Volumn 38, Issue 5, 2011, Pages 2398-2411

Investigation into the optimal linear time-invariant lag correction for radar artifact removal

Author keywords

a Si; flat panel; lag; radar artifact; x ray detector

Indexed keywords

AMORPHOUS SILICON; COMPUTERIZED TOMOGRAPHY; ERRORS; IMPULSE RESPONSE; PHANTOMS; RADAR; SILICON COMPOUNDS; SILICON DETECTORS; X RAY DETECTORS;

EID: 79959537742     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3574873     Document Type: Article
Times cited : (24)

References (23)
  • 1
    • 0032772594 scopus 로고    scopus 로고
    • A ghost story: Spatio-temporal response characteristics of an indirect- detection flat-panel imager
    • DOI 10.1118/1.598657
    • J. H. Siewerdsen and D. A. Jaffray, A ghost story: Spatio-temporal response characteristics of an indirect-detection flat-panel imager., Med. Phys. 26, 1624-1641 (1999). 10.1118/1.598657 (Pubitemid 29402274)
    • (1999) Medical Physics , vol.26 , Issue.8 , pp. 1624-1641
    • Siewerdsen, J.H.1    Jaffray, D.A.2
  • 2
    • 0033458079 scopus 로고    scopus 로고
    • Cone-beam computed tomography with a flat-panel imager: Effects of image lag
    • DOI 10.1118/1.598803
    • J. H. Siewerdsen and D. A. Jaffray, Cone-beam computed tomography with a flat-panel imager: Effects of image lag., Med. Phys. 26, 2635-2647 (1999). 10.1118/1.598803 (Pubitemid 30007709)
    • (1999) Medical Physics , vol.26 , Issue.12 , pp. 2635-2647
    • Siewerdsen, J.H.1    Jaffray, D.A.2
  • 3
    • 54949143655 scopus 로고    scopus 로고
    • An empirical method for lag correction in cone-beam CT
    • 10.1118/1.2977759
    • N. Mail, D. J. Moseley, J. H. Siewerdsen, and D. A. Jaffray, An empirical method for lag correction in cone-beam CT., Med. Phys. 35 (11), 5187-5196 (2008). 10.1118/1.2977759
    • (2008) Med. Phys. , vol.35 , Issue.11 , pp. 5187-5196
    • Mail, N.1    Moseley, D.J.2    Siewerdsen, J.H.3    Jaffray, D.A.4
  • 5
    • 85024787557 scopus 로고    scopus 로고
    • Th-e-330a-04: Investigation into the cause of a new artifact in cone beam CT reconstructions on a flat panel imager
    • 10.1118/1.2241936
    • J. Starman, G. Virshup, S. Bandy, J. Star-Lack, and R. Fahrig, Th-e-330a-04: Investigation into the cause of a new artifact in cone beam CT reconstructions on a flat panel imager., Med. Phys. 33 (6), 2288 (2006). 10.1118/1.2241936
    • (2006) Med. Phys. , vol.33 , Issue.6 , pp. 2288
    • Starman, J.1    Virshup, G.2    Bandy, S.3    Star-Lack, J.4    Fahrig, R.5
  • 6
    • 1242271376 scopus 로고    scopus 로고
    • Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device
    • DOI 10.1118/1.1637969
    • L. N. McDermott, R. J. W. Louwe, J. J. Sonke, M. B. van Herk, and B. J. Mijnheer, Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device., Med. Phys. 31, 285-295 (2004). 10.1118/1.1637969 (Pubitemid 38235398)
    • (2004) Medical Physics , vol.31 , Issue.2 , pp. 285-295
    • McDermott, L.N.1    Louwe, R.J.W.2    Sonke, J.-J.3    Van Herk, M.B.4    Mijnheer, B.J.5
  • 7
    • 34447301819 scopus 로고    scopus 로고
    • Computer modeling of the spatial resolution properties of a dedicated breast CT system
    • DOI 10.1118/1.2737263
    • K. Yang, A. L. C. Kwan, and J. M. Boone, Computer modeling of the spatial resolution properties of a dedicated breast ct system., Med. Phys. 34 (6), 2059-2069 (2007). 10.1118/1.2737263 (Pubitemid 47055741)
    • (2007) Medical Physics , vol.34 , Issue.6 , pp. 2059-2069
    • Yang, K.1    Kwan, A.L.C.2    Boone, J.M.3
  • 8
    • 0032594376 scopus 로고    scopus 로고
    • High-performance amorphous silicon image sensor for x-ray diagnostic medical imaging applications
    • 10.1117/12.349505
    • R. L. Weisfield, M. A. Hartney, R. Schneider, K. A., and R. Lujan, High-performance amorphous silicon image sensor for x-ray diagnostic medical imaging applications., Proc. SPIE 3659, 307-317 (1999). 10.1117/12.349505
    • (1999) Proc. SPIE , vol.3659 , pp. 307-317
    • Weisfield, R.L.1    Hartney, M.A.2    Schneider K A, R.3    Lujan, R.4
  • 9
    • 4244005128 scopus 로고
    • Effects of trapping in a-Si:H diodes
    • H. Wieczorek, Effects of trapping in a-Si:H diodes., Diffus. Defect Data, Pt. B 44, 957-972 (1995).
    • (1995) Diffus. Defect Data, Pt. B , vol.44 , pp. 957-972
    • Wieczorek, H.1
  • 10
    • 0033748906 scopus 로고    scopus 로고
    • Recursive correction algorithm for detector decay characteristics in CT
    • 10.1117/12.384505
    • J. Hsieh, O. E. Gurmen, and K. F. King, Recursive correction algorithm for detector decay characteristics in CT., Proc. SPIE 3977, 298-305 (2000). 10.1117/12.384505
    • (2000) Proc. SPIE , vol.3977 , pp. 298-305
    • Hsieh, J.1    Gurmen, O.E.2    King, K.F.3
  • 13
    • 43449118093 scopus 로고    scopus 로고
    • Photodiode forward bias to reduce temporal effects in a-Si based flat panel detectors
    • DOI 10.1117/12.769951, Medical Imaging 2008 - Physics of Medical Imaging
    • I. Mollov, C. Tognina, and R. Colbeth, Photodiode forward bias to reduce temporal effects in a-Si-based flat panel detectors., Proc. SPIE 6913, 69133S1-9 (2008). 10.1117/12.769951 (Pubitemid 351662660)
    • (2008) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.6913
    • Mollov, I.1    Tognina, C.2    Colbeth, R.3
  • 14
    • 44349183159 scopus 로고    scopus 로고
    • A moving slanted-edge method to measure the temporal modulation transfer function of fluoroscopic systems
    • DOI 10.1118/1.2919724
    • S. N. Friedman and I. A. Cunningham, A moving slanted-edge method to measure the temporal modulation transfer function of fluoroscopic systems., Med. Phys. 35 (6), 2473-2484 (2008). 10.1118/1.2919724 (Pubitemid 351748655)
    • (2008) Medical Physics , vol.35 , Issue.6 , pp. 2473-2484
    • Friedman, S.N.1    Cunningham, I.A.2
  • 15
    • 68149124534 scopus 로고    scopus 로고
    • A small-signal approach to temporal modulation transfer functions with exposure-rate dependence and its application to fluoroscopic detective quantum efficiency
    • 10.1118/1.3148668
    • S. N. Friedman and I. A. Cunningham, A small-signal approach to temporal modulation transfer functions with exposure-rate dependence and its application to fluoroscopic detective quantum efficiency., Med. Phys. 36 (8), 3775-3785 (2009). 10.1118/1.3148668
    • (2009) Med. Phys. , vol.36 , Issue.8 , pp. 3775-3785
    • Friedman, S.N.1    Cunningham, I.A.2
  • 16
    • 19744369868 scopus 로고    scopus 로고
    • Artifacts in CT: Recognition and avoidance1
    • 10.1148/rg.246045065
    • J. F. Barrett and N. Keat, Artifacts in CT: Recognition and avoidance1., Radiographics 24 (6), 1671-1691 (2004). 10.1148/rg.246045065
    • (2004) Radiographics , vol.24 , Issue.6 , pp. 1671-1691
    • Barrett, J.F.1    Keat, N.2
  • 17
    • 0033983855 scopus 로고    scopus 로고
    • An iterative approach to the beam hardening correction in cone beam CT
    • DOI 10.1118/1.598853
    • J. Hsieh, R. C. Molthen, C. A. Dawson, and R. H. Johnson, An iterative approach to the beam hardening correction in cone beam CT., Med. Phys. 27 (1), 23-29 (2000). 10.1118/1.598853 (Pubitemid 30051907)
    • (2000) Medical Physics , vol.27 , Issue.1 , pp. 23-29
    • Hsieh, J.1    Molthen, R.C.2    Dawson, C.A.3    Johnson, R.H.4
  • 18
    • 0035110308 scopus 로고    scopus 로고
    • Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter
    • DOI 10.1118/1.1339879
    • J. H. Siewerdsen and D. A. Jaffray, Cone-beam computed tomography with a flat-panel imager: Magnitude and effects of x-ray scatter., Med. Phys. 28 (2), 220-231 (2001). 10.1118/1.1339879 (Pubitemid 32165760)
    • (2001) Medical Physics , vol.28 , Issue.2 , pp. 220-231
    • Siewerdsen, J.H.1    Jaffray, D.A.2
  • 19
    • 33845562622 scopus 로고    scopus 로고
    • Scatter correction method for X-ray CT using primary modulation: Theory and preliminary results
    • DOI 10.1109/TMI.2006.884636
    • L. Zhu, N. R. Bennett, and R. Fahrig, Scatter correction method for x-ray CT using primary modulation: Theory and preliminary results., IEEE Trans. Med. Imag. 25 (12), 1573-1587 (2006). 10.1109/TMI.2006.884636 (Pubitemid 44936707)
    • (2006) IEEE Transactions on Medical Imaging , vol.25 , Issue.12 , pp. 1573-1587
    • Zhu, L.1    Bennett, N.R.2    Fahrig, R.3
  • 21
    • 0034257871 scopus 로고    scopus 로고
    • Investigation of a solid-state detector for advanced computed tomography
    • DOI 10.1109/42.887840
    • J. Hsieh, O. E. Gurmen, and K. F. King, Investigation of a solid-state detector for advanced computed tomography., IEEE Trans. Med. Imag. 19 (9), 930-940 (2000). 10.1109/42.887840 (Pubitemid 32084848)
    • (2000) IEEE Transactions on Medical Imaging , vol.19 , Issue.9 , pp. 930-940
    • Hsieh, J.1    Gurmen, O.E.2    King, K.F.3
  • 22
    • 23844485364 scopus 로고    scopus 로고
    • Current status of the CBCT project at Varian Medical Systems
    • DOI 10.1117/12.593418, 41, Medical Imaging 2005 - Physics of Medical Imaging
    • E. Matsinos, Current status of the CBCT project at varian medical systems., Proc. SPIE 5745, 340-351 (2005). 10.1117/12.593418 (Pubitemid 41146383)
    • (2005) Progress in Biomedical Optics and Imaging - Proceedings of SPIE , vol.5745 , Issue.1 , pp. 340-351
    • Matsinos, E.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.