Novel X-ray imaging technology enables significant patient dose reduction in interventional cardiology while maintaining diagnostic image quality

Author keywords

coronary angiography; coronary artery disease; imaging; radiation dose

Indexed keywords

CONTRAST MEDIUM;

EID: 84940454757     PISSN: 15221946     EISSN: 1522726X     Source Type: Journal    
DOI: 10.1002/ccd.25913     Document Type: Article
Times cited : (39)

References (33)
  • 1
    • 84912139288 scopus 로고    scopus 로고
    • Cardiovascular disease in Europe 2014: Epidemiological update
    • Melanie Nichols NT, Scarborough P, et al., Cardiovascular disease in Europe 2014: Epidemiological update. Eur Heart J 2014; 2950-2959.
    • (2014) Eur Heart J , pp. 2950-2959
    • Melanie Nichols, N.T.1    Scarborough, P.2
  • 2
    • 67649321444 scopus 로고    scopus 로고
    • Rotational coronary angiography
    • Klein AJ, Garcia JA,. Rotational coronary angiography. Cardiol Clin 2009; 27: 395-405.
    • (2009) Cardiol Clin , vol.27 , pp. 395-405
    • Klein, A.J.1    Garcia, J.A.2
  • 3
    • 77956381170 scopus 로고    scopus 로고
    • Comparison of rotational with conventional coronary angiography
    • Empen K, Kuon E, Hummel A, et al., Comparison of rotational with conventional coronary angiography. Am Heart J 2010; 160: 552-563.
    • (2010) Am Heart J , vol.160 , pp. 552-563
    • Empen, K.1    Kuon, E.2    Hummel, A.3
  • 4
    • 84944924712 scopus 로고    scopus 로고
    • Report of the United Nations scientific committee on the effects of atomic radiation
    • Annex A: Medical radiation exposure. United Nations Publications
    • UNSCEAR. Report of the United Nations scientific committee on the effects of atomic radiaion. Unscear 2008 report Vol I: Sources of ionizing radiation. Annex A: Medical radiation exposure. United Nations Publications. 2010;
    • (2010) Unscear 2008 Report Vol I: Sources of Ionizing Radiation
    • UNSCEAR1
  • 5
    • 84874999824 scopus 로고    scopus 로고
    • ICRP PUBLICATION 120: Radiological protection in cardiology
    • International Commission on Radiological P.:.
    • Cousins C, Miller DL, Bernardi G, et al,; and International Commission on Radiological P. ICRP PUBLICATION 120: Radiological protection in cardiology. Annals of the ICRP 2013; 42: 1-125. doi: 10.1016/j.icrp.2012.09.001.
    • (2013) Annals of the ICRP , vol.42 , pp. 1-125
    • Cousins, C.1    Miller, D.L.2    Bernardi, G.3    Al, E.4
  • 6
    • 69349096036 scopus 로고    scopus 로고
    • Exposure to low-dose ionizing radiation from medical imaging procedures
    • Fazel R, Krumholz HM, Wang Y, et al., Exposure to low-dose ionizing radiation from medical imaging procedures. New Engl J Med 2009; 361: 849-857.
    • (2009) New Engl J Med , vol.361 , pp. 849-857
    • Fazel, R.1    Krumholz, H.M.2    Wang, Y.3
  • 7
    • 84884232650 scopus 로고    scopus 로고
    • Radiation-induced noncancer risks in interventional cardiology: Optimisation of procedures and staff and patient dose reduction
    • Sun Z, AbAziz A, Yusof AK,. Radiation-induced noncancer risks in interventional cardiology: Optimisation of procedures and staff and patient dose reduction. BioMed Res Int 2013; 2013: 976-962.
    • (2013) BioMed Res Int , vol.2013 , pp. 976-962
    • Sun, Z.1    Abaziz, A.2    Yusof, A.K.3
  • 8
    • 84875061344 scopus 로고    scopus 로고
    • Evaluation of fluoroscopic cases qualifying as potential fluoroscopic sentinel events
    • Weinberg BD, Vance A, Arbique GM, et al., Evaluation of fluoroscopic cases qualifying as potential fluoroscopic sentinel events. Acad Radiol 2013; 20: 457-462.
    • (2013) Acad Radiol , vol.20 , pp. 457-462
    • Weinberg, B.D.1    Vance, A.2    Arbique, G.M.3
  • 9
    • 84899417230 scopus 로고    scopus 로고
    • Patient exposure to X-rays during coronary angiography and percutaneous transluminal coronary intervention: Results of a multicenter national survey
    • Georges JL, Belle L, Ricard C, et al., Patient exposure to X-rays during coronary angiography and percutaneous transluminal coronary intervention: Results of a multicenter national survey. Catheter Cardiovasc Interv 2014; 83: 729-738.
    • (2014) Catheter Cardiovasc Interv , vol.83 , pp. 729-738
    • Georges, J.L.1    Belle, L.2    Ricard, C.3
  • 10
    • 84865310261 scopus 로고    scopus 로고
    • Radiation dose reduction in the invasive cardiovascular laboratory: Implementing a culture and philosophy of radiation safety
    • Fetterly KA, Mathew V, Lennon R, et al., Radiation dose reduction in the invasive cardiovascular laboratory: Implementing a culture and philosophy of radiation safety. JACC Cardiovasc Interv 2012; 5: 866-873.
    • (2012) JACC Cardiovasc Interv , vol.5 , pp. 866-873
    • Fetterly, K.A.1    Mathew, V.2    Lennon, R.3
  • 11
    • 84901375738 scopus 로고    scopus 로고
    • Radiation dose reduction in the cardiac catheterization laboratory utilizing a novel protocol
    • Wassef AW, Hiebert B, Ravandi A, et al., Radiation dose reduction in the cardiac catheterization laboratory utilizing a novel protocol. JACC Cardiovasc Interv 2014; 7: 550-557.
    • (2014) JACC Cardiovasc Interv , vol.7 , pp. 550-557
    • Wassef, A.W.1    Hiebert, B.2    Ravandi, A.3
  • 12
    • 84885019840 scopus 로고    scopus 로고
    • Three-dimensional rotational X-ray acquisition technique is reducing patients' cancer risk in coronary angiography
    • Eloot L, Bacher K, Steenbeke F, et al., Three-dimensional rotational X-ray acquisition technique is reducing patients' cancer risk in coronary angiography. Catheter Cardiovasc Interv 2013; 82: E419-E427.
    • (2013) Catheter Cardiovasc Interv , vol.82 , pp. E419-E427
    • Eloot, L.1    Bacher, K.2    Steenbeke, F.3
  • 13
    • 84886651043 scopus 로고    scopus 로고
    • Image noise reduction algorithm for digital subtraction angiography: Clinical results
    • Soderman M, Holmin S, Andersson T, et al., Image noise reduction algorithm for digital subtraction angiography: Clinical results. Radiology 2013; 269: 553-560.
    • (2013) Radiology , vol.269 , pp. 553-560
    • Soderman, M.1    Holmin, S.2    Andersson, T.3
  • 14
    • 84888253334 scopus 로고    scopus 로고
    • Radiation dose in neuroangiography using image noise reduction technology: A population study based on 614 patients
    • Soderman M, Mauti M, Boon S, et al., Radiation dose in neuroangiography using image noise reduction technology: A population study based on 614 patients. Neuroradiology 2013; 55: 1365-1372.
    • (2013) Neuroradiology , vol.55 , pp. 1365-1372
    • Soderman, M.1    Mauti, M.2    Boon, S.3
  • 15
    • 84887018430 scopus 로고    scopus 로고
    • New image processing and noise reduction technology allows reduction of radiation exposure in complex electrophysiologic interventions while maintaining optimal image quality: A randomized clinical trial
    • Dekker LR, van der Voort PH, Simmers TA, et al., New image processing and noise reduction technology allows reduction of radiation exposure in complex electrophysiologic interventions while maintaining optimal image quality: A randomized clinical trial. Heart Rhythm 2013; 10: 1678-1682.
    • (2013) Heart Rhythm , vol.10 , pp. 1678-1682
    • Dekker, L.R.1    Van Der Voort, P.H.2    Simmers, T.A.3
  • 16
    • 11244304718 scopus 로고    scopus 로고
    • Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization
    • Bacher K, Bogaert E, Lapere R, et al., Patient-specific dose and radiation risk estimation in pediatric cardiac catheterization. Circulation 2005; 111: 83-89.
    • (2005) Circulation , vol.111 , pp. 83-89
    • Bacher, K.1    Bogaert, E.2    Lapere, R.3
  • 17
    • 77953355570 scopus 로고    scopus 로고
    • VIEWDEX: An efficient and easy-to-use software for observer performance studies
    • Hakansson M, Svensson S, Zachrisson S, et al., VIEWDEX: An efficient and easy-to-use software for observer performance studies. Radiat Prot Dosimetry 2010; 139: 42-51.
    • (2010) Radiat Prot Dosimetry , vol.139 , pp. 42-51
    • Hakansson, M.1    Svensson, S.2    Zachrisson, S.3
  • 18
    • 84865399491 scopus 로고    scopus 로고
    • Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems: Part II
    • Sanchez R, Vano E, Ubeda C, et al., Influence of image metrics when assessing image quality from a test object in cardiac X-ray systems: Part II. J Digit Imaging 2012; 25: 537-541.
    • (2012) J Digit Imaging , vol.25 , pp. 537-541
    • Sanchez, R.1    Vano, E.2    Ubeda, C.3
  • 19
    • 84455167783 scopus 로고    scopus 로고
    • Correlation of Contrast-detail analysis and clinical image quality assessment in chest radiography with a human cadaver study
    • De Crop A, Bacher K, Van Hoof T, et al., Correlation of Contrast-detail analysis and clinical image quality assessment in chest radiography with a human cadaver study. Radiology 2011; 262: 298-304.
    • Radiology 2011; , vol.262 , pp. 298-304
    • De Crop, A.1    Bacher, K.2    Van Hoof, T.3
  • 20
    • 51849124980 scopus 로고    scopus 로고
    • An investigation of flat panel equipment variables on image quality with a dedicated cardiac phantom
    • Dragusin O, Bosmans H, Pappas C, et al., An investigation of flat panel equipment variables on image quality with a dedicated cardiac phantom. Phys Med Biol 2008; 53: 4927-4940.
    • (2008) Phys Med Biol , vol.53 , pp. 4927-4940
    • Dragusin, O.1    Bosmans, H.2    Pappas, C.3
  • 21
    • 78650040512 scopus 로고    scopus 로고
    • Paediatric interventional cardiology: Flat detector versus image intensifier using a test object
    • Vano E, Ubeda C, Martinez LC, et al., Paediatric interventional cardiology: Flat detector versus image intensifier using a test object. Phys Med Biol 2010; 55: 7287-7297.
    • (2010) Phys Med Biol , vol.55 , pp. 7287-7297
    • Vano, E.1    Ubeda, C.2    Martinez, L.C.3
  • 22
    • 0035152717 scopus 로고    scopus 로고
    • Cardiac catheterization laboratory imaging quality assurance program
    • Wondrow MA, Laskey WK, Hildner FJ, et al., Cardiac catheterization laboratory imaging quality assurance program. Catheter Cardiovasc Interv 2001; 52: 59-66.
    • (2001) Catheter Cardiovasc Interv , vol.52 , pp. 59-66
    • Wondrow, M.A.1    Laskey, W.K.2    Hildner, F.J.3
  • 23
    • 14344270377 scopus 로고    scopus 로고
    • A method based on DIMOND quality criteria to evaluate imaging in diagnostic and interventional cardiology
    • Bernardi G, Padovani R, Morocutti G, et al., A method based on DIMOND quality criteria to evaluate imaging in diagnostic and interventional cardiology. Radiat Prot Dosimetry 2001; 94: 167-172.
    • (2001) Radiat Prot Dosimetry , vol.94 , pp. 167-172
    • Bernardi, G.1    Padovani, R.2    Morocutti, G.3
  • 24
    • 55249114896 scopus 로고    scopus 로고
    • Quality criteria for cardiac images: An update
    • Bernardi G, Bar O, Jezewski T, et al., Quality criteria for cardiac images: An update. Radiat Prot Dosimetry 2008; 129: 87-90.
    • (2008) Radiat Prot Dosimetry , vol.129 , pp. 87-90
    • Bernardi, G.1    Bar, O.2    Jezewski, T.3
  • 25
    • 54949086053 scopus 로고    scopus 로고
    • Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images
    • Richard S, Siewerdsen JH,. Comparison of model and human observer performance for detection and discrimination tasks using dual-energy x-ray images. Med Phys 2008; 35: 5043-5053.
    • (2008) Med Phys , vol.35 , pp. 5043-5053
    • Richard, S.1    Siewerdsen, J.H.2
  • 26
    • 45149087420 scopus 로고    scopus 로고
    • A large-scale multicentre study in Belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipment
    • Bogaert E, Bacher K, Thierens H,. A large-scale multicentre study in belgium of dose area product values and effective doses in interventional cardiology using contemporary X-ray equipment. Radiat Prot Dosimetry 2008; 128: 312-323.
    • (2008) Radiat Prot Dosimetry , vol.128 , pp. 312-323
    • Bogaert, E.1    Bacher, K.2    Thierens, H.3
  • 27
    • 55249102686 scopus 로고    scopus 로고
    • Patient dose in interventional radiology: A European survey
    • Vano E, Jarvinen H, Kosunen A, et al., Patient dose in interventional radiology: A European survey. Radiat Prot Dosimetry 2008; 129: 39-45.
    • (2008) Radiat Prot Dosimetry , vol.129 , pp. 39-45
    • Vano, E.1    Jarvinen, H.2    Kosunen, A.3
  • 28
    • 55249092932 scopus 로고    scopus 로고
    • Reference levels at european level for cardiac interventional procedures
    • Padovani R, Vano E, Trianni A, et al., Reference levels at european level for cardiac interventional procedures. Radiat Prot Dosimetry 2008; 129: 104-107.
    • (2008) Radiat Prot Dosimetry , vol.129 , pp. 104-107
    • Padovani, R.1    Vano, E.2    Trianni, A.3
  • 29
    • 4644296835 scopus 로고    scopus 로고
    • Identification of less-irradiating tube angulations in invasive cardiology
    • Kuon E, Dahm JB, Empen K, et al., Identification of less-irradiating tube angulations in invasive cardiology. J Am Coll Cardiol 2004; 44: 1420-1428.
    • (2004) J Am Coll Cardiol , vol.44 , pp. 1420-1428
    • Kuon, E.1    Dahm, J.B.2    Empen, K.3
  • 30
    • 84866120491 scopus 로고    scopus 로고
    • Evaluation of coronary angiographic projections to balance the clinical yield with the radiation risk
    • Smith IR, Cameron J, Mengersen KL, et al., Evaluation of coronary angiographic projections to balance the clinical yield with the radiation risk. Br J Radiol 2012; 85: e722-e728.
    • (2012) Br J Radiol , vol.85 , pp. e722-e728
    • Smith, I.R.1    Cameron, J.2    Mengersen, K.L.3
  • 31
    • 84901367472 scopus 로고    scopus 로고
    • Relationship of beam angulation and radiation exposure in the cardiac catheterization laboratory
    • Agarwal S, Parashar A, Bajaj NS, et al., Relationship of beam angulation and radiation exposure in the cardiac catheterization laboratory. JACC Cardiovasc Interv 2014; 7: 558-566.
    • (2014) JACC Cardiovasc Interv , vol.7 , pp. 558-566
    • Agarwal, S.1    Parashar, A.2    Bajaj, N.S.3
  • 32
    • 65349092929 scopus 로고    scopus 로고
    • A large-scale multicentre study of patient skin doses in interventional cardiology: Dose-area product action levels and dose reference levels
    • Bogaert E, Bacher K, Lemmens K, et al., A large-scale multicentre study of patient skin doses in interventional cardiology: Dose-area product action levels and dose reference levels. Br J Radiol 2009; 82: 303-312.
    • (2009) Br J Radiol , vol.82 , pp. 303-312
    • Bogaert, E.1    Bacher, K.2    Lemmens, K.3
  • 33
    • 10744232211 scopus 로고    scopus 로고
    • Preliminary reference levels in interventional cardiology
    • Neofotistou V, Vano E, Padovani R, et al., Preliminary reference levels in interventional cardiology. Eur Radiol 2003; 13: 2259-2263.
    • (2003) Eur Radiol , vol.13 , pp. 2259-2263
    • Neofotistou, V.1    Vano, E.2    Padovani, R.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.