메뉴 건너뛰기




Volumn 93, Issue 1, 2010, Pages 266-274

Characterization of parylene C as an encapsulation material for implanted neural prostheses

Author keywords

Adhesion; Crystallinity; Encapsulation; Neural prosthesis; Parylene C

Indexed keywords

ADHESION PROMOTION; CRYSTALLINITIES; CRYSTALLINITY; DEGREE OF CRYSTALLINITY; DIFFERENT SUBSTRATES; METALLIC STRUCTURES; NEURAL PROSTHESIS; PARYLENE C; PARYLENES; PROCESS PARAMETERS; STEAM STERILIZATION; TENSILE TESTS; THERMAL TREATMENT;

EID: 77949497612     PISSN: 15524973     EISSN: 15524981     Source Type: Journal    
DOI: 10.1002/jbm.b.31584     Document Type: Article
Times cited : (223)

References (42)
  • 2
    • 62349092532 scopus 로고    scopus 로고
    • Implant design and development
    • Cooper H, Craddock L, editors, 2nd ed. London: Wiley;
    • Grayden DB, Clark GM. Implant design and development. In: Cooper H, Craddock L, editors. Cochlear Implants: A Practical Guide, 2nd ed. London: Wiley; 2005. pp 1-20.
    • (2005) Cochlear Implants: A Practical Guide , pp. 1-20
    • Grayden, D.B.1    Clark, G.M.2
  • 4
    • 23844538849 scopus 로고    scopus 로고
    • An array of microactuated microelectrodes for monitoring single-neuronal activity in rodents
    • Muthuswamy J, Okandan M, Gilletti A, Baker MS, Jain T. An array of microactuated microelectrodes for monitoring single-neuronal activity in rodents. IEEE Trans Biomed Eng 2005;52: 1470-1477.
    • (2005) IEEE Trans Biomed Eng , vol.52 , pp. 1470-1477
    • Muthuswamy, J.1    Okandan, M.2    Gilletti, A.3    Baker, M.S.4    Jain, T.5
  • 6
    • 0020408457 scopus 로고
    • Sacral anterior root stimulators for bladder control in paraplegia
    • Brindley GS, Polkey CE, Rushton DN. Sacral anterior root stimulators for bladder control in paraplegia. Paraplegia 1982;20:365-381.
    • (1982) Paraplegia , vol.20 , pp. 365-381
    • Brindley, G.S.1    Polkey, C.E.2    Rushton, D.N.3
  • 7
    • 0024574737 scopus 로고
    • Encapsulating microelectronic implants in onepart silicone rubbers
    • Donaldson PEK. Encapsulating microelectronic implants in onepart silicone rubbers. Med Biol Eng Comput 1989;27:93-94.
    • (1989) Med Biol Eng Comput , vol.27 , pp. 93-94
    • Donaldson, P.E.K.1
  • 8
    • 0032125652 scopus 로고    scopus 로고
    • Chronic recording capability of the Utah intracortical electrode array in cat sensory cortex
    • Rousche PJ, Normann RA. Chronic recording capability of the Utah intracortical electrode array in cat sensory cortex. J Neurosci Methods 1998;82:1-15.
    • (1998) J Neurosci Methods , vol.82 , pp. 1-15
    • Rousche, P.J.1    Normann, R.A.2
  • 9
    • 0026205082 scopus 로고
    • A silicon-based, three-dimensional neural interface: Manufacturing processes for an intracortical electrode array
    • Campbell PK, Jones KE, Huber RJ, Horch KW, Normann RA. A silicon-based, three-dimensional neural interface: Manufacturing processes for an intracortical electrode array. IEEE Trans Biomed Eng 1991;38:758-768.
    • (1991) IEEE Trans Biomed Eng , vol.38 , pp. 758-768
    • Campbell, P.K.1    Jones, K.E.2    Huber, R.J.3    Horch, K.W.4    Normann, R.A.5
  • 12
    • 17444364109 scopus 로고    scopus 로고
    • Encapsulation of organic field effect transistors for flexible biomedical microimplants
    • Feili D, Schuettler M, Droege T, Kammer S, Stieglitz T. Encapsulation of organic field effect transistors for flexible biomedical microimplants. Sens Actuators A Phys 2005;120:101-109.
    • (2005) Sens Actuators A Phys , vol.120 , pp. 101-109
    • Feili, D.1    Schuettler, M.2    Droege, T.3    Kammer, S.4    Stieglitz, T.5
  • 14
    • 0030143820 scopus 로고    scopus 로고
    • The essential role played by adhesion in the technology of neurological prostheses
    • Donaldson PEK. The essential role played by adhesion in the technology of neurological prostheses. Int J Adhes Adhes 1996; 16:105-107.
    • (1996) Int J Adhes Adhes , vol.16 , pp. 105-107
    • Donaldson, P.E.K.1
  • 16
    • 18744384648 scopus 로고    scopus 로고
    • Parylene flexible neural probes integrated with microfluidic channels
    • Takeuchi S, Ziegler D, Yoshida Y, Mabuchi K, Suzuki T. Parylene flexible neural probes integrated with microfluidic channels. Lab Chip 2005;5:519-523.
    • (2005) Lab Chip , vol.5 , pp. 519-523
    • Takeuchi, S.1    Ziegler, D.2    Yoshida, Y.3    Mabuchi, K.4    Suzuki, T.5
  • 17
    • 34249846306 scopus 로고    scopus 로고
    • Neural probe design for reduced tissue encapsulation in CNS
    • Seymour JP, Kipke DR. Neural probe design for reduced tissue encapsulation in CNS. Biomaterials 2007;28:3594-3607.
    • (2007) Biomaterials , vol.28 , pp. 3594-3607
    • Seymour, J.P.1    Kipke, D.R.2
  • 19
    • 33746288952 scopus 로고    scopus 로고
    • Flexible organic field effect transistors for biomedical microimplants using polyimide and parylene C as substrate and insulator layers
    • Feili D, Schuettler M, Doerge T, Kammer S, Hoffmann KP, Stieglitz T. Flexible organic field effect transistors for biomedical microimplants using polyimide and parylene C as substrate and insulator layers. J Micromech Microeng 2006;16:1555-1561.
    • (2006) J Micromech Microeng , vol.16 , pp. 1555-1561
    • Feili, D.1    Schuettler, M.2    Doerge, T.3    Kammer, S.4    Hoffmann, K.P.5    Stieglitz, T.6
  • 20
    • 0017368305 scopus 로고
    • Parylene as a chronically stable. Reproducible microelectrode insulator
    • Loeb GE, Bak MJ, Salcman M, Schmidt EM. Parylene as a chronically stable. Reproducible microelectrode insulator. IEEE Trans Biomed Eng 1977;24:121-128.
    • (1977) IEEE Trans Biomed Eng , vol.24 , pp. 121-128
    • Loeb, G.E.1    Bak, M.J.2    Salcman, M.3    Schmidt, E.M.4
  • 21
    • 0029622323 scopus 로고
    • Toward the ultimate metal microelectrode
    • Loeb GE. Toward the ultimate metal microelectrode. J Microsci Methods 1995;63:175-183.
    • (1995) J Microsci Methods , vol.63 , pp. 175-183
    • Loeb, G.E.1
  • 25
    • 0037203991 scopus 로고    scopus 로고
    • Design and fabrication of high-density metal microelectrode array for neural recording
    • Xu C, Lemon W, Liu C. Design and fabrication of high-density metal microelectrode array for neural recording. Sens Actuators A Phys 2002;96:78-85.
    • (2002) Sens Actuators A Phys , vol.96 , pp. 78-85
    • Xu, C.1    Lemon, W.2    Liu, C.3
  • 26
    • 0345895607 scopus 로고    scopus 로고
    • Literature review: Biological safety of parylene C. Med Plast
    • Stark N. Literature review: Biological safety of parylene C. Med Plast Biomater 1996;3:30-35.
    • (1996) Biomater , vol.3 , pp. 30-35
    • Stark, N.1
  • 27
    • 26844533643 scopus 로고    scopus 로고
    • Assessing the performance and suitability of parylene coating
    • Wolgemuth L. Assessing the performance and suitability of parylene coating. Med Plast Biomater 2000;22:42-49.
    • (2000) Med Plast Biomater , vol.22 , pp. 42-49
    • Wolgemuth, L.1
  • 28
    • 0001419040 scopus 로고
    • A new general synthetic method for the preparation of linear poly-p-xylylenes
    • Gorham WF. A new general synthetic method for the preparation of linear poly-p-xylylenes. J Polym Sci Part A-1: Polym Chem 1966;4:3027-3039.
    • (1966) J Polym Sci Part A-1: Polym Chem , vol.4 , pp. 3027-3039
    • Gorham, W.F.1
  • 29
    • 0019588712 scopus 로고
    • Penetration of p-xylylene vapor into small channels prior to polymerization
    • Broer DJ, Luijks W. Penetration of p-xylylene vapor into small channels prior to polymerization. J Appl Polym Sci 1981;26: 2415-2422.
    • (1981) J Appl Polym Sci , vol.26 , pp. 2415-2422
    • Broer, D.J.1    Luijks, W.2
  • 31
    • 38349097111 scopus 로고    scopus 로고
    • Vapor deposition polymerization of para-xylylene derivatives-Mechanism and applications
    • Gazicki-Lipman M. Vapor deposition polymerization of para-xylylene derivatives-Mechanism and applications. J Vac Soc Jpn 2007;50:601-608.
    • (2007) J Vac Soc Jpn , vol.50 , pp. 601-608
    • Gazicki-Lipman, M.1
  • 32
    • 33645106481 scopus 로고    scopus 로고
    • Development of conformal PDMS and parylene coatings for microelectronics and MEMS packaging
    • ASME International Mechanical Engineering Congress and Exposition, Orlando, Florida, November 5-11
    • Lee H, Cho J. Development of conformal PDMS and parylene coatings for microelectronics and MEMS packaging. Proceedings of IMECE, ASME International Mechanical Engineering Congress and Exposition, Orlando, Florida, November 5-11, 2005.
    • (2005) Proceedings of IMECE
    • Lee, H.1    Cho, J.2
  • 33
    • 43949108535 scopus 로고    scopus 로고
    • Emergent properties of spatially organized poly(pxylylene) films fabricated by vapor deposition
    • Demirel MC. Emergent properties of spatially organized poly(pxylylene) films fabricated by vapor deposition. Colloid Surf A 2008;321:121-124.
    • (2008) Colloid Surf A , vol.321 , pp. 121-124
    • Demirel, M.C.1
  • 34
    • 46049095897 scopus 로고    scopus 로고
    • Effect of thermal and deposition processes on the surface morphology. Crystallinity, and adhesion of parylene-C
    • Hsu J-M, Rieth L, Kammer S, Orthner M, Solzbacher F. Effect of thermal and deposition processes on the surface morphology. Crystallinity, and adhesion of parylene-C. Sens Mater 2008;20: 87-102.
    • (2008) Sens Mater , vol.20 , pp. 87-102
    • Hsu, J.-M.1    Rieth, L.2    Kammer, S.3    Orthner, M.4    Solzbacher, F.5
  • 36
    • 77949495380 scopus 로고    scopus 로고
    • Merck KGaA. Safety data sheet: Silane A 174. Date of Issue: 2003/ 15/10. Available at http://www.merck-chemicals.co.uk/silane-a-174/ MDA-CHEM-107673/p- OYqb.s1LZKcAAAEWwuAfVhTl? attachments=MSDS. Accessed December 12, 2009.
    • Merck KGaA. Safety data sheet: Silane A 174. Date of Issue: 2003/ 15/10. Available at http://www.merck-chemicals.co.uk/silane-a-174/ MDA-CHEM-107673/p- OYqb.s1LZKcAAAEWwuAfVhTl? attachments=MSDS. Accessed December 12, 2009.
  • 37
    • 4444324684 scopus 로고    scopus 로고
    • Disinfection and sterilization in health care facilities: What clinicans need to know
    • Rutala WR, Weber DJ. Disinfection and sterilization in health care facilities: What clinicans need to know. Clin Inf Dis 2004;39: 702-709.
    • (2004) Clin Inf Dis , vol.39 , pp. 702-709
    • Rutala, W.R.1    Weber, D.J.2
  • 38
    • 0032771024 scopus 로고    scopus 로고
    • Morphology of poly(chloro-p-xylylene) CVD thin films
    • Senkevich JJ, Desu SB. Morphology of poly(chloro-p-xylylene) CVD thin films. Polymer 1999;40:5751-5759.
    • (1999) Polymer , vol.40 , pp. 5751-5759
    • Senkevich, J.J.1    Desu, S.B.2
  • 39
    • 77949523502 scopus 로고    scopus 로고
    • Peel tests
    • Packham DE, editor, 2nd ed. Chichester, UK: Wiley;
    • Packham DE. Peel tests. In: Packham DE, editor. Handbook of Adhesion, 2nd ed. Chichester, UK: Wiley; 2005. pp 311-315.
    • (2005) Handbook of Adhesion , pp. 311-315
    • Packham, D.E.1
  • 40
    • 69949183320 scopus 로고    scopus 로고
    • Deposition parameters determining insulation resistance and crystallinity of parylene c in neural implant encapsulation. In: Von Vander Sloten J
    • Verdonck P, Nyssen M, Haueisen J, editors, 1st ed. Berlin: Springer Berlin;
    • Hassler C, von Metzen R, Stieglitz T. Deposition parameters determining insulation resistance and crystallinity of parylene c in neural implant encapsulation. In: von Vander Sloten J, Verdonck P, Nyssen M, Haueisen J, editors. 4th European Conference of the International Federation for Medical and Biological Engineering, 1st ed. Berlin: Springer Berlin; 2008. pp 2439-2442.
    • (2008) 4th European Conference of the International Federation for Medical and Biological Engineering , pp. 2439-2442
    • Hassler, C.1    von Metzen, R.2    Stieglitz, T.3
  • 41
    • 8344238890 scopus 로고    scopus 로고
    • Evaluation of organic coatings with electrochemical impedance spectroscopy, Part 2: Application of EIS to coatings
    • Loveday D, Peterson P, Rodgers B. Evaluation of organic coatings with electrochemical impedance spectroscopy, Part 2: Application of EIS to coatings. J Coating Technol 2004;1:88-93.
    • (2004) J Coating Technol , vol.1 , pp. 88-93
    • Loveday, D.1    Peterson, P.2    Rodgers, B.3
  • 42
    • 0020207760 scopus 로고
    • Effect of glow discharge treatment of substrates on parylene-substrate adhesion
    • Sharma AK, Yasuda HK. Effect of glow discharge treatment of substrates on parylene-substrate adhesion. Am Vac Soc 1982;21: 994-998.
    • (1982) Am Vac Soc , vol.21 , pp. 994-998
    • Sharma, A.K.1    Yasuda, H.K.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.