Volumn 173, Issue 1, 2014, Pages 69-73

Delay and stability of central venous administration of norepinephrine in children: A bench study

Author keywords

Central venous catheter; Children; Dead volume; Delivery delay; Infusion stability; Norepinephrine

Indexed keywords

NORADRENALIN;

EID: 84893681868     PISSN: 03406199     EISSN: 14321076     Source Type: Journal    
DOI: 10.1007/s00431-013-2121-5     Document Type: Article
Times cited : (4)

References (14)
  • 1
    • 70349561988 scopus 로고    scopus 로고
    • An analysis of drug delivery dynamics via a pediatric central venous infusion system: Quantification of delays in achieving intended doses
    • 1:CAS:528:DC%2BD1MXhtFGnsLnE 19762743 10.1213/ane.0b013e3181b220c9
    • Bartels K, Moss DR, Peterfreund RA (2009) An analysis of drug delivery dynamics via a pediatric central venous infusion system: quantification of delays in achieving intended doses. Anesth Analg 109(4):1156-1161
    • (2009) Anesth Analg , vol.109 , Issue.4 , pp. 1156-1161
    • Bartels, K.1    Moss, D.R.2    Peterfreund, R.A.3
  • 2
    • 0029131840 scopus 로고
    • Fluctuations in syringe-pump infusions: Association with blood pressure variations in infants
    • 1:STN:280:DyaK28%2FhsVahuw%3D%3D 7583827
    • Capes DF, Dunster KR, Sunderland VB et al (1995) Fluctuations in syringe-pump infusions: association with blood pressure variations in infants. Am J Health Syst Pharm 52(15):1646-1653
    • (1995) Am J Health Syst Pharm , vol.52 , Issue.15 , pp. 1646-1653
    • Capes, D.F.1    Dunster, K.R.2    Sunderland, V.B.3
  • 3
    • 0033751074 scopus 로고    scopus 로고
    • Syringe pump assemblies and the natural history of clinical technology
    • 1:STN:280:DC%2BD3cvnvFKkuw%3D%3D 11032265 10.1007/BF03024861
    • Cook RI (2000) Syringe pump assemblies and the natural history of clinical technology. Can J Anaesth 47(10):929-935
    • (2000) Can J Anaesth , vol.47 , Issue.10 , pp. 929-935
    • Cook, R.I.1
  • 4
    • 70349581202 scopus 로고    scopus 로고
    • Impact of multiaccess infusion devices on in vitro drug delivery during multi-infusion therapy
    • 19762742 10.1213/ane.0b013e3181ae06e3
    • Décaudin B, Dewulf S, Lannoy D et al (2009) Impact of multiaccess infusion devices on in vitro drug delivery during multi-infusion therapy. Anesth Analg 109(4):1147-1155
    • (2009) Anesth Analg , vol.109 , Issue.4 , pp. 1147-1155
    • Décaudin, B.1    Dewulf, S.2    Lannoy, D.3
  • 5
    • 0023710746 scopus 로고
    • Determination of catecholamines in plasma by high-performance liquid chromatography
    • 1:STN:280:DyaL1czivVSrsQ%3D%3D 3416446
    • Guillemin A, Troupel S, Galli A (1988) Determination of catecholamines in plasma by high-performance liquid chromatography. Clin Chem 34(9):1913-1914
    • (1988) Clin Chem , vol.34 , Issue.9 , pp. 1913-1914
    • Guillemin, A.1    Troupel, S.2    Galli, A.3
  • 6
    • 84864115772 scopus 로고    scopus 로고
    • Ficts and facts of epinephrine and norepinephrine stability in injectable solutions
    • 1:CAS:528:DC%2BC38XnvFGgsrk%3D 22613065 10.1016/j.ijpharm.2012.05.017
    • Hoellein L, Holzgrabe U (2012) Ficts and facts of epinephrine and norepinephrine stability in injectable solutions. Int J Pharm 434(1-2):468-480
    • (2012) Int J Pharm , vol.434 , Issue.1-2 , pp. 468-480
    • Hoellein, L.1    Holzgrabe, U.2
  • 7
    • 0022492405 scopus 로고
    • Factors affecting delivery of drug through extension tubing
    • 1:STN:280:DyaL2s%2Fis1eqsg%3D%3D 3768227 10.1093/bja/58.10.1141
    • Hutton P, Thornberry EA (1986) Factors affecting delivery of drug through extension tubing. Br J Anaesth 58:1141-1148
    • (1986) Br J Anaesth , vol.58 , pp. 1141-1148
    • Hutton, P.1    Thornberry, E.A.2
  • 8
    • 15544389982 scopus 로고    scopus 로고
    • The impact of carrier flow rate and infusion set dead-volume on the dynamics of intravenous drug delivery
    • 1:CAS:528:DC%2BD2MXisVGrurw%3D 15781520 10.1213/01.ANE.0000146942.51020. 88
    • Lovich MA, Doles J, Peterfreund RA (2005) The impact of carrier flow rate and infusion set dead-volume on the dynamics of intravenous drug delivery. Anesth Analg 100(4):1048-1055
    • (2005) Anesth Analg , vol.100 , Issue.4 , pp. 1048-1055
    • Lovich, M.A.1    Doles, J.2    Peterfreund, R.A.3
  • 9
    • 33645504101 scopus 로고    scopus 로고
    • The delivery of drugs to patients by continuous intravenous infusion: Modeling predicts potential dose fluctuations depending on flow rates and infusion system dead volume
    • 1:CAS:528:DC%2BD28XislWktLo%3D 16551914 10.1213/01.ane.0000198670.02481. 6b
    • Lovich MA, Kinnealley ME, Sims NM et al (2006) The delivery of drugs to patients by continuous intravenous infusion: modeling predicts potential dose fluctuations depending on flow rates and infusion system dead volume. Anesth Analg 102(4):1147-1153
    • (2006) Anesth Analg , vol.102 , Issue.4 , pp. 1147-1153
    • Lovich, M.A.1    Kinnealley, M.E.2    Sims, N.M.3
  • 10
    • 36448996189 scopus 로고    scopus 로고
    • Central venous catheter infusions: A laboratory model shows large differences in drug delivery dynamics related to catheter dead volume
    • 1:CAS:528:DC%2BD2sXhsVGhsrvO 18074479 10.1097/01.CCM.0000295587.04882.4F
    • Lovich MA, Peterfreund GL, Sims NM et al (2007) Central venous catheter infusions: a laboratory model shows large differences in drug delivery dynamics related to catheter dead volume. Crit Care Med 35(12):2792-2798
    • (2007) Crit Care Med , vol.35 , Issue.12 , pp. 2792-2798
    • Lovich, M.A.1    Peterfreund, G.L.2    Sims, N.M.3
  • 11
    • 70350719515 scopus 로고    scopus 로고
    • An in vitro analysis of central venous drug delivery by continuous infusion: The effect of manifold design and port selection
    • 1:CAS:528:DC%2BD1MXht1OqsrjJ 19843791 10.1213/ANE.0b013e3181b7c359
    • Moss DR, Bartels K, Peterfreund GL et al (2009) An in vitro analysis of central venous drug delivery by continuous infusion: the effect of manifold design and port selection. Anesth Analg 109(5):1524-1529
    • (2009) Anesth Analg , vol.109 , Issue.5 , pp. 1524-1529
    • Moss, D.R.1    Bartels, K.2    Peterfreund, G.L.3
  • 12
    • 0035723595 scopus 로고    scopus 로고
    • Start-up delays of infusion syringe pumps
    • 1:STN:280:DC%2BD3Mnkt1WitQ%3D%3D 11696120 10.1046/j.1460-9592.2001.00730. x
    • Neff T, Fischer J, Fehr S et al (2001) Start-up delays of infusion syringe pumps. Paediatr Anaesth 11:561-565
    • (2001) Paediatr Anaesth , vol.11 , pp. 561-565
    • Neff, T.1    Fischer, J.2    Fehr, S.3
  • 13
    • 85047699866 scopus 로고    scopus 로고
    • Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps
    • 1:STN:280:DC%2BD2svgtV2lsw%3D%3D 17261217 10.1017/S0265021507000026
    • Neff SB, Neff TA, Gerber S et al (2007) Flow rate, syringe size and architecture are critical to start-up performance of syringe pumps. Eur J Anaesthesiol 24:602-608
    • (2007) Eur J Anaesthesiol , vol.24 , pp. 602-608
    • Neff, S.B.1    Neff, T.A.2    Gerber, S.3
  • 14
    • 0033936276 scopus 로고    scopus 로고
    • Influence of infusion line compliance on drug delivery rate during acute line loop formation
    • 1:STN:280:DC%2BD3M%2FlvFajsA%3D%3D 10945397 10.1007/s001340051246
    • Weiss M, Bänziger O, Neff T et al (2000) Influence of infusion line compliance on drug delivery rate during acute line loop formation. Intensive Care Med 26:776-779
    • (2000) Intensive Care Med , vol.26 , pp. 776-779
    • Weiss, M.1    Bänziger, O.2    Neff, T.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.