Volumn 34, Issue 2, 2011, Pages 261-266

The measurement of backscatter factors of kilovoltage X-ray beams using Gafchromic™ EBT2 film

Author keywords

Backscatter factors; Gafchromic film; Kilovoltage X rays

Indexed keywords

ACCURATE MEASUREMENT; BACKSCATTER FACTORS; ENERGY RANGES; FIELD SIZE; GAFCHROMIC EBT; GAFCHROMIC FILMS; ISO GUM; KILOVOLTAGE X-RAYS; MONTE CARLO CALCULATION; OPTICAL DENSITIES; RADIATION DOSE; X RAY BEAM;

EID: 80051679396     PISSN: 01589938     EISSN: None     Source Type: Journal    
DOI: 10.1007/s13246-011-0073-5     Document Type: Article
Times cited : (16)

References (48)
  • 2
    • 20044379949 scopus 로고    scopus 로고
    • Addendum to the IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL)
    • DOI 10.1088/0031-9155/50/12/001
    • RJ Aukett JE Burns AG Greener RM Harrison C Moretti AE Nahum KE Rosser 2005 Addendum to the IPEMB code of practice for the determination of absorbed dose for X-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL) Phys Med Biol 50 12 2739 2748 15930599 10.1088/0031-9155/50/12/001 1:STN:280:DC%2BD2M3os1Wmsg%3D%3D (Pubitemid 40771008)
    • (2005) Physics in Medicine and Biology , vol.50 , Issue.12 , pp. 2739-2748
    • Aukett, R.J.1    Burns, J.E.2    Greener, A.G.3    Harrison, R.M.4    Moretti, C.5    Nahum, A.E.6    Rosser, K.E.7
  • 3
    • 0029846001 scopus 로고    scopus 로고
    • The IPEMB code of practice for the determination of absorbed dose for x-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10-300 kV generating potential)
    • DOI 10.1088/0031-9155/41/12/002
    • SC Klevenhagen RJ Aukett RM Harrison C Moretti AE Nahum KE Rosser 1996 The IPEMB code of practice for the determination of absorbed dose for X-rays below 300 kV generating potential (0.035 mm Al-4 mm Cu HVL; 10-300 kV generating potential) Phys Med Biol 41 12 2605 2625 10.1088/0031-9155/41/12/002 1:CAS:528:DyaK2sXjsFCqug%3D%3D (Pubitemid 26412558)
    • (1996) Physics in Medicine and Biology , vol.41 , Issue.12 , pp. 2605-2625
    • Klevenhagen, S.C.1    Aukett, R.J.2    Harrison, R.M.3    Moretti, C.4    Nahum, A.E.5    Rosser, K.E.6
  • 4
    • 0034973266 scopus 로고    scopus 로고
    • AAPM protocol for 40-300 kV x-ray beam dosimetry in radiotherapy and radiobiology
    • CM Ma CW Coffey LA DeWerd C Liu R Nath SM Seltzer JP Seuntjens 2001 AAPM protocol for 40-300 kV X-ray beam dosimetry in radiotherapy and radiobiology Med Phys 28 6 868 893 11439485 10.1118/1.1374247 1:STN:280:DC%2BD38%2FhvVWmtw%3D%3D (Pubitemid 32574672)
    • (2001) Medical Physics , vol.28 , Issue.6 , pp. 868-893
    • Ma, C.-M.1    Coffey, C.W.2    DeWerd, L.A.3    Liu, C.4    Nath, R.5    Seltzer, S.M.6    Seuntjens, J.P.7
  • 5
    • 80051692064 scopus 로고    scopus 로고
    • Kilovoltage X-rays
    • P. Mayles A.E. Nahum J. Rosenwald (eds). CRC Press Boca Raton. 10.1201/9781420012026.ch21
    • Mayles P (2007) Kilovoltage X-rays. In: Mayles P, Nahum AE, Rosenwald J (eds) Handbook of radiotherapy physics. CRC Press, Boca Raton, pp 439-449
    • (2007) Handbook of Radiotherapy Physics , pp. 439-449
    • Mayles, P.1
  • 6
    • 0027402558 scopus 로고
    • Dependence of the photon backscatter factor for water on irradiation field size and source-to-phantom distances between 1.5 and 10 cm
    • B Grosswendt 1993 Dependence of the photon backscatter factor for water on irradiation field size and source-to-phantom distances between 1.5 and 10 cm Phys Med Biol 38 2 305 310 10.1088/0031-9155/38/2/007 (Pubitemid 23054595)
    • (1993) Physics in Medicine and Biology , vol.38 , Issue.2 , pp. 305-310
    • Grosswendt, B.1
  • 7
    • 58149269275 scopus 로고    scopus 로고
    • Study of the formalism used to determine the absorbed dose for low-energy X-ray beams
    • 19001702 10.1088/0031-9155/53/23/020 1:STN:280:DC%2BD1cjjt1OntQ%3D%3D
    • U Chica M Anguiano AM Lallena 2008 Study of the formalism used to determine the absorbed dose for low-energy X-ray beams Phys Med Biol 53 23 6963 6977 19001702 10.1088/0031-9155/53/23/020 1:STN:280:DC%2BD1cjjt1OntQ%3D%3D
    • (2008) Phys Med Biol , vol.53 , Issue.23 , pp. 6963-6977
    • Chica, U.1    Anguiano, M.2    Lallena, A.M.3
  • 8
    • 42249088602 scopus 로고    scopus 로고
    • Dose reduction from loss of backscatter in superficial x-ray radiation therapy with the Pantak SXT 150 unit
    • BJ Healy S Sylvander KN Nitschke 2008 Dose reduction from loss of backscatter in superficial X-ray radiation therapy with the Pantak SXT 150 unit Austral Phys Eng Sci Med 31 1 49 55 10.1007/BF03178453 1:STN:280: DC%2BD1czis1Gksg%3D%3D (Pubitemid 351546625)
    • (2008) Australasian Physical and Engineering Sciences in Medicine , vol.31 , Issue.1 , pp. 49-55
    • Healy, B.J.1    Sylvander, S.2    Nitschke, K.N.3
  • 9
    • 34347370459 scopus 로고    scopus 로고
    • An investigation of dose changes for therapeutic kilovoltage x-ray beams with underlying lead shielding
    • DOI 10.1118/1.2748103
    • R Hill B Healy L Holloway C Baldock 2007 An investigation of dose changes for therapeutic kilovoltage X-ray beams with underlying lead shielding Med Phys 34 7 3045 3053 17822012 10.1118/1.2748103 (Pubitemid 47019825)
    • (2007) Medical Physics , vol.34 , Issue.7 , pp. 3045-3053
    • Hill, R.1    Healy, B.2    Holloway, L.3    Baldock, C.4
  • 10
    • 0024355954 scopus 로고
    • Experimentally determined backscatter factors for x-rays generated at voltages between 16 and 140 kV
    • SC Klevenhagen 1989 Experimentally determined backscatter factors for X-rays generated at voltages between 16 and 140 kV Phys Med Biol 34 12 1871 1882 10.1088/0031-9155/34/12/010 (Pubitemid 20004954)
    • (1989) Physics in Medicine and Biology , vol.34 , Issue.12 , pp. 1871-1882
    • Klevenhagen, S.C.1
  • 11
    • 0344267664 scopus 로고    scopus 로고
    • Mass-energy absorption coefficient and backscatter factor ratios for kilovoltage x-ray beams
    • DOI 10.1088/0031-9155/44/1/011, PII S0031915599925435
    • CM Ma JP Seuntjens 1999 Mass-energy absorption coefficient and backscatter factor ratios for kilovoltage X-ray beams Phys Med Biol 44 1 131 143 10071880 10.1088/0031-9155/44/1/011 1:STN:280:DyaK1M7ntVSqtA%3D%3D (Pubitemid 29062232)
    • (1999) Physics in Medicine and Biology , vol.44 , Issue.1 , pp. 131-143
    • Ma, C.-M.1    Seuntjens, J.P.2
  • 12
    • 0025812545 scopus 로고
    • Memorandum from the Institute of Physical Sciences in Medicine Back-scatter and F-factors for low- and medium-energy X-ray beams in radiotherapy
    • 1913048 10.1259/0007-1285-64-765-836 1:STN:280:DyaK38%2FgsVGrsQ%3D%3D
    • SC Klevenhagen RJ Aukett JE Burns RM Harrison RT Knight AE Nahum KE Rosser 1991 Memorandum from the Institute of Physical Sciences in Medicine Back-scatter and F-factors for low- and medium-energy X-ray beams in radiotherapy Br J Radiol 64 765 836 841 1913048 10.1259/0007-1285-64-765-836 1:STN:280:DyaK38%2FgsVGrsQ%3D%3D
    • (1991) Br J Radiol , vol.64 , Issue.765 , pp. 836-841
    • Klevenhagen, S.C.1    Aukett, R.J.2    Burns, J.E.3    Harrison, R.M.4    Knight, R.T.5    Nahum, A.E.6    Rosser, K.E.7
  • 13
    • 0027512589 scopus 로고
    • Differences in reported backscatter factors for low-energy X-rays: A literature study
    • 10.1088/0031-9155/38/4/004
    • CA Carlsson 1993 Differences in reported backscatter factors for low-energy X-rays: a literature study Phys Med Biol 38 4 521 10.1088/0031-9155/38/4/004
    • (1993) Phys Med Biol , vol.38 , Issue.4 , pp. 521
    • Carlsson, C.A.1
  • 14
    • 0031943186 scopus 로고    scopus 로고
    • Lithium borate TLD for determining the backscatter factors for low- energy x rays: Comparison with chamber-based and Monte Carlo derived values
    • DOI 10.1118/1.598205
    • D Coudin G Marinello 1998 Lithium borate TLD for determining the backscatter factors for low- energy X rays: comparison with chamber-based and Monte Carlo derived values Med Phys 25 3 347 353 9547502 10.1118/1.598205 1:STN:280:DyaK1c3gvVOiug%3D%3D (Pubitemid 28146391)
    • (1998) Medical Physics , vol.25 , Issue.3 , pp. 347-353
    • Coudin, D.1    Marinello, G.2
  • 15
    • 0019948998 scopus 로고
    • The build-up of backscatter in the energy range 1 mm Al to 8 mm Al HVT
    • SC Klevenhagen 1982 The build-up of backscatter in the energy range 1 mm Al to 8 mm Al HVT (radiotherapy beams) Phys Med Biol 27 8 1035 1043 10.1088/0031-9155/27/8/005 1:CAS:528:DyaL38XltFOitL4%3D (Pubitemid 12042194)
    • (1982) Physics in Medicine and Biology , vol.27 , Issue.8 , pp. 1035-1043
    • Klevenhagen, S.C.1
  • 16
    • 0025178893 scopus 로고
    • Measurement of backscatter factors for low energy radiotherapy (0.1-2.0 mm Al HVL) using thermoluminescence dosimetry
    • DOI 10.1088/0031-9155/35/9/005
    • RM Harrison C Walker RJ Aukett 1990 Measurement of backscatter factors for low energy radiotherapy (0.1-2.0 mm Al HVL) using thermoluminescence dosimetry Phys Med Biol 35 9 1247 2236206 10.1088/0031-9155/35/9/005 1:STN:280:DyaK3M%2Fks1eisg%3D%3D (Pubitemid 20264598)
    • (1990) Physics in Medicine and Biology , vol.35 , Issue.9 , pp. 1247-1254
    • Harrison, R.M.1    Walker, C.2    Aukett, R.J.3
  • 17
    • 0032878123 scopus 로고    scopus 로고
    • Water phantom backscatter factors for x-rays in the 60 kV to 300 kV region
    • DOI 10.1088/0031-9155/44/9/307, PII S0031915599005357
    • II Tomljenovic MM Ninkovic D Bek-Uzarov SJ Stankovic M Kovacevic 1999 Water phantom backscatter factors for X-rays in the 60 kV to 300 kV region Phys Med Biol 44 9 2193 2200 10495114 10.1088/0031-9155/44/9/307 1:STN:280: DyaK1Mvitlejtw%3D%3D (Pubitemid 29424380)
    • (1999) Physics in Medicine and Biology , vol.44 , Issue.9 , pp. 2193-2200
    • Tomljenovic, I.I.1    Ninkovic, M.M.2    Bek-Uzarov, D.3    Stankovic, S.J.4    Kovacevic, M.5
  • 19
    • 0021131849 scopus 로고
    • Backscatter factors for x-rays generated at voltages between 10 and 100 kV
    • B Grosswendt 1984 Backscatter factors for X-rays generated at voltages between 10 and 100 kV Phys Med Biol 29 5 579 591 6739543 10.1088/0031-9155/29/5/ 010 1:STN:280:DyaL2c3kvValtg%3D%3D (Pubitemid 14054634)
    • (1984) Physics in Medicine and Biology , vol.29 , Issue.5 , pp. 579-591
    • Grosswendt, B.1
  • 20
    • 0031858317 scopus 로고    scopus 로고
    • Calculation of backscatter factors for diagnostic radiology using Monte Carlo methods
    • DOI 10.1088/0031-9155/43/8/017, PII S0031915598869479
    • N Petoussi-Henss M Zankl G Drexler W Panzer D Regulla 1998 Calculation of backscatter factors for diagnostic radiology using Monte Carlo methods Phys Med Biol 43 8 2237 9725601 10.1088/0031-9155/43/8/017 1:STN:280: DyaK1czptFCitQ%3D%3D (Pubitemid 28372620)
    • (1998) Physics in Medicine and Biology , vol.43 , Issue.8 , pp. 2237-2250
    • Petoussi-Henss, N.1    Zankl, M.2    Drexler, G.3    Panzer, W.4    Regulla, D.5
  • 22
    • 51849167775 scopus 로고    scopus 로고
    • Kilovoltage X-ray dosimetry-an experimental comparison between different dosimetry protocols
    • 18670053 10.1088/0031-9155/53/16/014
    • P Munck Af Rosenschold P Nilsson T Knoos 2008 Kilovoltage X-ray dosimetry-an experimental comparison between different dosimetry protocols Phys Med Biol 53 16 4431 4442 18670053 10.1088/0031-9155/53/16/014
    • (2008) Phys Med Biol , vol.53 , Issue.16 , pp. 4431-4442
    • Munck, P.1    Rosenschold, A.F.2    Nilsson, P.3    Knoos, T.4
  • 23
    • 8544270156 scopus 로고    scopus 로고
    • Experimental energy response verification of XR type T radiochromic film
    • DOI 10.1088/0031-9155/49/21/N02, PII S0031915504845603
    • T Cheung MJ Butson PKN Yu 2004 Experimental energy response verification of XR type T radiochromic film Phys Med Biol 49 21 N371 N376 15584536 10.1088/0031-9155/49/21/N02 (Pubitemid 39489941)
    • (2004) Physics in Medicine and Biology , vol.49 , Issue.21
    • Cheung, T.1    Butson, M.J.2    Yu, P.K.N.3
  • 24
    • 27644471150 scopus 로고    scopus 로고
    • Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range
    • DOI 10.1016/j.apradiso.2005.07.002, PII S0969804305002332
    • MJ Butson T Cheung PKN Yu 2006 Weak energy dependence of EBT gafchromic film dose response in the 50 kVp-10 MVp X-ray range Appl Radiat Isot 64 1 60 62 16105740 10.1016/j.apradiso.2005.07.002 1:CAS:528:DC%2BD2MXhtFGnu7fL (Pubitemid 41550172)
    • (2006) Applied Radiation and Isotopes , vol.64 , Issue.1 , pp. 60-62
    • Butson, M.J.1    Cheung, T.2    Yu, P.K.N.3
  • 25
    • 28044452190 scopus 로고    scopus 로고
    • Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies
    • DOI 10.1118/1.2065467
    • ST Chiu-Tsao Y Ho R Shankar L Wang LB Harrison 2005 Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies Med Phys 32 11 3350 3354 16370422 10.1118/1.2065467 (Pubitemid 41689286)
    • (2005) Medical Physics , vol.32 , Issue.11 , pp. 3350-3354
    • Chiu-Tsao, S.-T.1    Ho, Y.2    Shankar, R.3    Wang, L.4    Harrison, L.B.5
  • 26
    • 33846582373 scopus 로고    scopus 로고
    • Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter
    • DOI 10.1118/1.2431425
    • A Rink IA Vitkin DA Jaffray 2007 Energy dependence (75 kVp to 18 MV) of radiochromic films assessed using a real-time optical dosimeter Med Phys 34 2 458 463 17388161 10.1118/1.2431425 1:CAS:528:DC%2BD2sXhsFCgt7w%3D (Pubitemid 46185123)
    • (2007) Medical Physics , vol.34 , Issue.2 , pp. 458-463
    • Rink, A.1    Vitkin, I.A.2    Jaffray, D.A.3
  • 27
    • 38949201305 scopus 로고    scopus 로고
    • Measurement of radiotherapy superficial X-ray dose under eye shields with radiochromic film
    • 18162428 10.1016/j.ejmp.2007.11.001
    • MJ Butson T Cheung PK Yu S Price M Bailey 2008 Measurement of radiotherapy superficial X-ray dose under eye shields with radiochromic film Phys Med 24 1 29 33 18162428 10.1016/j.ejmp.2007.11.001
    • (2008) Phys Med , vol.24 , Issue.1 , pp. 29-33
    • Butson, M.J.1    Cheung, T.2    Yu, P.K.3    Price, S.4    Bailey, M.5
  • 29
    • 45749158708 scopus 로고    scopus 로고
    • An assessment of GafChromic film for measuring 50 kV and 100 kV percentage depth dose curves
    • DOI 10.1088/0031-9155/53/11/N02, PII S0031915508702124
    • CL Fletcher JA Mills 2008 An assessment of GafChromic film for measuring 50 kV and 100 kV percentage depth dose curves Phys Med Biol 53 11 N209 N218 18490809 10.1088/0031-9155/53/11/N02 (Pubitemid 351868626)
    • (2008) Physics in Medicine and Biology , vol.53 , Issue.11
    • Fletcher, C.L.1    Mills, J.A.2
  • 30
    • 76149125204 scopus 로고    scopus 로고
    • An investigation of backscatter factors for kilovoltage X-rays: A comparison between Monte Carlo simulations and Gafchromic EBT film measurements
    • 20071763 10.1088/0031-9155/55/3/016 1:STN:280:DC%2BC3c%2Fitl2gtg%3D%3D
    • J Kim R Hill E Claridge MacKonis Z Kuncic 2010 An investigation of backscatter factors for kilovoltage X-rays: a comparison between Monte Carlo simulations and Gafchromic EBT film measurements Phys Med Biol 55 3 783 797 20071763 10.1088/0031-9155/55/3/016 1:STN:280:DC%2BC3c%2Fitl2gtg%3D%3D
    • (2010) Phys Med Biol , vol.55 , Issue.3 , pp. 783-797
    • Kim, J.1    Hill, R.2    Claridge MacKonis, E.3    Kuncic, Z.4
  • 31
    • 77952395736 scopus 로고    scopus 로고
    • Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies
    • 20527528 10.1118/1.3373523 1:CAS:528:DC%2BC3cXlsFWjsbc%3D
    • B Arjomandy R Tailor A Anand N Sahoo M Gillin K Prado M Vicic 2010 Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies Med Phys 37 5 1942 1947 20527528 10.1118/1.3373523 1:CAS:528:DC%2BC3cXlsFWjsbc%3D
    • (2010) Med Phys , vol.37 , Issue.5 , pp. 1942-1947
    • Arjomandy, B.1    Tailor, R.2    Anand, A.3    Sahoo, N.4    Gillin, M.5    Prado, K.6    Vicic, M.7
  • 32
    • 77950551727 scopus 로고    scopus 로고
    • Technical note: Homogeneity of gafchromic® EBT2 film
    • 20443496 10.1118/1.3368601
    • B Hartmann M Martišíková O Jäkel 2010 Technical note: homogeneity of gafchromic® EBT2 film Med Phys 37 4 1753 1756 20443496 10.1118/1.3368601
    • (2010) Med Phys , vol.37 , Issue.4 , pp. 1753-1756
    • Hartmann, B.1    Martišíková, M.2    Jäkel, O.3
  • 33
    • 77951152389 scopus 로고    scopus 로고
    • Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy
    • 20393235 10.1088/0031-9155/55/9/012 1:STN:280:DC%2BC3c3mvFKgsg%3D%3D
    • L Richley AC John H Coomber S Fletcher 2010 Evaluation and optimization of the new EBT2 radiochromic film dosimetry system for patient dose verification in radiotherapy Phys Med Biol 55 9 2601 2617 20393235 10.1088/0031-9155/55/9/ 012 1:STN:280:DC%2BC3c3mvFKgsg%3D%3D
    • (2010) Phys Med Biol , vol.55 , Issue.9 , pp. 2601-2617
    • Richley, L.1    John, A.C.2    Coomber, H.3    Fletcher, S.4
  • 34
    • 77749292421 scopus 로고    scopus 로고
    • Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film
    • 20384246 10.1118/1.3301574 1:CAS:528:DC%2BC3cXivFaju7w%3D
    • JGH Sutherland DWO Rogers 2010 Monte Carlo calculated absorbed-dose energy dependence of EBT and EBT2 film Med Phys 37 3 1110 1116 20384246 10.1118/1.3301574 1:CAS:528:DC%2BC3cXivFaju7w%3D
    • (2010) Med Phys , vol.37 , Issue.3 , pp. 1110-1116
    • Sutherland, J.G.H.1    Rogers, D.W.O.2
  • 35
    • 75749154506 scopus 로고    scopus 로고
    • Investigation of energy dependence of EBT and EBT-2 Gafchromic film
    • 20229865 10.1118/1.3291622 1:CAS:528:DC%2BC3cXhsVGksLc%3D
    • P Lindsay A Rink M Ruschin D Jaffray 2010 Investigation of energy dependence of EBT and EBT-2 Gafchromic film Med Phys 37 2 571 576 20229865 10.1118/1.3291622 1:CAS:528:DC%2BC3cXhsVGksLc%3D
    • (2010) Med Phys , vol.37 , Issue.2 , pp. 571-576
    • Lindsay, P.1    Rink, A.2    Ruschin, M.3    Jaffray, D.4
  • 36
    • 33645519422 scopus 로고    scopus 로고
    • Accurate skin dose measurements using radiochromic film in clinical applications
    • 16696489 10.1118/1.2179169 1:STN:280:DC%2BD283mslWrsQ%3D%3D
    • S Devic J Seuntjens W Abdel-Rahman M Evans M Olivares EB Podgorsak T Vuong CG Soares 2006 Accurate skin dose measurements using radiochromic film in clinical applications Med Phys 33 4 1116 1124 16696489 10.1118/1.2179169 1:STN:280:DC%2BD283mslWrsQ%3D%3D
    • (2006) Med Phys , vol.33 , Issue.4 , pp. 1116-1124
    • Devic, S.1    Seuntjens, J.2    Abdel-Rahman, W.3    Evans, M.4    Olivares, M.5    Podgorsak, E.B.6    Vuong, T.7    Soares, C.G.8
  • 37
    • 78149480566 scopus 로고    scopus 로고
    • Enhanced epidermal dose caused by localized electron contamination from lead cutouts used in kilovoltage radiotherapy
    • 20879556 10.1118/1.3458722 1:STN:280:DC%2BC3cfmtVynsw%3D%3D
    • JE Lye DJ Butler DV Webb 2010 Enhanced epidermal dose caused by localized electron contamination from lead cutouts used in kilovoltage radiotherapy Med Phys 37 8 3935 3939 20879556 10.1118/1.3458722 1:STN:280:DC%2BC3cfmtVynsw%3D%3D
    • (2010) Med Phys , vol.37 , Issue.8 , pp. 3935-3939
    • Lye, J.E.1    Butler, D.J.2    Webb, D.V.3
  • 38
    • 78149484465 scopus 로고    scopus 로고
    • The water equivalence of solid phantoms for low energy photon beams
    • 20879595 10.1118/1.3462558 1:CAS:528:DC%2BC3cXps1aks70%3D
    • R Hill Z Kuncic C Baldock 2010 The water equivalence of solid phantoms for low energy photon beams Med Phys 37 8 4355 4363 20879595 10.1118/1.3462558 1:CAS:528:DC%2BC3cXps1aks70%3D
    • (2010) Med Phys , vol.37 , Issue.8 , pp. 4355-4363
    • Hill, R.1    Kuncic, Z.2    Baldock, C.3
  • 40
    • 27144551316 scopus 로고    scopus 로고
    • A dosimetric evaluation of water equivalent phantoms for kilovoltage x-ray beams
    • DOI 10.1088/0031-9155/50/21/N06
    • R Hill L Holloway C Baldock 2005 A dosimetric evaluation of water equivalent phantoms for kilovoltage X-ray beams Phys Med Biol 50 21 N331 N344 16237233 10.1088/0031-9155/50/21/N06 1:STN:280:DC%2BD2MrmtVOgtg%3D%3D (Pubitemid 41509169)
    • (2005) Physics in Medicine and Biology , vol.50 , Issue.21
    • Hill, R.1    Holloway, L.2    Baldock, C.3
  • 41
    • 78650581672 scopus 로고    scopus 로고
    • Study of dosimetric water equivalency of PRESAGE® for megavoltage and kilovoltage X-ray beams
    • T Gorjiara R Hill JH Kim Z Kuncic J Adamovics C Baldock 2010 Study of dosimetric water equivalency of PRESAGE® for megavoltage and kilovoltage X-ray beams J Phys Conf Ser 250 244 248
    • (2010) J Phys Conf ser , vol.250 , pp. 244-248
    • Gorjiara, T.1    Hill, R.2    Kim, J.H.3    Kuncic, Z.4    Adamovics, J.5    Baldock, C.6
  • 42
    • 0029362717 scopus 로고
    • Dose characteristics of a new 300 kVp orthovoltage machine
    • 1:STN:280:DyaK287hsVaquw%3D%3D
    • MJ Butson J Mathur PE Metcalfe 1995 Dose characteristics of a new 300 kVp orthovoltage machine Austral Phys Eng Sci Med 18 3 133 138 1:STN:280: DyaK287hsVaquw%3D%3D
    • (1995) Austral Phys Eng Sci Med , vol.18 , Issue.3 , pp. 133-138
    • Butson, M.J.1    Mathur, J.2    Metcalfe, P.E.3
  • 43
    • 0033979869 scopus 로고    scopus 로고
    • Electron contamination from the lead cutout used in kilovoltage radiotherapy
    • DOI 10.1088/0031-9155/45/1/301, PII S003191550005819X
    • CHM Lee KKD Chan 2000 Electron contamination from the lead cutout used in kilovoltage radiotherapy Phys Med Biol 45 1 1 8 10661579 10.1088/0031-9155/45/ 1/301 1:STN:280:DC%2BD3c7isFahsw%3D%3D (Pubitemid 30060656)
    • (2000) Physics in Medicine and Biology , vol.45 , Issue.1 , pp. 1-8
    • Lee, C.H.M.1    Chan, K.K.D.2
  • 44
    • 0025916626 scopus 로고
    • Complications in low energy X-ray dosimetry caused by electron contamination
    • 10.1088/0031-9155/36/8/006
    • SC Klevenhagen D D'Souza I Bonnefoux 1991 Complications in low energy X-ray dosimetry caused by electron contamination Phys Med Biol 36 8 1111 1116 10.1088/0031-9155/36/8/006
    • (1991) Phys Med Biol , vol.36 , Issue.8 , pp. 1111-1116
    • Klevenhagen, S.C.1    D'Souza, D.2    Bonnefoux, I.3
  • 47
    • 40049109564 scopus 로고    scopus 로고
    • The effective point of measurement of ionization chambers and the build-up anomaly in MV x-ray beams
    • DOI 10.1118/1.2839329
    • MR McEwen I Kawrakow CK Ross 2008 The effective point of measurement of ionization chambers and the build-up anomaly in MV X-ray beams Med Phys 35 3 950 958 18404931 10.1118/1.2839329 1:STN:280:DC%2BD1c3ks1Chuw%3D%3D (Pubitemid 351322676)
    • (2008) Medical Physics , vol.35 , Issue.3 , pp. 950-958
    • McEwen, M.R.1    Kawrakow, I.2    Ross, C.K.3
  • 48
    • 77950579345 scopus 로고    scopus 로고
    • Technical note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry
    • 20443477 10.1118/1.3355895
    • J Desroches H Bouchard F Lacroix 2010 Technical note: Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry Med Phys 37 4 1565 1570 20443477 10.1118/1.3355895
    • (2010) Med Phys , vol.37 , Issue.4 , pp. 1565-1570
    • Desroches, J.1    Bouchard, H.2    Lacroix, F.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.