Volumn 34, Issue 8, 2007, Pages 3249-3262

Verification of helical tomotherapy delivery using autoassociative kernel regression

Author keywords

Empirical modeling; Kernel regression; Quality assurance; Sinogram detector data; Tomotherapy

Indexed keywords

BIOMEDICAL ENGINEERING; ERROR DETECTION; FORECASTING; QUALITY ASSURANCE; UNCERTAINTY ANALYSIS;

EID: 34547455929     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.2754059     Document Type: Article
Times cited : (10)

References (49)
  • 1
    • 10744231470 scopus 로고    scopus 로고
    • Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee
    • G. A. Ezzell, J. M. Galvin, D. Low, J. R. Palta, I. Rosen, M. B. Sharpe, P. Xia, Y. Xiao, L. Xing, and C. X. Yu, " Guidance document on delivery, treatment planning, and clinical implementation of IMRT: Report of the IMRT subcommittee of the AAPM radiation therapy committee., " Med. Phys. 30, 2089-2097 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 2089-2097
    • Ezzell, G.A.1    Galvin, J.M.2    Low, D.3    Palta, J.R.4    Rosen, I.5    Sharpe, M.B.6    Xia, P.7    Xiao, Y.8    Xing, L.9    Yu, C.X.10
  • 2
    • 34547486010 scopus 로고    scopus 로고
    • NCI update on the use of IMRT in clinical trials
    • J. Deye, " NCI update on the use of IMRT in clinical trials., " AAPM Newsletter 27, 22-24 (2002).
    • (2002) AAPM Newsletter , vol.27 , pp. 22-24
    • Deye, J.1
  • 4
    • 24944537076 scopus 로고    scopus 로고
    • Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group
    • A. Molineu, D. S. Followill, P. A. Balter, W. F. Hanson, M. T. Gillin, M. S. Huq, A. Eisbruch, and G. S. Ibbott, " Design and implementation of an anthropomorphic quality assurance phantom for intensity-modulated radiation therapy for the radiation therapy oncology group., " Int. J. Radiat. Oncol., Biol., Phys. 63, 577-583 (2005).
    • (2005) Int. J. Radiat. Oncol., Biol., Phys. , vol.63 , pp. 577-583
    • Molineu, A.1    Followill, D.S.2    Balter, P.A.3    Hanson, W.F.4    Gillin, M.T.5    Huq, M.S.6    Eisbruch, A.7    Ibbott, G.S.8
  • 6
    • 1942516798 scopus 로고    scopus 로고
    • The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes
    • R. J. Louwe, R. Tilenburg, K. M. van Ingen, B. J. Mijnheer, and M. B. van Herk, " The stability of liquid-filled matrix ionization chamber electronic portal imaging devices for dosimetry purposes., " Med. Phys. 31, 819-827 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 819-827
    • Louwe, R.J.1    Tilenburg, R.2    Van Ingen, K.M.3    Mijnheer, B.J.4    Van Herk, M.B.5
  • 7
    • 0033867804 scopus 로고    scopus 로고
    • Leaf position verification during dynamic beam delivery: A comparison of three applications using electronic portal imaging
    • M. Partridge, P. M. Evans, M. B. van Herk, L. S. Ploeger, J. B. Geoffrey, and H. V. James, " Leaf position verification during dynamic beam delivery: A comparison of three applications using electronic portal imaging., " Med. Phys. 27, 1601-1609 (2000).
    • (2000) Med. Phys. , vol.27 , pp. 1601-1609
    • Partridge, M.1    Evans, P.M.2    Van Herk, M.B.3    Ploeger, L.S.4    Geoffrey, J.B.5    James, H.V.6
  • 8
    • 0035180108 scopus 로고    scopus 로고
    • An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factor
    • J. Chang, G. S. Mageras, C. L. Clifton, and W. Lutz, " An iterative EPID calibration procedure for dosimetric verification that considers the EPID scattering factor., " Med. Phys. 28, 2247-2257 (2001).
    • (2001) Med. Phys. , vol.28 , pp. 2247-2257
    • Chang, J.1    Mageras, G.S.2    Clifton, C.L.3    Lutz, W.4
  • 9
    • 0035989982 scopus 로고    scopus 로고
    • A method for gemetrical verification of dynamic intensity modulated radiotherapy using a scanning electronic portal imaging device
    • L. S. Ploeger, M. P. Smitsmans, K. G. Gilhuijs, and M. B. van Herk, " A method for gemetrical verification of dynamic intensity modulated radiotherapy using a scanning electronic portal imaging device., " Med. Phys. 29, 389-395 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 389-395
    • Ploeger, L.S.1    Smitsmans, M.P.2    Gilhuijs, K.G.3    Van Herk, M.B.4
  • 10
    • 0036735619 scopus 로고    scopus 로고
    • Fast and accurate leaf verification for dynamic multileaf collimator using an electronic portal imaging device
    • S. C. Vieira, L. P. Dirkx, K. L. Pasma, and B. J. M. Heijmen, " Fast and accurate leaf verification for dynamic multileaf collimator using an electronic portal imaging device., " Med. Phys. 29, 2034-2040 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 2034-2040
    • Vieira, S.C.1    Dirkx, L.P.2    Pasma, K.L.3    Heijmen, B.J.M.4
  • 11
    • 12244304896 scopus 로고    scopus 로고
    • Verification of multileaf collimator leaf positions using an electronic portal imaging device
    • S. S. Samant, W. Zheng, N. A. Parra, J. Chandler, A. Gopal, J. Wu, J. Jain, Y. Zhu, and M. Sontag, " Verification of multileaf collimator leaf positions using an electronic portal imaging device., " Med. Phys. 29, 2900-2912 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 2900-2912
    • Samant, S.S.1    Zheng, W.2    Parra, N.A.3    Chandler, J.4    Gopal, A.5    Wu, J.6    Jain, J.7    Zhu, Y.8    Sontag, M.9
  • 12
    • 0036797456 scopus 로고    scopus 로고
    • An investigation of a new amorphous silicon electronic portal imagining device for transit dosimetry
    • E. E. Grein, R. Lee, and K. Luchka, " An investigation of a new amorphous silicon electronic portal imagining device for transit dosimetry., " Med. Phys. 29, 2262-2268 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 2262-2268
    • Grein, E.E.1    Lee, R.2    Luchka, K.3
  • 13
    • 0038033842 scopus 로고    scopus 로고
    • Verification of dynamic IMRT
    • P. B. Greer and C. C. Popescu, " Verification of dynamic IMRT.. " Med. Phys. 30, 1618-1627 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 1618-1627
    • Greer, P.B.1    Popescu, C.C.2
  • 14
    • 0347625746 scopus 로고    scopus 로고
    • Dosimetric IMRT verification with a flat-panel EPID
    • B. Warkentin, S. Steciw, S. Rathee, and B. G. Fallone, " Dosimetric IMRT verification with a flat-panel EPID., " Med. Phys. 30, 3143-3155 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 3143-3155
    • Warkentin, B.1    Steciw, S.2    Rathee, S.3    Fallone, B.G.4
  • 15
    • 1242271376 scopus 로고    scopus 로고
    • Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device
    • L. N. McDermott, R. J. Louwe, J. J. Sonke, and M. B. van Herk, and B. J. Mijnheer, " Dose-response and ghosting effects of an amorphous silicon electronic portal imaging device., " Med. Phys. 31, 285-295 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 285-295
    • McDermott, L.N.1    Louwe, R.J.2    Sonke, J.J.3    Van Herk, M.B.4    Mijnheer, B.J.5
  • 16
    • 3142693958 scopus 로고    scopus 로고
    • Use of EPID for leaf position accuracy QA or dynamic multi-leaf collimator (DMLC) treatment
    • J. Chang, C. H. Obcemea, J. Sillanpaa, J. Mechalakos, and C. Burman, " Use of EPID for leaf position accuracy QA or dynamic multi-leaf collimator (DMLC) treatment., " Med. Phys. 31, 2091-2096 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 2091-2096
    • Chang, J.1    Obcemea, C.H.2    Sillanpaa, J.3    Mechalakos, J.4    Burman, C.5
  • 17
    • 17844399582 scopus 로고    scopus 로고
    • Comprehensive Monte Carlo calculation of the point spread function for a commercial a-Si EPID
    • C. Kirkby and R. Sloboda, " Comprehensive Monte Carlo calculation of the point spread function for a commercial a-Si EPID., " Med. Phys. 32, 1115-1127 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 1115-1127
    • Kirkby, C.1    Sloboda, R.2
  • 18
    • 23944453854 scopus 로고    scopus 로고
    • Consequences of the spectral response of an a-Si EPID and implications for dosimetric calibration
    • C. Kirkby and R. Sloboda, " Consequences of the spectral response of an a-Si EPID and implications for dosimetric calibration., " Med. Phys. 32, 2649-2658 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 2649-2658
    • Kirkby, C.1    Sloboda, R.2
  • 19
    • 27244439568 scopus 로고    scopus 로고
    • Dose-response characteristics of an amorphous silicon EPID
    • P. Winkler, A. Hefner, and G. Dietmar, " Dose-response characteristics of an amorphous silicon EPID., " Med. Phys. 32, 3095-3105 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 3095-3105
    • Winkler, P.1    Hefner, A.2    Dietmar, G.3
  • 20
    • 28044439624 scopus 로고    scopus 로고
    • Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager
    • G. J. Budgell, Q. Zhang, R. J. Trouncer, and R. I. Mackay, " Improving IMRT quality control efficiency using an amorphous silicon electronic portal imager., " Med. Phys. 32, 3267-3276 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 3267-3276
    • Budgell, G.J.1    Zhang, Q.2    Trouncer, R.J.3    MacKay, R.I.4
  • 21
    • 33644999533 scopus 로고    scopus 로고
    • Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry
    • P. B. Greer, " Correction of pixel sensitivity variation and off-axis response for amorphous silicon EPID dosimetry., " Med. Phys. 32, 3558-3568 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 3558-3568
    • Greer, P.B.1
  • 22
    • 0035081653 scopus 로고    scopus 로고
    • Accurate measurement of the dynamic response of a scanning electronic portal imaging device
    • L. S. Ploeger, M. P. Smitsmans, K. G. Gilhuijs, and M. B. van Herk, " Accurate measurement of the dynamic response of a scanning electronic portal imaging device., " Med. Phys. 28, 310-316 (2001).
    • (2001) Med. Phys. , vol.28 , pp. 310-316
    • Ploeger, L.S.1    Smitsmans, M.P.2    Gilhuijs, K.G.3    Van Herk, M.B.4
  • 23
    • 1242293815 scopus 로고    scopus 로고
    • Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager
    • J. J. Sonke, L. S. Ploeger, B. Brand, M. P. Smitsmans, and M. B. van Herk, " Leaf trajectory verification during dynamic intensity modulated radiotherapy using an amorphous silicon flat panel imager., " Med. Phys. 31, 389-395 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 389-395
    • Sonke, J.J.1    Ploeger, L.S.2    Brand, B.3    Smitsmans, M.P.4    Van Herk, M.B.5
  • 24
    • 14644390214 scopus 로고    scopus 로고
    • Three-dimensional IMRT verification with a flat panel EPID
    • S. Steciw, B. Warkentin, S. Rathee, and B. G. Fallone, " Three-dimensional IMRT verification with a flat panel EPID., " Med. Phys. 32, 600-612 (2005).
    • (2005) Med. Phys. , vol.32 , pp. 600-612
    • Steciw, S.1    Warkentin, B.2    Rathee, S.3    Fallone, B.G.4
  • 25
    • 31844436637 scopus 로고    scopus 로고
    • Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method
    • M. Wendlig, R. J. Louwe, L. N. McDermott, J. J. Sonke, M. B. van Herk, B. J. Mijnheer, " Accurate two-dimensional IMRT verification using a back-projection EPID dosimetry method., " Med. Phys. 33, 259-273 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 259-273
    • Wendlig, M.1    Louwe, R.J.2    McDermott, L.N.3    Sonke, J.J.4    Van Herk, M.B.5    Mijnheer, B.J.6
  • 26
    • 1642377027 scopus 로고    scopus 로고
    • Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device
    • O. A. Zeidan, J. G. Li, M. K. Ranade, A. M. Stell, and J. F. Dempsey, " Verification of step-and-shoot IMRT delivery using a fast video-based electronic portal imaging device., " Med. Phys. 31, 463-476 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 463-476
    • Zeidan, O.A.1    Li, J.G.2    Ranade, M.K.3    Stell, A.M.4    Dempsey, J.F.5
  • 27
    • 31044446718 scopus 로고    scopus 로고
    • A high-speed scintillation-based electronic portal imaging device to quantitatively characterize IMRT delivery
    • M. K. Ranade, B. D. Lynch, J. G. Li, and J. F. Dempsey, " A high-speed scintillation-based electronic portal imaging device to quantitatively characterize IMRT delivery., " Med. Phys. 33, 106-110 (2006).
    • (2006) Med. Phys. , vol.33 , pp. 106-110
    • Ranade, M.K.1    Lynch, B.D.2    Li, J.G.3    Dempsey, J.F.4
  • 30
    • 0032947653 scopus 로고    scopus 로고
    • Characterization for the output for helical delivery of intensity modulated slit beams
    • J. P. Balog, T. R. Mackie, P. Reckwerdt, M. Glass, and L. Angelos, " Characterization for the output for helical delivery of intensity modulated slit beams., " Med. Phys. 26, 55-64 (1999).
    • (1999) Med. Phys. , vol.26 , pp. 55-64
    • Balog, J.P.1    MacKie, T.R.2    Reckwerdt, P.3    Glass, M.4    Angelos, L.5
  • 33
    • 33745495381 scopus 로고    scopus 로고
    • History of tomotherapy
    • T. R. Mackie, " History of tomotherapy., " Phys. Med. Biol. 51, 427-449 (2006).
    • (2006) Phys. Med. Biol. , vol.51 , pp. 427-449
    • MacKie, T.R.1
  • 39
    • 33750348445 scopus 로고    scopus 로고
    • MSET performance optimization through regularization
    • J. W. Hines and A. A. Usynin, " MSET performance optimization through regularization., " Nuclear Engineering and Technology 37, 177-184 (2005).
    • (2005) Nuclear Engineering and Technology , vol.37 , pp. 177-184
    • Hines, J.W.1    Usynin, A.A.2
  • 40
    • 23344453352 scopus 로고    scopus 로고
    • Experiences and applications of PEANO for on-line monitoring in power plants
    • P. Fantoni, " Experiences and applications of PEANO for on-line monitoring in power plants., " Progress in Nuclear Energy for a special issue on "Lessons Learned from Computational Intelligence in Nuclear Application," 46, 3-4 (2005).
    • (2005) Progress in Nuclear Energy , vol.46 , pp. 3-4
    • Fantoni, P.1
  • 44
    • 2142806741 scopus 로고    scopus 로고
    • Uncertainty analysis of memory based sensor validation techniques
    • A. V. Gribok, A. M. Urmanov, and J. W. Hines, " Uncertainty analysis of memory based sensor validation techniques., " Real-Time Syst. 27, 7-26 (2004).
    • (2004) Real-Time Syst. , vol.27 , pp. 7-26
    • Gribok, A.V.1    Urmanov, A.M.2    Hines, J.W.3
  • 47
    • 0025472146 scopus 로고
    • Sequential probability ratio tests for reactor signal validation and sensor surveillance applications
    • K. E. Humenik and K. C. Gross, " Sequential probability ratio tests for reactor signal validation and sensor surveillance applications., " Nucl. Sci. Eng. 105, 383-390 (1990).
    • (1990) Nucl. Sci. Eng. , vol.105 , pp. 383-390
    • Humenik, K.E.1    Gross, K.C.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.