Volumn 32, Issue 4, 2011, Pages 271-277

Determination of dialysis dose: A clinical comparison of methods

(19)  Ahrenholz, Peter a   Taborsky, Petr j   Bohling, Margot b   Rawer, Peter c   Ibrahim, Noureddin d   Gajdos, Martin k   MacHek, Petr k   Sagova, Michaela j   Gruber, Hans e   Moucka, Pavel k   Rychlik, Ivan j   Leimenstoll, Gerd d   Vyskocil, Pavel l   Toenne, Gunter f   Possnickerova, Jindriska j   Woggan, Joerg c   Riegel, Werner g   Schneider, Helmut h   Wojke, Ralf i  

i FME   (Germany)

Author keywords

Comparison of methods; Delivered dialysis dose; Hemodialysis; Online clearance measurement; Online hemodiafiltration

Indexed keywords

UREA;

EID: 80051787563     PISSN: 02535068     EISSN: 14219735     Source Type: Journal    
DOI: 10.1159/000330340     Document Type: Article
Times cited : (13)

References (36)
  • 1
    • 0036052220 scopus 로고    scopus 로고
    • European best practice guidelines for hemodialysis: Part 1
    • European Best Practice Guidelines for Hemodialysis: Part 1. Nephrol Dial Transplant 2002; 17(suppl 7):S16-S31.
    • (2002) Nephrol Dial Transplant , vol.17 , Issue.SUPPL. 7
  • 2
    • 0030732432 scopus 로고    scopus 로고
    • Impaired delivery of hemodialysis prescriptions: An analysis of causes and an approach to evaluation
    • Coyne DW, Delmez J, Spence G, Windus DW: Impaired delivery of hemodialysis prescriptions: an analysis of causes and an approach to evaluation. J Am Soc Nephrol 1997; 8: 1315-1318. (Pubitemid 27493638)
    • (1997) Journal of the American Society of Nephrology , vol.8 , Issue.8 , pp. 1315-1318
    • Coyne, D.W.1    Delmez, J.2    Spence, G.3    Windus, D.W.4
  • 3
    • 33747110602 scopus 로고    scopus 로고
    • Quantifying the role of factors that limit attainment of K/DOQI urea reduction ratio dialytic goal
    • Press MH, Benz RL: Quantifying the role of factors that limit attainment of K/DOQI urea reduction ratio dialytic goal. Clin Nephrol 2006; 66: 98-102. (Pubitemid 44219435)
    • (2006) Clinical Nephrology , vol.66 , Issue.2 , pp. 98-102
    • Press, M.H.1    Benz, R.L.2
  • 5
    • 24044464516 scopus 로고    scopus 로고
    • Assessment of hemodialysis adequacy by ionic dialysance: Comparison to standard method of urea removal
    • DOI 10.1080/08860220500200262
    • Racki S, Zaputović L, Maleta I, Grzetić M, Mavrić Z, Devcić B, Vujicić B: Assessment of hemodialysis adequacy by ionic dialysance: comparison to standard method of urea removal. Ren Fail 2005; 27: 601-604. (Pubitemid 41216318)
    • (2005) Renal Failure , vol.27 , Issue.5 , pp. 601-604
    • Racki, S.1    Zaputovic, L.2    Maleta, I.3    Grzetic, M.4    Mavric, Z.5    Devcic, B.6    Vujicic, B.7
  • 6
    • 0032773635 scopus 로고    scopus 로고
    • Urea sensors - A world of possibilities
    • Sternby J: Urea sensors - a world of possibilities. Adv Ren Replace Ther 1999; 6: 265-272. (Pubitemid 29362496)
    • (1999) Advances in Renal Replacement Therapy , vol.6 , Issue.3 , pp. 265-272
    • Sternby, J.1
  • 7
    • 77958105220 scopus 로고    scopus 로고
    • Real-time Kt/V determination by ultraviolet absorbance in spent dialysate: Technique validation
    • E-pub ahead of print
    • Castellarnau A, Werner M, Günthner R, Jakob M: Real-time Kt/V determination by ultraviolet absorbance in spent dialysate: technique validation. Kidney Int 2010 (E-pub ahead of print).
    • (2010) Kidney Int
    • Castellarnau, A.1    Werner, M.2    Günthner, R.3    Jakob, M.4
  • 8
    • 0027688501 scopus 로고
    • Second-generation logarithmic estimates of single-pool variable volume Kt/V: An analysis of error
    • Daugirdas J: Second-generation logarithmic estimates of single-pool variable volume Kt/V: an analysis of error. J Am Soc Nephrol 1993; 4: 1205-1213.
    • (1993) J Am Soc Nephrol , vol.4 , pp. 1205-1213
    • Daugirdas, J.1
  • 10
    • 58849140431 scopus 로고    scopus 로고
    • Improvements in technology: A path to safer and more effective hemodialysis
    • Ward RA, Ronco C: Improvements in technology: a path to safer and more effective hemodialysis. Blood Purif 2009; 27: 6-10.
    • (2009) Blood Purif , vol.27 , pp. 6-10
    • Ward, R.A.1    Ronco, C.2
  • 11
    • 0018864615 scopus 로고
    • Total body water volumes for adult males and females estimated from simple anthropometric measurements
    • Watson PE, Watson ID, Batt RD: Total body water volumes for adult males and females estimated from simple anthropometric measurements. Am J Clin Nutr 1980; 33: 27-39. (Pubitemid 10141676)
    • (1980) American Journal of Clinical Nutrition , vol.33 , Issue.1 , pp. 27-39
    • Watson, P.E.1    Watson, I.D.2    Batt, R.D.3
  • 12
    • 0344945400 scopus 로고    scopus 로고
    • Determination of urea distribution volume for Kt/V assessed by conductivity monitoring
    • DOI 10.1046/j.1523-1755.2003.00327.x
    • Wuepper A, Tattersall J, Kraemer M, Wilkie M, Edwards L: Determination of urea distribution volume for Kt/V assessed by conductivity monitoring. Kidney Int 2003; 64: 2262-2271. (Pubitemid 37449563)
    • (2003) Kidney International , vol.64 , Issue.6 , pp. 2262-2271
    • Wuepper, A.1    Tattersall, J.2    Kraemer, M.3    Wilkie, M.4    Edwards, L.5
  • 15
    • 48049113307 scopus 로고    scopus 로고
    • Whole-body spectroscopy (BCM) in the assessment of normovolemia in hemodialysis patients
    • Wizemann V, Rode C, Wabel P: Whole-body spectroscopy (BCM) in the assessment of normovolemia in hemodialysis patients. Contrib Nephrol. Basel, Karger, 2008, vol 161, pp 115-118.
    • (2008) Contrib Nephrol. Basel Karger , vol.161 , pp. 115-118
    • Wizemann, V.1    Rode, C.2    Wabel, P.3
  • 16
    • 58049217457 scopus 로고    scopus 로고
    • A comparison of methods for determining urea distribution volume for routine use in on-line monitoring of hemodialysis adequacy
    • Lindley EJ, Chamney PW, Wuepper A, Ingles H, Tattersall JE, Will EJ: A comparison of methods for determining urea distribution volume for routine use in on-line monitoring of hemodialysis adequacy. Nephrol Dial Transplant 2009; 24: 211-216.
    • (2009) Nephrol Dial Transplant , vol.24 , pp. 211-216
    • Lindley, E.J.1    Chamney, P.W.2    Wuepper, A.3    Ingles, H.4    Tattersall, J.E.5    Will, E.J.6
  • 17
    • 58849150742 scopus 로고    scopus 로고
    • Importance of whole-body bioimpedance spectroscopy for the management of fluid balance
    • Wabel P, Chamney P, Moissl U, Jirka T: Importance of whole-body bioimpedance spectroscopy for the management of fluid balance. Blood Purif 2009; 27: 75-80.
    • (2009) Blood Purif , vol.27 , pp. 75-80
    • Wabel, P.1    Chamney, P.2    Moissl, U.3    Jirka, T.4
  • 18
    • 75649094885 scopus 로고    scopus 로고
    • Pucalka K: Evaluation of clinical dry weight assessment in haemodialysis patients using bioimpedance spectroscopy: A cross-sectional study
    • Passauer J, Petrov H, Schleser A, Leicht J: Pucalka K: Evaluation of clinical dry weight assessment in haemodialysis patients using bioimpedance spectroscopy: a cross-sectional study. Nephrol Dial Transplant 2010; 25: 545-551.
    • (2010) Nephrol Dial Transplant , vol.25 , pp. 545-551
    • Passauer, J.1    Petrov, H.2    Schleser, A.3    Leicht, J.4
  • 19
    • 34547980351 scopus 로고    scopus 로고
    • Evolution of hemodiafiltration
    • Ronco C: Evolution of hemodiafiltration. Contrib Nephrol. Basel, Karger, 2007, vol 158, pp 9-19.
    • (2007) Contrib Nephrol. Basel Karger , vol.158 , pp. 9-19
    • Ronco, C.1
  • 20
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • PII S0140673686908378
    • Bland JM, Altman DG: Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986; 1: 307-310. (Pubitemid 16134762)
    • (1986) Lancet , vol.1 , Issue.8476 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 21
    • 0024521543 scopus 로고
    • A concordance correlation coefficient to evaluate reproducibility
    • DOI 10.2307/2532051
    • Lin L: A concordance correlation coefficient to evaluate reproducibility. Biometrics 1989; 45: 255-268. (Pubitemid 19109172)
    • (1989) Biometrics , vol.45 , Issue.1 , pp. 255-268
    • I-Kuei Lin, L.1
  • 22
    • 32844474944 scopus 로고    scopus 로고
    • Double target comparison of blood-side methods for measuring the hemodialysis dose
    • Prado M, Roa LM, Palma A, Milán JA: Double target comparison of blood-side methods for measuring the hemodialysis dose. Kidney Int 2005; 68: 863-876.
    • (2005) Kidney Int , vol.68 , pp. 863-876
    • Prado, M.1    Roa, L.M.2    Palma, A.3    Milán, J.A.4
  • 23
    • 27744594208 scopus 로고    scopus 로고
    • Indices for assessment of hemodialysis adequacy: A comparison of different formulae
    • DOI 10.1111/j.1492-7535.2005.01148.x
    • Arogundade F, Barsoum RS: Indices for assessment of hemodialysis adequacy: a comparison of different formulae. Hemodial Int 2005 ;9: 325-331. (Pubitemid 41614641)
    • (2005) Hemodialysis International , vol.9 , Issue.4 , pp. 325-331
    • Arogundade, F.1    Barsoum, R.S.2
  • 24
    • 0031905911 scopus 로고    scopus 로고
    • Variation in blood sample collection for determination of hemodialysis adequacy
    • Beto JA, Bansal VK, Ing TS, Daugirdas JT: Variation in blood sample collection for determination of hemodialysis adequacy. Council on Renal Nutrition National Research Question Collaborative Study Group. Am J Kidney Dis 1998; 31: 135-141. (Pubitemid 28116695)
    • (1998) American Journal of Kidney Diseases , vol.31 , Issue.1 , pp. 135-141
    • Beto, J.A.1    Bansal, V.K.2    Ing, T.S.3    Daugirdas, J.T.4
  • 26
    • 0035409605 scopus 로고    scopus 로고
    • Future directions in dialysis quantification
    • Lindsay RM, Sternby J: Future directions in dialysis quantification. Semin Dial 2001; 14: 300-307. (Pubitemid 33720868)
    • (2001) Seminars in Dialysis , vol.14 , Issue.4 , pp. 300-307
    • Lindsay, R.M.1    Sternby, J.2
  • 27
    • 41849115905 scopus 로고    scopus 로고
    • Preanalytical venous blood sampling practices demand improvement - A survey of test-request management, test-tube labelling and information search procedures
    • Wallin O, Söderberg J, Van Guelpen B, Stenlund H, Grankvist K, Brulin C: Preanalytical venous blood sampling practices demand improvement - a survey of test-request management, test-tube labelling and information search procedures. Clin Chim Acta 2008; 391: 91-97.
    • (2008) Clin Chim Acta , vol.391 , pp. 91-97
    • Wallin, O.1    Söderberg, J.2    Van Guelpen, B.3    Stenlund, H.4    Grankvist, K.5    Brulin, C.6
  • 28
    • 43349084491 scopus 로고    scopus 로고
    • Stability of heparin blood samples during transport based on defined pre-analytical quality goals
    • Jensen EA, Stahl M, Brandslund I, Grinsted P: Stability of heparin blood samples during transport based on defined pre-analytical quality goals. Clin Chem Lab Med 2008; 46: 225-234.
    • (2008) Clin Chem Lab Med , vol.46 , pp. 225-234
    • Jensen, E.A.1    Stahl, M.2    Brandslund, I.3    Grinsted, P.4
  • 30
    • 0036200081 scopus 로고    scopus 로고
    • Accuracy and clinical utility of dialysis dose measurement using online ionic dialysance
    • Katopodis KP, Hoenich NA: Accuracy and clinical utility of dialysis dose measurement using online ionic dialysance. Clin Nephrol 2002; 57: 215-220. (Pubitemid 34213660)
    • (2002) Clinical Nephrology , vol.57 , Issue.3 , pp. 215-220
    • Katopodis, K.P.1    Hoenich, N.A.2
  • 31
    • 0033851929 scopus 로고    scopus 로고
    • Confounding factors in the assessment of delivered hemodialysis dose
    • Santoro A: Confounding factors in the assessment of delivered hemodialysis dose. Kidney Int 2000; 58:S19-S27. (Pubitemid 30612799)
    • (2000) Kidney International, Supplement , vol.58 , Issue.76
    • Santoro, A.1
  • 32
    • 0035093949 scopus 로고    scopus 로고
    • Anthropometry-based equations overestimate the urea distribution volume in hemodialysis patients
    • DOI 10.1046/j.1523-1755.2001.0590031165.x
    • Kloppenburg WD, Stegeman CA, de Jong PE, Huisman RM: Anthropometry-based equations overestimate the urea distribution volume in hemodialysis patients. Kidney Int 2001; 59: 1165-1174. (Pubitemid 32206882)
    • (2001) Kidney International , vol.59 , Issue.3 , pp. 1165-1174
    • Kloppenburg, W.D.1    Stegeman, C.A.2    De Jong, P.E.3    Huisman, R.M.4
  • 33
    • 1442310899 scopus 로고    scopus 로고
    • Body composition evaluation in peritoneal dialysis patients using anthropometric formulas estimating body water
    • Tzamaloukas AH, Murata GH, Vanderjagt DJ, Servilla KS, Glew RH: Body composition evaluation in peritoneal dialysis patients using anthropometric formulas estimating body water. Adv Perit Dial 2003; 19: 212-216.
    • (2003) Adv Perit Dial , vol.19 , pp. 212-216
    • Tzamaloukas, A.H.1    Murata, G.H.2    Vanderjagt, D.J.3    Servilla, K.S.4    Glew, R.H.5
  • 34
    • 0042431923 scopus 로고    scopus 로고
    • Anthropometrically estimated total body water volumes are larger than modeled urea volume in chronic hemodialysis patients: Effects of age, race, and gender
    • DOI 10.1046/j.1523-1755.2003.00179.x
    • Daugirdas JT, Greene T, Depner TA, Chumlea C, Rocco MJ, Chertow GM: Anthropometrically estimated total body water volumes are larger than modeled urea volume in chronic hemodialysis patients: effects of age, race, and gender. Kidney Int 2003; 64: 1108-1119. (Pubitemid 37013850)
    • (2003) Kidney International , vol.64 , Issue.3 , pp. 1108-1119
    • Daugirdas, J.T.1    Greene, T.2    Depner, T.A.3    Chumlea, C.4    Rocco, M.J.5    Chertow, G.M.6


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.