Volumn 25, Issue 5, 2005, Pages 436-440

Comparison of the ICare® rebound tonometer with the Goldmann tonometer in a normal population

Author keywords

Applanation tonometry; Impact tonometry; Induction; IOP; Non contact tonometry; Rebound tonometer

Indexed keywords

ADULT; ARTICLE; CONTROLLED STUDY; CORNEA; EQUIPMENT DESIGN; FEMALE; HUMAN; INTERMETHOD COMPARISON; INTRAOCULAR PRESSURE; MALE; NORMAL HUMAN; PRIORITY JOURNAL; SLIT LAMP; STATISTICAL ANALYSIS; TONOMETER; TONOMETRY; VISUAL SYSTEM EXAMINATION; ADOLESCENT; COMPARATIVE STUDY; EVALUATION; INSTRUMENTATION; OCULOPLETHYSMOGRAPHY; REFERENCE VALUE; REPRODUCIBILITY;

EID: 28244452880     PISSN: 02755408     EISSN: None     Source Type: Journal    
DOI: 10.1111/j.1475-1313.2005.00327.x     Document Type: Article
Times cited : (146)

References (25)
  • 2
    • 84908219488 scopus 로고
    • Statistical methods for assessing agreement between two methods of clinical measurement
    • Bland, J. M. and Altman, D. G. (1986) Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1, 307-310.
    • (1986) Lancet , vol.1 , pp. 307-310
    • Bland, J.M.1    Altman, D.G.2
  • 3
    • 0028791287 scopus 로고
    • Comparing methods of measurement: Why plotting difference against standard method is misleading
    • Bland, J. M. and Altman, D. G. (1995) Comparing methods of measurement: why plotting difference against standard method is misleading. Lancet 346, 1085-1087.
    • (1995) Lancet , vol.346 , pp. 1085-1087
    • Bland, J.M.1    Altman, D.G.2
  • 5
    • 0031042751 scopus 로고    scopus 로고
    • Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer
    • Cho, P. and Lui, T. (1997) Comparison of the performance of the Nidek NT-2000 noncontact tonometer with the Keeler Pulsair 2000 and the Goldmann applanation tonometer. Optom. Vis. Sci 74, 51-58.
    • (1997) Optom. Vis. Sci. , vol.74 , pp. 51-58
    • Cho, P.1    Lui, T.2
  • 8
    • 0036946949 scopus 로고    scopus 로고
    • Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer
    • Goldblum, D., Kontiola, A. L, Mittag, T., Chen, B. and Danias, J. (2002) Non-invasive determination of intraocular pressure in the rat eye. Comparison of an electronic tonometer (TonoPen), and a rebound (impact probe) tonometer. Graefes Arch. Clin. Exp. Ophthalmol. 240, 942-946.
    • (2002) Graefes Arch. Clin. Exp. Ophthalmol. , vol.240 , pp. 942-946
    • Goldblum, D.1    Kontiola, A.L.2    Mittag, T.3    Chen, B.4    Danias, J.5
  • 9
    • 11344262810 scopus 로고    scopus 로고
    • Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma
    • Horowitz, G. S., Byles, J., Lee, J. and D'Este, C. (2004) Comparison of the Tono-Pen and Goldmann tonometer for measuring intraocular pressure in patients with glaucoma. Clin. Experiment. Ophthalmol. 32, 584-589.
    • (2004) Clin. Experiment. Ophthalmol. , vol.32 , pp. 584-589
    • Horowitz, G.S.1    Byles, J.2    Lee, J.3    D'Este, C.4
  • 12
    • 0034115306 scopus 로고    scopus 로고
    • A new induction-based impact method for measuring intraocular pressure
    • Kontiola, A. I. (2000) A new induction-based impact method for measuring intraocular pressure. Acta Ophthalmol. Scand. 78, 142-145.
    • (2000) Acta Ophthalmol. Scand. , vol.78 , pp. 142-145
    • Kontiola, A.I.1
  • 13
    • 1142299596 scopus 로고    scopus 로고
    • Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic
    • Kontiola, A. and Puska, P. (2004) Measuring intraocular pressure with the Pulsair 3000 and Rebound tonometers in elderly patients without an anesthetic. Graefes Arch. Clin. Exp. Ophthalmol. 242, 3-7.
    • (2004) Graefes Arch. Clin. Exp. Ophthalmol. , vol.242 , pp. 3-7
    • Kontiola, A.1    Puska, P.2
  • 14
    • 0035704347 scopus 로고    scopus 로고
    • The induction/impact tonometer: A new instrument to measure intraocular pressure in the rat
    • Kontiola, A. L, Goldblum, D., Mittag, T. and Danias, J. (2001) The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat. Exp. Eye. Res. 73, 781-785.
    • (2001) Exp. Eye. Res. , vol.73 , pp. 781-785
    • Kontiola, A.L.1    Goldblum, D.2    Mittag, T.3    Danias, J.4
  • 16
    • 3543128942 scopus 로고    scopus 로고
    • Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen
    • Li, J., Herndon, L. W., Asrani, S. G., Stinnett, S. and Allingham, R. R. (2004) Clinical comparison of the Proview eye pressure monitor with the Goldmann applanation tonometer and the Tonopen. Arch Ophthalmol. 122, 1117-1121.
    • (2004) Arch. Ophthalmol. , vol.122 , pp. 1117-1121
    • Li, J.1    Herndon, L.W.2    Asrani, S.G.3    Stinnett, S.4    Allingham, R.R.5
  • 18
    • 20444472031 scopus 로고    scopus 로고
    • Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry
    • Losch, A., Scheuerle, A., Rupp, V., Auffarth, G. and Becker, M. (2005) Transpalpebral measurement of intraocular pressure using the TGDc-01 tonometer versus standard Goldmann applanation tonometry. Graefes Arch. Clin. Exp. Ophthalmol. 243, 313-316.
    • (2005) Graefes Arch. Clin. Exp. Ophthalmol. , vol.243 , pp. 313-316
    • Losch, A.1    Scheuerle, A.2    Rupp, V.3    Auffarth, G.4    Becker, M.5
  • 19
    • 0035173876 scopus 로고    scopus 로고
    • Twenty-four-hour time course of intra-ocular pressure in healthy and glaucomatous Africans: Relation to sleep patterns
    • Noel, C., Kabo, A. M., Romanet, J. P., Montmayeur, A. and Buguet, A. (2001) Twenty-four-hour time course of intra-ocular pressure in healthy and glaucomatous Africans: relation to sleep patterns. Ophthalmology 108, 139-144.
    • (2001) Ophthalmology , vol.108 , pp. 139-144
    • Noel, C.1    Kabo, A.M.2    Romanet, J.P.3    Montmayeur, A.4    Buguet, A.5
  • 20
    • 0030734884 scopus 로고    scopus 로고
    • The diurnal variation of intraocular pressure in non-glaucomatous subjects: Relevance in a clinical context
    • Pointer, J. S. (1997) The diurnal variation of intraocular pressure in non-glaucomatous subjects: relevance in a clinical context. Ophthalmic Physiol. Opt. 17, 456-465.
    • (1997) Ophthalmic Physiol. Opt. , vol.17 , pp. 456-465
    • Pointer, J.S.1
  • 21
    • 0032822190 scopus 로고    scopus 로고
    • Human intraocular pressure and its diurnal variation in healthy subjects
    • Pointer, J. S. (1999) Human intraocular pressure and its diurnal variation in healthy subjects. Ophthalmic Physiol. Opt. 19 (Suppl. 2), S43-S48.
    • (1999) Ophthalmic Physiol. Opt. , vol.19 , Issue.SUPPL. 2
    • Pointer, J.S.1
  • 22
    • 15244345047 scopus 로고    scopus 로고
    • Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intra-ocular pressure readings
    • Rai, S., Moster, M. R., Kesen, M., Fontanarosa, J., Spaeth, G. L., Steinmann, W. C., Wilson, R., Myers, J. and Henderer, J. (2005) Level of disagreement between Proview phosphene tonometer and Goldmann applanation tonometer intra-ocular pressure readings. J. Glaucoma 14, 120-123.
    • (2005) J. Glaucoma , vol.14 , pp. 120-123
    • Rai, S.1    Moster, M.R.2    Kesen, M.3    Fontanarosa, J.4    Spaeth, G.L.5    Steinmann, W.C.6    Wilson, R.7    Myers, J.8    Henderer, J.9
  • 23
    • 0017856159 scopus 로고
    • The accuracy of clinical applanation tonometry
    • Thorburn, W. (1978) The accuracy of clinical applanation tonometry. Acta Ophthalmol. (Copenh) 56, 1-5.
    • (1978) Acta Ophthalmol. (Copenh.) , vol.56 , pp. 1-5
    • Thorburn, W.1
  • 24
    • 0027291601 scopus 로고
    • Sources of error with use of Goldmann-type tonometers
    • Whitacre, M. M. and Stein, R. (1993) Sources of error with use of Goldmann-type tonometers. Surv. Ophthalmol. 38, 1-30.
    • (1993) Surv. Ophthalmol. , vol.38 , pp. 1-30
    • Whitacre, M.M.1    Stein, R.2


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.