Setup Reproducibility for Thoracic and Upper Gastrointestinal Radiation Therapy: Influence of Immobilization Method and On-Line Cone-Beam CT Guidance

Author keywords

Cone beam CT; Immobilization; Margins; Thoracic and upper gastrointestinal radiotherapy

Indexed keywords

ACCURACY; ANALYTICAL ERROR; ARTICLE; CANCER RADIOTHERAPY; COMPARATIVE STUDY; COMPUTER ASSISTED TOMOGRAPHY; COMPUTER PROGRAM; CONTROLLED STUDY; ESOPHAGUS CARCINOMA; HUMAN; IMAGE ANALYSIS; LUNG CARCINOMA; MAJOR CLINICAL STUDY; PANCREAS CARCINOMA; PRIORITY JOURNAL; STOMACH CARCINOMA; TREATMENT PLANNING;

EID: 78049462718     PISSN: 09583947     EISSN: None     Source Type: Journal    
DOI: 10.1016/j.meddos.2009.09.003     Document Type: Article
Times cited : (19)

References (34)
  • 1
    • 3142697244 scopus 로고    scopus 로고
    • Canadian Cancer Society/National Cancer Institute of Canada, Toronto, Canada
    • Canadian Cancer Statistics 2006 2006, 1-113. Canadian Cancer Society/National Cancer Institute of Canada, Toronto, Canada.
    • (2006) Canadian Cancer Statistics 2006 , pp. 1-113
  • 2
    • 33646247402 scopus 로고    scopus 로고
    • The potential for dose escalation in lung cancer as a result of systematically reducing margins used to generate planning target volume
    • Nelson C., Starkschall G., Chang J.Y. The potential for dose escalation in lung cancer as a result of systematically reducing margins used to generate planning target volume. Int. J. Radiat. Oncol. Biol. Phys. 2006, 65:573-586.
    • (2006) Int. J. Radiat. Oncol. Biol. Phys. , vol.65 , pp. 573-586
    • Nelson, C.1    Starkschall, G.2    Chang, J.Y.3
  • 3
    • 0035283667 scopus 로고    scopus 로고
    • Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning
    • de Boer H.C., van Sornsen de Koste J.R., Senan S., et al. Analysis and reduction of 3D systematic and random setup errors during the simulation and treatment of lung cancer patients with CT-based external beam radiotherapy dose planning. Int. J. Radiat. Oncol. Biol. Phys 2001, 49:857-868.
    • (2001) Int. J. Radiat. Oncol. Biol. Phys , vol.49 , pp. 857-868
    • de Boer, H.C.1    van Sornsen de Koste, J.R.2    Senan, S.3
  • 4
    • 0037249948 scopus 로고    scopus 로고
    • Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer
    • Erridge S.C., Seppenwoolde Y., Muller S.H., et al. Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer. Radiother. Oncol 2003, 66:75-85.
    • (2003) Radiother. Oncol , vol.66 , pp. 75-85
    • Erridge, S.C.1    Seppenwoolde, Y.2    Muller, S.H.3
  • 5
    • 0032899184 scopus 로고    scopus 로고
    • Setup reproducibility in radiation therapy for lung cancer: A comparison between T-bar and expanded foam immobilization devices
    • Halperin R., Roa W., Field M., et al. Setup reproducibility in radiation therapy for lung cancer: A comparison between T-bar and expanded foam immobilization devices. Int. J. Radiat. Oncol. Biol. Phys. 1999, 43:211-216.
    • (1999) Int. J. Radiat. Oncol. Biol. Phys. , vol.43 , pp. 211-216
    • Halperin, R.1    Roa, W.2    Field, M.3
  • 6
    • 0030793183 scopus 로고    scopus 로고
    • Impact of cradle immobilization on setup reproducibility during external beam radiation therapy for lung cancer
    • Bentel G.C., Marks L.B., Krishnamurthy R. Impact of cradle immobilization on setup reproducibility during external beam radiation therapy for lung cancer. Int. J. Radiat. Oncol. Biol. Phys. 1997, 38:527-531.
    • (1997) Int. J. Radiat. Oncol. Biol. Phys. , vol.38 , pp. 527-531
    • Bentel, G.C.1    Marks, L.B.2    Krishnamurthy, R.3
  • 7
    • 0027425957 scopus 로고
    • Immobilization improves the reproducibility of patient positioning during six-field conformal radiation therapy for prostate carcinoma
    • Rosenthal S.A., Roach M.3rd., Goldsmith B.J., et al. Immobilization improves the reproducibility of patient positioning during six-field conformal radiation therapy for prostate carcinoma. Int. J. Radiat. Oncol. Biol. Phys. 1993, 27:921-926.
    • (1993) Int. J. Radiat. Oncol. Biol. Phys. , vol.27 , pp. 921-926
    • Rosenthal, S.A.1    Roach, M.2    Goldsmith, B.J.3
  • 8
    • 0032843201 scopus 로고    scopus 로고
    • Radiotherapy of pelvic malignancies: Impact of two types of rigid immobilisation devices on localisation errors
    • Mitine C., Hoornaert M.T., Dutreix A., et al. Radiotherapy of pelvic malignancies: Impact of two types of rigid immobilisation devices on localisation errors. Radiother. Oncol. 1999, 52:19-27.
    • (1999) Radiother. Oncol. , vol.52 , pp. 19-27
    • Mitine, C.1    Hoornaert, M.T.2    Dutreix, A.3
  • 9
    • 0030033457 scopus 로고    scopus 로고
    • A comparison of four patient immobilization devices in the treatment of prostate cancer patients with three dimensional conformal radiotherapy
    • Song P.Y., Washington M., Vaida F., et al. A comparison of four patient immobilization devices in the treatment of prostate cancer patients with three dimensional conformal radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 1996, 34:213-219.
    • (1996) Int. J. Radiat. Oncol. Biol. Phys. , vol.34 , pp. 213-219
    • Song, P.Y.1    Washington, M.2    Vaida, F.3
  • 10
    • 0033621916 scopus 로고    scopus 로고
    • A randomized study of the use of a customized immobilization system in the treatment of prostate cancer with conformal radiotherapy
    • Nutting C.M., Khoo V.S., Walker V., et al. A randomized study of the use of a customized immobilization system in the treatment of prostate cancer with conformal radiotherapy. Radiother. Oncol 2000, 54:1-9.
    • (2000) Radiother. Oncol , vol.54 , pp. 1-9
    • Nutting, C.M.1    Khoo, V.S.2    Walker, V.3
  • 11
    • 0345435973 scopus 로고    scopus 로고
    • An analysis of anatomic landmark mobility and setup deviations in radiotherapy for lung cancer
    • Samson M.J., van Sornsen de Koste J.R., de Boer H.C., et al. An analysis of anatomic landmark mobility and setup deviations in radiotherapy for lung cancer. Int. J. Radiat. Oncol. Biol. Phys. 1999, 43:827-832.
    • (1999) Int. J. Radiat. Oncol. Biol. Phys. , vol.43 , pp. 827-832
    • Samson, M.J.1    van Sornsen de Koste, J.R.2    de Boer, H.C.3
  • 12
    • 20644454162 scopus 로고    scopus 로고
    • Emergent technologies for 3-dimensional image-guided radiation delivery
    • Jaffray D.A. Emergent technologies for 3-dimensional image-guided radiation delivery. Semin. Radiat. Oncol. 2005, 15:208-216.
    • (2005) Semin. Radiat. Oncol. , vol.15 , pp. 208-216
    • Jaffray, D.A.1
  • 13
    • 0034086714 scopus 로고    scopus 로고
    • Cone-beam computed tomography with a flat-panel imager: initial performance characterization
    • Jaffray D.A., Siewerdsen J.H. Cone-beam computed tomography with a flat-panel imager: initial performance characterization. Med. Phys 2000, 27:1311-1323.
    • (2000) Med. Phys , vol.27 , pp. 1311-1323
    • Jaffray, D.A.1    Siewerdsen, J.H.2
  • 14
    • 0036680147 scopus 로고    scopus 로고
    • Flat-panel cone-beam computed tomography for image-guided radiation therapy
    • Jaffray D.A., Siewerdsen J.H., Wong J.W., et al. Flat-panel cone-beam computed tomography for image-guided radiation therapy. Int. J. Radiat. Oncol. Biol. Phys. 2002, 53:1337-1349.
    • (2002) Int. J. Radiat. Oncol. Biol. Phys. , vol.53 , pp. 1337-1349
    • Jaffray, D.A.1    Siewerdsen, J.H.2    Wong, J.W.3
  • 15
    • 0035989949 scopus 로고    scopus 로고
    • A performance comparison of flat-panel imager-based MV and kV cone-beam CT
    • Groh B.A., Siewerdsen J.H., Drake D.G., et al. A performance comparison of flat-panel imager-based MV and kV cone-beam CT. Med. Phys. 2002, 29:967-975.
    • (2002) Med. Phys. , vol.29 , pp. 967-975
    • Groh, B.A.1    Siewerdsen, J.H.2    Drake, D.G.3
  • 16
    • 22444451154 scopus 로고    scopus 로고
    • Cone-beam-CT guided radiation therapy: A model for on-line application
    • Oldham M., Letourneau D., Watt L., et al. Cone-beam-CT guided radiation therapy: A model for on-line application. Radiother. Oncol 2005, 75:271-278.
    • (2005) Radiother. Oncol , vol.75 , pp. 271-278
    • Oldham, M.1    Letourneau, D.2    Watt, L.3
  • 17
    • 0037422708 scopus 로고    scopus 로고
    • Acquiring a four-dimensional computed tomography dataset using an external respiratory signal
    • Vedam S.S., Keall P.J., Kini V.R., et al. Acquiring a four-dimensional computed tomography dataset using an external respiratory signal. Phys. Med. Biol. 2003, 48:45-62.
    • (2003) Phys. Med. Biol. , vol.48 , pp. 45-62
    • Vedam, S.S.1    Keall, P.J.2    Kini, V.R.3
  • 18
    • 33744823315 scopus 로고    scopus 로고
    • Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy
    • Islam M.K., Purdie T.G., Norrlinger B.D., et al. Patient dose from kilovoltage cone beam computed tomography imaging in radiation therapy. Med. Phys 2006, 33:1573-1582.
    • (2006) Med. Phys , vol.33 , pp. 1573-1582
    • Islam, M.K.1    Purdie, T.G.2    Norrlinger, B.D.3
  • 19
    • 27144520502 scopus 로고    scopus 로고
    • Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy
    • Smitsmans M.H., de Bois J., Sonke J.J., et al. Automatic prostate localization on cone-beam CT scans for high precision image-guided radiotherapy. Int. J. Radiat. Oncol. Biol. Phys. 2005, 63:975-984.
    • (2005) Int. J. Radiat. Oncol. Biol. Phys. , vol.63 , pp. 975-984
    • Smitsmans, M.H.1    de Bois, J.2    Sonke, J.J.3
  • 20
    • 0024205301 scopus 로고
    • Models for longitudinal data: a generalized estimating equation approach
    • Zeger S.L., Liang K.Y., Albert P.S. Models for longitudinal data: a generalized estimating equation approach. Biometrics. 1988, 44:1049-1060.
    • (1988) Biometrics. , vol.44 , pp. 1049-1060
    • Zeger, S.L.1    Liang, K.Y.2    Albert, P.S.3
  • 21
    • 0842306308 scopus 로고    scopus 로고
    • Errors and margins in radiotherapy
    • van Herk M. Errors and margins in radiotherapy. Semin. Radiat. Oncol. 2004, 14:52-64.
    • (2004) Semin. Radiat. Oncol. , vol.14 , pp. 52-64
    • van Herk, M.1
  • 22
    • 0034237797 scopus 로고    scopus 로고
    • The probability of correct target dosage: Dose-population histograms for deriving treatment margins in radiotherapy
    • van Herk M., Remeijer P., Rasch C., et al. The probability of correct target dosage: Dose-population histograms for deriving treatment margins in radiotherapy. Int. J. Radiat. Oncol. Biol. Phys 2000, 47:1121-1135.
    • (2000) Int. J. Radiat. Oncol. Biol. Phys , vol.47 , pp. 1121-1135
    • van Herk, M.1    Remeijer, P.2    Rasch, C.3
  • 23
    • 34247117297 scopus 로고    scopus 로고
    • Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: Localization, verification, and intrafraction tumor position
    • Purdie T.G., Bissonnette J.P., Franks K., et al. Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: Localization, verification, and intrafraction tumor position. Int. J. Radiat. Oncol. Biol. Phys. 2007, 68:243-252.
    • (2007) Int. J. Radiat. Oncol. Biol. Phys. , vol.68 , pp. 243-252
    • Purdie, T.G.1    Bissonnette, J.P.2    Franks, K.3
  • 24
    • 34047145827 scopus 로고    scopus 로고
    • Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors
    • Guckenberger M., Meyer J., Wilbert J., et al. Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. Radiother. Oncol. 2007, 83:57-64.
    • (2007) Radiother. Oncol. , vol.83 , pp. 57-64
    • Guckenberger, M.1    Meyer, J.2    Wilbert, J.3
  • 25
    • 0031895830 scopus 로고    scopus 로고
    • Electronic portal imaging with on-line correction of setup error in thoracic irradiation: Clinical evaluation
    • Van de Steene J., Van den Heuvel F., Bel A., et al. Electronic portal imaging with on-line correction of setup error in thoracic irradiation: Clinical evaluation. Int. J. Radiat. Oncol. Biol. Phys. 1998, 40:967-976.
    • (1998) Int. J. Radiat. Oncol. Biol. Phys. , vol.40 , pp. 967-976
    • Van de Steene, J.1    Van den Heuvel, F.2    Bel, A.3
  • 26
    • 0034748364 scopus 로고    scopus 로고
    • Positioning errors and prostate motion during conformal prostate radiotherapy using on-line isocentre set-up verification and implanted prostate markers
    • Wu J., Haycocks T., Alasti H., et al. Positioning errors and prostate motion during conformal prostate radiotherapy using on-line isocentre set-up verification and implanted prostate markers. Radiother. Oncol. 2001, 61:127-133.
    • (2001) Radiother. Oncol. , vol.61 , pp. 127-133
    • Wu, J.1    Haycocks, T.2    Alasti, H.3
  • 27
    • 34247887718 scopus 로고    scopus 로고
    • Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation
    • White E.A., Cho J., Vallis K.A., et al. Cone beam computed tomography guidance for setup of patients receiving accelerated partial breast irradiation. Int. J. Radiat. Oncol. Biol. Phys 2007, 68:547-554.
    • (2007) Int. J. Radiat. Oncol. Biol. Phys , vol.68 , pp. 547-554
    • White, E.A.1    Cho, J.2    Vallis, K.A.3
  • 29
    • 0035425538 scopus 로고    scopus 로고
    • A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload
    • de Boer H.C., Heijmen B.J. A protocol for the reduction of systematic patient setup errors with minimal portal imaging workload. Int. J. Radiat. Oncol. Biol. Phys. 2001, 50:1350-1365.
    • (2001) Int. J. Radiat. Oncol. Biol. Phys. , vol.50 , pp. 1350-1365
    • de Boer, H.C.1    Heijmen, B.J.2
  • 30
    • 0027138281 scopus 로고
    • A verification procedure to improve patient set-up accuracy using portal images
    • Bel A., van Herk M., Bartelink H., et al. A verification procedure to improve patient set-up accuracy using portal images. Radiother. Oncol. 1993, 29:253-260.
    • (1993) Radiother. Oncol. , vol.29 , pp. 253-260
    • Bel, A.1    van Herk, M.2    Bartelink, H.3
  • 31
    • 34447321343 scopus 로고    scopus 로고
    • Image-guided patient positioning: If one cannot correct for rotational offsets in external-beam radiotherapy setup, how should rotational offsets be managed?
    • Murphy M.J. Image-guided patient positioning: If one cannot correct for rotational offsets in external-beam radiotherapy setup, how should rotational offsets be managed?. Med. Phys. 2007, 34:1880-1883.
    • (2007) Med. Phys. , vol.34 , pp. 1880-1883
    • Murphy, M.J.1
  • 32
    • 27144468304 scopus 로고    scopus 로고
    • Serial megavoltage CT imaging during external beam radiotherapy for non-small-cell lung cancer: Observations on tumor regression during treatment
    • Kupelian P.A., Ramsey C., Meeks S.L., et al. Serial megavoltage CT imaging during external beam radiotherapy for non-small-cell lung cancer: Observations on tumor regression during treatment. Int. J. Radiat. Oncol. Biol. Phys 2005, 63:1024-1028.
    • (2005) Int. J. Radiat. Oncol. Biol. Phys , vol.63 , pp. 1024-1028
    • Kupelian, P.A.1    Ramsey, C.2    Meeks, S.L.3
  • 33
    • 33746814036 scopus 로고    scopus 로고
    • Tumor volume changes on serial imaging with megavoltage CT for non-small-cell lung cancer during intensity-modulated radiotherapy: How reliable, consistent, and meaningful is the effect?
    • Siker M.L., Tome W.A., Mehta M.P. Tumor volume changes on serial imaging with megavoltage CT for non-small-cell lung cancer during intensity-modulated radiotherapy: How reliable, consistent, and meaningful is the effect?. Int. J. Radiat. Oncol. Biol. Phys. 2006, 66:135-141.
    • (2006) Int. J. Radiat. Oncol. Biol. Phys. , vol.66 , pp. 135-141
    • Siker, M.L.1    Tome, W.A.2    Mehta, M.P.3
  • 34
    • 34347269146 scopus 로고    scopus 로고
    • Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice
    • Guckenberger M., Meyer J., Wilbert J., et al. Precision of image-guided radiotherapy (IGRT) in six degrees of freedom and limitations in clinical practice. Strahlenther. Onkol 2007, 183:307-313.
    • (2007) Strahlenther. Onkol , vol.183 , pp. 307-313
    • Guckenberger, M.1    Meyer, J.2    Wilbert, J.3


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.