Volumn 59, Issue 6, 2014, Pages 599-614

Corneal assessment technologies: Current status

Author keywords

Confocal microscopy; Corneal morphology assessment; Dynamic applanation procedures; Optical coherence tomography (OCT); Orbscan topography; Placido disk; Scheimpflug imaging; Specular microscopy; Ultrasound pachymetry; Very high frequency (VHF) digital ultrasound biomicroscopy

Indexed keywords

ANALYZER; BIOMICROSCOPY; CONFOCAL MICROSCOPY; CORNEA EPITHELIUM; CORNEA THICKNESS; CORNEA TOPOGRAPHER; CORNEAL PACHYMETRY; DIAGNOSTIC PROCEDURE; EYE EXAMINATION; FLUORESCENCE IMAGING; FUNCTIONAL MORPHOLOGY; GALILEI DUAL SCHEIMPFLUG ANALYZER; HUMAN; KERATOSCOPY; OPHTHALMIC CAMERA; OPHTHALMOLOGICAL DIAGNOSTIC DEVICE; OPTICAL COHERENCE TOMOGRAPHY; REVIEW; SCHEIMPFLUG IMAGING; SLIT LAMP; SLIT SCANNING ELEVATION TOPOGRAPHY; SPECULAR MICROSCOPY; ACOUSTIC MICROSCOPY; CORNEA; CORNEAL DISEASES; KERATOMETRY; PATHOLOGY; PHOTOGRAPHY; PROCEDURES; VISUAL SYSTEM EXAMINATION;

EID: 84908211203     PISSN: 00396257     EISSN: 18793304     Source Type: Journal    
DOI: 10.1016/j.survophthal.2014.05.001     Document Type: Review
Times cited : (85)

References (121)
  • 1
    • 79960582849 scopus 로고    scopus 로고
    • Corneal endothelial cell loss in post-penetrating keratoplasty patients after cataract surgery: phacoemulsification versus planned extracapsular cataract extraction
    • Acar B.T., Buttanri I.B., Sevim M.S., et al. Corneal endothelial cell loss in post-penetrating keratoplasty patients after cataract surgery: phacoemulsification versus planned extracapsular cataract extraction. JCataract Refract Surg 2011, 37:1512-1516.
    • (2011) JCataract Refract Surg , vol.37 , pp. 1512-1516
    • Acar, B.T.1    Buttanri, I.B.2    Sevim, M.S.3
  • 2
    • 79959831017 scopus 로고    scopus 로고
    • Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter
    • Almubrad T.M., Osuagwu U.L., Alabbadi I., et al. Comparison of the precision of the Topcon SP-3000P specular microscope and an ultrasound pachymeter. Clin Ophthalmol 2011, 5:871-876.
    • (2011) Clin Ophthalmol , vol.5 , pp. 871-876
    • Almubrad, T.M.1    Osuagwu, U.L.2    Alabbadi, I.3
  • 3
    • 79955793147 scopus 로고    scopus 로고
    • Evaluation of corneal shape and biomechanics before LASIK
    • Ambrósio R., Nogueira L.P., Caldas D.L., et al. Evaluation of corneal shape and biomechanics before LASIK. Int Ophthalmol Clin 2011, 51:11-38.
    • (2011) Int Ophthalmol Clin , vol.51 , pp. 11-38
    • Ambrósio, R.1    Nogueira, L.P.2    Caldas, D.L.3
  • 4
    • 84873388798 scopus 로고    scopus 로고
    • Anterior chamber angle evaluation with fourier-domain optical coherence tomography
    • 103704
    • Bald M., Li Y., Huang D. Anterior chamber angle evaluation with fourier-domain optical coherence tomography. JOphthalmol 2012, 2012. 103704.
    • (2012) JOphthalmol , vol.2012
    • Bald, M.1    Li, Y.2    Huang, D.3
  • 5
    • 70449227246 scopus 로고
    • The application of ultrasonic locating techniques to ophthalmology. II. Ultrasonic slit-lamp in the ultrasonic visualization of soft tissues
    • Baum G., Greenwood I. The application of ultrasonic locating techniques to ophthalmology. II. Ultrasonic slit-lamp in the ultrasonic visualization of soft tissues. AMA Arch Ophthalmol 1958, 60:263-279.
    • (1958) AMA Arch Ophthalmol , vol.60 , pp. 263-279
    • Baum, G.1    Greenwood, I.2
  • 6
    • 63649104023 scopus 로고    scopus 로고
    • An introduction to understanding elevation-based topography: how elevation data are displayed - a review
    • Belin M.W., Khachikian S.S. An introduction to understanding elevation-based topography: how elevation data are displayed - a review. Clin Experiment Ophthalmol 2009, 37:14-29.
    • (2009) Clin Experiment Ophthalmol , vol.37 , pp. 14-29
    • Belin, M.W.1    Khachikian, S.S.2
  • 8
    • 0031425776 scopus 로고    scopus 로고
    • Long term contact lens wear induces a corneal degeneration with micro-dot deposits in the corneal stroma
    • Bohnke M., Masters B.R. Long term contact lens wear induces a corneal degeneration with micro-dot deposits in the corneal stroma. Ophthalmology 1997, 104:1887-1896.
    • (1997) Ophthalmology , vol.104 , pp. 1887-1896
    • Bohnke, M.1    Masters, B.R.2
  • 9
    • 70350570512 scopus 로고    scopus 로고
    • Average 3-dimensional models for the comparison of Orbscan II and Pentacam pachymetry maps in normal corneas
    • Bourges J.L., Alfonsi N., Laliberté J.F., et al. Average 3-dimensional models for the comparison of Orbscan II and Pentacam pachymetry maps in normal corneas. Ophthalmology 2009, 116:2064-2071.
    • (2009) Ophthalmology , vol.116 , pp. 2064-2071
    • Bourges, J.L.1    Alfonsi, N.2    Laliberté, J.F.3
  • 10
    • 77957605955 scopus 로고    scopus 로고
    • Corneal endothelium: past, present, and future
    • Bourne W.M. Corneal endothelium: past, present, and future. Eye Contact Lens 2010, 36:310-314.
    • (2010) Eye Contact Lens , vol.36 , pp. 310-314
    • Bourne, W.M.1
  • 11
    • 0842331154 scopus 로고    scopus 로고
    • Clinical responses of the corneal endothelium
    • Bourne W.M., McLaren J.W. Clinical responses of the corneal endothelium. Exp Eye Res 2004, 78:561-572.
    • (2004) Exp Eye Res , vol.78 , pp. 561-572
    • Bourne, W.M.1    McLaren, J.W.2
  • 12
    • 0028088105 scopus 로고
    • Corneal topography: history, technique and clinical uses
    • Brody J., Waller S., Wagoner M. Corneal topography: history, technique and clinical uses. Int Ophthalmol Clin 1994, 34:197-207.
    • (1994) Int Ophthalmol Clin , vol.34 , pp. 197-207
    • Brody, J.1    Waller, S.2    Wagoner, M.3
  • 13
    • 0026532536 scopus 로고
    • Specular microscopy in the identification of deep corneal opacities
    • Brooks A.M., Grant G., Gillies W.E. Specular microscopy in the identification of deep corneal opacities. Surv Ophthalmol 1992, 36:351-356.
    • (1992) Surv Ophthalmol , vol.36 , pp. 351-356
    • Brooks, A.M.1    Grant, G.2    Gillies, W.E.3
  • 14
    • 29544440497 scopus 로고    scopus 로고
    • Comparison of three methods of measuring corneal thickness and anterior chamber depth
    • Buehl W., Stojanac D., Sacu S., et al. Comparison of three methods of measuring corneal thickness and anterior chamber depth. Am J Ophthalmol 2006, 141:7-12.
    • (2006) Am J Ophthalmol , vol.141 , pp. 7-12
    • Buehl, W.1    Stojanac, D.2    Sacu, S.3
  • 15
    • 13944275573 scopus 로고    scopus 로고
    • Orbscan computerized topography: attributes, applications, and limitations
    • Cairns G., McGhee C.N.J. Orbscan computerized topography: attributes, applications, and limitations. JCataract Refract Surg 2005, 31:205-220.
    • (2005) JCataract Refract Surg , vol.31 , pp. 205-220
    • Cairns, G.1    McGhee, C.N.J.2
  • 16
    • 0033822579 scopus 로고    scopus 로고
    • Specular microscopy, confocal microscopy, and ultrasound biomicroscopy: diagnostic tools of the past quarter century
    • Cavanagh H.D., El-Agha M.S., Petroll W.M., Jester J.V. Specular microscopy, confocal microscopy, and ultrasound biomicroscopy: diagnostic tools of the past quarter century. Cornea 2000, 19:712-722.
    • (2000) Cornea , vol.19 , pp. 712-722
    • Cavanagh, H.D.1    El-Agha, M.S.2    Petroll, W.M.3    Jester, J.V.4
  • 17
    • 0035161383 scopus 로고    scopus 로고
    • Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes
    • Chakrabarti H.S., Craig J.P., Brahma A., et al. Comparison of corneal thickness measurements using ultrasound and Orbscan slit-scanning topography in normal and post-LASIK eyes. JCataract Refract Surg 2001, 27:1823-1828.
    • (2001) JCataract Refract Surg , vol.27 , pp. 1823-1828
    • Chakrabarti, H.S.1    Craig, J.P.2    Brahma, A.3
  • 18
    • 42149165577 scopus 로고    scopus 로고
    • Central corneal thickness measurements by ultrasound, Orbscan II, and Visante OCT after LASIK for myopia
    • Cheng A., Rao A.K., Lau S., et al. Central corneal thickness measurements by ultrasound, Orbscan II, and Visante OCT after LASIK for myopia. JRefrac Surg 2008, 24:361-365.
    • (2008) JRefrac Surg , vol.24 , pp. 361-365
    • Cheng, A.1    Rao, A.K.2    Lau, S.3
  • 19
    • 0038771128 scopus 로고    scopus 로고
    • Ocular toxicity of some corneal penetration enhancers evaluated by electrophysiology measurements on isolated rabbit corneas
    • Chetoni P., Burgalassi S., Monti D., et al. Ocular toxicity of some corneal penetration enhancers evaluated by electrophysiology measurements on isolated rabbit corneas. Toxicol InVitro 2003, 17:497-504.
    • (2003) Toxicol InVitro , vol.17 , pp. 497-504
    • Chetoni, P.1    Burgalassi, S.2    Monti, D.3
  • 20
    • 0842310253 scopus 로고    scopus 로고
    • Contribution of the ocular surface to visual optics
    • Courville C.B., Smolek M.K., Klyce S.D. Contribution of the ocular surface to visual optics. Exp Eye Res 2004, 78:417-425.
    • (2004) Exp Eye Res , vol.78 , pp. 417-425
    • Courville, C.B.1    Smolek, M.K.2    Klyce, S.D.3
  • 21
    • 78649372030 scopus 로고    scopus 로고
    • Invivo confocal microscopy of corneal nerves: analysis and clinical correlation
    • Cruzat A., Pavan-langston D., Hamrah P. Invivo confocal microscopy of corneal nerves: analysis and clinical correlation. Semin Ophthalmol 2010, 25:171-177.
    • (2010) Semin Ophthalmol , vol.25 , pp. 171-177
    • Cruzat, A.1    Pavan-langston, D.2    Hamrah, P.3
  • 22
    • 34547181745 scopus 로고    scopus 로고
    • Corneal epithelial stem cells in health and disease
    • Daniels J.T., Harris A.R., Mason C. Corneal epithelial stem cells in health and disease. Stem Cell Rev 2006, 2:247-254.
    • (2006) Stem Cell Rev , vol.2 , pp. 247-254
    • Daniels, J.T.1    Harris, A.R.2    Mason, C.3
  • 23
    • 77953653090 scopus 로고    scopus 로고
    • Value of optical coherence tomography for anterior segment surgery
    • Doors M., Berendschot T.T., de Brabander J., et al. Value of optical coherence tomography for anterior segment surgery. JCataract Refract Surg 2010, 36:1213-1229.
    • (2010) JCataract Refract Surg , vol.36 , pp. 1213-1229
    • Doors, M.1    Berendschot, T.T.2    de Brabander, J.3
  • 24
    • 77949448872 scopus 로고    scopus 로고
    • The orbscan acoustic (correction) factor for central corneal thickness measures of normal human corneas
    • Doughty M.J., Jonuscheit S. The orbscan acoustic (correction) factor for central corneal thickness measures of normal human corneas. Eye Contact Lens 2010, 36:106-115.
    • (2010) Eye Contact Lens , vol.36 , pp. 106-115
    • Doughty, M.J.1    Jonuscheit, S.2
  • 25
    • 0034603359 scopus 로고    scopus 로고
    • Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach
    • Doughty M.J., Zaman M.L. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol 2000, 44:367-408.
    • (2000) Surv Ophthalmol , vol.44 , pp. 367-408
    • Doughty, M.J.1    Zaman, M.L.2
  • 26
    • 43649098167 scopus 로고
    • Fundus photography by electronic flash III. A new fundus-anterior segment camera
    • Drews R.C. Fundus photography by electronic flash III. A new fundus-anterior segment camera. Am J Ophthalmol 1957, 44:522-525.
    • (1957) Am J Ophthalmol , vol.44 , pp. 522-525
    • Drews, R.C.1
  • 27
    • 0005679523 scopus 로고
    • Depth of field in slit image photography. An optical solution using the Scheimpflug principle
    • Drews R.C. Depth of field in slit image photography. An optical solution using the Scheimpflug principle. Ophthalmologica 1964, 148:143-150.
    • (1964) Ophthalmologica , vol.148 , pp. 143-150
    • Drews, R.C.1
  • 28
    • 40949107729 scopus 로고    scopus 로고
    • Multivariate model of refractive shift in Descemet-stripping automated endothe- lial keratoplasty
    • Dupps W.J., Qian Y., Meisler D.M. Multivariate model of refractive shift in Descemet-stripping automated endothe- lial keratoplasty. JCataract Refract Surg 2008, 34:578-584.
    • (2008) JCataract Refract Surg , vol.34 , pp. 578-584
    • Dupps, W.J.1    Qian, Y.2    Meisler, D.M.3
  • 29
    • 34248569309 scopus 로고    scopus 로고
    • Contact lens-induced changes in the anterior eye as observed invivo with the confocal microscope
    • Efron N. Contact lens-induced changes in the anterior eye as observed invivo with the confocal microscope. Prog Retin Eye Res 2007, 26:398-436.
    • (2007) Prog Retin Eye Res , vol.26 , pp. 398-436
    • Efron, N.1
  • 30
    • 79958849216 scopus 로고    scopus 로고
    • The Glenn A. Fry award lecture 2010: Ophthalmic markers of diabetic neuropathy
    • Efron N. The Glenn A. Fry award lecture 2010: Ophthalmic markers of diabetic neuropathy. Optom Vis Sci 2011, 88:661-683.
    • (2011) Optom Vis Sci , vol.88 , pp. 661-683
    • Efron, N.1
  • 33
    • 43649088112 scopus 로고    scopus 로고
    • Anterior chamber angle assessment techniques
    • Friedman D.S., He M. Anterior chamber angle assessment techniques. Surv Ophthalmol 2008, 53:250-273.
    • (2008) Surv Ophthalmol , vol.53 , pp. 250-273
    • Friedman, D.S.1    He, M.2
  • 34
    • 33751098788 scopus 로고    scopus 로고
    • Measurement of combined corneal, internal, and total ocular optical quality analysis in anterior segment pathology with the OPD-scan and OPD-station
    • Gatinel D., Hoang-Xuan T. Measurement of combined corneal, internal, and total ocular optical quality analysis in anterior segment pathology with the OPD-scan and OPD-station. JRefract Surg 2006, 22:1014-1020.
    • (2006) JRefract Surg , vol.22 , pp. 1014-1020
    • Gatinel, D.1    Hoang-Xuan, T.2
  • 35
    • 79952283349 scopus 로고    scopus 로고
    • Central corneal thickness measured with three optical devices and ultrasound pachometry
    • González-Pérez J., González-Méijome J.M., Rodríguez Ares M.T., et al. Central corneal thickness measured with three optical devices and ultrasound pachometry. Eye Contact Lens 2011, 37:66-70.
    • (2011) Eye Contact Lens , vol.37 , pp. 66-70
    • González-Pérez, J.1    González-Méijome, J.M.2    Rodríguez Ares, M.T.3
  • 36
    • 63849083776 scopus 로고    scopus 로고
    • The use of NIDEK OPD Scan II wavefront aberrometry in toric intraocular lens implantation
    • Gualdi L., Cappello V., Giordano C. The use of NIDEK OPD Scan II wavefront aberrometry in toric intraocular lens implantation. JRefract Surg 2009, 25:110-115.
    • (2009) JRefract Surg , vol.25 , pp. 110-115
    • Gualdi, L.1    Cappello, V.2    Giordano, C.3
  • 37
    • 84908207378 scopus 로고
    • Corneal topography and wavefront sensing
    • Elsevier India, Delhi, India, A.K. Gupta, V. Krishna (Eds.)
    • Gupta A.K., Krishna V. Corneal topography and wavefront sensing. Clinical Ophthalmology: Contemporary Perspectives 1993, 4-20. Elsevier India, Delhi, India. ed 1. A.K. Gupta, V. Krishna (Eds.).
    • (1993) Clinical Ophthalmology: Contemporary Perspectives , pp. 4-20
    • Gupta, A.K.1    Krishna, V.2
  • 38
    • 63649148290 scopus 로고    scopus 로고
    • Invivo confocal microscopy, an inner vision of the cornea - a major review
    • Guthoff R.F., Zhivov A., Stachs O. Invivo confocal microscopy, an inner vision of the cornea - a major review. Clin Experiment Ophthalmol 2009, 37:100-117.
    • (2009) Clin Experiment Ophthalmol , vol.37 , pp. 100-117
    • Guthoff, R.F.1    Zhivov, A.2    Stachs, O.3
  • 39
    • 63449103162 scopus 로고    scopus 로고
    • Pentacam and Orbscan II measurements of posterior corneal elevation before and after photorefractive keratectomy
    • Ha B.J., Kim S.W., Kim S.W., et al. Pentacam and Orbscan II measurements of posterior corneal elevation before and after photorefractive keratectomy. JRefract Surg 2009, 25:290-295.
    • (2009) JRefract Surg , vol.25 , pp. 290-295
    • Ha, B.J.1    Kim, S.W.2    Kim, S.W.3
  • 40
    • 73349102816 scopus 로고    scopus 로고
    • Midperipheral corneal thickness affects noncontact tonometry
    • Hamilton K. Midperipheral corneal thickness affects noncontact tonometry. JGlaucoma 2009, 18:623-627.
    • (2009) JGlaucoma , vol.18 , pp. 623-627
    • Hamilton, K.1
  • 41
    • 0041843611 scopus 로고    scopus 로고
    • Comparison of confocal biomicroscopy and non-contact specular microscopy for evaluation of the corneal endothelium
    • Hara M., Morishige N., Chikama T., et al. Comparison of confocal biomicroscopy and non-contact specular microscopy for evaluation of the corneal endothelium. Cornea 2003, 22:512-515.
    • (2003) Cornea , vol.22 , pp. 512-515
    • Hara, M.1    Morishige, N.2    Chikama, T.3
  • 42
    • 81355165893 scopus 로고    scopus 로고
    • Day to day clinically relevant corneal elevation, thickness, and curvature parameters using the orbscan II scanning slit topographer and the pentacam scheimpflug imaging device
    • Hashemi H., Mehravaran S. Day to day clinically relevant corneal elevation, thickness, and curvature parameters using the orbscan II scanning slit topographer and the pentacam scheimpflug imaging device. Middle East Afr J Ophthalmol 2010, 17:44-55.
    • (2010) Middle East Afr J Ophthalmol , vol.17 , pp. 44-55
    • Hashemi, H.1    Mehravaran, S.2
  • 43
    • 77955842202 scopus 로고    scopus 로고
    • The molecular basis of corneal transparency
    • Hassell J.R., Birk D.E. The molecular basis of corneal transparency. Exp Eye Res 2010, 91:326-335.
    • (2010) Exp Eye Res , vol.91 , pp. 326-335
    • Hassell, J.R.1    Birk, D.E.2
  • 44
    • 0024148769 scopus 로고
    • The specular microscope: its impact on laboratory and clinical studies of the cornea
    • Hodson S.A., Sherrard E.S. The specular microscope: its impact on laboratory and clinical studies of the cornea. Eye 1988, 2:81-97.
    • (1988) Eye , vol.2 , pp. 81-97
    • Hodson, S.A.1    Sherrard, E.S.2
  • 45
    • 17844366087 scopus 로고    scopus 로고
    • Correlation of the appearance of the keratoconic cornea invivo by confocal microscopy and invitro by light microscopy
    • Hollingsworth J.G., Bonshek R.E., Efron N. Correlation of the appearance of the keratoconic cornea invivo by confocal microscopy and invitro by light microscopy. Cornea 2005, 24:397-405.
    • (2005) Cornea , vol.24 , pp. 397-405
    • Hollingsworth, J.G.1    Bonshek, R.E.2    Efron, N.3
  • 46
    • 12444259908 scopus 로고
    • The cornea; permeability of the excised cornea to ions, as determined by measurements of impedance
    • Holt M., Cogan D.G. The cornea; permeability of the excised cornea to ions, as determined by measurements of impedance. Arch Ophthal 1946, 35:292-298.
    • (1946) Arch Ophthal , vol.35 , pp. 292-298
    • Holt, M.1    Cogan, D.G.2
  • 47
    • 0034068551 scopus 로고    scopus 로고
    • Quantitative assessment of the anterior segment using ultrasound biomicroscopy
    • Ishikawa H., Liebmann J.M., Ritch R. Quantitative assessment of the anterior segment using ultrasound biomicroscopy. Curr Opin Ophthalmol 2000, 11:133-139.
    • (2000) Curr Opin Ophthalmol , vol.11 , pp. 133-139
    • Ishikawa, H.1    Liebmann, J.M.2    Ritch, R.3
  • 48
    • 0027985314 scopus 로고
    • Micrometer-scale resolution imaging of the anterior eye invivo with optical coherence tomography
    • Izatt J.A., Hee M., Swanson E.A., et al. Micrometer-scale resolution imaging of the anterior eye invivo with optical coherence tomography. Arch Ophthalmol 1994, 112:1584-1589.
    • (1994) Arch Ophthalmol , vol.112 , pp. 1584-1589
    • Izatt, J.A.1    Hee, M.2    Swanson, E.A.3
  • 49
    • 84908214658 scopus 로고    scopus 로고
    • Pentacam: principle and clinical applications
    • Jain R., Grewal S. Pentacam: principle and clinical applications. Curr J Glaucoma Pract 2009, 3:20-32.
    • (2009) Curr J Glaucoma Pract , vol.3 , pp. 20-32
    • Jain, R.1    Grewal, S.2
  • 50
    • 71149091467 scopus 로고    scopus 로고
    • Discrepancy between central and midperipheral corneal thickness measurements obtained with slit-scanning pachymetry and noncontact specular microscopy
    • Jonuscheit S., Doughty M.J. Discrepancy between central and midperipheral corneal thickness measurements obtained with slit-scanning pachymetry and noncontact specular microscopy. JCataract Refract Surg 2009, 35:2127-2135.
    • (2009) JCataract Refract Surg , vol.35 , pp. 2127-2135
    • Jonuscheit, S.1    Doughty, M.J.2
  • 51
    • 33846236968 scopus 로고    scopus 로고
    • Regional repeatability measures of corneal thickness: Orbscan II and ultrasound
    • Jonuscheit S., Doughty M.J. Regional repeatability measures of corneal thickness: Orbscan II and ultrasound. Optom Vis Sci 2007, 84:52-58.
    • (2007) Optom Vis Sci , vol.84 , pp. 52-58
    • Jonuscheit, S.1    Doughty, M.J.2
  • 52
    • 33847225605 scopus 로고    scopus 로고
    • On the use of Orbscan II to assess the peripheral corneal thickness in humans: a comparison with ultrasound pachometry measures
    • Jonuscheit S., Doughty M.J., Button N.F. On the use of Orbscan II to assess the peripheral corneal thickness in humans: a comparison with ultrasound pachometry measures. Ophthalmic Physiol Opt 2007, 27:179-189.
    • (2007) Ophthalmic Physiol Opt , vol.27 , pp. 179-189
    • Jonuscheit, S.1    Doughty, M.J.2    Button, N.F.3
  • 53
    • 0029848304 scopus 로고    scopus 로고
    • Electrical impedance tomography of the eye: invitro measurements of the cornea and the lens
    • Jürgens I., Rosell J., Riu P.J. Electrical impedance tomography of the eye: invitro measurements of the cornea and the lens. Physiol Meas 1996, 17:187-195.
    • (1996) Physiol Meas , vol.17 , pp. 187-195
    • Jürgens, I.1    Rosell, J.2    Riu, P.J.3
  • 54
    • 84863428442 scopus 로고    scopus 로고
    • Correlation between epithelial thickness in normal corneas, untreated ectatic corneas, and ectatic corneas previously treated with CXL; is overall epithelial thickness a very early ectasia prognostic factor?
    • Kanellopoulos A.J., Aslanides I.M., Asimellis G. Correlation between epithelial thickness in normal corneas, untreated ectatic corneas, and ectatic corneas previously treated with CXL; is overall epithelial thickness a very early ectasia prognostic factor?. Clin Ophthalmol 2012, 6:789-800.
    • (2012) Clin Ophthalmol , vol.6 , pp. 789-800
    • Kanellopoulos, A.J.1    Aslanides, I.M.2    Asimellis, G.3
  • 55
    • 28044469557 scopus 로고    scopus 로고
    • Comparison of corneal endothelial cell images from a noncontact specular microscope and a scanning confocal microscope
    • Kitzmann A.S., Winter E.J., Nau C.B., et al. Comparison of corneal endothelial cell images from a noncontact specular microscope and a scanning confocal microscope. Cornea 2005, 24:980-984.
    • (2005) Cornea , vol.24 , pp. 980-984
    • Kitzmann, A.S.1    Winter, E.J.2    Nau, C.B.3
  • 56
    • 0037280735 scopus 로고    scopus 로고
    • Comparison of endothelial cell count using confocal and contact specular microscopy
    • Klais C.M., Bühren J., Kohnen T. Comparison of endothelial cell count using confocal and contact specular microscopy. Ophthalmologica 2003, 217:99-103.
    • (2003) Ophthalmologica , vol.217 , pp. 99-103
    • Klais, C.M.1    Bühren, J.2    Kohnen, T.3
  • 57
    • 0021716134 scopus 로고
    • Computer-assisted corneal topography. High-resolution graphic presentation and analysis of keratoscopy
    • Klyce S.D. Computer-assisted corneal topography. High-resolution graphic presentation and analysis of keratoscopy. Invest Ophthalmol Vis Sci 1984, 25:1426-1435.
    • (1984) Invest Ophthalmol Vis Sci , vol.25 , pp. 1426-1435
    • Klyce, S.D.1
  • 58
    • 77950691942 scopus 로고    scopus 로고
    • The architecture of the cornea and structural basis of its transparency
    • Knupp C., Pinali C., Lewis P.N., et al. The architecture of the cornea and structural basis of its transparency. Adv Protein Chem Struct Biol 2009, 78:25-49.
    • (2009) Adv Protein Chem Struct Biol , vol.78 , pp. 25-49
    • Knupp, C.1    Pinali, C.2    Lewis, P.N.3
  • 59
    • 34247239221 scopus 로고    scopus 로고
    • Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?
    • Konstantopoulos A., Hossain P., Anderson D.F. Recent advances in ophthalmic anterior segment imaging: a new era for ophthalmic diagnosis?. Br J Ophthalmol 2007, 91:551-557.
    • (2007) Br J Ophthalmol , vol.91 , pp. 551-557
    • Konstantopoulos, A.1    Hossain, P.2    Anderson, D.F.3
  • 60
    • 34248347895 scopus 로고    scopus 로고
    • Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer
    • Kotecha A., Elsheikh A., Roberts C.R., et al. Corneal thickness- and age-related biomechanical properties of the cornea measured with the ocular response analyzer. Invest Ophthalmol Vis Sci 2006, 47:5337-5347.
    • (2006) Invest Ophthalmol Vis Sci , vol.47 , pp. 5337-5347
    • Kotecha, A.1    Elsheikh, A.2    Roberts, C.R.3
  • 61
    • 0026090384 scopus 로고
    • Distinguishing features of the iridocorneal endothelial syndrome and posterior polymorphous dystrophy: value of endothelial specular microscopy
    • Laganowski H.C., Sherrard E.S., Muir M.G., et al. Distinguishing features of the iridocorneal endothelial syndrome and posterior polymorphous dystrophy: value of endothelial specular microscopy. Br J Ophthalmol 1991, 75:212-216.
    • (1991) Br J Ophthalmol , vol.75 , pp. 212-216
    • Laganowski, H.C.1    Sherrard, E.S.2    Muir, M.G.3
  • 62
    • 79952807351 scopus 로고    scopus 로고
    • Central corneal thickness and corneal endothelial cell changes caused by contact lens use in diabetic patients
    • Leem H.S., Lee K.J., Shin K.C. Central corneal thickness and corneal endothelial cell changes caused by contact lens use in diabetic patients. Yonsei Med J 2011, 52:322-325.
    • (2011) Yonsei Med J , vol.52 , pp. 322-325
    • Leem, H.S.1    Lee, K.J.2    Shin, K.C.3
  • 63
    • 33646143320 scopus 로고    scopus 로고
    • Corneal pachymetry mapping with high-speed optical coherence tomography
    • Li Y., Shekhar R., Huang D. Corneal pachymetry mapping with high-speed optical coherence tomography. Ophthalmology 2006, 113:792-799.
    • (2006) Ophthalmology , vol.113 , pp. 792-799
    • Li, Y.1    Shekhar, R.2    Huang, D.3
  • 64
    • 77951960898 scopus 로고    scopus 로고
    • Pachymetric mapping with Fourier-domain optical coherence tomography
    • Li Y., Tang M., Zhang X., et al. Pachymetric mapping with Fourier-domain optical coherence tomography. JCataract Refract Surg 2010, 36:826-831.
    • (2010) JCataract Refract Surg , vol.36 , pp. 826-831
    • Li, Y.1    Tang, M.2    Zhang, X.3
  • 65
    • 0019857035 scopus 로고
    • Electrical properties of rabbit corneal endothelium as determined from impedance measurements
    • Lim J.J., Fischbarg J. Electrical properties of rabbit corneal endothelium as determined from impedance measurements. Biophys J 1981, 36:677-695.
    • (1981) Biophys J , vol.36 , pp. 677-695
    • Lim, J.J.1    Fischbarg, J.2
  • 66
    • 0033999943 scopus 로고    scopus 로고
    • The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularity
    • Liu Z., Pflugfelder S.C. The effects of long-term contact lens wear on corneal thickness, curvature, and surface regularity. Ophthalmology 2000, 107:105-111.
    • (2000) Ophthalmology , vol.107 , pp. 105-111
    • Liu, Z.1    Pflugfelder, S.C.2
  • 67
    • 13844317411 scopus 로고    scopus 로고
    • Determining invivo biomechanical properties of the cornea with an ocular response analyzer
    • Luce D.A. Determining invivo biomechanical properties of the cornea with an ocular response analyzer. JCataract Refract Surg 2005, 31:156-162.
    • (2005) JCataract Refract Surg , vol.31 , pp. 156-162
    • Luce, D.A.1
  • 68
    • 84875430254 scopus 로고    scopus 로고
    • White-to-white corneal diameter differences in moderately and highly myopic eyes: partial coherence interferometry versus scanning-slit topography
    • Martin R., Ortiz S., Rio-San Cristobal A. White-to-white corneal diameter differences in moderately and highly myopic eyes: partial coherence interferometry versus scanning-slit topography. JCataract Refract Surg 2013, 39:585-589.
    • (2013) JCataract Refract Surg , vol.39 , pp. 585-589
    • Martin, R.1    Ortiz, S.2    Rio-San Cristobal, A.3
  • 69
    • 34548077624 scopus 로고    scopus 로고
    • Measurement of corneal swelling variations without removal of the contact lens during extended wear
    • Martin R., de Juan V., Rodríguez G., et al. Measurement of corneal swelling variations without removal of the contact lens during extended wear. Invest Ophthalmol Vis Sci 2007, 48:3043-3050.
    • (2007) Invest Ophthalmol Vis Sci , vol.48 , pp. 3043-3050
    • Martin, R.1    de Juan, V.2    Rodríguez, G.3
  • 70
    • 67449094277 scopus 로고    scopus 로고
    • Contact lens-induced corneal peripheral swelling: Orbscan repeatability
    • Martin R., de Juan V., Rodriguez G., et al. Contact lens-induced corneal peripheral swelling: Orbscan repeatability. Optom Vis Sci 2009, 86:340-349.
    • (2009) Optom Vis Sci , vol.86 , pp. 340-349
    • Martin, R.1    de Juan, V.2    Rodriguez, G.3
  • 71
    • 70449527662 scopus 로고    scopus 로고
    • Investigation of posterior corneal curvature in CL-induced corneal swelling
    • Martin R., Izquierdo M., Saber A. Investigation of posterior corneal curvature in CL-induced corneal swelling. Cont Lens Anterior Eye 2009, 32:288-293.
    • (2009) Cont Lens Anterior Eye , vol.32 , pp. 288-293
    • Martin, R.1    Izquierdo, M.2    Saber, A.3
  • 72
    • 79959699664 scopus 로고    scopus 로고
    • Constancy of the Orbscan acoustic factor to detect contact lens-induced corneal swelling
    • Martin R., Nuñez L., Sastre J., et al. Constancy of the Orbscan acoustic factor to detect contact lens-induced corneal swelling. Clin Exp Optom 2011, 94:352-360.
    • (2011) Clin Exp Optom , vol.94 , pp. 352-360
    • Martin, R.1    Nuñez, L.2    Sastre, J.3
  • 73
    • 84857797235 scopus 로고    scopus 로고
    • Stability of posterior corneal elevation one year after myopic laser in situ keratomileusis
    • Martin R., Rachidi H. Stability of posterior corneal elevation one year after myopic laser in situ keratomileusis. Clin Exp Optom 2012, 95:177-186.
    • (2012) Clin Exp Optom , vol.95 , pp. 177-186
    • Martin, R.1    Rachidi, H.2
  • 74
    • 84908210877 scopus 로고    scopus 로고
    • Exploración del polo anterior
    • Editorial Médica Panamericana, Madrid, Spain, R. Martin, G. Vecilla (Eds.)
    • Martin R., Vecilla G. Exploración del polo anterior. Manual de Optometría 2010, 313-330. Editorial Médica Panamericana, Madrid, Spain. ed 1. R. Martin, G. Vecilla (Eds.).
    • (2010) Manual de Optometría , pp. 313-330
    • Martin, R.1    Vecilla, G.2
  • 75
    • 0014418794 scopus 로고
    • Cellular membrane activity in the corneal endothelium of the intact eye
    • Maurice D.M. Cellular membrane activity in the corneal endothelium of the intact eye. Experientia 1968, 24:1094-1095.
    • (1968) Experientia , vol.24 , pp. 1094-1095
    • Maurice, D.M.1
  • 76
    • 40449133628 scopus 로고    scopus 로고
    • Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions
    • McCarey B.E., Edelhauser H.F., Lynn M.J. Review of corneal endothelial specular microscopy for FDA clinical trials of refractive procedures, surgical devices, and new intraocular drugs and solutions. Cornea 2008, 27:1-16.
    • (2008) Cornea , vol.27 , pp. 1-16
    • McCarey, B.E.1    Edelhauser, H.F.2    Lynn, M.J.3
  • 77
    • 2942623857 scopus 로고    scopus 로고
    • Corneal thickness measurement by confocal microscopy, ultrasound, and scanning slit methods
    • McLaren J.W., Nau C.B., Erie J.C., et al. Corneal thickness measurement by confocal microscopy, ultrasound, and scanning slit methods. Am J Ophthalmol 2004, 137:1011-1020.
    • (2004) Am J Ophthalmol , vol.137 , pp. 1011-1020
    • McLaren, J.W.1    Nau, C.B.2    Erie, J.C.3
  • 78
    • 33646762089 scopus 로고    scopus 로고
    • Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode
    • Meinhardt B., Stachs O., Stave J., et al. Evaluation of biometric methods for measuring the anterior chamber depth in the non-contact mode. Graefes Arch Clin Exp Ophthalmol 2006, 244:559-564.
    • (2006) Graefes Arch Clin Exp Ophthalmol , vol.244 , pp. 559-564
    • Meinhardt, B.1    Stachs, O.2    Stave, J.3
  • 79
    • 14644436938 scopus 로고
    • Corneal measurements and tests
    • W.B. Saunders Company, Philadelphia, PA, D.M. Albert, F.A. Jakobiec (Eds.)
    • Miller D., Greiner J.V. Corneal measurements and tests. Principles and Practice of Ophthalmology: Clinical Practice 1994, 4-13. W.B. Saunders Company, Philadelphia, PA. ed 1. D.M. Albert, F.A. Jakobiec (Eds.).
    • (1994) Principles and Practice of Ophthalmology: Clinical Practice , pp. 4-13
    • Miller, D.1    Greiner, J.V.2
  • 80
    • 84983965438 scopus 로고    scopus 로고
    • Memoir on inventing the confocal scanning microscope
    • Minsky M. Memoir on inventing the confocal scanning microscope. Scanning 1998, 10:128-138.
    • (1998) Scanning , vol.10 , pp. 128-138
    • Minsky, M.1
  • 81
    • 0000297590 scopus 로고
    • Ultrasonics in ocular diagnosis
    • Mundt G.H., Hughes W.F. Ultrasonics in ocular diagnosis. Am J Ophthalmol 1956, 41:488-498.
    • (1956) Am J Ophthalmol , vol.41 , pp. 488-498
    • Mundt, G.H.1    Hughes, W.F.2
  • 82
    • 79958144485 scopus 로고    scopus 로고
    • Comparison of anterior chamber depth measurements conducted with Pentacam HR and IOL Master
    • Nemeth G., Hassan Z., Modis L., et al. Comparison of anterior chamber depth measurements conducted with Pentacam HR and IOL Master. Ophthalmic Surg Lasers Imaging 2011, 42:144-147.
    • (2011) Ophthalmic Surg Lasers Imaging , vol.42 , pp. 144-147
    • Nemeth, G.1    Hassan, Z.2    Modis, L.3
  • 83
    • 74749103535 scopus 로고    scopus 로고
    • Clinical invivo confocal microscopy of the human cornea in health and disease
    • Niederer R.L., McGhee C.N.J. Clinical invivo confocal microscopy of the human cornea in health and disease. Prog Retin Eye Res 2010, 29:30-58.
    • (2010) Prog Retin Eye Res , vol.29 , pp. 30-58
    • Niederer, R.L.1    McGhee, C.N.J.2
  • 84
    • 84863467491 scopus 로고    scopus 로고
    • Repeatability and interobsever reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness
    • Ogbuehi K.C., Osuagwu U.L. Repeatability and interobsever reproducibility of Artemis-2 high-frequency ultrasound in determination of human corneal thickness. Clin Ophthalmol 2012, 6:761-769.
    • (2012) Clin Ophthalmol , vol.6 , pp. 761-769
    • Ogbuehi, K.C.1    Osuagwu, U.L.2
  • 85
    • 78650709901 scopus 로고    scopus 로고
    • Corneal imaging with slit-scanning and Scheimpflug imaging techniques
    • Oliveira C.M., Ribeiro C., Franco S. Corneal imaging with slit-scanning and Scheimpflug imaging techniques. Clin Exp Optom 2011, 94:33-42.
    • (2011) Clin Exp Optom , vol.94 , pp. 33-42
    • Oliveira, C.M.1    Ribeiro, C.2    Franco, S.3
  • 86
    • 33645336030 scopus 로고    scopus 로고
    • Tandem scanning confocal corneal microscopy in the diagnosis of suspected acanthamoeba keratitis
    • Parmar D.N., Awwad S.T., Petroll W.M., et al. Tandem scanning confocal corneal microscopy in the diagnosis of suspected acanthamoeba keratitis. Ophthalmology 2006, 113:538-547.
    • (2006) Ophthalmology , vol.113 , pp. 538-547
    • Parmar, D.N.1    Awwad, S.T.2    Petroll, W.M.3
  • 87
    • 33646015096 scopus 로고    scopus 로고
    • Mapping the corneal sub-basal nerve plexus in keratoconus by invivo laser scanning confocal microscopy
    • Patel D.V., McGhee C.N. Mapping the corneal sub-basal nerve plexus in keratoconus by invivo laser scanning confocal microscopy. Invest Ophthalmol Vis Sci 2006, 47:1348-1351.
    • (2006) Invest Ophthalmol Vis Sci , vol.47 , pp. 1348-1351
    • Patel, D.V.1    McGhee, C.N.2
  • 88
    • 33846707654 scopus 로고    scopus 로고
    • Contemporary invivo confocal microscopy of the living human cornea using white light and laser scanning techniques: a major review
    • Patel D.V., McGhee C.N. Contemporary invivo confocal microscopy of the living human cornea using white light and laser scanning techniques: a major review. Clin Experiment Ophthalmol 2007, 35:71-88.
    • (2007) Clin Experiment Ophthalmol , vol.35 , pp. 71-88
    • Patel, D.V.1    McGhee, C.N.2
  • 89
    • 0035146191 scopus 로고    scopus 로고
    • Normal human keratocyte density and corneal thickness measurement by using confocal microscopy invivo
    • Patel S., McLaren J., Hodge D., et al. Normal human keratocyte density and corneal thickness measurement by using confocal microscopy invivo. Invest Ophthalmol Vis Sci 2001, 42:333-339.
    • (2001) Invest Ophthalmol Vis Sci , vol.42 , pp. 333-339
    • Patel, S.1    McLaren, J.2    Hodge, D.3
  • 90
    • 10844270390 scopus 로고    scopus 로고
    • Corneal cell density measurement invivo by scanning slit confocal microscopy: method and validation
    • Popper M., Morgado A.M., Quadrado M.J., et al. Corneal cell density measurement invivo by scanning slit confocal microscopy: method and validation. Ophthalmic Res 2004, 36:270-276.
    • (2004) Ophthalmic Res , vol.36 , pp. 270-276
    • Popper, M.1    Morgado, A.M.2    Quadrado, M.J.3
  • 91
    • 67249126267 scopus 로고    scopus 로고
    • Central and peripheral corneal thickness measured with optical coherence tomography, scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes
    • Prospero Ponce C.M., Rocha K.M., Smith S.D., et al. Central and peripheral corneal thickness measured with optical coherence tomography, scheimpflug imaging, and ultrasound pachymetry in normal, keratoconus-suspect, and post-laser in situ keratomileusis eyes. JCataract Refract Surg 2009, 35:1055-1062.
    • (2009) JCataract Refract Surg , vol.35 , pp. 1055-1062
    • Prospero Ponce, C.M.1    Rocha, K.M.2    Smith, S.D.3
  • 92
    • 73149125178 scopus 로고    scopus 로고
    • Corneal transparency: genesis, maintenance and dysfunction
    • Qazi Y., Wong G., Monson B., et al. Corneal transparency: genesis, maintenance and dysfunction. Brain Res Bull 2010, 81:198-210.
    • (2010) Brain Res Bull , vol.81 , pp. 198-210
    • Qazi, Y.1    Wong, G.2    Monson, B.3
  • 93
    • 63649129753 scopus 로고    scopus 로고
    • Clinical and research applications of anterior segment optical coherence tomography-a review
    • Ramos J.L.B., Li Y., Huang D. Clinical and research applications of anterior segment optical coherence tomography-a review. Clin Experiment Ophthalmol 2009, 37:81-89.
    • (2009) Clin Experiment Ophthalmol , vol.37 , pp. 81-89
    • Ramos, J.L.B.1    Li, Y.2    Huang, D.3
  • 94
    • 33751033654 scopus 로고    scopus 로고
    • Combined Artemis very high-frequency digital ultrasound-assisted transepithelial phototherapeutic keratectomy and wavefront-guided treatment following multiple corneal refractive procedures
    • Reinstein D.Z., Archer T.J. Combined Artemis very high-frequency digital ultrasound-assisted transepithelial phototherapeutic keratectomy and wavefront-guided treatment following multiple corneal refractive procedures. JCataract Refract Surg 2006, 32:1870-1876.
    • (2006) JCataract Refract Surg , vol.32 , pp. 1870-1876
    • Reinstein, D.Z.1    Archer, T.J.2
  • 95
    • 67651036588 scopus 로고    scopus 로고
    • Corneal epithelial thickness profile in the diagnosis of keratoconus
    • Reinstein D.Z., Archer T.J., Gobbe M. Corneal epithelial thickness profile in the diagnosis of keratoconus. JRefract Surg 2009, 25:604-610.
    • (2009) JRefract Surg , vol.25 , pp. 604-610
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 96
    • 84882295326 scopus 로고    scopus 로고
    • Improved effectiveness of transepithelial PTK versus topography-guided ablation for stromal irregularities masked by epithelial compensation
    • Reinstein D.Z., Archer T.J., Gobbe M. Improved effectiveness of transepithelial PTK versus topography-guided ablation for stromal irregularities masked by epithelial compensation. JRefract Surg 2013, 29:526-533.
    • (2013) JRefract Surg , vol.29 , pp. 526-533
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 97
    • 79960167573 scopus 로고    scopus 로고
    • LASIK flap thickness profile and reproducibility of the standard vs zero compression Hansatome microkeratomes: three-dimensional display with Artemis VHF digital ultrasound
    • Reinstein D.Z., Archer T.J., Gobbe M. LASIK flap thickness profile and reproducibility of the standard vs zero compression Hansatome microkeratomes: three-dimensional display with Artemis VHF digital ultrasound. JRefract Surg 2011, 27:417-426.
    • (2011) JRefract Surg , vol.27 , pp. 417-426
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 98
    • 84908167294 scopus 로고    scopus 로고
    • Very high-frequency digital ultrasound biomicroscopy
    • Elsevier, Philadelphia, USA, A. Singh, B. Hayden (Eds.)
    • Reinstein D.Z., Archer T.J., Gobbe M., et al. Very high-frequency digital ultrasound biomicroscopy. Ophthalmic Ultrasonography 2012, 43-62. Elsevier, Philadelphia, USA. ed 1. A. Singh, B. Hayden (Eds.).
    • (2012) Ophthalmic Ultrasonography , pp. 43-62
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 99
    • 78449242405 scopus 로고    scopus 로고
    • Epithelial thickness after hyperopic LASIK: three-dimensional display with Artemis very high-frequency digital ultrasound
    • Reinstein D.Z., Archer T.J., Gobbe M., et al. Epithelial thickness after hyperopic LASIK: three-dimensional display with Artemis very high-frequency digital ultrasound. JRefract Surg 2010, 26:555-564.
    • (2010) JRefract Surg , vol.26 , pp. 555-564
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 100
    • 45249084956 scopus 로고    scopus 로고
    • Epithelial thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound
    • Reinstein D.Z., Archer T.J., Gobbe M., et al. Epithelial thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound. JRefract Surg 2008, 24:571-581.
    • (2008) JRefract Surg , vol.24 , pp. 571-581
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 101
    • 77957669157 scopus 로고    scopus 로고
    • Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner
    • Reinstein D.Z., Archer T.J., Gobbe M., et al. Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner. JRefract Surg 2010, 26:646-659.
    • (2010) JRefract Surg , vol.26 , pp. 646-659
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 102
    • 70350125378 scopus 로고    scopus 로고
    • Stromal thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound
    • Reinstein D.Z., Archer T.J., Gobbe M., et al. Stromal thickness in the normal cornea: three-dimensional display with Artemis very high-frequency digital ultrasound. JRefract Surg 2009, 25:776-786.
    • (2009) JRefract Surg , vol.25 , pp. 776-786
    • Reinstein, D.Z.1    Archer, T.J.2    Gobbe, M.3
  • 103
    • 60849115870 scopus 로고    scopus 로고
    • Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using artemis VHF digital ultrasound
    • Reinstein D.Z., Archer T.J., Silverman R.H., et al. Correlation of anterior chamber angle and ciliary sulcus diameters with white-to-white corneal diameter in high myopes using artemis VHF digital ultrasound. JRefract Surg 2009, 25:185-194.
    • (2009) JRefract Surg , vol.25 , pp. 185-194
    • Reinstein, D.Z.1    Archer, T.J.2    Silverman, R.H.3
  • 104
    • 33751039313 scopus 로고    scopus 로고
    • Direct residual stromal thickness measurement for assessing suitability for LASIK enhancement by Artemis 3D very high-frequency digital ultrasound arc scanning
    • Reinstein D.Z., Couch D.G., Archer T. Direct residual stromal thickness measurement for assessing suitability for LASIK enhancement by Artemis 3D very high-frequency digital ultrasound arc scanning. JCataract Refract Surg 2006, 32:1884-1888.
    • (2006) JCataract Refract Surg , vol.32 , pp. 1884-1888
    • Reinstein, D.Z.1    Couch, D.G.2    Archer, T.3
  • 105
    • 84864518897 scopus 로고    scopus 로고
    • Anterior segment biometry: a study and review of resolution and repeatability data
    • Reinstein D.Z., Gobbe M., Archer T.J. Anterior segment biometry: a study and review of resolution and repeatability data. JRefract Surg 2012, 28:509-520.
    • (2012) JRefract Surg , vol.28 , pp. 509-520
    • Reinstein, D.Z.1    Gobbe, M.2    Archer, T.J.3
  • 106
    • 0033873480 scopus 로고    scopus 로고
    • Arc-scanning very high-frequency digital ultrasound for 3D pachymetric mapping of the corneal epithelium and stroma in laser in situ keratomileusis
    • Reinstein D.Z., Silverman R.H., Raevsky T., et al. Arc-scanning very high-frequency digital ultrasound for 3D pachymetric mapping of the corneal epithelium and stroma in laser in situ keratomileusis. JRefract Surg 2000, 16:414-430.
    • (2000) JRefract Surg , vol.16 , pp. 414-430
    • Reinstein, D.Z.1    Silverman, R.H.2    Raevsky, T.3
  • 107
    • 33751118002 scopus 로고    scopus 로고
    • Probability model of the inaccuracy of residual stromal thickness prediction to reduce the risk of ectasia after LASIK part I: quantifying individual risk
    • Reinstein D.Z., Srivannaboon S., Archer T.J., et al. Probability model of the inaccuracy of residual stromal thickness prediction to reduce the risk of ectasia after LASIK part I: quantifying individual risk. JRefract Surg 2006, 22:851-860.
    • (2006) JRefract Surg , vol.22 , pp. 851-860
    • Reinstein, D.Z.1    Srivannaboon, S.2    Archer, T.J.3
  • 108
    • 4644258013 scopus 로고    scopus 로고
    • Very high frequency ultrasound biometry of the anterior and posterior chamber diameter
    • Rondeau M.J., Barcsay G., Silverman R.H., et al. Very high frequency ultrasound biometry of the anterior and posterior chamber diameter. JRefract Surg 2004, 20:454-464.
    • (2004) JRefract Surg , vol.20 , pp. 454-464
    • Rondeau, M.J.1    Barcsay, G.2    Silverman, R.H.3
  • 109
    • 66149192417 scopus 로고    scopus 로고
    • Pentacam Scheimpflug quantitative imaging of the crystalline lens and intraocular lens
    • Rosales P., Marcos S. Pentacam Scheimpflug quantitative imaging of the crystalline lens and intraocular lens. JRefract Surg 2009, 25:421-428.
    • (2009) JRefract Surg , vol.25 , pp. 421-428
    • Rosales, P.1    Marcos, S.2
  • 110
    • 79952937094 scopus 로고    scopus 로고
    • Repeatability of laser in situ keratomileusis flap thickness measurement by Fourier-domain optical coherence tomography
    • Rosas Salaroli C.H., Li Y., Zhang X., et al. Repeatability of laser in situ keratomileusis flap thickness measurement by Fourier-domain optical coherence tomography. JCataract Refract Surg 2011, 37:649-654.
    • (2011) JCataract Refract Surg , vol.37 , pp. 649-654
    • Rosas Salaroli, C.H.1    Li, Y.2    Zhang, X.3
  • 112
    • 67349170031 scopus 로고    scopus 로고
    • The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery
    • Shah S., Laiquzzaman M., Yeung I., et al. The use of the ocular response analyser to determine corneal hysteresis in eyes before and after excimer laser refractive surgery. Cont Lens Anterior Eye 2009, 32:123-128.
    • (2009) Cont Lens Anterior Eye , vol.32 , pp. 123-128
    • Shah, S.1    Laiquzzaman, M.2    Yeung, I.3
  • 113
    • 0030274740 scopus 로고    scopus 로고
    • Confocal invivo microscopy and confocal laser-scanning fluorescence microscopy in keratoconus
    • Somodi S., Hahnel C., Slowik C., et al. Confocal invivo microscopy and confocal laser-scanning fluorescence microscopy in keratoconus. Ger J Ophthalmol 1996, 5:518-525.
    • (1996) Ger J Ophthalmol , vol.5 , pp. 518-525
    • Somodi, S.1    Hahnel, C.2    Slowik, C.3
  • 114
    • 78449250993 scopus 로고    scopus 로고
    • Interchangeability between Placido disc and Scheimpflug system: quantitative and qualitative analysis
    • Stefano V.S., Melo Junior L.A., Mallmann F., et al. Interchangeability between Placido disc and Scheimpflug system: quantitative and qualitative analysis. Arq Bras Oftalmol 2010, 73:363-366.
    • (2010) Arq Bras Oftalmol , vol.73 , pp. 363-366
    • Stefano, V.S.1    Melo Junior, L.A.2    Mallmann, F.3
  • 115
    • 34250698061 scopus 로고    scopus 로고
    • Measuring the cornea: the latest developments in corneal topography
    • Swartz T., Marten L., Wang M. Measuring the cornea: the latest developments in corneal topography. Curr Opin Ophthalmol 2007, 18:325-333.
    • (2007) Curr Opin Ophthalmol , vol.18 , pp. 325-333
    • Swartz, T.1    Marten, L.2    Wang, M.3
  • 116
    • 84863386026 scopus 로고    scopus 로고
    • Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography
    • Tang M., Wang L., Koch D.D., et al. Intraocular lens power calculation after previous myopic laser vision correction based on corneal power measured by Fourier-domain optical coherence tomography. JCataract Refract Surg 2012, 38:589-594.
    • (2012) JCataract Refract Surg , vol.38 , pp. 589-594
    • Tang, M.1    Wang, L.2    Koch, D.D.3
  • 117
    • 36949028300 scopus 로고    scopus 로고
    • Acute corneal epithelial change after instillation of benzalkonium chloride evaluated using a newly developed invivo corneal transepithelial electric resistance measurement method
    • Uematsu M., Kumagami T., Kusano M., et al. Acute corneal epithelial change after instillation of benzalkonium chloride evaluated using a newly developed invivo corneal transepithelial electric resistance measurement method. Ophthalmic Res 2007, 39:308-314.
    • (2007) Ophthalmic Res , vol.39 , pp. 308-314
    • Uematsu, M.1    Kumagami, T.2    Kusano, M.3
  • 118
    • 84873545875 scopus 로고    scopus 로고
    • Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles
    • Ursea R., Feng M., Urs R., et al. Comparison of artemis 2 ultrasound and Visante optical coherence tomography corneal thickness profiles. JRefract Surg 2013, 29:36-41.
    • (2013) JRefract Surg , vol.29 , pp. 36-41
    • Ursea, R.1    Feng, M.2    Urs, R.3
  • 119
    • 63649088722 scopus 로고    scopus 로고
    • Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations - a review
    • Wegener A., Laser-Junga H. Photography of the anterior eye segment according to Scheimpflug's principle: options and limitations - a review. Clin Experiment Ophthalmol 2009, 37:144-154.
    • (2009) Clin Experiment Ophthalmol , vol.37 , pp. 144-154
    • Wegener, A.1    Laser-Junga, H.2
  • 120
    • 58349120605 scopus 로고    scopus 로고
    • Quantitative assessment of lens opacities with anterior segment optical coherence tomography
    • Wong A., Leung C.K., Weinreb R.N., et al. Quantitative assessment of lens opacities with anterior segment optical coherence tomography. Br J Ophthalmol 2009, 93:61-65.
    • (2009) Br J Ophthalmol , vol.93 , pp. 61-65
    • Wong, A.1    Leung, C.K.2    Weinreb, R.N.3
  • 121
    • 77952413389 scopus 로고    scopus 로고
    • Repeatability, reproducibility, and agreement of central anterior chamber depth measurements in pseudophakic and phakic eyes: optical coherence tomography versus ultrasound biomicroscopy
    • Zhang Q., Jin W., Wang Q. Repeatability, reproducibility, and agreement of central anterior chamber depth measurements in pseudophakic and phakic eyes: optical coherence tomography versus ultrasound biomicroscopy. JCataract Refract Surg 2010, 36:941-946.
    • (2010) JCataract Refract Surg , vol.36 , pp. 941-946
    • Zhang, Q.1    Jin, W.2    Wang, Q.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.