Time-resolved PIV technique for high temporal resolution measurement of mechanical prosthetic aortic valve fluid dynamics

Author keywords

Biofluid dynamics; Flow visualization; Particle image velocimetry; Prosthetic heart valves

Indexed keywords

FLOW FIELDS; FLOW VISUALIZATION; HEART VALVE PROSTHESES; TIME DOMAIN ANALYSIS; VALVES (MECHANICAL); VELOCIMETERS; VELOCITY;

EID: 34147095518     PISSN: 03913988     EISSN: None     Source Type: Journal    
DOI: 10.1177/039139880703000210     Document Type: Article
Times cited : (12)

References (52)
  • 1
    • 0025614529 scopus 로고
    • Estimation of turbulent shear stresses in pulsatile flow immediately downstream of two artificial aortic valves in vitro
    • Nygaard H, Giersiepen M, Hasenkam JM, Westphal D, Paulsen PK, Reul H. Estimation of turbulent shear stresses in pulsatile flow immediately downstream of two artificial aortic valves in vitro. J Biomech 1990; 23: 1231-8.
    • (1990) J Biomech , vol.23 , pp. 1231-1238
    • Nygaard, H.1    Giersiepen, M.2    Hasenkam, J.M.3    Westphal, D.4    Paulsen, P.K.5    Reul, H.6
  • 2
    • 0021685904 scopus 로고
    • Human red blood cell hemolysis in a turbulent shear flow: Contribution of Reynolds shear stresses
    • Sallam AM, Hwang NH. Human red blood cell hemolysis in a turbulent shear flow: contribution of Reynolds shear stresses. Biorheology 1984; 21: 783-97.
    • (1984) Biorheology , vol.21 , pp. 783-797
    • Sallam, A.M.1    Hwang, N.H.2
  • 3
    • 0030152830 scopus 로고    scopus 로고
    • Identification of peak stresses in cardiac prostheses. A comparison of two-dimensional versus three-dimensional principal stress analysis
    • Fontaine AA, Ellis JT, Healy TM, Hopmeyer J, Yoganathan AP. Identification of peak stresses in cardiac prostheses. A comparison of two-dimensional versus three-dimensional principal stress analysis. ASAIO J 1996; 42: 154-63.
    • (1996) ASAIO J , vol.42 , pp. 154-163
    • Fontaine, A.A.1    Ellis, J.T.2    Healy, T.M.3    Hopmeyer, J.4    Yoganathan, A.P.5
  • 4
    • 0018200398 scopus 로고
    • The Bjork-Shiley aortic prosthesis: Flow characteristics, thrombus formation and tissue overgrowth
    • Yoganathan AP, Corcoran WH, Harrison EC, Carl JR. The Bjork-Shiley aortic prosthesis: flow characteristics, thrombus formation and tissue overgrowth. Circulation 1978; 58: 70-6.
    • (1978) Circulation , vol.58 , pp. 70-76
    • Yoganathan, A.P.1    Corcoran, W.H.2    Harrison, E.C.3    Carl, J.R.4
  • 5
    • 0025082325 scopus 로고
    • Estimation of shear stress related blood damage in heart valve prostheses: In vitro comparison of 25 aortic valves
    • Giersiepen M, Wurzinger LJ, Opitz R, Reul H. Estimation of shear stress related blood damage in heart valve prostheses: in vitro comparison of 25 aortic valves. Int J Artif Organs 1990; 13: 300-6.
    • (1990) Int J Artif Organs , vol.13 , pp. 300-306
    • Giersiepen, M.1    Wurzinger, L.J.2    Opitz, R.3    Reul, H.4
  • 6
    • 0034791973 scopus 로고    scopus 로고
    • Pulsatile flow studies of a porcine bioprosthetic aortic valve in vitro: PIV measurements and shear-induced blood damage
    • Lim WL, Chew YT, Chew TC, Low HT. Pulsatile flow studies of a porcine bioprosthetic aortic valve in vitro: PIV measurements and shear-induced blood damage. J Biomech 2001; 34: 1417-27.
    • (2001) J Biomech , vol.34 , pp. 1417-1427
    • Lim, W.L.1    Chew, Y.T.2    Chew, T.C.3    Low, H.T.4
  • 7
    • 0033213797 scopus 로고    scopus 로고
    • A discussion on the threshold limit for hemolysis related to Reynolds shear stress
    • Grigioni M, Daniele C, D'Avenio G, Barbaro V. A discussion on the threshold limit for hemolysis related to Reynolds shear stress. J Biomech 1999; 32: 1107-12.
    • (1999) J Biomech , vol.32 , pp. 1107-1112
    • Grigioni, M.1    Daniele, C.2    D'Avenio, G.3    Barbaro, V.4
  • 9
    • 0031464974 scopus 로고    scopus 로고
    • 19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part I: Velocity profiles)
    • Barbaro V, Grigioni M, Daniele C, D'Avenio G, Boccanera G. 19 mm sized bileaflet valve prostheses' flow field investigated by bidimensional laser Doppler anemometry (part I: velocity profiles). Int J Artif Organs 1997; 20: 622-8.
    • (1997) Int J Artif Organs , vol.20 , pp. 622-628
    • Barbaro, V.1    Grigioni, M.2    Daniele, C.3    D'Avenio, G.4    Boccanera, G.5
  • 10
    • 0027703644 scopus 로고
    • Determination of principal Reynolds stresses in pulsatile flows after elliptical filtering of discrete velocity measurements
    • Baldwin JT, Deutsch S, Petrie, HL, Tarbell JM. Determination of principal Reynolds stresses in pulsatile flows after elliptical filtering of discrete velocity measurements. J Biomech Eng 1993; 115: 396-403.
    • (1993) J Biomech Eng , vol.115 , pp. 396-403
    • Baldwin, J.T.1    Deutsch, S.2    Petrie, H.L.3    Tarbell, J.M.4
  • 11
    • 0034132947 scopus 로고    scopus 로고
    • Laser Doppler anemometry study of bidimensional flows downstream of three 19-mm bileaflet valves in the mitral position, under kinematic similarity
    • Grigioni M, Daniele C, D'Avenio G, Barbaro V. Laser Doppler anemometry study of bidimensional flows downstream of three 19-mm bileaflet valves in the mitral position, under kinematic similarity. Ann Biomed Eng 2000; 28: 194-203.
    • (2000) Ann Biomed Eng , vol.28 , pp. 194-203
    • Grigioni, M.1    Daniele, C.2    D'Avenio, G.3    Barbaro, V.4
  • 12
    • 0022526420 scopus 로고
    • In vitro velocity measurements down stream from the Ionescu-Shiley aortic bioprosthesis in steady and pulsatile flow
    • Hanle DD, Harrison EC, Yoganathan AP, Corcoran WH. In vitro velocity measurements down stream from the Ionescu-Shiley aortic bioprosthesis in steady and pulsatile flow. Med Biol Eng Comput 1986; 24: 449-59.
    • (1986) Med Biol Eng Comput , vol.24 , pp. 449-459
    • Hanle, D.D.1    Harrison, E.C.2    Yoganathan, A.P.3    Corcoran, W.H.4
  • 13
    • 0027967939 scopus 로고
    • In vitro fluid dynamic evaluation of the Carbomedics bileaflet heart valve prosthesis in the aortic and mitral positions
    • Sung HW, Cape EG, Yoganathan AP. In vitro fluid dynamic evaluation of the Carbomedics bileaflet heart valve prosthesis in the aortic and mitral positions. J Heart Valve Dis 1994; 3: 673-83.
    • (1994) J Heart Valve Dis , vol.3 , pp. 673-683
    • Sung, H.W.1    Cape, E.G.2    Yoganathan, A.P.3
  • 14
    • 0029828882 scopus 로고    scopus 로고
    • Ellis JT, Healy TM, Fontaine AA, Weston MW, Jarret CA, Saxena R, Yoganathan AP. An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves. J Heart Valve Dis 1996; 5: 600-6.
    • Ellis JT, Healy TM, Fontaine AA, Weston MW, Jarret CA, Saxena R, Yoganathan AP. An in vitro investigation of the retrograde flow fields of two bileaflet mechanical heart valves. J Heart Valve Dis 1996; 5: 600-6.
  • 15
    • 12044260218 scopus 로고
    • Particle-imaging techniques for experimental fluid mechanics
    • Adrian RJ. Particle-imaging techniques for experimental fluid mechanics. Ann Rev Fluid Mech 1991; 23: 261-304.
    • (1991) Ann Rev Fluid Mech , vol.23 , pp. 261-304
    • Adrian, R.J.1
  • 16
    • 13544259955 scopus 로고    scopus 로고
    • Innovative technologies for the assessment of cardiovascular medical devices: State of the art techniques for artificial heart valves testing
    • Grigioni M, Daniele C, D'Avenio G, Morbiducci U, Del Gaudio C, Abbate M, Di Meo D. Innovative technologies for the assessment of cardiovascular medical devices: state of the art techniques for artificial heart valves testing. Expert Rev Med Devices 2004; 1: 89-101.
    • (2004) Expert Rev Med Devices , vol.1 , pp. 89-101
    • Grigioni, M.1    Daniele, C.2    D'Avenio, G.3    Morbiducci, U.4    Del Gaudio, C.5    Abbate, M.6    Di Meo, D.7
  • 17
    • 0001075219 scopus 로고
    • Effect of resolution on the speed and accuracy of particle image velocimetry interrogation
    • Prasad AK, Adrian RJ, Landreth CC, Offutt PW. Effect of resolution on the speed and accuracy of particle image velocimetry interrogation. Exp Fluids 1992; 13: 105-16.
    • (1992) Exp Fluids , vol.13 , pp. 105-116
    • Prasad, A.K.1    Adrian, R.J.2    Landreth, C.C.3    Offutt, P.W.4
  • 18
    • 1142264056 scopus 로고    scopus 로고
    • Cardiac valve prosthesis flow performances measured by 2D and 3D-stereo particle image velocimetry
    • Marassi M, Castellini P, Pinotti M, Scalise L. Cardiac valve prosthesis flow performances measured by 2D and 3D-stereo particle image velocimetry. Exp Fluids 2004; 36: 176-86.
    • (2004) Exp Fluids , vol.36 , pp. 176-186
    • Marassi, M.1    Castellini, P.2    Pinotti, M.3    Scalise, L.4
  • 21
    • 0015580040 scopus 로고
    • Pressure and flow generated by the left ventricule against different impedances
    • Westerhof N, Elzinga G. Pressure and flow generated by the left ventricule against different impedances. Circ Res 1973; 32: 178-86.
    • (1973) Circ Res , vol.32 , pp. 178-186
    • Westerhof, N.1    Elzinga, G.2
  • 22
    • 0033490803 scopus 로고    scopus 로고
    • Development of a flow feedback pulse duplicator system with rhesus monkey arterial input impedance characteristics
    • Jeremy D, Schaub JD, Koenig SC, Schroeder MJ, Ewert DL, Drew GA, Swope RD. Development of a flow feedback pulse duplicator system with rhesus monkey arterial input impedance characteristics. ASAIO J 1999; 45: 334-9.
    • (1999) ASAIO J , vol.45 , pp. 334-339
    • Jeremy, D.1    Schaub, J.D.2    Koenig, S.C.3    Schroeder, M.J.4    Ewert, D.L.5    Drew, G.A.6    Swope, R.D.7
  • 23
    • 1142272169 scopus 로고
    • Principle of fluid mechanics applied to some situations in the human circulation and particularly to testing of valves in a pulse duplicator
    • Temple LJ, Serafin R, Calvert NG, Drabble JM. Principle of fluid mechanics applied to some situations in the human circulation and particularly to testing of valves in a pulse duplicator. Thorax 1964; 19: 261-7.
    • (1964) Thorax , vol.19 , pp. 261-267
    • Temple, L.J.1    Serafin, R.2    Calvert, N.G.3    Drabble, J.M.4
  • 25
    • 24144462480 scopus 로고    scopus 로고
    • Symmetric phase only filtering: A new paradigm for DPIV data processing
    • Wernet M. Symmetric phase only filtering: a new paradigm for DPIV data processing. Meas Sci Technol 2005; 16: 601-18.
    • (2005) Meas Sci Technol , vol.16 , pp. 601-618
    • Wernet, M.1
  • 26
    • 33646859465 scopus 로고    scopus 로고
    • Chronology of Interpolation: From Ancient Astronomy to Modern Signal and Image Processing
    • Meijering E. A Chronology of Interpolation: From Ancient Astronomy to Modern Signal and Image Processing. Proceedings of IEEE, 2002; 90: 319-42.
    • (2002) Proceedings of IEEE , vol.90 , pp. 319-342
    • Meijering, E.A.1
  • 27
    • 0036144446 scopus 로고    scopus 로고
    • Iterative image deformation methods in PIV
    • Scarano F. Iterative image deformation methods in PIV, Meas Sci Technol 2002; 13: R1-R19.
    • (2002) Meas Sci Technol , vol.13
    • Scarano, F.1
  • 29
    • 33846219390 scopus 로고    scopus 로고
    • Advanced PIV algorithms with Image Distortion Validation and Comparison using Synthetic Images of Turbulent Flow
    • Workshop
    • Lecordier B, Trinité M. Advanced PIV algorithms with Image Distortion Validation and Comparison using Synthetic Images of Turbulent Flow, Proceedings of the Euro PIV 2 Workshop 2003.
    • (2003) Proceedings of the Euro PIV , vol.2
    • Lecordier, B.1    Trinité, M.2
  • 30
    • 34147113622 scopus 로고    scopus 로고
    • Thévenaz P, Blu T, Unser M. Image interpolation and resampling. In: Bankman IN, ed. Handbook of medical imaging, processing & analysis. San Diego, CA: Academic Press; 2000. p 393-420.
    • Thévenaz P, Blu T, Unser M. Image interpolation and resampling. In: Bankman IN, ed. Handbook of medical imaging, processing & analysis. San Diego, CA: Academic Press; 2000. p 393-420.
  • 31
    • 34147112069 scopus 로고    scopus 로고
    • Kaminsky R, Kallweit S, Simons AP, Weber HJ. PIV Measurements with high Temporal Resolution behind Artificial Heart Valves. Proceedings of 6th International Symposium on Particle Image Velocimetry, Sept 21-23 2005, Pasadena, California.
    • Kaminsky R, Kallweit S, Simons AP, Weber HJ. PIV Measurements with high Temporal Resolution behind Artificial Heart Valves. Proceedings of 6th International Symposium on Particle Image Velocimetry, Sept 21-23 2005, Pasadena, California.
  • 32
    • 0033621716 scopus 로고    scopus 로고
    • Experimental investigation of the steady flow downstream of the St. Jude bileaflet heart valve: A comparison between laser Doppler velocimetry and particle image velocimetry techniques
    • Browne P, Ramuzat A, Saxena R, Yoganathan AP. Experimental investigation of the steady flow downstream of the St. Jude bileaflet heart valve: a comparison between laser Doppler velocimetry and particle image velocimetry techniques. Ann Biomed Eng 2000; 28: 39-47.
    • (2000) Ann Biomed Eng , vol.28 , pp. 39-47
    • Browne, P.1    Ramuzat, A.2    Saxena, R.3    Yoganathan, A.P.4
  • 33
    • 0036646795 scopus 로고    scopus 로고
    • Unsteady flow through a new mechanical heart valve prosthesis analysed by digital particle image velocimetry
    • Brucker C, Steinseifer U, Schroder W, Reul H. Unsteady flow through a new mechanical heart valve prosthesis analysed by digital particle image velocimetry. Meas Sci Technol 2002; 13: 1043-9.
    • (2002) Meas Sci Technol , vol.13 , pp. 1043-1049
    • Brucker, C.1    Steinseifer, U.2    Schroder, W.3    Reul, H.4
  • 34
    • 0033774274 scopus 로고    scopus 로고
    • Particle image velocimetry investigation of intravalvular flow fields of a bileaflet mechanical heart valve in a pulsatile flow
    • Subramanian A, Mu H, Kadambi JR, Wernet MP, Brendzel AM, Harasaki H. Particle image velocimetry investigation of intravalvular flow fields of a bileaflet mechanical heart valve in a pulsatile flow. J Heart Valve Dis 2000; 9: 721-31.
    • (2000) J Heart Valve Dis , vol.9 , pp. 721-731
    • Subramanian, A.1    Mu, H.2    Kadambi, J.R.3    Wernet, M.P.4    Brendzel, A.M.5    Harasaki, H.6
  • 35
    • 0043236252 scopus 로고    scopus 로고
    • Regurgitant flow field characteristics of the St. Jude bileaflet mechanical heart valve under physiologic pulsatile flow using particle image velocimetry
    • Manning KB, Kini V, Fontaine AA, Deutsch S, Tarbell JM. Regurgitant flow field characteristics of the St. Jude bileaflet mechanical heart valve under physiologic pulsatile flow using particle image velocimetry. Artif Organs 2003; 27: 840-6.
    • (2003) Artif Organs , vol.27 , pp. 840-846
    • Manning, K.B.1    Kini, V.2    Fontaine, A.A.3    Deutsch, S.4    Tarbell, J.M.5
  • 36
    • 0035117322 scopus 로고    scopus 로고
    • Integrating particle image velocimetry and laser Doppler velocimetry measurements of the regurgitant flow field past mechanical heart valves
    • Kini V, Bachmann C, Fontaine AA, Deutsch S, Tarbell JM. Integrating particle image velocimetry and laser Doppler velocimetry measurements of the regurgitant flow field past mechanical heart valves. Artif Organs 2001; 25: 136-45.
    • (2001) Artif Organs , vol.25 , pp. 136-145
    • Kini, V.1    Bachmann, C.2    Fontaine, A.A.3    Deutsch, S.4    Tarbell, J.M.5
  • 37
    • 14344256283 scopus 로고    scopus 로고
    • Time-resolved DPIV analysis of vortex dynamics in a left ventricular model through bileaflet mechanical and porcine heart valve prostheses
    • Pierrakos O, Vlachos PP, Telionis DP Time-resolved DPIV analysis of vortex dynamics in a left ventricular model through bileaflet mechanical and porcine heart valve prostheses. J Biomech Eng 2004; 126: 714-26.
    • (2004) J Biomech Eng , vol.126 , pp. 714-726
    • Pierrakos, O.1    Vlachos, P.P.2    Telionis, D.P.3
  • 38
    • 26844497231 scopus 로고    scopus 로고
    • Time-resolved particle image velocimetry and laser doppler anemometry study of the turbulent flow field of bileaflet mechanical mitral prostheses
    • Akutsu T, Fukuda T. Time-resolved particle image velocimetry and laser doppler anemometry study of the turbulent flow field of bileaflet mechanical mitral prostheses. J Artif Organs 2005; 8: 171-83.
    • (2005) J Artif Organs , vol.8 , pp. 171-183
    • Akutsu, T.1    Fukuda, T.2
  • 40
    • 29844458743 scopus 로고    scopus 로고
    • Flow in Prosthetic Heart Valves: State-of-the-Art and Future Directions
    • Yoghanatan AP, Chandran KB, Sotiropoulos F. Flow in Prosthetic Heart Valves: State-of-the-Art and Future Directions. Ann Biomed Eng 2005: 33: 1689-94.
    • (2005) Ann Biomed Eng , vol.33 , pp. 1689-1694
    • Yoghanatan, A.P.1    Chandran, K.B.2    Sotiropoulos, F.3
  • 43
    • 18744405111 scopus 로고    scopus 로고
    • Three-dimensional numerical simulation of flow through an aortic bileaflet valve in a realistic model of aortic root
    • Grigioni M, Daniele C, Del Gaudio C, Morbiducci U, Balducci A, D'Avenio G, Barbaro V. Three-dimensional numerical simulation of flow through an aortic bileaflet valve in a realistic model of aortic root. ASAIO J 2005; 51: 176-83.
    • (2005) ASAIO J , vol.51 , pp. 176-183
    • Grigioni, M.1    Daniele, C.2    Del Gaudio, C.3    Morbiducci, U.4    Balducci, A.5    D'Avenio, G.6    Barbaro, V.7
  • 44
    • 0142169840 scopus 로고    scopus 로고
    • Numerical simulation of flow in mechanical heart valves: Grid resolution and the assumption of flow symmetry
    • Ge L, Jones SC, Sotiropoulos F, Healy TM, Yoganathan AP. Numerical simulation of flow in mechanical heart valves: grid resolution and the assumption of flow symmetry. J Biomech Eng 2003; 125: 709-18.
    • (2003) J Biomech Eng , vol.125 , pp. 709-718
    • Ge, L.1    Jones, S.C.2    Sotiropoulos, F.3    Healy, T.M.4    Yoganathan, A.P.5
  • 45
    • 0034176164 scopus 로고    scopus 로고
    • Turbulence characteristics downstream of bileaflet aortic valve prostheses
    • Liu JS, Lu PC, Chu SH. Turbulence characteristics downstream of bileaflet aortic valve prostheses. J Biomech Eng 2000; 122: 118-24.
    • (2000) J Biomech Eng , vol.122 , pp. 118-124
    • Liu, J.S.1    Lu, P.C.2    Chu, S.H.3
  • 47
    • 0014405754 scopus 로고
    • Fluid mechanics of the aortic root with application to coronary flow
    • Bellhouse BJ, Bellhouse FH, Reid KG. Fluid mechanics of the aortic root with application to coronary flow. Nature 1968; 219: 1059-61.
    • (1968) Nature , vol.219 , pp. 1059-1061
    • Bellhouse, B.J.1    Bellhouse, F.H.2    Reid, K.G.3
  • 48
    • 0034017795 scopus 로고    scopus 로고
    • Effect of the sinus of valsalva on the closing motion of bileaflet prosthetic heart valves
    • Ohta Y, Kikuta Y, Shimooka T, Mitamura Y, Yuhta T, Dohi T. Effect of the sinus of valsalva on the closing motion of bileaflet prosthetic heart valves. Artif Organs 2000; 24: 309-12.
    • (2000) Artif Organs , vol.24 , pp. 309-312
    • Ohta, Y.1    Kikuta, Y.2    Shimooka, T.3    Mitamura, Y.4    Yuhta, T.5    Dohi, T.6
  • 50
    • 0031697919 scopus 로고    scopus 로고
    • Medtronic Hall versus St. Jude Medical mechanical aortic valve: Downstream turbulences with respect to rotation in pigs
    • Kleine P, Perthel M, Nygaard H, Hansen SB, Paulsen PK, Riis C, Laas J. Medtronic Hall versus St. Jude Medical mechanical aortic valve: downstream turbulences with respect to rotation in pigs. J Heart Valve Dis 1998; 7: 548-55.
    • (1998) J Heart Valve Dis , vol.7 , pp. 548-555
    • Kleine, P.1    Perthel, M.2    Nygaard, H.3    Hansen, S.B.4    Paulsen, P.K.5    Riis, C.6    Laas, J.7
  • 51
    • 33745459275 scopus 로고    scopus 로고
    • Fluid dynamic assessment of three polymeric heart valves using particle image velocimetry
    • Leo HL, Dasi LP, Carberry J, Simon HA, Yoganathan AP. Fluid dynamic assessment of three polymeric heart valves using particle image velocimetry. Ann Biomed Eng 2006; 34: 936-52.
    • (2006) Ann Biomed Eng , vol.34 , pp. 936-952
    • Leo, H.L.1    Dasi, L.P.2    Carberry, J.3    Simon, H.A.4    Yoganathan, A.P.5
  • 52
    • 34147155562 scopus 로고    scopus 로고
    • Testing requirements for steroscopic Particle Image Velocimetry measurements of mechanical heart valves fluid dynamics
    • Nov 15; Rome, Italy. Rome: Istituto Superiore di Sanità, p
    • Morbiducci U, D'Avenio G, Del Gaudio C, Crigioni M. Testing requirements for steroscopic Particle Image Velocimetry measurements of mechanical heart valves fluid dynamics. Proceedings on III Workshop BioFluMen; 2004 Nov 15; Rome, Italy. Rome: Istituto Superiore di Sanità, p. 21-27.
    • (2004) Proceedings on III Workshop BioFluMen , pp. 21-27
    • Morbiducci, U.1    D'Avenio, G.2    Del Gaudio, C.3    Crigioni, M.4


* 이 정보는 Elsevier사의 SCOPUS DB에서 KISTI가 분석하여 추출한 것입니다.