Volumn 16, Issue 8, 2007, Pages 694-699

Clinical comparison of Pascal dynamic contour tonometry and Goldmann applanation tonometry in asymmetric open-angle glaucoma

Author keywords

Central corneal thickness; Corneal biomechanical properties; Dynamic contour tonometry; Glaucoma; Goldmann applanation tonometry; Intraocular pressure; Tonometry

Indexed keywords

ADULT; ARTICLE; COMPARATIVE STUDY; CONFIDENCE INTERVAL; CONTROLLED STUDY; DISEASE ASSOCIATION; FEMALE; HUMAN; INTRAOCULAR PRESSURE; MAJOR CLINICAL STUDY; MALE; MEASUREMENT; MULTIVARIATE ANALYSIS; OCULOPLETHYSMOGRAPHY; OPEN ANGLE GLAUCOMA; OPTIC NERVE DISEASE; PRIORITY JOURNAL; RETROSPECTIVE STUDY; SCORING SYSTEM; STUDENT T TEST; VISUAL FIELD; VISUAL FIELD DEFECT;

EID: 37349013501     PISSN: 10570829     EISSN: None     Source Type: Journal    
DOI: 10.1097/IJG.0b013e3180408dc6     Document Type: Article
Times cited : (21)

References (28)
  • 1
    • 23744463890 scopus 로고    scopus 로고
    • Dynamic contour tonometry for intraocular pressure measurement
    • Kamppeter BA, Jonas JB. Dynamic contour tonometry for intraocular pressure measurement. Am J Ophthalmol. 2005;140: 318-320.
    • (2005) Am J Ophthalmol , vol.140 , pp. 318-320
    • Kamppeter, B.A.1    Jonas, J.B.2
  • 2
    • 30344436327 scopus 로고    scopus 로고
    • Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry
    • Schneider E, Grehn F. Intraocular pressure measurement-comparison of dynamic contour tonometry and goldmann applanation tonometry. J Glaucoma. 2006;15:2-6.
    • (2006) J Glaucoma , vol.15 , pp. 2-6
    • Schneider, E.1    Grehn, F.2
  • 3
    • 28444475214 scopus 로고    scopus 로고
    • The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry
    • Koetcha A, White ET, Shewry JM, et al. The relative effects of corneal thickness and age on Goldmann applanation tonometry and dynamic contour tonometry. Br J Ophthalmol. 2005;89:1572-1575.
    • (2005) Br J Ophthalmol , vol.89 , pp. 1572-1575
    • Koetcha, A.1    White, E.T.2    Shewry, J.M.3
  • 4
    • 27744519845 scopus 로고    scopus 로고
    • Clinical comparison of contour and applanation tonometry and their relationship to pachymetry
    • Kniestedt C, Lin S, Choe J, et al. Clinical comparison of contour and applanation tonometry and their relationship to pachymetry. Arch Ophthalmol. 2005;123:1532-1537.
    • (2005) Arch Ophthalmol , vol.123 , pp. 1532-1537
    • Kniestedt, C.1    Lin, S.2    Choe, J.3
  • 5
    • 33751505840 scopus 로고    scopus 로고
    • Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry
    • Martinez-de-la-Casa JM, Garcia-Feijoo J, Vico E, et al. Effect of corneal thickness on dynamic contour, rebound, and goldmann tonometry. Ophthalmology. 2006;113:2156-2162.
    • (2006) Ophthalmology , vol.113 , pp. 2156-2162
    • Martinez-de-la-Casa, J.M.1    Garcia-Feijoo, J.2    Vico, E.3
  • 6
    • 33645823534 scopus 로고    scopus 로고
    • Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry
    • Ku JY, Danesh-Meyer HV, Craig JP, et al. Comparison of intraocular pressure measured by Pascal dynamic contour tonometry and Goldmann applanation tonometry. Eye. 2006;20:191-198.
    • (2006) Eye , vol.20 , pp. 191-198
    • Ku, J.Y.1    Danesh-Meyer, H.V.2    Craig, J.P.3
  • 7
    • 4344659799 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with goldmann applanation tonometry
    • Kaufmann C, Bachmann LM, Thiel MA. Comparison of dynamic contour tonometry with goldmann applanation tonometry. Invest Ophthalmol Vis Sci. 2004;45:3118-3121.
    • (2004) Invest Ophthalmol Vis Sci , vol.45 , pp. 3118-3121
    • Kaufmann, C.1    Bachmann, L.M.2    Thiel, M.A.3
  • 8
    • 33646761304 scopus 로고    scopus 로고
    • Correlation between intraocular pressure, central corneal thickness, stage of glaucoma, and demographic patient data: Prospective analysis of biophysical parameters in tertiary glaucoma practice populations
    • Kniestedt C, Lin S, Choe J, et al. Correlation between intraocular pressure, central corneal thickness, stage of glaucoma, and demographic patient data: prospective analysis of biophysical parameters in tertiary glaucoma practice populations. J Glaucoma. 2006;15: 91-97.
    • (2006) J Glaucoma , vol.15 , pp. 91-97
    • Kniestedt, C.1    Lin, S.2    Choe, J.3
  • 9
    • 20444489648 scopus 로고    scopus 로고
    • Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states
    • Kniestedt C, Nee M, Stamper RL. Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states. Graefes Arch Clin Exp Ophthalmol. 2005;243:359-366.
    • (2005) Graefes Arch Clin Exp Ophthalmol , vol.243 , pp. 359-366
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 10
    • 4444321895 scopus 로고    scopus 로고
    • Dynamic contour tonometry: A comparative study on human cadaver eyes
    • Kniestedt C, Nee M, Stamper RL. Dynamic contour tonometry: a comparative study on human cadaver eyes. Arch Ophthalmol. 2004;122:1287-1293.
    • (2004) Arch Ophthalmol , vol.122 , pp. 1287-1293
    • Kniestedt, C.1    Nee, M.2    Stamper, R.L.3
  • 11
    • 37349104476 scopus 로고    scopus 로고
    • The effect of corneal thickness, corneal curvature and axial length on IOP measurements taken by Pascal dynamic contour tonometry
    • E-Abstract 4444
    • Weber A, Boehm AG, Spoerl E, et al. The effect of corneal thickness, corneal curvature and axial length on IOP measurements taken by Pascal dynamic contour tonometry. Invest Ophthalmol Vis Sci. 2006;47:E-Abstract 4444.
    • (2006) Invest Ophthalmol Vis Sci , pp. 47
    • Weber, A.1    Boehm, A.G.2    Spoerl, E.3
  • 12
    • 37349092562 scopus 로고    scopus 로고
    • Dynamic contour tonometry vs. intracameral IOP measurements-first results
    • E-Abstract 4447
    • Neuburger M, Lautebach S, Pache M, et al. Dynamic contour tonometry vs. intracameral IOP measurements-first results. Invest Ophthalmol Vis Sci. 2006;47:E-Abstract 4447.
    • (2006) Invest Ophthalmol Vis Sci , pp. 47
    • Neuburger, M.1    Lautebach, S.2    Pache, M.3
  • 13
    • 0033822587 scopus 로고    scopus 로고
    • Confirmation of visual field abnormalities in the ocular hypertension treatment study
    • Keltner JL, Johnson CA, Quigg JM, et al. Confirmation of visual field abnormalities in the ocular hypertension treatment study. Arch Ophthalmol. 2000;118:1187-1194.
    • (2000) Arch Ophthalmol , vol.118 , pp. 1187-1194
    • Keltner, J.L.1    Johnson, C.A.2    Quigg, J.M.3
  • 14
    • 0028143590 scopus 로고
    • Advanced Glaucoma Intervention Study: 2. Visual Field Test Scoring and Reliability
    • The Advanced Glaucoma Intervention Study Investigators
    • The Advanced Glaucoma Intervention Study Investigators. Advanced Glaucoma Intervention Study: 2. Visual Field Test Scoring and Reliability. Ophthalmology. 1994;101:1445-1455.
    • (1994) Ophthalmology , vol.101 , pp. 1445-1455
  • 15
    • 22544486553 scopus 로고    scopus 로고
    • Comparison of dynamic contour tonometry with goldmann applanation tonometry over a wide range of central corneal thickness
    • Doyle A, Lachkar Y. Comparison of dynamic contour tonometry with goldmann applanation tonometry over a wide range of central corneal thickness. J Glaucoma. 2005;14:288-292.
    • (2005) J Glaucoma , vol.14 , pp. 288-292
    • Doyle, A.1    Lachkar, Y.2
  • 16
    • 33749158039 scopus 로고    scopus 로고
    • Asymmetries and visual field summaries as predictors of glaucoma in the Ocular Hypertension Treatment Study
    • and the Ocular Hypertension Treatment Study Group
    • Levine RA, Demirel S, Fan J, et al, and the Ocular Hypertension Treatment Study Group. Asymmetries and visual field summaries as predictors of glaucoma in the Ocular Hypertension Treatment Study. Invest Ophthalmol Vis Sci. 2006;47:3896-3903.
    • (2006) Invest Ophthalmol Vis Sci , vol.47 , pp. 3896-3903
    • Levine, R.A.1    Demirel, S.2    Fan, J.3
  • 17
    • 0034603359 scopus 로고    scopus 로고
    • Human corneal thickness and its impact on intraocular pressure measures: A review and meta-analysis approach
    • Doughty MJ, Zaman ML. Human corneal thickness and its impact on intraocular pressure measures: a review and meta-analysis approach. Surv Ophthalmol. 2000;44:367-408.
    • (2000) Surv Ophthalmol , vol.44 , pp. 367-408
    • Doughty, M.J.1    Zaman, M.L.2
  • 18
    • 0036272299 scopus 로고    scopus 로고
    • The Ocular Hypertension Treatment Study: Baseline factors that predict the onset of primary open-angle glaucoma
    • Gordon MO, Beiser JA, Brandt JD, et al. The Ocular Hypertension Treatment Study: baseline factors that predict the onset of primary open-angle glaucoma. Arch Ophthalmol. 2002;120:714-720.
    • (2002) Arch Ophthalmol , vol.120 , pp. 714-720
    • Gordon, M.O.1    Beiser, J.A.2    Brandt, J.D.3
  • 19
    • 33746637170 scopus 로고    scopus 로고
    • Relationship between asymmetric central corneal thickness and glaucomatous visual field loss within the same patient
    • Sullivan-Mee M, Gentry JM, Qualls C. Relationship between asymmetric central corneal thickness and glaucomatous visual field loss within the same patient. Optom Vis Sci. 2006;83: 516-519.
    • (2006) Optom Vis Sci , vol.83 , pp. 516-519
    • Sullivan-Mee, M.1    Gentry, J.M.2    Qualls, C.3
  • 20
    • 0347355073 scopus 로고    scopus 로고
    • Central corneal thickness as a risk factor for advanced glaucoma damage
    • Herndon LW, Weizer JS, Stinnett SS. Central corneal thickness as a risk factor for advanced glaucoma damage. Arch Ophthalmol. 2004;122:17-21.
    • (2004) Arch Ophthalmol , vol.122 , pp. 17-21
    • Herndon, L.W.1    Weizer, J.S.2    Stinnett, S.S.3
  • 21
    • 13844313434 scopus 로고    scopus 로고
    • Influence of corneal biomechanical properties on intraocular pressure measurement: Quantitative analysis
    • Liu J, Roberts CJ. Influence of corneal biomechanical properties on intraocular pressure measurement: quantitative analysis. J Cataract Refract Surg. 2005;31:146-155.
    • (2005) J Cataract Refract Surg , vol.31 , pp. 146-155
    • Liu, J.1    Roberts, C.J.2
  • 22
    • 0016663381 scopus 로고
    • Applanation tonometry and central corneal thickness
    • Ehlers N, Bramsen T, Sperling S. Applanation tonometry and central corneal thickness. Acta Ophthalmol. 1975;53:34-43.
    • (1975) Acta Ophthalmol , vol.53 , pp. 34-43
    • Ehlers, N.1    Bramsen, T.2    Sperling, S.3
  • 23
    • 0027269723 scopus 로고
    • The effect of corneal thickness on applanation tonometry
    • Whitacre MM, Stein RA, Hassanein K. The effect of corneal thickness on applanation tonometry. Am J Ophthalmol. 1993;115: 592-596.
    • (1993) Am J Ophthalmol , vol.115 , pp. 592-596
    • Whitacre, M.M.1    Stein, R.A.2    Hassanein, K.3
  • 24
    • 0035163029 scopus 로고    scopus 로고
    • Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings
    • Feltgen N, Leifert D, Funk J. Correlation between central corneal thickness, applanation tonometry, and direct intracameral IOP readings. Br J Ophthalmol. 2001;85:85-87.
    • (2001) Br J Ophthalmol , vol.85 , pp. 85-87
    • Feltgen, N.1    Leifert, D.2    Funk, J.3
  • 25
    • 0030963090 scopus 로고    scopus 로고
    • Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study
    • Wolfs RC, Klaver CC, Vingerling JR, et al. Distribution of central corneal thickness and its association with intraocular pressure: The Rotterdam Study. Am J Ophthalmol. 1997;123:767-772.
    • (1997) Am J Ophthalmol , vol.123 , pp. 767-772
    • Wolfs, R.C.1    Klaver, C.C.2    Vingerling, J.R.3
  • 26
    • 0033498048 scopus 로고    scopus 로고
    • Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic
    • Shah S, Chatterjee A, Mathai M, et al. Relationship between corneal thickness and measured intraocular pressure in a general ophthalmology clinic. Ophthalmology. 1999;106:2154-2160.
    • (1999) Ophthalmology , vol.106 , pp. 2154-2160
    • Shah, S.1    Chatterjee, A.2    Mathai, M.3
  • 27
    • 0034971648 scopus 로고    scopus 로고
    • Central corneal thickness, tonometry, and ocular dimensions in glaucoma and ocular hypertension
    • Singh RP, Goldberg I, Graham SL, et al. Central corneal thickness, tonometry, and ocular dimensions in glaucoma and ocular hypertension. J Glaucoma. 2001;10:206-210.
    • (2001) J Glaucoma , vol.10 , pp. 206-210
    • Singh, R.P.1    Goldberg, I.2    Graham, S.L.3
  • 28
    • 18644373590 scopus 로고    scopus 로고
    • High interocular corneal symmetry in average simulated keratometry, central corneal thickness, and posterior elevation
    • Myrowitz EH, Kouzis AC, O'Brien TP. High interocular corneal symmetry in average simulated keratometry, central corneal thickness, and posterior elevation. Optom Vis Sci. 2005;82:428-431.
    • (2005) Optom Vis Sci , vol.82 , pp. 428-431
    • Myrowitz, E.H.1    Kouzis, A.C.2    O'Brien, T.P.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.