Perfrmance evalution of the inveon dedicated PET preclinical tomograph based on the NEMA NU-4 standards

Author keywords

MicroPET; Molecular imaging; Performance evaluation; instrumentation; Small animal PET scanner

Indexed keywords

FLUORINE 18; FLUORODEOXYGLUCOSE F 18; LUTETIUM; LUTETIUM OXYORTHOSILICATE; SILICATE; UNCLASSIFIED DRUG; DIAGNOSTIC AGENT; RADIOPHARMACEUTICAL AGENT;

EID: 62449092307     PISSN: 01615505     EISSN: None     Source Type: Journal    
DOI: 10.2967/jnumed.108.056374     Document Type: Article
Times cited : (284)

References (27)
  • 1
    • 0034241796 scopus 로고    scopus 로고
    • Inaugural article: Positron emission tomography provides molecular imaging of biological processes
    • Phelps ME. Inaugural article: positron emission tomography provides molecular imaging of biological processes. Pmc Natl Acad Sci USA. 2000:97:9226-9233.
    • (2000) Pmc Natl Acad Sci USA , vol.97 , pp. 9226-9233
    • Phelps, M.E.1
  • 2
    • 0035222861 scopus 로고    scopus 로고
    • Use of positron emission tomography in animal research
    • Cherry SR, Gambhir SS. Use of positron emission tomography in animal research. ILAR J. 2001;42:219-232.
    • (2001) ILAR J , vol.42 , pp. 219-232
    • Cherry, S.R.1    Gambhir, S.S.2
  • 3
    • 62449218842 scopus 로고    scopus 로고
    • National Electrical Manufacturers Association NEMA, NEMA Standards Publication NU 4-2008. Rosslyn, VA: NEMA;
    • National Electrical Manufacturers Association (NEMA). Performance Measurements for Small Animal Positron Emission Tomographs (PETs). NEMA Standards Publication NU 4-2008. Rosslyn, VA: NEMA; 2008.
    • (2008) Performance Measurements for Small Animal Positron Emission Tomographs (PETs)
  • 4
    • 0023640412 scopus 로고
    • Treatment of axial data in three- dimensional PET
    • Daube-Witherspoon ME, Muehllehner G. Treatment of axial data in three- dimensional PET. J Nucl Med. 1987;28:1717-1724.
    • (1987) J Nucl Med , vol.28 , pp. 1717-1724
    • Daube-Witherspoon, M.E.1    Muehllehner, G.2
  • 6
    • 0024610693 scopus 로고
    • Analytic 3D image reconstruction using all delected events
    • Kinahan PE, Rogers JG. Analytic 3D image reconstruction using all delected events. IEEE Trans Nucl Sci. 1989;36:964-968.
    • (1989) IEEE Trans Nucl Sci , vol.36 , pp. 964-968
    • Kinahan, P.E.1    Rogers, J.G.2
  • 7
    • 0028700063 scopus 로고
    • Accelerated image reconstruction using ordered subsets of projection data
    • Hudson HM, Larkin RS. Accelerated image reconstruction using ordered subsets of projection data. IEEE Trans Med Imag. 1994:13:601-609.
    • (1994) IEEE Trans Med Imag , vol.13 , pp. 601-609
    • Hudson, H.M.1    Larkin, R.S.2
  • 9
    • 0031981253 scopus 로고    scopus 로고
    • Qi J, Leahy RM. Cherry SR, Chatziioannou A, FarquharTH. High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner. Phys Med Biol. 1998;43:1001-1013.
    • Qi J, Leahy RM. Cherry SR, Chatziioannou A, FarquharTH. High-resolution 3D Bayesian image reconstruction using the microPET small-animal scanner. Phys Med Biol. 1998;43:1001-1013.
  • 10
    • 0024143307 scopus 로고
    • An evaluation of a two-dimensional array detector for high resolution PET
    • Dahlbom M, Hoffman EJ. An evaluation of a two-dimensional array detector for high resolution PET. IEEE Trans Med Imag. 1988;7:264-272.
    • (1988) IEEE Trans Med Imag , vol.7 , pp. 264-272
    • Dahlbom, M.1    Hoffman, E.J.2
  • 13
    • 20244376008 scopus 로고    scopus 로고
    • The ECAT HRRT: An example of NEMA scatter estimation issues for LSO-based PET systems
    • Eriksson L, Watson CC, Wienhard K, et al. The ECAT HRRT: an example of NEMA scatter estimation issues for LSO-based PET systems. IEEE Trans Nucl Sci. 2005:52:90-94.
    • (2005) IEEE Trans Nucl Sci , vol.52 , pp. 90-94
    • Eriksson, L.1    Watson, C.C.2    Wienhard, K.3
  • 15
    • 19744363131 scopus 로고    scopus 로고
    • Investigation of single, random, and true counts from natural radioactivity in LSO-based clinical PET
    • Yamamoto S, Horii H, Hurutani M, Matsumoto K, Senda M. Investigation of single, random, and true counts from natural radioactivity in LSO-based clinical PET. Ann Nucl Med. 2005;19:109-114.
    • (2005) Ann Nucl Med , vol.19 , pp. 109-114
    • Yamamoto, S.1    Horii, H.2    Hurutani, M.3    Matsumoto, K.4    Senda, M.5
  • 16
    • 0032085732 scopus 로고    scopus 로고
    • Nuclear data sheets for A = 176
    • Browne E, Junde H. Nuclear data sheets for A = 176. Nucl Data Sheets. 1998;84:337-486.
    • (1998) Nucl Data Sheets , vol.84 , pp. 337-486
    • Browne, E.1    Junde, H.2
  • 17
    • 35348826758 scopus 로고    scopus 로고
    • Imaging of weak-source distributions in LSO-based small-animal PET scanners
    • Goerlzen AL. Suk JY, Thompson CJ. Imaging of weak-source distributions in LSO-based small-animal PET scanners. J Nucl Med. 2007;48:1692-1698.
    • (2007) J Nucl Med , vol.48 , pp. 1692-1698
    • Goerlzen, A.L.1    Suk, J.Y.2    Thompson, C.J.3
  • 18
    • 0025414253 scopus 로고
    • Measuring PET scanner sensitivity: Relating countrales to image signal-to-noise ratios using noise equivalents counts
    • Strother SC. Casey ME, Hoffman EJ. Measuring PET scanner sensitivity: relating countrales to image signal-to-noise ratios using noise equivalents counts. IEEE Trans Nucl Sci. 1990;37:783-788.
    • (1990) IEEE Trans Nucl Sci , vol.37 , pp. 783-788
    • Strother, S.C.1    Casey, M.E.2    Hoffman, E.J.3
  • 19
    • 11844291336 scopus 로고    scopus 로고
    • Count rate dependence of local signal to noise ratio in positron emission tomography
    • Watson CC. Count rate dependence of local signal to noise ratio in positron emission tomography. IEEE Nucl Sci Symp Conf Rec. 2003;4:2817-2821.
    • (2003) IEEE Nucl Sci Symp Conf Rec , vol.4 , pp. 2817-2821
    • Watson, C.C.1
  • 20
    • 22844452107 scopus 로고    scopus 로고
    • Attenuation correction for small animal PET tomographs
    • Chow PL, Rannou PR, Chatziioannou AF. Attenuation correction for small animal PET tomographs. Phys Med Biol. 2005:50:1837-1850.
    • (2005) Phys Med Biol , vol.50 , pp. 1837-1850
    • Chow, P.L.1    Rannou, P.R.2    Chatziioannou, A.F.3
  • 21
    • 0001320737 scopus 로고    scopus 로고
    • A single scatter simulation technique for scatter correction in 3D PET
    • Grangeat P, Amans JL, eds, Dordrecht, The Netherlands: Kluwer Academic Publishers;
    • Watson CC, Newport D. Casey ME. A single scatter simulation technique for scatter correction in 3D PET. In: Grangeat P, Amans JL, eds. Fully Three- Dimensional Image Reconstruction in Radiology and Nuclear Medicine. Dordrecht, The Netherlands: Kluwer Academic Publishers; 1996:255- 268.
    • (1996) Fully Three- Dimensional Image Reconstruction in Radiology and Nuclear Medicine , pp. 255-268
    • Watson, C.C.1    Newport, D.2    Casey, M.E.3
  • 22
    • 0034244929 scopus 로고    scopus 로고
    • New, faster, image-based scatter correction for 3D PET
    • Watson CC. New, faster, image-based scatter correction for 3D PET. IEEE Trans Nucl Sci. 2000:47:1587-1594.
    • (2000) IEEE Trans Nucl Sci , vol.47 , pp. 1587-1594
    • Watson, C.C.1
  • 23
    • 17844363989 scopus 로고    scopus 로고
    • Performance evaluation of the microPET focus: A third-generation microPET scanner dedicated to animal imaging
    • Tai YC, Ruangma A, Rowland D, et al. Performance evaluation of the microPET focus: a third-generation microPET scanner dedicated to animal imaging. J Nucl Med. 2005;46:455-463.
    • (2005) J Nucl Med , vol.46 , pp. 455-463
    • Tai, Y.C.1    Ruangma, A.2    Rowland, D.3
  • 24
    • 34548580353 scopus 로고    scopus 로고
    • Performance measurement of the microPET focus 120 scanner
    • Kim JS, Lee JS, lm KC, et al. Performance measurement of the microPET focus 120 scanner. J Nucl Med. 2007:48:1527-1535.
    • (2007) J Nucl Med , vol.48 , pp. 1527-1535
    • Kim, J.S.1    Lee, J.S.2    lm, K.C.3
  • 25
    • 0034913087 scopus 로고    scopus 로고
    • Performance evaluation of the microPETP4: A PET system dedicated to animal imaging
    • Tai C. Chatziioannou A, Siegel S, et al. Performance evaluation of the microPETP4: a PET system dedicated to animal imaging. Phys Med Biol. 2001;46:1845-1862.
    • (2001) Phys Med Biol , vol.46 , pp. 1845-1862
    • Tai, C.1    Chatziioannou, A.2    Siegel, S.3
  • 26
    • 0034178295 scopus 로고    scopus 로고
    • Comparison of 3-D maximum a posteriori and filtered back projection algorithms for high-resolution animal imaging with microPET
    • Chatziioannou A, Qi J, Moore A, et al. Comparison of 3-D maximum a posteriori and filtered back projection algorithms for high-resolution animal imaging with microPET. IEEE Trans Med Imaging. 2000;19:507-512.
    • (2000) IEEE Trans Med Imaging , vol.19 , pp. 507-512
    • Chatziioannou, A.1    Qi, J.2    Moore, A.3
  • 27
    • 0030362530 scopus 로고    scopus 로고
    • Accurate geometric and physical response modelling for statistical image reconstruction in high resolution PET
    • Mumcuoglu EU, Leahy RM, Cherry SR, Hoffman E. Accurate geometric and physical response modelling for statistical image reconstruction in high resolution PET. IEEE Nucl Sci Symp Conf Rec. 1996:3:1569-1573.
    • (1996) IEEE Nucl Sci Symp Conf Rec , vol.3 , pp. 1569-1573
    • Mumcuoglu, E.U.1    Leahy, R.M.2    Cherry, S.R.3    Hoffman, E.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.