High-throughput small animal PET imaging in cancer research: Evaluation of the capability of the Inveon scanner to image four mice simultaneously

Author keywords

cancer research; nude mice; small animal PET; tumour xenograft

Indexed keywords

FLUORODEOXYGLUCOSE F 18;

EID: 77957326584     PISSN: 01433636     EISSN: None     Source Type: Journal    
DOI: 10.1097/MNM.0b013e32833dc61d     Document Type: Article
Times cited : (16)

References (27)
  • 2
    • 60549115241 scopus 로고    scopus 로고
    • Early evaluation of the effects of chemotherapy with longitudinal FDG small-animal PET in human testicular cancer xenografts: Early flare response does not reflect refractory disease
    • Aide N, Poulain L, Briand M, Dutoit S, Allouche S, Labiche A, et al. Early evaluation of the effects of chemotherapy with longitudinal FDG small-animal PET in human testicular cancer xenografts: early flare response does not reflect refractory disease. Eur J Nucl Med Mol Imaging 2009; 36:396-405.
    • (2009) Eur J Nucl Med Mol Imaging , vol.36 , pp. 396-405
    • Aide, N.1    Poulain, L.2    Briand, M.3    Dutoit, S.4    Allouche, S.5    Labiche, A.6
  • 3
    • 27544494166 scopus 로고    scopus 로고
    • An in vivo tumor model exploiting metabolic response as a biomarker for targeted drug development
    • Cullinane C, Dorow DS, Kansara M, Conus N, Binns D, Hicks RJ, et al. An in vivo tumor model exploiting metabolic response as a biomarker for targeted drug development. Cancer Res 2005; 65:9633-9636.
    • (2005) Cancer Res , vol.65 , pp. 9633-9636
    • Cullinane, C.1    Dorow, D.S.2    Kansara, M.3    Conus, N.4    Binns, D.5    Hicks, R.J.6
  • 4
    • 33645853784 scopus 로고    scopus 로고
    • Multi-tracer small animal PET imaging of the tumour response to the novel pan-erb-B inhibitor CI-1033
    • Dorow DS, Cullinane C, Conus N, Roselt P, Binns D, McCarthy TJ, et al. Multi-tracer small animal PET imaging of the tumour response to the novel pan-Erb-B inhibitor CI-1033. Eur J Nucl Med Mol Imaging 2006; 33:441-452.
    • (2006) Eur J Nucl Med Mol Imaging , vol.33 , pp. 441-452
    • Dorow, D.S.1    Cullinane, C.2    Conus, N.3    Roselt, P.4    Binns, D.5    McCarthy, T.J.6
  • 5
    • 35348936707 scopus 로고    scopus 로고
    • Role of small animal PET for molecular imaging in pre-clinical studies
    • Nanni C, Rubello D, Fanti S. Role of small animal PET for molecular imaging in pre-clinical studies. Eur J Nucl Med Mol Imaging 2007; 34:1819-1822.
    • (2007) Eur J Nucl Med Mol Imaging , vol.34 , pp. 1819-1822
    • Nanni, C.1    Rubello, D.2    Fanti, S.3
  • 6
    • 34247609781 scopus 로고    scopus 로고
    • Role of small animal PET in stimulating the development of new radiopharmaceuticals in oncology
    • Nanni C, Rubello D, Khan S, Al-Nahhas A, Fanti S. Role of small animal PET in stimulating the development of new radiopharmaceuticals in oncology. Nucl Med Commun 2007; 28:427-429.
    • (2007) Nucl Med Commun , vol.28 , pp. 427-429
    • Nanni, C.1    Rubello, D.2    Khan, S.3    Al-Nahhas, A.4    Fanti, S.5
  • 7
    • 62449092307 scopus 로고    scopus 로고
    • Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards
    • Bao Q, Newport D, Chen M, Stout DB, Chatziioannou AF. Performance evaluation of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards. J Nucl Med 2009; 50:401-408.
    • (2009) J Nucl Med , vol.50 , pp. 401-408
    • Bao, Q.1    Newport, D.2    Chen, M.3    Stout, D.B.4    Chatziioannou, A.F.5
  • 8
    • 67649883194 scopus 로고    scopus 로고
    • Performance evaluation of an inveon PET preclinical scanner
    • Constantinescu CC, Mukherjee J. Performance evaluation of an Inveon PET preclinical scanner. Phys Med Biol 2009; 54:2885-2899.
    • (2009) Phys Med Biol , vol.54 , pp. 2885-2899
    • Constantinescu, C.C.1    Mukherjee, J.2
  • 10
    • 33644874714 scopus 로고    scopus 로고
    • Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the national electrical manufacturers association NU 2-2001 standard
    • Brambilla M, Secco C, Dominietto M, Matheoud R, Sachetti G, Inglese E. Performance characteristics obtained for a new 3-dimensional lutetium oxyorthosilicate-based whole-body PET/CT scanner with the National Electrical Manufacturers Association NU 2-2001 standard. J Nucl Med 2005; 46:2083-2091.
    • (2005) J Nucl Med , vol.46 , pp. 2083-2091
    • Brambilla, M.1    Secco, C.2    Dominietto, M.3    Matheoud, R.4    Sachetti, G.5    Inglese, E.6
  • 12
    • 77950951879 scopus 로고    scopus 로고
    • Image-quality assessment for several positron emitters using the NEMA NU 4-2008 standards in the siemens inveon small-animal PET scanner
    • Disselhorst JA, Brom M, Laverman P, Slump CH, Boerman OC, Oyen WJ, et al. Image-quality assessment for several positron emitters using the NEMA NU 4-2008 standards in the Siemens Inveon small-animal PET scanner. J Nucl Med 2001; 51:610-617.
    • (2001) J Nucl Med , vol.51 , pp. 610-617
    • Disselhorst, J.A.1    Brom, M.2    Laverman, P.3    Slump, C.H.4    Boerman, O.C.5    Oyen, W.J.6
  • 13
    • 34249781398 scopus 로고    scopus 로고
    • Retro-orbital injection is an effective route for radiopharmaceutical administration in mice during small-animal PET studies
    • Nanni C, Pettinato C, Ambrosini V, Spinelli A, Trespidi S, Rubello D, et al. Retro-orbital injection is an effective route for radiopharmaceutical administration in mice during small-animal PET studies. Nucl Med Commun 2007; 28:547-553.
    • (2007) Nucl Med Commun , vol.28 , pp. 547-553
    • Nanni, C.1    Pettinato, C.2    Ambrosini, V.3    Spinelli, A.4    Trespidi, S.5    Rubello, D.6
  • 14
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 8:307-310.
    • (1986) Lancet , vol.8 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 16
    • 54049102352 scopus 로고    scopus 로고
    • A rat head holder for simultaneous scanning of two rats in small animal PET scanners: Design, construction, feasibility testing and kinetic validation
    • Cheng TE,Yoder KK, Normandin MD, Risacher SL, Converse AK, Hampel JA, et al. A rat head holder for simultaneous scanning of two rats in small animal PET scanners: design, construction, feasibility testing and kinetic validation. J Neurosci Methods 2009; 176:24-33.
    • (2009) J Neurosci Methods , vol.176 , pp. 24-33
    • Cheng, T.E.1    Yoder, K.K.2    Normandin, M.D.3    Risacher, S.L.4    Converse, A.K.5    Hampel, J.A.6
  • 17
    • 77952093684 scopus 로고    scopus 로고
    • High throughput static and dynamic small animal imaging using clinical PET/CT: Potential preclinical applications
    • Aide N, Desmonts C, Beauregard JM, Beyer T, Kinross K, Roselt P, et al. High throughput static and dynamic small animal imaging using clinical PET/CT: potential preclinical applications. Eur J Nucl Med Mol Imaging 2010;37:991-1001.
    • (2010) Eur J Nucl Med Mol Imaging , vol.37 , pp. 991-1001
    • Aide, N.1    Desmonts, C.2    Beauregard, J.M.3    Beyer, T.4    Kinross, K.5    Roselt, P.6
  • 18
    • 77950361631 scopus 로고    scopus 로고
    • High-contrast PET of melanoma using (18)F-MEL050, a selective probe for melanin with predominantly renal clearance
    • Denoyer D, Greguric I, Roselt P, Neels OC, Aide N, Taylor SR, et al. High-contrast PET of melanoma using (18)F-MEL050, a selective probe for melanin with predominantly renal clearance. J Nucl Med 2010; 51:441-447.
    • (2010) J Nucl Med , vol.51 , pp. 441-447
    • Denoyer, D.1    Greguric, I.2    Roselt, P.3    Neels, O.C.4    Aide, N.5    Taylor, S.R.6
  • 20
  • 21
    • 34247229328 scopus 로고    scopus 로고
    • Performance evaluation of the GE healthcare eXplore VISTA dual-ring small-animal PET scanner
    • Wang Y, Seidel J, Tsui BM, Vaquero JJ, Pomper MG. Performance evaluation of the GE Healthcare eXplore VISTA dual-ring small-animal PET scanner. J Nucl Med 2006; 47:1891-1900.
    • (2006) J Nucl Med , vol.47 , pp. 1891-1900
    • Wang, Y.1    Seidel, J.2    Tsui, B.M.3    Vaquero, J.J.4    Pomper, M.G.5
  • 22
    • 33746191812 scopus 로고    scopus 로고
    • Fully 3-D PET reconstruction with system matrix derived from point source measurements
    • Panin VY, Kehren F, Michel C, Casey M. Fully 3-D PET reconstruction with system matrix derived from point source measurements. IEEE Trans Med Imaging 2006; 25:907-921.
    • (2006) IEEE Trans Med Imaging , vol.25 , pp. 907-921
    • Panin, V.Y.1    Kehren, F.2    Michel, C.3    Casey, M.4
  • 23
    • 0031981253 scopus 로고    scopus 로고
    • High resolution 3D Bayesian image reconstruction using the microPET small animal scanner
    • Qi J, Leahy RM, Cherry SR, Chatziioannou AF, Farquhar T. High resolution 3D Bayesian image reconstruction using the microPET small animal scanner. Phys Med Biol 1997; 43:1001-1013.
    • (1997) Phys Med Biol , vol.43 , pp. 1001-1013
    • Qi, J.1    Leahy, R.M.2    Cherry, S.R.3    Chatziioannou, A.F.4    Farquhar, T.5
  • 24
    • 6944235697 scopus 로고    scopus 로고
    • Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: A simulation study
    • Boellaard R, Krak NC, Hoekstra OS, Lammertsma AA. Effects of noise, image resolution, and ROI definition on the accuracy of standard uptake values: a simulation study. J Nucl Med 2004; 45:1519-1527.
    • (2004) J Nucl Med , vol.45 , pp. 1519-1527
    • Boellaard, R.1    Krak, N.C.2    Hoekstra, O.S.3    Lammertsma, A.A.4
  • 25
    • 17644421807 scopus 로고    scopus 로고
    • Effects of ROI definition and reconstruction method on quantitative outcome and applicability in a response monitoring trial
    • Krak NC, Boellaard R, Hoekstra OS, Twisk JW, Hoekstra CJ, Lammertsma AA. Effects of ROI definition and reconstruction method on quantitative outcome and applicability in a response monitoring trial. Eur J Nucl Med Mol Imaging 2005; 32:294-301.
    • (2005) Eur J Nucl Med Mol Imaging , vol.32 , pp. 294-301
    • Krak, N.C.1    Boellaard, R.2    Hoekstra, O.S.3    Twisk, J.W.4    Hoekstra, C.J.5    Lammertsma, A.A.6
  • 26
    • 33847159680 scopus 로고    scopus 로고
    • Quantification of FDG PET studies using standardised uptake values in multi-centre trials: Effects of image reconstruction, resolution and ROI definition parameters
    • Westerterp M, Pruim J, Oyen W, Hoekstra O, Paans A, Visser E, et al. Quantification of FDG PET studies using standardised uptake values in multi-centre trials: effects of image reconstruction, resolution and ROI definition parameters. Eur J Nucl Med Mol Imaging 2007; 34:392-404.
    • (2007) Eur J Nucl Med Mol Imaging , vol.34 , pp. 392-404
    • Westerterp, M.1    Pruim, J.2    Oyen, W.3    Hoekstra, O.4    Paans, A.5    Visser, E.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.