Accuracy of commercial devices and methods for noninvasive estimation of aortic systolic blood pressure a systematic review and meta-analysis of invasive validation studies

(16)  Papaioannou, Theodore G a   Karageorgopoulou, Theofani D a   Sergentanis, Theodoros N b   Protogerou, Athanase D c   Psaltopoulou, Theodora b   Sharman, James E d   Weber, Thomas e   Blacher, Jacques f   Daskalopoulou, Stella S g   Wassertheurer, Siegfried h   Khir, Ashraf W i   Vlachopoulos, Charalambos a   Stergiopulos, Nikolaos j   Stefanadis, Christodoulos k   Nichols, Wilmer W l   Tousoulis, Dimitrios a  


Author keywords

accuracy; aortic hemodynamics; applanation tonometry; calibration; fluid filled catheter manometer systems; micro tip pressure transducers; oscillometry; pressure wave; pulse wave analysis; transfer functions

Indexed keywords

AORTA PRESSURE; AORTIC SYSTOLIC BLOOD PRESSURE; ARTERY WALL; BLOOD PRESSURE MEASUREMENT; BRACHIAL ARTERY; CALIBRATION; CAROTID ARTERY; DIASTOLIC BLOOD PRESSURE; ECHOGRAPHY; HEMODYNAMIC PARAMETERS; HUMAN; INVASIVE PROCEDURE; MEAN ARTERIAL PRESSURE; MEASUREMENT ACCURACY; MEASUREMENT ERROR; META ANALYSIS; NON INVASIVE BLOOD PRESSURE MONITOR; NON INVASIVE MEASUREMENT; OSCILLOMETRY; PHYSIOLOGICAL PROCESS; PRIORITY JOURNAL; PULSE WAVE; RADIAL ARTERY; REVIEW; SYSTEMATIC REVIEW; SYSTOLIC BLOOD PRESSURE; TONOMETRY; AORTA; ARTERIAL PRESSURE; DEVICES; PHYSIOLOGY; PROCEDURES; REPRODUCIBILITY; STANDARDS; SYSTOLE;

EID: 84964749938     PISSN: 02636352     EISSN: 14735598     Source Type: Journal    
DOI: 10.1097/HJH.0000000000000921     Document Type: Review
Times cited : (126)

References (47)
  • 2
    • 84910640835 scopus 로고    scopus 로고
    • Blood pressure lowering and cardiovascular risk
    • Stoner L, Wu B. Blood pressure lowering and cardiovascular risk. Lancet 2014; 384: 1746.
    • (2014) Lancet , vol.384 , pp. 1746
    • Stoner, L.1    Wu, B.2
  • 3
    • 84925875800 scopus 로고    scopus 로고
    • New technological and clinical trends in blood pressure theranostics: Is it time to consider a spatio-Temporal approach?
    • Papaioannou TG, Vavuranakis M, Stefanadis C, Tousoulis D. New technological and clinical trends in blood pressure theranostics: is it time to consider a spatio-Temporal approach? Curr Pharm Des 2015; 21: 697-699.
    • (2015) Curr Pharm des , vol.21 , pp. 697-699
    • Papaioannou, T.G.1    Vavuranakis, M.2    Stefanadis, C.3    Tousoulis, D.4
  • 4
    • 45849141559 scopus 로고    scopus 로고
    • Central pressure: Variability and impact of cardiovascular risk factors: The anglo-Cardiff collaborative trial II
    • McEniery CM, Yasmin. McDonnell B, Munnery M, Wallace SM, Rowe CV, et al. Central pressure: variability and impact of cardiovascular risk factors: the Anglo-Cardiff Collaborative Trial II. Hypertension 2008; 51: 1476-1482.
    • (2008) Hypertension , vol.51 , pp. 1476-1482
    • McEniery, C.M.1    Yasmin McDonnell, B.2    Munnery, M.3    Wallace, S.M.4    Rowe, C.V.5
  • 6
    • 68549098079 scopus 로고    scopus 로고
    • Amplification of the pressure pulse in the upper limb in healthy, middle-Aged men and women
    • Segers P, Mahieu D, Kips J, Rietzschel E, De Buyzere M, De Bacquer D, et al. Amplification of the pressure pulse in the upper limb in healthy, middle-Aged men and women. Hypertension 2009; 54: 414-420.
    • (2009) Hypertension , vol.54 , pp. 414-420
    • Segers, P.1    Mahieu, D.2    Kips, J.3    Rietzschel, E.4    De Buyzere, M.5    De Bacquer, D.6
  • 7
    • 33645518432 scopus 로고    scopus 로고
    • Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: Principal results of the conduit artery function evaluation (cafe) study
    • Williams B, Lacy PS, Thom SM, Cruickshank K, Stanton A, Collier D, et al. Differential impact of blood pressure-lowering drugs on central aortic pressure and clinical outcomes: principal results of the Conduit Artery Function Evaluation (CAFE) study. Circulation 2006; 113: 1213-1225.
    • (2006) Circulation , vol.113 , pp. 1213-1225
    • Williams, B.1    Lacy, P.S.2    Thom, S.M.3    Cruickshank, K.4    Stanton, A.5    Collier, D.6
  • 8
    • 62249216401 scopus 로고    scopus 로고
    • The effect of antihypertensive drugs on central blood pressure beyond peripheral blood pressure Part I: Patho)-physiology, rationale and perspective on pulse pressure amplification
    • Protogerou AD, Papaioannou TG, Lekakis JP, Blacher J, Safar ME. The effect of antihypertensive drugs on central blood pressure beyond peripheral blood pressure. Part I: (Patho)-physiology, rationale and perspective on pulse pressure amplification. Curr Pharm Des 2009; 15: 267-271.
    • (2009) Curr Pharm des , vol.15 , pp. 267-271
    • Protogerou, A.D.1    Papaioannou, T.G.2    Lekakis, J.P.3    Blacher, J.4    Safar, M.E.5
  • 9
    • 77954814318 scopus 로고    scopus 로고
    • Prediction of cardiovascular events and all-cause mortality with central haemodynamics: A systematic review and meta-Analysis
    • Vlachopoulos C, Aznaouridis K, O'Rourke MF, Safar ME, Baou K, Stefanadis C. Prediction of cardiovascular events and all-cause mortality with central haemodynamics: a systematic review and meta-Analysis. Eur Heart J 2010; 31: 1865-1871.
    • (2010) Eur Heart J , vol.31 , pp. 1865-1871
    • Vlachopoulos, C.1    Aznaouridis, K.2    O'Rourke, M.F.3    Safar, M.E.4    Baou, K.5    Stefanadis, C.6
  • 10
    • 84886862518 scopus 로고    scopus 로고
    • Derivation and validation of diagnostic thresholds for central blood pressure measurements based on long-Term cardiovascular risks
    • Cheng HM, Chuang SY, Sung SH, Yu WC, Pearson A, Lakatta EG, et al. Derivation and validation of diagnostic thresholds for central blood pressure measurements based on long-Term cardiovascular risks. J Am Coll Cardiol 2013; 62: 1780-1787.
    • (2013) J Am Coll Cardiol , vol.62 , pp. 1780-1787
    • Cheng, H.M.1    Chuang, S.Y.2    Sung, S.H.3    Yu, W.C.4    Pearson, A.5    Lakatta, E.G.6
  • 11
    • 84870257287 scopus 로고    scopus 로고
    • Clinical assessment of central blood pressure
    • Miyashita H. Clinical assessment of central blood pressure. Curr Hypertens Rev 2012; 8: 80-90.
    • (2012) Curr Hypertens Rev , vol.8 , pp. 80-90
    • Miyashita, H.1
  • 12
    • 84925855696 scopus 로고    scopus 로고
    • Noninvasive estimation of aortic blood pressures: A close look at current devices and methods
    • Millasseau SAD. Noninvasive estimation of aortic blood pressures: a close look at current devices and methods. Curr Pharm Des 2015; 21: 709-718.
    • (2015) Curr Pharm des , vol.21 , pp. 709-718
    • Millasseau, S.A.D.1
  • 13
    • 84913593480 scopus 로고    scopus 로고
    • Central blood pressure and its amplification: A final breakthrough or do we need more?
    • Gillebert TC. Central blood pressure and its amplification: a final breakthrough or do we need more? Eur Heart J 2014; 35: 3088-3090.
    • (2014) Eur Heart J , vol.35 , pp. 3088-3090
    • Gillebert, T.C.1
  • 14
    • 0037658999 scopus 로고    scopus 로고
    • Pressure wave reflection assessed from the peripheral pulse: Is a transfer function necessary?
    • Millasseau SC, Patel SJ, Redwood SR, Ritter JM, Chowienczyk PJ. Pressure wave reflection assessed from the peripheral pulse: is a transfer function necessary? Hypertension 2003; 41: 1016-1020.
    • (2003) Hypertension , vol.41 , pp. 1016-1020
    • Millasseau, S.C.1    Patel, S.J.2    Redwood, S.R.3    Ritter, J.M.4    Chowienczyk, P.J.5
  • 15
    • 7944232464 scopus 로고    scopus 로고
    • Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform
    • Gallagher D, Adji A, O'Rourke MF. Validation of the transfer function technique for generating central from peripheral upper limb pressure waveform. Am J Hypertens 2004; 17: 1059-1067.
    • (2004) Am J Hypertens , vol.17 , pp. 1059-1067
    • Gallagher, D.1    Adji, A.2    O'Rourke, M.F.3
  • 16
    • 84906079410 scopus 로고    scopus 로고
    • Estimation of central aortic blood pressure: A systematic meta-Analysis of available techniques
    • Narayan O, Casan J, Szarski M, Dart AM, Meredith IT, Cameron JD. Estimation of central aortic blood pressure: a systematic meta-Analysis of available techniques. J Hypertens 2014; 32: 1727-1740.
    • (2014) J Hypertens , vol.32 , pp. 1727-1740
    • Narayan, O.1    Casan, J.2    Szarski, M.3    Dart, A.M.4    Meredith, I.T.5    Cameron, J.D.6
  • 17
    • 84883301130 scopus 로고    scopus 로고
    • Measurement accuracy of noninvasively obtained central blood pressure by applanation tonometry: A systematic review and meta-Analysis
    • Cheng HM, Lang D, Tufanaru C, Pearson A. Measurement accuracy of noninvasively obtained central blood pressure by applanation tonometry: a systematic review and meta-Analysis. Int J Cardiol 2013; 167: 1867-1876.
    • (2013) Int J Cardiol , vol.167 , pp. 1867-1876
    • Cheng, H.M.1    Lang, D.2    Tufanaru, C.3    Pearson, A.4
  • 18
    • 84925548115 scopus 로고    scopus 로고
    • Preferred reporting items for systematic reviews and meta-Analyses: The PRISMA statement
    • Moher D, Liberati A, Tetzlaff J, Altman DG. Preferred reporting items for systematic reviews and meta-Analyses: the PRISMA statement. J Clin Epidemiol 2009; 62: 1006-1012.
    • (2009) J Clin Epidemiol , vol.62 , pp. 1006-1012
    • Moher, D.1    Liberati, A.2    Tetzlaff, J.3    Altman, D.G.4
  • 20
    • 0037360953 scopus 로고    scopus 로고
    • Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure
    • Davies JI, Band MM, Pringle S, Ogston S, Struthers AD. Peripheral blood pressure measurement is as good as applanation tonometry at predicting ascending aortic blood pressure. J Hypertens 2003; 21: 571-576.
    • (2003) J Hypertens , vol.21 , pp. 571-576
    • Davies, J.I.1    Band, M.M.2    Pringle, S.3    Ogston, S.4    Struthers, A.D.5
  • 21
    • 0043092329 scopus 로고    scopus 로고
    • Estimation of central aortic pressure by SphygmoCor requires intra-Arterial peripheral pressures
    • Cloud GC, Rajkumar C, Kooner J, Cooke J, Bulpitt CJ. Estimation of central aortic pressure by SphygmoCor requires intra-Arterial peripheral pressures. Clin Sci (Lond) 2003; 105: 219-225.
    • (2003) Clin Sci (Lond) , vol.105 , pp. 219-225
    • Cloud, G.C.1    Rajkumar, C.2    Kooner, J.3    Cooke, J.4    Bulpitt, C.J.5
  • 22
    • 10244250292 scopus 로고    scopus 로고
    • Validation of a new noninvasive portable tonometer for determining arterial pressure wave and pulse wave velocity: The PulsePen device
    • Salvi P, Lio G, Labat C, Ricci E, Pannier B, Benetos A. Validation of a new noninvasive portable tonometer for determining arterial pressure wave and pulse wave velocity: the PulsePen device. J Hypertens 2004; 22: 2285-2293.
    • (2004) J Hypertens , vol.22 , pp. 2285-2293
    • Salvi, P.1    Lio, G.2    Labat, C.3    Ricci, E.4    Pannier, B.5    Benetos, A.6
  • 23
    • 34247228666 scopus 로고    scopus 로고
    • Relationship between radial and central arterial pulse wave and evaluation of central aortic pressure using the radial arterial pulse wave
    • Takazawa K, Kobayashi H, Shindo N, Tanaka N, Yamashina A. Relationship between radial and central arterial pulse wave and evaluation of central aortic pressure using the radial arterial pulse wave. Hypertens Res 2007; 30: 219-228.
    • (2007) Hypertens Res , vol.30 , pp. 219-228
    • Takazawa, K.1    Kobayashi, H.2    Shindo, N.3    Tanaka, N.4    Yamashina, A.5
  • 24
    • 59849084393 scopus 로고    scopus 로고
    • The noninvasive estimation of central aortic blood pressure in patients with aortic stenosis
    • Rajani R, Chowienczyk P, Redwood S, Guilcher A, Chambers JB. The noninvasive estimation of central aortic blood pressure in patients with aortic stenosis. J Hypertens 2008; 26: 2381-2388.
    • (2008) J Hypertens , vol.26 , pp. 2381-2388
    • Rajani, R.1    Chowienczyk, P.2    Redwood, S.3    Guilcher, A.4    Chambers, J.B.5
  • 25
    • 73249147117 scopus 로고    scopus 로고
    • The accuracy of central SBP determined from the second systolic peak of the peripheral pressure waveform
    • Hickson SS, Butlin M, Mir FA, Graggaber J, Cheriyan J, Khan F, et al. The accuracy of central SBP determined from the second systolic peak of the peripheral pressure waveform. J Hypertens 2009; 27: 1784-1788.
    • (2009) J Hypertens , vol.27 , pp. 1784-1788
    • Hickson, S.S.1    Butlin, M.2    Mir, F.A.3    Graggaber, J.4    Cheriyan, J.5    Khan, F.6
  • 26
    • 77957284980 scopus 로고    scopus 로고
    • Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity
    • Horvath IG, Nemeth A, Lenkey Z, Alessandri N, Tufano F, Kis P, et al. Invasive validation of a new oscillometric device (Arteriograph) for measuring augmentation index, central blood pressure and aortic pulse wave velocity. J Hypertens 2010; 28: 2068-2075.
    • (2010) J Hypertens , vol.28 , pp. 2068-2075
    • Horvath, I.G.1    Nemeth, A.2    Lenkey, Z.3    Alessandri, N.4    Tufano, F.5    Kis, P.6
  • 27
    • 77957608283 scopus 로고    scopus 로고
    • Validation of the central blood pressure estimation by the SphygmoCor system in Chinese
    • Zuo JL, Li Y, Yan ZJ, Zhang RY, Shen WF, Zhu DL, et al. Validation of the central blood pressure estimation by the SphygmoCor system in Chinese. Blood Press Monit 2010; 15: 268-274.
    • (2010) Blood Press Monit , vol.15 , pp. 268-274
    • Zuo, J.L.1    Li, Y.2    Yan, Z.J.3    Zhang, R.Y.4    Shen, W.F.5    Zhu, D.L.6
  • 28
    • 77649105206 scopus 로고    scopus 로고
    • Cuff and aortic pressure differences during dobutamine infusion: A study of the effects of systolic blood pressure amplification
    • Smulyan H, Mukherjee R, Sheehe PR, Safar ME. Cuff and aortic pressure differences during dobutamine infusion: a study of the effects of systolic blood pressure amplification. Am Heart J 2010; 159: 399-405.
    • (2010) Am Heart J , vol.159 , pp. 399-405
    • Smulyan, H.1    Mukherjee, R.2    Sheehe, P.R.3    Safar, M.E.4
  • 29
    • 79951603923 scopus 로고    scopus 로고
    • Development and validation of a novel method to derive central aortic systolic pressure from the radial pressure waveform using an n-point moving average method
    • Williams B, Lacy PS, Yan P, Hwee CN, Liang C, Ting CM. Development and validation of a novel method to derive central aortic systolic pressure from the radial pressure waveform using an n-point moving average method. J Am Coll Cardiol 2011; 57: 951-961.
    • (2011) J Am Coll Cardiol , vol.57 , pp. 951-961
    • Williams, B.1    Lacy, P.S.2    Yan, P.3    Hwee, C.N.4    Liang, C.5    Ting, C.M.6
  • 30
    • 83155172358 scopus 로고    scopus 로고
    • Validation of a brachial cuff-based method for estimating central systolic blood pressure
    • Weber T, Wassertheurer S, Rammer M, Maurer E, Hametner B, Mayer CC, et al. Validation of a brachial cuff-based method for estimating central systolic blood pressure. Hypertension 2011; 58: 825-832.
    • (2011) Hypertension , vol.58 , pp. 825-832
    • Weber, T.1    Wassertheurer, S.2    Rammer, M.3    Maurer, E.4    Hametner, B.5    Mayer, C.C.6
  • 31
    • 84865516120 scopus 로고    scopus 로고
    • Evaluation of a novel sphygmomanometer, which estimates central aortic blood pressure from analysis of brachial artery suprasystolic pressure waves
    • Lin AC, Lowe A, Sidhu K, Harrison W, Ruygrok P, Stewart R. Evaluation of a novel sphygmomanometer, which estimates central aortic blood pressure from analysis of brachial artery suprasystolic pressure waves. J Hypertens 2012; 30: 1743-1750.
    • (2012) J Hypertens , vol.30 , pp. 1743-1750
    • Lin, A.C.1    Lowe, A.2    Sidhu, K.3    Harrison, W.4    Ruygrok, P.5    Stewart, R.6
  • 32
    • 84861227791 scopus 로고    scopus 로고
    • Estimating central systolic blood pressure during oscillometric determination of blood pressure: Proof of concept and validation by comparison with intraaortic pressure recording and arterial tonometry
    • Brett SE, Guilcher A, Clapp B, Chowienczyk P. Estimating central systolic blood pressure during oscillometric determination of blood pressure: proof of concept and validation by comparison with intraaortic pressure recording and arterial tonometry. Blood Press Monit 2012; 17: 132-136.
    • (2012) Blood Press Monit , vol.17 , pp. 132-136
    • Brett, S.E.1    Guilcher, A.2    Clapp, B.3    Chowienczyk, P.4
  • 33
    • 84859905759 scopus 로고    scopus 로고
    • Estimation of central aortic systolic pressure using late systolic inflection of radial artery pulse and its application to vasodilator therapy
    • Takazawa K, Kobayashi H, Kojima I, Aizawa A, Kinoh M, Sugo Y, et al. Estimation of central aortic systolic pressure using late systolic inflection of radial artery pulse and its application to vasodilator therapy. J Hypertens 2012; 30: 908-916.
    • (2012) J Hypertens , vol.30 , pp. 908-916
    • Takazawa, K.1    Kobayashi, H.2    Kojima, I.3    Aizawa, A.4    Kinoh, M.5    Sugo, Y.6
  • 34
    • 84865736813 scopus 로고    scopus 로고
    • Comparison of two noninvasive devices for measurement of central systolic blood pressure with invasive measurement during cardiac catheterization
    • Ott C, Haetinger S, Schneider MP, Pauschinger M, Schmieder RE. Comparison of two noninvasive devices for measurement of central systolic blood pressure with invasive measurement during cardiac catheterization. J Clin Hypertens (Greenwich) 2012; 14: 575-579.
    • (2012) J Clin Hypertens (Greenwich) , vol.14 , pp. 575-579
    • Ott, C.1    Haetinger, S.2    Schneider, M.P.3    Pauschinger, M.4    Schmieder, R.E.5
  • 35
    • 84871713726 scopus 로고    scopus 로고
    • Comparison of the SphygmoCor and Omron devices in the estimation of pressure amplification against the invasive catheter measurement
    • Ding FH, Li Y, Zhang RY, Zhang Q, Wang JG. Comparison of the SphygmoCor and Omron devices in the estimation of pressure amplification against the invasive catheter measurement. J Hypertens 2013; 31: 86-93.
    • (2013) J Hypertens , vol.31 , pp. 86-93
    • Ding, F.H.1    Li, Y.2    Zhang, R.Y.3    Zhang, Q.4    Wang, J.G.5
  • 36
    • 84871711775 scopus 로고    scopus 로고
    • Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure
    • Pucci G, Cheriyan J, Hubsch A, Hickson SS, Gajendragadkar PR, Watson T, et al. Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure. J Hypertens 2013; 31: 77-85.
    • (2013) J Hypertens , vol.31 , pp. 77-85
    • Pucci, G.1    Cheriyan, J.2    Hubsch, A.3    Hickson, S.S.4    Gajendragadkar, P.R.5    Watson, T.6
  • 37
    • 84891598250 scopus 로고    scopus 로고
    • Assessment of central blood pressure in patients with type 2 diabetes: A comparison between SphygmoCor and invasively measured values
    • Laugesen E, Rossen NB, Peters CD, Maeng M, Ebbehoj E, Knudsen ST, et al. Assessment of central blood pressure in patients with type 2 diabetes: a comparison between SphygmoCor and invasively measured values. Am J Hypertens 2014; 27: 169-176.
    • (2014) Am J Hypertens , vol.27 , pp. 169-176
    • Laugesen, E.1    Rossen, N.B.2    Peters, C.D.3    Maeng, M.4    Ebbehoj, E.5    Knudsen, S.T.6
  • 38
    • 84898869348 scopus 로고    scopus 로고
    • Invasive validation of arteriograph estimates of central blood pressure in patients with type 2 diabetes
    • Rossen NB, Laugesen E, Peters CD, Ebbehoj E, Knudsen ST, Poulsen PL, et al. Invasive validation of arteriograph estimates of central blood pressure in patients with type 2 diabetes. Am J Hypertens 2014; 27: 674-679.
    • (2014) Am J Hypertens , vol.27 , pp. 674-679
    • Rossen, N.B.1    Laugesen, E.2    Peters, C.D.3    Ebbehoj, E.4    Knudsen, S.T.5    Poulsen, P.L.6
  • 39
    • 84925870139 scopus 로고    scopus 로고
    • Invasive validation of the Complior Analyse in the assessment of central artery pressure curves: A methodological study
    • Pereira T, Maldonado J, Coutinho R, Cardoso E, Laranjeiro M, Andrade I, Conde J. Invasive validation of the Complior Analyse in the assessment of central artery pressure curves: a methodological study. Blood Press Monit 2014; 19: 280-287.
    • (2014) Blood Press Monit , vol.19 , pp. 280-287
    • Pereira, T.1    Maldonado, J.2    Coutinho, R.3    Cardoso, E.4    Laranjeiro, M.5    Andrade, I.6    Conde, J.7
  • 40
    • 84899659435 scopus 로고    scopus 로고
    • Augmentation index does not reflect risk of coronary artery disease in elderly patients
    • Hayashi S, Yamada H, Bando M, Hotchi J, Ise T, Yamaguchi K, et al. Augmentation index does not reflect risk of coronary artery disease in elderly patients. Circ J 2014; 78: 1176-1182.
    • (2014) Circ J , vol.78 , pp. 1176-1182
    • Hayashi, S.1    Yamada, H.2    Bando, M.3    Hotchi, J.4    Ise, T.5    Yamaguchi, K.6
  • 41
    • 84926647870 scopus 로고    scopus 로고
    • Evaluation of a brachial cuff and suprasystolic waveform algorithm method to noninvasively derive central blood pressure
    • Costello BT, Schultz MG, Black JA, Sharman JE. Evaluation of a brachial cuff and suprasystolic waveform algorithm method to noninvasively derive central blood pressure. Am J Hypertens 2015; 28: 480-486.
    • (2015) Am J Hypertens , vol.28 , pp. 480-486
    • Costello, B.T.1    Schultz, M.G.2    Black, J.A.3    Sharman, J.E.4
  • 43
    • 0043268701 scopus 로고    scopus 로고
    • Clinical utility of aortic pulses and pressures calculated from applanated radial-Artery pulses
    • Smulyan H. Clinical utility of aortic pulses and pressures calculated from applanated radial-Artery pulses. Hypertension 2003; 42: 150-155.
    • (2003) Hypertension , vol.42 , pp. 150-155
    • Smulyan, H.1
  • 44
    • 81155133883 scopus 로고    scopus 로고
    • Quantification of the calibration error in the transfer function-derived central aortic blood pressures
    • Shih YT, Cheng HM, Sung SH, Hu WC, Chen CH. Quantification of the calibration error in the transfer function-derived central aortic blood pressures. Am J Hypertens 2011; 24: 1312-1317.
    • (2011) Am J Hypertens , vol.24 , pp. 1312-1317
    • Shih, Y.T.1    Cheng, H.M.2    Sung, S.H.3    Hu, W.C.4    Chen, C.H.5
  • 45
    • 33646729851 scopus 로고    scopus 로고
    • Transmission of calibration errors (input) by generalized transfer functions to the aortic pressures (output) at different hemodynamic states
    • Papaioannou TG, Lekakis JP, Karatzis EN, Papamichael CM, Stamatelopoulos KS, Protogerou AD, et al. Transmission of calibration errors (input) by generalized transfer functions to the aortic pressures (output) at different hemodynamic states. Int J Cardiol 2006; 110: 46-52.
    • (2006) Int J Cardiol , vol.110 , pp. 46-52
    • Papaioannou, T.G.1    Lekakis, J.P.2    Karatzis, E.N.3    Papamichael, C.M.4    Stamatelopoulos, K.S.5    Protogerou, A.D.6
  • 46
    • 84942626607 scopus 로고    scopus 로고
    • Brachial-To-radial systolic blood pressure amplification: Implications of age and estimated central blood pressure from radial tonometry
    • Picone DS CR, Ahuja KDK, Keske MA, Sharman JE. Brachial-To-radial systolic blood pressure amplification: implications of age and estimated central blood pressure from radial tonometry. J Hypertens 2015; 33: 1876-1883.
    • (2015) J Hypertens , vol.33 , pp. 1876-1883
    • Picone, D.S.C.R.1    Ahuja, K.D.K.2    Keske, M.A.3    Sharman, J.E.4
  • 47
    • 84942626600 scopus 로고    scopus 로고
    • Assessment of systolic aortic pressure and its association to all cause mortality critically depends on waveform calibration
    • Wassertheurer S, Baumann M. Assessment of systolic aortic pressure and its association to all cause mortality critically depends on waveform calibration. J Hypertens 2015; 33: 1884-1889.
    • (2015) J Hypertens , vol.33 , pp. 1884-1889
    • Wassertheurer, S.1    Baumann, M.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.