Lipemia: Causes, interference mechanisms, detection and management

Author keywords

Biological variation; Interference testing; Laboratory error; Lipemia interference; Preanalytical phase

Indexed keywords

C REACTIVE PROTEIN; CERULOPLASMIN; CHYLOMICRON; GAMMA GLUTAMYLTRANSFERASE; INTRALIPID; IVELIP; LIPID EMULSION; LOW DENSITY LIPOPROTEIN; UNCLASSIFIED DRUG; VERY LOW DENSITY LIPOPROTEIN;

EID: 84893618127     PISSN: 13300962     EISSN: None     Source Type: Journal    
DOI: 10.11613/BM.2014.008     Document Type: Review
Times cited : (191)

References (59)
  • 1
    • 0027960559 scopus 로고
    • Interference with clinical laboratory analyses
    • Kroll MH, Elin RJ. Interference with clinical laboratory analyses. Clin Chem 1994;40:1996-2005.
    • (1994) Clin Chem , vol.40 , pp. 1996-2005
    • Kroll, M.H.1    Elin, R.J.2
  • 3
    • 44949158707 scopus 로고    scopus 로고
    • Haemolysis: An overview of the leading cause of unsuitable specimens in clinical laboratories
    • Lippi G, Blanckaert N, Bonini P, Green S, Kitchen S, Palicka V, et al. Haemolysis: an overview of the leading cause of unsuitable specimens in clinical laboratories. Clin Chem Lab Med 2008;46:764-72. http://dx.doi.org/10.1515/CCLM.2008.170.
    • (2008) Clin Chem Lab Med , vol.46 , pp. 764-772
    • Lippi, G.1    Blanckaert, N.2    Bonini, P.3    Green, S.4    Kitchen, S.5    Palicka, V.6
  • 4
    • 0024316158 scopus 로고
    • Unreliable visual estimation of the incidence and amount of turbidity, hemolysis, and icterus in serum from hospitalized patients
    • Glick MR, Ryder KW, Glick SJ, Woods JR. Unreliable visual estimation of the incidence and amount of turbidity, hemolysis, and icterus in serum from hospitalized patients. Clin Chem 1989;35:837-9.
    • (1989) Clin Chem , vol.35 , pp. 837-839
    • Glick, M.R.1    Ryder, K.W.2    Glick, S.J.3    Woods, J.R.4
  • 5
    • 77953868631 scopus 로고    scopus 로고
    • The prevalence of preanalytical errors in Croatian ISO 15189 accredited laboratory
    • Simundic AM, Nikolac N, Vukasovic I, Vrkic N. The prevalence of preanalytical errors in Croatian ISO 15189 accredited laboratory. Clin Chem Lab Med 2010;48:1009-14. http://dx.doi.org/10.1515/CCLM.2010.221.
    • (2010) Clin Chem Lab Med , vol.48 , pp. 1009-1014
    • Simundic, A.M.1    Nikolac, N.2    Vukasovic, I.3    Vrkic, N.4
  • 6
    • 74549191326 scopus 로고    scopus 로고
    • Evaluation of errors in a clinical laboratory: A one-year experience
    • Goswami B, Singh B, Chawla R, Mallika V. Evaluation of errors in a clinical laboratory: a one-year experience. Clin Chem Lab Med 2010;48:63-6. http://dx.doi.org/10.1515/CCLM.2010.006.
    • (2010) Clin Chem Lab Med , vol.48 , pp. 63-66
    • Goswami, B.1    Singh, B.2    Chawla, R.3    Mallika, V.4
  • 8
    • 84865866502 scopus 로고    scopus 로고
    • Patient safety and quality in laboratory and hemostasis testing: A renewed loop?
    • Plebani M, Favaloro EJ, Lippi G. Patient safety and quality in laboratory and hemostasis testing: a renewed loop? Semin Thromb Hemost 2012;38:553-8. http://dx.doi.org/10.1055/s-0032-1315960.
    • (2012) Semin Thromb Hemost , vol.38 , pp. 553-558
    • Plebani, M.1    Favaloro, E.J.2    Lippi, G.3
  • 9
    • 0037317156 scopus 로고    scopus 로고
    • Effects of insulin resistance and type 2 diabetes on lipoprotein subclass particle size and concentration determined by nuclear magnetic resonance
    • Garvey WT, Kwon S, Zheng D, Shaughnessy S, Wallace P, Hutto A, et al. Effects of insulin resistance and type 2 diabetes on lipoprotein subclass particle size and concentration determined by nuclear magnetic resonance. Diabetes 2003;52:453-62. http://dx.doi.org/10.2337/diabetes.52.2.453.
    • (2003) Diabetes , vol.52 , pp. 453-462
    • Garvey, W.T.1    Kwon, S.2    Zheng, D.3    Shaughnessy, S.4    Wallace, P.5    Hutto, A.6
  • 10
    • 84884211546 scopus 로고    scopus 로고
    • Interference in coagulation testing: Focus on spurious hemolysis, icterus, and lipemia
    • Lippi G, Plebani M, Favaloro EJ. Interference in coagulation testing: focus on spurious hemolysis, icterus, and lipemia. Semin Thromb Hemost 2013;39:258-66.
    • (2013) Semin Thromb Hemost , vol.39 , pp. 258-266
    • Lippi, G.1    Plebani, M.2    Favaloro, E.J.3
  • 11
    • 84885064166 scopus 로고    scopus 로고
    • Are patients well informed about the fasting requirements for laboratory blood testing?
    • Kackov S, Simundic AM, Gatti-Drnic A. Are patients well informed about the fasting requirements for laboratory blood testing? Biochem Med 2013;23:326-31. http://dx.doi.org/10.11613/BM.2013.040.
    • (2013) Biochem Med , vol.23 , pp. 326-331
    • Kackov, S.1    Simundic, A.M.2    Gatti-Drnic, A.3
  • 12
    • 84885028011 scopus 로고    scopus 로고
    • Croatian Society of Medical Biochemistry and Laboratory Medicine: National recommendations for venous blood sampling
    • Nikolac N, Šupak-Smolčič V, Šimundić AM, Ćelap I. Croatian Society of Medical Biochemistry and Laboratory Medicine: national recommendations for venous blood sampling. Biochem Med 2013;23:242-54. http://dx.doi.org/10.11613/BM.2013.031.
    • (2013) Biochem Med , vol.23 , pp. 242-254
    • Nikolac, N.1    Šupak-Smolčič, V.2    Šimundić, A.M.3    Ćelap, I.4
  • 14
    • 84893553246 scopus 로고    scopus 로고
    • Standardisation of collection requirements for fasting samples
    • in press
    • Simundic AM, Cornes M, Grankvist K, Lippi G, Nybo M. Standardisation of collection requirements for fasting samples. Clin Chim Acta 2013; in press. http://dx.doi.org/10.1016/j.cca.2013.11.008.
    • (2013) Clin Chim Acta
    • Simundic, A.M.1    Cornes, M.2    Grankvist, K.3    Lippi, G.4    Nybo, M.5
  • 15
    • 33845998545 scopus 로고    scopus 로고
    • Does visual turbidity correlate with serum triglyceride levels in babies on total parenteral nutrition?
    • Lim KH, Lian WB, Yeo CL. Does visual turbidity correlate with serum triglyceride levels in babies on total parenteral nutrition? Ann Acad Med Singapore 2006;35:790-3.
    • (2006) Ann Acad Med Singapore , vol.35 , pp. 790-793
    • Lim, K.H.1    Lian, W.B.2    Yeo, C.L.3
  • 16
    • 84886602919 scopus 로고    scopus 로고
    • Lipid emulsions in parenteral nutrition: Current applications and future developments
    • Ren T, Cong L, Wang Y, Tang Y, Tian B, Lin X, et al. Lipid emulsions in parenteral nutrition: current applications and future developments. Expert Opin Drug Deliv 2013;10:1533-49. http://dx.doi.org/10.1517/17425247.2013.824874.
    • (2013) Expert Opin Drug Deliv , vol.10 , pp. 1533-1549
    • Ren, T.1    Cong, L.2    Wang, Y.3    Tang, Y.4    Tian, B.5    Lin, X.6
  • 17
    • 84886635967 scopus 로고    scopus 로고
    • Prolonged lipemia and pancreatitis due to extended infusion of lipid emulsion in bupropion overdose
    • Bucklin MH, Gorodetsky RM, Wiegand TJ. Prolonged lipemia and pancreatitis due to extended infusion of lipid emulsion in bupropion overdose. Clin Toxicol (Phila) 2013;51:896-8. http://dx.doi.org/10.3109/15563650.2013.831436.
    • (2013) Clin Toxicol (Phila) , vol.51 , pp. 896-898
    • Bucklin, M.H.1    Gorodetsky, R.M.2    Wiegand, T.J.3
  • 18
    • 84879395481 scopus 로고    scopus 로고
    • Use of intravenous lipid emulsion in the resuscitation of a patient with cardiovascular collapse after a severe overdose of quetiapine
    • Bartos M, Knudsen K. Use of intravenous lipid emulsion in the resuscitation of a patient with cardiovascular collapse after a severe overdose of quetiapine. Clin Toxicol (Phila) 2013;51:501-4. http://dx.doi.org/10.3109/15563650.2013.803229.
    • (2013) Clin Toxicol (Phila) , vol.51 , pp. 501-504
    • Bartos, M.1    Knudsen, K.2
  • 19
    • 84893617548 scopus 로고    scopus 로고
    • Intralipid. Available at, Accessed October 22
    • Intralipid. Available at: http://www.drugs.com/pro/intralipid.html. Accessed October 22, 2013.
    • (2013)
  • 20
    • 0031924428 scopus 로고    scopus 로고
    • Serum protein electrophoresis by CZE 2000 clinical capillary electrophoresis system
    • Bossuyt X, Schiettekatte G, Bogaerts A, Blanckaert N. Serum protein electrophoresis by CZE 2000 clinical capillary electrophoresis system. Clin Chem 1998;44:749-59.
    • (1998) Clin Chem , vol.44 , pp. 749-759
    • Bossuyt, X.1    Schiettekatte, G.2    Bogaerts, A.3    Blanckaert, N.4
  • 21
    • 84880040339 scopus 로고    scopus 로고
    • Interferences in Immunoassays
    • Chiu NHL (Ed.), InTech, Available at, Accessed October 10, 2013
    • Schiettecatte J, Anckaert E, Smitz J. Interferences in Immunoassays. In: Advances in Immunoassay Technology, Chiu NHL (Ed.), InTech, 2012. Available at: http://www.intechopen.com/books/advances-in-immunoassay-technology/interference-in-immunoassays. Accessed October 10, 2013.
    • (2012) Advances In Immunoassay Technology
    • Schiettecatte, J.1    Anckaert, E.2    Smitz, J.3
  • 23
    • 7044241205 scopus 로고    scopus 로고
    • Evaluating interference caused by lipemia
    • Kroll MH. Evaluating interference caused by lipemia. Clin Chem 2004;50:1968-9. http://dx.doi.org/10.1373/clinchem.2004.038075.
    • (2004) Clin Chem , vol.50 , pp. 1968-1969
    • Kroll, M.H.1
  • 24
    • 84884298434 scopus 로고    scopus 로고
    • Heterogeneity of manufacturers' declarations for lipemia interference urgent call for standardization
    • Nikolac N, Simundic AM, Miksa M, Lima-Oliveira G, Salvagno GL, Caruso B, Guidi GC. Heterogeneity of manufacturers' declarations for lipemia interference urgent call for standardization. Clin Chim Acta 2013;426:33-40. http://dx.doi.org/10.1016/j.cca.2013.08.015.
    • (2013) Clin Chim Acta , vol.426 , pp. 33-40
    • Nikolac, N.1    Simundic, A.M.2    Miksa, M.3    Lima-Oliveira, G.4    Salvagno, G.L.5    Caruso, B.6    Guidi, G.C.7
  • 25
    • 67651121968 scopus 로고    scopus 로고
    • Interference on the laboratory measurement of bilirubin: The effect of in vitro interactions
    • de Cordova CM, Nogara MS, de Cordova MM. Interference on the laboratory measurement of bilirubin: the effect of in vitro interactions. Clin Chim Acta 2009;407:77-9. http://dx.doi.org/10.1016/j.cca.2009.06.037.
    • (2009) Clin Chim Acta , vol.407 , pp. 77-79
    • de Cordova, C.M.1    Nogara, M.S.2    de Cordova, M.M.3
  • 26
    • 42449101821 scopus 로고    scopus 로고
    • Haemolysis and turbidity influence on three analysis methods of quantitative determination of total and conjugated bilirubin on ADVIA 1650
    • (in French)
    • Gobert De Paepe E, Munteanu G, Schischmanoff PO, Porquet D. Haemolysis and turbidity influence on three analysis methods of quantitative determination of total and conjugated bilirubin on ADVIA 1650. (in French). Ann Biol Clin (Paris) 2008;66:175-82.
    • (2008) Ann Biol Clin (Paris) , vol.66 , pp. 175-182
    • De Paepe, G.E.1    Munteanu, G.2    Schischmanoff, P.O.3    Porquet, D.4
  • 27
    • 55949089038 scopus 로고    scopus 로고
    • Effects of hemolysis and lipemia on the COBAS salicylate and acetaminophen assays compared to GDS assays
    • Meany D, Schowinsky J, Clarke W. Effects of hemolysis and lipemia on the COBAS salicylate and acetaminophen assays compared to GDS assays. Clin Biochem 2008;41:1486-8. http://dx.doi.org/10.1016/j.clinbiochem.2008.09.111
    • (2008) Clin Biochem , vol.41 , pp. 1486-1488
    • Meany, D.1    Schowinsky, J.2    Clarke, W.3
  • 28
    • 84893609847 scopus 로고    scopus 로고
    • Available at, Accessed December 2nd
    • Volume Displacement of Electrolytes by Lipemia. Available at: https://ahdc.vet.cornell.edu/clinpath/modules/chem/volume.htm. Accessed December 2nd, 2013.
    • (2013) Volume Displacement of Electrolytes By Lipemia
  • 29
    • 0037406485 scopus 로고    scopus 로고
    • Pseudohyponatremia in a Myeloma Patient: Direct Electrode Potentiometry is a Method Worth its Salt
    • Lyon AW, Baskin LB. Pseudohyponatremia in a Myeloma Patient: Direct Electrode Potentiometry is a Method Worth its Salt. LabMedicine 2003;34:357-60.
    • (2003) LabMedicine , vol.34 , pp. 357-360
    • Lyon, A.W.1    Baskin, L.B.2
  • 30
    • 79955103353 scopus 로고    scopus 로고
    • Pseudohyponatremia revisited: A modern-day pitfall
    • Fortgens P, Pillay TS. Pseudohyponatremia revisited: a modern-day pitfall. Arch Pathol Lab Med 2011;135:516-9.
    • (2011) Arch Pathol Lab Med , vol.135 , pp. 516-519
    • Fortgens, P.1    Pillay, T.S.2
  • 31
    • 33751249016 scopus 로고    scopus 로고
    • Pseudohyponatremia: Does It Matter in Current Clinical Practice?
    • Kim GH. Pseudohyponatremia: Does It Matter in Current Clinical Practice? Electrolyte & Blood Pressure 2006;4:77-82. http://dx.doi.org/10.5049/EBP.2006.4.2.77.
    • (2006) Electrolyte & Blood Pressure , vol.4 , pp. 77-82
    • Kim, G.H.1
  • 32
    • 84893550566 scopus 로고    scopus 로고
    • Hemolysis, Lipemia, and High Bilirubin: Effect on Laboratory Tests
    • In: Dasgupta A, Sepulveda JL, eds., Amsterdam: Elsevier
    • Kazmierczak SC. Hemolysis, Lipemia, and High Bilirubin: Effect on Laboratory Tests. In: Dasgupta A, Sepulveda JL, eds. Accurate Results in the Clinical Laboratory: A Guide to Error Detection and Correction. Amsterdam: Elsevier 2013; p. 53-62. http://dx.doi.org/10.1016/B978-0-12-415783-5.00005-0.
    • (2013) Accurate Results In the Clinical Laboratory: A Guide to Error Detection and Correction , pp. 53-62
    • Kazmierczak, S.C.1
  • 33
    • 84857654702 scopus 로고    scopus 로고
    • Hemolysis, lipaemia and icterus in specimens for arterial blood gas analysis
    • Salvagno GL, Lippi G, Gelati M, Guidi GC. Hemolysis, lipaemia and icterus in specimens for arterial blood gas analysis. Clin Biochem 2012;45:372-3. http://dx.doi.org/10.1016/j.clinbiochem.2011.12.005
    • (2012) Clin Biochem , vol.45 , pp. 372-373
    • Salvagno, G.L.1    Lippi, G.2    Gelati, M.3    Guidi, G.C.4
  • 34
    • 73349087188 scopus 로고    scopus 로고
    • Comparison of visual versus automated detection of lipemic, icteric and hemolyzed specimens: Can we rely on a human eye?
    • Simundic AM, Nikolac N, Ivankovic V, Ferenec-Ruzic D, Magdic B, Kvaternik M, Topic E. Comparison of visual versus automated detection of lipemic, icteric and hemolyzed specimens: can we rely on a human eye? Clin Chem Lab Med 2009;47:1361-5. http://dx.doi.org/10.1515/CCLM.2009.306.
    • (2009) Clin Chem Lab Med , vol.47 , pp. 1361-1365
    • Simundic, A.M.1    Nikolac, N.2    Ivankovic, V.3    Ferenec-Ruzic, D.4    Magdic, B.5    Kvaternik, M.6    Topic, E.7
  • 35
    • 0242659853 scopus 로고    scopus 로고
    • Unreliability of triglyceride measurement to predict turbidity induced interference
    • Twomey PJ, Don-Wauchope AC, McCullough D. Unreliability of triglyceride measurement to predict turbidity induced interference. J Clin Pathol 2003;56:861-2. http://dx.doi.org/10.1136/jcp.56.11.861.
    • (2003) J Clin Pathol , vol.56 , pp. 861-862
    • Twomey, P.J.1    Don-Wauchope, A.C.2    McCullough, D.3
  • 37
    • 0037216249 scopus 로고    scopus 로고
    • Glycerol metabolism and the determination of triglycerides--clinical, biochemical and molecular findings in six subjects
    • Hellerud C, Burlina A, Gabelli C, Ellis JR, Nyholm PG, Lindstedt S. Glycerol metabolism and the determination of triglycerides--clinical, biochemical and molecular findings in six subjects. Clin Chem Lab Med 2003;41:46-55. http://dx.doi.org/10.1515/CCLM.2003.009.
    • (2003) Clin Chem Lab Med , vol.41 , pp. 46-55
    • Hellerud, C.1    Burlina, A.2    Gabelli, C.3    Ellis, J.R.4    Nyholm, P.G.5    Lindstedt, S.6
  • 38
    • 32544438605 scopus 로고    scopus 로고
    • Comparison of triglyceride concentration with lipemic index in disorders of triglyceride and glycerol metabolism
    • De Haene H, Taes Y, Christophe A, Delanghe J. Comparison of triglyceride concentration with lipemic index in disorders of triglyceride and glycerol metabolism. Clin Chem Lab Med 2006;44:220-2. http://dx.doi.org/10.1515/CCLM.2006.040.
    • (2006) Clin Chem Lab Med , vol.44 , pp. 220-222
    • de Haene, H.1    Taes, Y.2    Christophe, A.3    Delanghe, J.4
  • 39
    • 84867274925 scopus 로고    scopus 로고
    • Investigation of unusual high serum indices for lipemia in clear serum samples on Siemens analysers Dimension
    • Fliser E, Jerkovic K, Vidovic T, Gorenjak M. Investigation of unusual high serum indices for lipemia in clear serum samples on Siemens analysers Dimension. Biochem Med 2012;22:352-62. http://dx.doi.org/10.11613/BM.2012.037.
    • (2012) Biochem Med , vol.22 , pp. 352-362
    • Fliser, E.1    Jerkovic, K.2    Vidovic, T.3    Gorenjak, M.4
  • 41
    • 44249124482 scopus 로고    scopus 로고
    • Interference with serum indices measurement, but not chemical analysis, on the Roche Modular by Patent Blue V
    • Darby D, Broomhead C. Interference with serum indices measurement, but not chemical analysis, on the Roche Modular by Patent Blue V. Ann Clin Biochem 2008;45:289-92. http://dx.doi.org/10.1258/acb.2007.007176.
    • (2008) Ann Clin Biochem , vol.45 , pp. 289-292
    • Darby, D.1    Broomhead, C.2
  • 43
    • 79960137736 scopus 로고    scopus 로고
    • Lipemia interferences in routine clinical biochemical tests
    • Calmarza P, Cordero J. Lipemia interferences in routine clinical biochemical tests. Biochem Med 2011;21:160-6. http://dx.doi.org/10.11613/BM.2011.025.
    • (2011) Biochem Med , vol.21 , pp. 160-166
    • Calmarza, P.1    Cordero, J.2
  • 44
    • 79951711347 scopus 로고    scopus 로고
    • Lipaemic samples: Effective process for lipid reduction using high speed centrifugation compared with ultracentrifugation
    • Dimeski G, Jones BW. Lipaemic samples: effective process for lipid reduction using high speed centrifugation compared with ultracentrifugation. Biochem Med 2011;21:86-92. http://dx.doi.org/10.11613/BM.2011.016.
    • (2011) Biochem Med , vol.21 , pp. 86-92
    • Dimeski, G.1    Jones, B.W.2
  • 45
    • 0025273693 scopus 로고
    • Flocculation of serum lipoproteins with cyclodextrins: Application to assay of hyperlipidemic serum
    • Sharma A, Anderson K, Baker JW. Flocculation of serum lipoproteins with cyclodextrins: application to assay of hyperlipidemic serum. Clin Chem 1990;36:529-32.
    • (1990) Clin Chem , vol.36 , pp. 529-532
    • Sharma, A.1    Anderson, K.2    Baker, J.W.3
  • 46
    • 84893546925 scopus 로고    scopus 로고
    • Lipoclear. Available at, Accessed October 27
    • Lipoclear. Available at: http://www.statspin.com/pdfs/64-003957-001D.pdf. Accessed October 27, 2013.
    • (2013)
  • 48
    • 33846528579 scopus 로고    scopus 로고
    • Correction of patient results for Beckman Coulter LX-20 assays affected by interference due to hemoglobin, bilirubin or lipids: A practical approach
    • Vermeer HJ, Steen G, Naus AJ, Goevaerts B, Agricola PT, Schoenmakers CH. Correction of patient results for Beckman Coulter LX-20 assays affected by interference due to hemoglobin, bilirubin or lipids: a practical approach. Clin Chem Lab Med 2007;45:114-9. http://dx.doi.org/10.1515/CCLM.2007.004.
    • (2007) Clin Chem Lab Med , vol.45 , pp. 114-119
    • Vermeer, H.J.1    Steen, G.2    Naus, A.J.3    Goevaerts, B.4    Agricola, P.T.5    Schoenmakers, C.H.6
  • 49
    • 84896028351 scopus 로고    scopus 로고
    • The evaluation and comparison of consecutive high speed centrifugation and LipoClear® reagent for lipemia removal
    • pii:S0009-9120(14)00004-6
    • Saracevic A, Nikolac N, Simundic AM. The evaluation and comparison of consecutive high speed centrifugation and LipoClear® reagent for lipemia removal. Clin Biochem 2014;pii:S0009-9120(14)00004-6.
    • (2014) Clin Biochem
    • Saracevic, A.1    Nikolac, N.2    Simundic, A.M.3
  • 50
    • 0022885477 scopus 로고
    • Graphical comparisons of interferences in clinical chemistry instrumentation
    • Glick MR, Ryder KW, Jackson SA. Graphical comparisons of interferences in clinical chemistry instrumentation. Clin Chem 1986;32:470-5.
    • (1986) Clin Chem , vol.32 , pp. 470-475
    • Glick, M.R.1    Ryder, K.W.2    Jackson, S.A.3
  • 51
    • 33750447404 scopus 로고    scopus 로고
    • Clinical Laboratory Standards Institute, Second Edition. CLSI document EP7-A2. Clinical Laboratory Standards Institute, Wayne, Pennsylvania, USA
    • Clinical Laboratory Standards Institute. Interference Testing in Clinical Chemistry; Approved Guideline, Second Edition. CLSI document EP7-A2. Clinical Laboratory Standards Institute, Wayne, Pennsylvania, USA, 2005.
    • (2005) Interference Testing In Clinical Chemistry; Approved Guideline
  • 52
    • 0024084033 scopus 로고
    • Lipaemic interference: Effects of lipaemic serum and intralipid
    • Nanji AA, Poon R, Hinberg I. Lipaemic interference: effects of lipaemic serum and intralipid. J Clin Pathol 1988;41:1026-7. http://dx.doi.org/10.1136/jcp.41.9.1026.
    • (1988) J Clin Pathol , vol.41 , pp. 1026-1027
    • Nanji, A.A.1    Poon, R.2    Hinberg, I.3
  • 53
    • 7044260841 scopus 로고    scopus 로고
    • Assay-specific differences in lipemic interference in native and intralipid-supplemented samples
    • Bornhorst JA, Roberts RF, Roberts WL. Assay-specific differences in lipemic interference in native and intralipid-supplemented samples. Clin Chem 2004;50:2197-201. http://dx.doi.org/10.1373/clinchem.2004.040154.
    • (2004) Clin Chem , vol.50 , pp. 2197-21201
    • Bornhorst, J.A.1    Roberts, R.F.2    Roberts, W.L.3
  • 54
    • 0033369894 scopus 로고    scopus 로고
    • Current databases on biological variation: Pros, cons and progress
    • This database was most recently updated in 2012. Available at, Accessed October 27
    • Ricos C, Alvarez V, Cava F, García-Lario JV, Hernández A, Jiménez CV, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest 1999;59:491-500. This database was most recently updated in 2012. Available at: http://www.westgard.com/biodatabase1.htm. Accessed October 27, 2013.
    • (1999) Scand J Clin Lab Invest , vol.59 , pp. 491-500
    • Ricos, C.1    Alvarez, V.2    Cava, F.3    García-Lario, J.V.4    Hernández, A.5    Jiménez, C.V.6
  • 55
    • 0016254486 scopus 로고
    • Criteria for judging precision and accuracy in method development and evaluation
    • Westgard JO, Carey RN, Wold S. Criteria for judging precision and accuracy in method development and evaluation. Clin Chem USA 1974;20:825-33.
    • (1974) Clin Chem USA , vol.20 , pp. 825-833
    • Westgard, J.O.1    Carey, R.N.2    Wold, S.3
  • 56
    • 33645763031 scopus 로고    scopus 로고
    • Multicenter evaluation of the interference of hemoglobin, bilirubin and lipids on Synchron LX-20 assays
    • Steen G, Vermeer HJ, Naus AJ, Goevaerts B, Agricola PT, Schoenmakers CH. Multicenter evaluation of the interference of hemoglobin, bilirubin and lipids on Synchron LX-20 assays. Clin Chem Lab Med 2006;44:413-9. http://dx.doi.org/10.1515/CCLM.2006.067.
    • (2006) Clin Chem Lab Med , vol.44 , pp. 413-419
    • Steen, G.1    Vermeer, H.J.2    Naus, A.J.3    Goevaerts, B.4    Agricola, P.T.5    Schoenmakers, C.H.6
  • 57
    • 84867720074 scopus 로고    scopus 로고
    • Analytical interferences resulting from intravenous lipid emulsion
    • Grunbaum AM, Gilfix BM, Gosselin S, Blank DW. Analytical interferences resulting from intravenous lipid emulsion. Clin Toxicol (Phila) 2012;50:812-7. http://dx.doi.org/10.3109/15563650.2012.731509.
    • (2012) Clin Toxicol (Phila) , vol.50 , pp. 812-817
    • Grunbaum, A.M.1    Gilfix, B.M.2    Gosselin, S.3    Blank, D.W.4
  • 58
    • 79959295985 scopus 로고    scopus 로고
    • Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assays
    • Ji JZ, Meng QH. Evaluation of the interference of hemoglobin, bilirubin, and lipids on Roche Cobas 6000 assays. Clin Chim Acta 2011;412:1550-3. http://dx.doi.org/10.1016/j.cca.2011.04.034.
    • (2011) Clin Chim Acta , vol.412 , pp. 1550-1553
    • Ji, J.Z.1    Meng, Q.H.2
  • 59
    • 82955163112 scopus 로고    scopus 로고
    • Hemoglobin, bilirubin, and lipid interference on Roche Cobas 6000 assays
    • author reply 342-3
    • Szoke D, Braga F, Valente C, Panteghini M. Hemoglobin, bilirubin, and lipid interference on Roche Cobas 6000 assays. Clin Chim Acta 2012;413:339-41; author reply 342-3. http://dx.doi.org/10.1016/j.cca.2011.09.044.
    • (2012) Clin Chim Acta , vol.413 , pp. 339-341
    • Szoke, D.1    Braga, F.2    Valente, C.3    Panteghini, M.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.