A phantom study of an in vivo dosimetry system using plastic scintillation detectors for real-time verification of 192Ir HDR brachytherapy

Author keywords

brachytherapy; high dose rate; in vivo dosimetry; plastic scintillation detectors; stem effect

Indexed keywords

CATHETERS; ERROR DETECTION; NEUTRON DETECTORS; ONCOLOGY; ONLINE SYSTEMS; OPTICAL FIBER FABRICATION; PHANTOMS; PLASTIC OPTICAL FIBERS; RADIOACTIVITY; RADIOTHERAPY; SCINTILLATION; WALLS (STRUCTURAL PARTITIONS); WATER TREATMENT;

EID: 79959564142     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3572229     Document Type: Article
Times cited : (81)

References (26)
  • 1
    • 0023770696 scopus 로고
    • 192Ir using LiF TLD
    • 10.1088/0031-9155/33/10/005
    • 192Ir using LiF TLD., Phys. Med. Biol. 33, 1159-1170 (1988). 10.1088/0031-9155/33/10/005
    • (1988) Phys. Med. Biol , vol.33 , pp. 1159-1170
    • Meigooni, A.1    Meli, J.2    Nath, R.3
  • 2
    • 0029563617 scopus 로고
    • TLD, diode and Monte Carlo dosimetry of an 192-Ir source for high dose-rate brachytherapy
    • 10.1088/0031-9155/40/12/002
    • A. S. Kirov, J. F. Williamson, A. S. Meigooni, and Y. Zhu, TLD, diode and Monte Carlo dosimetry of an 192-Ir source for high dose-rate brachytherapy., Phys. Med. Biol. 40, 2015-2036 (1995). 10.1088/0031-9155/40/12/002
    • (1995) Phys. Med. Biol , vol.40 , pp. 2015-2036
    • Kirov, A.S.1    Williamson, J.F.2    Meigooni, A.S.3    Zhu, Y.4
  • 3
    • 0031853415 scopus 로고    scopus 로고
    • Application of a diamond detector to brachytherapy dosimetry
    • DOI 10.1088/0031-9155/43/8/007, PII S0031915598895183
    • S. N. Rustgi, Application of a diamond detector to brachytherapy dosimetry., Phys. Med. Biol. 43, 2085-2094 (1998). 10.1088/0031-9155/43/8/007 (Pubitemid 28372610)
    • (1998) Physics in Medicine and Biology , vol.43 , Issue.8 , pp. 2085-2094
    • Rustgi, S.N.1
  • 4
    • 0033744527 scopus 로고    scopus 로고
    • In vivo urethral dose measurements: A method to verify high dose rate prostate treatments
    • 10.1118/1.1312811
    • I. A. Brezovich, J. Duan, P. N. Pareek, J. Fiveash, and M. Ezekiel, In vivo urethral dose measurements: A method to verify high dose rate prostate treatments., Med. Phys. 27, 2297-2301 (2000). 10.1118/1.1312811
    • (2000) Med. Phys , vol.27 , pp. 2297-2301
    • Brezovich, I.A.1    Duan, J.2    Pareek, P.N.3    Fiveash, J.4    Ezekiel, M.5
  • 5
    • 28944443904 scopus 로고    scopus 로고
    • In-vivo dosimetry for gynaecological brachytherapy: Physical and clinical considerations
    • DOI 10.1016/j.radonc.2005.09.004, PII S0167814005003907
    • C. Waldhausl, A. Wamberside, R. Potter, and D. Georg, In-vivo dosimetry for gynaecological brachytherapy: Physical and clinical considerations., Radiother. Oncol. 77, 310-317 (2005). 10.1016/j.radonc.2005.09.004 (Pubitemid 41785342)
    • (2005) Radiotherapy and Oncology , vol.77 , Issue.3 , pp. 310-317
    • Waldhausl, C.1    Wambersie, A.2    Potter, R.3    Georg, D.4
  • 6
    • 33144482782 scopus 로고    scopus 로고
    • Dosimetric evaluation of a new OneDose MOSFET for Ir-192 energy
    • DOI 10.1088/0031-9155/51/5/015, PII S0031915506138142
    • R. A. Kinhikar, P. K. Sharma, C. M. Tambe, and D. D. Deshpande, Dosimetric evaluation of a new OneDose MOSFET for Ir-192 energy., Phys. Med. Biol. 51, 1261-1268 (2006). 10.1088/0031-9155/51/5/015 (Pubitemid 43267193)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.5 , pp. 1261-1268
    • Kinhikar, R.A.1    Sharma, P.K.2    Tambe, C.M.3    Deshpande, D.D.4
  • 7
    • 33751168473 scopus 로고    scopus 로고
    • A plastic scintillation dosimeter for high dose rate brachytherapy
    • DOI 10.1088/0031-9155/51/21/008, PII S0031915506261293, 008
    • J. Lambert, D. R. McKenzie, S. Law, J. Elsey, and N. Suchowerska, A plastic scintillation dosimeter for high dose rate brachytherapy., Phys. Med. Biol. 51, 5505-5516 (2006). 10.1088/0031-9155/51/21/008 (Pubitemid 44773328)
    • (2006) Physics in Medicine and Biology , vol.51 , Issue.21 , pp. 5505-5516
    • Lambert, J.1    McKenzie, D.R.2    Law, S.3    Elsey, J.4    Suchowerska, N.5
  • 9
    • 34247616143 scopus 로고    scopus 로고
    • In vivo dosimeters for HDR brachytherapy: A comparison of a diamond detector, MOSFET, TLD, and scintillation detector
    • DOI 10.1118/1.2727248
    • J. Lambert, T. Nakano, S. Law, J. Elsey, D. R. McKenzie, and N. Suchowerska, In vivo dosimeters for HDR brachytherapy: A comparison of a diamond detector, MOSFET, TLD, and scintillation detector., Med. Phys. 34, 1759-1765 (2007). 10.1118/1.2727248 (Pubitemid 46669170)
    • (2007) Medical Physics , vol.34 , Issue.5 , pp. 1759-1765
    • Lambert, J.1    Nakano, T.2    Law, S.3    Elsey, J.4    McKenzie, D.R.5    Suchowerska, N.6
  • 10
    • 56749186073 scopus 로고    scopus 로고
    • Response of an implantable MOSFET dosimeter to 192-Ir HDR radiation
    • 10.1118/1.3013574
    • J. M. Fagerstrom, J. A. Micka, and L. A. DeWerd, Response of an implantable MOSFET dosimeter to 192-Ir HDR radiation., Med. Phys. 35, 5729-5737 (2008). 10.1118/1.3013574
    • (2008) Med. Phys , vol.35 , pp. 5729-5737
    • Fagerstrom, J.M.1    Micka, J.A.2    Dewerd, L.A.3
  • 11
    • 0031739032 scopus 로고    scopus 로고
    • Monte Carlo-aided dosimetry of a new high dose-rate brachytherapy source
    • DOI 10.1118/1.598418
    • G. M. Daskalov, E. Loffler, and J. F. Williamson, Monte-Carlo-aided dosimetry of a new high dose-rate brachytherapy source., Med. Phys. 25, 2200-2208 (1998). 10.1118/1.598418 (Pubitemid 28530111)
    • (1998) Medical Physics , vol.25 , Issue.11 , pp. 2200-2208
    • Daskalov, G.M.1    Loffler, E.2    Williamson, J.F.3
  • 12
    • 34447301820 scopus 로고    scopus 로고
    • Verification of the plan dosimetry for high dose rate brachytherapy using metal-oxide-semiconductor field effect transistor detectors
    • DOI 10.1118/1.2736288
    • Z.-Y. Qi, X.-W. Deng, S.-M. Huang, J. Lu, M. Lerch, D. Cutajar, and A. Rosenfeld, Verification of the plan dosimetry for high dose rate brachytherapy using metal-oxide-semiconductor field effect transistor detectors., Med. Phys. 34, 2007-2013 (2007). 10.1118/1.2736288 (Pubitemid 47055736)
    • (2007) Medical Physics , vol.34 , Issue.6 , pp. 2007-2013
    • Qi, Z.-Y.1    Deng, X.-W.2    Huang, S.-M.3    Lu, J.4    Lerch, M.5    Cutajar, D.6    Rosenfeld, A.7
  • 15
    • 70449379227 scopus 로고    scopus 로고
    • Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy
    • 10.1118/1.3238102
    • C. E. Andersen, S. K. Nielsen, J. C. Lindegaard, and K. Tanderup, Time-resolved in vivo luminescence dosimetry for online error detection in pulsed dose-rate brachytherapy., Med. Phys. 36, 5033-5043 (2009). 10.1118/1.3238102
    • (2009) Med. Phys , vol.36 , pp. 5033-5043
    • Andersen, C.E.1    Nielsen, S.K.2    Lindegaard, J.C.3    Tanderup, K.4
  • 16
    • 77952390957 scopus 로고    scopus 로고
    • Dose mapping of the rectal wall during brachytherapy with an array of scintillation dosimeters
    • 10.1118/1.3397446
    • L. E. Cartwright, N. Suchowerska, Y. Yin, J. Lambert, M. Haque, and D. R. McKenzie, Dose mapping of the rectal wall during brachytherapy with an array of scintillation dosimeters., Med. Phys. 37, 2247-2255 (2010). 10.1118/1.3397446
    • (2010) Med. Phys , vol.37 , pp. 2247-2255
    • Cartwright, L.E.1    Suchowerska, N.2    Yin, Y.3    Lambert, J.4    Haque, M.5    McKenzie, D.R.6
  • 18
    • 0026793771 scopus 로고
    • Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements
    • 10.1088/0031-9155/37/10/007
    • A. S. Beddar, T. R. Mackie, and F. H. Attix, Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: II. Properties and measurements., Phys. Med. Biol. 37, 1901-1913 (1992). 10.1088/0031-9155/37/10/ 007
    • (1992) Phys. Med. Biol , vol.37 , pp. 1901-1913
    • Beddar, A.S.1    MacKie, T.R.2    Attix, F.H.3
  • 19
    • 0026804696 scopus 로고
    • Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerations
    • 10.1088/0031-9155/37/10/006
    • A. S. Beddar, T. R. Mackie, and F. H. Attix, Water-equivalent plastic scintillation detectors for high-energy beam dosimetry: I. Physical characteristics and theoretical considerations., Phys. Med. Biol. 37, 1883-1900 (1992). 10.1088/0031-9155/37/10/006
    • (1992) Phys. Med. Biol , vol.37 , pp. 1883-1900
    • Beddar, A.S.1    MacKie, T.R.2    Attix, F.H.3
  • 20
    • 31044441579 scopus 로고    scopus 로고
    • Measurement accuracy and Cerenkov removal for high performance, high spatial resolution scintillation dosimetry
    • DOI 10.1118/1.2138010
    • L. Archambault, A. S. Beddar, L. Gingras, R. Roy, and L. Beaulieu, Measurement accuracy and Cerenkov removal for high performance, high spatial resolution scintillation dosimetry., Med. Phys. 33, 128-135 (2006). 10.1118/1.2138010 (Pubitemid 43121042)
    • (2006) Medical Physics , vol.33 , Issue.1 , pp. 128-135
    • Archambault, L.1    Sam Beddar, A.2    Gingras, L.3    Roy, R.4    Beaulieu, L.5
  • 21
    • 0026555508 scopus 로고
    • Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams
    • 10.1088/0031-9155/37/4/007
    • A. S. Beddar, T. R. Mackie, and F. H. Attix, Cerenkov light generated in optical fibres and other light pipes irradiated by electron beams., Phys. Med. Biol. 37, 925-935 (1992). 10.1088/0031-9155/37/4/007
    • (1992) Phys. Med. Biol , vol.37 , pp. 925-935
    • Beddar, A.S.1    MacKie, T.R.2    Attix, F.H.3
  • 22
    • 79953671467 scopus 로고    scopus 로고
    • Technical note: Removing the stem effect when performing Ir-192 HDR brachytherapy in vivo dosimetry using plastic scintillation detectors-A relevant and necessary step
    • 38, 2176-2179 (2011). 10.1118/1.3562902
    • F. Therriault-Proulx, S. Beddar, T. M. Briere, L. Archambault, and L. Beaulieu, Technical note: Removing the stem effect when performing Ir-192 HDR brachytherapy in vivo dosimetry using plastic scintillation detectors-A relevant and necessary step., Med. Phys. 38, 2176-2179 (2011). 10.1118/1.3562902
    • Med. Phys
    • Therriault-Proulx, F.1    Beddar, S.2    Briere, T.M.3    Archambault, L.4    Beaulieu, L.5
  • 23
    • 0028919251 scopus 로고
    • Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43
    • 10.1118/1.597458
    • R. Nath, L. L. Anderson, G. Luxton, K. A. Weaver, J. F. Williamson, and A. S. Meigooni, Dosimetry of interstitial brachytherapy sources: Recommendations of the AAPM Radiation Therapy Committee Task Group No. 43., Med. Phys. 22, 209-234 (1995). 10.1118/1.597458
    • (1995) Med. Phys , vol.22 , pp. 209-234
    • Nath, R.1    Anderson, L.L.2    Luxton, G.3    Weaver, K.A.4    Williamson, J.F.5    Meigooni, A.S.6
  • 25
    • 79953670862 scopus 로고    scopus 로고
    • Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: A comparison study of calibration procedures and validation in Cerenkov light-dominated situations
    • 38, 2140-2150 (2011). 10.1118/1.3562896
    • M. Guillot, L. Gingras, L. Archambault, S. Beddar, and L. Beaulieu, Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: A comparison study of calibration procedures and validation in Cerenkov light-dominated situations., Med. Phys. 38, 2140-2150 (2011). 10.1118/1.3562896
    • Med. Phys
    • Guillot, M.1    Gingras, L.2    Archambault, L.3    Beddar, S.4    Beaulieu, L.5
  • 26
    • 0034844177 scopus 로고    scopus 로고
    • Measurement of transit time of a remote after-loading high dose rate brachytherapy source
    • DOI 10.1118/1.1386779
    • N. Sahoo, Measurement of transit time of a remote after-loading high dose rate brachytherapy source., Med. Phys. 28, 1786-1790 (2001). 10.1118/1.1386779 (Pubitemid 32783224)
    • (2001) Medical Physics , vol.28 , Issue.8 , pp. 1786-1790
    • Sahoo, N.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.