Volumn 37, Issue 12, 2010, Pages 6147-6156

High-rate x-ray spectroscopy in mammography with a CdTe detector: A digital pulse processing approach

Author keywords

CdTe detectors; digital pulse processing; high photon counting rate; mammography; x ray spectroscopy

Indexed keywords

CADMIUM TELLURIDE; II-VI SEMICONDUCTORS; MAMMOGRAPHY; METAL DETECTORS; MOLYBDENUM; PHOTONS; PILES; QUALITY CONTROL; X RAY SPECTROGRAPHS; X RAY SPECTROSCOPY; X RAY TUBES;

EID: 78650106604     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3512804     Document Type: Article
Times cited : (33)

References (42)
  • 1
    • 0028393103 scopus 로고
    • A Monte Carlo program for the calculation of contrast, noise and absorbed dose in diagnostic radiology
    • DOI 10.1016/0169-2607(94)90127-9
    • M. Sandborg, D. R. Dance, G. Alm Carlsson, and J. Persliden, " A Monte Carlo for the simulation of image quality and absorbed dose in diagnostic radiology.," Comput. Methods Programs Biomed. CMPBEX 0169-2607 42, 167-180 (1994). 10.1016/0169-2607(94)90127-9 (Pubitemid 24134794)
    • (1994) Computer Methods and Programs in Biomedicine , vol.42 , Issue.3 , pp. 167-180
    • Sandborg, M.1    Dance, D.R.2    Persliden, J.3    Carlsson, G.A.4
  • 2
    • 0031466296 scopus 로고    scopus 로고
    • Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography
    • DOI 10.1118/1.598100
    • J. M. Boone, T. R. Fewell, and R. J. Jennings, " Molybdenum, rhodium, and tungsten anode spectral models using interpolating polynomials with application to mammography.," Med. Phys. MPHYA6 0094-2405 24, 1863-1874 (1997). 10.1118/1.598100 (Pubitemid 28013501)
    • (1997) Medical Physics , vol.24 , Issue.12 , pp. 1863-1874
    • Boone, J.M.1    Fewell, T.R.2    Jennings, R.J.3
  • 3
    • 21644449154 scopus 로고    scopus 로고
    • A robust method of x-ray source spectrum estimation from transmission measurements: Demonstrated on computer simulated, scatter-free transmission data
    • DOI 10.1063/1.1928312, 124701
    • E. Y. Sidky, L. Yu, X. Pan, Y. Zou, and M. Vannier, " A robust method of x-ray source spectrum estimation from transmission measurements: Demonstrated on computer simulated, scatter-free transmission data.," J. Appl. Phys. JAPIAU 0021-8979 97, 124701 (2005). 10.1063/1.1928312 (Pubitemid 40925370)
    • (2005) Journal of Applied Physics , vol.97 , Issue.12 , pp. 1-11
    • Sidky, E.Y.1    Yu, L.2    Pan, X.3    Zou, Y.4    Vannier, M.5
  • 4
    • 0033695031 scopus 로고    scopus 로고
    • Determination of the voltage applied to x-ray tubes from the bremsstrahlung spectrum obtained with a silicon PIN photodiode
    • MPHYA6 0094-2405. 10.1118/1.1318222
    • M. C. Silva, " Determination of the voltage applied to x-ray tubes from the bremsstrahlung spectrum obtained with a silicon PIN photodiode.," Med. Phys. MPHYA6 0094-2405 27, 2617-2623 (2000). 10.1118/1.1318222
    • (2000) Med. Phys. , vol.27 , pp. 2617-2623
    • Silva, M.C.1
  • 5
    • 13444271654 scopus 로고    scopus 로고
    • Comparison of mammography radiation dose values obtained from direct incident air kerma measurements with values from measured X-ray spectral data
    • DOI 10.1016/j.apradiso.2004.09.010, PII S0969804304005524
    • M. Assiamah, T. L. Nam, and R. J. Keddy, " Comparison of mammography radiation dose values obtained from direct incident air kerma measurements with values from measured x-ray spectral data.," Appl. Radiat. Isot. ARISEF 0969-8043 62, 551-560 (2005). 10.1016/j.apradiso.2004.09.010 (Pubitemid 40215550)
    • (2005) Applied Radiation and Isotopes , vol.62 , Issue.4 , pp. 551-560
    • Assiamah, M.1    Nam, T.L.2    Keddy, R.J.3
  • 6
    • 0142057579 scopus 로고    scopus 로고
    • Computation of beam quality parameters for Mo/Mo, Mo/Rh, Rh/Rh, and W/Al target/filter combinations in mammography
    • DOI 10.1118/1.1608498
    • H. Kharrati and B. Zarrad, " Computation of beam quality parameters for Mo/Mo, Mo/Rh, Rh/Rh, and W/Al target/filter combinations in mammography.," Med. Phys. MPHYA6 0094-2405 30, 2638-2642 (2003). 10.1118/1.1608498 (Pubitemid 37266761)
    • (2003) Medical Physics , vol.30 , Issue.10 , pp. 2638-2642
    • Kharrati, H.1    Zarrad, B.2
  • 7
    • 35648976045 scopus 로고    scopus 로고
    • Dual-energy approach to contrast-enhanced mammography using the balanced filter method: Spectral optimization and preliminary phantom measurement
    • DOI 10.1118/1.2790841
    • M. Saito, " Dual-energy approach to contrast-enhanced mammography using the balanced filter method: Spectral optimization and preliminary phantom measurement.," Med. Phys. MPHYA6 0094-2405 34, 4236-4246 (2007). 10.1118/1.2790841 (Pubitemid 350029645)
    • (2007) Medical Physics , vol.34 , Issue.11 , pp. 4236-4246
    • Saito, M.1
  • 8
    • 0030981982 scopus 로고    scopus 로고
    • Computation of energy imparted in diagnostic radiology
    • DOI 10.1118/1.597939
    • N. A. Gkanatsios and W. Huda, " Computation of energy imparted in diagnostic radiology.," Med. Phys. MPHYA6 0094-2405 24, 571-579 (1997). 10.1118/1.597939 (Pubitemid 27193544)
    • (1997) Medical Physics , vol.24 , Issue.4 , pp. 571-579
    • Gkanatsios, N.A.1    Huda, W.2
  • 9
    • 0025759912 scopus 로고
    • Molybdenum target x-ray spectra: A semiempirical model
    • MPHYA6 0094-2405. 10.1118/1.596686
    • D. M. Tucker, G. T. Barnes, and X. Wu, " Molybdenum target x-ray spectra: A semiempirical model.," Med. Phys. MPHYA6 0094-2405 18, 402-407 (1991). 10.1118/1.596686
    • (1991) Med. Phys. , vol.18 , pp. 402-407
    • Tucker, D.M.1    Barnes, G.T.2    Wu, X.3
  • 10
    • 0004152630 scopus 로고    scopus 로고
    • Catalogue of diagnostic x-ray spectra and other data
    • (CD-Rom Edition) (electronic version prepared by D. Sutton) [York: The Institute of Physics and Engineering in Medicine (IPEM)].
    • K. Cranley, " Catalogue of diagnostic x-ray spectra and other data.," IPEM Report No. 78 (CD-Rom Edition 1997) (electronic version prepared by D. Sutton) [York: The Institute of Physics and Engineering in Medicine (IPEM)].
    • (1997) IPEM Report , Issue.78
    • Cranley, K.1
  • 11
    • 0021133498 scopus 로고
    • Monte Carlo calculations of mammographic x-ray spectra
    • PHMBA7 0031-9155. 10.1088/0031-9155/29/2/013
    • R. Kulkarni and S. Supe, " Monte Carlo calculations of mammographic x-ray spectra.," Phys. Med. Biol. PHMBA7 0031-9155 29, 185-190 (1984). 10.1088/0031-9155/29/2/013
    • (1984) Phys. Med. Biol. , vol.29 , pp. 185-190
    • Kulkarni, R.1    Supe, S.2
  • 12
    • 8544271686 scopus 로고    scopus 로고
    • Monte Carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C
    • DOI 10.1088/0031-9155/49/21/004, PII S0031915504778422
    • R. M. Ay, " Monte Carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C.," Phys. Med. Biol. PHMBA7 0031-9155 49, 4897-4917 (2004). 10.1088/0031-9155/49/21/004 (Pubitemid 39489931)
    • (2004) Physics in Medicine and Biology , vol.49 , Issue.21 , pp. 4897-4917
    • Ay, M.R.1    Shahriari, M.2    Sarkar, S.3    Adib, M.4    Zaidi, H.5
  • 13
    • 33750556634 scopus 로고    scopus 로고
    • Suitability of new anode materials in mammography: Dose and subject contrast considerations using Monte Carlo simulation
    • DOI 10.1118/1.2362874
    • H. Delis, " Suitability of new anode materials in mammography: Dose and subject contrast considerations using Monte Carlo simulation.," Med. Phys. MPHYA6 0094-2405 33, 4221-4235 (2006). 10.1118/1.2362874 (Pubitemid 44673112)
    • (2006) Medical Physics , vol.33 , Issue.11 , pp. 4221-4235
    • Delis, H.1    Spyrou, G.2    Costaridou, L.3    Tzanakos, G.4    Panayiotakis, G.5
  • 15
    • 0017654625 scopus 로고
    • Photon energy distribution of some typical diagnostic x ray beams
    • DOI 10.1118/1.594364
    • T. R. Fewell and R. E. Shuping, " Photon energy distribution of some typical diagnostic x-ray beams.," Med. Phys. MPHYA6 0094-2405 4, 187-196 (1977). 10.1118/1.594364 (Pubitemid 8138312)
    • (1977) Medical Physics , vol.4 , Issue.3 , pp. 187-197
    • Fewell, T.R.1    Shuping, R.E.2
  • 16
    • 0018399326 scopus 로고
    • Computation of bremsstrahlung x-ray spectra and comparison with spectra measured with a Ge(Li) detector
    • PHMBA7 0031-9155. 10.1088/0031-9155/24/3/002
    • R. Birch and M. Marshall, " Computation of bremsstrahlung x-ray spectra and comparison with spectra measured with a Ge(Li) detector.," Phys. Med. Biol. PHMBA7 0031-9155 24, 505-517 (1979). 10.1088/0031-9155/24/3/002
    • (1979) Phys. Med. Biol. , vol.24 , pp. 505-517
    • Birch, R.1    Marshall, M.2
  • 17
    • 0035011178 scopus 로고    scopus 로고
    • A comparison of mammography spectral measurements with spectra produced using several different mathematical models
    • DOI 10.1088/0031-9155/46/5/316, PII S0031915501194480
    • L. E. Wilkinson, P. N. Johnston, and J. C. P. Heggie, " A comparison of mammography spectral measurements with spectra produced using several different mathematical models.," Phys. Med. Biol. PHMBA7 0031-9155 46, 1575-1589 (2001). 10.1088/0031-9155/46/5/316 (Pubitemid 32440205)
    • (2001) Physics in Medicine and Biology , vol.46 , Issue.5 , pp. 1575-1589
    • Wilkinson, L.E.1    Johnston, P.N.2    Heggie, J.C.P.3
  • 18
    • 0035996550 scopus 로고    scopus 로고
    • CdZnTe detector in diagnostic x-ray spectroscopy
    • DOI 10.1118/1.1485975
    • S. Miyajima, K. Imagawa, and M. Matsumoto, " CdZnTe detector in diagnostic x-ray spectroscopy.," Med. Phys. MPHYA6 0094-2405 29, 1421-1429 (2002). 10.1118/1.1485975 (Pubitemid 34774573)
    • (2002) Medical Physics , vol.29 , Issue.7 , pp. 1421-1429
    • Miyajima, S.1    Imagawa, K.2    Matsumoto, M.3
  • 19
    • 0038248801 scopus 로고    scopus 로고
    • Thin CdTe detector in diagnostic x-ray spectroscopy
    • DOI 10.1118/1.1566388
    • S. Miyajima, " Thin CdTe detector in diagnostic x-ray spectroscopy.," Med. Phys. MPHYA6 0094-2405 30, 771-777 (2003). 10.1118/1.1566388 (Pubitemid 36578678)
    • (2003) Medical Physics , vol.30 , Issue.5 , pp. 771-777
    • Miyajima, S.1
  • 20
    • 7244254289 scopus 로고    scopus 로고
    • Direct analysis of molybdenum target generated x-ray spectra with a portable device
    • DOI 10.1118/1.1785891
    • S. Stumbo, " Direct analysis of molybdenum target generated x-ray spectra with a portable device.," Med. Phys. MPHYA6 0094-2405 31, 2763-2770 (2004). 10.1118/1.1785891 (Pubitemid 39431001)
    • (2004) Medical Physics , vol.31 , Issue.10 , pp. 2763-2770
    • Stumbo, S.1    Bottigli, U.2    Golosio, B.3    Oliva, P.4    Tangaro, S.5
  • 24
    • 67650261448 scopus 로고    scopus 로고
    • Progress in the development of CdTe and CdZnTe semiconductor radiation detectors for astrophysical and medical applications
    • SNSRES 0746-9462. 10.3390/s90503491
    • S. Del Sordo, L. Abbene, E. Caroli, A. M. Mancini, A. Zappettini, and P. Ubertini, " Progress in the development of CdTe and CdZnTe semiconductor radiation detectors for astrophysical and medical applications.," Sensors SNSRES 0746-9462 9, 3491-3526 (2009). 10.3390/s90503491
    • (2009) Sensors , vol.9 , pp. 3491-3526
    • Del Sordo, S.1    Abbene, L.2    Caroli, E.3    Mancini, A.M.4    Zappettini, A.5    Ubertini, P.6
  • 26
    • 72349094091 scopus 로고    scopus 로고
    • Pileup correction algorithms for very-high-count-rate gamma-ray spectrometry with NaI(Tl) detectors
    • IEIMAO 0018-9456. 10.1109/TIM.2009.2022107
    • M. Bolic, V. Drndarevic, and W. Gueaieb, " Pileup correction algorithms for very-high-count-rate gamma-ray spectrometry with NaI(Tl) detectors.," IEEE Trans. Instrum. Meas. IEIMAO 0018-9456 59, 122-130 (2010). 10.1109/TIM.2009.2022107
    • (2010) IEEE Trans. Instrum. Meas. , vol.59 , pp. 122-130
    • Bolic, M.1    Drndarevic, V.2    Gueaieb, W.3
  • 27
    • 77957895627 scopus 로고    scopus 로고
    • Performance of a digital CdTe x-ray spectrometer in low and high counting rate environment
    • NIMAER 0168-9002. 10.1016/j.nima.2010.06.002
    • L. Abbene, G. Gerardi, S. Del Sordo, and G. Raso, " Performance of a digital CdTe x-ray spectrometer in low and high counting rate environment.," Nucl. Instrum. Methods Phys. Res. A NIMAER 0168-9002 621, 447-452 (2010). 10.1016/j.nima.2010.06.002
    • (2010) Nucl. Instrum. Methods Phys. Res. A , vol.621 , pp. 447-452
    • Abbene, L.1    Gerardi, G.2    Del Sordo, S.3    Raso, G.4
  • 29
    • 85027228207 scopus 로고    scopus 로고
    • The LAX laboratory homepage.
    • The LAX laboratory homepage (http://www.ifc.inaf.it).
  • 30
    • 0017194431 scopus 로고
    • Spectroscopy of diagnostic x rays by a Compton-scatter method
    • MPHYA6 0094-2405. 10.1118/1.594263
    • M. Yaffe, K. W. Taylor, and H. E. Johns, " Spectroscopy of diagnostic x rays by a Compton-scatter method.," Med. Phys. MPHYA6 0094-2405 3, 328-334 (1976). 10.1118/1.594263
    • (1976) Med. Phys. , vol.3 , pp. 328-334
    • Yaffe, M.1    Taylor, K.W.2    Johns, H.E.3
  • 31
    • 0020540081 scopus 로고
    • Effect of collimators on the measurement of diagnostic x-ray spectra
    • PHMBA7 0031-9155. 10.1088/0031-9155/28/7/006
    • Y. Kodera, H. P. Chan, and K. Doi, " Effect of collimators on the measurement of diagnostic x-ray spectra.," Phys. Med. Biol. PHMBA7 0031-9155 28, 841-852 (1983). 10.1088/0031-9155/28/7/006
    • (1983) Phys. Med. Biol. , vol.28 , pp. 841-852
    • Kodera, Y.1    Chan, H.P.2    Doi, K.3
  • 32
    • 0030442807 scopus 로고    scopus 로고
    • Comparison of x-ray cross sections for diagnostic and therapeutic medical physics
    • DOI 10.1118/1.597899
    • J. M. Boone and A. E. Chavez, " Comparison of x-ray cross sections for diagnostic and therapeutic medical physics.," Med. Phys. MPHYA6 0094-2405 23, 1997-2005 (1996). 10.1118/1.597899 (Pubitemid 27023697)
    • (1996) Medical Physics , vol.23 , Issue.12 , pp. 1997-2005
    • Boone, J.M.1    Chavez, A.E.2
  • 35
    • 51849086474 scopus 로고    scopus 로고
    • Dosimetry in diagnostic radiology: An international code of practice
    • " Dosimetry in diagnostic radiology: An international code of practice.," IAEA Technical Reports Series No. 457.
    • IAEA Technical Reports Series No. 457
  • 36
    • 0025204669 scopus 로고
    • On the measurement of half value layer in film screen mammography
    • MPHYA6 0094-2405. 10.1118/1.596444
    • L. K. Wagner, B. R. Archer, and F. Cerra, " On the measurement of half value layer in film screen mammography.," Med. Phys. MPHYA6 0094-2405 17, 989-997 (1990). 10.1118/1.596444
    • (1990) Med. Phys. , vol.17 , pp. 989-997
    • Wagner, L.K.1    Archer, B.R.2    Cerra, F.3
  • 37
    • 9244230589 scopus 로고    scopus 로고
    • X-ray spectroscopy in mammography with a silicon PIN photodiode with application to the measurement of tube voltage
    • DOI 10.1118/1.1803432
    • R. Künzel, S. B. Herdade, R. A. Terini, and P. R. Costa, " X-ray spectroscopy in mammography with a silicon PIN photodiode with application to the measurement of tube voltage.," Med. Phys. MPHYA6 0094-2405 31, 2996-3003 (2004). 10.1118/1.1803432 (Pubitemid 39552689)
    • (2004) Medical Physics , vol.31 , Issue.11 , pp. 2996-3003
    • Kunzel, R.1    Herdade, S.B.2    Terini, R.A.3    Costa, P.R.4
  • 38
    • 77953330647 scopus 로고    scopus 로고
    • Detectors for the future of x-ray imaging
    • RPDODE 0144-8420. 10.1093/rpd/ncq074
    • M. Aslund, E. Fredenberg, M. Telman, and M. Danielsson, " Detectors for the future of x-ray imaging.," Radiat. Prot. Dosim. RPDODE 0144-8420 139, 327-333 (2010). 10.1093/rpd/ncq074
    • (2010) Radiat. Prot. Dosim. , vol.139 , pp. 327-333
    • Aslund, M.1    Fredenberg, E.2    Telman, M.3    Danielsson, M.4
  • 42
    • 77952404690 scopus 로고    scopus 로고
    • Contrast-enhanced spectral mammography with a photon-counting detector
    • MPHYA6 0094-2405. 10.1118/1.3371689
    • E. Fredenberg, M. Hemmendorff, B. Cederstrom, M. Aslund, and M. Danielsson, " Contrast-enhanced spectral mammography with a photon-counting detector.," Med. Phys. MPHYA6 0094-2405 37, 2017-2029 (2010). 10.1118/1.3371689
    • (2010) Med. Phys. , vol.37 , pp. 2017-2029
    • Fredenberg, E.1    Hemmendorff, M.2    Cederstrom, B.3    Aslund, M.4    Danielsson, M.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.