Technical Evaluation of the Novel Preanalytical Module on Instrumentation Laboratory ACL TOP: Advancing Automation in Hemostasis Testing

Author keywords

automation; interference; preanalytical variability; quality

Indexed keywords

CLINICAL CHEMISTRY; COAGULATION ASSAYS; COMPARISON STUDY; INSTRUMENTATION LABORATORIES; INTERFERING SUBSTANCES; OPTIMAL PRECISION; PREANALYTICAL VARIABILITY; TECHNICAL EVALUATION;

EID: 84887459288     PISSN: 22110682     EISSN: 22110690     Source Type: Journal    
DOI: 10.1177/2211068213491747     Document Type: Article
Times cited : (36)

References (24)
  • 1
    • 84865372924 scopus 로고    scopus 로고
    • Preanalytical Phase-A Continuous Challenge for Laboratory Professionals
    • Simundic A. M.,Lippi G.Preanalytical Phase-A Continuous Challenge for Laboratory Professionals.Biochem. Med. (Zagreb. 2012;22:145-149.
    • (2012) Biochem. Med. (Zagreb , vol.22 , pp. 145-149
    • Simundic, A.M.1    Lippi, G.2
  • 2
    • 79960272717 scopus 로고    scopus 로고
    • Preanalytical Quality Improvement: From Dream to Reality
    • Lippi G.,Chance J. J.,Church S.,et al.Preanalytical Quality Improvement: From Dream to Reality.Clin. Chem. Lab. Med. 2011;49:1113-1126.
    • (2011) Clin. Chem. Lab. Med , vol.49 , pp. 1113-1126
    • Lippi, G.1    Chance, J.J.2    Church, S.3
  • 3
    • 84865866502 scopus 로고    scopus 로고
    • Patient Safety and Quality in Laboratory and Hemostasis Testing: A Renewed Loop?
    • Plebani M.,Favaloro E. J.,Lippi G.Patient Safety and Quality in Laboratory and Hemostasis Testing: A Renewed Loop?.Semin. Thromb. Hemost. 2012;38:553-558.
    • (2012) Semin. Thromb. Hemost , vol.38 , pp. 553-558
    • Plebani, M.1    Favaloro, E.J.2    Lippi, G.3
  • 4
    • 84866085262 scopus 로고    scopus 로고
    • Quality Standards for Sample Processing, Transportation, and Storage in Hemostasis Testing
    • Adcock Funk D. M.,Lippi G.,Favaloro E. J.Quality Standards for Sample Processing, Transportation, and Storage in Hemostasis Testing.Semin. Thromb. Hemost. 2012;38:576-585.
    • (2012) Semin. Thromb. Hemost , vol.38 , pp. 576-585
    • Adcock Funk, D.M.1    Lippi, G.2    Favaloro, E.J.3
  • 5
    • 84866078801 scopus 로고    scopus 로고
    • Quality Standards for Sample Collection in Coagulation Testing
    • Lippi G.,Salvagno G. L.,Montagnana M.,et al.Quality Standards for Sample Collection in Coagulation Testing.Semin. Thromb. Hemost. 2012;38:565-575.
    • (2012) Semin. Thromb. Hemost , vol.38 , pp. 565-575
    • Lippi, G.1    Salvagno, G.L.2    Montagnana, M.3
  • 6
    • 84867272037 scopus 로고    scopus 로고
    • Preanalytical Variables in Coagulation Testing Associated to Diagnostic Errors in Hemostasis
    • Favaloro E. J.,Funk (Adcock) D. M.,Lippi G.Preanalytical Variables in Coagulation Testing Associated to Diagnostic Errors in Hemostasis.Lab. Med. 2012;43:54-60.
    • (2012) Lab. Med , vol.43 , pp. 54-60
    • Favaloro, E.J.1    Funk Adcock, D.M.2    Lippi, G.3
  • 7
    • 84884211546 scopus 로고    scopus 로고
    • Interference in Coagulation Testing: Focus on Spurious Hemolysis, Icterus, and Lipemia
    • Lippi G.,Plebani M.,Favaloro E. J.Interference in Coagulation Testing: Focus on Spurious Hemolysis, Icterus, and Lipemia.Semin. Thromb. Hemost. 2013;39:258-266.
    • (2013) Semin. Thromb. Hemost , vol.39 , pp. 258-266
    • Lippi, G.1    Plebani, M.2    Favaloro, E.J.3
  • 8
    • 84878332967 scopus 로고    scopus 로고
    • Continuous-Flow Automation and Hemolysis Index: A Crucial Combination
    • Lippi G.,Plebani M.Continuous-Flow Automation and Hemolysis Index: A Crucial Combination.J. Lab. Autom. 2013;18:184-188.
    • (2013) J. Lab. Autom , vol.18 , pp. 184-188
    • Lippi, G.1    Plebani, M.2
  • 9
    • 34547934654 scopus 로고    scopus 로고
    • Evaluation and Performance Characteristics of the automated coagulation analyzer ACL TOP
    • Appert-Flory A.,Fischer F.,Jambou D.,et al.Evaluation and Performance Characteristics of the automated coagulation analyzer ACL TOP.Thromb. Res. 2007;120:733-743.
    • (2007) Thromb. Res , vol.120 , pp. 733-743
    • Appert-Flory, A.1    Fischer, F.2    Jambou, D.3
  • 10
    • 58449134379 scopus 로고    scopus 로고
    • Evaluation and Performance Characteristics of the Coagulation System: ACL TOP Analyzer-HemosIL Reagents
    • Milos M.,Herak D.,Kuric L.,et al.Evaluation and Performance Characteristics of the Coagulation System: ACL TOP Analyzer-HemosIL Reagents.Int. J. Lab. Hematol. 2009;31:26-35.
    • (2009) Int. J. Lab. Hematol , vol.31 , pp. 26-35
    • Milos, M.1    Herak, D.2    Kuric, L.3
  • 11
    • 84859812541 scopus 로고    scopus 로고
    • Interference Studies: Focus on Blood Cell Lysates Preparation and Testing
    • Lippi G.Interference Studies: Focus on Blood Cell Lysates Preparation and Testing.Clin. Lab. 2012;58:351-355.
    • (2012) Clin. Lab , vol.58 , pp. 351-355
    • Lippi, G.1
  • 12
    • 84857441940 scopus 로고    scopus 로고
    • Influence of In Vitro Hemolysis on Hematological Testing on Advia 2120
    • Lippi G.,Musa R.,Avanzini P.,et al.Influence of In Vitro Hemolysis on Hematological Testing on Advia 2120.Int. J. Lab. Hematol. 2012;34:179-184.
    • (2012) Int. J. Lab. Hematol , vol.34 , pp. 179-184
    • Lippi, G.1    Musa, R.2    Avanzini, P.3
  • 13
    • 35348958435 scopus 로고    scopus 로고
    • Evaluation of the Unicel DXC 600 (Beckman Coulter) and Comparison with Integra 800 (Roche Diagnostics)]
    • Servonnet A.,Théfenne H.,Boukhira A.,et al.Evaluation of the Unicel DXC 600 (Beckman Coulter) and Comparison with Integra 800 (Roche Diagnostics)].Ann. Biol. Clin. (Paris). 2007;65:555-562.
    • (2007) Ann. Biol. Clin. (Paris) , vol.65 , pp. 555-562
    • Servonnet, A.1    Théfenne, H.2    Boukhira, A.3
  • 14
    • 70349484951 scopus 로고    scopus 로고
    • Multicenter Evaluation of the Hemolysis Index in Automated Clinical Chemistry Systems
    • Lippi G.,Luca Salvagno G.,Blanckaert N.,et al.Multicenter Evaluation of the Hemolysis Index in Automated Clinical Chemistry Systems.Clin. Chem. Lab. Med. 2009;47:934-939.
    • (2009) Clin. Chem. Lab. Med , vol.47 , pp. 934-939
    • Lippi, G.1    Luca Salvagno, G.2    Blanckaert, N.3
  • 15
    • 84858400592 scopus 로고    scopus 로고
    • Evaluation of the Analytical Performances of the Novel Beckman Coulter AU5800
    • Lippi G.,Dipalo M.,Musa R.,et al.Evaluation of the Analytical Performances of the Novel Beckman Coulter AU5800.Clin. Biochem. 2012;45:502-504.
    • (2012) Clin. Biochem , vol.45 , pp. 502-504
    • Lippi, G.1    Dipalo, M.2    Musa, R.3
  • 16
    • 0033369894 scopus 로고    scopus 로고
    • Current Databases on Biologic Variation: Pros, Cons and Progress
    • Ricos C.,Alvarez V.,Cava F.,et al.Current Databases on Biologic Variation: Pros, Cons and Progress.Scand. J. Clin. Lab. Invest. 1999;59:491-500.
    • (1999) Scand. J. Clin. Lab. Invest , vol.59 , pp. 491-500
    • Ricos, C.1    Alvarez, V.2    Cava, F.3
  • 17
    • 80052289159 scopus 로고    scopus 로고
    • Hemolyzed Specimens: A Major Challenge for Emergency Departments and Clinical Laboratories
    • Lippi G.,Plebani M.,Di Somma S.,et al.Hemolyzed Specimens: A Major Challenge for Emergency Departments and Clinical Laboratories.Crit. Rev. Clin. Lab. Sci. 2011;48:143-153.
    • (2011) Crit. Rev. Clin. Lab. Sci , vol.48 , pp. 143-153
    • Lippi, G.1    Plebani, M.2    Di Somma, S.3
  • 19
    • 34147208730 scopus 로고    scopus 로고
    • Emerging Technologies in Hemostasis Diagnostics: A Report from the Australasian Society of Thrombosis and Haemostasis Emerging Technologies Group
    • Adams M.,Ward C.,Thom J.,et al.Emerging Technologies in Hemostasis Diagnostics: A Report from the Australasian Society of Thrombosis and Haemostasis Emerging Technologies Group.Semin. Thromb. Hemost. 2007;33:226-234.
    • (2007) Semin. Thromb. Hemost , vol.33 , pp. 226-234
    • Adams, M.1    Ward, C.2    Thom, J.3
  • 20
    • 84887431481 scopus 로고    scopus 로고
    • Advancing Haemostasis Automation: Successful Implementation of Robotic Centrifugation and Sample Processing in a tertiary service hospital
    • Sédille-Mostafaie N.,Engler H.,Lutz S.,et al.Advancing Haemostasis Automation: Successful Implementation of Robotic Centrifugation and Sample Processing in a tertiary service hospital.Clin. Chem. Lab. Med. 2012;15:1-6.
    • (2012) Clin. Chem. Lab. Med , vol.15 , pp. 1-6
    • Sédille-Mostafaie, N.1    Engler, H.2    Lutz, S.3
  • 21
    • 84872951853 scopus 로고    scopus 로고
    • Laboratory Hemostasis: Milestones in Clinical Chemistry and Laboratory Medicine
    • Lippi G.,Favaloro E. J.Laboratory Hemostasis: Milestones in Clinical Chemistry and Laboratory Medicine.Clin. Chem. Lab. Med. 2013;51:91-97.
    • (2013) Clin. Chem. Lab. Med , vol.51 , pp. 91-97
    • Lippi, G.1    Favaloro, E.J.2
  • 22
    • 18844444286 scopus 로고    scopus 로고
    • High Throughput Coagulation Analyzers Review
    • Qari M. H.High Throughput Coagulation Analyzers Review.Comb. Chem. High Throughput Screen. 2005;8:353-360.
    • (2005) Comb. Chem. High Throughput Screen , vol.8 , pp. 353-360
    • Qari, M.H.1
  • 23
    • 82255169401 scopus 로고    scopus 로고
    • Evaluation of the Automated Coagulation Analyzer Sysmex® CS-2100i (Siemens)
    • Martin-Toutain I.,Jobic L.,Mancic T.,et al.Evaluation of the Automated Coagulation Analyzer Sysmex® CS-2100i (Siemens).Ann. Biol. Clin. (Paris). 2011;69:699-704.
    • (2011) Ann. Biol. Clin. (Paris) , vol.69 , pp. 699-704
    • Martin-Toutain, I.1    Jobic, L.2    Mancic, T.3
  • 24
    • 80052823578 scopus 로고    scopus 로고
    • Pre-acquisition System Assessment of the Sysmex(®) Coagulation System CS-2100i and Comparison with End-User Verification: A Model for the Regional Introduction of New Analysers and Methods
    • Molenaar P. J.,Leyte A.Pre-acquisition System Assessment of the Sysmex(®) Coagulation System CS-2100i and Comparison with End-User Verification: A Model for the Regional Introduction of New Analysers and Methods.Clin. Chem. Lab. Med. 2011;49:1479-1489.
    • (2011) Clin. Chem. Lab. Med , vol.49 , pp. 1479-1489
    • Molenaar, P.J.1    Leyte, A.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.