Volumn 36, Issue 5, 2009, Pages 1768-1774

Measurement of relative output factors for the 8 and 4 mm collimators of Leksell Gamma Knife Perfexion by film dosimetry

Author keywords

Leksell Gamma Knife Perfexion ; Relative output factor; Small field dosimetry

Indexed keywords

CALIBRATION; DOSIMETRY;

EID: 65449117589     PISSN: 00942405     EISSN: None     Source Type: Journal    
DOI: 10.1118/1.3113904     Document Type: Article
Times cited : (43)

References (39)
  • 1
    • 34548665085 scopus 로고    scopus 로고
    • The Leksell Gamma Knife Perfexion and comparisons with its predecessors
    • 0148-396X,. 10.1227/01.neu.0000289726.35330.8a
    • C. Lindquist and I. Paddick, " The Leksell Gamma Knife Perfexion and comparisons with its predecessors.," Neurosurgery 0148-396X 61, 130-140 (2007). 10.1227/01.neu.0000289726.35330.8a
    • (2007) Neurosurgery , vol.61 , pp. 130-140
    • Lindquist, C.1    Paddick, I.2
  • 5
    • 0024492354 scopus 로고
    • Stereotactic radiosurgery of the brain using the first United States 201 cobalt-60 source gamma knife
    • 0148-396X,. 10.1097/00006123-198902000-00001
    • L. D. Lunsford, J. Flickinger, G. Lindner, and A. Maitz, " Stereotactic radiosurgery of the brain using the first United States 201 cobalt-60 source gamma knife.," Neurosurgery 0148-396X 24, 151-159 (1989). 10.1097/00006123-198902000-00001
    • (1989) Neurosurgery , vol.24 , pp. 151-159
    • Lunsford, L.D.1    Flickinger, J.2    Lindner, G.3    Maitz, A.4
  • 6
    • 0034433343 scopus 로고    scopus 로고
    • First clinical experience with the automatic positioning system and Leksell gamma knife model C, Technical note
    • 0022-3085.
    • G. A. Horstmann, H. Schöpgens, A. T. C. J. Van Eck, H. J. Kreiner, and W. Hertz, " First clinical experience with the automatic positioning system and Leksell gamma knife model C, Technical note.," J. Neurosurg. 0022-3085 93, 193-197 (2000).
    • (2000) J. Neurosurg. , vol.93 , pp. 193-197
    • Horstmann, G.A.1    Schöpgens, H.2    Van Eck, A.T.C.J.3    Kreiner, H.J.4    Hertz, W.5
  • 7
    • 0036484633 scopus 로고    scopus 로고
    • An evaluation of the Model C gamma knife with automatic patient positioning
    • 0148-396X,. 10.1097/00006123-200202000-00041
    • D. Kondziolka, A. H. Maitz, A. Niranjan, J. C. Flickinger, and L. D. Lunsford, " An evaluation of the Model C gamma knife with automatic patient positioning.," Neurosurgery 0148-396X 50, 429-431 (2002). 10.1097/00006123-200202000-00041
    • (2002) Neurosurgery , vol.50 , pp. 429-431
    • Kondziolka, D.1    Maitz, A.H.2    Niranjan, A.3    Flickinger, J.C.4    Lunsford, L.D.5
  • 8
    • 3042819108 scopus 로고    scopus 로고
    • The Leksell gamma knife Model U versus Model C: A quantitative comparison of radiosurgical treatment parameters
    • 0148-396X,. 10.1227/01.NEU.0000126880.33125.E6
    • J. S. Kuo, C. Yu, S. L. Giannotta, Z. Petrovich, and M. L. Apuzzo, " The Leksell gamma knife Model U versus Model C: A quantitative comparison of radiosurgical treatment parameters.," Neurosurgery 0148-396X 55, 168-172 (2004). 10.1227/01.NEU.0000126880.33125.E6
    • (2004) Neurosurgery , vol.55 , pp. 168-172
    • Kuo, J.S.1    Yu, C.2    Giannotta, S.L.3    Petrovich, Z.4    Apuzzo, M.L.5
  • 9
    • 13744258554 scopus 로고    scopus 로고
    • A comparison of the gamma knife model C and the automatic positioning system with Leksell model B
    • 0022-3085.
    • D. Tlachacova, M. Schmitt, J. Novotny, Jr., J. Novotny, M. Majali, and R. Liscak, " A comparison of the gamma knife model C and the automatic positioning system with Leksell model B.," J. Neurosurg. 0022-3085 102, 25-28 (2005).
    • (2005) J. Neurosurg. , vol.102 , pp. 25-28
    • Tlachacova, D.1    Schmitt, M.2    Novotny Jr., J.3    Novotny, J.4    Majali, M.5    Liscak, R.6
  • 10
    • 85027239742 scopus 로고    scopus 로고
    • Leksell Gamma Plan 8.2 Online Reference Manual, ELEKTA Instruments AB, Stockholm, Sweden, Document No. 1005472, Rev.01.
    • Leksell Gamma Plan 8.2 Online Reference Manual, ELEKTA Instruments AB, Stockholm, Sweden, Document No. 1005472, Rev.01, 2007.
    • (2007)
  • 11
    • 0037152478 scopus 로고    scopus 로고
    • Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams
    • 0031-9155 10.1088/0031-9155/47/20/306.
    • G. Bednarz, M. S. Huq, and U. Rosenow, " Deconvolution of detector size effect for output factor measurement for narrow Gamma Knife radiosurgery beams.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/47/20/306 47, 3643-3649 (2002).
    • (2002) Phys. Med. Biol. , vol.47 , pp. 3643-3649
    • Bednarz, G.1    Huq, M.S.2    Rosenow, U.3
  • 12
    • 0036732644 scopus 로고    scopus 로고
    • Precision dosimetry for narrow photon beams used in radiosurgery- Determination of Gamma Knife output factors
    • 0094-2405 10.1118/1.1501138.
    • A. Mack, S. Scheib, J. Major, S. Gianolini, G. Pazmandi, H. Feist, H. Czempiel, and H. J. Kreiner, " Precision dosimetry for narrow photon beams used in radiosurgery-Determination of Gamma Knife output factors.," Med. Phys. 0094-2405 10.1118/1.1501138 29, 2080-2089 (2002).
    • (2002) Med. Phys. , vol.29 , pp. 2080-2089
    • MacK, A.1    Scheib, S.2    Major, J.3    Gianolini, S.4    Pazmandi, G.5    Feist, H.6    Czempiel, H.7    Kreiner, H.J.8
  • 13
    • 0038247942 scopus 로고    scopus 로고
    • Determination of the 4 mm Gamma Knife helmet relative output factor using a variety of detectors
    • 0094-2405 10.1118/1.1567736.
    • J. Tsai, M. J. Rivard, M. J. Engler, J. E. Mignano, D. E. Wazer, and W. A. Schucart, " Determination of the 4 mm Gamma Knife helmet relative output factor using a variety of detectors.," Med. Phys. 0094-2405 10.1118/1.1567736 30, 986-992 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 986-992
    • Tsai, J.1    Rivard, M.J.2    Engler, M.J.3    Mignano, J.E.4    Wazer, D.E.5    Schucart, W.A.6
  • 14
    • 0041387395 scopus 로고    scopus 로고
    • Measurements of Gamma-Knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector
    • 0094-2405 10.1118/1.1587451.
    • F. Araki, T. Ikegami, I. Ishidoya, and H. D. Kubo, " Measurements of Gamma-Knife helmet output factors using a radiophotoluminescent glass rod dosimeter and a diode detector.," Med. Phys. 0094-2405 10.1118/1.1587451 30, 1976-1981 (2003).
    • (2003) Med. Phys. , vol.30 , pp. 1976-1981
    • Araki, F.1    Ikegami, T.2    Ishidoya, I.3    Kubo, H.D.4
  • 15
    • 33847655528 scopus 로고    scopus 로고
    • Measurement of Gamma Knife helmet factors using MOSFETs
    • 0094-2405 10.1118/1.2437282.
    • L. Kurjewicz and A. Berndt, " Measurement of Gamma Knife helmet factors using MOSFETs.," Med. Phys. 0094-2405 10.1118/1.2437282 34, 1007-1012 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 1007-1012
    • Kurjewicz, L.1    Berndt, A.2
  • 16
    • 0032762524 scopus 로고    scopus 로고
    • Monte Carlo calculated output factors of a Leksell Gamma Knife unit
    • 0031-9155,. 10.1088/0031-9155/44/12/401
    • J. Y. C. Cheung, K. N. Yu, C. P. Yu, and R. T. K. Ho, " Monte Carlo calculated output factors of a Leksell Gamma Knife unit.," Phys. Med. Biol. 0031-9155 44, N247-N249 (1999). 10.1088/0031-9155/44/12/401
    • (1999) Phys. Med. Biol. , vol.44
    • Cheung, J.Y.C.1    Yu, K.N.2    Yu, C.P.3    Ho, R.T.K.4
  • 17
    • 0037150458 scopus 로고    scopus 로고
    • Monte Carlo simulation of the Leksell GammaKnife: I. Source modeling and calculation in homogeneous media
    • 0031-9155 10.1088/0031-9155/47/12/301.
    • V. Moskvin, C. DesRosiers, L. Papiez, R. Timmerman, M. Randall, and P. DesRosiers, " Monte Carlo simulation of the Leksell GammaKnife: I. Source modeling and calculation in homogeneous media.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/47/12/301 47, 1995-2011 (2002).
    • (2002) Phys. Med. Biol. , vol.47 , pp. 1995-2011
    • Moskvin, V.1    Desrosiers, C.2    Papiez, L.3    Timmerman, R.4    Randall, M.5    Desrosiers, P.6
  • 18
    • 33845920500 scopus 로고    scopus 로고
    • MCNP-based computational model for the Leksell Gamma Knife
    • 0094-2405 10.1118/1.2401054.
    • J. Trnka, J. Novotny, Jr., and J. Kluson, " MCNP-based computational model for the Leksell Gamma Knife.," Med. Phys. 0094-2405 10.1118/1.2401054 34, 63-75 (2007).
    • (2007) Med. Phys. , vol.34 , pp. 63-75
    • Trnka, J.1    Novotny Jr., J.2    Kluson, J.3
  • 19
    • 0008796868 scopus 로고    scopus 로고
    • Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine
    • 0094-2405 10.1118/1.598407.
    • A. Niroomand-Rad, " Radiochromic film dosimetry: Recommendations of AAPM Radiation Therapy Committee Task Group 55. American Association of Physicists in Medicine.," Med. Phys. 0094-2405 10.1118/1.598407 25, 2093-2115 (1998).
    • (1998) Med. Phys. , vol.25 , pp. 2093-2115
    • Niroomand-Rad, A.1
  • 20
    • 0029909070 scopus 로고    scopus 로고
    • High-resolution dosimetry using radiochromic film and a document scanner
    • 0031-9155 10.1088/0031-9155/41/11/008.
    • M. A. Stevens, J. R. Turner, R. P. Hugtenburg, and P. H. Butler, " High-resolution dosimetry using radiochromic film and a document scanner.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/41/11/008 41, 2357-2365 (1996).
    • (1996) Phys. Med. Biol. , vol.41 , pp. 2357-2365
    • Stevens, M.A.1    Turner, J.R.2    Hugtenburg, R.P.3    Butler, P.H.4
  • 21
    • 0034037793 scopus 로고    scopus 로고
    • Quality assurance for an image-guided frameless radiosurgery system using radiochromic film
    • 0288-2043.
    • H. Shiomi, T. Inoue, and S. Nakamura, " Quality assurance for an image-guided frameless radiosurgery system using radiochromic film.," Radiat. Med. 0288-2043 18, 107-113 (2000).
    • (2000) Radiat. Med. , vol.18 , pp. 107-113
    • Shiomi, H.1    Inoue, T.2    Nakamura, S.3
  • 22
    • 0033010557 scopus 로고    scopus 로고
    • Dosimetry of Gamma Knife and linac-based radiosurgery using radiochromic and diode detectors
    • 0031-9155 10.1088/0031-9155/44/4/006.
    • A. Somigliana, " Dosimetry of Gamma Knife and linac-based radiosurgery using radiochromic and diode detectors.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/44/4/006 44, 887-897 (1999).
    • (1999) Phys. Med. Biol. , vol.44 , pp. 887-897
    • Somigliana, A.1
  • 23
    • 2442701589 scopus 로고    scopus 로고
    • GAFChromic film dosimetry with a flatbed color scanner for Leksell Gamma Knife therapy
    • 0094-2405 10.1118/1.1712393.
    • M. Yamauchi, T. Tominaga, O. Nakamura, R. Ueda, and M. Hoshi, " GAFChromic film dosimetry with a flatbed color scanner for Leksell Gamma Knife therapy.," Med. Phys. 0094-2405 10.1118/1.1712393 31, 1243-1248 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 1243-1248
    • Yamauchi, M.1    Tominaga, T.2    Nakamura, O.3    Ueda, R.4    Hoshi, M.5
  • 24
    • 0000274968 scopus 로고    scopus 로고
    • Verification of dose plans using film dosimetry for quality assurance in radiosurgery
    • 10.1159/000062344
    • A. Mack, G. Mack, D. Weltz, H. Czempiel, and H. J. Kreiner, " Verification of dose plans using film dosimetry for quality assurance in radiosurgery.," Radiosurgery 4, 213-227 (2002). 10.1159/000062344
    • (2002) Radiosurgery , vol.4 , pp. 213-227
    • MacK, A.1    MacK, G.2    Weltz, D.3    Czempiel, H.4    Kreiner, H.J.5
  • 25
    • 33749872451 scopus 로고    scopus 로고
    • Evaluation of the EDR-2 film for relative dosimetry of high-energy photon and electron beams
    • 0144-8420,. 10.1093/rpd/ncj006
    • M. Ahmad, Z. Chen, H. Song, J. Deng, and R. Nath, " Evaluation of the EDR-2 film for relative dosimetry of high-energy photon and electron beams.," Radiat. Prot. Dosim. 0144-8420 120, 159-162 (2006). 10.1093/rpd/ncj006
    • (2006) Radiat. Prot. Dosim. , vol.120 , pp. 159-162
    • Ahmad, M.1    Chen, Z.2    Song, H.3    Deng, J.4    Nath, R.5
  • 26
    • 0031466420 scopus 로고    scopus 로고
    • GafChromic MD-55: Investigated as a precision dosimeter
    • 0094-2405 10.1118/1.598106.
    • N. V. Klassen and L. Van der Zvan, " GafChromic MD-55: Investigated as a precision dosimeter.," Med. Phys. 0094-2405 10.1118/1.598106 24, 1924-1934 (1997).
    • (1997) Med. Phys. , vol.24 , pp. 1924-1934
    • Klassen, N.V.1    Van Der Zvan, L.2
  • 27
    • 34247503351 scopus 로고    scopus 로고
    • Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry
    • 0031-9155 10.1088/0031-9155/52/1/015.
    • L. Paelinck, W. De Neve, and C. De Wagter, " Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry.," Phys. Med. Biol. 0031-9155 10.1088/0031-9155/52/1/015 52, 231-242 (2007).
    • (2007) Phys. Med. Biol. , vol.52 , pp. 231-242
    • Paelinck, L.1    De Neve, W.2    De Wagter, C.3
  • 28
    • 0032727433 scopus 로고    scopus 로고
    • The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry
    • 0094-2405 10.1118/1.598730.
    • J. L. Robar and B. G. Clark, " The use of radiographic film for linear accelerator stereotactic radiosurgical dosimetry.," Med. Phys. 0094-2405 10.1118/1.598730 26, 2144-2150 (1999).
    • (1999) Med. Phys. , vol.26 , pp. 2144-2150
    • Robar, J.L.1    Clark, B.G.2
  • 29
    • 0034088508 scopus 로고    scopus 로고
    • A practical technique for verification of three-dimensional conformal dose distributions in stereotactic radiosurgery
    • 0094-2405 10.1118/1.598962.
    • J. L. Robar and B. G. Clark, " A practical technique for verification of three-dimensional conformal dose distributions in stereotactic radiosurgery.," Med. Phys. 0094-2405 10.1118/1.598962 27, 978-987 (2000).
    • (2000) Med. Phys. , vol.27 , pp. 978-987
    • Robar, J.L.1    Clark, B.G.2
  • 32
    • 3142727652 scopus 로고    scopus 로고
    • EDR2 film dosimetry for IMRT verification using low-energy photon filters
    • 0094-2405 10.1118/1.1760190.
    • I. J. Yeo, A. B. Ardakani, Y. B. Cho, M. Heydarian, T. Zhang, and M. Islam, " EDR2 film dosimetry for IMRT verification using low-energy photon filters.," Med. Phys. 0094-2405 10.1118/1.1760190 31, 1960-1963 (2004).
    • (2004) Med. Phys. , vol.31 , pp. 1960-1963
    • Yeo, I.J.1    Ardakani, A.B.2    Cho, Y.B.3    Heydarian, M.4    Zhang, T.5    Islam, M.6
  • 33
    • 34347377755 scopus 로고    scopus 로고
    • Analysis of the sources of uncertainty for EDR2 film-based IMRT quality assurance
    • 1526-9914 10.1120/jacmv7i2.2230.
    • C. Shi, N. Papanikolaou, Y. Yan, X. Weng, and H. Jiang, " Analysis of the sources of uncertainty for EDR2 film-based IMRT quality assurance.," J. Appl. Clin. Med. Phys. 1526-9914 10.1120/jacmp.v7i2.2230 7, 1-8 (2006).
    • (2006) J. Appl. Clin. Med. Phys. , vol.7 , pp. 1-8
    • Shi, C.1    Papanikolaou, N.2    Yan, Y.3    Weng, X.4    Jiang, H.5
  • 34
    • 33746898695 scopus 로고    scopus 로고
    • Evaluation of Gafchromic EBT film for intensity modulated radiation therapy dose distribution verification
    • A. Sankar, P. G. G. Kurup, V. Murali, K. M. Ayyangar, R. M. Nehru, and J. Velmurugan, " Evaluation of Gafchromic EBT film for intensity modulated radiation therapy dose distribution verification.," Journal of Medical Physics 31, 78-82 (2006).
    • (2006) Journal of Medical Physics , vol.31 , pp. 78-82
    • Sankar, A.1    Kurup, P.G.G.2    Murali, V.3    Ayyangar, K.M.4    Nehru, R.M.5    Velmurugan, J.6
  • 35
    • 33750510893 scopus 로고    scopus 로고
    • Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification
    • 0739-0211,. 10.1016/j.meddos.2006.06.001
    • A. Sankar, K. M. Ayyangar, R. M. Nehru, P. G. G. Kurup, V. Murali, C. A. Enke, and J. Velmurugan, " Comparison of Kodak EDR2 and Gafchromic EBT film for intensity-modulated radiation therapy dose distribution verification.," Med. Dosim. 0739-0211 31, 273-282 (2006). 10.1016/j.meddos.2006.06.001
    • (2006) Med. Dosim. , vol.31 , pp. 273-282
    • Sankar, A.1    Ayyangar, K.M.2    Nehru, R.M.3    Kurup, P.G.G.4    Murali, V.5    Enke, C.A.6    Velmurugan, J.7
  • 37
    • 58149210333 scopus 로고
    • PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter
    • 0168-583X 10.1016/0168-583X(95)00349-5.
    • J. Baro, J. Sempau, J. M. Fernandez-Varea, and F. Salvat, "' PENELOPE: An algorithm for Monte Carlo simulation of the penetration and energy loss of electrons and positrons in matter.," Nucl. Instrum. Methods Phys. Res. B 0168-583X 10.1016/0168-583X(95)00349-5 100, 31-46 (1995).
    • (1995) Nucl. Instrum. Methods Phys. Res. B , vol.100 , pp. 31-46
    • Baro, J.1    Sempau, J.2    Fernandez-Varea, J.M.3    Salvat, F.4
  • 39
    • 63849324254 scopus 로고    scopus 로고
    • Unintended attenuation in the Leksell Gamma Knife Perfexion calibration-phantom adaptor and its effect on dose calibration
    • 10.1118/1.3093240
    • J. P. Bhatnagar, J. Novotny Jr., M. A. Quader, G. Bednarz, and M. S. Huq, " Unintended attenuation in the Leksell Gamma Knife Perfexion calibration-phantom adaptor and its effect on dose calibration.," Med. Phys. 36, 1208-1211 (2009). 10.1118/1.3093240
    • (2009) Med. Phys. , vol.36 , pp. 1208-1211
    • Bhatnagar, J.P.1    Novotny Jr., J.2    Quader, M.A.3    Bednarz, G.4    Huq, M.S.5


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.