Volumn 34, Issue 3, 2008, Pages 146-150

The clinical investigation of the base curve and comfort rate of a new prototype silicone hydrogel contact lens

Author keywords

Comfort; Contact lenses; Lens fitting; Silicone hydrogel

Indexed keywords

ADULT; ARTICLE; CLINICAL TRIAL; CONTACT LENS; CONTROLLED CLINICAL TRIAL; CONTROLLED STUDY; FEMALE; HUMAN; HYDROGEL; MALE; PRIORITY JOURNAL; RANDOMIZED CONTROLLED TRIAL; SCORING SYSTEM; SILICONE PROSTHESIS;

EID: 43249107909     PISSN: 15422321     EISSN: None     Source Type: Journal    
DOI: 10.1097/ICL.0b013e3181506bbd     Document Type: Article
Times cited : (7)

References (38)
  • 1
    • 0024792026 scopus 로고
    • The Glenn A. Fry Award Lecture 1988: The ocular response to contact lens wear
    • Holden BA. The Glenn A. Fry Award Lecture 1988: The ocular response to contact lens wear. Optom Vis Sci 1989;66:717-733.
    • (1989) Optom Vis Sci , vol.66 , pp. 717-733
    • Holden, B.A.1
  • 2
    • 0034351442 scopus 로고    scopus 로고
    • Microcyst response to high Dk/t silicone hydrogel contact lenses
    • Keay L, Sweeney DF, Jalbert I, et al. Microcyst response to high Dk/t silicone hydrogel contact lenses. Optom Vis Sci 2000;77:582-585.
    • (2000) Optom Vis Sci , vol.77 , pp. 582-585
    • Keay, L.1    Sweeney, D.F.2    Jalbert, I.3
  • 3
    • 0035106234 scopus 로고    scopus 로고
    • Hypoxic effects of high Dk soft contact lens wearers are negligible
    • Covey M, Sweeney DF, Terry R, et al. Hypoxic effects of high Dk soft contact lens wearers are negligible. Optom Vis Sci 2001;78:95-99.
    • (2001) Optom Vis Sci , vol.78 , pp. 95-99
    • Covey, M.1    Sweeney, D.F.2    Terry, R.3
  • 4
    • 0030955386 scopus 로고    scopus 로고
    • High-oxygen-transmissibility soft contact lenses do not induce limbal hyperaemia
    • Papas EB, Vajdic CM, Austen R, et al. High-oxygen-transmissibility soft contact lenses do not induce limbal hyperaemia. Curr Eye Res 1997;16:942-948.
    • (1997) Curr Eye Res , vol.16 , pp. 942-948
    • Papas, E.B.1    Vajdic, C.M.2    Austen, R.3
  • 5
    • 33644617061 scopus 로고    scopus 로고
    • Long-term changes in ocular physiology, tear film characteristics and symptomatology with silicone hydrogel contact lens wear
    • Santodomingo-Rubido J, Wolffsohn JS, Gilmartin B. Long-term changes in ocular physiology, tear film characteristics and symptomatology with silicone hydrogel contact lens wear. Optom Vis Sci 2006;83:73-81.
    • (2006) Optom Vis Sci , vol.83 , pp. 73-81
    • Santodomingo-Rubido, J.1    Wolffsohn, J.S.2    Gilmartin, B.3
  • 6
    • 36049019585 scopus 로고    scopus 로고
    • Adverse events and discontinuations over 18 months of silicone hydrogel contact lens wear
    • Santodomingo-Rubido J, Wolffsohn JS, Gilmartin B. Adverse events and discontinuations over 18 months of silicone hydrogel contact lens wear. Eye Contact Lens 2007;33:288-292.
    • (2007) Eye Contact Lens , vol.33 , pp. 288-292
    • Santodomingo-Rubido, J.1    Wolffsohn, J.S.2    Gilmartin, B.3
  • 7
    • 21344462338 scopus 로고    scopus 로고
    • Adverse events and discontinuations with rigid and soft hyper Dk contact lenses used for continuous wear
    • Morgan PB, Efron N, Maldonado-Codina C, et al. Adverse events and discontinuations with rigid and soft hyper Dk contact lenses used for continuous wear. Optom Vis Sci 2005;82:528-535.
    • (2005) Optom Vis Sci , vol.82 , pp. 528-535
    • Morgan, P.B.1    Efron, N.2    Maldonado-Codina, C.3
  • 8
    • 16244405548 scopus 로고    scopus 로고
    • Incidence of keratitis of varying severity among contact lens wearers
    • Morgan PB, Efron N, Hill EA, et al. Incidence of keratitis of varying severity among contact lens wearers. Br J Ophthalmol 2005;89:430-436.
    • (2005) Br J Ophthalmol , vol.89 , pp. 430-436
    • Morgan, P.B.1    Efron, N.2    Hill, E.A.3
  • 9
    • 28344434590 scopus 로고    scopus 로고
    • The incidence of microbial keratitis among wearers of a 30-day silicone hydrogel extended-wear contact lens
    • Schein OD, McNally JJ, Katz J, et al. The incidence of microbial keratitis among wearers of a 30-day silicone hydrogel extended-wear contact lens. Ophthalmology 2005;112:2172-2179.
    • (2005) Ophthalmology , vol.112 , pp. 2172-2179
    • Schein, O.D.1    McNally, J.J.2    Katz, J.3
  • 10
    • 21344432328 scopus 로고    scopus 로고
    • The size, location, and clinical severity of corneal infiltrative events associated with contact lens wear
    • Efron N, Morgan P, Hill EA, et al. The size, location, and clinical severity of corneal infiltrative events associated with contact lens wear. Optom Vis Sci 2005;82:519-527.
    • (2005) Optom Vis Sci , vol.82 , pp. 519-527
    • Efron, N.1    Morgan, P.2    Hill, E.A.3
  • 11
    • 34247281033 scopus 로고    scopus 로고
    • The epidemiology of contact lens related infiltrates
    • Stapleton F, Keay L, Jalbert I, et al. The epidemiology of contact lens related infiltrates. Optom Vis Sci 2007;84:257-272.
    • (2007) Optom Vis Sci , vol.84 , pp. 257-272
    • Stapleton, F.1    Keay, L.2    Jalbert, I.3
  • 12
    • 0035105094 scopus 로고    scopus 로고
    • Superior arcuate lesions with soft contact lens wear
    • Holden BA, Stephenson A, Stretton S, et al. Superior arcuate lesions with soft contact lens wear. Optom Vis Sci 2001;78:9-12.
    • (2001) Optom Vis Sci , vol.78 , pp. 9-12
    • Holden, B.A.1    Stephenson, A.2    Stretton, S.3
  • 13
    • 33644847286 scopus 로고    scopus 로고
    • Two presentations of contact lens-induced papillary conjunctivitis (CLPC) in hydrogel lens wear: Local and general
    • Skotnitsky C, Naduvilath TJ, Sweeney DF, et al. Two presentations of contact lens-induced papillary conjunctivitis (CLPC) in hydrogel lens wear: Local and general. Optom Vis Sci 2006;83:27-36.
    • (2006) Optom Vis Sci , vol.83 , pp. 27-36
    • Skotnitsky, C.1    Naduvilath, T.J.2    Sweeney, D.F.3
  • 14
    • 38049007615 scopus 로고    scopus 로고
    • Conjunctival epithelial flaps with 18 months silicone hydrogel contact lens wear
    • Santodomingo-Rubido J, Wolffsohn JS, Gilmartin B. Conjunctival epithelial flaps with 18 months silicone hydrogel contact lens wear. Eye Contact Lens 2007;34:35-38.
    • (2007) Eye Contact Lens , vol.34 , pp. 35-38
    • Santodomingo-Rubido, J.1    Wolffsohn, J.S.2    Gilmartin, B.3
  • 15
    • 1442323315 scopus 로고
    • Happy patients: Connecting comfort and compliance
    • Schlanger JL, Schwartz CA, Leser E. Happy patients: Connecting comfort and compliance. Spectrum 1993;8:45-47.
    • (1993) Spectrum , vol.8 , pp. 45-47
    • Schlanger, J.L.1    Schwartz, C.A.2    Leser, E.3
  • 16
    • 33750984226 scopus 로고    scopus 로고
    • Prevalence of ocular surface symptoms, signs, and uncomfortable hours of wear in contact lens wearers: The effect of refitting with daily-wear silicone hydrogel lenses (senofilcon)
    • Riley C, Young G, Chalmers R. Prevalence of ocular surface symptoms, signs, and uncomfortable hours of wear in contact lens wearers: The effect of refitting with daily-wear silicone hydrogel lenses (senofilcon). Eye Contact Lens 2006;32:281-286.
    • (2006) Eye Contact Lens , vol.32 , pp. 281-286
    • Riley, C.1    Young, G.2    Chalmers, R.3
  • 17
    • 0033967626 scopus 로고    scopus 로고
    • Responses of contact lens wearers to a dry eye survey
    • Begley CG, Caffery B, Nichols KK, et al. Responses of contact lens wearers to a dry eye survey. Optom Vis Sci 2000;77:40-46.
    • (2000) Optom Vis Sci , vol.77 , pp. 40-46
    • Begley, C.G.1    Caffery, B.2    Nichols, K.K.3
  • 18
    • 0032872146 scopus 로고    scopus 로고
    • The frequency of ocular symptoms during spectacle and daily soft and rigid contact lens wear
    • Vajdic C, Holden BA, Sweeney DF, et al. The frequency of ocular symptoms during spectacle and daily soft and rigid contact lens wear. Optom Vis Sci 1999;76:705-711.
    • (1999) Optom Vis Sci , vol.76 , pp. 705-711
    • Vajdic, C.1    Holden, B.A.2    Sweeney, D.F.3
  • 19
    • 2942672703 scopus 로고    scopus 로고
    • Impact of manufacturing technology and material composition on the clinical performance of hydrogel lenses
    • Maldonado-Codina C, Efron N. Impact of manufacturing technology and material composition on the clinical performance of hydrogel lenses. Optom Vis Sci 2004;81:442-454.
    • (2004) Optom Vis Sci , vol.81 , pp. 442-454
    • Maldonado-Codina, C.1    Efron, N.2
  • 20
    • 0142089777 scopus 로고    scopus 로고
    • In vivo dehydration of silicone hydrogel contact lenses
    • Morgan PB, Efron N. In vivo dehydration of silicone hydrogel contact lenses. Eye Contact Lens 2003;29:173-176.
    • (2003) Eye Contact Lens , vol.29 , pp. 173-176
    • Morgan, P.B.1    Efron, N.2
  • 21
    • 0038678950 scopus 로고    scopus 로고
    • Lysozyme and lipid deposition on silicone hydrogel contact lens materials
    • Jones L, Senchyna M, Glasier MA, et al. Lysozyme and lipid deposition on silicone hydrogel contact lens materials. Eye Contact Lens 2003;29:S75-S79.
    • (2003) Eye Contact Lens , vol.29
    • Jones, L.1    Senchyna, M.2    Glasier, M.A.3
  • 22
    • 0001981525 scopus 로고    scopus 로고
    • Silicone hydrogel materials: How do they work?
    • Sweeney DF, ed, Oxford, Butterworth-Heinemann
    • Tighe B. Silicone hydrogel materials: How do they work? In: Sweeney DF, ed. Silicone Hydrogels: The Rebirth of Continuous Wear Contact Lenses. Oxford, Butterworth-Heinemann, 2000, pp 1-21.
    • (2000) Silicone Hydrogels: The Rebirth of Continuous Wear Contact Lenses , pp. 1-21
    • Tighe, B.1
  • 23
    • 43249122869 scopus 로고    scopus 로고
    • A next-generation silicone hydrogel lens for daily wear
    • Steffen R, Schnider C. A next-generation silicone hydrogel lens for daily wear. Optician 2004;227:10-13.
    • (2004) Optician , vol.227 , pp. 10-13
    • Steffen, R.1    Schnider, C.2
  • 24
    • 33747107008 scopus 로고    scopus 로고
    • Cytotoxicity and antimicrobial activity of six multipurpose soft contact lens disinfecting solutions
    • Santodomingo-Rubido J, Mori O, Kawaminami S. Cytotoxicity and antimicrobial activity of six multipurpose soft contact lens disinfecting solutions. Ophthalmic Physiol Opt 2006;26:476-482.
    • (2006) Ophthalmic Physiol Opt , vol.26 , pp. 476-482
    • Santodomingo-Rubido, J.1    Mori, O.2    Kawaminami, S.3
  • 25
    • 33847189977 scopus 로고    scopus 로고
    • The comparative clinical performance of a new polyhexamethylene biguanide- vs a polyquad-based contact lens regime with two silicone hydrogel contact lenses
    • Santodomingo-Rubido J. The comparative clinical performance of a new polyhexamethylene biguanide- vs a polyquad-based contact lens regime with two silicone hydrogel contact lenses. Ophthalmic Physiol Opt 2007;27:168-173.
    • (2007) Ophthalmic Physiol Opt , vol.27 , pp. 168-173
    • Santodomingo-Rubido, J.1
  • 26
    • 18744391374 scopus 로고    scopus 로고
    • Bulbar conjunctival staining in contact lens wearers and non lens wearers and its association with symptomatology
    • Guillon M, Maissa C. Bulbar conjunctival staining in contact lens wearers and non lens wearers and its association with symptomatology. Contact Lens Anterior Eye 2005;28:67-73.
    • (2005) Contact Lens Anterior Eye , vol.28 , pp. 67-73
    • Guillon, M.1    Maissa, C.2
  • 27
    • 0023008824 scopus 로고
    • Determinants of the initial comfort of hydrogel contact lenses
    • Efron N, Brennan NA, Currie JM, et al. Determinants of the initial comfort of hydrogel contact lenses. Am J Optom Physiol Opt 1986;63:819-823.
    • (1986) Am J Optom Physiol Opt , vol.63 , pp. 819-823
    • Efron, N.1    Brennan, N.A.2    Currie, J.M.3
  • 28
    • 2142835594 scopus 로고    scopus 로고
    • The critical comfort of soft contact lenses
    • McMonnies CW. The critical comfort of soft contact lenses. Clin Exp Optom 1997;80:53-58.
    • (1997) Clin Exp Optom , vol.80 , pp. 53-58
    • McMonnies, C.W.1
  • 29
    • 0036799178 scopus 로고    scopus 로고
    • Effect of lens base curve on subjective comfort and assessment of fit with silicone hydrogel continuous wear contact lenses
    • Dumbleton KA, Chalmers R, McNally J, et al. Effect of lens base curve on subjective comfort and assessment of fit with silicone hydrogel continuous wear contact lenses. Optom Vis Sci 2002;79:633-637.
    • (2002) Optom Vis Sci , vol.79 , pp. 633-637
    • Dumbleton, K.A.1    Chalmers, R.2    McNally, J.3
  • 30
    • 0021123317 scopus 로고
    • Critical oxygen levels to avoid corneal edema for daily and extended wear contact lenses
    • Holden BA, Mertz GW. Critical oxygen levels to avoid corneal edema for daily and extended wear contact lenses. Invest Ophthalmol Vis Sci 1984;25:1161-1167.
    • (1984) Invest Ophthalmol Vis Sci , vol.25 , pp. 1161-1167
    • Holden, B.A.1    Mertz, G.W.2
  • 31
    • 0032843148 scopus 로고    scopus 로고
    • Re-evaluation of the oxygen diffusion model for predicting minimum contact lens Dk/t values needed to avoid corneal anoxia
    • Harvitt DM, Bonanno JA. Re-evaluation of the oxygen diffusion model for predicting minimum contact lens Dk/t values needed to avoid corneal anoxia. Optom Vis Sci 1999;76:712-719.
    • (1999) Optom Vis Sci , vol.76 , pp. 712-719
    • Harvitt, D.M.1    Bonanno, J.A.2
  • 33
    • 10644257886 scopus 로고    scopus 로고
    • Short-term physiologic response in neophyte subjects fitted with hydrogel and silicone hydrogel contact lenses
    • Maldonado-Codina C, Morgan PB, Schnider CM, et al. Short-term physiologic response in neophyte subjects fitted with hydrogel and silicone hydrogel contact lenses. Optom Vis Sci 2004;81:911-921.
    • (2004) Optom Vis Sci , vol.81 , pp. 911-921
    • Maldonado-Codina, C.1    Morgan, P.B.2    Schnider, C.M.3
  • 34
    • 34247113209 scopus 로고    scopus 로고
    • A 12-month prospective clinical trial of comfilcon A silicone-hydrogel contact lenses worn on a 30-day continuous wear basis
    • Brennan NA, Chantal Coles M-L, Conner HRM, et al. A 12-month prospective clinical trial of comfilcon A silicone-hydrogel contact lenses worn on a 30-day continuous wear basis. Contact Lens Anterior Eye 2007;30:108-118.
    • (2007) Contact Lens Anterior Eye , vol.30 , pp. 108-118
    • Brennan, N.A.1    Chantal Coles, M.-L.2    Conner, H.R.M.3
  • 35
    • 0035999965 scopus 로고    scopus 로고
    • A comparison of three different scales for rating contact lens handling
    • du Toit R, Pritchard N, Heffernan S, et al. A comparison of three different scales for rating contact lens handling. Optom Vis Sci 2002;79:313-320.
    • (2002) Optom Vis Sci , vol.79 , pp. 313-320
    • du Toit, R.1    Pritchard, N.2    Heffernan, S.3
  • 36
    • 33847103565 scopus 로고    scopus 로고
    • Use of silicone hydrogel material for daily wear
    • Guillon M, Maissa C. Use of silicone hydrogel material for daily wear. Contact Lens Anterior Eye 2007;30:5-10.
    • (2007) Contact Lens Anterior Eye , vol.30 , pp. 5-10
    • Guillon, M.1    Maissa, C.2
  • 37
    • 43249093641 scopus 로고    scopus 로고
    • A new interference-based in vitro technique for the wettability assessment of soft contact lenses
    • Poster presented at: May, Manchester, United Kingdom
    • Maruyama K, Yokoi N, Kinoshita S. A new interference-based in vitro technique for the wettability assessment of soft contact lenses. Poster presented at: 31st British Contact Lens Association Conference; May 2007; Manchester, United Kingdom.
    • (2007) 31st British Contact Lens Association Conference
    • Maruyama, K.1    Yokoi, N.2    Kinoshita, S.3
  • 38
    • 34247220326 scopus 로고    scopus 로고
    • Targeting contact lens induced dryness and discomfort: What properties will make lenses more comfortable
    • Fonn D. Targeting contact lens induced dryness and discomfort: What properties will make lenses more comfortable. Optom Vis Sci 2007;84:279-285.
    • (2007) Optom Vis Sci , vol.84 , pp. 279-285
    • Fonn, D.1


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.