Volumn 38, Issue 10, 2011, Pages 5432-5440

In vivo dosimetry with optically stimulated luminescent dosimeters, OSLDs, compared to diodes; The effects of buildup cap thickness and fabrication material

Author keywords

buildup caps; diodes; in vivo dosimetry; optically stimulated luminescent dosimeters; OSLD

Indexed keywords

DIODES; DOSIMETERS; DOSIMETRY; LUMINESCENCE; NANODOTS; NANOSTRUCTURED MATERIALS; PHANTOMS; SHEET METAL; SURFACE MEASUREMENT; WALLS (STRUCTURAL PARTITIONS); X RAYS;

EID: 80053583970     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3633939     Document Type: Article
Times cited : (28)

References (50)
  • 1
    • 0025166575 scopus 로고
    • Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure
    • G. Leunens, J. Van Dam, A. Dutreix, and E. Van Der Shueren, Quality assurance in radiotherapy by in vivo dosimetry. 1. Entrance dose measurements, a reliable procedure., Radiother. Oncol. 17, 141-151 (1990). 10.1016/0167- 8140(90)90102-3 (Pubitemid 20062409)
    • (1990) Radiotherapy and Oncology , vol.17 , Issue.2 , pp. 141-151
    • Leunens, G.1    Van Dam, J.2    Dutreix, A.3    Van Der Schueren, E.4
  • 2
    • 0026615792 scopus 로고
    • Assesment of dose inhomogeneity at target level by in vivo dosimetry: Can the recommended 5% accuracy in the dose delivered to the target volume be fulfilled in daily practice?
    • DOI 10.1016/0167-8140(92)90243-N
    • G. Leunens, J. Verstraete, A. Dutreix, and E. van der Schueren, Assessment of dose inhomogeneity at target level by in vivo dosimetry: Can the recommended 5 accuracy in the dose delivered to the target volume be fulfilled in daily practice. ? Radiother. Oncol. 25, 242-250 (1992). 10.1016/0167-8140(92) 90243-N (Pubitemid 23000766)
    • (1992) Radiotherapy and Oncology , vol.25 , Issue.4 , pp. 242-250
    • Leunens, G.1    Verstraete, J.2    Dutreix, A.3    Van Der Schueren, E.4
  • 3
    • 0032843083 scopus 로고    scopus 로고
    • In vivo dosimetry during conformal radiotherapy. Requirements for and findings of a routine procedure
    • DOI 10.1016/S0167-8140(99)00074-2, PII S0167814099000742
    • J. H. Lanson, M. Essers, G. J. Meijer, A. W. H. Minken, G. J. Uiterwaal, and B. J. Mijnheer, In vivo dosimetry during conformal radiotherapy: Requirements for and findings of a routine procedure., Radiother. Oncol. 52, 51-59 (1999). 10.1016/S0167-8140(99)00074-2 (Pubitemid 29461096)
    • (1999) Radiotherapy and Oncology , vol.52 , Issue.1 , pp. 51-59
    • Lanson, J.H.1    Essers, M.2    Meijer, G.J.3    Minken, A.W.H.4    Uiterwaal, G.J.5    Mijnheer, B.J.6
  • 4
    • 0033556681 scopus 로고    scopus 로고
    • In vivo dosimetry during external photon beam radiotherapy
    • DOI 10.1016/S0360-3016(98)00341-1, PII S0360301698003411
    • M. Essers and B. J. Mijnheer, In vivo dosimetry during external photon beamradiotherapy., Int. J. Radiat. Oncol. Biol. Phys. 43, 245-249 (1999). 10.1016/S0360-3016(98)00341-1 (Pubitemid 29091462)
    • (1999) International Journal of Radiation Oncology Biology Physics , vol.43 , Issue.2 , pp. 245-259
    • Essers, M.1    Mijnheer, B.J.2
  • 5
    • 0034858227 scopus 로고    scopus 로고
    • Implementation of an in vivo diode dosimetry program and changes in diode characteristics over a 4-year clinical history
    • DOI 10.1118/1.1388217
    • P. A. Jursinic, Implementation of an in vivo diode dosimetry program and changes in diode characteristics over a four-year clinical history., Med. Phys. 28, 1718-1726 (2001). 10.1118/1.1388217 (Pubitemid 32783216)
    • (2001) Medical Physics , vol.28 , Issue.8 , pp. 1718-1726
    • Jursinic, P.A.1
  • 11
    • 0028213977 scopus 로고
    • Evaluation of a dual bias dual metal oxide-silicon semiconductor field effect transistor detector as radiation dosimeter
    • DOI 10.1118/1.597314
    • M. Soubra, J. Cygler, and G. Mackay, Evaluation of a dual bias dual metal-oxide-silicon semiconductor field effect transistor detector as radiation detector., Med. Phys. 21, 567-572 (1994). 10.1118/1.597314 (Pubitemid 24129936)
    • (1994) Medical Physics , vol.21 , Issue.4 , pp. 567-572
    • Soubra, M.1    Cygler, J.2    Mackay, G.3
  • 12
    • 0021043262 scopus 로고
    • Effects of radiation damage on p-type silicon detectors
    • G. Rikner and E. Grusell, Effect of radiation damage on p-type silicon detectors., Phys. Med. Biol. 28, 1261-1267 (1983). 10.1088/0031-9155/28/11/006 (Pubitemid 14244043)
    • (1983) Physics in Medicine and Biology , vol.28 , Issue.11 , pp. 1261-1267
    • Rikner, G.1    Grusell, E.2
  • 13
    • 0023620571 scopus 로고
    • Patient dose measurements in photon fields by means of silicon semiconductor detectors
    • DOI 10.1118/1.596015
    • G. Rikner and E. Grusell, Patient dose measurements in photon fields by means of silicon semiconductor detectors., Med. Phys. 14, 870-873 (1987). 10.1118/1.596015 (Pubitemid 18021203)
    • (1987) Medical Physics , vol.14 , Issue.5 , pp. 870-873
    • Rikner, G.1    Grusell, E.2
  • 14
    • 0030292867 scopus 로고    scopus 로고
    • Entrance and exit dose measurements with semiconductors and thermoluminescent dosemeters: A comparison of methods and in vivo results
    • DOI 10.1016/S0167-8140(96)01826-9, PII S0167814096018269
    • T. Loncol, J. L. Greffe, S. Vynckier, and V. Scalliet, Entrance and exit dose measurements with semiconductors and thermoluminescent dosemeters: A comparison of methods and in vivo results., Radiother. Oncol. 41, 179-187 (1996). 10.1016/S0167-8140(96)01826-9 (Pubitemid 26423630)
    • (1996) Radiotherapy and Oncology , vol.41 , Issue.2 , pp. 179-187
    • Loncol, Th.1    Greffe, J.L.2    Vynckier, S.3    Scalliet, P.4
  • 19
    • 36749049838 scopus 로고    scopus 로고
    • Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements
    • DOI 10.1118/1.2804555
    • P. Jursinic, Characterization of optically stimulated luminescent dosimeters, OSLDs, for clinical dosimetric measurements., Med. Phys. 34, 4594-4604 (2007). 10.1118/1.2804555 (Pubitemid 350199438)
    • (2007) Medical Physics , vol.34 , Issue.12 , pp. 4594-4604
    • Jursinic, P.A.1
  • 20
    • 34548794727 scopus 로고    scopus 로고
    • 3:C optically stimulated luminescence dosimeter element at radiation oncology and nuclear accident dose levels
    • 10.1093/rpd/ncl500
    • 3:C optically stimulated luminescence dosimeter element at radiation oncology and nuclear accident dose levels., Radiat. Prot. Dosim. 123, 435-442 (2007). 10.1093/rpd/ncl500
    • (2007) Radiat. Prot. Dosim. , vol.123 , pp. 435-442
    • Miller, S.D.1    Murphy, M.K.2
  • 21
    • 34447299153 scopus 로고    scopus 로고
    • Optically stimulated luminescence (OSL) of carbon-doped aluminum oxide (Al2 O3:C) for film dosimetry in radiotherapy
    • DOI 10.1118/1.2737160
    • 3:C) for film dosimetry radiotherapy., Med. Phys. 34, 2113-2118 (2007). 10.1118/1.2737160 (Pubitemid 47055746)
    • (2007) Medical Physics , vol.34 , Issue.6 , pp. 2113-2118
    • Schembri, V.1    Heijmen, B.J.M.2
  • 22
    • 58149291758 scopus 로고    scopus 로고
    • Optically stimulated luminescence (OSL) dosimetry in medicine
    • 10.1088/0031-9155/53/20/R01
    • E. G. Yukihara and S. W. S. McKeever, Optically stimulated luminescence (OSL) dosimetry in medicine., Phys. Med. Biol. 53, R351-R379 (2008). 10.1088/0031-9155/53/20/R01
    • (2008) Phys. Med. Biol. , vol.53
    • Yukihara, E.G.1    McKeever, S.W.S.2
  • 23
    • 37549059594 scopus 로고    scopus 로고
    • Evaluation of Al2O3:C optically stimulated luminescence (OSL) dosimeters for passive dosimetry of high-energy photon and electron beams in radiotherapy
    • 10.1118/1.2816106
    • E. G. Yukihara, G. Mardirossian, M. Mirzasadeghi, S. Guduru, and S. Ahmad, Evaluation of Al2O3:C optically stimulated luminescence (OSL) dosimeters for passive dosimetry of high-energy photon and electron beams in radiotherapy., Med. Phys. 35, 260-269 (2008). 10.1118/1.2816106
    • (2008) Med. Phys. , vol.35 , pp. 260-269
    • Yukihara, E.G.1    Mardirossian, G.2    Mirzasadeghi, M.3    Guduru, S.4    Ahmad, S.5
  • 24
    • 65449186980 scopus 로고    scopus 로고
    • The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovoltage photon, megavoltage photon, and electron, proton, and carbon beams
    • 10.1118/1.3097283
    • C. S. Reft, The energy dependence and dose response of a commercial optically stimulated luminescent detector for kilovoltage photon, megavoltage photon, and electron, proton, and carbon beams., Med. Phys. 36, 1690-1699 (2009). 10.1118/1.3097283
    • (2009) Med. Phys. , vol.36 , pp. 1690-1699
    • Reft, C.S.1
  • 25
    • 41449098750 scopus 로고    scopus 로고
    • Radiotherapy dosimetry using a commercial OSL system
    • DOI 10.1118/1.2841940
    • A. Viamonte, L. A. R. da Rosa, L. A. Buckley, A. Cherpak, and J. E. Cygler, Radiotherapy dosimetry using a commercial OSL system., Med. Phys. 35, 1261-1266 (2008). 10.1118/1.2841940 (Pubitemid 351457491)
    • (2008) Medical Physics , vol.35 , Issue.4 , pp. 1261-1266
    • Viamonte, A.1    Da Rosa, L.A.R.2    Buckley, L.A.3    Cherpak, A.4    Cygler, J.E.5
  • 26
    • 77955308531 scopus 로고    scopus 로고
    • Optically stimulated luminescence (OSL) dosimetry in radiotherapy
    • in, edited by D. W. O. Rogers and J. Cygler (Medical Physics Publishing, Madison, WI)
    • J. E. Cygler and E. G. Yukihara, Optically stimulated luminescence (OSL) dosimetry in radiotherapy., in AAPM 2009 Summer School, Clinical Dosimetry Measurements in Radiotherapy, edited by, D. W. O. Rogers, and, J. Cygler, (Medical Physics Publishing, Madison, WI, 2009).
    • (2009) AAPM 2009 Summer School, Clinical Dosimetry Measurements in Radiotherapy
    • Cygler, J.E.1    Yukihara, E.G.2
  • 27
    • 84880249496 scopus 로고    scopus 로고
    • edited by P. Mayles, J. C. Rosenwald, and A. Nahum (CRC, Boca RatonTaylor and Francis, New York)
    • G. Marinello, Handbook of Radiotherapy Physics, edited by, P. Mayles, J. C. Rosenwald, and, A. Nahum, (CRC, Boca RatonTaylor and Francis, New York, 2007), pp. 303-320.
    • (2007) Handbook of Radiotherapy Physics , pp. 303-320
    • Marinello, G.1
  • 28
    • 0033010139 scopus 로고    scopus 로고
    • Build-up modification of commercial diodes for entrance dose measurements in 'higher energy' photon beams
    • DOI 10.1016/S0167-8140(99)00058-4, PII S0167814099000584
    • D. Georg, B. De Ost, M. T. Hoornaert, P. Pilette, J. Van Dam, M. Van Dycke, and D. Huyskens, Build-up modification of commercial diodes for entrance dose measurements in higher energy' photon beams., Radiother. Oncol. 51, 249-256 (1999). 10.1016/S0167-8140(99)00058-4 (Pubitemid 29305709)
    • (1999) Radiotherapy and Oncology , vol.51 , Issue.3 , pp. 249-256
    • Georg, D.1    De Ost, B.2    Hoornaert, M.-T.3    Pilette, P.4    Van Dam, J.5    Van Dycke, M.6    Huyskens, D.7
  • 29
    • 0029874870 scopus 로고    scopus 로고
    • Calibration of semiconductor detectors for dose assessment in total body irradiation
    • DOI 10.1016/0167-8140(95)01677-5
    • N. Jornet, M. Ribas, and T. Eudaldo, Calibration of semiconductor detectors for dose assessment in total body irradiation.. Radiother. Oncol. 38, 247-251 (1996). 10.1016/0167-8140(95)01677-5 (Pubitemid 26080182)
    • (1996) Radiotherapy and Oncology , vol.38 , Issue.3 , pp. 247-251
    • Jornet, N.1    Ribas, M.2    Eudaldo, T.3
  • 31
    • 33744782342 scopus 로고    scopus 로고
    • Measurement of head scatter factors of linear accelerators with columnar miniphantoms
    • DOI 10.1118/1.2201148
    • P.A. Jursinic, Measurement of head scatter factors of linear accelerators with columnar miniphantoms., Med. Phys. 33, 1720-1728 (2006). 10.1118/1.2201148 (Pubitemid 43830260)
    • (2006) Medical Physics , vol.33 , Issue.6 , pp. 1720-1728
    • Jursinic, P.A.1
  • 32
    • 0025727879 scopus 로고
    • The determination of phantom and collimator scatter components of the output of megavoltage photon beams: Measurements of the collimator scatter part with a beam-coxial narrow cylindrical phantom
    • 10.1016/0167-8140(91)90124-Y
    • J. J. M. van Gasteren, S. Heukelom, H. J. van Kleffens, R. van der Laarse, J. L. M. Venselaar, and C. F. Westermann, The determination of phantom and collimator scatter components of the output of megavoltage photon beams: Measurements of the collimator scatter part with a beam-coxial narrow cylindrical phantom., Radiother. Oncol. 20, 250-257 (1991). 10.1016/0167- 8140(91)90124-Y
    • (1991) Radiother. Oncol. , vol.20 , pp. 250-257
    • Van Gasteren, J.J.M.1    Heukelom, S.2    Van Kleffens, H.J.3    Van Der Laarse, R.4    Venselaar, J.L.M.5    Westermann, C.F.6
  • 33
    • 0027145156 scopus 로고
    • In vivo dosimetry during conformal therapy of prostatic cancer
    • DOI 10.1016/0167-8140(93)90258-A
    • M. Essers, J. H. Lanson, and B. J. Mijnheer, In vivo dosimetry during conformal therapy of prostatic cancer., Radiother. Oncol. 29, 271-279 (1993). 10.1016/0167-8140(93)90258-A (Pubitemid 24007870)
    • (1993) Radiotherapy and Oncology , vol.29 , Issue.2 , pp. 271-279
    • Essers, M.1    Lanson, J.H.2    Mijnheer, B.J.3
  • 34
    • 0027971376 scopus 로고
    • Dosimetric control of conformal treatment of parotid gland tumours
    • DOI 10.1016/0167-8140(94)90102-3
    • M. Essers, R. Keus, J. H. Lanson, and B. J. Mijnheer, Dosimetric control of conformal treatment of parotid gland tumors., Radiother. Oncol. 32, 154-162 (1994). 10.1016/0167-8140(94)90102-3 (Pubitemid 24270737)
    • (1994) Radiotherapy and Oncology , vol.32 , Issue.2 , pp. 154-162
    • Essers, M.1
  • 35
    • 0030332980 scopus 로고    scopus 로고
    • The use of diode dosimetry in quality improvement of patient care in radiation therapy
    • DOI 10.1016/S0958-3947(96)00081-7, PII S0958394796000817
    • D. P. Fontenla, R. Yaparpalvi, C. S. Chui, and E. Briot, The use of diode dosimeters in quality improvement of patient care in radiation therapy., Med. Dosim. 4, 235-241 (1996). 10.1016/S0958-3947(96)00081-7 (Pubitemid 126380772)
    • (1996) Medical Dosimetry , vol.21 , Issue.4 , pp. 235-241
    • Fontenla, D.P.1    Yaparpalvi, R.2    Chui, C.-S.3    Briot, E.4
  • 36
    • 0032874911 scopus 로고    scopus 로고
    • AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    • DOI 10.1118/1.598691
    • P. R. Almond, P. J. Biggs, B. M. Coursey, W. F. Hanson, M. S. Huq, R. Nath, and D. W. O. Rogers, AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams., Med. Phys. 26, 1847-1870 (1999). 10.1118/1.598691 (Pubitemid 29442268)
    • (1999) Medical Physics , vol.26 , Issue.9 , pp. 1847-1870
    • Almond, P.R.1    Biggs, P.J.2    Coursey, B.M.3    Hanson, W.F.4    Huq, M.S.5    Nath, R.6    Rogers, D.W.O.7
  • 39
    • 73649095518 scopus 로고    scopus 로고
    • Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose
    • 10.1118/1.3267489
    • P. Jursinic, Changes in optically stimulated luminescent dosimeter (OSLD) dosimetric characteristics with accumulated dose., Med. Phys. 37, 132-140 (2010). 10.1118/1.3267489
    • (2010) Med. Phys. , vol.37 , pp. 132-140
    • Jursinic, P.1
  • 40
    • 66449113635 scopus 로고    scopus 로고
    • Angular dependence of dose sensitivity of surface diodes
    • 10.1118/1.3125644
    • P. Jursinic, Angular dependence of dose sensitivity of surface diodes., Med. Phys. 36, 2165-2171 (2009). 10.1118/1.3125644
    • (2009) Med. Phys. , vol.36 , pp. 2165-2171
    • Jursinic, P.1
  • 41
    • 0018125662 scopus 로고
    • Tissue substitutes in experimental radiation physics
    • DOI 10.1118/1.594456
    • D. R. White, Tissue substitutes in experimental radiation physics., Med. Phys. 5, 467-479 (1978). 10.1118/1.594456 (Pubitemid 9067455)
    • (1978) Medical Physics , vol.5 , Issue.6 , pp. 467-479
    • White, D.R.1
  • 42
    • 0018176837 scopus 로고
    • Electron contamination of a high-energy X-ray beam
    • T. N. Padikal and J. A. Deye, Electron contamination of a high-energy x-ray beam., Phys. Med. Biol. 23, 1086-1092 (1978). 10.1088/0031-9155/23/6/004 (Pubitemid 9055646)
    • (1978) Physics in Medicine and Biology , vol.23 , Issue.6 , pp. 1086-1092
    • Padikal, T.N.1    Deye, J.A.2
  • 43
    • 0027285239 scopus 로고
    • On the cause of the variation in tissue-maximum ratio values with source- to-detector distance
    • DOI 10.1118/1.597022
    • B. R. Thomadsen, S. Kubsad, B. R. Paliwal, S. Shahabi, and T. R. Mackie, On the cause of the variation in tissue-maximum ration values with source-to-detector distance., Med. Phys. 20, 723-727 (1993). 10.1118/1.597022 (Pubitemid 23195713)
    • (1993) Medical Physics , vol.20 , Issue.3 , pp. 723-727
    • Thomadsen, B.R.1    Kubsad, S.S.2    Paliwal, B.R.3    Shahabi, S.4    Mackie, T.R.5
  • 44
    • 0032939033 scopus 로고    scopus 로고
    • Measurements of head-scatter factors with cylindrical build-up caps and columnar miniphantoms
    • DOI 10.1118/1.598550
    • P. A. Jursinic and B. R. Thomadsen, Measurements of head-scatter factors with cylindrical build-up caps and columnar miniphantoms., Med. Phys. 26, 512-517 (1999). 10.1118/1.598550 (Pubitemid 29186268)
    • (1999) Medical Physics , vol.26 , Issue.4 , pp. 512-517
    • Jursinic, P.A.1    Thomadsen, B.R.2
  • 45
    • 0030440153 scopus 로고    scopus 로고
    • Characteristics of secondary electrons produced by 6, 10 and 24 MV x-ray beams
    • DOI 10.1088/0031-9155/41/8/016
    • P. A. Jursinic and T. R. Mackie, Characteristics of secondary electrons produced by 6, 10 and 24 MV x-ray beams., Phys. Med. Biol. 41, 1499-1509 (1996). 10.1088/0031-9155/41/8/016 (Pubitemid 27230174)
    • (1996) Physics in Medicine and Biology , vol.41 , Issue.8 , pp. 1499-1509
    • Jursinic, P.A.1    Mackie, T.R.2
  • 47
    • 0023870307 scopus 로고
    • Characteristics of silicon diodes as patient dosemeters in external radiation therapy
    • B. Nilsson, B. I. Ruden, and B. Sorcini, Characteristics of silicon diodes as patient dosimeters in external radiation therapy., Radiother. Oncol. 11, 279-288 (1988). 10.1016/0167-8140(88)90011-4 (Pubitemid 18086884)
    • (1988) Radiotherapy and Oncology , vol.11 , Issue.3 , pp. 279-288
    • Nilsson, B.1    Ruden, B.-I.2    Sorcini, B.3
  • 48
    • 0030887102 scopus 로고    scopus 로고
    • Dose perturbations due to in vivo dosimetry with diodes
    • DOI 10.1016/S0167-8140(96)01872-5, PII S0167814096018725
    • R. Alecu, J. J. Feldmeier, and M. Alecu, Dose perturbations due to in vivo dosimetry with diodes., Radiother. Oncol. 42, 289-291 (1997). 10.1016/S0167-8140(96)01872-5 (Pubitemid 27179684)
    • (1997) Radiotherapy and Oncology , vol.42 , Issue.3 , pp. 289-291
    • Alecu, R.1    Feldmeier, J.J.2    Alecu, M.3
  • 49
    • 0025960367 scopus 로고
    • Comparison of entrance and exit dose measurements using ionization chambers and silicon diodes
    • 10.1088/0031-9155/36/1/005
    • S. Heukelom, J. H. Lanson, and B. J. Mijnheer, Comparison of entrance and exit dose measurements using ionization chambers and silicon diodes., Phys. Med. Biol. 36, 47-59 (1991). 10.1088/0031-9155/36/1/005
    • (1991) Phys. Med. Biol. , vol.36 , pp. 47-59
    • Heukelom, S.1    Lanson, J.H.2    Mijnheer, B.J.3
  • 50
    • 85024948361 scopus 로고    scopus 로고
    • Entrance dose measurements for in-vivo diode dosimetry: Comparison of correction factors for two types of commercial silicon diode detectors
    • 10.1120/1.308253
    • X. R. Zhu, Entrance dose measurements for in-vivo diode dosimetry: Comparison of correction factors for two types of commercial silicon diode detectors., J. Appl. Clin. Med. Phys. 1, 100-107 (2000). 10.1120/1.308253
    • (2000) J. Appl. Clin. Med. Phys. , vol.1 , pp. 100-107
    • Zhu, X.R.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.