Volumn 22, Issue 4, 2013, Pages 325-329

Comparison of intraocular pressure measurements and assessment of intraobserver and interobserver reproducibility with the portable icare rebound tonometer and goldmann applanation tonometer in glaucoma patients

Author keywords

Glaucoma; Goldmann applanation tonometer; Intraocular pressure; Rebound tonometer; Reproducibility

Indexed keywords

ADULT; AGED; ARTICLE; CENTRAL CORNEAL THICKNESS; CORRELATION COEFFICIENT; FEMALE; GLAUCOMA; GOLDMANN APPLANATION TONOMETER; HUMAN; INTRAOCULAR PRESSURE; INTRAOCULAR PRESSURE MEASURING DEVICE; LOW TENSION GLAUCOMA; MAJOR CLINICAL STUDY; MALE; OPEN ANGLE GLAUCOMA; PRIORITY JOURNAL; PROSPECTIVE STUDY; REBOUND TONOMETER; REPRODUCIBILITY; TONOMETER;

EID: 84879152718     PISSN: 10570829     EISSN: 1536481X     Source Type: Journal    
DOI: 10.1097/IJG.0b013e318237caa2     Document Type: Article
Times cited : (57)

References (24)
  • 3
    • 33645464184 scopus 로고    scopus 로고
    • Clinical evaluation of rebound tonometer
    • Davies LN, Bartlett H, Mallen EAH, et al. Clinical evaluation of rebound tonometer. Acta Ophthal Scand. 2006;84:206-209.
    • (2006) Acta Ophthal Scand. , vol.84 , pp. 206-209
    • Davies, L.N.1    Bartlett, H.2    Eah, M.3
  • 4
    • 34247633160 scopus 로고    scopus 로고
    • Reproducibility and tolerability of the ICare rebound tonometer in school children
    • Sahin A, Basmak H, Niyaz L, et al. Reproducibility and tolerability of the ICare rebound tonometer in school children. J Glaucoma. 2007;16:185-188.
    • (2007) J Glaucoma. , vol.16 , pp. 185-188
    • Sahin, A.1    Basmak, H.2    Niyaz, L.3
  • 5
    • 34548410983 scopus 로고    scopus 로고
    • Rebound tonometry in a clinical setting. Comparison with applanation tonometry
    • Lopez-Caballero C, Contreras I, Munoz-Negrete FJ, et al. Rebound tonometry in a clinical setting. Comparison with applanation tonometry. Arch Soc Esp Oftalmol. 2007;82: 273-278.
    • (2007) Arch Soc Esp Oftalmol. , vol.82 , pp. 273-278
    • Lopez-Caballero, C.1    Contreras, I.2    Munoz-Negrete, F.J.3
  • 6
    • 33746277068 scopus 로고    scopus 로고
    • Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers
    • Nakamura M, Darhad U, Tatsumi Y, et al. Agreement of rebound tonometer in measuring intraocular pressure with three types of applanation tonometers. Am J Ophthalmol. 2006;142:332-334.
    • (2006) Am J Ophthalmol. , vol.142 , pp. 332-334
    • Nakamura, M.1    Darhad, U.2    Tatsumi, Y.3
  • 7
    • 33745282996 scopus 로고    scopus 로고
    • Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients
    • Brusini P, Salvetat ML, Zeppieri M, et al. Comparison of ICare tonometer with Goldmann applanation tonometer in glaucoma patients. J Glaucoma. 2006;15:213-217.
    • (2006) J Glaucoma. , vol.15 , pp. 213-217
    • Brusini, P.1    Salvetat, M.L.2    Zeppieri, M.3
  • 9
    • 0027269723 scopus 로고
    • The effect of corneal thickness on applanation tonometry
    • Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol. 1993;115:592-596.
    • (1993) Am J Ophthalmol. , vol.115 , pp. 592-596
    • Whitacre, M.M.1    Stein, R.A.2    Hassanein, K.3
  • 10
    • 40049104046 scopus 로고    scopus 로고
    • Clinical comparison of the ICare tonometer and Goldmann applanation tonometry
    • Pakrou N, Gray T, Mills R, et al. Clinical comparison of the ICare tonometer and Goldmann applanation tonometry. J Glaucoma. 2008;17:43-47.
    • (2008) J Glaucoma. , vol.17 , pp. 43-47
    • Pakrou, N.1    Gray, T.2    Mills, R.3
  • 11
    • 70350148529 scopus 로고    scopus 로고
    • The ICare rebound tonometer: Comparisons with Goldmann tonometry, and influence of central corneal thickness
    • Poostchi A, Mitchell R, Nicholas S, et al. The ICare rebound tonometer: comparisons with Goldmann tonometry, and influence of central corneal thickness. Clin Exp Ophthalmol. 2009;37:687-691.
    • (2009) Clin Exp Ophthalmol. , vol.37 , pp. 687-691
    • Poostchi, A.1    Mitchell, R.2    Nicholas, S.3
  • 12
    • 37349001245 scopus 로고    scopus 로고
    • The influence of corneal properties on rebound tonometry
    • Chui WS, Lam A, Chen D, et al. The influence of corneal properties on rebound tonometry. Ophthalmology. 2008;115: 80-84.
    • (2008) Ophthalmology. , vol.115 , pp. 80-84
    • Chui, W.S.1    Lam, A.2    Chen, D.3
  • 13
    • 58149123724 scopus 로고    scopus 로고
    • Correlations between corneal biomechanical properties measured with the ocular response analyzer and ICare rebound tonometry
    • Jorge JM, Gonzalez-Meijome JM, Queiros A, et al. Correlations between corneal biomechanical properties measured with the ocular response analyzer and ICare rebound tonometry. J Glaucoma. 2008;17:442-448.
    • (2008) J Glaucoma. , vol.17 , pp. 442-448
    • Jorge, J.M.1    Gonzalez-Meijome, J.M.2    Queiros, A.3
  • 14
    • 0036669531 scopus 로고    scopus 로고
    • Effects of prostaglandins on the aqueous humor outflow pathways
    • Weinreb RN, Toris CB, Gabelt BT, et al. Effects of prostaglandins on the aqueous humor outflow pathways. Surv Ophthalmol. 2002;47(suppl 1):S53-S64.
    • (2002) Surv Ophthalmol. , vol.47 , Issue.SUPPL. 1
    • Weinreb, R.N.1    Toris, C.B.2    Gabelt, B.T.3
  • 15
    • 33846003126 scopus 로고    scopus 로고
    • Relationship between travoprost and central corneal thickness in ocular hypertension and open-angle glaucoma
    • Harasymowycz PJ, Pappamatheakis DG, Ennis M, et al. Relationship between travoprost and central corneal thickness in ocular hypertension and open-angle glaucoma. Cornea. 2007;26:34-41.
    • (2007) Cornea. , vol.26 , pp. 34-41
    • Harasymowycz, P.J.1    Pappamatheakis, D.G.2    Ennis, M.3
  • 16
    • 0028273231 scopus 로고
    • Reliability of intraocular pressure measurement with the Goldmann appla-nation tonometer in epidemiological studies
    • Dielemans I, Vingerling JR, Hofman A, et al. Reliability of intraocular pressure measurement with the Goldmann appla-nation tonometer in epidemiological studies. Graefes Arch Clin Exp Ophthalmol. 1994;232:141-144.
    • (1994) Graefes Arch Clin Exp Ophthalmol. , vol.232 , pp. 141-144
    • Dielemans, I.1    Vingerling, J.R.2    Hofman, A.3
  • 17
    • 4344659799 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with Goldmann applanation tonometry
    • Kaufman C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with Goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004;45:3118-3121.
    • (2004) Invest Ophthalmol Vis Sci. , vol.45 , pp. 3118-3121
    • Kaufman, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 19
    • 0014314194 scopus 로고
    • Repeated applanation tonometry
    • Moses RA, Liu CH. Repeated applanation tonometry. Am J Ophthalmol. 1968;66:89-91.
    • (1968) Am J Ophthalmol. , vol.66 , pp. 89-91
    • Moses, R.A.1    Liu, C.H.2
  • 22
    • 1142299596 scopus 로고    scopus 로고
    • Measuring intraocular pressure with the Pulsair 3000 and rebound tonometers in elderly patients without an anesthetic
    • Kontiola A, Puska P. Measuring intraocular pressure with the Pulsair 3000 and rebound tonometers in elderly patients without an anesthetic. Graefes Arch Clin Exp Ophthalmol. 2004;242:3-7.
    • (2004) Graefes Arch Clin Exp Ophthalmol. , vol.242 , pp. 3-7
    • Kontiola, A.1    Puska, P.2
  • 23
    • 37349013501 scopus 로고    scopus 로고
    • Clinical comparison of PASCAL dynamic contour tonometry and Goldmann applanation tonometry in asymmetric open-angle glaucoma
    • Sullivan-Mee M, Halverson KD, Qualis C. Clinical comparison of PASCAL dynamic contour tonometry and Goldmann applanation tonometry in asymmetric open-angle glaucoma. J Glaucoma. 2007;8:694-699.
    • (2007) J Glaucoma. , vol.8 , pp. 694-699
    • Sullivan-Mee, M.1    Halverson, K.D.2    Qualis, C.3
  • 24
    • 77949269541 scopus 로고    scopus 로고
    • The influence of refractive errors on IOP measurements by rebound tonometer (ICare) and Goldmann applanation tonometry
    • Avitabile T, Longo A, Rocca D, et al. The influence of refractive errors on IOP measurements by rebound tonometer (ICare) and Goldmann applanation tonometry. Graefes Arch Clin Exp Ophthalmol. 2010;248:585-591.
    • (2010) Graefes Arch Clin Exp Ophthalmol. , vol.248 , pp. 585-591
    • Avitabile, T.1    Longo, A.2    Rocca, D.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.