Volumn 248, Issue , 2016, Pages 196-202

Adherence to guidelines strongly improves reproducibility of brachial artery flow-mediated dilation

Author keywords

Atherosclerosis; Cardiovascular disease; Endothelial function; Methodology; Reproducibility

Indexed keywords

ARTERY BLOOD FLOW; ARTERY DIAMETER; ARTERY DILATATION; ARTICLE; ATHEROSCLEROSIS; BRACHIAL ARTERY; HUMAN; MEASUREMENT ERROR; PRACTICE GUIDELINE; PRIORITY JOURNAL; REPRODUCIBILITY; SYSTEMATIC REVIEW; VASCULAR ENDOTHELIUM; ADULT; AGED; BLOOD FLOW; BLOOD FLOW VELOCITY; CARDIOVASCULAR DISEASE; DILATATION; MIDDLE AGED; PATHOPHYSIOLOGY; PROTOCOL COMPLIANCE; REGRESSION ANALYSIS; RISK FACTOR; VASODILATATION; YOUNG ADULT;

EID: 84961615908     PISSN: 00219150     EISSN: 18791484     Source Type: Journal    
DOI: 10.1016/j.atherosclerosis.2016.03.011     Document Type: Article
Times cited : (64)

References (44)
  • 1
    • 33947716172 scopus 로고    scopus 로고
    • Endothelial function and dysfunction: testing and clinical relevance
    • Deanfield J.E., Halcox J.P., Rabelink T.J. Endothelial function and dysfunction: testing and clinical relevance. Circulation 2007, 115:1285-1295.
    • (2007) Circulation , vol.115 , pp. 1285-1295
    • Deanfield, J.E.1    Halcox, J.P.2    Rabelink, T.J.3
  • 2
    • 0026478022 scopus 로고
    • Non-invasive detection of endothelial dysfunction in children and adults at risk of atherosclerosis
    • Celermajer D.S., Sorensen K.E., Gooch V.M., et al. Non-invasive detection of endothelial dysfunction in children and adults at risk of atherosclerosis. Lancet 1992, 340:1111-1115.
    • (1992) Lancet , vol.340 , pp. 1111-1115
    • Celermajer, D.S.1    Sorensen, K.E.2    Gooch, V.M.3
  • 3
    • 0032535699 scopus 로고    scopus 로고
    • Endothelium-dependent flow-mediated vasodilation in coronary and brachial arteries in suspected coronary artery disease
    • A1537-A1538
    • Takase B., Uehata A., Akima T., et al. Endothelium-dependent flow-mediated vasodilation in coronary and brachial arteries in suspected coronary artery disease. Am. J. Cardiol. 1998, 82:1535-1539. A1537-1538.
    • (1998) Am. J. Cardiol. , vol.82 , pp. 1535-1539
    • Takase, B.1    Uehata, A.2    Akima, T.3
  • 4
    • 84883790321 scopus 로고    scopus 로고
    • Flow-mediated dilation and cardiovascular risk prediction: a systematic review with meta-analysis
    • Ras R.T., Streppel M.T., Draijer R., et al. Flow-mediated dilation and cardiovascular risk prediction: a systematic review with meta-analysis. Int. J. Cardiol. 2013, 168:344-351.
    • (2013) Int. J. Cardiol. , vol.168 , pp. 344-351
    • Ras, R.T.1    Streppel, M.T.2    Draijer, R.3
  • 5
    • 77957933610 scopus 로고    scopus 로고
    • Prediction of future cardiovascular outcomes by flow-mediated vasodilatation of brachial artery: a meta-analysis
    • Inaba Y., Chen J.A., Bergmann S.R. Prediction of future cardiovascular outcomes by flow-mediated vasodilatation of brachial artery: a meta-analysis. Int. J. Cardiovasc. Imaging 2010, 26:631-640.
    • (2010) Int. J. Cardiovasc. Imaging , vol.26 , pp. 631-640
    • Inaba, Y.1    Chen, J.A.2    Bergmann, S.R.3
  • 6
    • 84877834826 scopus 로고    scopus 로고
    • Functional and structural alterations of large arteries: methodological issues
    • Bianchini E., Giannarelli C., Bruno R.M., et al. Functional and structural alterations of large arteries: methodological issues. Curr. Pharm. Des. 2013, 19:2390-2400.
    • (2013) Curr. Pharm. Des. , vol.19 , pp. 2390-2400
    • Bianchini, E.1    Giannarelli, C.2    Bruno, R.M.3
  • 7
    • 38549122312 scopus 로고    scopus 로고
    • Importance of measuring the time course of flow-mediated dilatation in humans
    • Black M.A., Cable N.T., Thijssen D.H., et al. Importance of measuring the time course of flow-mediated dilatation in humans. Hypertension 2008, 51:203-210.
    • (2008) Hypertension , vol.51 , pp. 203-210
    • Black, M.A.1    Cable, N.T.2    Thijssen, D.H.3
  • 8
    • 0035208280 scopus 로고    scopus 로고
    • Flow-mediated dilatation following wrist and upper arm occlusion in humans: the contribution of nitric oxide
    • Doshi S.N., Naka K.K., Payne N., et al. Flow-mediated dilatation following wrist and upper arm occlusion in humans: the contribution of nitric oxide. Clin. Sci. Lond. 2001, 101:629-635.
    • (2001) Clin. Sci. Lond. , vol.101 , pp. 629-635
    • Doshi, S.N.1    Naka, K.K.2    Payne, N.3
  • 9
    • 0035793340 scopus 로고    scopus 로고
    • Heterogenous nature of flow-mediated dilatation in human conduit arteries in vivo: relevance to endothelial dysfunction in hypercholesterolemia
    • Mullen M.J., Kharbanda R.K., Cross J., et al. Heterogenous nature of flow-mediated dilatation in human conduit arteries in vivo: relevance to endothelial dysfunction in hypercholesterolemia. Circ. Res. 2001, 88:145-151.
    • (2001) Circ. Res. , vol.88 , pp. 145-151
    • Mullen, M.J.1    Kharbanda, R.K.2    Cross, J.3
  • 10
    • 0037116573 scopus 로고    scopus 로고
    • Guidelines for the ultrasound assessment of endothelial-dependent flow-mediated vasodilation of the brachial artery: a report of the International Brachial Artery Reactivity Task Force
    • Corretti M.C., Anderson T.J., Benjamin E.J., et al. Guidelines for the ultrasound assessment of endothelial-dependent flow-mediated vasodilation of the brachial artery: a report of the International Brachial Artery Reactivity Task Force. J. Am. Coll. Cardiol. 2002, 39:257-265.
    • (2002) J. Am. Coll. Cardiol. , vol.39 , pp. 257-265
    • Corretti, M.C.1    Anderson, T.J.2    Benjamin, E.J.3
  • 11
    • 78651308651 scopus 로고    scopus 로고
    • Assessment of flow-mediated dilation in humans: a methodological and physiological guideline
    • Thijssen D.H., Black M.A., Pyke K.E., et al. Assessment of flow-mediated dilation in humans: a methodological and physiological guideline. Am. J. Physiol. Heart Circ. Physiol. 2011, 300:H2-H12.
    • (2011) Am. J. Physiol. Heart Circ. Physiol. , vol.300 , pp. H2-H12
    • Thijssen, D.H.1    Black, M.A.2    Pyke, K.E.3
  • 13
    • 0033936333 scopus 로고    scopus 로고
    • Measures of reliability in sports medicine and science
    • Hopkins W.G. Measures of reliability in sports medicine and science. Sports Med. 2000, 30:1-15.
    • (2000) Sports Med. , vol.30 , pp. 1-15
    • Hopkins, W.G.1
  • 14
    • 84897052755 scopus 로고    scopus 로고
    • New methodological approach to improve reproducibility of brachial artery flow-mediated dilatation
    • Kanahara M., Harada H., Katoh A., et al. New methodological approach to improve reproducibility of brachial artery flow-mediated dilatation. Echocardiography 2014, 31:197-202.
    • (2014) Echocardiography , vol.31 , pp. 197-202
    • Kanahara, M.1    Harada, H.2    Katoh, A.3
  • 15
    • 84890537763 scopus 로고    scopus 로고
    • Variability and reproducibility of flow-mediated dilatation in a multicentre clinical trial
    • Charakida M., de Groot E., Loukogeorgakis S.P., et al. Variability and reproducibility of flow-mediated dilatation in a multicentre clinical trial. Eur. Heart J. 2013, 34:3501-3507.
    • (2013) Eur. Heart J. , vol.34 , pp. 3501-3507
    • Charakida, M.1    de Groot, E.2    Loukogeorgakis, S.P.3
  • 16
    • 84862758115 scopus 로고    scopus 로고
    • Assessment of flow-mediated dilation reproducibility: a nationwide multicenter study
    • Ghiadoni L., Faita F., Salvetti M., et al. Assessment of flow-mediated dilation reproducibility: a nationwide multicenter study. J. Hypertens. 2012, 30:1399-1405.
    • (2012) J. Hypertens. , vol.30 , pp. 1399-1405
    • Ghiadoni, L.1    Faita, F.2    Salvetti, M.3
  • 17
    • 84861819113 scopus 로고    scopus 로고
    • Reproducibility of different methods to measure the endothelial function
    • Onkelinx S., Cornelissen V., Goetschalckx K., et al. Reproducibility of different methods to measure the endothelial function. Vasc. Med. 2012, 17:79-84.
    • (2012) Vasc. Med. , vol.17 , pp. 79-84
    • Onkelinx, S.1    Cornelissen, V.2    Goetschalckx, K.3
  • 18
    • 78449296165 scopus 로고    scopus 로고
    • Pulsatility index change of brachial artery shows better reproducibility than flow-mediated vasodilation
    • Lima J.C., Martins W.P., Nastri C.O., et al. Pulsatility index change of brachial artery shows better reproducibility than flow-mediated vasodilation. Ultrasound Med. Biol. 2010, 36:2036-2041.
    • (2010) Ultrasound Med. Biol. , vol.36 , pp. 2036-2041
    • Lima, J.C.1    Martins, W.P.2    Nastri, C.O.3
  • 19
    • 67049158579 scopus 로고    scopus 로고
    • Retrograde flow and shear rate acutely impair endothelial function in humans
    • Thijssen D.H., Dawson E.A., Tinken T.M., et al. Retrograde flow and shear rate acutely impair endothelial function in humans. Hypertension 2009, 53:986-992.
    • (2009) Hypertension , vol.53 , pp. 986-992
    • Thijssen, D.H.1    Dawson, E.A.2    Tinken, T.M.3
  • 20
    • 43049181357 scopus 로고    scopus 로고
    • Methodological approaches to optimize reproducibility and power in clinical studies of flow-mediated dilation
    • Donald A.E., Halcox J.P., Charakida M., et al. Methodological approaches to optimize reproducibility and power in clinical studies of flow-mediated dilation. J. Am. Coll. Cardiol. 2008, 51:1959-1964.
    • (2008) J. Am. Coll. Cardiol. , vol.51 , pp. 1959-1964
    • Donald, A.E.1    Halcox, J.P.2    Charakida, M.3
  • 21
    • 38049051455 scopus 로고    scopus 로고
    • Interobserver and intraobserver variability of flow-mediated vasodilatation of the brachial artery
    • Simova I., Nossikoff A., Denchev S. Interobserver and intraobserver variability of flow-mediated vasodilatation of the brachial artery. Echocardiography 2008, 25:77-83.
    • (2008) Echocardiography , vol.25 , pp. 77-83
    • Simova, I.1    Nossikoff, A.2    Denchev, S.3
  • 22
    • 33845372780 scopus 로고    scopus 로고
    • New monitoring software for larger clinical application of brachial artery flow-mediated vasodilatation measurements
    • Craiem D., Chironi G., Gariepy J., et al. New monitoring software for larger clinical application of brachial artery flow-mediated vasodilatation measurements. J. Hypertens. 2007, 25:133-140.
    • (2007) J. Hypertens. , vol.25 , pp. 133-140
    • Craiem, D.1    Chironi, G.2    Gariepy, J.3
  • 23
    • 34548842475 scopus 로고    scopus 로고
    • Reproducibility of the flow-mediated dilation response to acute exercise in overweight men
    • Harris R.A., Padilla J., Hanlon K.P., et al. Reproducibility of the flow-mediated dilation response to acute exercise in overweight men. Ultrasound Med. Biol. 2007, 33:1579-1585.
    • (2007) Ultrasound Med. Biol. , vol.33 , pp. 1579-1585
    • Harris, R.A.1    Padilla, J.2    Hanlon, K.P.3
  • 24
    • 37449016016 scopus 로고    scopus 로고
    • Reliability of brachial artery flow-mediated dilatation measurement using ultrasound
    • Meirelles Cde M., Leite S.P., Montenegro C.A., et al. Reliability of brachial artery flow-mediated dilatation measurement using ultrasound. Arq. Bras. Cardiol. 2007, 89(160-167):176-183.
    • (2007) Arq. Bras. Cardiol. , vol.89 , Issue.160-167 , pp. 176-183
    • Meirelles Cde, M.1    Leite, S.P.2    Montenegro, C.A.3
  • 25
    • 33750463408 scopus 로고    scopus 로고
    • Non-invasive assessment of endothelial function: which technique?
    • Donald A.E., Charakida M., Cole T.J., et al. Non-invasive assessment of endothelial function: which technique?. J. Am. Coll. Cardiol. 2006, 48:1846-1850.
    • (2006) J. Am. Coll. Cardiol. , vol.48 , pp. 1846-1850
    • Donald, A.E.1    Charakida, M.2    Cole, T.J.3
  • 26
    • 33644785903 scopus 로고    scopus 로고
    • Variability of flow-mediated dilation measurements with repetitive reactive hyperemia
    • Harris R.A., Padilla J., Rink L.D., et al. Variability of flow-mediated dilation measurements with repetitive reactive hyperemia. Vasc. Med. 2006, 11:1-6.
    • (2006) Vasc. Med. , vol.11 , pp. 1-6
    • Harris, R.A.1    Padilla, J.2    Rink, L.D.3
  • 27
    • 33644829343 scopus 로고    scopus 로고
    • Cardiovascular magnetic resonance imaging for non-invasive assessment of vascular function: validation against ultrasound
    • Leeson C.P., Robinson M., Francis J.M., et al. Cardiovascular magnetic resonance imaging for non-invasive assessment of vascular function: validation against ultrasound. J. Cardiovasc Magn. Reson 2006, 8:381-387.
    • (2006) J. Cardiovasc Magn. Reson , vol.8 , pp. 381-387
    • Leeson, C.P.1    Robinson, M.2    Francis, J.M.3
  • 28
    • 84961622206 scopus 로고    scopus 로고
    • Reproducibility of B-mode ultrasound brachial Flow Mediated Dilation measurements
    • Elsen B.M., Scholten H.J., Schilder J.C.M., et al. Reproducibility of B-mode ultrasound brachial Flow Mediated Dilation measurements. Atheroscler. Suppl. 2005, 6. 126-126.
    • (2005) Atheroscler. Suppl. , vol.6 , pp. 126
    • Elsen, B.M.1    Scholten, H.J.2    Schilder, J.C.M.3
  • 29
    • 13544264771 scopus 로고    scopus 로고
    • Limitations of non-invasive endothelial function assessment by brachial artery flow-mediated dilatation
    • Sejda T., Pit'ha J., Svandova E., et al. Limitations of non-invasive endothelial function assessment by brachial artery flow-mediated dilatation. Clin. Physiol. Funct. Imaging 2005, 25:58-61.
    • (2005) Clin. Physiol. Funct. Imaging , vol.25 , pp. 58-61
    • Sejda, T.1    Pit'ha, J.2    Svandova, E.3
  • 30
    • 2442680416 scopus 로고
    • Relationship between blood velocity and conduit artery diameter and the effects of smoking on vascular responsiveness
    • Stoner L., Sabatier M., Edge K., et al. Relationship between blood velocity and conduit artery diameter and the effects of smoking on vascular responsiveness. J. Appl. Physiol. 1985, 2004(96):2139-2145.
    • (1985) J. Appl. Physiol. , vol.2004 , Issue.96 , pp. 2139-2145
    • Stoner, L.1    Sabatier, M.2    Edge, K.3
  • 31
    • 5644272818 scopus 로고    scopus 로고
    • Biological correlates of day-to-day variation in flow-mediated dilation in individuals with Type 2 diabetes: a study of test-retest reliability
    • West S.G., Wagner P., Schoemer S.L., et al. Biological correlates of day-to-day variation in flow-mediated dilation in individuals with Type 2 diabetes: a study of test-retest reliability. Diabetologia 2004, 47:1625-1631.
    • (2004) Diabetologia , vol.47 , pp. 1625-1631
    • West, S.G.1    Wagner, P.2    Schoemer, S.L.3
  • 32
    • 0036721230 scopus 로고    scopus 로고
    • A rapid and reproducible on line automated technique to determine endothelial function
    • Sidhu J.S., Newey V.R., Nassiri D.K., et al. A rapid and reproducible on line automated technique to determine endothelial function. Heart 2002, 88:289-292.
    • (2002) Heart , vol.88 , pp. 289-292
    • Sidhu, J.S.1    Newey, V.R.2    Nassiri, D.K.3
  • 33
    • 0035712717 scopus 로고    scopus 로고
    • Automatic evaluation of arterial diameter variation from vascular echographic images
    • Beux F., Carmassi S., Salvetti M.V., et al. Automatic evaluation of arterial diameter variation from vascular echographic images. Ultrasound Med. Biol. 2001, 27:1621-1629.
    • (2001) Ultrasound Med. Biol. , vol.27 , pp. 1621-1629
    • Beux, F.1    Carmassi, S.2    Salvetti, M.V.3
  • 34
    • 0034888489 scopus 로고    scopus 로고
    • Flow-mediated vasodilation is not impaired when HDL-cholesterol is lowered by substituting carbohydrates for monounsaturated fat
    • de Roos N.M., Bots M.L., Siebelink E., et al. Flow-mediated vasodilation is not impaired when HDL-cholesterol is lowered by substituting carbohydrates for monounsaturated fat. Br. J. Nutr. 2001, 86:181-188.
    • (2001) Br. J. Nutr. , vol.86 , pp. 181-188
    • de Roos, N.M.1    Bots, M.L.2    Siebelink, E.3
  • 35
    • 0034776872 scopus 로고    scopus 로고
    • Brachial flow-mediated vasodilator responses in population-based research: methods, reproducibility and effects of age, gender and baseline diameter
    • Herrington D.M., Fan L., Drum M., et al. Brachial flow-mediated vasodilator responses in population-based research: methods, reproducibility and effects of age, gender and baseline diameter. J. Cardiovasc. Risk 2001, 8:319-328.
    • (2001) J. Cardiovasc. Risk , vol.8 , pp. 319-328
    • Herrington, D.M.1    Fan, L.2    Drum, M.3
  • 36
    • 0034924532 scopus 로고    scopus 로고
    • Improved analysis of brachial artery ultrasound using a novel edge-detection software system
    • Woodman R.J., Playford D.A., Watts G.F., et al. Improved analysis of brachial artery ultrasound using a novel edge-detection software system. J. Appl. Physiol. 2001, 91:929-937.
    • (2001) J. Appl. Physiol. , vol.91 , pp. 929-937
    • Woodman, R.J.1    Playford, D.A.2    Watts, G.F.3
  • 37
    • 0033745671 scopus 로고    scopus 로고
    • Evaluation of endothelium-dependent vasodilation in the human peripheral circulation
    • Lind L., Hall J., Larsson A., et al. Evaluation of endothelium-dependent vasodilation in the human peripheral circulation. Clin. Physiol. 2000, 20:440-448.
    • (2000) Clin. Physiol. , vol.20 , pp. 440-448
    • Lind, L.1    Hall, J.2    Larsson, A.3
  • 38
    • 0033740535 scopus 로고    scopus 로고
    • Automated ultrasonic measurement of human arteries for the determination of endothelial function
    • Preik M., Lauer T., Heiss C., et al. Automated ultrasonic measurement of human arteries for the determination of endothelial function. Ultraschall Med. 2000, 21:195-198.
    • (2000) Ultraschall Med. , vol.21 , pp. 195-198
    • Preik, M.1    Lauer, T.2    Heiss, C.3
  • 39
    • 0032430788 scopus 로고    scopus 로고
    • Non-invasive measurements of arterial structure and function: repeatability, interrelationships and trial sample size
    • Liang Y.L., Teede H., Kotsopoulos D., et al. Non-invasive measurements of arterial structure and function: repeatability, interrelationships and trial sample size. Clin. Sci. Lond. 1998, 95:669-679.
    • (1998) Clin. Sci. Lond. , vol.95 , pp. 669-679
    • Liang, Y.L.1    Teede, H.2    Kotsopoulos, D.3
  • 40
    • 0031415554 scopus 로고    scopus 로고
    • Reproducibility of brachial ultrasonography and flow-mediated dilatation (FMD) for assessing endothelial function
    • Hardie K.L., Kinlay S., Hardy D.B., et al. Reproducibility of brachial ultrasonography and flow-mediated dilatation (FMD) for assessing endothelial function. Aust. N. Z. J. Med. 1997, 27:649-652.
    • (1997) Aust. N. Z. J. Med. , vol.27 , pp. 649-652
    • Hardie, K.L.1    Kinlay, S.2    Hardy, D.B.3
  • 41
    • 33847790738 scopus 로고    scopus 로고
    • A system for real-time measurement of the brachial artery diameter in B-mode ultrasound images
    • Gemignani V., Faita F., Ghiadoni L., et al. A system for real-time measurement of the brachial artery diameter in B-mode ultrasound images. IEEE Trans. Med. Imaging 2007, 26:393-404.
    • (2007) IEEE Trans. Med. Imaging , vol.26 , pp. 393-404
    • Gemignani, V.1    Faita, F.2    Ghiadoni, L.3
  • 42
    • 0036826355 scopus 로고    scopus 로고
    • Automated analysis of brachial ultrasound image sequences: early detection of cardiovascular disease via surrogates of endothelial function
    • Sonka M., Liang W., Lauer R.M. Automated analysis of brachial ultrasound image sequences: early detection of cardiovascular disease via surrogates of endothelial function. IEEE Trans. Med. Imaging 2002, 21:1271-1279.
    • (2002) IEEE Trans. Med. Imaging , vol.21 , pp. 1271-1279
    • Sonka, M.1    Liang, W.2    Lauer, R.M.3
  • 43
    • 69249113063 scopus 로고    scopus 로고
    • Impact of age, sex, and exercise on brachial artery flow-mediated dilatation
    • Black M.A., Cable N.T., Thijssen D.H., et al. Impact of age, sex, and exercise on brachial artery flow-mediated dilatation. Am. J. Physiol. Heart Circ. Physiol. 2009, 297:H1109-H1116.
    • (2009) Am. J. Physiol. Heart Circ. Physiol. , vol.297 , pp. H1109-H1116
    • Black, M.A.1    Cable, N.T.2    Thijssen, D.H.3
  • 44
    • 84937919957 scopus 로고    scopus 로고
    • Arterial structure and function in vascular ageing: are you as old as your arteries?
    • Thijssen D.H., Carter S.E., Green D.J. Arterial structure and function in vascular ageing: are you as old as your arteries?. J. Physiol. 2015, 10.1113/JP270597.
    • (2015) J. Physiol.
    • Thijssen, D.H.1    Carter, S.E.2    Green, D.J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.