet Dent 2006;95:243-248.
    • (2006) J Prosthet Dent , vol.95 , pp. 243-248
    • De Oliveira Correa, G.1    Henriques, G.E.2    Mesquita, M.F.3    Sobrinho, L.C.4
  • 48
    • 0031203639 scopus 로고    scopus 로고
    • Soldering and remelting influence on fatigue strength of cobalt-chromium alloys
    • Henriques GE, Consani S, Rollo JM, Andrade e Silva F. Soldering and remelting influence on fatigue strength of cobalt-chromium alloys. J Prosthet Dent 1997;78:146-152.
    • (1997) J Prosthet Dent , vol.78 , pp. 146-152
    • Henriques, G.E.1    Consani, S.2    Rollo, J.M.3    Andrade, E.4    Silva, F.5
  • 49
    • 34047251477 scopus 로고    scopus 로고
    • A comparison of the accuracy of fit of 2 methods for fabricating implant-prosthodontic frameworks
    • Al-Fadda SA, Zarb GA, Finer Y. A comparison of the accuracy of fit of 2 methods for fabricating implant-prosthodontic frameworks. Int J Prosthodont 2007;20:125-131.
    • (2007) Int J Prosthodont , vol.20 , pp. 125-131
    • Al-Fadda, S.A.1    Zarb, G.A.2    Finer, Y.3
  • 50
    • 33749673379 scopus 로고    scopus 로고
    • Physical and metallurgical considerations of failures of soldered bars in bar attachment systems for implant overdentures: A review of the literature
    • Waddell JN, Payne AG, Swain MV. Physical and metallurgical considerations of failures of soldered bars in bar attachment systems for implant overdentures: A review of the literature. J Prosthet Dent 2006;96:283-288.
    • (2006) J Prosthet Dent , vol.96 , pp. 283-288
    • Waddell, J.N.1    Payne, A.G.2    Swain, M.V.3
  • 51
    • 0028389987 scopus 로고
    • Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants
    • Goheen KL, Vermilyea SG, Vossoughi J, Agar JR. Torque generated by handheld screwdrivers and mechanical torquing devices for osseointegrated implants. Int J Oral Maxillofac Implants 1994;9:149-155.
    • (1994) Int J Oral Maxillofac Implants , vol.9 , pp. 149-155
    • Goheen, K.L.1    Vermilyea, S.G.2    Vossoughi, J.3    Agar, J.R.4
  • 52
    • 0030238164 scopus 로고    scopus 로고
    • Effects of fabrication, finishing, and polishing procedures on preload in prostheses using conventional " gold' and plastic cylinders
    • Carr AB, Brunski JB, Hurley E. Effects of fabrication, finishing, and polishing procedures on preload in prostheses using conventional "gold' and plastic cylinders. Int J Oral Maxillofac Implants 1996;11:589-598.
    • (1996) Int J Oral Maxillofac Implants , vol.11 , pp. 589-598
    • Carr, A.B.1    Brunski, J.B.2    Hurley, E.3
  • 53
    • 0018352075 scopus 로고
    • Bite force and oral function in complete denture wearers
    • Haraldson T, Karlsson U, Carlsson GE. Bite force and oral function in complete denture wearers. J Oral Rehabil 1979;6:41-48.
    • (1979) J Oral Rehabil , vol.6 , pp. 41-48
    • Haraldson, T.1    Karlsson, U.2    Carlsson, G.E.3
  • 54
    • 38449120467 scopus 로고    scopus 로고
    • Influence of bite force and tongue pressure on oro-pharyngeal residue in the elderly
    • Ono T, Kumakura I, Arimoto M, et al. Influence of bite force and tongue pressure on oro-pharyngeal residue in the elderly. Gerodontology 2007;24:143-150.
    • (2007) Gerodontology , vol.24 , pp. 143-150
    • Ono, T.1    Kumakura, I.2    Arimoto, M.3
  • 56
    • 0038284246 scopus 로고    scopus 로고
    • Mandibular flexure and stress build-up in mandibular full-arch fixed prostheses supported by osseointegrated implants
    • Zarone F, Apicella A, Nicolais L, Aversa R, Sorrentino R. Mandibular flexure and stress build-up in mandibular full-arch fixed prostheses supported by osseointegrated implants. Clin Oral Implants Res 2003;14:103-114.
    • (2003) Clin Oral Implants Res , vol.14 , pp. 103-114
    • Zarone, F.1    Apicella, A.2    Nicolais, L.3    Aversa, R.4    Sorrentino, R.5
  • 57
    • 41649119154 scopus 로고    scopus 로고
    • Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws
    • Al Jabbari YS, Fournelle R, Ziebert G, Toth J.lacopino AM. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 1: Characterization of adhesive wear and structure of retaining screws. J Prosthodont 2007;17:168-180.
    • (2007) J Prosthodont , vol.17 , pp. 168-180
    • Al Jabbari, Y.S.1    Fournelle, R.2    Ziebert, G.3    Toth, J.4    Lacopino, A.M.5
  • 58
    • 41649100086 scopus 로고    scopus 로고
    • Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 4: Failure analysis of 10 fractured retaining screws retrieved from three patients
    • Al Jabbari YS, Fournelle R, Ziebert G, Toth J, lacopino AM. Mechanical behavior and failure analysis of prosthetic retaining screws after long-term use in vivo. Part 4: Failure analysis of 10 fractured retaining screws retrieved from three patients. J Prosthodont 2008;17:201-210.
    • (2008) J Prosthodont , vol.17 , pp. 201-210
    • Al Jabbari, Y.S.1    Fournelle, R.2    Ziebert, G.3    Toth, J.4    Lacopino, A.M.5
  • 59
    • 0035430649 scopus 로고    scopus 로고
    • Clinical evaluation and prosthetic complications of single tooth replacements by non-submerged implants
    • Mericske-Stern R, Grutter L, Rosch R, Mericske E. Clinical evaluation and prosthetic complications of single tooth replacements by non-submerged implants. Clin Oral Implants Res 2001;12:309-318.
    • (2001) Clin Oral Implants Res , vol.12 , pp. 309-318
    • Mericske-Stern, R.1    Grutter, L.2    Rosch, R.3    Mericske, E.4
  • 60
    • 0036357495 scopus 로고    scopus 로고
    • Relationship between implant preload and screw loosening on implant-supported prostheses
    • Siamos G, Winkler S, Boberick KG. Relationship between implant preload and screw loosening on implant-supported prostheses. J Oral Implantol 2002;28:67-73.
    • (2002) J Oral Implantol , vol.28 , pp. 67-73
    • Siamos, G.1    Winkler, S.2    Boberick, K.G.3
  • 61
    • 0025657547 scopus 로고
    • 3-year followup study of early single implant restorations ad modum Branemark
    • Jemt T, Lekholm U, Gröndahl K. 3-year followup study of early single implant restorations ad modum Branemark. Int J Periodontics Restorative Dent 1990;10:340-349.
    • (1990) Int J Periodontics Restorative Dent , vol.10 , pp. 340-349
    • Jemt, T.1    Lekholm, U.2    Gröndahl, K.3
  • 62
    • 0030089626 scopus 로고    scopus 로고
    • Ultimate tensile strength of five interchangeable prosthetic retaining screws
    • Jaarda MJ, Razzoog ME, Gratton DG. Ultimate tensile strength of five interchangeable prosthetic retaining screws. Implant Dent 1996;5:16-19.
    • (1996) Implant Dent , vol.5 , pp. 16-19
    • Jaarda, M.J.1    Razzoog, M.E.2    Gratton, D.G.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.