Volumn 7, Issue 2, 2014, Pages 337-346

Defibrillation testing is necessary at the time of implantable cardioverter defibrillator implantation

Author keywords

Arrhythmias; Cardiac; Defibrillators; Implantable; Ventricular fibrillation

Indexed keywords

ARTICLE; CARDIAC RESYNCHRONIZATION THERAPY; CARDIOVASCULAR MORTALITY; CLINICAL PRACTICE; CONVULSIVE THERAPY; DEFIBRILLATION; DEFIBRILLATION TESTING; DISEASE SEVERITY; EVIDENCE BASED MEDICINE; HEART ARREST; HEART ATRIUM FIBRILLATION; HEART FAILURE; HEART LEFT VENTRICLE EJECTION FRACTION; HEART LEFT VENTRICLE FAILURE; HEART VENTRICLE FIBRILLATION; HEART VENTRICLE TACHYCARDIA; HUMAN; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR; LONG TERM SURVIVAL; MEDICAL DEVICE COMPLICATION; NONISCHEMIC CARDIOMYOPATHY; OBSERVATIONAL STUDY; ORGAN PERFUSION; PEROPERATIVE COMPLICATION; PNEUMOTHORAX; POST HOC ANALYSIS; PRIORITY JOURNAL; R WAVE; RESPIRATION DEPRESSION; SINUS RHYTHM; SUDDEN CARDIAC DEATH; SUPERIOR CAVA VEIN; THORACOTOMY; TRANSIENT ISCHEMIC ATTACK; WAVEFORM;

EID: 84899697423     PISSN: 19413149     EISSN: 19413084     Source Type: Journal    
DOI: 10.1161/CIRCEP.113.000371     Document Type: Article
Times cited : (20)

References (79)
  • 1
    • 9844224486 scopus 로고    scopus 로고
    • A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near fatal ventricular arrhythmias
    • Antiarrhythmics versus Implantable Defibrillator (AVID) Investigators
    • Antiarrhythmics versus Implantable Defibrillator (AVID) Investigators. A comparison of antiarrhythmic drug therapy with implantable defibrillators in patients resuscitated from near fatal ventricular arrhythmias. N Engl J. 1997;337:1576-1583.
    • (1997) N Engl J. , vol.337 , pp. 1576-1583
  • 2
    • 17744375757 scopus 로고    scopus 로고
    • Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. Avid, cash and cids studies. Antiarrhythmics vs implantable defibrillator study. Cardiac arrest study hamburg. Canadian implantable defibrillator study
    • Connolly SJ, Hallstrom AP, Cappato R, Schron EB, Kuck KH, Zipes DP, Greene HL, Boczor S, Domanski M, Follmann D, Gent M, Roberts RS. Meta-analysis of the implantable cardioverter defibrillator secondary prevention trials. AVID, CASH and CIDS studies. Antiarrhythmics vs Implantable Defibrillator study. Cardiac Arrest Study Hamburg. Canadian Implantable Defibrillator Study. Eur Heart J. 2000;21:2071-2078.
    • (2000) Eur Heart J. , vol.21 , pp. 2071-2078
    • Connolly, S.J.1    Hallstrom, A.P.2    Cappato, R.3    Schron, E.B.4    Kuck, K.H.5    Zipes, D.P.6    Greene, H.L.7    Boczor, S.8    Domanski, M.9    Follmann, D.10    Gent, M.11    Roberts, R.S.12
  • 4
    • 0037149716 scopus 로고    scopus 로고
    • Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction
    • Multicenter Automatic Defibrillator Implantation Trial II Investigators
    • Moss AJ, Zareba W, Hall WJ, Klein H, Wilber DJ, Cannom DS, Daubert JP, Higgins SL, Brown MW, Andrews ML; Multicenter Automatic Defibrillator Implantation Trial II Investigators. Prophylactic implantation of a defibrillator in patients with myocardial infarction and reduced ejection fraction. N Engl J Med. 2002;346:877-883.
    • (2002) N Engl J Med. , vol.346 , pp. 877-883
    • Moss, A.J.1    Zareba, W.2    Hall, W.J.3    Klein, H.4    Wilber, D.J.5    Cannom, D.S.6    Daubert, J.P.7    Higgins, S.L.8    Brown, M.W.9    Andrews, M.L.10
  • 5
    • 0033576804 scopus 로고    scopus 로고
    • A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter unsustained tachycardia trial investigators
    • Buxton AE, Lee KL, Fisher JD, Josephson ME, Prystowsky EN, Hafley G. A randomized study of the prevention of sudden death in patients with coronary artery disease. Multicenter Unsustained Tachycardia Trial Investigators. N Engl J Med. 1999;341:1882-1890.
    • (1999) N Engl J Med. , vol.341 , pp. 1882-1890
    • Buxton, A.E.1    Lee, K.L.2    Fisher, J.D.3    Josephson, M.E.4    Prystowsky, E.N.5    Hafley, G.6
  • 8
    • 0028296757 scopus 로고
    • Effect of biphasic waveform pulse on endocardial defibrillation efficacy in humans
    • Neuzner J, Pitschner HF, Huth C, Schlepper M. Effect of biphasic waveform pulse on endocardial defibrillation efficacy in humans. Pacing Clin Electrophysiol. 1994;17:207-212.
    • (1994) Pacing Clin Electrophysiol. , vol.17 , pp. 207-212
    • Neuzner, J.1    Pitschner, H.F.2    Huth, C.3    Schlepper, M.4
  • 9
    • 0030237562 scopus 로고    scopus 로고
    • Effects of an active pectoral-pulse generator shell on defibrillation efficacy with a transvenous lead system
    • Gold MR, Foster AH, Shorofsky SR. Effects of an active pectoral-pulse generator shell on defibrillation efficacy with a transvenous lead system. Am J Cardiol. 1996;78:540-543.
    • (1996) Am J Cardiol. , vol.78 , pp. 540-543
    • Gold, M.R.1    Foster, A.H.2    Shorofsky, S.R.3
  • 10
    • 0030220823 scopus 로고    scopus 로고
    • Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillator. Active can investigators
    • Bardy GH, Yee R, Jung W. Multicenter experience with a pectoral unipolar implantable cardioverter-defibrillator. Active Can Investigators. J Am Coll Cardiol. 1996;28:400-410.
    • (1996) J Am Coll Cardiol. , vol.28 , pp. 400-410
    • Bardy, G.H.1    Yee, R.2    Jung, W.3
  • 11
    • 0028136303 scopus 로고
    • Preliminary experience with a hybrid nonthoracotomy defibrillating system that includes a biphasic device: Comparison with a standard monophasic device using the same lead system
    • Natale A, Sra J, Axtell K, Maglio C, Dhala A, Deshpande S, Jazayeri M, Wase A, Akhtar M. Preliminary experience with a hybrid nonthoracotomy defibrillating system that includes a biphasic device: comparison with a standard monophasic device using the same lead system. J Am Coll Cardiol. 1994;24:406-412.
    • (1994) J Am Coll Cardiol. , vol.24 , pp. 406-412
    • Natale, A.1    Sra, J.2    Axtell, K.3    Maglio, C.4    Dhala, A.5    Deshpande, S.6    Jazayeri, M.7    Wase, A.8    Akhtar, M.9
  • 12
    • 0023127807 scopus 로고
    • Defibrillation threshold: A simple and quantitative estimate of the ability to defibrillate
    • Rattes MF, Jones DL, Sharma AD, Klein GJ. Defibrillation threshold: a simple and quantitative estimate of the ability to defibrillate. Pacing Clin Electrophysiol. 1987;10(1 Pt 1):70-77.
    • (1987) Pacing Clin Electrophysiol. , vol.10 , Issue.1 PART 1 , pp. 70-77
    • Rattes, M.F.1    Jones, D.L.2    Sharma, A.D.3    Klein, G.J.4
  • 13
    • 0026877545 scopus 로고
    • Defibrillation threshold: Clinical utility and therapeutic implications
    • Singer I, Lang D. Defibrillation threshold: clinical utility and therapeutic implications. Pacing Clin Electrophysiol. 1992;15:932-949.
    • (1992) Pacing Clin Electrophysiol. , vol.15 , pp. 932-949
    • Singer, I.1    Lang, D.2
  • 14
    • 0023786718 scopus 로고
    • Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillator
    • Marchlinski FE, Flores B, Miller JM, Gottlieb CD, Hargrove WC 3rd. Relation of the intraoperative defibrillation threshold to successful postoperative defibrillation with an automatic implantable cardioverter defibrillator. Am J Cardiol. 1988;62:393-398.
    • (1988) Am J Cardiol. , vol.62 , pp. 393-398
    • Marchlinski, F.E.1    Flores, B.2    Miller, J.M.3    Gottlieb, C.D.4    Hargrove III, W.C.5
  • 16
    • 0026733276 scopus 로고
    • Measurement of ventricular electrogram amplitude during intraoperative induction of ventricular tachyarrhythmias
    • Ellenbogen KA, Wood MA, Stambler BS, Welch WJ, Damiano RJ. Measurement of ventricular electrogram amplitude during intraoperative induction of ventricular tachyarrhythmias. Am J Cardiol. 1992;70:1017-1022.
    • (1992) Am J Cardiol. , vol.70 , pp. 1017-1022
    • Ellenbogen, K.A.1    Wood, M.A.2    Stambler, B.S.3    Welch, W.J.4    Damiano, R.J.5
  • 17
    • 0035336187 scopus 로고    scopus 로고
    • Implantation of cardioverter defibrillators without induction of ventricular fibrillation
    • Swerdlow CD. Implantation of cardioverter defibrillators without induction of ventricular fibrillation. Circulation. 2001;103:2159-2164.
    • (2001) Circulation. , vol.103 , pp. 2159-2164
    • Swerdlow, C.D.1
  • 19
    • 44649093349 scopus 로고    scopus 로고
    • High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems
    • INTRINSIC RV Study Investigators
    • Day JD, Olshansky B, Moore S, Brown S, Stolen KQ, Lerew DR; INTRINSIC RV Study Investigators. High defibrillation energy requirements are encountered rarely with modern dual-chamber implantable cardioverter-defibrillator systems. Europace. 2008;10:347-350.
    • (2008) Europace. , vol.10 , pp. 347-350
    • Day, J.D.1    Olshansky, B.2    Moore, S.3    Brown, S.4    Stolen, K.Q.5    Lerew, D.R.6
  • 22
  • 25
    • 84871256608 scopus 로고    scopus 로고
    • A randomized-controlled pilot study comparing icd implantation with and without intraoperative defibrillation testing in patients with heart failure and severe left ventricular dysfunction: A substudy of the raft trial
    • Healey JS, Gula LJ, Birnie DH, Sterns L, Connolly SJ, Sapp J, Crystal E, Simpson C, Exner DV, Kus T, Philippon F, Wells G, Tang AS. A randomized-controlled pilot study comparing ICD implantation with and without intraoperative defibrillation testing in patients with heart failure and severe left ventricular dysfunction: a substudy of the RAFT trial. J Cardiovasc Electrophysiol. 2012;23:1313-1316.
    • (2012) J Cardiovasc Electrophysiol. , vol.23 , pp. 1313-1316
    • Healey, J.S.1    Gula, L.J.2    Birnie, D.H.3    Sterns, L.4    Connolly, S.J.5    Sapp, J.6    Crystal, E.7    Simpson, C.8    Exner, D.V.9    Kus, T.10    Philippon, F.11    Wells, G.12    Tang, A.S.13
  • 26
    • 84894886175 scopus 로고    scopus 로고
    • Effect of number of shock coils on defibrillation testing outcome: Analysis from the medtronic discovery™ hub database
    • Havel WJ, Johnson J. Effect of number of shock coils on defibrillation testing outcome: analysis from the Medtronic Discovery™ Hub database. Heart Rhythm. 2010;7:S75.
    • (2010) Heart Rhythm. , vol.7
    • Havel, W.J.1    Johnson, J.2
  • 28
    • 33845220084 scopus 로고    scopus 로고
    • Effect of defibrillation testing on management during implantable cardioverter-defibrillator implantation
    • Leong-Sit P, Gula LJ, Diamantouros P, Krahn AD, Skanes AC, Yee R, Klein GJ. Effect of defibrillation testing on management during implantable cardioverter-defibrillator implantation. Am Heart J. 2006;152:1104-1108.
    • (2006) Am Heart J. , vol.152 , pp. 1104-1108
    • Leong-Sit, P.1    Gula, L.J.2    Diamantouros, P.3    Krahn, A.D.4    Skanes, A.C.5    Yee, R.6    Klein, G.J.7
  • 31
    • 0037245915 scopus 로고    scopus 로고
    • High defibrillation thresholds in transvenous biphasic implantable defibrillators: Clinical predictors and prognostic implications
    • Shukla HH, Flaker GC, Jayam V, Roberts D. High defibrillation thresholds in transvenous biphasic implantable defibrillators: clinical predictors and prognostic implications. Pacing Clin Electrophysiol. 2003;26(1 Pt 1):44-48.
    • (2003) Pacing Clin Electrophysiol. , vol.26 , Issue.1 PART 1 , pp. 44-48
    • Shukla, H.H.1    Flaker, G.C.2    Jayam, V.3    Roberts, D.4
  • 33
    • 0027442215 scopus 로고
    • Determinants of successful nonthoracotomy cardioverter-defibrillator implantation: Experience in 101 patients using two different lead systems
    • Brooks R, Garan H, Torchiana D, Vlahakes GJ, Jackson G, Newell J, McGovern BA, Ruskin JN. Determinants of successful nonthoracotomy cardioverter-defibrillator implantation: experience in 101 patients using two different lead systems. J Am Coll Cardiol. 1993;22:1835-1842.
    • (1993) J Am Coll Cardiol. , vol.22 , pp. 1835-1842
    • Brooks, R.1    Garan, H.2    Torchiana, D.3    Vlahakes, G.J.4    Jackson, G.5    Newell, J.6    McGovern, B.A.7    Ruskin, J.N.8
  • 34
    • 84877583002 scopus 로고    scopus 로고
    • Echocardiographic parameters to predict inadequate defibrillation safety margin in patients receiving implantable cardioverter defibrillators for primary prevention
    • Jain SK, Ghanbari H, Hourani R, Larsen TR, Daccarett M, Machado C. Echocardiographic parameters to predict inadequate defibrillation safety margin in patients receiving implantable cardioverter defibrillators for primary prevention. J Interv Card Electrophysiol. 2013;37:79-85.
    • (2013) J Interv Card Electrophysiol. , vol.37 , pp. 79-85
    • Jain, S.K.1    Ghanbari, H.2    Hourani, R.3    Larsen, T.R.4    Daccarett, M.5    Machado, C.6
  • 35
    • 0029032493 scopus 로고
    • Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system
    • Raitt MH, Johnson G, Dolack GL, Poole JE, Kudenchuk PJ, Bardy GH. Clinical predictors of the defibrillation threshold with the unipolar implantable defibrillation system. J Am Coll Cardiol. 1995;25:1576-1583.
    • (1995) J Am Coll Cardiol. , vol.25 , pp. 1576-1583
    • Raitt, M.H.1    Johnson, G.2    Dolack, G.L.3    Poole, J.E.4    Kudenchuk, P.J.5    Bardy, G.H.6
  • 37
    • 0026657534 scopus 로고
    • Clinical characteristics and outcome of patients with high defibrillation thresholds. A multicenter study
    • Epstein AE, Ellenbogen KA, Kirk KA, Kay GN, Dailey SM, Plumb VJ. Clinical characteristics and outcome of patients with high defibrillation thresholds. A multicenter study. Circulation. 1992;86:1206-1216.
    • (1992) Circulation. , vol.86 , pp. 1206-1216
    • Epstein, A.E.1    Ellenbogen, K.A.2    Kirk, K.A.3    Kay, G.N.4    Dailey, S.M.5    Plumb, V.J.6
  • 38
    • 0031913093 scopus 로고    scopus 로고
    • Comparison of right-and left-sided pectoral implantation parameters with the jewel active can cardiodefibrillator
    • The World Wide Jewel Investigators.
    • Flaker GC, Tummala R, Wilson J. Comparison of right-and left-sided pectoral implantation parameters with the Jewel active can cardiodefibrillator. The World Wide Jewel Investigators. Pacing Clin Electrophysiol. 1998;21:447-451.
    • (1998) Pacing Clin Electrophysiol. , vol.21 , pp. 447-451
    • Flaker, G.C.1    Tummala, R.2    Wilson, J.3
  • 41
    • 33747070573 scopus 로고    scopus 로고
    • Effect of amiodarone and sotalol on ventricular defibrillation threshold: The optimal pharmacological therapy in cardioverter defibrillator patients (optic) trial
    • Hohnloser SH, Dorian P, Roberts R, Gent M, Israel CW, Fain E, Champagne J, Connolly SJ. Effect of amiodarone and sotalol on ventricular defibrillation threshold: the optimal pharmacological therapy in cardioverter defibrillator patients (OPTIC) trial. Circulation. 2006;114:104-109.
    • (2006) Circulation. , vol.114 , pp. 104-109
    • Hohnloser, S.H.1    Dorian, P.2    Roberts, R.3    Gent, M.4    Israel, C.W.5    Fain, E.6    Champagne, J.7    Connolly, S.J.8
  • 43
    • 0033922484 scopus 로고    scopus 로고
    • Amiodarone and defibrillation thresholds: A clinical conundrum
    • Dorian P. Amiodarone and defibrillation thresholds: a clinical conundrum. J Cardiovasc Electrophysiol. 2000;11:741-743.
    • (2000) J Cardiovasc Electrophysiol. , vol.11 , pp. 741-743
    • Dorian, P.1
  • 44
    • 78049415612 scopus 로고    scopus 로고
    • Incidence of very high defibrillation thresholds (dft) and efficacy of subcutaneous (sq) array insertion during implantable cardioverter defibrillator (icd) implantation
    • Verma A, Kaplan AJ, Sarak B, Oosthuizen R, Beardsall M, Higgenbottam J, Wulffhart Z, Khaykin Y. Incidence of very high defibrillation thresholds (DFT) and efficacy of subcutaneous (SQ) array insertion during implantable cardioverter defibrillator (ICD) implantation. J Interv Card Electrophysiol. 2010;29:127-133.
    • (2010) J Interv Card Electrophysiol. , vol.29 , pp. 127-133
    • Verma, A.1    Kaplan, A.J.2    Sarak, B.3    Oosthuizen, R.4    Beardsall, M.5    Higgenbottam, J.6    Wulffhart, Z.7    Khaykin, Y.8
  • 45
    • 0036231640 scopus 로고    scopus 로고
    • Clinical predictors of defibrillation thresholds with an active pectoral pulse generator lead system
    • Hodgson DM, Olsovsky MR, Shorofsky SR, Daly B, Gold MR. Clinical predictors of defibrillation thresholds with an active pectoral pulse generator lead system. Pacing Clin Electrophysiol. 2002;25(4 Pt 1):408-413.
    • (2002) Pacing Clin Electrophysiol. , vol.25 , Issue.4 PART 1 , pp. 408-413
    • Hodgson, D.M.1    Olsovsky, M.R.2    Shorofsky, S.R.3    Daly, B.4    Gold, M.R.5
  • 47
    • 34247499412 scopus 로고    scopus 로고
    • Relationship between defibrillation threshold, programmed energy and first shock efficacy for fast ventricular tachycardia and ventricular fibrillation in painfre rx ii
    • Investigators PRI II-503A
    • Sweeney M, DeGroot P, Stark A, Otterness M, Buxton A, Wathen M, Investigators PRI. Relationship between defibrillation threshold, programmed energy and first shock efficacy for fast ventricular tachycardia and ventricular fibrillation in PainFRE Rx II. Circulation. 2004;110:II-503A.
    • (2004) Circulation. , vol.110
    • Sweeney, M.1    DeGroot, P.2    Stark, A.3    Otterness, M.4    Buxton, A.5    Wathen, M.6
  • 48
    • 33644778037 scopus 로고    scopus 로고
    • Intraoperative testing of the implantable cardioverter-defibrillator: How much is enough?
    • Pires LA, Johnson KM. Intraoperative testing of the implantable cardioverter-defibrillator: how much is enough? J Cardiovasc Electrophysiol. 2006;17:140-145.
    • (2006) J Cardiovasc Electrophysiol. , vol.17 , pp. 140-145
    • Pires, L.A.1    Johnson, K.M.2
  • 49
    • 0037123144 scopus 로고    scopus 로고
    • Sudden death in patients with implantable cardioverter defibrillators: The importance of post-shock electromechanical dissociation
    • Mitchell LB, Pineda EA, Titus JL, Bartosch PM, Benditt DG. Sudden death in patients with implantable cardioverter defibrillators: the importance of post-shock electromechanical dissociation. J Am Coll Cardiol. 2002;39:1323-1328.
    • (2002) J Am Coll Cardiol. , vol.39 , pp. 1323-1328
    • Mitchell, L.B.1    Pineda, E.A.2    Titus, J.L.3    Bartosch, P.M.4    Benditt, D.G.5
  • 50
    • 0019022393 scopus 로고
    • Therapeutic indices for transchest defibrillator shocks: Effective, damaging, and lethal electrical doses
    • Babbs CF, Tacker WA, VanVleet JF, Bourland JD, Geddes LA. Therapeutic indices for transchest defibrillator shocks: effective, damaging, and lethal electrical doses. Am Heart J. 1980;99:734-738.
    • (1980) Am Heart J. , vol.99 , pp. 734-738
    • Babbs, C.F.1    Tacker, W.A.2    VanVleet, J.F.3    Bourland, J.D.4    Geddes, L.A.5
  • 51
    • 0024215748 scopus 로고
    • Death and damage caused by multiple direct current shocks: Studies in an animal model
    • Wilson CM, Allen JD, Bridges JB, Adgey AA. Death and damage caused by multiple direct current shocks: studies in an animal model. Eur Heart J. 1988;9:1257-1265.
    • (1988) Eur Heart J. , vol.9 , pp. 1257-1265
    • Wilson, C.M.1    Allen, J.D.2    Bridges, J.B.3    Adgey, A.A.4
  • 52
    • 84871188679 scopus 로고    scopus 로고
    • Defibrillation testing can reveal 'concealed' lead fracture
    • Bun SS, Duytschaever M, Tavernier R. Defibrillation testing can reveal 'concealed' lead fracture. Europace. 2013;15:54.
    • (2013) Europace. , vol.15 , pp. 54
    • Bun, S.S.1    Duytschaever, M.2    Tavernier, R.3
  • 53
    • 84883490504 scopus 로고    scopus 로고
    • Is high-voltage lead integrity measurement adequate during defibrillator generator replacement?
    • Doshi R, Ceballos S, Mendez F. Is high-voltage lead integrity measurement adequate during defibrillator generator replacement? Journal of Innovations in Cardiac Rhythm Management. 2012;3:1016-1019.
    • (2012) Journal of Innovations in Cardiac Rhythm Management. , vol.3 , pp. 1016-1019
    • Doshi, R.1    Ceballos, S.2    Mendez, F.3
  • 54
    • 84872004351 scopus 로고    scopus 로고
    • Failure to deliver therapy by a riata lead with internal wire externalization and normal electrical parameters during routine interrogation
    • Shah P, Singh G, Chandra S, Schuger CD. Failure to deliver therapy by a Riata Lead with internal wire externalization and normal electrical parameters during routine interrogation. J Cardiovasc Electrophysiol. 2013;24:94-96.
    • (2013) J Cardiovasc Electrophysiol. , vol.24 , pp. 94-96
    • Shah, P.1    Singh, G.2    Chandra, S.3    Schuger, C.D.4
  • 58
    • 33644693959 scopus 로고    scopus 로고
    • Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients
    • Alter P, Waldhans S, Plachta E, Moosdorf R, Grimm W. Complications of implantable cardioverter defibrillator therapy in 440 consecutive patients. Pacing Clin Electrophysiol. 2005;28:926-932.
    • (2005) Pacing Clin Electrophysiol. , vol.28 , pp. 926-932
    • Alter, P.1    Waldhans, S.2    Plachta, E.3    Moosdorf, R.4    Grimm, W.5
  • 62
    • 84889686156 scopus 로고    scopus 로고
    • Patient, physician, and procedural factors influencing the use of defibrillation testing during initial implantable cardioverter insertion: Findings of the ncdr®
    • Russo A, Wang Y, Curtis J, Al-Khatib, S, Lampert R. Patient, physician, and procedural factors influencing the use of defibrillation testing during initial implantable cardioverter insertion: findings of the NCDR®. Pacing Clin Electrophysiol. 2013;36:1522-1531.
    • (2013) Pacing Clin Electrophysiol. , vol.36 , pp. 1522-1531
    • Russo, A.1    Wang, Y.2    Curtis, J.3    Al-Khatib, S.4    Lampert, R.5
  • 63
    • 0032856728 scopus 로고    scopus 로고
    • Sudden death in recipients of transvenous implantable cardioverter defibrillator systems: Terminal events, predictors, and potential mechanisms
    • Pires LA, Hull ML, Nino CL, May LM, Ganji JR. Sudden death in recipients of transvenous implantable cardioverter defibrillator systems: terminal events, predictors, and potential mechanisms. J Cardiovasc Electrophysiol. 1999;10:1049-1056.
    • (1999) J Cardiovasc Electrophysiol. , vol.10 , pp. 1049-1056
    • Pires, L.A.1    Hull, M.L.2    Nino, C.L.3    May, L.M.4    Ganji, J.R.5
  • 64
  • 65
    • 79955436789 scopus 로고    scopus 로고
    • Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation
    • Michowitz Y, Lellouche N, Contractor T, Bourke T, Wiener I, Buch E, Boyle N, Bersohn M, Shivkumar K. Defibrillation threshold testing fails to show clinical benefit during long-term follow-up of patients undergoing cardiac resynchronization therapy defibrillator implantation. Europace. 2011;13:683-688.
    • (2011) Europace. , vol.13 , pp. 683-688
    • Michowitz, Y.1    Lellouche, N.2    Contractor, T.3    Bourke, T.4    Wiener, I.5    Buch, E.6    Boyle, N.7    Bersohn, M.8    Shivkumar, K.9
  • 67
    • 82455194861 scopus 로고    scopus 로고
    • Shocks burden and increased mortality in implantable cardioverter- defibrillator patients
    • Larsen GK, Evans J, Lambert WE, Chen Y, Raitt MH. Shocks burden and increased mortality in implantable cardioverter-defibrillator patients. Heart Rhythm. 2011;8:1881-1886.
    • (2011) Heart Rhythm. , vol.8 , pp. 1881-1886
    • Larsen, G.K.1    Evans, J.2    Lambert, W.E.3    Chen, Y.4    Raitt, M.H.5
  • 72
    • 84882274660 scopus 로고    scopus 로고
    • Effect of defibrillation threshold testing on heart failure hospitalization or death in the multicenter automatic defibrillator implantation trial-cardiac resynchronization therapy (madit-crt)
    • Aktas MK, Huang DT, Daubert JP, Schuger CD, McNitt S, Goldenberg I, Moss AJ, Zareba W. Effect of defibrillation threshold testing on heart failure hospitalization or death in the Multicenter Automatic Defibrillator Implantation Trial-Cardiac Resynchronization Therapy (MADIT-CRT). Heart Rhythm. 2013;10:193-199.
    • (2013) Heart Rhythm. , vol.10 , pp. 193-199
    • Aktas, M.K.1    Huang, D.T.2    Daubert, J.P.3    Schuger, C.D.4    McNitt, S.5    Goldenberg, I.6    Moss, A.J.7    Zareba, W.8
  • 73
    • 77952529329 scopus 로고    scopus 로고
    • The prognostic impact of shocks for clinical and induced arrhythmias on morbidity and mortality among patients with implantable cardioverter- defibrillators
    • Bhavnani SP, Kluger J, Coleman CI, White CM, Guertin D, Shafi NA, Yarlagadda RK, Clyne CA. The prognostic impact of shocks for clinical and induced arrhythmias on morbidity and mortality among patients with implantable cardioverter-defibrillators. Heart Rhythm. 2010;7:755-760.
    • (2010) Heart Rhythm. , vol.7 , pp. 755-760
    • Bhavnani, S.P.1    Kluger, J.2    Coleman, C.I.3    White, C.M.4    Guertin, D.5    Shafi, N.A.6    Yarlagadda, R.K.7    Clyne, C.A.8
  • 79
    • 79951494024 scopus 로고    scopus 로고
    • Defibrillation testing: Philosophy and science
    • Fisher JD. Defibrillation testing: philosophy and science. Pacing Clin Electrophysiol. 2011;34:185.
    • (2011) Pacing Clin Electrophysiol. , vol.34 , pp. 185
    • Fisher, J.D.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.